- Home
- Do-It-Yourself tools
- Hand tools
- Measuring & layout tools
- MICRO-EPSILON
- optoNCDT 1700
- Operating instructions
advertisement
Appendix
1
10
3
2
12
11
13
14
4
5
6
9
7
8
View: Solder-pin side male cable connector, insulator
15.2 Pin Assignment Sensor Cable
Pin Designation
5 +U
B
6 GND
13 Analog output
Characteristics
Power supply (11 ... 30 VDC)
System ground for power supply switch signals (Laser on/off, Midpoint, Limits)
Current 4 ... 20 mA or
Voltage 0 ... 10 V
14 AGND
9 Laser on/off
10 Zero
8
7
3
4
1
2
12
11
Switching output 1
Switching output 2
Sync +
Sync
Tx -
Rx -
Tx +
Rx +
-
1
1
Switching input Laser ON / OFF
Switching input for reset
Error or limit output
Limit output
Symmetrical synchron output
(Master) or input (Slave)
RS422 - Output (symmetric)
RS422 - Input (symmetric)
Plug connector:
ODU MINI-SNAP, 14-pin, series B, Dimension 2, Code 0, IP 68
More information on www.odu.de
1) Used as trigger inputs in mode “Triggering“, see Chap. 6.14
optoNCDT 1700
Color Sensor Cable
PC1700-x red black
Coaxial inner conductor, white red and blue white and green gray and pink violet blue pink green brown gray yellow
Page 102
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 7 1. Safety
- 7 Symbols Used
- 7 1.2 Warnings
- 8 Notes on CE Identification
- 9 Proper Use
- 9 Proper Environment
- 10 Laser class
- 12 Functional Principle, Technical Data
- 12 Functional Principle
- 12 Diffuse Reflection
- 13 Direct Reflection
- 13 Real Time Control
- 14 Exposure Control
- 14 Technical Data
- 19 Control and Indicator Elements
- 20 4. Delivery
- 20 Scope of Delivery
- 20 5. Installation
- 20 Sensor Mounting Diffuse Reflection
- 27 Sensor Mounting Direct Reflection
- 30 Connector and Sensor Cable
- 31 Switching Inputs Laser On/Off, Setting Masters and the Mid-point
- 32 6. Operation
- 32 Getting Ready for Operation
- 33 Membrane Keys
- 34 LED-Functions
- 34 Inputs and Outputs
- 35 Menue, Setting the Parameters
- 63 8.5 Commands
- 63 8.5.1 Overview
- 65 Reading out the Sensor Parameters
- 66 Reading out the Sensor Settings
- 70 Set Average Number
- 73 Set Average Type
- 74 Starting and Stopping the Measurement Value Output
- 75 Set Limit Values
- 76 Assignment of the Limits to the Switch Outputs
- 77 Operation Mode
- 78 Set the Measurement Value Output Type
- 78 Set Measurement Frequency (Speed)
- 80 Error Output (Analog Output)
- 81 Synchronous and Trigger Mode
- 82 Switching off the Laser (External)
- 83 Switching the Data Format
- 84 Key Lock
- 84 Set Factory Setting
- 85 Reset Sensor
- 86 Reading out the Measurements
- 87 Enable / Lock the Flash for Setting Masters and the Mid-point
- 88 Mastering or Setting Mid-point
- 90 Instruction for Operating
- 90 Reflection Factor of the Target Surface
- 91 Error Influences
- 91 Light from other Sources
- 91 Color Differences
- 91 Temperature Influences
- 91 Mechanical Vibration
- 91 Movement Blurs
- 92 Surface Roughness
- 92 Sensor Tilting
- 93 Optimizing the Measuring Accuracy
- 94 Protective Housing
- 97 ILD1700 Tool
- 97 Installation and Preparation for Measurements
- 97 System Requirements
- 97 Cable and Program Routine Requirements
- 98 10.2 Measurement
- 99 Software Support with MEDAQLib
- 100 12. Warranty
- 100 Service, Repair
- 100 Decommissioning, Disposal
- 101 15. Appendix
- 101 Output Rate optoNCDT
- 102 Pin Assignment Sensor Cable
- 103 Pin Assignment RS422 Connection
- 104 Factory Setting
- 104 Pin Assignment PC1700-x/x/USB/OE/IND
- 105 15.6 Accessory