12. Other Input Related Spurious Responses
A synthesized source and the spectrum analyzer are set to the same frequency and the amplitude of the source is set to -10 dBm. A marker-amplitude reference is set on the analyzer. The source is then tuned to several different frequencies which should generate image responses. At each source frequency, the source amplitude is set to -10 dBm and the amplitude of the response, if any, is measured using the spectrum analyzer marker function.
The marker-amplitude difference is then compared to the specification.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter)
Power sensor
Adapter, Type N (f) to APC 3.5 (f)
Adapter, Type N (f) to Type N (f)
Cable, Type N, 183 cm (72 in)
Additional Equipment for Option 001
Power sensor
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 0
Adapter, Type N (f) to Type N (f)
l-50 Calibrating HP 859OD Spectrum Analyzer
12. Other Input Related Spurious Responses
Caution
Use only 75 cables, connectors, or adapters on the 75 62 input of an Option
001 or damage to the input connector will occur.
SYNTHESIZED SWEEPER
MEASURING RECEIVER
SPECTRUM ANALYZER
\
I
I
I
I
’ * l ADAPTER
T Y P E N C A B L E
I
I
I
I
I
MINIMUM ;
LOSS
POWER SENSOR
ADAPTER
’ MINIMUM ADAPTER I
; ADAPTER I
I LOSS
- - - - - - - - - - - -
I
I
-----------
I
;:
OPT ION 001 ONLY ;
------------d
Figure 1-17. Other Input Related Spurious Test Setup
XY118
Procedure
1. Zero and calibrate the measuring receiver and the power sensor in log mode so that power is read out in dBm. Enter the power sensor’s 542.8 MHz Cal Factor into the measuring receiver.
Option 001 only:
Use 75 61 power sensor.
2. Press INSTRUMENT PRESET on the synthesized sweeper and set the controls as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 542.8 MHz
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -10 dBm
Option 001 only:
POWER LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -14.3 dBm
3. Connect the equipment as shown in Figure l-17. Connect the output of the synthesizer to the power sensor using adapters.
Option 001 only:
Use the minimum loss adapter and 75 D adapter to connect to the 75 Q power sensor.
4. Adjust the synthesized sweeper power level for a -10 dBm *O. 1 dB reading on the measuring receiver.
5. On the synthesized sweeper, press SAVE 1.
6. Enter the power sensor’s Cal Factor for 1142.8 MHz into the measuring receiver.
7. Set the CW frequency on the synthesized sweeper to 1142.8 MHz.
HP 859OD Spectrum Analyzer Calibrating l-51
12.
Other Input Related Spurious Responses
8.
Adjust the synthesized sweeper power level for a - 10 dBm fO.1 dB reading on the measuring receiver.
9.
On the synthesized sweeper, press SAVE 2.
10.
Enter the power sensor’s Cal Factor for 500 MHz into the measuring receiver.
11.
Set the CW frequency on the synthesized sweeper to 500 MHz.
12.
Adjust the synthesized sweeper power level for a - 10 dBm fO.1 dB reading on the measuring receiver.
13.
Connect the synthesized sweeper to the RF INPUT of the spectrum analyzer using the appropriate cable and adapters.
Option
001
only:
Use the minimum loss adapter and 75 Q adapter as shown in Figure 1-17.
14.
On the spectrum analyzer, press (j-1 and wait for the preset to finish then set the controls as follows:
FREQUENCY)
500 @ hlO@
(
AMPLITUDE
)
0
(-dBm)
0ption 001 only:
Press
(
AMPLITUDE
],
More 1 of 2, Amptd Units, then dBm .
15.
On the spectrum analyzer, press the following keys:
(
PEAK
SEARCH)
(jMKR- MK TRACK ON OFF (ON)
=200(kHz)
Wait for the AUTO ZOOM message to disappear. Press the following spectrum analyzer keys:
[
PEAK
SEARCH)
(jj] MARKER -> REF LVL
(MKR] MK TRACK OM OFF (OFF)
(PEAK SEARCH)
MARKER A
AMPLITUDE
]
(iJJ (step-down key).
&Em-)
16.
For each of the frequencies listed in ‘Ruble 1-16, do the following: a. Set the synthesized sweeper to the listed CW frequency by pressing RECALL 1 for a CW frequency of 542.8 MHz or RECALL 2 for a CW frequency of 1148.8 MHz.
b.
Press [SGL) and wait for the completion of a new sweep.
c. On the spectrum analyzer, press
(
PEAK SEARCH
)
and record the marker-delta amplitude reading in ‘Iable 1-16 as the Actual MKR AAmplitude.
The Actual MKR A Amplitude should be less than the Maximum MKR A Amplitude listed in the table below.
The Maximum MKR A Amplitude is 10 dB more positive than the specification. This is due to the 10 dB change in reference level made in step 15.
l-52 Calibrating
HP 8590D Spectrum Analyzer
12. Other Input Related Spurious Responses
‘I&ble 1-16. Image Responses Worksheet
,
Synthesized Sweeper
CW Frequency
Actual MKR A
Maximum MKR A
L
Amplitude (d&z) Amplitude (dBc)
542.8 MHz
1142.8 MHz
- 5 5
- 5 5
17. Record the Maximum MKR A Amplitude from lhble 1-16 as TR Entry 12-1 in the performance test record.
HP 859OD Spectrum Analyzer
Calibrating 1-53
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 Notice
- 5 Certification
- 5 Warranty
- 5 Assistance
- 6 Safety Symbols
- 6 General Safety Considerations
- 7 HP 8590 Series Spectrum Analyzer Documentation Description
- 8 How to Use This Guide
- 13 Calibration
- 13 Operation Verification
- 15 Safety
- 15 Before You Start
- 21 1. Frequency Readout Accuracy
- 21 Equipment Required
- 21 Additional Equipment for Option 001
- 22 Procedure
- 23 2. Frequency Readout and Marker Count Accuracy for Option 013
- 23 Equipment Required
- 23 Additional Equipment for Option 001
- 24 Procedure
- 24 Part 1: Frequency Readout Accuracy
- 25 Part 2: Marker Count Accuracy
- 26 3. Noise Sidebands
- 26 Equipment Required
- 26 Additional Equipment for Option 001
- 27 Procedure
- 29 4. System Related Sidebands
- 29 Equipment Required
- 29 Additional Equipment for Option 001
- 30 Procedure
- 31 5. Frequency Span Readout Accuracy
- 31 Equipment Required
- 31 Additional Equipment for Option 001
- 32 Procedure
- 32 Part 1: Span 1800 MHz
- 33 Part 2: Spans <500 MHz
- 34 6. Sweep Time Accuracy
- 34 Equipment Required
- 34 Additional Equipment for Option 001
- 34 Procedure
- 36 7. Scale Fidelity
- 36 Equipment Required
- 36 Additional Equipment for Option 001
- 37 Procedure
- 37 Log Scale
- 38 Linear Scale
- 40 Log to Linear Switching
- 42 8. Reference Level Accuracy
- 42 Equipment Required
- 42 Additional Equipment for Option 001
- 43 Procedure
- 43 Log Scale
- 44 Linear Scale
- 46 9. Absolute Amplitude Calibration and Resolution Bandwidth Switching Uncertainties
- 46 Equipment Required
- 46 Additional Equipment for Option 001
- 47 Procedure
- 49 10. Calibrator Amplitude and Frequency Accuracy
- 49 Equipment Required
- 49 Additional Equipment for Option 001
- 51 Procedure
- 51 Part 1: LPF, Attenuator, and Adapter Insertion Loss Characterization
- 52 Part 2: Calibrator Amplitude Accuracy
- 53 Part 3: Calibrator Frequency Accuracy
- 54 11. Frequency Response
- 54 Equipment Required
- 54 Additional Equipment for Option 001
- 55 System Characterization Procedure for Option 001
- 56 Procedure
- 58 Frequency Response
- 59 Test Results
- 62 12. Other Input Related Spurious Responses
- 62 Equipment Required
- 62 Additional Equipment for Option 001
- 63 Procedure
- 66 13. Spurious Response
- 66 Equipment Required
- 66 Additional Equipment for Option 001
- 67 Procedure
- 67 Part 1: Second Harmonic Distortion, 30 MHz
- 69 Part 2: Third Order Intermodulation Distortion, 50 MHz
- 72 14. Gain Compression
- 72 Equipment Required
- 72 Additional Equipment for Option 001
- 73 Procedure
- 75 15. Displayed Average Noise Level
- 75 Equipment Required
- 75 Additional Equipment Required for Option 001
- 76 Procedure
- 77 400 kHz
- 78 1 MHz
- 79 1 MHz to 1.5 GHz
- 80 1.5 GHz to 1.8 GHz
- 81 16. Residual Responses
- 81 Equipment Required
- 81 Additional Equipment for Option 001
- 81 Procedure
- 81 150 kHz to 1 MHz
- 82 1 MHz to 1.8 GHz
- 84 17. Absolute Amplitude, Vernier, and Power Sweep Accuracy
- 84 Equipment Required
- 84 Additional Equipment for Option 011
- 85 Procedure
- 87 18. Tracking Generator Level Flatness
- 87 Equipment Required
- 87 Additional Equipment for Option 011
- 88 Procedure
- 91 19. Harmonic Spurious Outputs
- 91 Equipment Required
- 91 Additional Equipment for Option 011
- 92 Procedure
- 94 20. Non-Harmonic Spurious Outputs
- 94 Equipment Required
- 94 Additional Equipment for Option 011
- 95 Procedure
- 98 21. Tracking Generator Feedthrough
- 98 Equipment Required
- 98 Additional Equipment for Option 011
- 99 Procedure
- 102 22. 10 MHz Reference Output Accuracy
- 102 Equipment Required
- 102 Procedure
- 104 Performance Verification Test Record
- 112 General Specifications
- 113 Frequency Specifications
- 115 Amplitude Specifications
- 118 Option Specifications
- 118 Tracking Generator Specifications (Option 010 or 011)
- 120 Frequency Characteristics
- 121 Amplitude Characteristics
- 123 Option Characteristics
- 124 Physical Characteristics
- 124 Front-Panel Inputs and Outputs
- 124 Rear-Panel Inputs
- 128 Regulatory Information
- 129 Declaration of Conformity
- 130 Notice for Germany: Noise Declaration
- 131 Calling HP Sales and Service Offices
- 131 Before calling Hewlett-Packard
- 132 Check the Basics
- 134 Returning the Spectrum Analyzer for Service