Operation
Fig. 5 Calibration Setup
(1) To bypass this setup, press the or button; otherwise, continue as follows.
(2) Press the q button to enter the submenu.
The calibration gas concentration of the first sensor is shown.
(3) Press the or button to change the value.
(4) Press the q button to store the value.
Setup screen for the next sensor is displayed.
(5) Repeat the procedure for all other sensors.
After setting the last sensor, the user is prompted to set CALDUE.
(6) Press the or button to enable or disable CALDUE.
(7) Confirm with the q button.
(8) If CALDUE is set on, press the or button to select the number of days
(9) Confirm with the q button.
(10) Press the or button to enable or disable the calibration password (CAL PW) screen.
(11) Confirm with the q button.
(12) After confirmation, continue to Alarm setup.
Alarm Setup
The user can switch all alarms on or off and change the alarm set points for each sensor. For more
information, see the flow charts in chapter 7.8 "Alarm Setup".
See chapter 5.1 "Factory-set Alarm Thresholds and Setpoints" for alarm adjustment limits. The value
of the High Alarm can only be set to a value that is higher than the Low Alarm set point.
US
Fig. 6 Alarm Setup
(1) To bypass this setup, press the or button; otherwise, continue as follows.
(2) Press the q button to enter the submenu.
ALTAIR
®
4XR
30
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 Correct Use
- 6 Liability Information
- 6 Safety and Precautionary Measures to be Adopted
- 9 Warranty
- 10 Overview
- 11 Device Hardware Interfaces
- 12 On-Screen Indicators
- 13 Battery Care
- 15 Viewing Additional Pages
- 18 Sensor Missing Alarm
- 18 Sensor End of Life Warning
- 18 Sensor End of Life Indicator
- 18 Monitoring Toxic Gases
- 19 2.10 Monitoring Oxygen Concentration
- 20 2.11 Monitoring Combustible Gases
- 21 Environmental Factors
- 22 Turning ON and Fresh Air Setup
- 26 Fresh Air Setup (FAS)
- 27 Measurement Mode [Normal Operation]
- 28 Device Setup
- 29 Sensor Setup
- 29 Calibration Setup
- 30 Alarm Setup
- 31 Setup Time and Date
- 32 Enable Bluetooth Operation
- 32 Bluetooth Operation
- 33 Data Logging
- 33 Function Tests
- 34 Bump Test
- 35 Bump LED
- 35 3.10 Calibration
- 35 Fresh Air Setup and Zero Calibration
- 37 Attach the Calibration Cap
- 37 Span Calibration
- 39 3.11 Time of Day Testing
- 41 Troubleshooting
- 42 Live Maintenance Procedure - Replacing and Adding a Sensor
- 43 Cleaning
- 43 Storage
- 43 Scope of Delivery
- 45 Factory-set Alarm Thresholds and Setpoints
- 46 Performance Specification
- 50 XCell Sensor Patents
- 52 Start Up Sequence (Power on)
- 53 Fresh Air Setup (FAS)
- 54 Reset Screen Controls
- 56 Bump Test
- 57 Options Setup
- 58 Sensor Setup
- 59 Calibrations
- 61 Alarm Setup
- 62 Time and Date Setup
- 63 7.10 Bluetooth Setup