Maintenance and Storage
General Maintenance
This instrument performs an internal check of the unit just before the unit indicates that it is ready to measure. If the unit displays that it is ready to measure, then the system check was successfully completed and the unit is ready for use.
Battery
Replace the iPac Battery when it stops holding a charge.
Battery Replacement:
1. On the back of the iPac is the battery door. Open it by pushing the latch towards the door and lifting the door.
2. Disconnect the battery harness from the iPac.
3. Replace the battery and connect the battery harness to the iPac.
Note: Be sure that the battery is oriented correctly so that the door can close properly. If the battery is oriented incorrectly, the harness may become damaged and the door may not close.
If the battery harness becomes damaged, the battery will need to be replaced. The battery should connect at the harness, the wires should lay flat along the side of the battery compartment, and then attach to the battery at the bottom. If the wires are bunched up at the top, near the point of connection to the iPac, then the battery is oriented incorrectly.
4. Attach the door and ensure that it latches closed.
5. Charge the battery for approximately 10 hours before using.
Note: Refer to Disposal section of this manual for your local laws and ordinances regarding the proper disposal of the battery.
Battery Harness
Correct
Incorrect
Battery Orientation
Storage
If the instrument is to be stored for an extended period or prepared for transportation, remove the iPac battery to avoid possible damage to the instrument due to battery leakage.
30 16040-101 Rev. H
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 Warnings and Cautions
- 7 Symbol Information
- 8 Introduction
- 10 Instrument Setup
- 10 Unpacking Instructions
- 11 Parts Identification
- 11 Optional Accessories
- 12 Icon Description
- 13 Charging the iPac Pachymeter
- 14 Power Up Mode
- 14 Measurement Mode
- 14 Sleep Mode
- 17 iPac Menu Options
- 17 Bluetooth
- 18 About
- 20 Display
- 21 Instrument Operation
- 21 Measurement Screen
- 23 Operation
- 23 Operational Check
- 23 Patient Preparation
- 24 Measurement Process
- 26 Bluetooth Printing
- 27 iPac Bluetooth Printer
- 28 Cleaning and Disinfection
- 28 Measurement Tip Cleaning Instructions
- 29 Measurement Tip High Level Disinfection Instructions
- 30 Maintenance and Storage
- 30 General Maintenance
- 30 Battery
- 30 Storage
- 31 Troubleshooting
- 32 Specifications
- 32 Disposal
- 32 Software Revision
- 37 Regulatory Compliance - Bluetooth
- 38 Acoustic Output
- 40 Warranty
- 41 Appendix A