Figure 5-31 SEQUENCE EDIT MODE:Additional Test Step. Wayne Kerr 3260B

Figure 5-31 SEQUENCE EDIT MODE:Additional Test Step. Wayne Kerr 3260B

Add to My manuals
225 Pages

The Wayne Kerr 3260B is a Precision Magnetics Analyzer that allows for the 2-terminal and 4-terminal (Kelvin) measurement of inductors and transformers. It also offers DC resistance measurement capabilities. The frequency range for the device spans 20Hz to 3MHz, and the drive level can be adjusted between 1mV and 10V rms. The analyzer supports ALC to maintain drive level at the component. In addition, the 3260B has a DC bias current adjustable between 1mA and 1A and provides the option to use external 3265B DC Bias Units to achieve up to 125A DC bias current.

advertisement

Figure 5-31 SEQUENCE EDIT MODE:Additional Test Step. Wayne Kerr 3260B | Manualzz

5–28

Advanced Operation

5.10.2 Sequence Mode Test Program

5.10.2.1 Adding a Test Step

Figure 5-31 SEQUENCE EDIT MODE:Additional Test Step

1) Open the program to be edited. See section 5.10.1.2 Loading an Existing Program.

2) Use the and navigation keys to select the insertion point of the new test step in the test program. The new step will be added to the program after the highlighted test step.

3) Press the Add soft key to list the available tests

4) Use the and navigation keys to highlight the type of test to be added.

5) Pressing the Ok soft key adds the test step to the program.

Notes:

• New primary winding test steps are always added after the last primary winding test in the list.

• New secondary winding test steps are always added after the last secondary winding test.

• New test steps mimic the settings of the step which was highlighted when the new step was added.

advertisement

Key Features

  • 2/4 terminal measurement
  • Frequency range: 20Hz to 3MHz
  • DC bias current up to 125A
  • Automatic level control (ALC)
  • Telecom function
  • In-circuit measurements
  • Measurement of very small capacitors/inductors
  • Measurement of iron-cored/ferrite inductors
  • Measurement of transformers
  • BINNING MODE

Frequently Answers and Questions

What is the frequency range of the 3260B?
The frequency range of the 3260B is 20Hz to 3MHz.
What types of components can be measured with the 3260B?
The 3260B can measure inductors and transformers using 2-terminal or 4-terminal (Kelvin) connections.
What is the maximum DC bias current that can be supplied by the 3260B?
The 3260B can provide up to 125A DC bias current when using an external 3265B DC Bias Unit.
What are some of the advanced measurement modes available on the 3260B?
The 3260B offers advanced measurement modes such as BINNING MODE, SEQUENCE MODE, MULTI FREQ MODE, GRAPH MODE, RESONANCE MODE, and DEMAG MODE.
What is the purpose of the Telecom function on the 3260B?
The Telecom function allows for the measurement of insertion loss and return loss of line matching transformers, along with the ability to select damped network components and a blocking capacitor, if needed.