V8plus Body Shaping System User Manual

Add to My manuals
29 Pages

The V8plus is a body shaping system that combines four technologies: bipolar radio frequency, near-infrared laser 940nm, endermologie, and pulse mechanical tissue manipulation. These technologies work together to promote the metabolism of fat cells and fiber cells, resulting in body contouring and cellulite reduction.

advertisement

V8plus Body Shaping System User Manual | Manualzz

Chapter 4 Maintenance and Troubleshooting

Warning:Ten minutes later turn off the equipment, you draw the pin of power supply to vindicate the machine. Do not leave the machine with the cover board opened up and no attendance. Or it can harm human being or the machine itself.

4.1 Protective Maintenance

The preventative maintains should be taken by our authorized staff.

4.2 To Clean and Maintain Handpiece

1) After each treatment, the handpiece shall be cleaned by ultrasonic cleaner;

2) Cleaning procedures: a) Fill 2/3 capacity of ultrasonic cleaner with warm water (the temperature of water is about 30 Degrees

Centigrade); b) Put handpiece in ultrasonic cleaner; c) After handpiece has been cleaned for 3 minutes, you dry the handpiece with tissue paper or warm wind.

4.3 Trouble Elimination Directory

Trouble

Start-up failure

Reasons

Trouble with power switch

Didn’t insert power line well

Blown fuse indicator

Measures

Change power switch

Insert power line well

Change fuse

Key on handpiece has no response

Trouble with control board

Trouble with switch power supply

Trouble with control system

Change control board

Change switch power supply

1. Shut off power switch and start-up again;

No response of touch screen Trouble with touch screen

Trouble with vacuum pump

Something wrong with negative pressure

Trouble with AC contactor

Trouble with electromagnetic valve

Trouble with 24V switch power supply

Trouble with air connection

Trouble with handpiece

RF level is 0

Change switch power supply

Change socket connector

Check the handpiece

Increase level

Trouble with PA board Change PA board

Trouble with chip of barometer on control board Change chip

No RF

2. Contact with authorized stuff

Contact with authorized stuff

Change vacuum pump

Change vacuum pump

Change electromagnetic valve

Trouble with control board

Trouble with relay for exchanging RF

Change control board

Change interposer

Trouble with inner parts of handpiece Check the handpiece

No laser

Roller doesn’t work

Laser level is 0

Trouble with inner parts of handpiece

Trouble with on control board

Speed of roller is 0

Add to level

Check the handpiece

Change control board

Add to level

Poor contact in handpiece or trouble with motor Check the handpiece

Trouble with control board Change control board

17

Chapter 5 Clinical Guide

5.1 Treatment Principle and Functions

a) Infrared laser reduces skin impedance by heating skin and RF energy penetrates deeply into connective tissue;

The synergistic combination of infrared laser and conducted RF energies increase the oxygen intracellular diffusion by heating the skin b) Vacuum plus specially designed rollers manipulate leads RF penetration to be even 5-15mm. At the same time, vacuum and roller mechanical tissue manipulation nips and stretches fibrillar connective tissue, effectively breaks down subcutaneous fat as well as extruded capillary vessel, Increase lymphatic drainage, promotes metabolism and reduces or shrinks the size of the actual fat chamber and greatly improved body contouring effect.

c) The technology that vacuum folds skin makes RF energy penetrate a specific folded skin, greatly improve effect and safety, even for the upper eyelid treatment.

5.2 Indications:

1) Black circle around eyelid removal, wrinkle around eyelid improvement and pouch improvement;

2) Upper eyelid lifting and wrinkle removal;

3) Body contouring after birth;

4) Body re-shaping, reduce waist;

5) Generalized obesity, localized adiposity, fat dissolving, skin tightening ( arm, leg, shoulder and back, mountaineering shoe, buttocks);

6) Alleviate arthritis pain and whole body physical therapy;

7) Stretch mark improvement

5.3 Contraindications

1) Pacemaker and defibrillator in body;

2) Metal under treated skin;

3) With skin diseases or cancer history;

4) With other serious disease, such as heart disease or diabetes;

5) Gestation period and breast-feed;

6) With thermal stimulus history;

7) Openness wound;

8) Light-sensitiveness;

9) Bronzer skin because of solarization;

10) Menses.

5.4 To Set Course of Treatment:

1) Face: 30 minutes / treatment, 1 treatment /week,10 times/course;

2) Eyelid: 15 minutes / treatment, one treatment / week, 10 times/course;

18

advertisement

Key Features

  • Bipolar RF
  • Near-infrared laser 940nm
  • Endermologie
  • Pulse mechanical tissue manipulation
  • Wireless IC card controlling system
  • Body contouring
  • Cellulite removal
  • Skin tightening
  • Upper canthus lifting
  • Black circle around eyes removal

Related manuals

Frequently Answers and Questions

What are the technologies used by the V8plus?
The V8plus uses four technologies: bipolar radio frequency, near-infrared laser 940nm, endermologie, and pulse mechanical tissue manipulation.
What can I expect to see after using the V8plus?
You can expect to see body contouring, cellulite reduction, skin tightening, upper canthus lifting, black circle around eyes removal, face lifting, and stretch mark improvement.
How long does a treatment take?
A full body treatment takes 45-60 minutes, while a localized treatment takes 15 minutes.
How many treatments are in a course?
A course of treatment consists of 10 treatments.
What is the treatment frequency?
For the body, treatments are done 2 times a week, and for the eye, they are done 1 time a week.