Step 1: Unpack and Inspect the Modules
Step 1: Unpack and Inspect the Modules
The modules are shipped in materials which prevent damage from static. The module should only be removed from the packaging in an anti-static area ensuring that correct anti-static precautions are taken. Store all modules in anti-static envelopes when not in use.
ESD
Electrostatic discharge (ESD) can damage or destroy electronic components. Use a static-safe work station to perform all work on electronic assemblies. The figure (left) shows a static-safe work station using two types of ESD protection: conductive table-mat and wrist-strap combination, and conductive floor-mat and heel-strap combination. Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone. To ensure user safety, the static-safe accessories must provide at least 1 MΩ of isolation from ground.
DO NOT use these techniques for a static-safe work station when working on circuitry with a voltage potential greater than 500 volts.
Inspect for Damage
After unpacking an instrument, inspect it for any shipping damage. Report any damage to the shipping agent immediately, as such damage is not covered by the warranty (see warranty information at beginning of this document).
To avoid damage when handling a module, do not touch exposed connector pins.
See www.keysight.com/find/tips for information on preventing damage to your keysight equipment.
Return an Instrument for Service
Should it become necessary to return an instrument for repair or service, follow the steps below:
All component modules for an M9380A, M9381A, M9391A or M9393A instrument are factory tested, aligned, calibrated and shipped as a "bundle". It is important that you maintain the bundle when installing modules or
Keysight M9393A PXIe Performance Vector Signal Analyzer
11
Step 1: Unpack and Inspect the Modules returning an instrument for repair.
See "Step 5: Verify Operation of the Keysight M9393A PXIe Vector Signal
Analyzer" on page 29
.
1. Review the warranty information shipped with your product.
2. Contact Keysight to obtain a Return Material Authorization (RMA) and return address. For assistance finding contact information, go to www.keysight.com/find/assist .
3. Write the following information on a tag and attach it to the malfunctioning equipment: l
Name and address of owner. A P.O. box is not acceptable as a return address.
l
Description of failure or service required.
4. Pack the instrument in its original packaging. Include all cables. If the original packaging material is not available, use anti-static bubble wrap or packing peanuts and place the instrument in a sealed container and mark the container
“FRAGILE”.
5. On the shipping label, write ATTENTION REPAIR DEPARTMENT and the RMA number.
In your correspondence, refer to the modules by serial number and the instrument by model number.
12
Keysight M9393A PXIe Performance Vector Signal Analyzer
Step 2: Verify M9393A Shipment Contents and Model Options
Step 2: Verify M9393A Shipment Contents and Model Options
The Keysight M9393A PXIe Performance Vector Signal Analyzer is housed in a PXIe chassis. The minimum
Vector Signal Analyzer consists of the software, chassis, a Keysight M9365A PXIe Downconverter, a Keysight M9214A PXIe IF Digitizer, a Keysight M9308A PXIe Synthesizer, and a Keysight M9300A PXIe Frequency Reference. The M9300A may be used in this and other configurations. You may also configure the
Keysight M9380A PXIe CW Source, the Keysight M9381A PXIe Vector Signal Generator and the Keysight
M9391A PXIe Vector Signal Analyzer in the same chassis and also use the same M9300A.
Qty Keysight Part Number Description
1 M9393-10002 Software and Product Information CD, contains: Soft Front Panels, drivers, and all printed documentation in PDF format (also available at www.keysight.com/find/M9393A-driver
1 M9393-90002
1 E2094-60003
Keysight M9393A PXIe Performance Vector Signal Analyzer Startup Guide in hard copy
Keysight IO Libraries Suite CD
1 5023-1450
1 5002-3361
1 5972-3335
1 5061-7383
Wrench, socket,extension, 5/16 inch, SMA
Cable removal tool, SMB/SMP/MMCX
PXI Modular Product Startup Quick Reference
South Korean Class A EMC Declaration
1 9320-6698
1 M9300A
1 M9308A
1 M9365A
1 M9214A
1 1250-3807
1 8121-2063
1 8120-5091
China RoHS Addendum for Vector Signal Analyzer
Keysight M9300A PXIe Frequency Reference Option M9393A-300
Keysight M9308A PXIe Synthesizer
Keysight M9365A PXIe Downconverter
Keysight M9214A PXIe IF Digitizer
Adapter, T-type, SMB (male) - SMB (male) - SMB (female)
Cable, coaxial, BNC (male) - SMB (female) 1200 mm
Cable, coaxial, SMB (female ) - SMB (female) 120 mm
1 8121-2169
1 8121-2174
1 8121-2723
1 8121-2554
1 8121-2636
1 M9365-20039
1 M9365-20040
Cable, coaxial, SMB (female) - SMB (female) 165 mm
Cable, coaxial SMB (female) - SMB (female) 255 mm
Cable, coaxial, SMB (female) - SMP (female) 150 mm
Cable, coaxial, SMP (female) - SMP (female) 150 mm
Cable, coaxial, SMB (female) - SMP (female) 240 mm
Cable, semi-rigid, SMA (male) - SMA (male)
Cable, semi-rigid, SMA (male) - SMP (male)
All of the files contained on the CDs are available for downloading at the Keysight M9202A PXIe IF Digitizerwebsite at www.keysight.com/find/M9393A .
Keysight M9393A PXIe Performance Vector Signal Analyzer
13
Step 2: Verify M9393A Shipment Contents and Model Options
M9393A-300
M9393A-F08
M9393A-F14
M9393A-F18
M9393A-F27
M9393A-B04
M9393A-B10
M9393A-B16
M9393A-M01
M9393A-M05
M9393A-M10
M9393A-P08
M9393A-P14
M9393A-P18
M9393A-P27
M9393A-UNZ
M9393A-UK6
Model – Option List for the Keysight M9393A PXIe Performance Vector Signal Analyzer
M9393A
M9393A
Description
PXIe Vector Signal Analyzer - comprised of: M9308A PXIe Synthesizer, M9365A PXIe
Downconverter and M9214A PXIe IF Digitizer
Adds M9300A PXIe Frequency Reference (required for warranted specifications)
Frequency range, 9 kHz to 8.4 GHz
Frequency range, 9 kHz to 14 GHz
Frequency range, 9 kHz to 18 GHz
Frequency range, 9 kHz to 27 GHz
Analysis bandwidth, 40 MHz
Analysis bandwidth, 100 MHz
Analysis bandwidth, 160 MHz
Memory, 128 MSa
Memory, 512 MSa
Memory, 1024 MSa
Preamplifier, 8.4 GHz
Preamplifier, 14 GHz
Preamplifier, 18 GHz
Preamplifier, 27 GHz
Fast tuning
Commercial calibration certificate with test data
14
Keysight M9393A PXIe Performance Vector Signal Analyzer