B. Braun Infusomat Space
Manuals and User Guides for B. Braun Infusomat Space. We found 6 manuals for free downloads Instructions for use, Operating instructions, User manual
Below you will find brief information for Infusomat Space. This document serves as the Instructions for Use for the Infusomat® Space volumetric infusion pump system and its accessories. It outlines detailed operational procedures, including starting and ending infusions, bolus application, and line changes. The manual also covers advanced features like dose rate calculation, drug library integration, Patient Controlled Analgesia (PCA), Target Controlled Infusion (TCI), and various infusion modes, providing comprehensive guidance for safe and effective use in clinical settings.
Key Features
Volumetric infusion pump system for parenteral and enteral fluids
Suitable for adults, pediatrics, and neonates
Supports intermittent or continuous delivery
Integrated Dose Rate Calculation
Comprehensive Drug Library with categories and patient profiles
Patient Controlled Analgesia (PCA) functionality
Target Controlled Infusion (TCI) for specific drugs (Propofol, Remifentanil)
Supports Barcoding for therapy setup
Piggyback Function for sequential infusions
Advanced infusion modes: Ramp/Taper, Program, Intermittent, Dose Over Time
Pages: 70 B. Braun Infusomat Space Instructions for use
Brand: B. Braun Size: 1 MB
Languages: English
Open manualOther manuals and documents
Frequently Answers and Questions
How do I start an infusion with the Infusomat Space?
To start an infusion, first ensure the pump is properly installed and the line is correctly inserted and primed. Close the pump door, select the line type, open the roller clamp, and establish patient connection. Then, in the Main Menu, open the rate parameter, set the desired rate, and press the Start/Stop button to begin the infusion.
What are the ways to administer a bolus with this pump?
There are three ways: Manual Bolus (press and hold the bolus button), Bolus with Volume Preselection (set a dose limit before starting), and Bolus with Rate Calculation (set a dose and time for the pump to calculate the rate). The bolus unit can also be selected.
How do I change an infusion line and start a new therapy?
First, stop the current infusion and close the roller clamp, then interrupt the patient connection. Press the door open button and open the pump door. Remove the old line and insert a new one, ensuring it's properly seated. Close the pump door, confirm the line, and if needed, prime the new line. Re-establish patient connection and then start the new infusion.
What is the intended use of the Infusomat Space?
The Infusomat Space Volumetric Infusion Pump System is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral and enteral fluids through clinically accepted routes. This includes medications, colloids, crystalloids, blood and blood components, TPN, lipids, and enteral fluids.
Questions & Answers
M M G
How should the Infusomat® Space be cleaned and disinfected?
Clean the external surface using mild soap suds, avoiding spraying near the mains connection. Use recommended wiping disinfectants, such as Meliseptol® Foam pure or Melsept SF 10%. Detailed procedures exist for the line guide element and peristalsis. Do not use Hexaquart® or any other alkylamine-containing disinfectants.
E E U
How is the Infusomat® Space pump fully powered off?
To fully power off the device, ensure the infusion line is removed, then press and hold the power button for 3 seconds while the countdown bar completes. If an infusion line is still inserted, the pump will only enter standby mode instead of shutting down.
V V C
What to do if the pump door cannot be closed to begin infusion?
If you are unable to close the pump door, press the stop button, then the open button. Press down the green opening lever completely until it locks in place, remove the line and insert a new line. In some cases, use the crank from inside the battery compartment cover to remove the emergency aperture cover. Place the crank in the aperture and turn clockwise until the pump door opens.
What to check if the pump door still won't close after trying the initial steps?
Ensure the line is properly inserted into the sensors, especially the air-sensor. Confirm that the freeflow clamp is correctly inserted into the opened aperture until the opening lever locks in place and the safety clamp squeezes the line, causing the flashing signal lamp to go out.
B B L
How are configuration parameters managed for the SpaceStation system when using the optional SpaceCom module?
Configuration parameters (such as IP address, WLAN/Ethernet settings, BCC protocol settings, user settings, and time synchronization via SNTP) can be adjusted via a web access application called SpaceOnline, which is integrated into SpaceCom. The default IP address for the Ethernet port is 192.168.100.41.
How is the loudspeaker volume adjusted or battery maintenance initiated on the SpaceStation?
Operational elements on the SpaceCover Comfort allow for controlling the 9-step volume adjustment for the integrated loudspeaker and initiating the battery maintenance program (when connected to the mains).