Chapter 7 Specifications. Ascon tecnologic RZ10000
Advertisement
Advertisement
Chapter 7 Specifications
7.1
Input Specifications
Number of Inputs and Scan Interval on the Pen Model
Item
Number of inputs
Scan interval
Specifications
1, 2, 3, or 4
125 ms
Number of Inputs and Scan Interval on the Dot Model
Item
Number of inputs
Scan interval
Specifications
6
1 s (when the integration time of the A/D converter is 20 ms or 16.7 ms)
2.5 s (when the integration time of the A/D converter is 100 ms)
Input Type
Item
Input type
Specifications
DC voltage, 1-5V, TC, RTD, DI (ON/OFF input), DC current (by adding an external shunt resistor)
Input Range and Measurable Range
Input
Volt
TC
RTD
Range Type
E
1
J
1
T
1
N
1
R
1
S
1
B
1
K
1
W
2
L
3
U
3
WRe
4
20 mV
60 mV
200 mV
2 V
6 V
20 V
50 V
1-5V
PT (Pt100)
5
JPT (JPt100)
5
Measurable Range
–20.00 to 20.00 mV
–60.00 to 60.00 mV
–200.0 to 200.0 mV
–2.000 to 2.000 V
–6.000 to 6.000 V
–20.00 to 20.00 V
–50.00 to 50.00 V
(0.800 to 1.200 V range) to (4.800 to 5.200 V range)
0.0 to 1760.0°C
0.0 to 1760.0°C
0.0 to 1820.0°C
−200.0 to 1370.0°C
−200.0 to 800.0°C
−200.0 to 1100.0°C
−200.0 to 400.0°C
0.0 to 1300.0°C
0.0 to 2315.0°C
−200.0 to 900.0°C
−200.0 to 400.0°C
0.0 to 2400.0°C
32 to 3200°F
32 to 3200°F
32 to 3308°F
−328 to 2498°F
−328.0 to 1472.0°F
−328.0 to 2012.0°F
−328.0 to 752.0°F
32 to 2372°F
32 to 4199°F
−328.0 to 1652.0°F
−328.0 to 752.0°F
32 to 4352°F
–200.0 to 600.0°C
–200.0 to 550.0°C
−328.0 to 1112.0°F
−328.0 to 1022.0°F
0: Less than 2.4 V, 1: Greater than or equal to 2.4 V
0: Open, 1: Closed
DI Level
Contact
1 R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995
2 W: W-5% Rd/W-26% Rd (Hoskins Mfg. Co.), ASTM E988
3 L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710
4 WRe: W-3%Re/W-25%Re (Hoskins Mfg Co.)
5 Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996
JPt100: JIS C1604-1989, JIS C1606-1989
Measurement current: i = 1 mA (Pt100, JPt100)
7
7-1
IM 04P03B01-01E
7.1 Input Specifications
Item
Input type
Specifications
Floating unbalanced input.
Isolation between channels (except, b terminal is shared for RTD input).
Burnout Detection of TC Can be set for each channel.
Upscale/Downscale switchable.
Input Type
TC input
1-5V input
Operating Conditions
2 kΩ or less: normal, 10 MΩ or more: burnout, detection current: approx. 10 µA
0.2 V or less: burnout
RJC of TC input Set whether to use the recorder RJC function or an external RJC function for each channel.
Set the compensation voltage when using an external RJC function.
Compensation voltage range: –19999 µV to 20000 µV.
A/D converter
Resolution
Integration time
16 bits
Pen model: Selectable from 20 ms (50 Hz), 16.7 ms (60 Hz), and Auto (auto switch between 20 ms and 16.7 ms according to the power supply frequency).
Dot model: Selectable from 20 ms (50 Hz), 16.7 ms (60 Hz), 100 ms and Auto (auto switch between 20 ms and 16.7 ms according to the power supply frequency).
Filter function (pen model) Filter ON/OFF switchable for each channel.
Time constant selectable from 2, 5, and 10 s.
Moving average function Moving average ON/OFF switchable for each channel.
(dot model) The number of samples of the moving average is selectable between 2 and 16.
Input Computation
Item
Delta computation
Linear scaling
Specifications
Calculates the difference between two channel inputs using the following equation.
Delta = (the input value of the channel set to delta computation) – (the input value of the reference channel) where (reference channel NO.) < (channel No. set to delta computation).
Computable input type: DC voltage, TC, and RTD. The channel set to delta computation and the reference channel is set to the same range type.
Scales the input value to a value with a unit suitable for the application.
Computable input type: DC voltage, TC, RTD, DI
Value after scaling
Mantissa: –19999 to 30000
Decimal position: Can be set arbitrarily
Unit: Can be set arbitrarily (up to 6 characters)
Displayable and printable range
–19999 to 31500
Computation on the 1-5V input
Linear scaling Linearly scales the 1-5V signal.
Value after scaling: Same as linear scaling.
Displayable and printable range: Same as linear scaling.
Low-cut function Sets values below 0% of the recording span to the leftmost value of the scale. The leftmost value of scale must be less than the rightmost value of scale.
Square root computation Takes the square root of the input value and linearly scales the result.
Computable input type: DC voltage
Linear scaling Value after scaling: Same as linear scaling.
Displayable and printable range: Same as linear scaling.
Low-cut function Sets values below the low-cut point to the leftmost value of scale. The leftmost value of scale must be less than the rightmost value of scale.
Low-cut point range: 0.0 to 5.0% of the recording span (0.1% steps)
Bias Adds a bias to the measuring input value.
Bias value range: ±10% of the span of the measurable range at the specified range type.
±10% of the scaling span if linear scaling is used.
7-2
IM 04P03B01-01E
7.2
Alarm Function Specifications
Item
Number of alarms
Alarm type
Hysteresis
Specifications
Up to four alarms (level) for each measurement channels.
High limit (H), low limit (L), difference high limit (h), and difference low limit (l). The symbol indicating the alarm is given in parentheses.
Set a width to the value for detecting alarm occurrence/release (common to all channels and all levels)
Applied to high limit alarm and low limit alarm.
Hysteresis range: Approx. 0.0% to 1.0% of the recording span (0.1% steps)
Display Indicates the alarm occurrence status on the display.
Display for each channel: Displays the alarm type (display precedence: , , , and ).
Alarm summary display: The ALM indicator illuminates.
Alarm relay contact output (optional specifications)
See section 7.5
7
7-3
IM 04P03B01-01E
7-4
7.3
Recording Function Specifications
Trend Recording (Pen Model)
Item
Recording pen
Step response time
Number of pens
Specifications
Disposable felt pen
Approx. 1 s (using the IEC61143 measurement method)
Up to 4
Recording color
Trend recording
Pen 1: Red, pen 2: Green, pen 3: Blue, pen 4: Violet
Updates the data at the scan interval. Continuous recording.
Pen offset compensation Records by compensating for the offset of each pen along the time axis.
Chart speed 10 to 12000 mm/h (40 steps).
Zone recording Recording zone can be specified for each channel.
Recording width: 5 mm or more (1 mm steps)
Partial expanded recording Records by expanding the right or left side of the boundary position (the other section is reduced).
Boundary position: 1 to 99%
Boundary value: Within the recording span range
Trend Recording (Dot Model)
Item
Recording method
Specifications
Trend recording by a 6-color wire dot printer
Recording color Channel 1: Purple, channel 2: Red, channel 3: Green, channel 4: Blue, channel 5: Brown, channel 6: Black
Trend recording interval Records at the following interval through a raster scan system.
AUTO: Synchronizes to the chart speed and automatically sets to the interval
FIX: 10 s/6 channels
Chart speed
Recording ON/OFF
10 to 1500 mm/h (28 steps)
Recording can be turned ON/OFF for each channel.
Zone recording Same as the pen model.
Partial expanded recording Same as the pen model.
Chart Paper
Item
Printing width
Specifications
100 mm
Type/Length Z-fold type. Approx. 16 m.
Chart paper feeding accuracy
Within ±0.1% After feeding 1000 mm and with respect to the print scale on the chart paper.
IM 04P03B01-01E
7.3 Recording Function Specifications
Printouts (Pen Model)
Item
Recording pen (color)
Specifications
Plotter pen (purple)
Alarm printout
Printout contents
Time printout format
Alarm printout buffer
Prints alarm occurrence/release.
Occurrence ( )/release ( ) marks, channel No. or tag, alarm type, alarm level, time, printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute, month:day:hour:minute:second, and year:month:day:hour:minute:second.
Stores 8 occurrences of alarm information waiting to be printed.
Periodic printout
Interval
Mode
Printout contents
Prints at specified time intervals from the reference time.
Reference time: 00 hour 00 minute to 23 hour 00 minute (1 hour steps, minute fixed)
Auto: Automatically determined from the chart speed.
Manual: Select 10 min, 12 min, 15 min, 20 min, 30 min, 1 hour, 2 hours, 3 hours, 4 hours, 6 hours,
8 hours, 12 hours, or 24 hours.
Select no periodic printout or print instantaneous values.
Date/Time, offset compensation ON/OFF, channel information (measured value, alarm status, unit, delta computation ON/OFF, scale, channel number or tag, recording color), chart speed (with time ticks)
Message printout
Number of messages
Printout contents
Time printout format
Prints preset message strings.
5
Time, message (up to 16 characters), printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute, month:day:hour:minute:second, year:month:day:hour:minute:second, and not print.
Message printout buffer Stores 5 messages waiting to be printed.
Recording start printout
Printout contents
Time printout format
Prints at the start of recording.
Time and chart speed, printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month:day:hour:minute, month:day:hour:minute:second, and year:month:day:hour:minute:second.
New chart speed printout Prints when the chart speed is changed.
Printout contents Chart speed and time, printout buffer overflow mark.
Time printout format Selectable from hour:minute, hour:minute:second, month:day:hour:minute, month:day:hour:minute:second, and year:month:day:hour:minute:second.
Manual printout
Printout contents
Prints the measured values of all channels.
Measured values of all channels at that point.
Trend recording is temporarily suspended.
Setting printout (List)
Printout contents
Prints the settings of items in Setting mode.
Range setting, alarm setting, etc.
Trend recording is temporarily suspended.
Setting printout (setup list) Prints the settings of items in Basic Setting mode.
Printout contents Basic alarm specifications, basic recording specifications, etc.
Trend recording is temporarily suspended.
Items dependent on the chart speed
Chart speed
10 to 1500 mm/h
1800 mm/h or higher
Periodic printout
Alarm printout
Message printout
Recording start printout
New chart speed printout
Printed
Not printed
Date printout format
Chart speed
10 to 15 mm/h
20 to 30 mm/h
40 to 60 mm/h
75 to 120 mm/h
150 to 180 mm/h
200 to 300 mm/h
360 to 1500 mm/h
Periodic printout interval
8 hours
4 hours
2 hours
1 hour
30 minutes
20 minutes
10 minutes
Select the printout format of the date from Y/M/D, M/D/Y, D/M/Y, D.M.Y, and M.D.Y. The format applies to the date printout of all printout items .
7
IM 04P03B01-01E
7-5
7.3 Recording Function Specifications
Printout (Dot Model)
Item
Recording
Channel printout
Alarm printout
Printout contents
Time printout format
Alarm printout buffer
Periodic printout
Interval
Measured value
Printout contents
Specifications
Dot printing.
Prints the channel number beside the trend recording every approx. 25 mm of chart paper.
Channel printout ON/OFF selectable.
Prints alarm occurrence/release.
Occurrence ( , red)/release ( , blue) marks, channel No. or tag, alarm type, alarm level, time, printout buffer overflow mark.
Selectable from hour:minute, hour:minute:second, month day hour:minute, month day hour:minute:second, and month day year hour:minute:second.
Stores 12 occurrences of alarm information waiting to be printed.
Prints at specified time intervals from the reference time.
Same as the pen model.
Same as the pen model.
Same as the pen model except offset compensation ON/OFF and recording color are not printed.
Message printout
Recording start printout
Same as the pen model.
Same as the pen model. With time ticks.
New chart speed printout Same as the pen model. With time ticks.
Manual printout Same as the pen model.
Setting printout (List)
Setting printout (Setup)
Same as the pen model.
Same as the pen model.
Items dependent on the chart speed
Chart speed
Channel printout
10 to 100 mm/h
120 mm/h or higher
Printed
Not printed
Periodic printout
Alarm printout
Message printout
Recording start printout
New chart speed printout
Printed
Not printed
Chart speed
10 to 15 mm/h
20 to 30 mm/h
40 to 75 mm/h
80 to 100 mm/h
Periodic printout interval
8 hours
4 hours
2 hours
1 hour
Date printout format Same as the pen model.
Values for Special Measured Values
Same as “Special values” for the digital display on page 7-7.
7-6
IM 04P03B01-01E
7.4
Display Function Specifications
Display and Displayed Contents
This section explains the specifications of the display functions of measurement channels. Display examples are illustrations used to explain the displayed contents and differ in appearance from the actual displays.
Item
Display
Specifications
7-segment LED (orange)
Number of screens
Displayed contents
Display type
5 (switched using keys)
Digital display and channel ID alarm status display (automatically/manually switch the displayed channel)
Date display
Time display
OFF
Display update of measured values / computed values
When displayed channels are fixed
Dot model: Updated at the scan interval
Pen model: Updated every 1 s
When displayed channels are automatically switched
Switches the channel/alarm status and measured value at 2-s intervals.
Status display
RCD
ALM
Display type specifications
This indicator turns ON when recording is started and turns OFF when stopped.
The indicator turns ON when an alarm occurs and turns OFF when the alarm is released.
Digital display
Channel number (1 st
digit) Displayed with 1 character as follows:
Alarm (2 nd
digit)
Measurement channels: 1, 2, 3, 4, 5, 6
Displayed with the symbols below. If multiple alarms are occurring simultaneously, the alarm with higher precedence is shown.
Measured value (3
(Higher order of precedence) , , , (lower order of precedence) rd
to 7 th
digits)
• DC voltage input: Displays values between “the min. measurable value – 5% of the span of the measurable range” and “the max. measurable value + 5% of the span of the measurable range” of the range (example: –2.200 to 2.200 V for the 2 V range).
• TC and RTD input: Displays values for input values between “the min. measurable value – 10°C” and “the max. measurable value + 10°C” (example: –10.0°C to 1770.0°C for TC R)
• ON/OFF input (DI): Displays the input value 0 or 1.
• Linear scaling (1-5V, scaling, and square root): Displays value for input values between “the leftmost value of scale – 5% of scaling span” and “the rightmost value of scale + 5% of scaling span” (example: –5.0 to 105.0 when scaled to 0.000 to 100.0). However, the displayable range during scaling is from –19999 to 31500 (excluding the decimal). Values below –19999 and values above 31500 are displayed as –Over and +Over, respectively.
7
7-7
IM 04P03B01-01E
7.4 Display Function Specifications
Item
Special values
Date/Time display
Lights off
Name
Digital Display
Date display
Specifications
Measurement channel
Status
+Over
–Over
Skip
Error
+Burnout
–Burnout
Display
OVER
–OVER
SKIP
ERROR
B.OUT
B.OUT
Description
See below.
See below.
A value for channels set to skip.
Values such as when both the reference channel and measurement channel are +Over or –Over in delta computation.
A value displayed when a burnout is detected on a channel set to burnout up scale.
A value displayed when a burnout is detected on a channel set to burnout down scale.
+Over and –Over of measurement channels
A condition in which the value is outside the display range indicated in “Measured value or computed value” above. For example,
• For 2 V range, values below –2.200 V are –Over and values over 2.200 V are +Over.
• For TC R, values below –10.0°C are –Over and values over 1770.0°C are +Over.
• If scaled to 0.0 to 100.0, values below –5.0 are –Over and values above 105.0 are +Over.
However, the displayable range during scaling is from –19999 to 31500 (excluding the decimal).
Values below –19999 and values above 31500 are displayed as –Over and +Over, respectively.
Displays the date or the time.
The date display format can be selected. The format is common with the date printout format.
Displays nothing.
Display Example Description
Channel number. (1 digit), alarm (1 digit), and measured value (5 digits)
Alarm information Measured value
Channel number
Year/Month/Day Displays the lower two digits for the year.
Day
Year
Month/Day/Year
Month
Month
Day/Month/Year
Day
Day
Month
Year
Year
Time display
Hour Minute Second
7-8
IM 04P03B01-01E
7.5
Specifications of Optional Functions
Alarm Output Relay (/A1, /A2, and /A3)
Item
Operation
Number of outputs
Relay contact rating
Output format
Auxiliary functions
Diagnosis output
Specifications
Outputs relay contact signals from the dedicated terminals on the rear panel when alarms occur.
2 outputs (/A1), 4 outputs (/A2), 6 outputs (/A3)
250 VDC/0.1 A (for resistance load)
250 VAC (50/60 Hz)/3 A
NO-C-NC
Energized/De-energized operation selectable (common to all relays)
Reflash operation (Fixed to relay I01, I02, and I03. Relay release time is approx. 500 ms.)
Relay I01 can be used for diagnosis output.
Diagnosis output: Outputs a relay signal when an error in the recording section (plotter on the pen model), a burnout, or an error in the A/D converter is detected. Relay operation is fixed to de-energized.
RS-422A/485 Communication Interface (/C3)
Item
Connection
Connection type
Transmission mode
Synchronization
Baud rate
Start bit
Data length
Stop bit
Parity
Received buffer length
Escape sequence
Protocol
Communication mode
Communication distance
Specifications
Complies with EIA RS-422A/485
Multi-drop connection 1:32 (four-wire system), 1:31 (two-wire system)
Half-duplex
Start-stop synchronization
1200, 2400, 4800, 9600, 19200, or 38400 bps
Fixed to 1 bit
7 bits or 8 bits
Fixed to 1 bit
Odd, even, or none (no parity)
2047 bytes
Open and close
Dedicated protocol, Modbus slave protocol
ASCII for input/output for control and setting
ASCII or binary for input/output of measured data
1.2 km
Ethernet Communication Interface (/C7)
Item Specifications
Function
Setting/Measurement server Uses a dedicated protocol
Maintenance/Test server
Measured/computed data, setup data, status byte
Connection information, network statistics, etc.
Instrument information server Model, serial number, etc.
Electrical and mechanical specifications
Conforms to IEEE 802.3.
Interface Basic specifications
Connection
Protocol
Ethernet (10BASE-T)
TCP, IP, UDP, ICMP, and ARP
The maximum number of simultaneous connections and the number of simultaneous use
Function
Administrators
Setting/Measurement server 3
Maximum Number of Number of
Connections Users Simultaneous Users Port Number
1 2 34260/tcp
Maintenance/Test server
Instrument information server -
1
-
1
-
1 34261/tcp
34264/udp
Miscellaneous
User authentication
Keepalive
Administrator: 1, users: 6 (limitations placed on users)
Drops the connection if there is no response to an inspection packet. Use or not use is selectable.
7-9
7
IM 04P03B01-01E
7.5 Specifications of Optional Functions
Green Display (/D6)
Item
Color
Specifications
1 st
digit (channel display) and 3 rd
to 7 th
digits (data display) of the 7- segment LED: Green
2 nd
digit (alarm display): Orange
Calibration Correction (/CC1)
Item Specifications
Calibration Correction Method
Converts the measured values of the measurement channels to correction values.
Specify using revise values or absolute values.
Number of correction points: 2 to 16 (including the start and end points)
Selectable input ranges DC voltage, TC, RTD, and 1-5V
Cu10, Cu25 RTD Input (/N1)
Item Specifications
Cu10, Cu25 RTD This option allows Cu10 and Cu25 inputs to be added to the standard input types.
Input type and measurable range
Input Type
Cu10 (GE)
Measurable Range
RTD
(Measurement current i: 2 mA)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10: α = 0.00392 at 20°C
Cu10: α = 0.00393 at 20°C
Cu25*: α = 0.00425 at 0°C
* Measurement current i = 1 mA
–200 to 300°C
–328.0 to 572.0°F
Measurement and recording accuracy
Input Type
Cu10 (GE)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10: α = 0.00392 at 20°C
Cu10: α = 0.00393 at 20°C
Cu25: α = 0.00425 at 0°C
Measurement Accuracy
±(0.4% of rdg + 1.0°C)
±(0.3% of rdg + 0.8°C)
Recording Accuracy
Measurement accuracy
± (0.3% of recording span)
7-10
Expansion Inputs (/N3)
Item
Input type
Specifications
Adds the input types shown in the table below.
Thermocouple
i = 1 mA)
Input Type
PR40-20
PLATINEL
NiNiMo
W/WRe26
Type N (AWG14)
Measurement Range
0.0 to 1900.0
°
C
0.0 to 1400.0
°
C
0.0 to 1310.0
°
C
0.0 to 2400.0
°
C
0.0 to 1300.0
°
C
32 to 3452
32 to 2552
32 to 2390
32 to 4352
32 to 2372
°
°
°
°
°
F
F
F
F
F
Kp vs Au7Fe 0.0 to 300.0 K
RTD
(Measuring current
Pt25 –200.0
°
F
Pt50 –200.0
°
F
Ni100 (SAMA)
Ni100 (DIN)
Ni120
–200.0 to 250.0
°
C
–60.0 to 180.0
°
C
–70.0 to 200.0
°
C
–328.0 to 482.0
–76.0 to 356.0
–94.0 to 392.0
°
°
°
F
F
F
J263*B 0.0 to 300.0 K
Cu53 –50.0 to 150.0
°
C
Cu100:
α
= 0.00425 at 0
°
C
–50.0 to 150.0
°
C
–58.0 to 302.0
–58.0 to 302.0
°
°
F
F
IM 04P03B01-01E
7.5 Specifications of Optional Functions
Item Specifications
Measurement and recording accuracy
Input Type Measurement Accuracy Recording Accuracy
PR40-20 0 to 450
°
C Not
450 to 750
°
C
±
(0.9% of rdg + 3.2
°
C)
750 to 1100
°
C
±
(0.9% of rdg + 1.3
°
C)
1100 to 1900
°
C
±
(0.9% of rdg + 0.4
°
C)
PLATINEL
±
(0.25% of rdg + 2.3
°
C)
NiNiMo
W/WRe26 0 to 400
°
C
±
(0.25% of rdg + 0.7
°
C)
±
15.0
°
C or less
400 to 2400
°
C
±
(0.2% of rdg + 2.0
°
C)
Type N(AWG14)
Kp vs Au7Fe 0 to 20 K
±
(0.2% of rdg + 1.3
°
C) Measurement
±
4.5 K accuracy
20 to 300 K
±
2.5 K
±
(0.3% of
Pt25
±
(0.15% of rdg + 0.6
°
C) recording
Pt50
±
(0.3% of rdg + 0.6
°
C)
Ni100(SAMA)
Ni100(DIN)
±
(0.15% of rdg + 0.4
°
C)
Ni120
J263*B 0 to 40 K
40 to 300 K
±
3.0 K
±
1.0 K
Cu53
±
(0.15% of rdg + 0.8
°
C)
Cu100
±
(0.2% of rdg + 1.0
°
C)
Note) No RJC for PR40-20 (fixed to 0
°
C)
Remote Control 5 Points (/R1)
Item Specifications
Number of input terminals 5
Input type Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input terminals, and shared common
Input type and signal level
Voltage-free contact
Open collector
Signal type
Operation
Contact closed at 200 Ω or less and contact open at 100 kΩ or greater
ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 A or less
Edge, level, or trigger (250 ms or more)
Perform the following control operations by applying a given signal to the remote signal input terminal.
• Start/stop recording (edge)
• Set the clock (adjust the clock to the nearest hour with the contact input, trigger)
Time When Signal Is Input Procedure
00 min 00 s to 01 min 59 s Truncates the minutes and seconds. Example: 10:00:50 → 10:00:00
58 min 00 s to 59 min 59 s Rounds up the minutes and seconds. Example: 10:59:50 → 11:00:00
02 min 00 s to 57 min to 59 s No operation
• Manual printout (trigger)
• Message 1 to 5 printout (trigger)
• Switch the chart speed (level)
7
IM 04P03B01-01E
7-11
7.5 Specifications of Optional Functions
24-VDC/AC Power Supply Operation (/P1)
Item Specifications
Rated supply voltage 24 VDC/AC
Allowable power supply voltage range
Withstand voltage
21.6 to 26.4 VDC/AC
1000 VAC at 50/60 Hz for one minute (between the power terminal and the ground terminal)
Rated power supply frequency
50/60 Hz (for AC operation)
Allowable power supply frequency range
50 Hz ± 2% or 60 Hz ± 2% (for AC operation)
Effects of power supply voltage fluctuation
The fluctuation in the measured values is within ±1 digit and that in the recording is within ±0.1% in the recording span for the 21.6 to 26.4 VAC/DC range.
Effects of power supply frequency fluctuation
Rated power
The fluctuation in the measured values and recording is ±(0.1% of rdg + 1 digit) at the rated frequency ±2 Hz.
25 VA (for DC operation) or 35 VA (for AC operation)
Power consumption
Model 24 VDC 24 VAC Maximum
1 to 4 pen models
6 dot model
* When balanced
Approx. 7 VA*
Approx. 8 VA*
Approx. 13 VA*
Approx. 13 VA*
Approx. 35 VA
Approx. 35 VA
7-12
IM 04P03B01-01E
7.6
General Specifications
Construction
Item Specifications
Mounting
Mounting angle
Flush panel mounting (on a vertical plane)
Inclined backward up to 30 degrees from a horizontal plane.
Allowable panel thickness 2 to 26 mm
Material Case: drawn steel
Color
Front panel
Dimensions
Weight
Front door: Aluminum die-cast
Case: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Front door: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Water and dust-proof (complies with DIN40050-IP54)
144(W) × 144(H) × 220(D) mm (D: depth from the panel mounting plane)
1-pen: approx. 2.1 kg, 2-pen: approx. 2.2 kg, 3-pen: approx. 2.3 kg, 4-pen: approx. 2.4 kg
6-dot: approx. 2.5 kg
Normal Operating Conditions
Item Specifications
Ambient temperature
Ambient humidity
0 to 50°C
20 to 80%RH (at 5 to 40°C)
Allowable power supply voltage range
90 to 132, 180 to 264 VAC
Power supply frequency 50 Hz±2%, 60 Hz±2%
Vibration 10 to 60 Hz 0.2 m/s
2
or less
Shock
Magnetic field
Noise
Mounting position
Warm-up time
Altitude
Not acceptable.
400 A/m or less (DC and 50/60 Hz)
Normal mode (50/60 Hz)
Volt: The peak value including the signal must be less than 1.2 times the rated value of the range
(Example: 2 V for 2 V range).
TC: The peak value including the signal must be less than 1.2 times the rated value of the range.
RTD: 50 mV or less
Common mode noise (50/60 Hz): 250 VACrms or less for all ranges
Maximum noise voltage between channels (50/60 Hz): 250 VACrms or less.
Up to 30° of backward tilt angle allowed, no tilt to the left or right
At least 30 minutes after power on
2000 m or less above sea level
Power Supply
Item Specifications
Rated supply voltage 100 to 240 VAC
Allowable power supply voltage range
90 to 132, 180 to 264 VAC
Rated power supply frequency
Rated power
Power consumption
50 Hz, 60 Hz
40 VA
Model 100 VAC
1- to 4-pen model Approx. 12 VA*
6-dot model
* When balanced
Approx. 13 VA*
240 VAC
Approx. 17 VA*
Approx. 18 VA*
Maximum
Approx. 40 VA
Approx. 40 VA
7
7-13
IM 04P03B01-01E
7.6 General Specifications
Isolation
Item
Insulation resistance
Dielectric strength
Grounding
Specifications
Each terminal to ground terminal: 20 MΩ or more (at 500 VDC)
Power supply to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Contact output terminal to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Measuring input terminal to ground terminal: 1000 VAC (50/60 Hz), 1 minute
Between measuring input terminals: 1000 VAC (50/60 Hz), 1 minute (except for RTD input terminal)
Remote input terminal to ground terminal: 500 VDC, 1 minute
Grounding resistance: 100 Ω or less
Transport and Storage Conditions
Item
Ambient temperature
Ambient humidity
Vibration
Shock
Specifications
–25 to 60°C
5 to 95%RH (no condensation)
10 to 60 Hz, 4.9 m/s
2
maximum
392 m/s
2
maximum (in packaged condition)
Supported Standards
Item
CSA
Specifications
Certified by CSA22.2 No. 61010-1, Installation category II
1
, Measurement category II degree 2
3
2
, Pollution
Certified by UL61010B-1 (CSA NRTL/C) UL
CE
EMC
Low voltage directive
C-Tick
Noise
EN61326(Emission: Class A, Immunity: Annex A)
EN61000-3-2
EN61000-3-3
EN55011 Class A Group 1
EN61010-1, Installation category II, Measurement category II, Pollution degree 2
AS/NZS CISPR 11 Class A Group 1
Machine Noise Information Ordinance 3.GSGV, Jan 18, 1991
Maximum noise level: 60 dB (A) or less (complies with ISO7779)
1 Installation category (overvoltage category) describes a number which defines a transient overvoltage condition. It implies the regulation for impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation like a distribution board.
2 Applies to measuring circuits connected to low voltage installation, and electrical instruments supplied with power from fixed equipment such as electric switchboards.
3 “Pollution degree” describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.
7-14
IM 04P03B01-01E
7.6 General Specifications
Standard Performance
Item
Measurement and recording accuracy
The following specifications apply to operation of the recorder under standard operation conditions:
Temperature:
Humidity:
23 ± 2°C
55% ± 10%RH
Power supply voltage: 90 to 132 or 180 to 264 VAC
Power supply frequency: 50/60 Hz ± 1%
Warm-up time: At least 30 minutes.
Other ambient conditions such as vibration should not adversely affect the operation
Measurement (Digital Display) Recording (Trend)
Input Type Range Type
DC voltage
20 mV
60 mV
200 mV
2 V
Measurement Accuracy
±(0.1% of rdg + 2 digits)
Highest Res.
Recording Accuracy
10 µV
10 µV
100 µV
1 mV
Measurement accuracy
± (0.3% of recording span)
Resolution
Pen model:
Dead band 0.2% of recording span
Dot model:
Resolution 0.1 mm
DC voltage
TC
(excludes
RJC accuracy)
6 V
20 V
50 V
1-5 V
R
S
B
±(0.1% of rdg + 3 digits)
±(0.1% of rdg + 2 digits)
±(0.15% of rdg + 1°C) except R,S: 0 to 100°C, ±3.7°C
100 to 300°C, ±1.5°C
B: 400 to 600°C, ±2°C
Accuracy not warranted for values less than 400°C
1 mV
10 mV
10 mV
1 mV
0.1°C
K
±(0.15% of rdg + 0.7°C) except ±(0.15% of rdg + 1°C) for –200 to –100°C
E
J
T
±(0.15% of rdg + 0.5°C) except ±(0.15% of rdg + 0.7°C) for: –200 to –100°C
N
W
L
U
Specifications
±(0.15% of rdg + 0.7°C)
±(0.15% of rdg + 1°C)
±(0.15% of rdg + 0.5°C) except ±(0.15% of rdg + 0.7°C) for: –200 to –100°C
WRe
RTD
ON/OFF input
Pt100
JPt100
Level
Contact
±(0.2% of rdg + 1°C)
±(0.15% of rdg + 0.3°C)
Threshold level (2.4 V) accuracy ± 0.1 V
≤ 1 KΩ: ON, ≥ 100 kΩ: OFF
(Parallel capacitance
0.01 µF or less)
–
–
Measuring accuracy in case of linear scaling
Accuracy during scaling (digits) = measuring accuracy (digits) × multiplier + 2 digits (rounded up) where the multiplier = scaling span (digits)/measuring span (digits).
Example: Assuming that
• range type: 1-5 V
• measuring span: 1.000 to 5.000 V
• scaling span: 0.000 to 2.000 Then,
Measuring accuracy = ±(0.1% × 5 V + 2 digits)
= ±(0.005 V [5 digits] + 2 digits)
= ±7 digits
Multiplier = {2000 digits (0.000 to 2.000)}/{4000 digits (1.000 to 5.000)} = 0.5
Accuracy during scaling = ±(7 × 0.5 + 2) digits = 6 digits (rounded up)
Reference junction compensation
Internal/External selectable for each channel
IM 04P03B01-01E
7-15
7
7.6 General Specifications
Item Specifications
Reference junction compensation accuracy
Maximum input voltage
Above 0°C with input terminal temperature balanced (60 minutes after power on)
Type R, S, B, W, WRe: ±1.0°C
Type K, J, E, T, N, L, U: ±0.5°C
±10 VDC (continuous) for ranges of 200 mV or less, TC, RTD, and DI ranges
Input resistance
Input source resistance
±60 VDC (continuous) for 2 VDC or higher ranges
Approximately 10 MΩ or more for ranges of 200 mV or less and TC
Approximately 1 MΩ for 2 VDC or higher ranges
Volt, TC: 2 kΩ or less
RTD input: 10 Ω or less per wire (The resistance of all three wires must be equal).
10 nA or less (except when burnout detection function is enabled) Bias current
Common mode rejection ratio 120 dB (50/60 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)
Normal mode rejection ratio 40 dB or more (50/60 Hz ± 0.1%)
Noise rejection By the integrating A/D converter: The integration time is 16.7 ms, 20 ms, or 100 ms (dot model)
Low-pass filtering (pen model): Filter time constant is 2 s, 5 s, or 10 s
Moving average (dot model): The number of samples is 2 to 16.
Effects of Operating Conditions
Item Specifications
Ambient temperature With temperature variation of 10°C:
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording: Measurement fluctuation + ±0.2% of the recording span or less
* Excluding the error of reference junction compensation
Power supply fluctuation With variation within 90 to 132 V and 180 to 264 VAC (50/60 Hz):
Magnetic field
Measurement: ±1 digit or less
Recording:
±0.1% of the recording span or less
With variation of ±2 Hz from rated power frequency (power supply voltage 90 to 132, 180 to 264 VAC):
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording: Same as the measurement fluctuation
AC (50/60 Hz) and DC 400 A/m fields:
Measurement: ±(0.1% of rdg 10 digits) or less
Recording:
±0.5% of the recording span or less
Input source resistance
DC voltage range
TC range
RTD range
Mounting position
Vibration
With variation of +1 kΩ:
Ranges of 200 mV or less: Within ±10 µV
Ranges of 2V or more:
±0.1% of rdg
With variation of +1 kΩ: ±10 µV or less
With variation of 10 Ω per wire (resistance of all three wires must be equal):
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording: Measurement fluctuation + ±0.1% of the recording span or less
With maximum difference of 40 mΩ between wires: Approx. 0.1°C (for Pt100)
With a backward tilt of 30° or less
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording: Indication fluctuation + ±0.1% of the recording span or less
When a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an acceleration of 0.2 m/s
2
is applied for 2 hours
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording: Indication fluctuation + ±0.1% of the recording span or less
Other Specifications
Item
Clock
Accuracy of clock
DST
Memory backup
Key Lock Function
Key Lock Items
Specifications
With a calendar function
±100 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.
The date/time for switching between standard time and DST can be specified.
A built-in lithium battery backs up the setup parameters and runs the clock
(battery life: approximately ten years at room temperature).
Password for releasing the key lock can be set.
RCD key, MENU key, FEED key, and function under the FUNC key (manual printout, list printout, setup list printout, message printout, printout buffer clear, and pen/ribbon cassette exchange)