3.2 Specifi cations. Ohio Medical Push-To-Set Continuous Vacuum Regulator
Below you will find brief information for Vacuum Regulator Push-To-Set Continuous Vacuum Regulator. The Push-To-Set Continuous Vacuum Regulator is a lightweight, compact unit used throughout the hospital primarily for pharyngeal/tracheal suctioning (airway management). Various models provide regulated or full-line vacuum for hospital suction procedures.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
3/Specifi cations
3.2 Specifi cations
Flow Rate:
Positive Pressure Safety
Relief Valve:
0 to 80 LPM without fi ttings at full increase setting depending on supply
Located in the vacuum supply line to prevent pressurization of the patient connection by failed injector vacuum (venturi) units, or inadvertent
Standard Models
High Models
Low Models
0 to 200 mm Hg/0 to 26 kPa 0 to 760 mm Hg/0 to 100 kPa
0 to 760 mm Hg/0 to 100 kPa 0 to 760 mm Hg/0 to 100 kPa
0 to 160 mm Hg/0 to 21 kPa 0 to 135 mm Hg/0 to 18 kPa*
* When measured with an independent measuring device.
Vacuum Relief Valve
(Low Vacuum models ONLY):
140 mm Hg ± 5 mm Hg/18.7 kPa ± 0.7 kPa
11 grams
(Less fi ttings)
Weight
Low Vacuum models:
(Less fi ttings)
12 oz/340 grams
Latex tubing, 0.25 in (6.4 mm) I.D. X
12 in (30 cm) supplied:
Disposable tubing (Available separate in some markets;
6 mm I.D. X 450 mm, 750 mm and 2M) to connect regulator and collection bottle
Disposable Suction Filter:
0 to full vacuum
Flow dependent on source and setup
0 to full vacuum
Flow dependent on source and setup
0 to full vacuum
0 to 100 Lpm @ 650 mmHg (-87 kPa)
40 o
F (4 o
C) to 120 o
F (49 o
C)
0 o
F (-18 o
C) to 150 o
F (71 o
C)
Operating and Storage Relative Humidity: 5 to 95%
8700-0006-000 Rev.3
4
advertisement
Key Features
- Lightweight and compact
- Provides regulated or full-line vacuum
- Includes vacuum gauge
- Features positive pressure safety relief valve
- Low models have vacuum relief valve to limit maximum suction
- Operates in regulated or non-regulated mode
- Automatically adjusts suction level
- Easy to use and adjust
- Safe and reliable
Frequently Answers and Questions
What is the purpose of the Push-To-Set Continuous Vacuum Regulator?
How do I set the suction level on the Push-To-Set Continuous Vacuum Regulator?
What is the difference between the regulated and non-regulated modes on the Push-To-Set Continuous Vacuum Regulator?
How do I clean the Push-To-Set Continuous Vacuum Regulator?
How do I sterilize the Push-To-Set Continuous Vacuum Regulator?
Related manuals
advertisement
Table of contents
- 4 1.1 Defi nitions
- 5 1.2 Warnings
- 5 1.3 Cautions
- 5 2.1 North American Vacuum Regulators
- 5 2.2 International Vacuum Regulators
- 6 3.1 Description
- 7 3.2 Specifi cations
- 8 4.1 Equipment Set-up
- 8 4.2 Attaching the Safety Trap
- 9 4.3 Mode Selection
- 9 4.4 Setting the Suction Level
- 9 4.5 Pre-Use Checkout Procedure
- 11 4.6 Patient Set-up
- 12 5.1 Cleaning
- 12 5.1.1 Routine Exterior Cleaning
- 12 5.1.2 Internal Component Cleaning
- 12 5.1.3 Cold Flush Procedures
- 12 5.1.4 Disinfecting
- 12 5.2 Sterilization
- 14 6/Troubleshooting
- 16 7.1 Service Tools and Equipment
- 16 7.2 Continuous Vacuum Regulators (All Models)
- 16 7.2.1 Disassembly
- 16 7.2.2 Assembly
- 17 7.3 Suction Level Limit Setting - Low Models ONLY
- 17 7.3.1 Low Vacuum Limiting Screws
- 17 7.3.2 Vacuum Relief Valve Adjustment
- 17 7.4 Regulator Module
- 17 7.4.1 Disassembly
- 17 7.4.2 Assembly
- 18 7.5 Digital gauge
- 18 7.5.1 Removal of batteries
- 18 7.5.2 Inserting Batteries
- 19 8.2 Flow Test
- 19 8.3 Gauge Test
- 19 8.3.1 High Vacuum Gauges ONLY
- 19 Gauges ONLY
- 19 8.4 Regulation Test
- 19 8.5 Low Vacuum Regulators ONLY
- 19 8.6 Bleed Test
- 19 8.7 Leak Test - Supply Side
- 19 8.8 Leak Test - Patient Side
- 20 9.1 General Maintenance of Suction Equipment
- 20 9.2 Recommended Maintenance Schedule
- 20 9.2.1 Maintenance Schedule
- 21 9.3 Repair Policy
- 21 9.4 Technical Assistance
- 21 9.5 Return Instructions
- 22 9.6 Warranty
- 22 and Fittings
- 23 10.1 Illustrated Parts
- 23 10.2 Service Kits
- 23 10.3 Fittings and Adapters
- 23 10.4 North American
- 23 10.4.1 Regulator Options
- 23 10.4.2 Other Options
- 23 10.5 International
- 23 10.5.1 Regulator Options
- 23 10.5.2 Other Options