TREND 300
TRE
USER`S MANUAL
ND
300
Serial Number
Each HOFFRICHTER device is provided with a serial number which guarantees the product can be traced back.
The serial number is on the type identification plate to
be found on the reverse of the device.
In case of any query or complaint, please always state
the serial number.
Conformity
0123
Respiration Therapy Device
2
The device complies with the
requirements of the directives
93/42/CEE.
TREND 300
Contents
Legend of Symbols
04
Safety Instructions
05
Intended Use
08
Technology of TREND 300
09
Hints to be observed before starting
10
Description of the Device
11
Proper Use of the Device
14
Use of Oxygen
16
Cleaning Instruction, Changing of Filter
17
Alarm Functions of TREND 300
18
Functions of TREND 300
19
Trouble Shooting
23
Display Indications
26
Inspection and Check
27
Approved Accessories
28
Technical Data
29
TREND 300 - Patient Pass
30
Respiration Therapy Device
TREND 300
3
Legend of Symbols
Information on the Type Identification Plate:
Pay attention to user´s manual
Applied part type BF
Protection class II (insulation - enclosed)
CE - Conformity Statement
Do not dispose of the device with normal
domestic waste.
For information on the proper disposal of the
device, pleasa contact your responsible
customer service.
Respiration Therapy Device
4
TREND 300
Safety Instructions
Before first use of your CPAP-device, please read this
manual thoroughly.
Keep the manual in a safe place, so that you will be able
to consult it whenever necessary.
Attention!
HOFFRICHTER GmbH shall not be held liable for any
damages or consequential damages, consequences
for safety, reliability and performance of the device, in
case of:
Power Supply
TREND 300 has a switched mode power pack with
a wide power supply range of 100 V~ … 240 V~,
50 ... 60 Hz. This allows connection to power supply
lines all over the world. It should be noted that this unit
must be connected only to a grounded contact socket
outlet.
This device is a product of protection class II (protective
insulation). The grounding wire of the power cord is
internal used as functional earth.
DC operation is also possible with 24 V DC, e.g. for
mobile use in a truck or caravan. Connect the unit to a
DC power source only with the connecting cable supplied.
■ Tampering, modifying, adding expansion features,
adjustment, repair and maintenance executed by
persons who have not been authorized by us.
■ Building in accessory and spare parts which have
been fabricated by other manufactures and not
been released by us for use in the respiration
therapy device TREND 300 .
■ Using the therapy device in a different way than
what has been described in the manual.
Respiration Therapy Device
TREND 300
5
Safety Instructions
Please observe the following points in order to avoid
dangerous situations which may occur:
1. Use the device exclusively for your own respiration
therapy and only according to what has been
directed by your physician.
2. Only use accessory recommended and offered
by your authorized dealer.
3. Check the alarm functions of the device at
regular intervals (see page 18).
4. Make sure the small opening in the mask or in the
connecting piece between mask and tube is
always free so that the exhaled air containing
CO2 can escape.
5. Never place TREND 300 in a cupboard shelf or
under the bed. Make sure the air circulation
around the device is not hindered.
6. Keep the unit away from heaters and heated
surefaces.
Respiration Therapy Device
6
TREND 300
7. In case oxygen is given to you in connection
with the respiration therapy make sure that all
regulations in view of general and fire safety during
dealing with oxygen will be observed.
8. During mobile use, the device is only to
switched on in the parked car.
9. You should immediately consult your physician if
due to use of the device problems such as dryness
of the mucous membranes in nose and pharynx,
trouble of the frontal sinus, earache, a runny nose,
hypersensitiviy of the skin, irritability, change of
mood, disturbance of the sense of direction or
memory defects occur.
10. You are requested to contact your authorized dealer
immediately, if the device does not function in a
proper way.
Safety Instructions
In order to maintain electric safety please observe the
following points:
■ Do not switch on the device if casing or cable
are damaged.
■ Do not open the casing of TREND 300. In case of
failures always contact your authorized dealer.
Stand-by-Operation
When the turbine is switched off the unit is in stand-by
mode and this is indicated by the illuminated display
with date and time.
When an air humidifier is connected to the unit, this is
in standby mode; the present heater setting is indicated.
The power consumption in this operating mode is
approx. 3,5 Watts.
■ The device is always to be kept dry.
■ Never use the device in damp rooms
or in a bathroom.
■ Never place vessels or glasses
filled with liquids on the device.
■ In case the device falls into water, do not touch it.
■ Do not place the device in the near of
water-containing vessels (or a bath-tub).
■ Before cleaning the device the AC power plug
is always to be pulled out of the socket.
By switching off the device with the main switch on the
reverse of the device TREND 300 is totally switched
off.
In this case the internal clock is supplied by
a built-in battery.
■ TREND 300 can safely remain
in standby operation all the time.
NOTE:
The memory of the digital trigger may be inaccurate if
the mask is put on and taken off repeatedly while the
fan is running. If the digital trigger does not change the
pressure correctly there after, it is recommended to set
back the trigger computer memory. Turning the ventilator off and on once is sufficient.
Respiration Therapy Device
TREND 300
7
Intended Use
Intended Use
Contraindications
The TREND 300 is for treatment of nighttime sleep
apnea in patient weighing at least 30 kg.
The respiration therapy may be contraindicated for patients with the following pre-existing medical conditions:
This device is not intended for use with ventilatordependent patients.
■
■
■
■
The TREND 300 provides positive airway pressure in
two pressure levels.
This device is designed to be used with nasal breathing
masks equipped with an exhalation port to allow a continuous flow of air through the mask. Do not wear the
mask with the device turned off as rebreating of the air
occurs. Refer to the manufacturer´s instructions for the
mask and exhalation port.
Devices equipped with an AquaTREND accessory
humidifier provide humidification of the air delivered to
the patient.
Respiration Therapy Device
8
TREND 300
Bullous lung diseases
Pneumothorax
Extremely low blood pressure
Pneumocephalus or pre-existing CSF leaks or head
trauma.
Acute sinus or middle ear infection may be an indication
to suspend the therapy. Contact your physician if you
have questions about this condition.
Devices equipped with an AquaTREND accessory
humidifier are not intended for use with medications,
aromatic oils or other substances added to the humidifier water.
Not for use with patients whose supraglottic airways
have been bypassed.
Technology of TREND 300
TREND 300 is a respiration BILEVEL therapy unit for
treatment of nighttime sleep apnea. An air humidifier
can be obtained as an accessory.
For the pressure generation TREND 300 has an
electronically controlled blower. In order to keep
the patient's burdening as low as possible the
turbine has been equipped with sufficient power
reserves as well as a fast control circuit.
The pressure control is applied directly in
the breathing-mask. So the breathing musculature
needs not balance out the flow losses of the
breathing air in tube and humidifier.
That means the breathing work is made as
easy as possible.
NOTE:
No negative pressure is been generated during
exspiration phase.
Trigger device
The respiratory trigger recognises the patients efforts to
breathe in and out and transmits them to a control
mechanism.
The trigger mechanism of the TREND 300 has its own
microcontroller to optimise the release of the trigger by
dynamic adjustment of the release criteria to various
sleeping phases.
In conjunction with optimal flank adjustment the patient
experiences a pleasant ventilation which gives him or
her the impression that the unit pursues the natural
breathing mechanisms without any inertia.
TREND 300 has separate air ductings for breathing
air and convection air for the electronics.
The cold temperature technology applied in the power
electronics ensures a low temperature and a long life of
the device.
For the control of all functions TREND 300 has a
built-in microcomputer. The integrated quartz
alarm clock increases the comfort during use
of the device.
The air heats up marginally on passing through the unit
and has a higher water uptake capacity. Particularly in
winter when the room air is dry the mucous membranes
in the mouth and nose can dry out. This is unpleasent
and in isolated cases also leads to infections. Air humidification may therefore also be necessary in conjunction
with respiration therapy.
AquaTREND accessory devices are available which
provide humidification. The humidifier simply clicks into
position and can be removed by pressing the release
button.
Refer to the AquaTREND user´s manual or contact your
dealer for more information.
Respiration Therapy Device
TREND 300
9
Hints to be observed before starting
Before operating the device, please pay attention to the
following:
Please pay attention to the following
advices for setting up:
1. Never switch on the device if it has recently
been in a very cold environment. Wait
approximately one hour for temperature
compensation.
1. Never place the device near a heater
or radiator. Make sure the air circulation especially
under and behind the device is not hindered by
furniture, curtains or pillows.
2. Check the filter cassette at regular intervals and
change it regular (see pages 17, 26). Never use
the device without filter cassette.
2. Place the device beside your bed on a firm and even
surface such as a nightstand or bedside table.
3. Clean the mask system regularly, and check the
accessory especially the therapeutic tube, mask
and fixing straps. We recommend to wash the
face before going to bed as skin oils and
cosmetic should be removed.
Thus, the life-span of the mask will be increased
and skin irritations can be avoided.
4. Verify that the measuring tube (smaller tube) is
correctly attached, open and free of moisture or
obstructions. If water enters this tube during
cleaning, take special care to remove it.
Respiration Therapy Device
10
TREND 300
3. During mobile operation the device is also to
be placed on an even and firm surface which
should have a border preventing from falling down.
Front View
Description of the Device
Release Button
Air Outlet
Control Key
(Top) for accessories
ON/OFF
Display
HOFFRICHTER
TREND 300
Input Key
Connectors for
AquaTREND (accessory)
Measuring Tube
Connection
Program Keys
Respiration Therapy Device
TREND 300
11
Back of the Device
Description of the Device
Air Input
Filter Cassette (Top)
TREND 300
DC IN 24 V
Fernbedienung
Remote Control
AC Power Plug
AC Power Switch
Type Identification Plate
Respiration Therapy Device
12
TREND 300
DC-Input
RS232-Interface
Accessory and Spare Parts
Description of the Device
Cable for
Motor Car
Measuring Tube
Filter Cassette
Stoppers
Carrier Bag
Power Supply Cable
Exhalation Port
(additional accessory)
Breathing-Mask
(additional accessory)
Therapy Tube
(1,80 m)
Respiration Therapy Device
TREND 300
13
Proper Use of the Device
1. As soon as the device is placed properly
please connect it with the power supply cable
to an earthing contact socket.
2
3a
2. Switch on the AC power switch on the back of the
device. The device welcomes you with a little text
in the display. Afterwards the display switches
over to indication of current date and time.
ON/OF
F
TREND 300
HOFFRICH
DC IN 24 V
TER
TREND
300
Fernbedienung
Remote Control
3. Connect the therapeutic tube
a) to the air outlet and
b) to the breathing-mask.
The device must be used with an exhalation port.
4. Connect the measuring tube
to the sensor socket (measuring tube connection).
5. Put on the breathing-mask. If the automatic
operating mode has been selected the device
will start working upon the first breath. In case
of manual operating mode you must press
the control key (ON/OFF) for starting the device.
6. Now first TREND 300 starts with the prescribed
pressure for the certain period of time you have
selected for mask test. During this time you
should fit the mask correctly in order to eliminate
possible leakages.
Respiration Therapy Device
14
TREND 300
3b
4
ON/OFF
HO FF RI
CH TE R
TREND
300
Proper Use of the Device
7.Place the therapy tube in such a way that
there is no tension onto the mask when you
lay down.
8.Breathe deeply and calmly and only through
the nose. If the softstart-function is activated
the device will after leakage test automatically
reduce the therapeutic pressure first and then
gradually increase it until the value prescribed is
reached. This enables you to fall asleep while
the device is working with a reduced pressure.
ON/OFF
TREND 300
NOTE:
Read in the section "Programming" how the time
for softstart and mask test can be adjusted to your
individual requirements.
Respiration Therapy Device
TREND 300
15
Use of Oxygen
Please observe the following advices if you
must additionally inhale oxygen during the
respiration-therapy:
If you want to lead the oxygen directly into the
breathing-mask please use a non-buckling tube
made from material approved for medical applications.
Consult your authorized dealer.
NOTE:
At a fixes flow rate of supplemental oxygen flow, the
inhaled oxygen concentration will vary, depending on the
pressure settings.
Refer to the oxygen suppliers instruction for use.
WARNING!
Oxygen promotes combustion processes.
Observe the valid rules for fire safety during use
of oxygen, to reduce the risk of fire. Keep the oxygen
instruments free from oils. All sources of ignition including smoking or smoking materials must be kept
away from the device and preferably out of the room.
Respiration Therapy Device
16
TREND 300
WARNING!
If oxygen is used with this therapy device, the oxygen
flow must be turned off when the therapy device is
not operating.
■ For therapy first switch on turbine and then switch
on oxygene.
■ For ending therapy first turn off oxygene then turn off
turbine. Please program "Automatic OFF" as
described on pages 19 - 22.
■ Before each use, check the proper fit of
the tube connections.
Cleaning Instructions, Changing of Filter
Daily Cleaning of the Breathing-Mask
■ Disconnect the mask and exhalation port from the
therapy tube.
■ Clean the breathing mask in accordance with the
instructions of the manufacturer of the mask.
Weekly cleaning of the device
■ Unplug the AC power cord or switch off the main
switch.
■ Clean the device with a cloth moistened with
soap water. Afterwards wipe the device dry.
Weekly Cleaning of the Therapy Tube
■ Disconnect the tubing from the TREND 300 and the
mask and exhalation port.
■ Before cleaning:
The measuring tube has to be closed tightly
with the stoppers.
■ Use mild soap water and rinse the measuring tube
very carefully with clear water.
Air-dry the tube very well.
■ After cleaning:
Remove the stoppers.
Cleaning of the fixing straps (if necessary)
■ Separate the fixing straps from the breathing-mask.
■ Clean the fixing straps in accordance with the
instructions of the manufacturer of the straps.
A chemical or thermic desinfection of the device or parts
of the device by the user is not necessary.
Before change of patient the hygienical reprocessing of
the device has to be realised by the authorised dealer or
manufacturer.
Changing the filter
■ If the filter is dirty or the display reads > FI LTER <, the
filter cassette must be changed (see pages 26 and
27). Press on the underside of the unit on the filter
cassette protrusion and remove the cassette from
the top. (see picture)
■ To reset the message please keep the input key
pressed while switching on the device.
TREND 300
V
DC IN 24
Fernbedienung
Remote Control
Changing the filter cassette
Respiration Therapy Device
TREND 300
17
Alarm Functions of TREND 300
Power Failure
Wake-up Alarm
Probably you won't notice a power failure at night-time.
As a consequence you would permanently inhale the
exhausted air from the therapy tube. Therefore TREND
300 has an alarm mechanism responding if a power failure occurs at night-time or if a fuse becomes defective.
In this case an acoustic alarm would wake you up
so that you can take off the mask and breathe fresh
air. The alarm can be switched off by pressing the
control key (ON/OFF).
After the end of power failure the turbine of TREND
300 starts again automatically.
For higher comfort TREND 300 has an integrated
alarm-clock. It can be activated and deactivated with
the program keys.
The alarm-clock raises alarm at the pre-set wake time.
If you press the control key once the alarm will stop
for the next five minutes and if you press it twice
the alarm will be totally switched off.
Mask Alarm
This alarm function is active only if the automatic
function is switched off. If at night-time the mask
slips down or the therapeutic tube is pulled off the
device won't switch off automatically but raise an
acoustic alarm. This additional safety function,
however, requires that the device is switched on and
off with the control key (ON/OFF) each morning and
evening.
Respiration Therapy Device
18
TREND 300
Check of the Power Failure Alarm:
Use the device with running turbine and pull out the AC
power plug or switch off the AC power switch. Now the
acoustic alarm should be occur. Check every month if
the acoustic signal lasts long enough (guide value 1 min).
After pressing the control key or switching on the power
supply the acoustic signal will stop.
Check of the Mask Alarm:
Use the device with running turbine and switch off
automatic function. Do not fix the mask to the face.
After 10 seconds an acoustic signal will occur.
Functions of TREND 300
For programming the functions of TREND 300
three keys are available:
= Input key
= Program key
= Program key
By holding the INPUT key pressed, you reach
the input mode after 1 second, and then
you see the menu of TREND 300 in the display.
The following 2 different kinds of menu can
be activated:
Standard-Menu
Hints for (1) to (4):
The display first indicates the menu line (1). After adjustment of I-Pressure the lines (2) to (4) will be indicated step
by step and can be set by the doctor.
Complete-Menu
WAKE TIME ........04:45
WAKE TIME .......04:45
TIME ........................ 16:24
TIME ..........................16:24
DATE .................
DATE ....................12.11.02
12.11.02
EXIT MENU
Example: Setting alarm clock
Press key for 1 second. Display indicates:
RAMP................0s/0min
WAKE TIME.......... 04:45
AUTOMATIC...........OFF
(1)
NOTE:
If you prefer instead of
the standard-menu the
complete-menu (or
reverse) contact your
dealer please.
The display always shows two lines. With the keys
and you can choose which lines are to be indicated. A triangle symbol in front of a line means that this
line has been chosen.
I-SET.......... ......12,0hPa
EXIT MENU!
(PIN-protected)
(2) E-SET. . . . . . . 8,0hPa
(PIN-protected)
(3) I-SLOPE.......... 0,5s
(PIN-protected)
(4) E-SLOPE........0,5s
(PIN-protected)
TRIGGER ................ 3
MODE............BILEV-S
BRIGHTNESS ...100%
After pressing key again the hour of the alarm time is
flashing and can be changed by the keys and .
Press key again for changing alarm minute according
to same procedure. To come back to the menu press
the key again. For leaving the menu select point
" E X I T M E N U !" :
OPERATION......01937h
EXIT MENU!
Respiration Therapy Device
TREND 300
19
Functions of TREND 300
WAKE TIME.......... 06:00
All functions are programmed in the same way:
EXIT MENU!
Press for 1 s, and choose the line requested with the
keys and :
Press Input key to leave the programming mode.
During standard operation the display indicates
date and time:
SA 28.FEB
12:42
The alarm clock can be activated or deactivated by the
keys and .
With the key the alarm clock can be switched on. For
controlling the programmed alarm time is indicated for a short
time in the display. After this a bell symbol shows that
the alarm clock is active.
SA 28.FEB 12:42
With the key the alarm clock can be switched off (e.g. on
Sunday). In this case the bell symbol is not indicated.
NOTE:
If there is no key pressed during programming for
30 seconds the programming mode will be left
automatically.
Respiration Therapy Device
20
TREND 300
By pressing the key the following options can be
chosen:
WAKE TIME
WAKE HOUR, WAKE MINUTE
TIME
HOUR, MINUTE
DATE
YEAR, MONTH, DAY, WEEKDAY
RAMP
1.
mask test time:
2. softstart time:
AUTOMATIC
ON, OFF
I-SET
1.
PIN INPUT!
0...90 s
(in 5 s-steps)
0 . . . 6 0 m i n (in 1 min-steps)
following selection after:
2 . 3 , 0 . . . 1 8 , 0 h P a (in 0,5 hPa-steps)
E-SET
3 , 0 . . . 1 8 , 0 h P a (in 0,5 hPa-steps)
I-SLOPE
0,0 ... 1,5 s
(in 0,1 s-steps)
E-SLOPE
0,0 ... 1,5 s
(in 0,1 s-steps)
TRIGGER
MODE
1 ... 5
BILEV-S
BRIGHTNESS
100%, 50%
COUNTER
POWER, BLOWER, FILTER, THERAPY (time indicate in h)
(only indicate, not selection possibility)
EXIT MENU!
If you want to finish the programming, choose the menu
line “EXIT MENU!” and press key .
Functions of TREND 300
Date, Time
Date and time of the internal clock can be adjusted. The
internal clock has a built-in battery with reserve power
for 8 years without mains connection. You can adjust
the weekdays (Mo thru Su), day (1 till 31), month (1 till
12), year (00 till 99), hour (00 till 24) and minute (00 till
59). The clock serves for indicating the right date, for
the right function of the alarm-clock and for the exact
recording of current events.
Ramp
After each turbine start the computer of TREND 300
checks whether a softstart-ramp has been programmed. During the time which you have preset the softstart function slowly raises the pressure from 3 hPa
(mbar) up to the pressure prescribed by your physician.
This should ease your falling asleep. Before starting softstart TREND 300 checks the proper fitting of the mask
using the full pressure for the selected time.
Especially if you feel you have not yet got used to the
respiration-therapy the softstart can give you relief.
The delay time of TREND 300 can be adjusted
between zero (in this case the therapeutic pressure
prescribed is instantly given) and 60 minutes (the
therapeutic pressure prescribed is built-up within 60
min).
be programmed. This mask test runs under full therapy
pressure before starting ramp and checks whether there
is a leakage.
pressure
in hPa
10
pressure
prescribed:
10 hPa (mbar)
The pressure is gradually
increased up to the
therapeutic pressure
mask test
Softstart time programmed in min
0-90 s
3
time in min
0
15
30
45
60
To ensure that there is no mask leakage even during the
high therapy pressure a mask test time up to 90 s can
Respiration Therapy Device
TREND 300
21
Functions of TREND 300
Automatic Operation
In case of automatic operation you have only to put on
the mask. As soon as you breathe the turbine of
TREND 300 will be switched on.
If you take off the mask the turbine will be switched off
automatically after a delay time of 5 seconds. During
automatic operation the device also will be switched off
if mask or tube drop off unnoticed at night-time.
That is why TREND 300 optionally offers the function
"Automatic OFF" with mask-alarm.
If you choose this function you must press the CONTROL key (ON/OFF) for switching on the device each
evening and switching off each morning. In this case
you get an acoustic alarm if the mask drops off at nighttime.
Pressure (I-Set, E-Set, I-Slope, E-Slope)
The adjustment of the desired pressures and slopes is
done by the physician who has a special remote control.
Without remote control, however, it is also possible to
change the pressure adjustment of the device provided
that a respective PIN has been programmed. The PIN
will enable the service company to do the respective
work if necessary.
Please, never try to "crack" the PIN.
You should better talk to your physician in case of any
doubt about the pressure prescribed.
Respiration Therapy Device
22
TREND 300
Trigger
The trigger sensitivity is adjustable in 5 steps.
Example:
TRIGGER .... 1 (lowest trigger threshold
= highest trigger sensitivity)
TRIGGER .... 3 (medium trigger threshold
* factory adjustment)
TRIGGER .... 5 (highest trigger threshold
= lowest trigger sensitivity)
NOTE:
The adjustment of the trigger sensitivity should ever be
made by your physician.
Brightness
You can adjust the brightness of the display to suit your
needs by selecting between two stages:
■ 100% = maximum brightness, the display is easily
legible also in brightly lit rooms.
■ 50% = minimum brightness, the display is easily
legible in slightly darkened rooms.
In the event of an alarm, the display always switches automatically to 100% brightness for the duration of the alarm.
Power
Under this point you find the running time of the turbine,
the operation time of the filter, the total operation time
including stand-by operation and the therapy time of
TREND 300.
Trouble Shooting
PROBLEM
POSSIBLE CAUSE
HELP
Pain in nose, nasal
cavity or ears
Inflammation
Stop therapy and contact
your doctor
Unpleasant feeling of too
high pressure
Problem with
prescribed high pressure
Contact your doctor
You are not yet used to the
respiration-therapy
Relax and use softstart function
Dryness and irritation
of nose and throat
Dry treatment air
Consult your doctor about prescribing the
addition of AquaTREND humidifier
Symptoms of sleep apneas
appear again
Your personal life
conditions have changed
Contact your doctor
Devise is not operating properly
Check device by an
authorized company
Respiration Therapy Device
TREND 300
23
Trouble Shooting
PROBLEM
POSSIBLE CAUSE
HELP
Irriteted or dry eyes
Leakage between mask
and skin
Check correct position
of mask
New mask if defective
Wet nose
Reaction to the air flow
Contact your doctor about use of a humidifier
Head cold, sinus drainage or
discomfort
Normal cold or sinus infection
Stop use if your condition interferes with
treatment, contact your doctor
Cold nose
Low room temperature
Increase room temperature
Warm up tube under your
pillow
Inflammation of the
skin under the mask
Wrong mask size
Contact your doctor
Fixing straps too short
Relax the straps
Allergic reaction
Contact your doctor
Respiration Therapy Device
24
TREND 300
Trouble Shooting
PROBLEM
POSSIBLE CAUSE
Air too warm
Dirty filter
Exchange filter
Air input is blocked
Make sure air input is clear
Heating too close to
TREND 300
Check position of TREND 300
Device is defective
Ask your authorized dealer
Water in measuring tube
Make sure air input is clear
Softstart function is active
Decrease softstart time
Air input is blocked
Make sure air input is clear
Measuring tube is
disconnected or blocked
Check measuring tube
Leakage in device
Ask your authorized dealer
Water in measuring tube
Dry measuring tube
Water in measuring tube
Dry measuring tube
Device is defective
Contact service company
No air flow
Low air flow
Turbine works always
with maximum speed
Therapy pressure is not constant
HELP
Respiration Therapy Device
TREND 300
25
Display Indications
In the following table the most important display indications are shown. The sense off all other indications can be taken from context.
MASK TEST. . .55s
You still have 55 s for fitting the mask correctly.
After that time the softstart function starts.
SOFTSTART 10:33
You still have 10 min and 33 s until the full therapeutic pressure is reached.
CHECK BREATHING
MASK
The mask has dropped down from your face or you haven't put it on during "Automatic OFF"
and running turbine.
ALARM-CLOCK ON
In normal operation you have pressed the key and got the message that you will be
woken up at a quater to six o'clock.
ALARM-CLOCK OFF
In normal operation you have pressed the key and got the message that the alarm-clock
has been switched off right now.
WAKE ALARM
Wake alarm has been triggered. Press the control key once in order to start the slumber.
If you press the main key for the second time during the slumber phase the alarm is finally switched off.
SLUMBER. . . 1:18
You still have 1 min and 18 s until the alarm-clock is ringing again.
WAKE ALARM OFF
You have pressed the control key and finished the slumber phase.
> FILTER <
You are required to exchange the filter cassette (see page 17). If switching on the device after filter exchange please
keep the input key pressed in order to reset the message.
> SERVICE <
TREND 300 needs a security check. Please inform your after-sales service as only the
authorized service people can reset this message. The device can be used furtheron.
PUT ON MASK!
You are required to put on the mask and to breathe through the nose.
TAKE OFF MASK!
During activated automatic one has tried to switch off the device by hand. Please take off the mask.
The device switches off the therapeutic pressure after a few seconds.
BAT: 24,6 V
Voltage on board of a vehicle (if battery operation)
POWER FAILURE
After the end of power failure the turbine of TREND 300 restarts automatically.
RESTART TURBINE
Respiration Therapy Device
26
TREND 300
Inspection and Check
Although all parts of TREND 300 are proved to have a
long life the device should be examined from time to
time. A device check is necessary if the message
" S E R V I C E !" appears in the display. In this case we
urgently recommend you to have the device examined
and maintained by the authorized after-sales service
because only these people have the necessary tools
and measuring devices. Also in case of the " S E R V I C E !"
display, the device can be used furtheron.
Please pay attention to the following
check-list:
■ weekly check of the alarm function
"Power failure" (see page 18)
■ weekly check of "Mask Alarm"
(see page 18)
■ check of the filter cassette and if necessary
replacement by a new one
Attention!
The display message " S E R V I C E !" cannot be
removed if the maintenance is executed by
an unauthorized workshop.
For a long lifetime of your TREND 300 device
note the following points:
1. Please follow the cleaning instruction on the
page 17.
2. Check the system at regular intervals.
Respiration Therapy Device
TREND 300
27
Approved Accessories
Specification
HOFFRICHTER Order-No.
Filter Cassette
AC Power Supply Cable
Therapy Tube
Stoppers
00007870
31100001
00007116
44000012
(for closing measuring tube during tube cleaning)
Carrier Bag
User´s Manual
Cable for motor car
00007860
00007877
00009212
Plug-In-Humidifier
AquaTREND III-H0
AquaTREND III-H1
AquaTREND III-H5
00007992
00007993
00007990
NOTE:
The TREND 300 should be used only with therapy
tubes provided by the manufactuer.
Respiration Therapy Device
28
TREND 300
Concerning available approved therapy masks and
additional accessories please contact your authorized
dealer.
Technical Data
Dimensions (B X T X H)
230 x 225 x 110 mm
Mass
ca. 2,5 kg
Power supply
100 V~ … 240 V~, 50 ... 60 Hz
24 V DC
Pressure range
IPAP, EPAP: 3 … 18 hPa (mbar) in steps of 0,5 hPa
Power consumption without humidifier 22 W (average value)
Power consumption of stand-by
<5W
Pressure variance
< ± 0,08 hPa (mbar) at 10 hPa
according to DIN EN ISO 17510
Inner resistance
Noise level without mask
< 0,12 hPa s/l
< 27 dB(A) at 10 hPa (mbar)
according to DIN EN ISO 17510
Constant display
pressure value, bargraph, date, time, alarm
clock state
Operating Altitude
- 400 m ... 3500 m above sea level
Operating Temperature
5 °C ... 40 °C
Storage Temperature
- 40 °C ... 70 °C
CE-mark according to the EC-directives 93/42/ECC, concerning medical devices, class IIa
Data subject to technical alteration
Respiration Therapy Device
TREND 300
29
TREND 300 - patient pass
name
serial-No.
software-version
first name
type of mask
mask size
street
Therapy process
postal zip code/place
date
telephon
weight
high
birth. on
health insurance company
blood pressure
apnea index
medicine
IPAP-Set
hospital-stamp
EPAP-Set
I-Slope
E-Slope
family doctor-stamp
therapy standing
next service
executed
customer service-stamp
Respiration Therapy Device
30
TREND 300
Notes
Respiration Therapy Device
TREND 300
31
MANUFACTURER:
SALES & SERVICE:
HOFFRICHTER GmbH
Mettenheimer Straße 12/14
19061 Schwerin, Germany
Tel.: +49-385-39925-0
Fax: +49-385-39925-25
Mail: [email protected]
URL: www.hoffrichter.de
HOFFRICHTER type names:
TREND 110 Monolevel
Typ 5CP 102
TREND 210 Monolevel
Typ 5CP 210
TREND 300 Bilevel
Typ 5CP 301
TREND 400 Bilevel ST20
Typ 5CP 401
TREND 500 Bilevel ST33
Typ 5CP 501
AutoTREND Auto CPAP
Typ 5CP 253
AquaTREND III-H0 humidifier
Typ 5CP 931
AquaTREND III-H1 humidifier
Typ 5CP 941
AquaTREND III-H5 humidifier
Typ 5CP 961
TREND 300-eng-1208-06