- No category
advertisement
IN071
Rev B
IN071
Rev B
RadStar GM 1-2 Grab Sampler/Manometer
Turning Grab Sampler On:
Connect radon sample and filter tube to Sample Port. Turn knob to left [Grab Sampler]. The unit will beep; the pump and display will activate.
RadStar GM-1,2
HV Charging…..
After the High Voltage charges for 60 seconds* the bottom display line will display live results.
RadStar GM-1,2
000cnts 360secs
Live Display
(Note: Pressing [NEXT/ZERO] key switch will reset Live Display counter.)
After 360 seconds the top display line will display the previous 6 minute sample results and the bottom line will reset for another sample:
010cpm 10pCi/l
000cnts 000secs
To Select Grab Sampler Options:
Press [ENTER]
Unit will display:
Previous Sample
RadStar GM-1,2
Grab Sample?
Press [NEXT/ZERO]
Press [ENTER] to return to Grab Sample Mode.
RadStar GM-1,2
Radon Units?
Press [NEXT/ZERO]
Press [ENTER] to select pCi/l or Bq/m3
RadStar GM-1,2
Beep? Press [ENTER] to select Beep On or Off.
Press [NEXT/ZERO]
CAUTION: DO NOT SUCTION DUST, DIRT OR WATER WITH RADON SAMPLE TUBE.
*NOTE: [ENTER] or [ZERO/NEXT] Pushbutton switches will not respond while HV is charging.
IN071
Rev B
Turning Manometer On :
Connect one end of pressure tubing to Low manometer port** and the other end to pressure source to be measured. The High manometer port is a reference open to atmosphere. Turn knob to right [Manometer].
The unit will beep and the display will activate.
RadStar GM-1,2
+.000 Inches WC
To Zero Display Reading:
Disconnect any pressure source from High and Low Ports.
Press [NEXT/ZERO]
Low Pressure Range (0 to .250 Inches WC or 0 to 62 Pascals)
RadStar GM-1,2
Zeroing…………
AutoRange:
When a pressure is applied that exceeds the Low Pressure range the unit will automatically change to the
High Pressure setting. Press [RESET] to change back to Low Pressure range.
RadStar GM-1,2
+0.00 Inches WC High Pressure Range (0 to 8.00 inches WC or 0 to 2000 Pascals)
To Select Manometer Options:
Press [ENTER]
Unit will display:
RadStar GM-1,2
Manometer?
Press [NEXT/ZERO]
Press [ENTER] to return to Manometer Mode
RadStar GM-1,2
Pressure Units?
Press [NEXT/ZERO]
Press [ENTER] to select Inches WC or Pascals
RadStar GM-1,2
Time Average?
Press [NEXT/ZERO]
Press [ENTER] to select Time Average On or Off
**NOTE: The Low manometer port will measure positive or negative pressures.
Measuring pressure by connecting to the High manometer port will display opposite polarity (+ or -) readings.
IN071
Rev B
RadStar GM1-2 Instrument Specifications
Radon Mode: 6 minute
Range:
Grab Sampler, Brushed DC Sampling Pump rated for 500 hrs
0 to 999 picoCuries per liter (pCi/L or
0 to 37,000 Bequerels per cubic meter (Bq/m3)
Manometer Mode: (2) Ranges- 0 to .250 Inches WC or 0 to 62 Pascals
0 to 8.00 Inches WC or 0 to 2000 Pascals
User Selectable Options:
Power:
Units, Beep, Time Average
120VAC 60Hz Transformer, rechargeable NiCad battery.
Indicators:
Rotary Switch:
Communication Port:
Pushbuttons:
Operating Range:
2 X 16 LCD Display, Green AC Power LED
2 Position, [Grab Sample] [Manometer]
RS-232 port, for use by manufacturer only.
Momentary [Enter] [Next/Zero] [Reset]
Temperature 55 ̊ F to 85 ̊ F
Size:
Length
Width
Height
Weight
10 3/4 inches
9 3/4 inches
4 7/8 inches
Color:
Warranty:
Federal Safety Orange/Black
1 year