Elgar TerraSAS

Programmable Solar Array Simulator

• Simulate dynamic irradiance and temperature ranging from a clear day to cloud cover conditions

• Ramp the voltage, temperature or irradiance level over a programmed time interval

• Readback of voltage, current, irradiance level and temperature setting

• Tests for inverter Maximum Power Point Tracking

(MPPT)

• Provides programmable I-V curves for PV Inverter testing

• Simulates different types of solar cell material

• Multi-Channel, Up to 1MW

Why power supply is critical for PV simulations

Many solar inverters generate AC ripple on their

DC input, which is connected to the photovoltaic array. For single phase inverters, the frequency of this ripple is twice the line frequency (120 Hz for US models). The simulator’s power supplies must not supress this ripple as a function of their regulation loop. An increasing number of inverters (and virtually all micro-inverters) accurately measure amplitude and phase of the ripple voltage and current to quickly track the

MPP of the array. This approach allows tracking the MPP at a much higher speed when compared to conventional dithering techniques (also called perturbate-and-observe). Faster tracking of the

MPP results in a much higher overall efficiency in cloudy conditions, where the irradiance is constantly changing. It is likely that all solar inverters will use this approach in the near future, since end users are very sensitive to the overall efficiency of their solar energy installations.

To satisfy this requirement, the PV simulator must be capable of reproducing the voltage / current behavior of a solar array at the ripple frequency.

Most standard switching power supplies employ very large output capacitors and inductors in their output circuits and are unable to deliver the required performance - regardless of the response speed of the I/V curve controller.

Elgar’s line of PV simulators are based on high speed versions of our standard products, where output capacitors and other speed-limiting components have been adjusted. This results in a speed improvement of 10 times or better.

Proprietary features built into the PV controller hardware and firmware, combined with our high speed power supplies, deliver the required performance. This technology was extensively tested on micro-inverters and is ready to test the next generation of inverters.

Strengths of using DSP signal processing

Our technology avoids using linear amplifiers, which are fast but bulky and inefficient. The required performance is delivered by high speed switching power supplies and advanced DSP signal processing techniques. Competitors data sheets mentions that speed requirements may not be met in some conditions, “...depending also on the type of MPP tracking principles”. An additional linear module is required to satisfy the new requirements. Some competitor’s power supplies specifications say that it uses “...

innovative IGBT and transformer technology”.

Our power supplies use Power MOSFETs, which typically switch ten times as fast as the most recent IGBTs. Higher switching frequency translates to smaller output capacitors and inductors - which is the key to a successful high speed power supply design.

Product Overview

The Elgar TerraSAS System, (TSAS) provides an easily programmable means of simulating the characteristic behavior of a PV array. The system provides a turn-key approach to testing the maximum peak power tracking (MPPT) characteristics for grid-tied inverters and DC charge controllers. The ability to simulate virtually any fill factor or solar cell material allows the customer to validate the MPPT algorithm with a power source. Hardware control is accomplished by an application running on the local controller that communicates directly to the PV simulator using RS422, which operate as a dedicated IV curve generation processor. The local Graphical

User Interface (GUI) is accomplished via another application that provides all of the user controls to the TerraSAS system. Imbedded in the application is the Ethernet (LAN) parser for remote communication and control. All of the

858.458.0223

[email protected]

1kW-1MW

80-1000 V

115 208 400

480

AMETEK

Programmable Power

9250 Brown Deer Road

San Diego, CA 92121-2267

USA

1

Elgar TerraSAS

functions available locally through the controller are also available remotely.

Description

As shown in the rack drawing, the TerraSAS consists of programmable DC power supplies, a rack mounted controller, keyboard and LCD display with control software and GUI interface, output isolation and polarity reversing relays and a unique PV simulation engine that controls the power supply. This combination of hardware allows the TerraSAS to simulate most test protocols or combination of events that a solar installation will be subjected to. Power supplies are available in 1-15KW increments to simulate arrays up to 1MW.

The included software, as displayed below, allows modeling of a PV panel without an extensive knowledge of solar array parameters. The only parameters required for a simulation are the open circuit voltage and short circuit current. The slope of the VI curve can then be modified by the peak power parameters, Vmpp and Impp. Changes to these parameters will allow the shape of the VI curve to be adapted to any fill factor between

0.5 and 1. Once an IV curve has been generated, changes to the irradiation level or temperature can be changed on the fly so that the behavior of a grid tied inverter can be tested under realistic conditions for cloud shadowing and panel temperature rise. Long term weather simulations can be run to determine the amount of energy delivered in a given situation. Inverters can be optimized for real MPP search modes, because shadowing and temperature changes can be simulated realistically.

The PV simulation software allows definition of key parameters like Voc, Isc, Vmpp and Isc at 25

°C and 1000W/m2, so that the resulting VI curve is calculated according to a standard solar cell model.

The PV simulator has the ability to simulate ideal IV curves as well as irregular characteristics for peak power tracking that result when solar panels with different output characteristics are paralleled as shown on the following two graphs below. With the simulator programmed for different values of irradiance or temperature, the characteristic “multiple hump” IV curve will result. By programming the changes in irradiance and temperature in a table, dynamic simulation of compressed time profiles of a 24 hour day can be run in a loop to simulate the day and night periods for extended periods of time.

Dynamic simulation showing changes in Irradiance and Temperature over time

2 www.ProgrammablePower.com

TerraSAS

IV Curve Control Interface : Parameters are programmable through sliding scale or direct input of value

Control Displays

The graphic above shows the GUI interface displays. The entered set of IV curves is displayed as soon as the parameters are entered. The actual measured data is then overlaid on the screen so that the operating point can be viewed in real time. The display times can be set from minutes to days to allow for long term testing.

Programmable Parameters

Set a specified irradiance level

Set a specified temperature value

Set a specified voltage level

Set a specified current level

Set a specified temperature coefficient

Ramp of voltage, temperature or irradiance level over a programmed time interval Readback of voltage, current, irradiance level, and temperature setting Programmable calibration of system

Curve Formula

The PV curves for the simulator are derived from the formula shown below.

Io as a function of Vo:

Io=Isc (1-C1 (exp (V/(C2 x Voc))-1))

C1=(1-(Imp/Isc)) (exp(-Vmp/(C2 x Voc)))

C2=((Vmp/Voc)-1)/(ln(1-Imp/Isc))

Where the Reference Irradiance conditions for the simulated arrays is 1000W/m2 and the Reference

Array Temperature is 25°C

The simulated PV arrays are provided in terms of array fill factor, Maximum Power Point Voltage and Maximum Power Point Power. The curves generated are based on the Sandia Labs simplified

PV Array model defining the relationship between these values and other parameters as provided below:

Where:

858.458.0223

[email protected]

3

TerraSAS - Specifications 1kW-1MW

Where:

β Array temperature Coefficient, %/°C

T= Cell temperature, °C

V= Voltage, V

I= Current, A

FF= Fill Factor

Subscripts:

Ref= Reference (i.e., at reference or rated conditions)

MP= Maximum Power

OC= Open Circuit

SC= Short Circuit

DC Output Connections

The output connections will use finger safe, pressure type connectors or terminal blocks of suitable ampacity on the rear I/O panel depending on output current requirements.

“Multiple Hump” IV Curve

Utilizing data gathered from the Solar Advisor

Model (SAM) data base, the TerraSAS allows the user to model systems made up of two or more subsystems. For example, a PV system that consists of three arrays with different orientations, thus creating a “Multiple Hump” as shown below.

Safety

The system includes a shutdown function that will disable the output with an open interlock contact. In the event of an open interlock, the

PV simulator chassis will program down the DC output and open the output relays, and provides complete qalvamic isolation

The benefit of simulators is simply that they offer the ability to test and invert without reliance on a real array and can simulate PV behaviors that a real array cannot be easily manipulated to do.

This assumes that the PV simulator can behave like a real panel of course.

Characteristic “multiple hump” IV curve results when three PV profiles are added

4 www.ProgrammablePower.com

TerraSAS - Specifications

Specifications

AC Power

DC Output

Programmable Parameters

Accuracy

Programming Interface

AC Input Connections

DC Output Connections

AC Input Voltage: 115V (for DCS) 208VAC three phase Std, 400VAC and 480VAC three phase are optional

(Input current depends on power rating)

Open circuit voltage, Voc: 0 - 1000VDC

Short circuit current, Isc: 0 – 1000A

Maximum output power at MPP: 1MW

(Lower voltage ranges will provide proportionately higher currents) 1-6 channel output, consult factory for additional channels.

Irradiance level: 0 to 2000 W/m²

Temperature: -40 to 90°C

Temperature Coefficient: 0 to -65,000 mV/°C

Simulation Times: 0 to 65,000 seconds

Isolation relay and polarity relay closure

Voltage Readback: 0.2% of rated max voltage

Current Readback: 0.5% of max current

Ethernet with RJ-45 connector / LAN

Finger safe, pressure type connectors three phase AC four wire plus safety ground stud AC input circuit breaker

Finger safe, pressure type connectors positive and negative

Safety

Output Voltage and Current Ranges

The output isolation relay operates as a disconnect relay in the event of a malfunction or an open interlock contact

Power (MPP) 1 kW 5 kW 10 kW 15 kW RMS P-P DC Leakage Current

Voc

80V

600Voc

1000Voc

Isc=12.5

Isc=1.6A

N/A

Isc=83A

Isc=8A

Isc=5A

Isc=167A

Isc=16A

Isc=10A

Isc=250A

Isc=25A

N/A

MMPT

Scalable (MPPT) 1000W to 1.0MW

Response to MPPT

Current Slew Rate

Control Loop Sampling Rate

Up to 120Hz

3msec/A

1usec / 10kHz

Static and Dynamic Programmable PV Array Parameters

Irradiance Level 0-2,000W/m2

Temperature

Voltage Level

-100 to +100*C

0-600/1,000V 80V - Consult factory for other voltages

0-Rated Output (see MPP Chart)

0 to -2% / *C

Up to 4096 data points

20 mV

60 mV

100 mV

350 mV 335mA

<1mA

Current level to rated output current

Voltage Temperature Coefficient

Arbitrary VI Curve

Programmable Setpoints

Voc

Fill Factor

Vmp

Imp

ISC

Over Voltage Protection (OVP)

VI Curve Set Point Accuracy

Voltage

Current

Programming Resolution

Programming Resolution

Voltage / Current

0-Rated output voltage

0.5 to 0.95

0-Voc

0-Isc

0-Rated output current

0.1% to 110% of Voc Max

<0.1%, FS

<0.5%, FS

<0.002% of FS

<0.002% of FS

858.458.0223

[email protected]

5

TerraSAS

VI Curve Readback Accuracy

Voltage

Current

Output Sampling Rate

IV Curve Update Rate

IV Curve Interpolation rate

Stability

CC

Temperature Coefficient

CC

Misc

Simulation PV Array Channels

Preloaded Formula

SAM Database

<0.1%, FS

<0.5%, FS

100usec

1sec

7.8msec

0.05

0.03

1-250

LUFT

Over 100 pre-loaded PV Panels, Series & Parallel capability

Over 100 pre-loaded PV Panels, Series & Parallel capability

1kW-1MW

6 www.ProgrammablePower.com

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement