advertisement
C
HAPTER
E
IGHTEEN
: A
REA
D
ENSITY
You can also Update the results from the Table menu on the Area Density Results window.
S
AVING AND
P
RINTING THE
R
ESULTS AND
D
ATA
Save the Results
From the Area Density Results window > File menu you can select from multiple options if you need to save the analysis data:
Save Outlines: Lets you save the file
Load Outlines
Print the Results
From the Area Density Results window > File menu you can select from multiple options if you need to save the analysis data:
•
Page Setup
•
Print Preview
•
Export the Results
From the Area Density Results window > Data menu you can select from multiple options if you need to save the analysis data:
•
Copy to Clipboard
•
Export to CVS
•
Export to Excel
I
NTENSITY
C
ALIBRATION
C
URVES
Intensity Calibration is the method of creating a mapping (and subsequently a curve) of input intensities to output intensities. Ideally, such a mapping would be linear. However, if you require finely tuned results, you may need to take into account, various errors introduced in the process
164 -
C
HAPTER
E
IGHTEEN
: A
REA
D
ENSITY of lighting and imaging. These errors may change the mapping of original input intensity value to higher or lower than that. E.g. a pixel value of 100 might look slightly brighter, say 120. Or possibly the camera being used has a noise step of 40, which will make all values below 40 look as dark as 0.
One can use Area Density tool in LS Software to carry out analysis based on two different metrics of light intensity:
LS software lets you create curves for both types of calibrations.
Optical Density
Standard Optical Density (OD) is used when the sample of interest is imaged with transmitted light. (i.e. light going thru the sample, into the camera for imaging.) OD value of an area gives an idea of how much light can pass through that area. If the area belongs to a sample in question, it simply how much of sample might be present in that area. Higher OD means less light can get through, suggesting presence of higher quantity of sample.
Following Beer’s Law, the Optical Density of a given pixel P (say at position x,y) is calculated by LS software in the following way:
OD = - Log [ (P(x,y) – Black) / (Incident – Black) ], if P(x,y) < Black
= - Log [ 1/ (Incident - Black) ], otherwise
Where
White = value of brightest white pixel in the imaging environment
Black = value of darkest black pixel in the imaging environment
Total Optical Density of an area is simply the sum total of OD values of all pixels.
Grey Levels
Grey Level calculations are used when the sample is imaged using reflective light. Grey level of an area is simply the sum of grey levels of all pixels in the area.
165 -
C
HAPTER
E
IGHTEEN
: A
REA
D
ENSITY
To Calibrate the Intensity
To add points on the Intensity Calibration Curve, click Add on the Points List
Collection window. Click OK. You can also add points by clicking the Edit List button on
Points List Collection window.
Select the Free Form curve for use on fluorescent/luminescent samples to produce results in Grey Levels.
Select the Std. Optical Density for colorimetric samples imaged using transmitted white light to produce results in Optical Density units.
The scroll-box Number of Samples lets you select the total samples you want displayed on the X-axis of the curve. By default, it is set to the highest value of the dynamic range of the image (8 bit=>256, 16bit =>65536).
To Calibrate Free Form Intensity Curve
Click on the Data Points tab in the Intensity Calibration Curve window. This window lets you manually add or edit each point on the curve (hence the name Free Form). For each intensity volume there is a calibrated volume value that can be assigned. You can also fit a proper curve (in the Graph tab > Curve Model).
166 -
C
HAPTER
E
IGHTEEN
: A
REA
D
ENSITY
Changes are reflected on the curve after the window is closed.
Special Cases of Free Form Curves
Calibration Always Positive
A curve of the following type (left) has a significant section in the negative ‘Y’. If you may not want such a curve, click on Calibration Always Positive checkbox shown above in the Intensity Calibration Curve > Graph Options tab. Result is the curve on the right.
167 -
C
HAPTER
E
IGHTEEN
: A
REA
D
ENSITY
Monotonic Calibration
A curve of the following type (left) can have two values of input (X) for some values of output (Y). In order to avoid that, check the box Monotonic Calibration from the
Intensity Calibration Curve > Graph Options tab to turn the slope of the curve all positive (or negative). Corrected curve looks like the one on the right below.
Saving and Loading calibration
In LS software, Intensity Calibration can be saved as well as loaded back when required. The calibration is stored in a .cal file. It stores information about all calibration curves created.
168 -
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 50 Ftp Transfer
- 51 Darkroom and Lens
- 52 Overview
- 52 Darkroom connectivity
- 54 Selecting a Camera
- 55 OptiChemi
- 60 BioChemi HR/Chemi HR
- 62 Fluor / Chemi Cameras
- 66 Digi Camera
- 70 Scanning Images
- 72 Purpose
- 72 How to Access the Player
- 72 Player Features
- 73 Player Options
- 74 Extract or Delete individual Image files (frames)
- 75 Create A Sequence (.sqv) file by Merging images
- 76 .AVI files
- 77 Loading Images
- 77 Saving Images
- 78 Image Files
- 79 Transferring Images Using FTP
- 81 Overview
- 81 Undo and Redo
- 82 Using Selection Tools
- 84 Paste
- 84 Paste Special
- 87 Overview
- 89 Rotate
- 90 Flip Horizontally
- 90 Flip Vertically
- 90 Resize
- 91 Reduce To Mono
- 91 Convert
- 92 Remove Noise
- 92 Despeckle
- 93 Sharpen
- 93 Emboss
- 94 Starfield Subtraction
- 95 Duplicating Images
- 95 Burn Changes on New Images
- 95 Background Correction
- 95 Background Subtraction
- 96 Overview
- 97 Viewing And Hiding Annotations
- 97 Types Of Annotation
- 98 Spatial Calibration
- 100 Rulers
- 100 Creating Annotations
- 101 The Text Annotation Window
- 102 Selecting Annotations
- 102 Moving And Resizing Annotations
- 103 Formatting Annotations
- 104 Deleting Annotations
- 105 Overview
- 106 Calibrating Image Scale
- 106 Image History
- 107 Understand a Pixel
- 109 Reports
- 110 Print
- 111 Overview
- 111 Navigation
- 114 Finding 1D Gel Lanes and Bands
- 117 Modifying Bands
- 119 Clearing All Lane and Band Information
- 120 Overview
- 120 Retardation factor (Rf) Lines
- 122 Molecular Weight Calibration
- 124 Applying a Standard to a Lane
- 125 Removing a Standard from a Lane
- 126 Overview
- 126 Background Correction
- 128 Concentration Calibration
- 130 Selecting Data Points
- 132 Selecting Curve Type
- 133 Removing Concentration Calibration
- 134 Overview
- 134 Lane Information
- 135 Band Information
- 136 Lane Profile Graph
- 139 Data Explorer
- 143 Fixed Image and Analysis Reports
- 146 Overview
- 146 Automatic Counting
- 147 Manual Colony Counting
- 151 Results Window and Explanation of Statistics
- 157 Purpose
- 157 Area Density Options
- 160 Steps to follow
- 164 Saving and Printing the Results and Data
- 164 Intensity Calibration Curves
- 169 Spatial Calibration
- 170 Purpose
- 170 Features
- 170 Usage
- 172 Secure User Accounts