Input. ENDRESS+HAUSER Levelflex FMP55

Add to My manuals
52 Pages

The Levelflex FMP55 is a premium device with a coated multiparameter probe for interface measurement in the oil & gas, chemical and power industries. This device is designed for safe and reliable measurement both for the interface layer and the overall level by a redundant measuring system. The Levelflex FMP55 is factory precalibrated and features a menu-guided operating concept in national languages for intuitive operation and easy integration into control or asset management systems. It also comes with advanced diagnostics and approvals for various hazardous areas.

advertisement

Input. ENDRESS+HAUSER Levelflex FMP55 | Manualzz

Measured variable

Measuring range

Blocking distance

Levelflex FMP55

Input

The measured variable is the distance between the reference point and the product surface.

Subject to the empty distance entered "E" the level is calculated.

Alternatively, the level can be converted into other variables (volume, mass) by means of linearization (32 points).

The following table describes the media groups and the possible measuring range as a function of the media group.

The maximum measuring range is:

• for rope probe in bypass or stilling well up to 10 m (33 ft)

• for rod probe in bypass or stilling well up to 4 m (13 ft)

• for coax probe up to 6 m (20 ft)

• Reduction of the max. possible measuring range through buildup, above all of moist products.

• Due to the high diffusion rate of ammonia it is recommended with gas-tight bushing

2)

for measurements

in this medium.

The upper blocking distance (= UB) is the minimum distance from the reference point of the measurement

(mounting flange) to the maximum level.

R

SD

UB

100%

E

F

LN

Measuring frequency spectrum

0%

A0011279

R = reference point of measurement

LN = probe length

UB = upper blocking distance

E = empty calibration (= zero)

F = full calibration (= span)

SD = safety distance

Blocking distance (factory setting):

• with coax probes: 0 mm (0 in)

• with rod and rope probes up to 8 m (26 ft): 200 mm (8 in)

• with rod and rope probes exceeding a length of 8 m (26 ft): 0.025 * (length of probe)

The specified blocking distances are preset on delivery. Depending on the application these settings can be changed.

Within the blocking distance, a reliable measurement can not be guaranteed.

A safety distance SD can be defined in addition to the blocking distance. A warning is generated if the level rises into this safety distance.

100 MHz to 1.5 GHz

2)

8 for FMP55 optionally available

Endress+Hauser

advertisement

Key Features

  • Coated multiparameter probe
  • Redundant measuring system
  • Measuring range up to 10m
  • Factory precalibrated
  • Menu-guided operating concept
  • Simple integration
  • Advanced diagnostics
  • ATEX, IEC Ex, CSA, FM approvals

Related manuals

Frequently Answers and Questions

What is the measuring range of the Levelflex FMP55?
The measuring range of the Levelflex FMP55 is up to 10 m (33 ft).
What types of interfaces are available for system integration with the Levelflex FMP55?
The Levelflex FMP55 offers the following interfaces for system integration: HART with 4...20 mA analog and PROFIBUS PA (Profile 3.02).
What are the temperature and pressure ranges of the Levelflex FMP55?
The temperature range of the Levelflex FMP55 is -50 to +200 °C (-58 to +392 °F). The pressure range is -1 to 40 bar (-14.5 to 580 psi).
Download PDF

advertisement

Table of contents