Performance characteristics. ENDRESS+HAUSER Levelflex FMP55

Add to My manuals
52 Pages

The Levelflex FMP55 is a premium device with a coated multiparameter probe for interface measurement in the oil & gas, chemical and power industries. This device is designed for safe and reliable measurement both for the interface layer and the overall level by a redundant measuring system. The Levelflex FMP55 is factory precalibrated and features a menu-guided operating concept in national languages for intuitive operation and easy integration into control or asset management systems. It also comes with advanced diagnostics and approvals for various hazardous areas.

advertisement

Performance characteristics. ENDRESS+HAUSER Levelflex FMP55 | Manualzz

Levelflex FMP55

Reference operating conditions

Maximum measured error

Performance characteristics

• Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)

• Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)

• Humidity = 60 % ±15 %

• Reflection factor ≥ 0,8 (surface of the water for coax probe, metal plate for rod and rope probe with min.

1 mm (0.04 in) diameter)

• Flange for rod or rope probe ³ 300 mm (12 in) diameter

• Distance to obstacles ³ 1 m (40 in)

• For interface measurement:

– Coax probe

– DC of the lower medium = 80 (water)

– DC of the upper medium = 2 (oil)

Typical data under reference operating conditions: DIN EN 61298-2, percentage values in relation to the span.

Output: digital

Sum of non-linearity, nonrepeatability and hysteresis

Level measurement:

Measuring range up to 10 m (33 ft): ±2 mm (0.08 in)

Offset / Zero

Interface measurement:

• Measuring range up to 500 mm (19.7 in): ±20 mm (0.79 in)

• Measuring range >500 mm (19.7 in): ±10 mm (0.39 in)

• If the thickness of the interface is <100 mm (3.94 in):

±40 mm (1.57 in)

±4 mm (0.16 in)

1) Add error of the analogous value to the digital value.

analog

1)

±0.02 %

±0.03 %

If the reference conditions are not met, the offset/zero point arising from the mounting situation may be up to

±12 mm (0.47 in) for rope and rod probes. This additional offset/zero point can be compensated for by entering a correction (parameter "level correction") during commissioning.

Differing from this, the following measuring error is present in the vicinity of the lower probe end:

rod probe and coax probe

80

60

40

20

0

-20

-40

-60

-80

0

DC = 2

DC > 7

50 100 150 200 250

distance from probe end [mm]

300

rope probe

80

60

40

DC > 7

20

0

-20

-40

-60

-80

0 50 100 150 200 250 300

distance from probe end [mm]

A0014154-EN

If for rope probes the DC value is less than 7, then measurement is not possible in the area of the straining weight (0 to 250 mm from end of probe; lower blocking distance).

20 Endress+Hauser

Levelflex FMP55

Differing from this, the following measuring error is present in the vicinity of the upper probe end

(rod/rope only):

Resolution

Reaction time

Influence of ambient temperature

R

40 mm

30 mm

20 mm

10 mm

D

2 mm

0 mm

- 2 mm

- 10 mm

- 20 mm

- 30 mm

- 40 mm

DC = 2

DC > 7

100 mm 200 mm

A0015091

D Sum of non-linearity, non-repeatability and hysteresis

R Reference point of measurement

DC Dielectric constant

• digital: 1 mm

• analog: 1 mA

The reaction time can be parametrized. The following step response times (as per DIN EN 61298-2)

3)

if the damping is switched off:

are valid

Probe length

< 10 m (33 ft)

Level measurement

Sampling rate

3.6 measurements/second

Step response time

< 0.8 s

Probe length

< 10 m (33 ft)

Interface measurement

Sampling rate

³ 1.1 measurements/second

The measurements are carried out in accordance with EN 61298-3

• digital (HART, PROFIBUS PA): average T

K

= 0.6 mm/10 K

• analog (current output):

– zero point (4 mA): average T

– span (20 mA): average T

K

K

= 0.02 %/10 K

= 0.05 %/10 K

Step response time

< 2.2 s

3) According to DIN EN 61209-2 the response time is the time which passes after a sudden change of the input signal until the output signal for the first time assumes 90% of the steady-state value.

Endress+Hauser 21

advertisement

Key Features

  • Coated multiparameter probe
  • Redundant measuring system
  • Measuring range up to 10m
  • Factory precalibrated
  • Menu-guided operating concept
  • Simple integration
  • Advanced diagnostics
  • ATEX, IEC Ex, CSA, FM approvals

Related manuals

Frequently Answers and Questions

What is the measuring range of the Levelflex FMP55?
The measuring range of the Levelflex FMP55 is up to 10 m (33 ft).
What types of interfaces are available for system integration with the Levelflex FMP55?
The Levelflex FMP55 offers the following interfaces for system integration: HART with 4...20 mA analog and PROFIBUS PA (Profile 3.02).
What are the temperature and pressure ranges of the Levelflex FMP55?
The temperature range of the Levelflex FMP55 is -50 to +200 °C (-58 to +392 °F). The pressure range is -1 to 40 bar (-14.5 to 580 psi).