Levelflex FMP55
Storage temperature
Climate class
Geometric height
–40 to +80 °C (–40 to +176 °F)
DIN EN 60068-2-38 (test Z/AD)
Degree of protection
Up to 2 000 m (6 600 ft) above MSL.
Can be expanded to 3 or HAW569.
000 m (9 800 ft) above MSL by application of an overvoltage protection, e.g. HAW562
• With closed housing tested according to:
– IP68, NEMA6P (24 h at 1.83 m under water surface)
– For plastic housing with transparent cover (display module): IP68 (24 h at 1.00 m under water surface)
4)
– IP66, NEMA4X
• With open housing: IP20, NEMA1 (also ingress protection of the display)
Degree of protection IP68 NEMA6P applies for M12 PROFIBUS PA plugs only when the PROFIBUS cable is plugged in and is also rated IP68 NEMA6P.
Vibration resistance
Cleaning the probe
DIN EN 60068-2-64 / IEC 68-2-64: 20 to 2 000 Hz, 1 (m/s 2 ) 2 /Hz
Depending on the application, contamination or buildup can accumulate on the probe. A thin, even layer only influences measurement slightly. Thick layers can dampen the signal and then reduce the measuring range.
Severe, uneven buildup, adhesion e.g. through crystallization, can lead to incorrect measurement. In this case, we recommend that you use a non-contact measuring principle, or check the probe regularly for soiling.
Electromagnetic compatibility
(EMC)
Electromagnetic compatibility to EN 61326 and NAMUR Recommendation EMC (NE21). Details are provided in the Declaration of Conformity
5)
. A standard installation cable is sufficient if only the analog signal is used.
Use a shielded cable when working with a superimposed communications signal (HART).
Maximum measured error: < 0.5 % of the span.
When installing the probes in metal and concrete tanks and when using a coax probe:
• Interference emission to EN 61326 - x series, electrical equipment Class B.
• Interference immunity to EN 61326 - x series, requirements for industrial areas and NAMUR
Recommendation NE 21 (EMC)
The measured value can be affected by strong electromagnetic fields when installing rod and rope probes without a shielding/metallic wall, e.g. in plastic and wooden silos.
• Interference emission to EN 61326 - x series, electrical equipment Class A.
• Interference Immunity: the measured value can be affected by strong electromagnetic fields.
Process temperature range
Process pressure limits
Operating conditions: Process
The maximum permitted temperature at the process connection is determined by the O-ring version ordered:
Device
FMP55
O-ring material
—
Process temperature
–50 to +200 °C (–58 to +392 °F); completely coated
Device
FMP55
Process pressure
–1 to 40 bar (–14.5 to 580 psi)
4)
5)
28
This restriction is valid if the following options of the product structure have been selected at the same time: 030("Display, Operation") = C("SD02") or E("SD03");
040("Housing") = A("GT19").
Can be downloaded from www.endress.com.
Endress+Hauser
Levelflex FMP55
Materials in contact with process
This range may be reduced by the selected process connection. The pressure rating (PN) specified on the flanges refers to a reference temperature of 20 °C, for ASME flanges 100 °F. Pay attention to pressuretemperature dependencies.
Please refer to the following standards for the pressure values permitted for higher temperatures:
• EN 1092-1: 2001 Tab. 18
With regard to their temperature stability properties, the materials 1.4435 and 1.4404 are grouped under 13E0 in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
• ASME B 16.5a - 1998 Tab. 2-2.2 F316
• ASME B 16.5a - 1998 Tab. 2.3.8 N10276
• JIS B 2220
• Endress+Hauser supplies DIN/EN flanges made of stainless steel according to AISI 316L (DIN/EN material number 1.4404 or 1.4435). With regard to their temperature stability properties, the materials
1.4404 and 1.4435 are grouped under 13E0 in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
• Further material specifications (® ä 33)
Flange
EN/ASME/JIS
1
2
Levelflex FMP55
No. Material
1 304 (1.4301)
2 316L (1.4435/1.4404)
4 Coating 2 mm (0.08 in): PTFE (Dyneon TFM1600)
4
A0014650
Rod probe
Æ 16 mm (2/3")
coated
Rope probe
Æ 4 mm (1/6") coated
Levelflex FMP55
Coax probe
1
2
1
2
3
1
1
No. Material
1 316L (1.4404)
2 Coating 2 mm (0.08 in): PFA (Daikin
PFA AP230)
3 Rope: 316L (1.4404)
Coating 0.75mm (0.03 in): PFA
(Daikin PFA AP230)
4 Core: 316L (1.4435)
5 PFA (Daikin PFA AP230), centering disk
5
2
2
4
2
A0013871
2
5
A0013870 A0013887
Dielectric constant (DC) and conductivity
Extension of the rope probes through temperature
• DC (upper medium) £ 10
• DC (lower medium) - DK (upper medium) ³ 10
• Interface thickness ³ 60 mm (2.4 in)
• Conductivity (upper medium): £ 1 mS/cm
• Conductivity (lower medium): ³ 100 mS/cm
Elongation through temperature increase from 30 °C (86 °F) to 150 °C (302 °F): 2 mm / m rope length
Endress+Hauser 29