Certificates and approvals. ENDRESS+HAUSER Levelflex FMP55

Add to My manuals
52 Pages

The Levelflex FMP55 is a premium device with a coated multiparameter probe for interface measurement in the oil & gas, chemical and power industries. This device is designed for safe and reliable measurement both for the interface layer and the overall level by a redundant measuring system. The Levelflex FMP55 is factory precalibrated and features a menu-guided operating concept in national languages for intuitive operation and easy integration into control or asset management systems. It also comes with advanced diagnostics and approvals for various hazardous areas.

advertisement

Certificates and approvals. ENDRESS+HAUSER Levelflex FMP55 | Manualzz

Levelflex FMP55

CE mark

C-Tick symbol

Ex approval

Functional Safety

AD2000

Telecommunications

CRN approval

Other standards and guidelines

Certificates and approvals

The measuring system meets the legal requirements of the applicable EC guidelines. These are listed in the corresponding EC Declaration of Conformity together with the standards applied.

Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

The measuring system meets the EMC requirements of the "Australian Communications and Media Authority

(ACMA)".

The devices are certified for use in hazardous areas and the relevant safety instructions are provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on the nameplate.

The separate documentation "Safety Instructions" (XA) containing all the relevant explosion protection data is available from your Endress+Hauser Sales Center. Correlation of documentations to the device

(® ä 51).

Used for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous redundancy), independently assessed by TÜV Rhineland as per IEC 61508. Other information see documentation SD00326F: "Functional Safety

Manual".

The pressure retaining material 316L (1.4435/1.4404) corresponds to AD2000 - W2/W10.

Complies with part 15 of the FCC rules for an unintentional radiator. All probes meet the requirements for a

Class A digital device.

In addition, all probes in metallic tanks as well as the coax probe meet the requirements for a Class B digital device.

Some device versions have CRN approval. For a CRN-approved device, a CRN-approved process connection has to be ordered with a CSA approval.

(® ä 42), Product structure, Features 010 "Approval" and 100 "Process Connection".

These devices are marked with the registration number 0F14480.5 on the nameplate.

• EN 60529

Degrees of protection by housing (IP code)

• EN 61010-1

Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory

Procedures.

• IEC/EN 61326

"Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC requirements)

• NAMUR NE 21

Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.

• NAMUR NE 43

Standardization of the signal level for the breakdown information of digital transmitters with analog output signal.

• NAMUR NE 53

Software of field devices and signal-processing devices with digital electronics

• NAMUR NE 107

Status classification as per NE107

• NAMUR NE 131

Requirements for field devices for standard applications

• IEC61508

Functional safety of electrical/electronic/programmable electronic safety-related systems

Endress+Hauser 41

advertisement

Key Features

  • Coated multiparameter probe
  • Redundant measuring system
  • Measuring range up to 10m
  • Factory precalibrated
  • Menu-guided operating concept
  • Simple integration
  • Advanced diagnostics
  • ATEX, IEC Ex, CSA, FM approvals

Related manuals

Frequently Answers and Questions

What is the measuring range of the Levelflex FMP55?
The measuring range of the Levelflex FMP55 is up to 10 m (33 ft).
What types of interfaces are available for system integration with the Levelflex FMP55?
The Levelflex FMP55 offers the following interfaces for system integration: HART with 4...20 mA analog and PROFIBUS PA (Profile 3.02).
What are the temperature and pressure ranges of the Levelflex FMP55?
The temperature range of the Levelflex FMP55 is -50 to +200 °C (-58 to +392 °F). The pressure range is -1 to 40 bar (-14.5 to 580 psi).