Figure 5-29 SEQUENCE EDIT MODE:New Program. Wayne Kerr 3260B

Figure 5-29 SEQUENCE EDIT MODE:New Program. Wayne Kerr 3260B

Add to My manuals
225 Pages

The Wayne Kerr 3260B is a Precision Magnetics Analyzer that allows for the 2-terminal and 4-terminal (Kelvin) measurement of inductors and transformers. It also offers DC resistance measurement capabilities. The frequency range for the device spans 20Hz to 3MHz, and the drive level can be adjusted between 1mV and 10V rms. The analyzer supports ALC to maintain drive level at the component. In addition, the 3260B has a DC bias current adjustable between 1mA and 1A and provides the option to use external 3265B DC Bias Units to achieve up to 125A DC bias current.

advertisement

Figure 5-29 SEQUENCE EDIT MODE:New Program. Wayne Kerr 3260B | Manualzz

5–26

Advanced Operation

3) To return to the SEQUENCE EDIT MODE press the Edit soft key.

5.10.1.3 Creating a New Program

Figure 5-29 SEQUENCE EDIT MODE:New Program

1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.

2) Select the New soft key.

3) Type a unique program number between 1 and 65535, followed by the keypad Enter key.

A single step test program will be created assigned the number entered.

5.10.1.4 Adding a Program Label

Figure 5-30 SEQUENCE EDIT MODE:Program Label

Advanced Operation 5–27

Any test program may have a label assigned with up to 10 characters.

1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.

2) Use the and navigation keys to scroll through the program list and highlight the program to be labelled.

3) Select the Label soft key.

4) Use the navigation keys to select each label character followed by the keypad Enter key for confirmation. To cancel the last character entered use the Delete soft key.

5) To store the label with the program select the Ok softkey.

5.10.1.5 Saving an Existing Program

It is a good idea to regularly save a program during editing.

1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.

2) Select the Save softkey.

3) To overwrite the loaded test file press the Enter key on the keypad. Select the Yes softkey when prompted to confirm an existing file overwrite or No to abort.

5.10.1.6 Copying an Existing Program

1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.

2) Use the and navigation keys to scroll through the list and press the Load soft key when the desired program is highlighted.

3) Select the Save softkey.

4) Enter a unique program number in the Save As field at the bottom of the screen and confirm the program number by pressing the Enter key. A label may be added to the program number. See section Adding a Test Program Label.

5.10.1.7 Deleting a Program

1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.

2) Use the and navigation keys to scroll through the program list and highlight the program to be deleted.

3) Select the Delete softkey.

4) Program deletion is confirmed by pressing the Yes softkey. Abort the deletion process by pressing the No softkey.

advertisement

Key Features

  • 2/4 terminal measurement
  • Frequency range: 20Hz to 3MHz
  • DC bias current up to 125A
  • Automatic level control (ALC)
  • Telecom function
  • In-circuit measurements
  • Measurement of very small capacitors/inductors
  • Measurement of iron-cored/ferrite inductors
  • Measurement of transformers
  • BINNING MODE

Frequently Answers and Questions

What is the frequency range of the 3260B?
The frequency range of the 3260B is 20Hz to 3MHz.
What types of components can be measured with the 3260B?
The 3260B can measure inductors and transformers using 2-terminal or 4-terminal (Kelvin) connections.
What is the maximum DC bias current that can be supplied by the 3260B?
The 3260B can provide up to 125A DC bias current when using an external 3265B DC Bias Unit.
What are some of the advanced measurement modes available on the 3260B?
The 3260B offers advanced measurement modes such as BINNING MODE, SEQUENCE MODE, MULTI FREQ MODE, GRAPH MODE, RESONANCE MODE, and DEMAG MODE.
What is the purpose of the Telecom function on the 3260B?
The Telecom function allows for the measurement of insertion loss and return loss of line matching transformers, along with the ability to select damped network components and a blocking capacitor, if needed.