Below you will find brief information for Gas Sensor/Monitor GDC-350. This economical, self-contained gas detector is ideal for non-hazardous commercial applications. It's available in single or dual-sensor models, with options for on-board and/or remote sensors. The GDC-350 features LEDs for power, fail, and gas alarms, an audible alarm, and two alarm relays, making it a versatile solution for various gas detection needs.

Key Features

  • Economical gas detection
  • Single or dual-sensor configurations
  • On-board and remote sensor options
  • Audible and visual alarms
  • Two alarm relays for equipment control
  • Calibration Extending Firmware (CEF) for electrochemical sensors

Pages: 40 Bacharach GDC 350 Operation manual

Brand: Bacharach Category: Oxygen Equipment Size: 1 MB

Languages: English

Open manual

Frequently Answers and Questions

What types of sensors are compatible with the GDC-350?
Electrochemical sensors for toxic gases and oxygen, and MOS solid-state sensors for combustible gases, TVOCs, and refrigerants are available.
What is the maximum distance between the controller and a remote sensor?
500 feet maximum using minimum 18 gauge wire.
How often should the GDC-350 be calibrated?
Calibration frequency depends on the application. Parking garage detectors may require calibration once a year, while OHS applications may require calibration every six months.
What happens if a sensor fails?
The red 'Fail' LED will illuminate, both relays will de-energize, and the audible alarm will pulse.
Can the alarm setpoints be adjusted?
Yes, the alarm setpoints are field-adjustable.