8 Perform service routines via the control panel
8.1 Overview of service routines
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8
Perform service routines via the control panel
Perform service routines via the control panel
8.1
Overview of service routines
Overview of service routines
8.1.1
List of all service routines available for selection
List of all service routines available for selection
Function
Radiation without rotary movement, selectable kV/mA level and maximum radiation time
Long-term exposure with fixed radiation intervals from any position
General X-ray tube assembly service
(for internal Sirona purposes only)
(for internal Sirona purposes only)
Fan test
Temperature sensor test, single tank
Automatic adjustment of pulse preheating
Error logging memory
Display error logging memory
Clearing error logging memory
Enable the CAN bus logging in the Miniweb
Update service
Overview of the module software versions
Confirming the unit serial number
Flash file system
Formatting flash file system
Test flash file system
Save/restore DX89 data
Dosimetry (without ring movement)
Current measurement (unpulsed)
Dosimetry with pulsed radiation
CAN bus service
Presence display of modules
Inquiry of the CAN status register of the modules
Resetting the CAN status register of the modules
Display of CAN bus cycle on the LEDs of the modules
Configuration service
Configuring the hardware version
Enter the country group code
Select a language see
S.
S.
S.
174
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.1 Overview of service routines
S017.15
S017.25
S018
S018.2
S018.3
S018.4
S037
S037.1
S037.2
Service routine
S017.5
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S017.6
S017.7
S017.9
S017.13
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S017.14
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S037.3
S037.4
Function
Select a language set
Activate the remote control display
Configure the switching plate for the swivel arm
(for internal Sirona purposes only)
Enable/disable the welcome screen
Enable/disable certain lines of the welcome screen
Activate/deactivate the acoustic signal for end of exposure
Select the diaphragm type
Service for height adjustment
Set the maximum travel height
Undo the maximum travel height setting
Check the height adjustment sensor system
Network service
Displaying the network data
Setting the default IP address, default gateway address and default subnet mask
Configuring boot mode: DYNAMIC (DHCP/AutoIP) / STATIC
(fixed address)
Manual input of static network settings (IP address, default gateway address, and subnet mask) see
S.
S.
S.
S.
S.
S.
8.1.2
Alphabetical list of service routine functions
Alphabetical list of service routine functions
Function
Acoustic signal for end of exposure, activate/deactivate
Automatic adjustment of pulse preheating
Modules, presence display
Module software version, show overview
Welcome screen, enable/disable
Welcome screen, enable/disable lines
Select the diaphragm type
Configuring boot mode: DYNAMIC (DHCP/AutoIP) / STATIC
(fixed address)
CAN bus logging in the web interface, enable
CAN bus cycle, display on the LEDs of the modules
Service routine
S017.15
S005.8
S012.1
S008.2
S017.13
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S017.14
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S017.25
S037.3
S007.5
S012.4
see
S.
S.
S.
S.
64 26 345 D3437
D3437.076.03.02.02 04.2013
175
8 Perform service routines via the control panel
8.1 Overview of service routines
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Function
CAN bus, service
CAN status register for the modules, query
CAN status register for the modules, reset
Long-term exposure with fixed radiation intervals from any position
Default gateway address, set to defaults
Default gateway address, manual input
Dosimetry, with pulsed radiation
Dosimetry (without ring movement)
Save/restore DX89 data
Error logging memory
Error logging memory, display
Error logging memory, clear
Remote control, enable/disable display
Flash file system
Flash file system, format
Flash file system, test
Unit serial number, confirm
Unit variant, configure
Country group code, enter
Height adjustment, sensor system test
Height adjustment, service
IP address, set to defaults
IP address, enter manually
Configuration, service
Fan, test
Network, service
Displaying the network data
Configure the switching plate for the swivel arm
Select a language set
Select a language
Service routine
S012
S012.2
S012.3
S002.5
Radiation without rotary movement, selectable kV/mA level and maximum radiation time
Tube assembly service, general
Current measurement, unpulsed
Subnet mask, set to defaults
Subnet mask, manual input
Temperature sensor test, single tank
Update, service
S005
S011.9
S037.2
S037.4
S005.5
S008
S009.4
S009.5
S008.3
S017.2
S017.3
S018.4
S018
S037.2
S037.2
S037.4
S011.12
S011
S009.7
S007
S007.1
S007.2
S017.6
S009
S037.4
S017
S005.4
S037
S037.1
S017.7
S017.5
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S017.4
("GALILEOS Comfort"/"
GALILEOS Comfort
PLUS
")
S002
176 see
S.
S.
S.
S.
S.
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.2 Service menu and service routines
Function
Maximum travel height, set
Maximum travel height, undo setting
Service routine
S018.2
S018.3
see
8.2
Service menu and service routines
Service menu and service routines
You can use the service routines to check the function of certain unit components and modules, as well as to set important unit parameters.
8.2.1
Displays and symbols in the service menu
8.2.1.1
Displays and symbols in the service menu
Easypad
Easypad
There are many different control symbols and display fields on the touchscreen; these are activated on a context-sensitive basis depending on the procedure step.
64 26 345 D3437
D3437.076.03.02.02 04.2013
A
B
C
D
E
X RAY
X RAY Active!
Selection field 1
Selection field 2
Selection field 3
Arrow keys
F
G
H
S1 - S37
1 - n
Patient symbol keys
Radiation can be released.
Caution! Radiation is being released.
Display fields for service routines, test steps, values, unit parameters, etc.
Touch the "+" and "-" arrow keys to
select unit parameters in the selection fields [ → 187].
Selected service routine.
Selected test step.
Different functions, depending on service routine.
177
8 Perform service routines via the control panel
8.2 Service menu and service routines
I
J
Memory key
Service key
K
L
T(est rotation) key
R(eturn) key
M Double arrow key
N Wrench symbol
8.2.1.2
Multipad
Multipad
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Save selection.
Different functions, depending on service routine. Most, however, confirm a selection or the activation of the next test step.
Start a test.
Move the unit to the starting position or confirm a save operation.
Return to the main menu.
Displayed if level 4 (service menu) is activated.
A
178
K L H J I
E
B C D
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.2 Service menu and service routines
K
L
A
B
C
D
E
X-ray lamp
Selection field 1
(8 digits)
Selection field 2
(4 digits)
Selection field 3
(4 digits)
Arrow keys
H
I
J
Patient symbol keys
Memory key
Service key
T(est rotation) key
R(eturn) key
Caution! Radiation is being released.
Display fields for service routines, test steps, values, unit parameters, etc.
Touch the "+" and "-" arrow keys to select unit parameters in the selection
fields (Select parameters [ → 187]).
Different functions, depending on service routine.
Save selection.
Different functions, depending on service routine. Most, however, confirm a selection or the activation of the next test step.
Start a test.
Move the unit to the starting position or confirm a save operation.
64 26 345 D3437
D3437.076.03.02.02 04.2013
179
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8.3
Basic operating procedures in the service menu
Basic operating procedures in the service menu
8.3.1
Activating the service menu
8.3.1.1
Activating the service menu
Easypad
Easypad
The structure of the touchscreen user interface on the control panel is subdivided into 4 levels:
● Level 1: Main menu
● Level 2: Program Settings
● Level 3: Basic Settings
● Level 4: Service menu
When the unit starts up, the main menu appears on the touchscreen.
1. To select level 2 (
"Select program settings"
the top right corner of the touchscreen (A).
), touch the blue arrow in
2. To select level 3 (
"Select basic settings"
), touch the left-hand blue arrow in the top right corner of the touchscreen (B).
180
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
3. To select level 4 (service menu/access), touch the wrench symbol
(C).
4. Switch to the service menu:
Press and hold down the Service key (D) until the patient symbol keys light up (E-H) (approx. 2 s). Then press the patient symbol keys in the sequence F – H – E within the next 4 s.
After you have entered the key combination correctly, the service menu is displayed. You can return to the next higher level with the double arrow key (I) at any time.
64 26 345 D3437
D3437.076.03.02.02 04.2013
181
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8.3.1.2
Multipad
1. Press and hold down Service key (D) until the patient symbol keys light up (E-H) (approx. 2 s).
D
2. Then press the patient symbol keys in the sequence F – H – E within the next 4 s.
E F G H
After you have entered the key combination correctly, the service menu is displayed. You can return to the next higher level with the arrow key above selection field 3 (I) at any time.
8.3.2
Selecting service routines and test steps
Selecting service routines and test steps
8.3.2.1
Selecting a service routine
Selecting a service routine
✔
The service menu must be selected [ → 180].
182
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
A A
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
➢ Select the desired service routine via the arrow keys in the selection field 1 (A) and confirm the selection via the service key (B).
If the selected service routine has several test steps, the first selectable test step is displayed in selection field 2 (test step 1 in the example).
-
-
S005
1
+
+
S5
A
B
S005_______1____
A
B
8.3.2.2
Selecting a test step
✔
The service menu must be selected [ → 180].
✔
The required service routine must be selected [ → 182].
64 26 345 D3437
D3437.076.03.02.02 04.2013
183
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
➢ Select the required test step in selection field 2 with arrow keys (C) and confirm your selection by pressing Service key (B).
-
01
+
S5
4
Easypad:
The selected service routine as well as the test step chosen are displayed in the right-hand column (S005.4 in the example).
184
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Multipad:
The parameters or IDs of the selected service routine are displayed on the Multipad. The Multipad does not show which service routine or test step is currently active.
8.3.2.3
Service routines with security access
Service routines with security access
A security code is required for accessing service routines involving functions such as radiation release or editing of configuration data or stored values. This procedure prevents the inadvertent selection or activation of these service routines.
To select a service routine or test step with security access, proceed as follows:
After you have confirmed your selection, a "0" appears in selection field 2.
64 26 345 D3437
D3437.076.03.02.02 04.2013
185
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. Confirm security access by once again selecting the number of the main routine (2 in the example) with the arrow keys in selection field
2 (C) and press the Service key (B) to confirm your selection.
186
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Following this double selection and confirmation via the Service key, the service routine is activated.
Example
8.3.3
Select parameters
Select parameters
Easypad touchscreen
Select parameter XG 3D / XG 3D ready / GALILEOS
If arrow keys are displayed in the selection fields once the required service routine has been selected, you can use these arrow keys to choose between different parameters.
You want to run service routine S017.6 to activate the remote control.
✔
Once you have selected service routine S017.6, the code "00" is preselected for the "Remote control disabled" option.
64 26 345 D3437
D3437.076.03.02.02 04.2013
187
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
➢ Touch the + or - arrow keys (A) to select the code 01 (B) for the
"Remote control enabled" option.
Once the selected parameter has been changed (in this case the code for the activation of the remote control), the Memory key (C) lights up.
Title: Multipad GALILEOS
Example
Multipad
Select parameter GALILEOS, XG 3 / 5
You want to run service routine S017.6 to activate the remote control.
✔
Once you have selected service routine S017.6, the code "00" is preselected for the "Remote control disabled" option.
➢ Press the UP or DOWN arrow key (A) to select the code 01 (B) for the
"Remote control enabled" option.
Once the selected parameter has been changed (in this case the code for the activation of the remote control), the LED above
Memory key (C) lights up.
188
8.3.4
Saving parameters
Saving parameters
Once one or a number of parameters have been selected via a service routine, the current selection must be saved so that it is applied in the unit.
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Example
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Title: Saving parameters in GALILEOS
Easypad touchscreen
Save Easypad parameter: GALILEOS / XG 3D / XG 3D ready
You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔
The Memory key (C) lights up.
1. Touch the Memory key (C).
The R key (D) lights up.
2. Touch the R key (D).
The selected setting is saved to non-volatile memory.
Title: Save parameter GALILEOS
Example
Multipad
Save parameter GALILEOS / XG 3 / 5
You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔
The LED above Memory key (C) lights up.
1. Press Memory key (C).
The LED above the R key lights up.
2. Press R key (D).
The selected setting is saved to non-volatile memory.
64 26 345 D3437
D3437.076.03.02.02 04.2013
189
8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8.3.5
Exiting the test step and service routine
Exiting the test step and service routine
Easypad
Touch the Service key (A) or the double arrow key (B) to go back to the menu for selecting service routines.
Touch the double arrow key (B) in the service menu to go back to the main menu.
Exception: Service routine S017
In service menu S017, touch the Service key (A) to go to the next test step in the service routine.
Title: Exit service routine test steps GALILEOS 1
Multipad
Press the Service key (A) or the up arrow key above selection field 3 (B) to go back to the menu for selecting service routines.
In the service menu, touch the up arrow key above selection field 3 (B) to go back to the main menu.
Exception: Service routine S017
In service menu S017, press the Service key (A) to go to the next test step in the service routine.
190
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH 8 Perform service routines via the control panel
Service Manual (as of February 2013) GALILEOS 8.4 S002: Radiation without rotary movement, selectable kV/mA level and maximum
SR*
S002
S002.5
SHZ**
Yes
8.4
S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
Function
X-ray beam test
Long-term exposure with fixed radiation intervals from any position
* SR=service routine, ** SHZ=security access
8.4.1
S002: Test step 5
S002: Test step 5
Long-term exposure with fixed radiation intervals from any position
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away from the unit as the coiled cable for the release button allows you to.
Selection field
1
Parameters kV/mA level
Range of values
85kV/5mA
85 kV / 7 mA*
0.1 s – 4 s* 2 Radiation time
* Factory setting
1. Call service routine S002.5 [ → 182].
64 26 345 D3437
D3437.076.03.02.02 04.2013
191
8 Perform service routines via the control panel
8.5 S005: General X-ray tube assembly service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. Use the arrow keys (A) to select the required kV/mA level and the required radiation time (see table).
SR*
S005
S005.4
S005.5
S005.8
192
SA**
No
No
Yes
3. Initiate the radiation.
The maximum set radiation time has elapsed.
IMPORTANT: If you let go of the release button before the maximum radiation time has elapsed, radiation is terminated prematurely and the exposure is interrupted. The actual radiation time is
not
displayed.
When you release radiation during the cool-down interval, a countdown of the remaining waiting time is displayed in the header of the control panel or in selection field 2 on the Multipad
(automatic exposure blocking).
4. Exit the service routine [ → 190].
8.5
S005: General X-ray tube assembly service
S005: General X-ray tube assembly service
Function
General X-ray tube assembly service
Fan test
Temperature sensor test, single tank
Automatic adjustment of pulse preheating
* SR=service routine, ** SA=security access
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.5 S005: General X-ray tube assembly service
8.5.1
S005: Test step 4
S005: Test step 4
Fan test
Selection field
1
Code
00
01
Function
Fan off*
Fan on
* Factory setting
1. Call service routine S005.4 [ → 182].
2. Use the arrow keys (A) in selection field 1 to select the code "01" (see table).
3. Confirm your selection by pressing the R key (B).
The fan starts up.
4. Check the fan for running noise.
5. Exit the service routine [ → 190].
Upon exiting the service routine, the fan is automatically switched off again.
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.5.2
S005: Test step 5
S005: Test step 5
Temperature sensor test, single tank
1. Call service routine S005.5 [ → 182].
193
8 Perform service routines via the control panel
8.5 S005: General X-ray tube assembly service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
After the service routine has been selected, selection field 1 displays the single tank temperature in °C. The display is updated once per second.
2. Exit the service routine [ → 190].
8.5.3
S005: Test step 8
S005: Test step 8
IMPORTANT
After performing this service routine, the unit must be completely recalibrated.
Automatic adjustment of pulse preheating
1. Call service routine S005.8 [ → 182].
194
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.5 S005: General X-ray tube assembly service
An inactive progress indicator in selection field 1 and the message "FFFF" in selection field 2 signal that the system is ready for compensation.
2. Start the automatic adjustment by pressing and holding the release button.
IMPORTANT: Keep pressing the release button until adjustment is completed and the new offset value for preheating is displayed. If you interrupt the adjustment procedure prematurely by letting go of the release button, the message "EEEE" appears in selection field 2. This message must be acknowledged by pressing the R key.
When pressing the release button, radiation is released for 2 s to warm up the tube assembly to operating temperature. This is followed by the automatic tuning routine.
64 26 345 D3437
D3437.076.03.02.02 04.2013
195
8 Perform service routines via the control panel
8.6 S007: Error logging memory
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
A progress indicator is displayed during the service function.
3. After the adjustment has been performed, exit the service routine [ → 190].
SR*
S007
S007.1
S007.2
S007.5
196
SHZ**
No
Yes
No
8.6
S007: Error logging memory
S007: Error logging memory
Function
Error logging memory
Display error logging memory
Clearing error logging memory
Enabling CAN bus logging in the web interface
* SR=service routine, ** SHZ=security access
8.6.1
S007: Test step 1
S007: Test step 1
Display error logging memory
In addition to service routine S007.1, you can also use the extended detail query in SiXABCon to check the error logging memory.
Title only for GALILEOS
Easypad touchscreen
Symbol on the control panel
Status Function
Patient symbol key 1 (B) is selected Step width for scrolling between error events = 1*
Patient symbol key 2 (C) is selected Step width for scrolling between error events = 10
Patient symbol key 3 (D) is selected Step width for scrolling between error events = 100
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.6 S007: Error logging memory
64 26 345 D3437
D3437.076.03.02.02 04.2013
Multipad
1. Call service routine S007.1 [ → 182].
2. Use the arrow keys (A) in selection field 1 to select the required error event (66 in the example).
Use the patient symbol keys to set the increment for scrolling between the error numbers:
Patient symbol key 1 (left) = increment 1 (factory setting)
Patient symbol key 2 = increment 10
Patient symbol key 3 = increment 100
The LED above the selected patient symbol key is lit up.
197
8 Perform service routines via the control panel
8.6 S007: Error logging memory
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2.
4. Exit the service routine [ → 190].
198
8.6.2
S007: Test step 2
S007: Test step 2
Clearing error logging memory
1. Call service routine S007.2 [ → 182].
The system's readiness to clear the memory is indicated by the display message "FFFF" in selection field 1. If the error logging memory does not contain any data, "0000" is displayed.
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.6 S007: Error logging memory
2. To clear the memory, press the Memory key (A) (R key (B) (Easypad) or LED above the R key (B) (Multipad) lights up) followed by the R key
(B).
64 26 345 D3437
D3437.076.03.02.02 04.2013
199
8 Perform service routines via the control panel
8.6 S007: Error logging memory
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Once the memory has been cleared, the message "0000" is displayed in selection field 1.
3. Exit the service routine [ → 190].
200
8.6.3
S007: Test step 5
S007: Test step 5
Enabling CAN bus logging in the web interface
NOTICE
This service routine may only be called up subject to the approval of and with the support of the Sirona Customer Service Center (CSC).
Selection field Parameters
1 WS CAN bus off
Function
Logging off*
WS CAN bus on Logging on
WS CAN bus ex. on Extended logging on
* Factory setting
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.6 S007: Error logging memory
2. Use the arrow keys (A) to select the required setting (see table).
Once the required setting has been selected, the T key (B)
(Easypad) or the LED above the T key (B) (Multipad) lights up.
3. Touch the T key (B) to enable the selected setting.
All CAN bus events occurring from now on during operation of the unit will be logged and can be displayed with a web browser (e.g.
Internet Explorer). This log will help you when consulting the
Sirona Customer Service Center (CSC) for error diagnosis.
4. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.6.3.1
Displaying the log with a web browser
Displaying the log with a web browser
1. Enter the following web address on a PC (with internet access) integrated in a system network:
A: IP address of the unit
201
8 Perform service routines via the control panel
8.6 S007: Error logging memory
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
The CAN bus browser opens.
2. In the lower area, select the "CAN bus" link.
202
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.7 S008: Update service
SR*
S008
S008.2
S008.3
64 26 345 D3437
D3437.076.03.02.02 04.2013
SHZ**
No
No
The CAN bus protocol is displayed in the browser and can be saved as an HTML page, printed out, or sent to the Sirona
Customer Service Center (CSC).
8.7
S008: Update service
S008: Update service
Function
Checking the software versions
Overview of the module software versions
Input/confirm/query unit serial number
* SR=service routine, ** SHZ=security access
8.7.1
S008: Test step 2
S008: Test step 2
Overview of module software versions
Title only for GALILEOS
Easypad touchscreen
1. Call service routine S008.2 [ → 182].
203
8 Perform service routines via the control panel
8.7 S008: Update service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
The software versions currently installed on the modules are displayed on an info screen on the touchscreen display.
2. Exit the service routine [ → 190].
S008_2 GALILEOS
Multipad
1. Call service routine S008.2 [ → 182].
"SYSTEMSOFTWARE" is displayed on the display line of the
Multipad.
2. Select the required module in selection field 1 with the arrow keys (A) and confirm your selection by pressing the Memory key (B).
204
The software version of the selected module is displayed in selection field 1.
3. Exit the service routine [ → 190].
8.7.2
S008: Test step 3
S008: Test step 3
Confirming the unit serial number
IMPORTANT
If the backup copy of the old unit serial number does not match the new one after replacing a module, the entry of the serial number is activated.
If an incorrect serial number is entered, the message "FFFF" appears on the display. In this case, the service routine can be run again.
1. Call service routine S008.3 [ → 182].
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.8 S009: Flash file system
2. Confirm the serial number displayed by pressing the R key (A).
3. Exit the service routine [ → 190].
SR*
S009
S009.4
S009.5
S009.7
SHZ**
Yes
No
Yes
8.8
S009: Flash file system
S009: Flash file system
IMPORTANT
the error logging memory is lost.
S009 GALILEOS Function
Flash file system
Initializing the flash file system
Test flash file system
Save/restore DX89 data
* SR=service routine, ** SHZ=security access
8.8.1
S009: Test step 4
S009: Test step 4
Formatting flash file system
64 26 345 D3437
D3437.076.03.02.02 04.2013
205
8 Perform service routines via the control panel
8.8 S009: Flash file system
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. To initialize the flash file system, press the Memory key (A) (R key
(Easypad) or LED above R key (Multipad) lights up) followed by the
R key (B).
206
Flash file system formatting in progress. This process takes approx. 5-6 mins and is visualized by a progress indicator.
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.8 S009: Flash file system
The end of this process is indicated by the message "0000" in selection field 2.
3. Exit the service routine [ → 190].
8.8.2
S009: Test step 5
S009: Test step 5
Test flash file system
64 26 345 D3437
D3437.076.03.02.02 04.2013
207
8 Perform service routines via the control panel
8.8 S009: Flash file system
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. To test the flash file system, press the Memory key (A) (R key
(Easypad) or LED above R key (Multipad) lights up) followed by the
R key (B).
208
Once the system has passed the test without errors, "OK" appears in selection field 1.
3. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.8 S009: Flash file system
8.8.3
S009: Test step 7
S009: Test step 7
Save/restore DX89 data
1. Call service routine S009.7 [ → 182].
After you select this service routine, the following can be displayed in selection field 1:
DX89 ⇒ DX11:
Data is transferred from DX89 to DX11, memory key (A)
(Easypad) or LED above the memory key (A) (Multipad) is lit up.
DX11 ⇒ DX89:
Data is transferred from DX11 to DX89, memory key (A)
(Easypad) or LED above the memory key (A) (Multipad) is lit up.
"---":
Data on both boards (DX11 and DX89) is valid or the data transfer is not possible, all keys (Easypad) or LEDs (Multipad) are not lit.
Only one practical direction of data transfer is offered at any one time. If both locations contain valid data,
"---" is displayed.
64 26 345 D3437
D3437.076.03.02.02 04.2013
209
8 Perform service routines via the control panel
8.8 S009: Flash file system
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. To trigger the memory process, press the Memory key (A) (R key (B)
(Easypad) or LED above the R key (B) (Multipad) lights up) followed by the R key (B).
210
The data are transferred. During the data transfer, a progress indicator is displayed in selection field 1.
3. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
SR*
S011
S011.9
S011.12
8 Perform service routines via the control panel
8.9 S011: Dosimetry (without ring movement)
SA**
Yes
Yes
8.9
S011: Dosimetry (without ring movement)
S011: Dosimetry (without ring movement)
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away from the unit as the coiled cable for the release button allows you to.
S011: GALILEOS, new
Function
Dosimetry (without ring movement)
For the GALILEOS Compact and GALILEOS Comfort:
4 s continuous radiation with 85 kV/7 mA (for current measurement)
For the GALILEOS Comfort PLUS:
4 s continuous radiation with 98 kV/6 mA (for current measurement)
Dosimetry with pulsed radiation
* SR=service routine, ** SA=security access
8.9.1
S011: Test step 9
S011: Test step 9
Current measurement (unpulsed)
1. Call service routine S011.9 [ → 182].
Selection field 1 displays the kVmA level, and selection field 2 displays the maximum radiation time.
The kVmA level and the maximum radiation time are preset and cannot be changed:
– GALILEOS Compact and GALILEOS Comfort: 85 kV/7 mA/4.0 s
– GALILEOS Comfort
PLUS
: 98 kV/6 mA/4.0 s
64 26 345 D3437
D3437.076.03.02.02 04.2013
211
8 Perform service routines via the control panel
8.9 S011: Dosimetry (without ring movement)
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. Initiate the radiation.
The maximum set radiation time has elapsed.
IMPORTANT: If you let go of the release button before the maximum radiation time has elapsed, radiation is terminated prematurely and the exposure is interrupted. The actual radiation time is
not
displayed.
If you release radiation during the cool-down interval, a countdown of the remaining waiting time is displayed in the header of the Easypad or in selection field 2 on the Multipad
(automatic exposure blocking).
3. Exit the service routine [ → 190].
8.9.2
S011: Test step 12
S011: Test step 12
Dosimetry with pulsed radiation
1. Call service routine S011.12 [ → 182].
Selection field 1 displays the kVmA level:
– GALILEOS Compact and GALILEOS Comfort: 85 kV/42 mAs
– GALILEOS Comfort
PLUS
: 98 kV/12 mAs
212
2. Initiate the radiation.
Radiation uses 200 pulses and 85 kV/42 mAs or 98 kV/12 mAs.
IMPORTANT: If you let go of the release button before the maximum radiation time has elapsed, radiation is terminated prematurely and the exposure is interrupted. The actual radiation time is
not
displayed.
If you release radiation during the cool-down interval, a countdown of the remaining waiting time is displayed in the header of the Easypad or in selection field 2 on the Multipad
(automatic exposure blocking).
3. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
SR*
S012
S012.1
8 Perform service routines via the control panel
8.10 S012: CAN bus service
SHZ**
No
8.10
S012: CAN bus service
S012: CAN bus service
Function
CAN bus service
Presence display of modules
* SR=service routine, ** SHZ=security access
IMPORTANT
The CAN bus service is not yet implemented for the module DX11!
8.10.1
S012: Test step 1
S012: Test step 1
Presence display of modules
Selection field Parameter/Display Range of values
DX1 - DX88 1
2
Subassembly
● Counter value of
CAN bus events
● Presence code behind the counter value:
P = module present
L = module lost
➢ Call service routine S012.1 [ → 182].
64 26 345 D3437
D3437.076.03.02.02 04.2013
213
8 Perform service routines via the control panel
8.10 S012: CAN bus service
Checking the module
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
S012.1: GALILEOS
➢ Use the arrow keys (A) in selection field 1 to select the required module.
The counter value of the CAN bus events processed so far (since the last switch-on of the unit) of the selected module is displayed in selection field 2 with the presence code of the module ("L" or
"P") (see table).
Once the module has been selected, the T key (B) (Easypad) or the LED above the T key (B) (Multipad) lights up.
214
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Clearing the counter for the module
8 Perform service routines via the control panel
8.10 S012: CAN bus service
1. To delete the counter, press the T key (B).
The counter is then reset to "0".
2. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.10.2
S012: Test step 2
S012: Test step 2
Querying the CAN status register for the modules
IMPORTANT
1. Call service routine S012.2 [ → 182].
The currently selected module is displayed in selection field 1
(DX7 in the example).
215
8 Perform service routines via the control panel
8.10 S012: CAN bus service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. Use the arrow keys (A) in selection field 1 to select the required module.
Once the module has been selected, the T key (B) lights up.
3. Touch the T key (B).
The CAN status registers for the module called, e.g.
CAN-State
DX6
, are displayed.
If values deviate from "zero", it points to a problematic CAN bus connection.
4. Exit the service routine [ → 190].
8.10.3
S012: Test step 3
S012: Test step 3
Resetting the CAN status registers
IMPORTANT
Run this service routine as required before service routine S012.2.
216
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.10 S012: CAN bus service
1. Call service routine S012.3 [ → 182].
Once the service routine has been selected, the Memory key (A) lights up.
2. To delete the CAN bus registers, press the Memory key (A) (R key lights up) followed by the R key (B).
3. Exit the service routine [ → 190].
8.10.4
S012: Test step 4
S012: Test step 4
Display of CAN bus cycle on LEDs of modules
Selection field
1
Code
00
01
Function
LED display on the modules is switched off*
LED display on the modules is switched on
* Factory setting
1. Call service routine S012.4 [ → 182].
After selecting the service routine, selection field 1 displays the code for the current setting of the display (see table).
2. Use the arrow keys (A) in selection field 1 to select the required code and confirm your selection by pressing the T key (B).
3. Exit the service routine [ → 190].
The LEDs on the modules normally flash slowly (1 Hz) (for code 00).
When code 01 has been selected and confirmed, the CAN bus clock pulse of the TTP protocol, which is output by the master module as a broadcast with a frequency of 20 Hz, is output on the LEDs (the green life
LED flashes on DX7). Thus by "rocking through" the connectors, you can detect directly the contact loss of the module on the CAN bus. The activation or deactivation of this function simultaneously acts on all modules.
64 26 345 D3437
D3437.076.03.02.02 04.2013
217
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
SR*
S017
S017.2
S017.3
S017.4
(for "GALILEOS Comfort" and
"GALILEOS Comfort
PLUS
")
S017.5
(for "GALILEOS Comfort" and
"GALILEOS Comfort
PLUS
")
S017.6
S017.7
S017.13
(for "GALILEOS Comfort" and
"GALILEOS Comfort
PLUS
")
S017.14
(for "GALILEOS Comfort" and
"GALILEOS Comfort
PLUS
")
S017.15
S017.25
Yes
Yes
Yes
Yes
Yes
Yes
Yes
8.11
S017: Configuration service
SA**
Yes
Yes
Yes
S017: Configuration service
Function
Unit configuration
Confirming the unit version
Enter the country group code
Select a language
Select a language set
Enable/disable the remote control
Configure the switching plate for the swivel arm
Enable/disable the welcome screen
Enable/disable certain lines of the welcome screen
Activate/deactivate the acoustic signal for end of exposure
Select the diaphragm type
* SR=service routine, ** SA=security access
8.11.1
S017: Test step 2
S017: Test step 2
Configuring the hardware version
Selection field
1
Code
0100
0500
0800
0C00
Function
GALILEOS Comfort /
GALILEOS Compact
GALILEOS Comfort /
GALILEOS Compact incl. FACESCAN
GALILEOS Comfort
PLUS
GALILEOS Comfort
PLUS
FACESCAN
incl.
Factory setting for each configuration
218
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
1. Call service routine S017.2 [ → 190].
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
Memory key (A) (Easypad) or the LED above Memory key (A)
(Multipad) lights up.
2. Use the arrow keys (A) to select the code for the required hardware version in selection field 1 (see table).
Once the hardware version has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
4. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
219
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8.11.2
S017: Test step 3
S017: Test step 3
Enter the country group code
Selection field
1
Code
00
01
02
03
* Factory setting
Function
Worldwide*
Asia
USA
France CTDI-DAP
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
220
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
2. Use arrow keys (A) to select the required country group code in selection field 1 (see table).
Once the country group code has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
4. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.11.3
S017: Test step 4
S017: Test step 4
Select a language
Selection field
1
10
11
12
13
04
05
06
08
Code
00
01
02
03
* Factory setting varies by order
1. Call service routine S017.4 [ → 182].
Function*
English
German
French
Italian
Dutch
Spanish
Russian
Portuguese
Chinese (PRC)
Korean
Japanese
Chinese (Taiwan)
221
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
222
2. Use the arrow keys (A) to select the code for the required language in selection field 1 (see table).
Once the language has been selected, the Memory key (B) lights up.
IMPORTANT: If the selected language is not in the installed language
set (S017: Test step 5 [ → 222]), "English" is set by default.
4. Exit the service routine [ → 190].
8.11.4
S017: Test step 5
S017: Test step 5
Select a language set
IMPORTANT
A software update must be performed [ → 47] every time the language
set changes, in order to install the corresponding languages in the system.
Selection field Code
1 00
01
02
03
04
05
Function*
German, English, French, Italian
German, English, French, Dutch
German, English, Spanish, Russian
German, English, Korean, Japanese
German, English, Spanish, Portuguese
German, English, Chinese (PRC),
Chinese (Taiwan)
* Factory setting varies by order
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
1. Call service routine S017.5 [ → 182].
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the code for the required language set in selection field 1 (see table).
Once the language set has been selected, the Memory key (B) lights up.
4. Exit the service routine [ → 190].
8.11.5
S017: Test step 6
S017: Test step 6
Enable/disable the remote control
Selection field
1
Code
00
01
* Factory setting
Function
Remote control disabled*
Remote control enabled
64 26 345 D3437
D3437.076.03.02.02 04.2013
223
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the code for the required setting in selection field 1 (see table).
Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
4. Exit the service routine [ → 190].
224
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
8.11.6
S017: Test step 7
S017: Test step 7
Configuring the switching plate for the swivel arm
Selection field
1
Code
01
02
Function up to unit serial number 1079 unit serial number
1080 or higher*
* Factory setting
1. Call service routine S017.7 [ → 182].
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
64 26 345 D3437
D3437.076.03.02.02 04.2013
225
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. Use arrow keys (A) to select the required country group code in selection field 1 (see table).
Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
4. Exit the service routine [ → 190].
226
8.11.7
S017: Test step 13
S017: Test step 13
Enable/disable the welcome screen
Selection field Code
01 00
01
Function
Welcome screen disabled
Welcome screen enabled*
* Factory setting
1. Call service routine S017.13 [ → 182].
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, the Memory key (B) lights up.
4. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.11.8
S017: Test step 14
S017: Test step 14
Enable/disable certain lines of the welcome screen
Selection field
1
2
4
0
1
2
3
Code
1
* Factory setting
1. Call service routine S017.14 [ → 182].
Meaning/Function
First name
Last name
Date of birth
Patient number
Inactive*
Active
227
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Once the service routine has been selected, the code for the line currently selected is displayed in selection field 1.
2. Use the arrow keys (A) to select the required line in selection field 1
(see table).
The activation status code is displayed in selection field 2.
3. Use the arrow keys (B) to select the code for the required state of the line selected in selection field 1 in selection field 2 (see table).
Once the required setting has been selected, the Memory key (C) lights up.
5. Exit the service routine [ → 190].
228
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
8.11.9
S017: Test step 15
S017: Test step 15
Selection field Code
1 00
01
Function
Acoustic signal indicating the end of the exposure is disabled
Acoustic signal indicating the end of the exposure is enabled*
* Factory setting
Activate/deactivate the acoustic signal for end of exposure
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
64 26 345 D3437
D3437.076.03.02.02 04.2013
229
8 Perform service routines via the control panel
8.11 S017: Configuration service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
4. Exit the service routine [ → 190].
230
8.11.10
S017: Test step 25
S017: Test step 25
Select the diaphragm type
Selection field
1
Code
00
01
02
03
07
Function
Type 1 diaphragm
("GALILEOS Compact")
Type 1/Type 2 diaphragm
("GALILEOS Comfort")
Type 3 diaphragm
("GALILEOS Compact")*
Type 3 diaphragm
("GALILEOS Comfort")**
Type 3 diaphragm
("GALILEOS Comfort
PLUS
")***
* Factory setting for GALILEOS Compact
** Factory setting for GALILEOS Comfort
*** Factory setting for GALILEOS Comfort
PLUS
1. Call service routine S017.25. [ → 182]
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.11 S017: Configuration service
Once the service routine has been selected, the code for the current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required code in selection field 1 (see table).
Once the required setting has been selected, Memory key (B)
(Easypad) or the LED above Memory key (B) (Multipad) lights up.
4. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
231
SR*
S018
S018.2
S018.3
S018.4
S018.5
S018.6
8 Perform service routines via the control panel
8.12 S018: Service for height adjustment
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
No
No
No
No
No
SA**
8.12
S018: Service for height adjustment
S018: Service for height adjustment
Function
Service for height adjustment
Set the maximum travel height
Undo the maximum travel height setting
Check the height adjustment sensor system
Setting the minimum travel height
Undoing the minimum travel height setting
* SR=service routine, ** SHZ=security access
8.12.1
S018: Test step 2
S018: Test step 2
Set the maximum travel height
1. Move the unit to the required maximum travel height by pressing the
Up/Down keys in the user mode on the control panel.
2. Call service routine S018.2 [ → 182].
Once the service routine has been selected, the current height position is displayed in selection field 1.
Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
232
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.12 S018: Service for height adjustment
3. To save the maximum travel height, press Memory key (B) (R key (C) lights up) followed by R key (C).
4. Set the mechanical limit stop at the unit:
Loosen nut (D) and move mechanical limit stop (E) for the limit switch until it engages. Tighten nut (D) again.
The next time the UP key is pressed, the unit stops 10 mm below the limit switch.
5. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.12.2
S018: Test step 3
S018: Test step 3
Undo the maximum travel height setting
Once the service routine has been selected, the current height position is displayed in selection field 1.
233
8 Perform service routines via the control panel
8.12 S018: Service for height adjustment
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
2. To undo the maximum travel height setting, press Memory key (B) (R key (C (Easypad) or LED above R key (C) lights up) followed by R key
(C).
3. Exit the service routine [ → 190].
234
8.12.3
S018: Test step 4
S018: Test step 4
Check the height adjustment sensor system
This service routine is used to move the unit up or down as far as the limit switches using the Up/Down keys on the control panel. The "soft limit positions" set by the software are ignored in this case.
Display on the control panel
Patient symbol key 1
Status lit not lit
Meaning
Patient symbol key 2
Patient symbol key 3 lit not lit lit not lit
Correction switch activated
Correction switch not activated
Lower limit switch activated
Lower limit switch not activated
Upper limit switch activated
Upper limit switch not activated
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.12 S018: Service for height adjustment
Once the service routine has been selected, the current height position is displayed in selection field 1.
Patient symbol keys 1 to 3 (F) show the switching state of the limit switches (see table).
If the patient symbol key (Easypad) or the LED above the patient symbol key (Multipad) is lit, the corresponding switch is activated, i.e. the unit is at a position value greater than 1500.
64 26 345 D3437
D3437.076.03.02.02 04.2013
2. Use the UP/DOWN keys on the control panel to move the unit up and down and use the patient symbol keys (F) to check the switching states.
3. Exit the service routine [ → 190].
8.12.4
S018: Test step 5
S018: Test step 5
Setting the minimum travel height
Setting the minimum travel height is possible only for a unit height that is below the lower correction switch level (< position value of 1500)!
1. In user mode, move the unit to the required minimum travel height by pressing the Up/Down keys (A).
2. Call service routine S018.5 [ → 182].
Once the service routine has been selected, the current height position is displayed in selection field 1.
Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
235
8 Perform service routines via the control panel
8.12 S018: Service for height adjustment
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
3. To save the maximum travel height, press Memory key (B) (R key (C) lights up) followed by R key (C).
4. Exit the service routine [ → 190].
The limitation of the minimum travel height is purely software based. The lower limit switch is not mechanically adapted to the new minimum travel height!
236
8.12.5
S018: Test step 6
S018: Test step 6
Undoing the minimum travel height setting
Once the service routine has been selected, the current height position is displayed in selection field 1.
Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.13 S037: Network service
2. To undo the minimum travel height setting, press Memory key (B) (R key (C (Easypad) or LED above R key (C) lights up) followed by R key
(C).
3. Exit the service routine [ → 190].
SR*
S037
S037.1
S037.2
S037.3
S037.4
No
Yes
Yes
Yes
SA**
8.13
S037: Network service
S037: Network service
Function
Network service
Displaying the network data
Delete network addresses or set them to factory defaults
Set boot mode: DYNAMIC (DHCP/AutoIP) / STATIC (fixed address)
Manual input of static network settings (IP address, default gateway address, and subnet mask)
* SR=service routine, ** SHZ=security access
8.13.1
S037: Test step 1
S037: Test step 1
Displaying the network data
S037_1 GALILEOS
If all network data is set to default, the system is in UDP boot mode.
64 26 345 D3437
D3437.076.03.02.02 04.2013
237
8 Perform service routines via the control panel
8.13 S037: Network service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Symbol on the control panel
(LED above) patient symbol key 1 (A)
(LED above) patient symbol key 2 (B)
(LED above) patient symbol key 3 (C)
Status lit lit lit
Meaning
The IP address is displayed in selection field 1*
The default gateway is displayed in selection field 1
The subnet mask is displayed in selection field 1
* Factory setting
Selection field Parameter/Display Meaning
1 IP address, default gateway, or subnet mask of the unit
2 default static
Fixed address*
Fixed address, modified setting dynamic Automatic address assignment
* Factory setting
1. Call service routine S037.1 [ → 182].
Once the service routine has been selected, the IP address of the unit is displayed in selection field 1.
Easypad:
"Default", "static" or "dynamic" is displayed in selection field 2 (see table).
238
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.13 S037: Network service
2. You can display various items of network data in selection field 1 by pressing the patient symbol keys (A, B, or C) (see table).
The patient symbol key selected in each case (Easypad) or the
LED above the patient symbol key selected in each case
(Multipad) lights up.
3. Exit the service routine [ → 190].
8.13.2
S037: Test step 2
S037: Test step 2
Setting the default IP address, default gateway address and default subnet mask
S037.2 GALILEOS
IMPORTANT
The network address can only be restored to the factory setting (default value) in fixed address boot mode (STATIC or no DHCP).
Symbol on the control panel
(LED above) patient symbol key 1 (A)
(LED above) patient symbol key 2 (B)
(LED above) patient symbol key 3 (C)
Status lit lit lit
Meaning
The IP address is displayed in selection field 1*
The default gateway is displayed in selection field 1
The subnet mask is displayed in selection field 1
1. Call service routine S037.2 [ → 182].
64 26 345 D3437
D3437.076.03.02.02 04.2013
239
8 Perform service routines via the control panel
8.13 S037: Network service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Once the service routine has been selected, the network data is displayed as in test step 1.
Easypad:
The Memory key and the R key also become visible.
The Memory key (Easypad) or the LED above the Memory key
(Multipad) lights up.
2. Before restoring the factory settings, check the network data that is still in the system:
A = Show IP address
B = Show default gateway
C = Show subnet mask
The patient symbol key selected in each case (Easypad) or the
LED above the patient symbol key selected in each case
(Multipad) lights up.
240
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.13 S037: Network service
3. To reset the network data, press Memory key (E) (R key (Easypad) or LED above R key (Multipad) lights up) followed by R key (F).
The default network data (factory default setting) is displayed. To switch between the display of the different network data, proceed as in test step 1.
4. Perform a restart of the device.
5. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.13.3
S037: Test step 3
S037: Test step 3
Configuring boot mode
S037.3 GALILEOS
Selection field
1
Parameter
DYNAMIC
STATIC
Meaning
Automatic address assignment
(DHCP/AutoIP)
Fixed address*
* Factory setting
1. Call service routine S037.3 [ → 182].
241
8 Perform service routines via the control panel
8.13 S037: Network service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Once the service routine has been selected, the current boot mode of the unit is displayed in selection field 1.
242
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.13 S037: Network service
2. Use arrow keys (A) to select the required boot mode "automatic address assignment" (DYNAMIC) or "fixed address" (STATIC) in selection field 1 (see table).
Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
4. Exit the service routine [ → 190].
64 26 345 D3437
D3437.076.03.02.02 04.2013
8.13.4
S037: Test step 4
S037: Test step 4
Manual input of static network settings (IP address, default gateway address, and subnet mask)
S037.4 GALILEOS
This service routine cannot run in DYNAMIC mode (T key is blocked).
Symbol on the control panel
(LED above) patient symbol key 1 (A)
Status Function lit
(LED above) patient symbol key 2 (B) lit
The IP address is displayed in selection field 1* or - after pressing the T key - number pad B1 is selected
The default gateway is displayed in selection field 1 or - after pressing the T key - number pad B2 is selected
243
8 Perform service routines via the control panel
8.13 S037: Network service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
Symbol on the control panel
(LED above) patient symbol key 3 (C)
(LED above) patient symbol key 4 (D)
Status lit
Function
The subnet mask is displayed in selection field 1 or - after pressing the T key - number pad B3 is selected or - after pressing the T key - number pad B4 is selected
* Factory setting
Selection field Parameter/Display Meaning
1 IP address, default gateway, or subnet mask of the unit
2 or - after pressing the T key - selected digit default Fixed address* static dynamic
Fixed address, modified setting
Automatic address assignment
* Factory setting
1. Call service routine S037.4 [ → 182].
244
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.13 S037: Network service
Once the service routine has been selected, the IP address of the unit is displayed in selection field 1.
Easypad:
"DEFAULT", "STATIC" or "DYNAMIC" is displayed in selection field 2 (see table).
64 26 345 D3437
D3437.076.03.02.02 04.2013
2. You can display various items of network data in selection field 1 by pressing the patient symbol keys (A, B, or C) (see table).
The patient symbol key selected in each case (Easypad) or the
LED above the patient symbol key selected in each case
(Multipad) lights up.
3. To change the selected parameter, first press the T key (E).
4. Now use the patient symbol keys to select the required number pad
1-4 (A-D) (see also table):
A = Number pad B1
B = Number pad B2
C = Number pad B3
D = Number pad B4
The patient symbol key selected in each case (Easypad) or the
LED above the patient symbol key selected in each case
(Multipad) lights up.
245
8 Perform service routines via the control panel
8.13 S037: Network service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
The digit currently selected for changing is displayed in selection field 1 ("Digit No. 3" in the example).
Important: The number of the digit always refers to the currently selected number pad.
The current value of the corresponding digit is displayed in selection field 2 ("2" in the example).
246
64 26 345 D3437
D3437.076.03.02.02 04.2013
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
8 Perform service routines via the control panel
8.13 S037: Network service
5. Use arrow keys (E) to select the digit to be changed in selection field
1 ("Digit No. 12" in the example).
The corresponding patient symbol key (D) or the LED above the corresponding patient symbol key (D) lights up.
Selection field 2 displays the value of the currently selected digit.
64 26 345 D3437
D3437.076.03.02.02 04.2013
247
8 Perform service routines via the control panel
8.13 S037: Network service
Sirona Dental Systems GmbH
Service Manual (as of February 2013) GALILEOS
6. To change the value for the digit, use arrow keys (F) in selection field
2.
Memory key (G) (Easypad) or the LED above Memory key (G)
(Multipad) lights up.
8. Exit the service routine [ → 190].
9. Perform a restart of the device.
248
64 26 345 D3437
D3437.076.03.02.02 04.2013