3. Scale Fidelity (Log)

3. Scale Fidelity

(Log)

Specification

5 *l.O dB max. over 0 to 80 dB display

< f 1.5 dB max. over 0 to 90 dB display

Related Performance

Amplitude Fidelity Test

Test

Description

The specification listed is for cumulative error. Only cumulative error is measured in this procedure.

A signal source of known amplitude is connected to the spectrum analyzer and the analyzer is adjusted for a reference. The signal source is stepped down in 1 dB steps and the displayed signal amplitude on the spectrum analyzer measured at each step. This measurement is performed in both the 3 kHz and 300 kHz bandwidths.