Unit description. Sirona GALILEOS COMPACT, GALILEOS ComfortPLUS, GALILEOS COMFORT
GALILEOS is a 3D CBCT imaging system used in dentistry. It provides high-resolution images of the teeth, jaw, and surrounding structures. This allows dentists to diagnose and treat problems more effectively, and to plan complex procedures with greater accuracy.
Advertisement
Advertisement
3 Unit description
3.1 Unit classes and versions
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3
Unit description
Unit description
3.1
Unit classes and versions
Unit classes and versions
A
C
A
B
C
B
GALILEOS unit*
Control panel
FaceScan (optional)
16
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3 Unit description
3.2 Hardware
3.1.1
Control panels
The "Comfort/Comfort
PLUS
" and "Compact" unit classes differ by the equipment of the control panel (B). While the GALILEOS Comfort class features a control panel with a color touchscreen (Easypad), the Compact class is equipped with a simpler control panel with a single-line display
(Multipad). Due to their different control panels, the operating procedures for these two system classes also vary.
Accompanying documents
3.2
Hardware
Hardware
3.2.1
Information on the unit
Information on the unit
The following symbols are applied to the unit:
Accompanying documents
This symbol is affixed next to the unit rating plate.
Meaning: When operating the unit, observe the operating instructions.
This symbol is affixed on the unit rating plate.
Meaning: The accompanying documents are available on the homepage of Sirona.
Electrostatic discharge (ESD)
Connector pins or sockets bearing ESD warning labels must not be touched or interconnected without ESD protective measures. See also
"Electrostatic Discharge" and "Electromagnetic Compatibility" [ → 14].
64 26 345 D3437
D3437.076.03.02.02 04.2013
17
3 Unit description
3.2 Hardware
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Single use hygienic protective sleeves
Identification of single use devices
Prior to each exposure, the hygienic protective sleeves (single use devices) must be fitted.
Single use devices are identified with the symbol shown on the left. They must be disposed of immediately after use. Do not use single use devices more than once.
3.2.2
Installation versions
Installation versions
A B
C
18
The unit can be equipped with...
● a 1-3 m coiled cable with release button inside the treatment room (A) or ...
● a remote control with or without coiled cable (B+C) located outside the X-ray room (see also installation instructions).
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3.2.3
Modules and components
3 Unit description
3.2 Hardware
64 26 345 D3437
D3437.076.03.02.02 04.2013
C
C
D
A
B
Slide
Stand
D A
B
19
3 Unit description
3.2 Hardware
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
LS
3.2.3.1
Slide
Slide
MU
A
DX6*
DX7*
LS*
B
DX11
DX1
DX89
GALILEOS Comfort / Comfort PLUS
DX71*
GALILEOS Compact
Component
Boards
Motor
Designation
DX1
DX11
DX6*
DX6*
DX7*
DX71*
DX89
MU
Light barriers LS
LS
(A)
(B)
Function
Open loop/closed loop control in general
Controller board
Open loop/closed loop tube assembly (GALILEOS Comfort/
Compact)
Open loop/closed loop tube assembly (GALILEOS
Comfort
PLUS
)
Easypad touchscreen
(GALILEOS Comfort)
LED display on Multipad
(GALILEOS Compact)
Image memory of the X-ray detector
Rotary movement of rotating element
Position control of the ring cycle
Position control of the swivel arm
*) not available as individual repair part (see spare parts list).
20
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3.2.3.2
Stand
MHV
DX32
3 Unit description
3.2 Hardware
DX42
Component
Boards
Motor
Designation
DX32
MHV
Function
Power supply board
Linear movement of height adjustment
3.2.3.3
Remote control
Remote control
Component
Board
Designation
DX42
Function
Display board for remote control
64 26 345 D3437
D3437.076.03.02.02 04.2013
21
3 Unit description
3.2 Hardware
3.2.3.4
FaceScan
FaceScan
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
22
Component
Boards
Designation
FACESCAN
PoE
Function
Modular board
Power supply board
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3.2.4
Cabling overview
Cabling overview
L1
L2
L3
DX6
H1
L3
A
L5
L4
L1
L5
C
A
B
C
Power switch
Line filter
Wago terminal
L4
DX32
L2
B
3 Unit description
3.2 Hardware
DX11
DX1
DX89
64 26 345 D3437
D3437.076.03.02.02 04.2013
23
3 Unit description
3.2 Hardware
5922526
X1/J307
5922526 DX1
L6
L9
L10
L15
L9
DX6
L10
H1
A
DX7/DX71
L6
L15
A Power switch
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
DX11
DX1 DX89
24
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Cabling overview2: GALILEOS, new
L13
L16
L19
L27
L28*
L29
LS
DX6
L16
HV
L19
L29
H1
A
DX7/DX71
D
* Cable L28 cannot be replaced.
A Power switch
3 Unit description
3.2 Hardware
DX11
DX1
DX89
L13 L27 L28*
64 26 345 D3437
D3437.076.03.02.02 04.2013
25
3 Unit description
3.2 Hardware
L7
L108
L117
L25/L26
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
DX1
DX42
L108
L117
L7
L25
E
E Media converter
26
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
FaceScan cabling overview
FaceScan cabling
L77
L78
3 Unit description
3.2 Hardware
A
L77 L78
DX11
DX1
DX89
FACESCAN
L77
PoE
A Power switch
64 26 345 D3437
D3437.076.03.02.02 04.2013
27
3 Unit description
3.2 Hardware
Boards DX1/DX11
3.2.5
Board photos
3.2.5.1
Board photos
Boards in the slide
Boards in the slide
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Board DX6
28
DX6: GALILEOS, new
This board is not available as a spare part or a repair part. X-ray tube assemblies can only be ordered as complete units.
GALILEOS Compact and GALILEOS Comfort:
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3 Unit description
3.2 Hardware
64 26 345 D3437
D3437.076.03.02.02 04.2013
GALILEOS Comfort PLUS:
29
3 Unit description
3.2 Hardware
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Board DX7
DX7: Restriction for GALILEOS, new
This board is only used in the "GALILEOS Comfort" and the "GALILEOS
Comfort
PLUS
" (not in the "GALILEOS Compact").
The board is not available as a spare part or a repair part. The Easypad can only be ordered as a complete unit.
30
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3 Unit description
3.2 Hardware
Board DX71
DX71 GALILEOS / XG 3 / XG 5
DX71: Restriction for GALILEOS, new
This board is only used in the "GALILEOS Compact" (not in the
"GALILEOS Comfort" or the "GALILEOS Comfort
PLUS
").
64 26 345 D3437
D3437.076.03.02.02 04.2013
31
3 Unit description
3.2 Hardware
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
32
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Board DX89
DX89: GALILEOS
3 Unit description
3.2 Hardware
Board DX32
3.2.5.2
Boards in the stand
Boards in the stand
64 26 345 D3437
D3437.076.03.02.02 04.2013
A Line filter
33
3 Unit description
3.2 Hardware
Board DX42
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3.2.5.3
Board in the remote control
Board in the remote control
This board is not available as a spare part or a repair part.
FACESCAN modular board
3.2.5.4
FaceScan boards
FaceScan boards
This modular board is not available as a spare part or a repair part.
FaceScan can only be ordered as a complete unit.
34
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3 Unit description
3.2 Hardware
64 26 345 D3437
D3437.076.03.02.02 04.2013
35
3 Unit description
3.2 Hardware
PoE power supply board
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
GALILEOS
3.2.6
Covers
Covers
When removing covers, always remember that direct sunlight or bright room lighting can cause system malfunctions due to activated light barriers. Therefore: avoid direct sunlight and bright room lighting above the unit!
Reattach all covers. When attaching the covers: be sure to screw the sheet metal cover back on.
IMPORTANT: For reasons of electromagnetic compatibility, be sure to fasten all screws.
Covers: GALILEOS, new
IMPORTANT
The covers of your unit may differ slightly from those shown in the following illustration, depending on the unit class/unit serial number.
36
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3 Unit description
3.2 Hardware
64 26 345 D3437
D3437.076.03.02.02 04.2013
C
P
E
A
B
Profile covers, top and bottom
Intermediate piece
Tube assembly cover, rear
Tube assembly cover
Cover for ring center (in units without head fixation device)
37
3 Unit description
3.2 Hardware
Chassis:
X-ray tube assembly:
Detector:
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
F
A
N
O
I
J
G
H
K
L
Acquisition unit for head fixation device (in units with head fixation device)
Ring cover
Support cover
Swivel arm cover
Arm cover
Slide cover, front
Slide cover, top rear
Slide cover, center rear
Slide cover, bottom rear
X-ray detector cover
3.2.7
Technical data
3.2.7.1
Technical data
GALILEOS Compact/GALILEOS Comfort
GALILEOS Compact/GALILEOS Comfort
Model designation
Nominal voltage:
Permissible fluctuation:
Permissible drop under load:
Rated current:
Nominal power output:
Current time product:
Nominal frequency:
Internal line impedance:
Main building fuse:
Power consumption:
GALILEOS Compact/Comfort
200 V – 240 V
±10%
10%
6 A
0.6 kW at 85 kV/7mA
42 mAs
50 Hz / 60 Hz max. 0.8 ohms
25 A slow-blow (16 A for single line)
0.9 kVA
Focal spot size acc. to IEC 60336, measured in the central X-ray beam: 0.5
kV: 85 kV mA:
Pulsed mode:
5 mA / 7 mA
10 ms – 30 ms
Total filtration of X-ray tube assembly > 2.5 Al / 90 IEC 60522
Integrated filter: 0.4 mm Cu
Cone-beam angle:
High voltage generation frequency: collimated to approx. 24°
80 kHz – 100 kHz
Type:
Active input window size:
Camera:
Image intensifier (I.I.),
Thales or Siemens
215 mm (8 1/2") diameter
Pixels: 1000
2
38
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Geometry:
Scanning process:
3 Unit description
3.2 Hardware
FPS: 15 – 30
Dynamics: 12 bits,
(4096 brightness values), 60 dB
Source-I.I. converter coating distance
(central X-ray beam)
Source-isocenter distance
(central X-ray beam)
Source-skin distance
(minimum distance)
510 mm (20 1/16")
333 mm (13 1/8") approx. 220 mm (8 5/8")
Orbital angle
Scan time
Number of single exposures
204° approx. 14 s
200
Reconstruction:
Marking of focal spot:
64 26 345 D3437
D3437.076.03.02.02 04.2013
Automatic exposure blocking:
Class I device
Degree of protection against electric shock:
Degree of protection against ingress of water:
Type B device
Ordinary equipment
(without protection against ingress of water)
Year of manufacture:
The duration of automatic exposure blocking (cooling period) depends on the set kV/mA level and the actual exposure time. Depending on the tube load, interval times of 8 s to 300 s are automatically set by the system.
Operating mode:
Long-term power output:
Anode material:
Exposure parameters for determining leakage radiation:
Continuing current for leakage radiation measurements:
Transport and storage temperature:
Basic unit
(on the rating plate)
Continuous operation
100 W
Tungsten
7mA / 85 kV
0.14 mA
-40°C – +70°C (-40°F – 158°F)
39
3 Unit description
3.2 Hardware
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Detector
Air humidity:
Admissible operating temperature:
Operating altitude:
-30°C – +55°C (-22°F – 131°F)
10% – 95% without condensation from +10°C to +35°C (50°F – 95°F)
≤ 3000 m
X-ray tube: Toshiba DF-151R or
Siemens SR 120/15/60
Minimum requirements for reconstruction PC (included in the scope of supply):
Processor:
RAM:
Hard disks:
Operating system:
External drive:
DualCore from 2 GHz
Minimum requirements for
SIDEXIS visualization PC
(not included in the scope of supply):
See SIDEXIS XG Operator’s Manual.
The system requirements are also listed under www.sidexis.com
Network:
Chassis:
Network:
Communication interface:
100 MB Ethernet, 1 Gbit Ethernet recommended
RJ45 for LAN cable
3.2.7.2
GALILEOS Comfort PLUS
GALILEOS Comfort PLUS
Model designation
Nominal voltage:
Permissible fluctuation:
Permissible drop under load:
Rated current:
Nominal power output:
Current time product:
Nominal frequency:
Internal line impedance:
Main building fuse:
Power consumption:
GALILEOS Comfort
PLUS
200 V – 240 V
±10%
10%
6 A
0.6 kW at 98 kV/6mA
30 mAs
50 Hz / 60 Hz max. 0.8 ohms
25 A slow-blow (16 A for single line)
0.9 kVA
X-ray tube assembly:
4 GB RAM
> 500 GB
Windows XP Professional SP3 or
Windows 7 Professional
1x DVD-ROM, dual-layer
Focal spot size acc. to IEC 60336, measured in the central X-ray beam: 0.5
kV: mA:
Pulsed mode:
98 kV
3 mA - 6 mA
10 ms – 14 ms
40
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Detector:
Geometry:
Scanning process:
3 Unit description
3.2 Hardware min. permanent filtering: min. half-value layer:
Integrated filter:
Cone-beam angle:
High voltage generation frequency:
2.8 mm Al / 98 IEC 60522
3.9 mm Al
0.2 mm Cu collimated to approx. 24°
80 kHz – 100 kHz
Type:
Active input window size:
Camera:
Image intensifier (I.I.), Siemens
215 mm (8 1/2") diameter
Pixels: 1000
2
FPS: 15 – 30
Dynamics: 12 bits,
(4096 brightness values), 63 dB
Source-I.I. converter coating distance
(central X-ray beam)
Source-isocenter distance
(central X-ray beam)
Source-skin distance
(minimum distance)
590 mm (23 1/4")
413 mm (16 1/4") approx. 300 mm (11 13/16")
Orbital angle
Scan time
Number of single exposures
204° approx. 14 s
200
HD mode 357
Reconstruction:
Marking of focal spot:
64 26 345 D3437
D3437.076.03.02.02 04.2013
Automatic exposure blocking: The duration of automatic exposure blocking (cooling period) depends on the set kV/mA level and the actual exposure time. Depending on the tube load, interval times of 8 s to 300 s are automatically set by the system.
Class I device
Degree of protection against electric shock:
Degree of protection against ingress of water:
Type B device
Ordinary equipment
(without protection against ingress of water)
41
3 Unit description
3.2 Hardware
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Year of manufacture:
Operating mode:
Long-term power output:
Anode material:
Exposure parameters for determining leakage radiation:
Continuing current for leakage radiation measurements:
Transport and storage temperature:
Basic unit
Detector
Air humidity:
Admissible operating temperature:
Operating altitude:
(on the rating plate)
Continuous operation
100 W
Tungsten
6mA / 98 kV
0.14 mA
-40°C – +70°C (-40°F – 158°F)
-30°C – +55°C (-22°F – 131°F)
10% – 95% without condensation between +10°C and +35°C
(50°F – 95°F)
At a room temperature of > 35°C (> 95°F),
Sirona recommends using an air conditioning system.
≤ 3000 m
X-ray tube: Siemens SR 120/15/60
Minimum requirements for reconstruction PC (included in the scope of supply):
Processor:
RAM:
Hard disks:
Operating system:
External drive:
Graphics card:
QuadCore Intel i5
8 GB RAM
1 TByte
Windows 7 64-bit (mandatory)
1x DVD-ROM, dual-layer explicit (not onboard), at least 1 GB RAM
Minimum requirements for
SIDEXIS visualization PC
(not included in the scope of supply):
See SIDEXIS XG Operator’s Manual.
The system requirements are also listed under www.sidexis.com
Network: Network:
Communication interface:
100 MB Ethernet, 1 Gbit Ethernet recommended
RJ45 for LAN cable
42
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3 Unit description
3.3 Software/compatibility
Main software V04.12.00
SIDEXIS XG GALILEOS software
V2.5.6 or higher
V2.0
Main software V04.14.00
3.3
Software/compatibility
Software/compatibility
3.3.1
GALILEOS firmware
GALILEOS firmware
Any software combinations other than those listed here are not allowed.
If a module software version does not match the main software version, the main software version is identified with an asterisk on the info screen
(e.g. 04.12.00*).
GALILEOS
Board
DX6
DX6NG
DX7
DX7-L0
DX7-L1
DX7-L2
DX7-L3
DX7-L4
DX7-L5
DX71
DX11
DX11-FPGA
DX89
DX89 FPGA
Software
03.07.04
04.09.00
02.80.07
02.29.00
02.29.00
02.27.00
02.27.00
02.04.00
02.02.00
02.54.03
04.12.00
01.04.00
01.59.02
01.58.00
Remote control
Board
DX42
Software
02.56.04
RCU server software
V2.3
GALILEOS
Implant
V1.9SP1
FaceScan unit software
V1.15.369–
V1.14369
FaceScan
PC software
V1.2
GALILEOS
Board
DX6
DX6NG
DX7
DX7-L0
DX7-L1
DX7-L2
DX7-L3
DX7-L4
DX7-L5
DX71
DX11
Software
03.08.00
04.09.07
02.82.01
02.31.00
02.31.00
02.29.00
02.29.00
02.06.00
02.02.00
02.55.02
04.14.00
FaceScan
USB stick
FS0004
Remote control
Board
DX42
Software
02.58.02
64 26 345 D3437
D3437.076.03.02.02 04.2013
43
3 Unit description
3.3 Software/compatibility
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
GALILEOS
Board
DX11-FPGA
DX89
DX89 FPGA
Software
01.04.00
01.60.02
01.58.00
Remote control
Board Software
SIDEXIS XG GALILEOS software
V2.5.6 or higher
V2.1
FaceScan
FS 0004 or higher
RCU server software
V2.4
GALILEOS
Implant
V1.9 SP1
FaceScan unit software
V1.15.369 –
V1.14369
3.3.2
FaceScan firmware
GALILEOS main unit software
V04.12.00 or higher
FaceScan firmware
GALILEOS Software
V2.0 or higher
FaceScan
PC software
V1.2
FaceScan
USB stick
FS0004
SIDEXIS XG
V2.5.6 or higher
3.3.3
GALILEOS Software
GALILEOS
Software
V2.0
CD index Remarks
006 Requires main unit software V 04.12.00 or higher and SIDEXIS 2.5.6 or higher.
44
64 26 345 D3437
D3437.076.03.02.02 04.2013
Advertisement
Key Features
- 3D CBCT imaging
- High-resolution images
- Accurate diagnosis and treatment planning
- Improved patient care