Keysight 89600 Vector Signal Analysis Software for M9393A
Procedure for a Single Channel Configuration
This procedure is for setting up and configuring a single channel system. To set up and configure multiple channel systems:
See "Procedure for a Two Channel Configuration" on page 50
See "Procedure for a Three or Four Channel Configuration" on page 55
To create a new M9393A instrument, use the Keysight 89600 VSA Software and follow these steps:
1. Open the "New Hardware Configuration" window in the Keysight 89600 VSA software by selecting: Utilities > Hardware > Configurations from the menu bar. When you hover over the last element in this sequence you will see a listing of the relevant configurations. Click on the "Add New Hardware" icon .
2. In the "New Hardware Configuration" window:
46
Keysight M9393A PXIe Performance Vector Signal Analyzer
Keysight 89600 Vector Signal Analysis Software for M9393A a. Select the Keysight M9393A Analyzer entry and click on the icon to copy the entry to the "Configuration" window.
b. Configure the new M9393A instrument by selecting the modules identified by their VISA address in the "2. Select the instrument(s) to use for each logical instrument in the configuration" section.
c. Name the new configuration in the "3. Name the configuration" section. The default name
"Analyzer1" may be changed. Use a meaningful name for later identification.
d. Save the new configuration by selecting the "OK" button.
Keysight M9393A PXIe Performance Vector Signal Analyzer
47
Keysight 89600 Vector Signal Analysis Software for M9393A
Refer to the Keysight 89600 Vector Signal Analysis Software for further details about making measurements with your Keysight M9393A PXIe Performance Vector Signal Analyzer.
Parameter Control and Input Extensions
The Keysight 89600 role allows the VSA software to set the following parameters: l
Expected Input Power (Range) l
Center Frequency l
Bandwidth l
Acquisition Time l
Trigger Style (Free Run, External, IF Mag) l
Trigger Delay l
Trigger Slope l
Trigger Level
Extensions
Input extensions are the hardware configuration and setup parameters that apply to the Keysight 89600
VSA software with M9393A. To modify a hardware parameter, open the input extension dialog (Input > Extensions tab), click the parameter name and specify the value. For True/False values, a selected check box sets the value to True. A cleared check box sets the value to False.
Extensions: l
ADC Dither Enabled l
ADC Dither Level l
Auto IF Bandwidth l
Conversion Mode l
Digitizer Level Offset l
Extwernal TTL Trigger Source l
IF Bandwidth l
Peak to Average l
Preamplifier l
Trigger Termination
48
Keysight M9393A PXIe Performance Vector Signal Analyzer
Keysight 89600 Vector Signal Analysis Software for M9393A
For additional information on each extension item, highlight the item and a summary is displayed near the bottom of the extension dialog box.
Running Calibration on the M9393A from the 89600 VSA
Software
It is possible to run calibrations on the M9393A VSA from the 89600 VSA software when the two are correctly configured. The calibration path in the 89600 VSA software is Utilities > Calibration.
Before running a calibration on the M9393A, it is important to make sure that there is no RF signal connected to the RF In connector of the M9365A Downconverter.module in the M9393A instrument. A signal present at
Keysight M9393A PXIe Performance Vector Signal Analyzer
49