2.0 SPECIFICATIONS. Signametrics SMX2040


Add to my manuals
125 Pages

advertisement

2.0 SPECIFICATIONS. Signametrics SMX2040 | Manualzz

2.0 Specifications

2.1 DC Voltage Measurement

Input Characteristics

Input Resistance 330 mV & 3.3 V Ranges: >10 GΩ ,

Input Resistance 33 V & 330 V Ranges: 10.0 MΩ

Accuracy ± (% of reading + Volts) [1]

6 ½ Digits

Resolution 24 hours

23

°C ± 1°C

330 mV 330.0000 mV

100

ηV

3.3 V 3.300000 V

1

µV

33 V 33.00000 V

10

µV

330 V 330.0000 V

100

µV

0.003 + 4.5

0.002 + 10

0.003 + 250

µV

µV

µV

0.004 + 1 mV

90 Days

23

°C ± 5°C

One Year 23

°C

± 5°C

0.004 + 5.5

µV

0.007 + 8

µV

0.0025 + 12

µV

0.0045 + 17

µV

0.004 + 280

µV

0.007 + 330

µV

0.005 + 1.2 mV 0.008 + 1.5 mV

[1] With reading rate set to 10 readings per second (rps) or slower, and within one hour of DCV zero, using Relative control.

DCV Noise Rejection Normal Mode Rejection, at 50, 60, or 400 Hz ± 0.5%, is better than 95 dB for reading rates of 10 rps or lower. Common Mode Rejection (with 1 k

Ω lead imbalance) is better than 120 dB for these conditions.

2.2 DC Current Measurement

Input Characteristics

Burden Voltage < 350 mV for all ranges

Protected with 2.5A fuse ( 5x20mm, 250 V Fast)

Accuracy ± (% of reading + Amps) [1]

3.3 mA

33 mA

330 mA

2.5 A

5 ½ Digits

3.30000 mA

33.0000 mA

330.000 mA

2.50000 A

Resolution

10

ηA

100

ηA

1

µA

10

µA

24 hours

23

°C ± 5°C

90 Days

23

°C ± 5°C

0.052 + 200

ηA

0.07 + 350

ηA

0.04 + 1

µA

0.05 + 30

µA

0.55 + 50

µA

0.06 + 2

µA

0.055 + 40

µA

0.6 + 200

µA

One Year 23

°C

± 5°C

0.1 + 400

ηA

0.1 + 3

µA

0.075 + 60

µA

0.65 + 350

µA

[1] With reading rate set to 10 rps or slower, and within one hour of DCI zero, using Relative control.

2.3 AC Voltage Measurements

Input Characteristics

Input Resistance 1.0 MΩ, shunted by < 100 pF, all ranges

Crest Factor 3 at Full Scale, increasing to 7 at Lowest Specified Voltage

AC coupled Specified range: 10 Hz to 100 kHz

Typical Settling time < 0.5 sec to within 0.1% of final value

2.3.1 AC Voltage True RMS Measurement

Range

330 mV

3.3 V

33 V

250 V [2]

Full Scale 6 ½ Digits

330.0000 mV

3.300000 V

33.00000 V

250.0000 V

Lowest specified Voltage

5 mV [1]

10 mV

100 mV

1 V

Resolution

100

ηV

1

µV

10

µV

100

µV

[1] Between 5 mV and 10 mV, add 100

µV additional error to the accuracy table below. In many computer installations, if the DMM is not near a noisy board, usable voltage measurements of 1 mV can be obtained.

[2] Signal is limited to 8x10

6

Volt Hz Product. For example, the largest frequency input at 250 V is

32 kHz, or 8x10

6

Volt x Hz.

9 Signametrics

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • 6½-digit resolution for precise measurements
  • True RMS AC measurements for accurate readings of non-sinusoidal waveforms
  • 4-wire resistance measurements for low-resistance measurements
  • Built-in temperature sensor for temperature measurements
  • Fast measurement speed for efficient testing
  • Wide range of measurement functions for versatility
  • Easy-to-use interface for quick setup and operation
  • Compact and portable design for convenient use in various environments
  • Optional accessories for enhanced functionality

Related manuals

Frequently Answers and Questions

What is the accuracy of the SMX2040?
The accuracy of the SMX2040 varies depending on the measurement function and range. Please refer to the specifications section of the manual for detailed accuracy information.
Can the SMX2040 measure AC current?
Yes, the SMX2040 can measure AC current using a current shunt or clamp-on probe (not included).
What is the maximum input voltage that the SMX2040 can measure?
The maximum input voltage that the SMX2040 can measure depends on the input protection mode. Please refer to the manual for more information.
Can the SMX2040 be used for data logging?
The SMX2040 does not have built-in data logging capabilities. However, it can be connected to a computer via the USB interface to transfer measurement data.
What type of warranty is provided with the SMX2040?
The SMX2040 comes with a standard one-year warranty.

advertisement

Table of contents