The Wayne Kerr 3260B is a Precision Magnetics Analyzer that allows for the 2-terminal and 4-terminal (Kelvin) measurement of inductors and transformers. It also offers DC resistance measurement capabilities. The frequency range for the device spans 20Hz to 3MHz, and the drive level can be adjusted between 1mV and 10V rms. The analyzer supports ALC to maintain drive level at the component. In addition, the 3260B has a DC bias current adjustable between 1mA and 1A and provides the option to use external 3265B DC Bias Units to achieve up to 125A DC bias current.

advertisement

Figure 5-44 MULTI FREQ – Run. Wayne Kerr 3260B | Manualzz

Advanced Operation 5–41 values will be displayed. If either Abs or % was selected in MULTI FREQ – Set mode, the analyzer will report PASS, FAIL, HI or LO according to the table below. Figure 5-44 shows the results of running the multi-frequency test set up in section 5.14.1.1.

PASS

Major and minor terms are within the limits set.

FAIL

Major and minor terms are outside of the limits set.

HI (X), e.g. HI D, HI C

The parameter indicated is above the upper limit.

LO (X), e.g. LO L, LO Q The parameter indicated is below the lower limit.

Figure 5-44 MULTI FREQ – Run

When the bin handler option is fitted, the bin handler Pass/Fail output corresponds to the

, , and results. The Pass/Fail output goes low only when a measurement has passed all set limits, see section 4.1.11.5 for the bin handler interface pin assignment.

5.15 GRAPH MODE

GRAPH MODE allows both major and minor term component or circuit characteristics to be viewed in graphical form (linear or logarithmic) across a user-defined frequency range. Only one term can be viewed at a time but it is possible to rapidly toggle between the major and minor term views. The major term graph may be in absolute units, or as a percentage from a nominal value. The vertical axes for both the major and minor term may be pre-defined by the user. After plotting the graph, the FIT function may be used to autoscale either vertical axis for the best available resolution. Other functions within GRAPH MODE can be used to find resonance and to identify the highest peak and lowest trough across the frequency range. A marker is displayed with the graph and can move along the graph outline and give x and y coordinate readouts. Results can be printed to an Epson-compatible printer if required, after entering Code 30 from the MAIN MENU.

When first entering GRAPH MODE from the MAIN MENU, the GRAPH MODE – Set screen is displayed (Figure 5-45) which defines the parameters necessary to plot the graph.

advertisement

Key Features

  • 2/4 terminal measurement
  • Frequency range: 20Hz to 3MHz
  • DC bias current up to 125A
  • Automatic level control (ALC)
  • Telecom function
  • In-circuit measurements
  • Measurement of very small capacitors/inductors
  • Measurement of iron-cored/ferrite inductors
  • Measurement of transformers
  • BINNING MODE

Frequently Answers and Questions

What is the frequency range of the 3260B?
The frequency range of the 3260B is 20Hz to 3MHz.
What types of components can be measured with the 3260B?
The 3260B can measure inductors and transformers using 2-terminal or 4-terminal (Kelvin) connections.
What is the maximum DC bias current that can be supplied by the 3260B?
The 3260B can provide up to 125A DC bias current when using an external 3265B DC Bias Unit.
What are some of the advanced measurement modes available on the 3260B?
The 3260B offers advanced measurement modes such as BINNING MODE, SEQUENCE MODE, MULTI FREQ MODE, GRAPH MODE, RESONANCE MODE, and DEMAG MODE.
What is the purpose of the Telecom function on the 3260B?
The Telecom function allows for the measurement of insertion loss and return loss of line matching transformers, along with the ability to select damped network components and a blocking capacitor, if needed.