Figure 5-7 INSULATION MODE: Current Too Large. Wayne Kerr 3260B

Figure 5-7 INSULATION MODE: Current Too Large. Wayne Kerr 3260B

Add to My manuals
225 Pages

The Wayne Kerr 3260B is a Precision Magnetics Analyzer that allows for the 2-terminal and 4-terminal (Kelvin) measurement of inductors and transformers. It also offers DC resistance measurement capabilities. The frequency range for the device spans 20Hz to 3MHz, and the drive level can be adjusted between 1mV and 10V rms. The analyzer supports ALC to maintain drive level at the component. In addition, the 3260B has a DC bias current adjustable between 1mA and 1A and provides the option to use external 3265B DC Bias Units to achieve up to 125A DC bias current.

advertisement

Figure 5-7 INSULATION MODE: Current Too Large. Wayne Kerr 3260B | Manualzz

Advanced Operation 5–11 repetitive measurements until the key is released. To minimize the risk of electric shock, high voltages are applied to the test leads only during testing.

WARNING!

A V O I D T O U C H

I

I N G T H E T E S T C O N N E C T

I

I O N S

D

U R

I

I N G M E

A

S U R E M E N

T

O

R

T R I I M M I I N G O P E R A T I I O N

S. Although the current is limited to a safe level, there is a risk of electric shock, especially at 200V and 500V test levels. Unless energy is stored in components, e.g. capacitors, connected to the measurement terminals, the voltage at the measurement terminals will be removed within 0.5 seconds of test completion.

Unless global measurement conditions are selected (see section 5.18—The SETTINGS Page), the test voltage and other settings may be set differently for each of the three test types.

The insulation test cannot be performed if the test current exceeds 1mA, i.e. if the impedance is too low. This corresponds to a minimum resistance of 100kΩ at 100V, 200kΩ at 200V or

500kΩ at 500V. In this case a message, shown in Figure 5-7, will be displayed.

Figure 5-7 INSULATION MODE: Current Too Large

Note

If the INSULATION soft key is pressed when the insulation option is not fitted, the analyzer will report Unit Not Available (Figure 5-8).

Figure 5-8 Unit Not Available

5.8.1 Connections

Except when using a handler or 4-wire scanner, the standard connection protocol for

TRANSFORMER/TELECOM MODE may be used for all insulation test types. This is shown in Figure 5-2. The tables below show the active connection for each test. Leads marked

not used may be left connected to the analyzer but are not used for the insulation test.

For operation with a 4-wire scanner the scanner relays should be configured as for turns ratio measurement. During Insulation mode operation, set the instrument to 2-terminal operation.

Since a 4-wire scanner is connected to the analyzer with HANDLER MODE lead configuration (see Figure 5-3), it is not possible to perform a primary to secondary winding test.

advertisement

Key Features

  • 2/4 terminal measurement
  • Frequency range: 20Hz to 3MHz
  • DC bias current up to 125A
  • Automatic level control (ALC)
  • Telecom function
  • In-circuit measurements
  • Measurement of very small capacitors/inductors
  • Measurement of iron-cored/ferrite inductors
  • Measurement of transformers
  • BINNING MODE

Frequently Answers and Questions

What is the frequency range of the 3260B?
The frequency range of the 3260B is 20Hz to 3MHz.
What types of components can be measured with the 3260B?
The 3260B can measure inductors and transformers using 2-terminal or 4-terminal (Kelvin) connections.
What is the maximum DC bias current that can be supplied by the 3260B?
The 3260B can provide up to 125A DC bias current when using an external 3265B DC Bias Unit.
What are some of the advanced measurement modes available on the 3260B?
The 3260B offers advanced measurement modes such as BINNING MODE, SEQUENCE MODE, MULTI FREQ MODE, GRAPH MODE, RESONANCE MODE, and DEMAG MODE.
What is the purpose of the Telecom function on the 3260B?
The Telecom function allows for the measurement of insertion loss and return loss of line matching transformers, along with the ability to select damped network components and a blocking capacitor, if needed.

Related manuals

Download PDF

advertisement

Table of contents