4. Maintenance
Important Matters
Dispose of replacement parts in accordance with local regulations.
4.1 Replacing the illumination bulb
• Turn the main power switch (8) off.
• Unplug the connector on the lamp housing.
• Rotate the lamp cap (26) counter clockwise and remove it from the illumination
part (A).
33
Plug
28
• Remove the old bulb and replace it with a new one. The groove in the bulb
fixation disc should be aligned with the flange of the lamp base. Improper
alignment may cause uneven illumination.
Bulb
Fixation disc
Flange
Important Matters
The bulb is hot
• Place the lamp cap in the original position and rotate it clockwise and insert
the connector.
• Turn on the main power switch and ensure that the new bulb illuminates.
4.2 Replacing the reflecting mirror
• Set the angle between the microscope and the illumination arm to exceed
30º.
29
• Incline the illumination arm more than 10º.
• Remove the long mirror by holding the extended surface.
• Insert new long or short reflecting mirror.
4.3 Replacing the fuse
• Turn off the main power switch (8) and unplug the power cord from the power
socket.
• Unscrew the fuse holder cover with the screw driver (U).
Fuse holder
30
• Remove and replace the fuse, replace the fuse cover.
• The fuse specifications are as follows:
110V 1A, 250V
220V 0,5A, 250V
Important Matters
Please select the fuse of the same type, specification and rated value.
4.4 Replacing the chin-rest paper
When the paper is exhausted, remove the pins by pulling them upward out of
the chin-rest. Install a new package of paper and replace the pins.
Pin
4.5 Adjusting the tightness of the slit width knob
If the slit width control knob is too loose, the slit width may change. Loosen the
screw on the right knob with the screw driver (W), and then hold the left knob
firmly with one hand, while rotating the right knob clockwise for tightening.
31
Screw
Right knob Left knob
When the desired tightness has been achieved, tighten the locking screw on
the right knob.
4.6 Adjusting the inclination of the illumination part
If the inclination mechanism of the illumination part is too loose, tighten the screws
on both sides of the pivot point with the screw driver (U).
4.7 Cleaning
1. Cleaning the lenses and mirrors
Remove any dust on the lenses or reflecting mirrors by brushing them with
the supplied brush (S). If any dust remains, remove it with soft cotton
moistened with absolute alcohol.
32
Important Matters
Never touch the lenses of mirrors with your fingers or any hard materials.
2. Cleaning the slide plate, rail and shaft
Wipe the dirty segment with clean soft cloth.
3. Cleaning the plastic parts
Clean the plastic parts such as the chin-rest bracket and forehead belt with
soft cloth moistened with a mild detergent or water. Disinfect patient contact
surfaces with isopropanol.
Shaft
Rail
Slide plate
Important Matters
To prevent damage to surfaces do not use corrosive cleaners.
4.8 Protection of Equipment
Always ensure that the protective cap (N) is installed over the main shaft hole
during operation. This will prevent the ingress of dust and physiological fluids
that could cause improper operation. Only remove the dust cap when assembling,
disassembling, or repairing the instrument.
When not in use always cover the instrument.
33
N
4.9 Consumable Parts
Please specify names and quantities when ordering following consumable parts.
Outlook Part name
Illumination Bulb
SL 1400
Slit Lamp
Short Mirror
Long Mirror
Chin-rest paper
Fuse 1 A (110 V)
Fuse 0,5 A (220V)
34
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 1. Name of Parts
- 8 2. Assembly
- 19 3. Operation
- 28 4. Maintenance
- 35 5. Common Trouble Shooting
- 37 6. Optional Photographic Attachment
- 65 7. Optional Observation Tube
- 67 8. Optional TV Accessories
- 68 9. Optional Barrier Filter
- 69 10. Applanation Tonometer
- 70 11. 12.5x measuring eyepieces
- 71 12. Resopnsibility
- 71 13. Transportation
- 72 14. Specifications