advertisement
EN Instructions for use for devices of type WM 100 TD prisma20C prisma20A prismaCR prisma25S
Sleep therapy devices
prisma25S-C prisma25ST prismaLAB prisma30ST
Contents
Contents
2
|
EN
4.2 Connecting up the power supply ............................................................ 19
5.2 Switching on the therapy device ............................................................. 23
5.10 Viewing therapy data and device information ......................................... 32
Contents
7.3 Hygienic preparation of the therapy device ............................................. 43
7.4 Hygienic preparation of the respiration hose ........................................... 45
9 Alarms and error messages 47
13.2 Emission of electromagnetic interference ............................................... 61
13.3 Electromagnetic interference immunity .................................................. 61
13.4 Electromagnetic interference immunity for ME equipment and
EN
|
3
4
|
EN
1 Introduction
1 Introduction
1.1 Intended use
The WM 100 TD devices are pressure-controlled, non-invasive, non-life-sustaining therapy devices for the treatment of sleep disordered breathing (SDB) or intermittent treatment of respiratory insufficiency by means of a mask.
The devices can be used on persons weighing above 30 kg. The CPAP mode can be used on persons above the age of 3 years. The device may only be used on the instruction of a physician.
The WM 100 TD devices are used in clinical facilities and in domestic situations. In domestic situations, the devices are also taken on trips.
1.2 Function
The fan in the therapy device sucks ambient air in through a filter, compresses it, and routes it to the device outlet.
From here, the air flows through the hose system and the mask to the patient. The exhalation system in front of the mask, or optionally integrated in the mask, prevents
CO
2
-enriched exhaled air from collecting in the hose system.
The therapy device determines and analyzes the pressure and respiratory flow signal.
This allows respiratory events to be recognized.
The device can function with one pressure level (CPAP) or with two or three pressure levels (BiLevel or inspiratory pressure, expiratory pressure, and end-expiratory pressure). Depending on the version employed, the pressure levels can be set automatically by the device within preset limits, or they can be set manually.
Depending on the mode, the pressure can be continually applied at one level, or triggered by the patient, or applied with time controls. Pressure signals, respiratory flow signals, and respiratory events can be saved and/or can be output in analog form on a PSG system.
The therapy data are saved in the device and on an SD card for therapy monitoring.
The device is operated via an On/Off button and a touchscreen.
The device can be remotely controlled using the prismaTS therapy software.
In the case of a power failure, the settings are retained and the therapy is continued once the power supply is restored.
1 Introduction
1.3 User qualifications
The person operating the device is referred to in these instructions for use as the
"user". In contrast, a "patient" is the person receiving the therapy. Always perform all the operating steps in accordance with these instructions for use.
1.4 Indications prisma20C
CPAP therapy device for the treatment of patients with obstructive sleep apnea with a constant pressure requirement.
prisma20A
APAP therapy device for the treatment of patients with obstructive sleep apnea with a variable pressure requirement. The therapy pressure adjusts automatically to suit the patient's pressure requirement.
prismaCR
Therapy device for the treatment of patients with periodic breathing or Cheyne-Stokes respiration (e.g., in cases of heart failure) as well as with central, mixed, or complex sleep apnea. The therapy device adjusts the ventilation automatically and continually to the changing requirements of the patient.
prisma25S
BiLevel therapy device for the treatment of patients with obstructive, mixed, or complex sleep apnea, and a high and/or fluctuating pressure requirement, poor CPAP compliance The device has different pressure levels during inspiration and expiration.
prisma25S-C
BiLevel therapy device for the treatment of patients with obstructive, mixed, or complex sleep apnea and a high pressure requirement, poor CPAP compliance The device has different pressure levels during inspiration and expiration.
prisma25ST
BiLevel therapy device for the treatment of patients with obstructive, mixed, or complex sleep apnea and a high and/or fluctuating pressure requirement, poor CPAP compliance, central apneas, sleep-related or position-dependent hypoventilation (e.g.,
OHS), respiratory insufficiency, coprevalent respiratory insufficiency (e.g., COPD/ overlap). The device has different pressure levels during inspiration and expiration and a backup frequency for the treatment of central events.
EN
|
5
6
|
EN
1 Introduction
prisma30ST
BiLevel therapy device for the treatment of patients with obstructive, mixed, or complex sleep apnea and/or chronically reduced respiratory drive (e.g., sleep-related or position-dependent hypoventilation or chronically stable OHS), respiratory insufficiency, e.g., COPD.
1.5 Contraindications
The following contraindications are known – the physician in charge is responsible for deciding whether to use the therapy device in each individual case.
Acute cardiac decompensation, severe arrhythmia, severe hypotension, particularly in combination with intravascular volume depletion, severe epistaxis, high risk of a barotrauma, severe chronic/decompensated pulmonary conditions, pneumothorax or pneumomediastinum, pneumocephalus, cranial trauma, status following brain surgery or surgical intervention on the pituitary gland or the middle/inner ear, acute sinus infection (sinusitis), middle ear infection (otitis media) or perforated eardrum, dehydration.
prismaCR
Symptomatic chronic systolic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea (AHI 15/h, CAHI/AHI 50% and CAI 10/h)
1.6 Side effects
The following undesirable side effects may occur when using the therapy device for short or long periods of time: pressure marks from the respiratory mask and the forehead cushion on the face, flush of the facial skin, nasal congestion, dry nose, morning xerostomia (dry mouth), sensation of pressure in the sinuses, irritated conjunctiva, gastrointestinal air insufflation ("bloating"), epistaxis.
These side effects are general side effects associated with therapy using a sleep therapy device and are not specially linked to the use of WM 100 TD devices.
2 Safety
2 Safety
Please read these instructions for use carefully. They form part of the devices described, and must be available at all times.
Use the unit for the designated purpose only
(see "1.1 Intended use", page 4) .
For your own safety and that of your patients, and in accordance with the requirements of Directive 93/42/EEC, please observe the following safety instructions.
2.1 Safety information
2.1.1 Safe use of the therapy device, components, and accessories
Risk of injury due to device or component malfunction!
A damaged device or damaged components may result in injury to the patient, user or bystanders.
⇒ Only operate the device and components if they are externally undamaged.
⇒ Only operate the device and components if the function check has been successfully completed.
⇒ Only operate the device if the display is functional.
Risk of injury if the device is operated outside the prescribed ambient conditions!
Use of the device outside the prescribed ambient conditions can result in failure to comply with tolerances, device failures, and injury to the patient.
⇒ Only operate the device within the prescribed ambient conditions
(see "13.1 Technical Data", page 52) .
Risk of injury if disposable items are reused!
Disposable items are only intended to be used once. Reused disposable items may be contaminated and/or not function correctly and thus cause patient injury.
⇒ Do not reuse disposable items.
Risk of infection when reusing therapy device!
When the therapy device is used by multiple patients, infections may be passed on to the next patient.
⇒ Use a bacteria filter
⇒ When the device is used without a bacteria filter: Have the device hygienically prepared by the manufacturer or an authorized dealer.
EN
|
7
8
|
EN
2 Safety
Treatment prevented due to increased resistance when bacteria filters are used!
Misting or moistening can increase the resistance of the bacteria filters, thereby modifying the output of the therapy pressure.
⇒ Check bacteria filters regularly for increased resistance and blockages and rectify these.
2.1.2 Power supply
Risk of injury due to inaccessible power plug!
An obstructed power plug cannot be pulled out in an emergency and can thus result in injury.
⇒ Keep the power plug and power supply accessible at all times.
Risk of injury and material damage as a result of insufficient power supply!
Operation of the device outside the specified power supply range can injure the user and damage the device.
⇒ Only operate the device with the supplied power supply unit at voltages from
100 V to 240 V.
⇒ Use the DC adapter for operation at voltages from 12 V or 24 V.
2.1.3 Transport
Water in the device can cause material damage!
If the device is tilted severely, the residual water from the respiratory air humidifier can enter the device and damage it.
⇒ Do not transport or tilt the device when the respiratory air humidifier is filled.
Dirt in the device can cause material damage!
Dirt entering the device during transport can damage the device.
⇒ Only transport the device with the cover in position.
⇒ Transport the device in the corresponding transport bag.
2 Safety
2.1.4 Therapy
The use of oxygen in combination with flammable substances poses a fire hazard!
Oxygen in combination with flammable substances can result in spontaneous explosions. In cases of insufficient ventilation, oxygen in the surrounding area (e.g., clothes, hair, bedclothes) can become enriched and cause fires and thus injuries to the patient, user, and others in the immediate vicinity.
⇒ Do not smoke.
⇒ Do not use naked flames.
⇒ Ensure sufficient ventilation.
⇒ Use an oxygen safety valve.
⇒ Keep the device and screwed unions free from oil and grease.
⇒ Always replace splashguards immediately after use.
The use of oxygen in combination with flammable substances poses a fire hazard!
Supplying oxygen without special safety equipment can cause fires and injure people.
⇒ Always use an oxygen safety valve.
⇒ Observe the instructions for use for the oxygen safety valve and the oxygen supply unit.
⇒ Set up oxygen sources more than 1 m from the device.
Prevented therapy and material damage due to dirt in the device or respiratory air humidifier!
Dirt entering the device can impair the success of the therapy and damage the device.
⇒ Use the gray air filter.
⇒ If necessary, use the white pollen filter (optional accessory).
Risk of injury if the patient connection opening becomes hot when uses a hose heating system!
In combination with the device, the hose heating system generates a somewhat higher temperature at the patient connection opening.
⇒ Observe the instructions for use for the hose heating system.
EN
|
9
10
|
EN
2 Safety
2.2 General information
• Use of third-party products may lead to functional failures and restricted usability.
Biocompatibility may also be compromised. Please note that in these cases, any claim under warranty and liability will be void if neither the accessories nor original spare parts recommended in the instructions for use are used.
• Repairs, servicing, and maintenance should only be carried out by the manufacturer or by a technician expressly authorized by the manufacturer.
• Only connect up the devices and modules permitted in accordance with these instructions for use. The devices must satisfy their respective product standard.
Position non-medical devices outside of the patient's immediate vicinity.
• The device is subject to special precautions with regard to EMC (electromagnetic compatibility). Maintain a minimum distance of 30 cm between the device and equipment that emits HF radiation (e.g. cell phones). This also applies to accessories such as antenna cables and external antennas, for example. Ignoring this requirement may lead to the device exhibiting reduced performance characteristics.
• Do not operate the device outside the EMC environment specified for this device (see
"1.1 Intended use", page 4) in order to prevent undesired events for the patient or
operator due to electromagnetic interference. Do not operate the device if the housing, cables or other equipment for electromagnetic shielding are damaged.
• Do not operate the device in the immediate vicinity of other devices or in a stacked arrangement, otherwise there may be malfunctions. If it is necessary to operate the device in the immediate vicinity of other devices or in a stacked arrangement, keep all the devices under observation to ensure that they are all operating properly.
• Only operate device within the specified ambient conditions
.
• The operator is responsible for ensuring the compatibility of the therapy device and all the connected components and accessories prior to the application with the patient.
• Only use accessory parts from the manufacturer. Third-party electrical connecting cables, in particular, may cause the device to malfunction.
• Only have modifications to the unit carried out by the manufacturer or by a technician expressly authorized by the manufacturer.
• Please observe the chapter on hygienic preparation in order to avoid infection or bacterial contamination
(see "7 Hygienic preparation", page 42) .
2 Safety
• The operator is responsible for ensuring that the therapy pressure setting is specified individually for each patient according to the device configuration, including accessories, that is to be used.
• The operator is required to regularly assess the effectiveness of the therapy settings.
• Also observe the respective instructions for use for the therapy device, the components, and the accessories.
• Always carry out a function check before using the unit
• Keep the therapy device and accessories out of the reach of children and pets. Keep the therapy device in its transport case when not in use and when it is being transported.
2.3 Warnings in this document
Warnings are used to flag up safety-relevant information.
You will find a warning preceding any action that entails a hazard for persons or equipment.
Danger!
Designates an extremely dangerous situation. Failure to observe this warning will lead to serious, irreversible injury, or death.
Warning!
Designates an extremely dangerous situation. Failure to observe this warning may lead to serious, irreversible, or fatal injury.
Caution!
Designates a dangerous situation. Failure to observe this warning may lead to minor or moderately serious injury.
Notice!
Indicates a harmful situation. Failure to observe this warning may lead to damage to equipment.
Designates useful information relating to a particular action.
EN
|
11
3 Product description
3.1 Therapy device overview
3
1
2
4 5
6
12
|
EN
7
17 16
14
13
12 8
15 11
10
9
NO. DESIGNATION
Cover
Unlocking button therapy device
DESCRIPTION
Covers the humidifier connection when no respiratory air humidifier is connected.
Allows the cover to be removed for connecting the humidifier.
Display
System interface
Handle
SD card
Allows operation of the therapy device and the respiratory air humidifier.
Displays settings and current values.
Connects the therapy device to modules.
Allows lifting and transporting of the therapy device.
Records therapy data.
Filter compartment in the suction area
Houses the air filter and, where applicable, the pollen filter. The respiratory air is sucked in here and the dust particles are filtered out.
Connects the therapy device to the power supply unit.
Power input
Power supply unit with connection cable
Supplies power to the device. Connects the power supply unit to the therapy device.
3 Product description
NO. DESIGNATION
Mounting holes
Power supply cable
SD card slot
Micro USB port
On/Off button
Respiration hose
Hose heating system connection
Device outlet
DESCRIPTION
For attaching and securing a module to the therapy device.
Connects the power supply unit to the power socket.
For inserting an SD card. The symbol on the display indicates the communication between the SD card and the therapy device.
Used for point-to-point connection with a PC on which prismaTS is installed. Allows settings to be changed on the therapy device and data to be exported.
Switches the therapy device on and off.
Switches the therapy device to standby mode.
Starts and stops the therapy.
Connects the therapy device to the respiratory mask.
Electrical power supply connection for a heatable hose.
Connection for the respiratory hose, through which the patient is supplied with respiratory air.
3.2 Display
The information shown on the display depends on the current status of the therapy device:
• Standby mode (no therapy in progress)
The therapy device operating hours since therapy began are shown for the first
30 seconds. Then the device switches to the start screen automatically.
The start screen shows the clock and the wake-up time if the alarm clock is set.
(see "3.2 Display", page 13) .
Settings can be performed on the therapy device (see "6 Settings in the menu", page 38)
.
• Therapy mode (therapy in progress)
Therapy is in progress (see "3.2.2 Display in Therapy mode", page 15) .
You can perform the mask test and start the softSTART sleep aid
(see "5 Operation", page 23) .
• Energy-saving mode
The therapy device is supplied with a very low level of power; nothing is shown on the display. You can return to Standby mode by pressing the On/Off button .
EN
|
13
14
|
EN
3 Product description
3.2.1 Display in Standby mode (Start screen)
5
NO.
DESIGNATION
Info menu button
Alarm clock with wake-up time
Menu button
Dimmer button
Time
4
DESCRIPTION
Provides access to the info menu.
Alarm clock is set.
Displays the set wake-up time.
Provides access to the settings menus.
Dims the display.
Displays the current time.
3
2
1
3 Product description
3.2.2 Display in Therapy mode
1 2
3
4
9 8
5
7 6
NO.
DESIGNATION
Time
SD card symbol
Info button
Alarm clock with wake-up time softSTART button
DESCRIPTION
Displays the current time.
The SD card is in the therapy device.
Provides access to the info screen with detailed information on the therapy currently in progress.
Alarm clock is set.
Displays the set wake-up time.
Switches the softSTART function on or off.
Displays the time remaining.
If the softSTART is off, the set softSTART period is displayed.
If there is no softSTART button, the physician or authorized dealer has disabled this function.
Respiration status symbol Indicates the current respiration status.
Mask status symbol with leak indicator
Indicates how well the respiratory mask is positioned.
Humidifier button for respiratory air humidifier
Indicates that the respiratory air humidifier is connected and switched on.
Shows the set humidifier level of the respiratory air humidifier.
Function buttons for the respiratory air humidifier
Allow the humidifier level to be increased/ decreased.
EN
|
15
16
|
EN
3 Product description
3.2.3 Symbols on the display
SYMBOL DESCRIPTION
Bacteria filter is connected and active.
If this symbol is displayed even though you are not using a bacteria filter, contact your authorized dealer.
Air filter replacement required. (Symbol only appears if the authorized dealer has activated the reminder to change the air filter)
Maintenance required (symbol only appears when maintenance function is active).
USB port prismaCONNECT module is plugged in
Module prismaPSG is connected (Green symbol)
No connection to prismaPSG module established (Gray symbol)
Network connection available (Green symbol)
No network connection available (Gray symbol)
SD card in SD card slot.
Symbol flashes: Data is being saved to the SD card or read off the SD card.
Respiratory air humidifier is connected and switched off.
Respiratory air humidifier is connected and switched on.
The set humidifier level is displayed.
The choice of humidifier levels 1-7 can be limited by the physician.
Respiratory air humidifier is connected and empty of water.
3 Product description
SYMBOL DESCRIPTION
Alarm clock is set.
If no alarm clock symbol is shown: The alarm clock is off.
Displays the respiration status:
• Arrow pointing upward: inhalation
• Arrow pointing downward: exhalation
• Green arrow, spontaneous respiration
• Orange arrow, assisted breathing
Apnea
Mask position is good, no leaks
Mask is not well positioned, considerable leaks, the efficacy of the therapy is not guaranteed
Indicates the diameter of the hose in mm.
Indicates which menu level you are currently in:
The more green dots, the deeper you are in the menu structure.
Alarm window
Low-priority alarm triggered.
Alarm paused for 2 minutes.
Indicates that the acoustic signal for an alarm can be muted (Black symbol)
Acoustic signal for alarm is muted (Orange symbol)
EN
|
17
3.3 Accessories
1
3
2
3 Product description
4
5
8
18
|
EN
7
6
NO. DESIGNATION DESCRIPTION
Respiratory air humidifier Humidifies the respiratory air
Pollen filter
Inverter
Filters the suctioned respiratory air and prevents the ingress of fine dust particles, pollen and fungal spores.
Enables operation of the device via a DC power socket
(12 V/24 V)
Respiratory hose with
15 mm diameter
Connects the therapy device to the respiratory mask.
Communication module
SpO
2
and nurse call module
PSG module
Heatable hose
Creates a connection between the therapy device and a
PC or the PSG module
Connects the therapy device with a call system and acquires SPO
2
and pulse frequency data.
Converts digital signals from the therapy device into analog data. Is used in sleep laboratories.
Avoids condensation in the respiration hose.
4 Preparation
4 Preparation
4.1 Setting up the therapy device
Material damage due to overheating!
Temperatures which are too high can cause the therapy device to overheat and damage the device.
⇒ The therapy device and power supply unit must not be covered with textiles
(e.g., bedclothes).
⇒ Do not operate the therapy device close to heating systems.
⇒ Do not expose the therapy device to direct sunlight.
⇒ Do not operate the therapy device in the transport bag.
1. Place the therapy device on a flat surface (e.g., a bedside table).
2. Leave the suction area of the therapy device uncovered.
3. Keep the power plug and power socket accessible at all times.
4. Pull the protective foil off the device.
4.2 Connecting up the power supply
Risk of injury due to electric shock when connecting an incorrect power supply unit to the line power!
The power supply unit contains a safety device to prevent electric shock. The use of a non-original power supply unit may result in injury to the user and the patient.
⇒ Only operate the device on line power using the power supply unit recommended by the manufacturer.
EN
|
19
1. Connect the power supply cable with the power supply unit.
4 Preparation
20
|
EN
2. Insert the free connector of the power supply unit's connection cable into the power supply port on the therapy device. When doing so, pay attention to the alignment of the connector.
4 Preparation
If you want to operate the therapy device at 12 V or 24 V, connect the optionally available inverter WM 24616 (12 V) or WM 24617 (24 V) to the device.
3. Plug the free end of the power supply cable in the power socket.
The power supply unit adjusts to the line voltage (110 V or 240 V) automatically.
The LED on the power supply unit lights up green.
If you want to disconnect the therapy device from the power supply, press the clip on the connector and pull the connector out.
Do not pull on the power supply cable.
4.2.1 Connecting up the respiration hose
Risk of injury due to contaminated or infected patient hose system!
A patient hose system contaminated or infected due to lack of or incorrectly performed hygienic preparation procedures can pass contamination or infection on to the next patient and cause injuries.
⇒ Do not reprepare disposable hose systems.
⇒ Use a bacteria filter
⇒
1. Connect the respiration hose to the device outlet.
EN
|
21
22
|
EN
4 Preparation
Risk of asphyxia when using full face masks without exhalation system!
When using full face masks without an integrated exhalation system, the CO
2 concentration can increase to critical values and endanger the patient.
⇒ Use full face masks with an external exhalation system if there is no exhalation system integrated.
⇒ Observe the instructions for use of the exhalation system.
2. If not integrated: Insert the external exhalation system between the respiratory mask and the respiration hose (see instructions for use of the respiratory mask and the exhalation system).
Risk of injury due to incorrectly positioned respiration hose!
An incorrectly positioned respiration hose can injure the patient.
⇒ Never place the respiration hose around the neck.
⇒ Do not use any small parts to fix the respiration hose in position as they might be accidentally swallowed.
⇒ Do not squash the respiration hose.
3. Connect the mask with the respiration hose.
Correct positioning and alignment of the mask on the patient's face is critical for consistent operation of the device.
4. Check whether the hose diameter used is set in the therapy device
(see "6.2 Setting accessories parameters", page 39) .
5. Put on the respiratory mask (see instructions for use of the respiratory mask).
6. Start the therapy
(see "5.4 Starting the therapy", page 25) .
.
5 Operation
5 Operation
5.1 Navigating the menu
You configure all the settings in the menu via the display. Press the required field directly on the display.
BUTTON FUNCTION
Go back a screen
Go forward a screen
Select values:
• If the parameter can have exactly 2 possible values (e.g., on/off):
Press the button.
The value changes to the other one.
• If the parameter can have a range of different values, press the button and select the value from the overview.
Increase or decrease value
Confirm value
Reject value
Go back to start screen (Standby or Therapy mode)
5.2 Switching on the therapy device
5.2.1 Switching on the therapy device for the first time
Before the first therapy is performed, the therapy device must be configured. If your authorized dealer has not done so already, configure the following settings.
Material damage if power supply is interrupted during configuration!
If the power supply is interrupted prematurely, the configuration will not be performed correctly.
⇒ Leave the therapy device connected to the power supply throughout the configuration.
⇒ Only disconnect the power supply once the Configuration successful message has appeared.
EN
|
23
5 Operation
1. Connect up the power supply (see "4.2 Connecting up the power supply", page 19)
.
2. Select your preferred language.
3. Select your time zone with the arrow keys and .
24
|
EN
4. Set the time:
• Select daylight saving time or standard time . Click on the symbol with the gray background to select it. The background turns green when the setting is activated.
• Use the arrow keys on the right to set the minutes.
• Select the clock version: 24 h (0-24) or 12 h (0-12)
5. Confirm the set time with the button.
If you have received an SD card from your authorized dealer with the configuration,
5 Operation
5.2.2 Switching the therapy device on each time
The therapy device can assume 3 different modes:
• Standby mode (no therapy in progress)
• Therapy mode (therapy in progress)
• Energy saving mode (display is off to save energy during the day)
1. To switch the therapy device to Standby mode, connect up the power supply
(see "4.2 Connecting up the power supply", page 19)
.
2. If the display remains off, the therapy device is in Energy saving mode: Press the
On/Off button .
After being switched on, the device displays the patient-related operating hours for
30 seconds.
5.3 Switching off the therapy device
1. To save energy during the day, keep the On/Off button
3 seconds.
depressed for or
If the automatic energy saving function is activated: The therapy device switches to the Energy saving mode automatically 15 minutes after the user has performed the last action.
The automatic energy saving function can be activated in the menu Main menu |
Device | Energy saving
(see "6.4 Setting device parameters", page 40)
.
5.4 Starting the therapy
1. Connect the components (see "6.1 Setting comfort parameters", page 38)
.
2. Connect the power supply (see "4.2 Connecting up the power supply", page 19)
.
3. If the display remains off, the therapy device is in Energy saving mode: Press the
On/Off button .
The therapy device switches to the Standby mode.
4. Press the On/Off button or
.
If the auto START -STOP function is active: Breathe into the mask.
You can activate the auto START -STOP function in the menu Main menu | Comfort | autoSTART-STOP
(see "6.1 Setting comfort parameters", page 38)
.
The therapy starts.
The start screen is shown in the Therapy mode.
EN
|
25
5 Operation
If you want to view detailed information on your therapy: Press the info button .
26
|
EN
To allow you to sleep undisturbed, the display automatically turns dark after
30 seconds. The therapy continues normally. As soon as you press the display, the start screen is shown in the Therapy mode again.
5.5 Ending the therapy
1. Press the On/Off button or
.
If the autoSTART-STOP function is active: Remove the respiratory mask.
The therapy is automatically ended after 5 seconds.
You can activate the auto START -STOP function in the menu Main menu | Comfort | auto START -STOP
(see "6.1 Setting comfort parameters", page 38) .
The therapy data for the last therapy session is shown briefly if the physician or authorized dealer has enabled this function. In all other cases, the usage time is displayed.
5 Operation
The more green checks are shown (max. 3), the better the result.
If you want to end the therapy prematurely during the night, you can use the dimmer button on the start screen to turn the display dark and sleep undisturbed.
The therapy device is still supplied with power and the alarm function remains activated. As soon as you touch the display, the start screen is shown in the Standby mode again.
5.6 Performing a mask test
The therapy device is equipped with a mask test function. To minimize the risk of leaks and test the correct positioning of the mask even at higher pressures, you can perform a mask test before starting the therapy.
Requirement
• The mask test function has been enabled by the physician or authorized dealer.
• The therapy device is in Therapy mode.
1. Press the button.
2. To start the mask test, press the mask test
The remaining time in seconds is shown.
button.
3. Check the seal of the mask against what is shown on the display:
SYMBOL MEANING
Mask position is good, no leaks
Mask is not well positioned, considerable leaks, the efficacy of the therapy is not guaranteed
4. If necessary: Adjust the mask straps.
EN
|
27
28
|
EN
5 Operation
5. Wait until the therapy device automatically ends the mask test after 30 seconds.
or
To end the mask test prematurely, press the mask test button .
If you switch the softSTART on during the mask test, the mask test is automatically switched off.
5.7 Switching softSTART on/off
The softSTART function makes it easier to get used to the ventilation pressure when falling asleep. You can set a pressure different to the prescribed therapy pressure.
When switched on, the therapy device sets this softSTART pressure. The pressure then increases slowly within the specified period or drops after the specified period
(maximum 45 minutes) to the therapy level.
This function is suitable for patients who find a high or low pressure uncomfortable when awake and cannot fall asleep.
Requirement
• The softSTART function has been enabled by the physician or authorized dealer.
• A softSTART pressure is set (see "6.1 Setting comfort parameters", page 38) .
1. Start the therapy
(see "5.4 Starting the therapy", page 25) .
2. If softSTART was activated during the last therapy: softSTART starts automatically when the therapy starts.
or
Press the softSTART button to switch softSTART on.
The remaining time in minutes is shown.
3. Press the softSTART button to switch softSTART off.
The set softSTART time in minutes is shown.
When running, a mask test will only interrupt softSTART and it will be restarted after the mask test.
5 Operation
5.8 Setting the respiratory air humidifier
5.8.1 Switching on the respiratory air humidifier
The respiratory air humidifier switches on automatically when you start the therapy
(see "5.4 Starting the therapy", page 25) .
You can also preheat the humidifier to ensure that the water in the respiratory air humidifier has already reached the required temperature by the start of the therapy.
Please note that the respiratory air humidifier will switch itself off again automatically after 30 minutes of preheating.
Requirement
• The therapy device is in Standby mode.
• The respiratory air humidifier is filled with water.
• The respiratory air humidifier is connected.
The humidifier button is gray .
1. Press the humidifier button .
5.8.2 Switching off the respiratory air humidifier
The respiratory air humidifier switches off automatically when you end the therapy
(see "5.5 Ending the therapy", page 26) .
You can also switch the respiratory air humidifier off during the therapy.
Requirement
• The therapy device is in the Therapy mode.
• The respiratory air humidifier is connected to the therapy device.
• The respiratory air humidifier is switched on.
The humidifier button is green .
EN
|
29
5 Operation
1. Press the humidifier button .
If there is no more water left in the respiratory air humidifier, the respiratory air humidifier switches off automatically. The humidifier button is orange .
5.8.3 Setting the humidifier level
Requirement
• The therapy device is in the Standby or Therapy mode.
• The respiratory air humidifier is filled with water.
• The respiratory air humidifier is connected to the therapy device.
• The respiratory air humidifier is switched on.
The humidifier button is green and the humidifier level is shown .
30
|
EN
1. The and buttons can be used to increase or decrease the humidifier level.
5 Operation
There are seven humidifier levels available (1-7). The level which is suitable for you depends on the room temperature and the humidity. The standard setting is level 4.
If you wake up with dry airways, the heating is set too low. If there is condensation in the respiration hose in the morning, the heating is too high.
The choice of humidifier levels 1-7 can be limited by the physician.
To reduce condensation in the respiration hose, we recommend using a hose heating system.
5.9 Setting the alarm
5.9.1 Setting the wake-up time and switching on the alarm
Requirement
The therapy device is in Standby mode.
1. Press the time display on the start screen.
or
Press the menu button .
2. Press the Time field.
3. Press the Wake-up time field.
4. To switch the alarm on, press the alarm button .
5. To set the wake-up time, use the left arrow keys to select the hours and the right arrow keys to select the minutes.
6. Confirm the settings with the button.
7. To return to the start screen, press the Home button .
EN
|
31
32
|
EN
5 Operation
5.9.2 Switching off the alarm
Requirement
The alarm is ringing.
1. To snooze the alarm for 5 minutes, press the Pause field.
2. To turn the alarm off for today, press the Off field.
The alarm will go off the following day again at the set wake-up time.
5.9.3 Deactivating the alarm
Requirement
• The therapy device is in Standby mode.
• The alarm is switched on
(see "5.9.1 Setting the wake-up time and switching on the alarm", page 31)
.
1. Press the time display on the start screen.
or
Press the menu button .
2. Press the Time field.
3. Press the Wake-up time field.
4. Press the alarm button .
5. Confirm the setting with the button.
6. To return to the start screen, press the Home button .
5.10 Viewing therapy data and device information
In the info menu you can view information about the therapy (usage time, mask fit, therapy quality) within a selectable period of time and general information about the device and network.
Requirement
The therapy device is in Standby mode.
1. Press the info button .
5 Operation
2. If necessary: To view therapy data from a night other than the previous night, select the desired date in the list .
3. If necessary: To view a longer period of time, navigate to the second screen .
EN
|
33
5 Operation
4. Select the required period.
5. To go back a screen, press the arrow key .
34
|
EN
6. If required, save all the data to the SD card
(see " Saving the therapy data manually", page 36)
.
7. To view the device information, navigate to the next screen using the arrow keys
and .
8. To exit the info menu, press the Home button .
5 Operation
5.11 Using the SD card
An SD card is not necessarily required for the operation of the therapy device. The therapy data and settings are stored internally in the device.
Loss of data due to incorrect SD card!
SD cards not purchased from the manufacturer may have reduced functionality or result in the loss of data.
⇒ Only use SD cards from brand manufacturers which comply with the
specifications (see "13.1 Technical Data", page 52)
.
Do not use the SD card for third-party files.
5.11.1 Inserting the SD card
Requirement
The therapy device is in Standby mode.
1. Open the SD card slot cover.
2. Slide the SD card into the SD card slot until it audibly clicks into place.
Note: The beveled corner of the SD card must be at the top and facing the device during insertion.
3. Close the SD card slot cover.
EN
|
35
36
|
EN
5 Operation
5.11.2 Saving therapy data to the SD card
Data loss in case of power loss!
Data may be lost if the therapy device is disconnected from the power supply during the saving process.
⇒ Keep the therapy device connected to the power supply during the saving process (SD card symbol flashes).
Autosave
The therapy device saves the therapy data automatically in the following events:
• Each time you end a therapy.
• Each time you insert an SD card. Only insert an SD card when the device is in
Standby mode.
• When the therapy device is reconnected to the power supply after a saving process is interrupted.
Saving the therapy data manually
Requirement
• The SD card is inserted in the therapy device (see "5.11.1 Inserting the SD card", page 35)
.
• The info menu with the therapy data for the requested period is open
(see "5.10 Viewing therapy data and device information", page 32)
.
1. To save all the therapy data to the SD card, press the SD card button .
2. Press the Save all data field and confirm with the OK field.
5.11.3 Removing the SD card
Requirement
• The therapy device is in the Standby mode.
• The SD card symbol is no longer flashing.
1. Open the SD card slot cover.
2. Briefly press in the SD card.
The SD card is ejected slightly.
5 Operation
3. Remove the SD card.
4. Close the cover of the SD card slot.
5.11.4 Setting the device with the SD card
You can set the device with the help of an SD card provided by your physician or authorized dealer.
Requirement
The therapy device is in the Standby mode.
1. Insert the SD card with the saved device settings (see "5.11.1 Inserting the SD card", page 35) .
The message Configuration via SD card was successful appears on the display.
You can continue the therapy with the new settings.
If the new settings for your device were not suitable or could not be read, the message Configuration via SD card has failed appears on the display. Contact your authorized dealer to obtain new settings.
EN
|
37
38
|
EN
6 Settings in the menu
6 Settings in the menu
You can configure settings for the comfort, accessories, and time parameters in the settings menu when the therapy device is in Standby mode.
6.1 Setting comfort parameters
Comfort parameters facilitate handling of the therapy device and components for the patient and ensure a comfortable therapy.
Requirement
The therapy device is in Standby mode.
1. Press the menu button .
2. Press the Comfort field.
3. Configure the desired settings and confirm.
PARAMETER autoSTART-
STOP
Mask test pressure softSTART
Pressure softSTART time
POSSIBLE
VALUES
DESCRIPTION
On
Off
8 cmH
2
20 cmH
O-
2
O
(depending on the therapy pressure currently set)
Here you can activate/deactivate the automatic on/off function autoSTART-STOP.
If the automatic on/off function is activated, you can switch the therapy device on with a breath.
If there is no pressure for 5 seconds (e.g., because the mask has been removed), the therapy device switches itself off again automatically.
Here you can set the pressure at which the mask test
is performed (see "5.6 Performing a mask test", page 27) .
Leaks due to a poorly sitting mask often only occur at higher pressures.
Intervals of 0.5 in the range prescribed by the physician or authorized dealer
(e.g., at least
4 cmH
2
8 cmH
2
O to
O).
Intervals of
5 minutes in the range prescribed by the physician or authorized dealer
(e.g., 5 mins to max. 45 mins).
The softSTART function makes it easier to get used to the ventilation pressure when falling asleep.
You can set the desired softSTART pressure here.
If it is not possible to select this function, it must be enabled by your physician or authorized dealer.
Here you can set the period of time during which the ventilation pressure increases until it reaches the therapy pressure when the softSTART function is used.
If it is not possible to select this function, it must be enabled by your physician or authorized dealer.
6 Settings in the menu
PARAMETER softPAP
POSSIBLE
VALUES
Off
1
2
3
DESCRIPTION
Settings 1 and 2 of the softPAP breathing relief function are intended for patients who find exhaling against high pressure uncomfortable. The breathing relief function reduces the pressure early during the transition to expiration, allowing you to breathe out more easily.
Setting 3 is suitable for patients who experience respiratory distress with a low pressure setting. The pressure is raised slightly during inspiration.
You can select the setting for the softPAP breathing relief here or deactivate it if you do not wish to use the function anymore.
• Setting 1: Low breathing relief
• Setting 2: Normal breathing relief
• Setting 3: Breathing relief with inhalation assistance
This function is only available in CPAP and APAP mode. If it is not possible to select this function in one of these modes, it must be enabled by your physician or authorized dealer.
6.2 Setting accessories parameters
The accessories parameters are used to set the use of the accessories.
Requirement
The therapy device is in Standby mode.
1. Press the menu button .
2. Press the Accessories field.
3. Configure the desired settings and confirm.
PARAMETER
Tube type
Air filter
Change
POSSIBLE
VALUES
15 mm
19-22 mm
Changed
Cancel
DESCRIPTION
Here you select the diameter of the hose type used. If it is not possible to select this function, it must be enabled by your physician or authorized dealer.
Here you specify whether you have changed the air filter. For this function, the authorized dealer must have activated the air filter reminder.
EN
|
39
40
|
EN
6 Settings in the menu
6.3 Setting time parameters
In the time parameters you set the minutes of the current time, the time zone, and the desired wake-up time.
Requirement
The therapy device is in Standby mode.
1. Press the menu button
2. Press the Time field.
.
3. Configure the desired settings and confirm.
PARAMETER
Time
Time zone
Wake-up time
POSSIBLE
VALUES
DESCRIPTION
Here you can set the current time:
• Select daylight saving time or standard time.
The green background of the symbol shows that this setting is active.
• Use the arrow keys on the right to set the minutes.
• To set the hours: Select another time zone.
• Select the clock version:
24 hours (0-24)
12 hours (0-12)
You can reset the time to the end of the last therapy at most.
UTC -12 to
UTC +12
Here you select the desired time zone.
00:00 -12:00 /
23:59
Here you set the time at which you want to be woken
up (see "5.9.1 Setting the wake-up time and switching on the alarm", page 31) .
6.4 Setting device parameters
You can use the device parameters to set the brightness of the display and the volume of the acoustic signals among other things as you wish.
Requirement
The therapy device is in Standby mode.
1. Press the menu button
2. Press the Device field.
.
3. Configure the desired settings and confirm.
6 Settings in the menu
PARAMETER
Display alert brightness
Leakage
Energy saving
Key tone volume
Alarm volume
Alarm clock volume
POSSIBLE
VALUES
1
2
3
Off
On
Off
On
Off
1
2
3
1
2
3
Off
1
2
3
DESCRIPTION
Here you can set the brightness of the display.
• Level 1: Dark
• Level 2: Normal
• Level 3: Bright
Here you can set whether an alarm should be triggered in case of a leak. This allows you to change the position of your mask at night. By doing so you avoid side effects or a reduced therapy quality due to severe leaks.
If it is not possible to select this function, it must be enabled by your physician or authorized dealer.
Here you can activate or deactivate whether the therapy device automatically switches to Energy saving mode 15 minutes after the therapy has finished.
You save electricity if the therapy device is in Energy saving mode during the day.
Here you can set the volume of the acoustic signal for every time a key is pressed or switch the signal off.
• Level 1: Quiet
• Level 2: Normal
• Level 3: Loud
Here you can set the volume of the alarms.
• Level 1: Quiet
• Level 2: Normal
• Level 3: Loud
Here you can set the volume of the alarm.
• Level 1: Quiet
• Level 2: Normal
• Level 3: Loud
EN
|
41
7 Hygienic preparation
42
|
EN
7.1 General information
• This product may contain disposable items. Disposable items are intended to be used only once.
Use these items only once and do not reprocess them.
Reprocessing disposable items may impair the functionality and safety of the product and lead to unforeseeable reactions as a result of aging, embrittlement, wear, thermal load, the effects of chemical processes, etc.
• Wear suitable protective equipment for disinfection work.
• Please refer to the instructions for use supplied with the disinfectant used.
• Also observe the respective instructions for use for the therapy device, the components, and the accessories.
• The therapy device is suitable for subsequent use on further patients following hygienic preparation by the authorized dealer.
7.2 Cleaning intervals
INTERVAL
Weekly
Monthly
Every 6 months
Annually
As necessary
When changing patients
ACTION
Clean the therapy device
(see "7.3 Hygienic preparation of the therapy device", page 43)
Clean the respiration hose (see "7.4 Hygienic preparation of the respiration hose", page 45)
Clean the respiratory air humidifier
In clinical areas: Disinfect the respiratory air humidifier
Clean the air filter
(see "7.3.1 Cleaning the air filter (gray filter)", page 44)
Replace the air filter
Replace the respiration hose
Descale the respiratory air humidifier.
In clinical areas: Disinfect the respiration hose.
For hygienic reasons: Replace the housing components of the respiratory air humidifier if they are in poor condition (e.g., if cracks appear).
If the therapy device or respiratory air humidifier has been used without a bacteria filter: Have professional hygienic preparation performed before using the device again. Send the therapy device to your authorized dealer.
7 Hygienic preparation
7.3 Hygienic preparation of the therapy device
Risk of injury from electric shock!
Any liquids entering the device can cause a short circuit, injure the user, and damage the therapy device.
⇒ Disconnect the therapy device from the power supply before starting the hygienic preparation.
⇒ Do not immerse the therapy device and the components in liquids.
⇒ Do not pour liquids over the therapy device and the components.
1. Switch off the therapy device
(see "5.3 Switching off the therapy device", page 25)
.
2. Disconnect the therapy device from the power supply.
3. If present: Remove the respiratory air humidifier.
4. Prepare the therapy device and the components hygienically in accordance with the following table:
PART
Housing
High-gloss surfaces on the housing
Display
Power supply unit
Power supply cable
CLEANING
Wipe with a damp cloth using water or mild soap
Wipe with a damp cloth using water or mild soap; do not use microfiber cloths
Wipe with a dry cloth: do not use water, mild soap or microfiber cloths
Wipe with a damp cloth using water or mild soap
Wipe with a damp cloth using water or mild soap
DISINFECTION
Wipe disinfection
(Recommendation: terralin
®
protect or perform advanced
Alcohol EP)
STERILIZATION
Not permitted
5. If present: Connect the respiratory air humidifier to the therapy device.
6. Reconnect to power supply.
7. Perform a function check
(see "8 Function check", page 46) .
EN
|
43
7.3.1 Cleaning the air filter (gray filter)
7 Hygienic preparation
1. Remove the air filter.
2. Clean the air filter under running water.
3. Leave the air filter to dry.
4. Replace the air filter in the holding bracket.
7.3.2 Replacing the optional pollen filter (white filter)
1. Remove the air filter.
44
|
EN
2. Remove and dispose of the pollen filter.
3. Insert the new pollen filter in the holding bracket.
4. Replace the air filter in the holding bracket.
7 Hygienic preparation
7.4 Hygienic preparation of the respiration hose
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the device.
⇒ Only use the respiration hose when it is completely dry.
1. Remove the respiration hose from the therapy device.
2. Carry out hygienic preparation of the respiration hose as specified in the following table:
PART CLEANING
Respiration hose
With warm water and washing-up liquid
DISINFECTION
Immersion disinfection
(Recommendation: gigasept FF
®
)
STERILIZATION
Not permitted
3. Rinse respiration hose off with clean water.
4. Shake respiration hose out thoroughly.
5. Hang up the respiration hose and leave it to drip dry.
6. Dry the respiration hose.
If you use a heatable respiratory hose, please observe the instructions for use for the respiratory hose.
EN
|
45
46
|
EN
8 Function check
8.1 Intervals
Carry out a functional check after every hygienic preparation, after every repair, and at least every 6 months.
8.2 Checking the therapy device
Requirement
• The therapy device is disconnected from the patient.
• The therapy device is connected to the power supply.
• The therapy device is in Standby mode.
1. Inspect the therapy device for external damage.
If damaged: Do not use the therapy device.
2. Inspect the plug and cable for external damage.
If damaged: Contact the authorized dealer and have the parts replaced.
3. Check that the components are connected to the therapy device correctly in accordance with these instructions for use.
4. Switch on the therapy device
(see "5.2 Switching on the therapy device", page 23)
.
5. If softSTART is activated: Press the softSTART button to stop softSTART.
6. Close the opening on the respiratory mask (e.g., using the elbow).
7. Press the info button .
8. Compare the pressure shown in the display with the prescribed pressure.
If the pressure variance is > 1 cmH the authorized dealer.
2
O: Do not use the therapy device and contact
9 Alarms and error messages
9 Alarms and error messages
If you are not able to clear an error message with the aid of the table below, or in the event of unexpected operation or an incident, you should have the device repaired by the manufacturer or your authorized dealer. To avoid serious damage, do not continue using the device.
9.1 Alarms
Alarms can be categorized into three priority levels (low, medium, high). This device only has low-priority alarms, which are indicated by the symbol .
9.1.1 Alarm messages
ALARM MESSAGE CAUSE REMEDY
Pressure build-up not possible!
Please connect the mask and hose.
No respiration hose and/or mask connected.
Connect the mask and respiration hose correctly
"4.2.1 Connecting up the respiration hose", page
High Leak!
Please check the mask fit.
Mask has slipped or is not tight.
Reposition mask. If the mask is faulty, exchange it.
Apnea! Please check the ventilation settings and the course of the respiration hose.
Low tidal volume!
Please check the ventilation settings and the course of the respiration hose.
The respiratory volume output by the device is lower than the target value.
The respiratory volume output by the device is lower than the target value.
Check that the respiration hose is neither blocked nor kinked.
Reposition the mask and breathe through it.
If the alarm continues to show:
Have the settings checked by the attending physician.
Check that the respiration hose is neither blocked nor kinked.
Reposition the mask and breathe through it.
If the alarm continues to show:
Have the settings checked by the attending physician.
EN
|
47
48
|
EN
9 Alarms and error messages
ALARM MESSAGE CAUSE
Low minute volume!
Please check the ventilation settings and the course of the respiration hose.
The respiratory volume output by the device is lower than the target value.
REMEDY
Check that the respiration hose is neither blocked nor kinked.
Reposition the mask and breathe through it.
If the alarm continues to show:
Have the settings checked by the attending physician.
9.1.2 Muting the alarm
If an alarm sounds, you can mute the audible alarm for 2 minutes.
Requirement
An alarm has been triggered.
.
1. Press the mute symbol .
The alarm is muted for 2 minutes. The symbol turns orange. After 2 minutes, the audible alarm sounds again.
If your physician has activated this function, you can also deactivate the High Leak
alarm permanently (see "6.4 Setting device parameters", page 40) .
9.1.3 Pausing the alarm
If an alarm sounds, you can pause the alarm for 2 minutes to operate the device normally in the meantime.
Requirement
The Apnea, Low minute volume, or Low tidal volume alarm has been triggered.
1. Press the PAUSE field.
The alarm is paused for 2 minutes. The symbol appears in the status line. After
2 minutes, the audible alarm sounds again.
If your physician has activated this function, you can also deactivate the High Leak
alarm permanently (see "6.4 Setting device parameters", page 40) .
9 Alarms and error messages
9.2 Faults in the therapy device
FAULT
No running noise, no information on the display.
It is not possible to start therapy with a breath.
The therapy device does not switch off after approx. 5 seconds when the mask is removed.
softSTART cannot be switched on.
Therapy device does not reach the lower pressure limit.
CAUSE
No power supply.
SD card defective.
autoSTART-STOP function is not active.
REMEDY
Check that the power supply cable is connected properly.
Check the function of the socket-outlet.
Remove the SD card (see 5.11.3, p.
from the power supply, reconnect it and switch it back on.
If the device can be switched on:
Replace SD card.
If the error persists: Contact your authorized dealer.
Activate the autoSTART-STOP function
autoSTART-STOP function can be impaired by accessories with a high level of resistance.
Contact your authorized dealer.
softSTART function is disabled.
Air filter is dirty.
Respiratory mask is leaking.
Ask the physician whether the function can be enabled.
Clean the air filter. If necessary:
Replace filter
(see "7 Hygienic preparation", page 42) .
Adjust the headband until the mask fits tightly.
If necessary, replace defective mask.
EN
|
49
9 Alarms and error messages
50
|
EN
9.3 Display messages
If the message Error (xxx): Please follow the instructions in the Instructions for use appears on the display, locate the displayed error code in the table. Rectify the error as described.
ERROR
CODE
(108)
CAUSE
The therapy device does not display the set time.
REMEDY
(204)
(205)
(206)
(702)
The respiratory air humidifier is not working correctly.
The power supply voltage is not within the permitted range.
Error in the prismaCONNECT module.
Device output is blocked. /
Water in therapy device.
All other error codes
Problems with the electronics
Contact the authorized dealer and have the device repaired.
Remove the respiratory air humidifier from the therapy device and connect it again.
If the message is still shown, contact the authorized dealer and have the device and the respiratory air humidifier checked.
Check whether the correct power supply unit is connected (WM 29657). Contact the authorized dealer and have the device and power supply unit checked.
Remove and reconnect the prismaCONNECT module. If the fault persists: Contact the authorized dealer and have the prismaCONNECT module replaced.
Ensure that the respiration hose and device output are not blocked.
If the fault persists:
• Check whether there is water in the device. To do so, remove the respiratory air humidifier and side part and tilt the device with the open side facing downward.
• If water comes out: Wait until all the water has escaped.
• Allow the device to dry until the message is no longer displayed. In future, do not transport the device with water in the respiratory air humidifier.
• If water collects in the respiration hose: Reduce the humidifier level to avoid condensation.
Disconnect the therapy device from the power supply and reconnect it
If the message is still shown, contact the authorized dealer and have the device and the respiratory air humidifier checked.
10 Maintenance
10 Maintenance
The therapy device is designed to have a useful service life of 6 years.
If the therapy device is used as intended in accordance with the instructions for use, it does not require any maintenance within this period.
If the therapy device is used beyond this period, we recommend having it checked by an authorized dealer.
If you identify faulty parts during the function check
(see "8 Function check", page 46)
, contact your authorized dealer.
11 Transport and storage
Store and transport the device under the specified ambient conditions (see "13.1
12 Disposal
Do not dispose of the product in the household waste.
Consult an authorized, certified electronic waste recycling company for proper disposal. You can find out their address from your environmental officer or from your local council.
The device packaging (cardboard box and inserts) can be disposed of as waste paper.
EN
|
51
13 Appendix
52
|
EN
13 Appendix
13.1 Technical Data
13.1.1 Technical data of therapy device
SPECIFICATION THERAPY DEVICE
Product class according to 93/42/EEC IIa
Dimensions W x H x D in cm 17 x 13.5 x 18
1.4 kg Weight
Temperature range
- Operation
- Transport and storage
+5°C to +40°C
-25°C to +70°C
Permissible humidity during operation, transport and storage
Rel. humidity 15% to 93%, non-condensing
Air pressure range
700 cmH
2
O to 1060 cmH
2
O corresponds to a height of 3000 m above sea level adapts automatically to altitude
Connection diameter of respiration hose in mm
Electrical output
System interface
19.5 (to fit standard cone)
Max. 40 VA
12 V DC
Max. 10 VA
Current consumption during operation
(Therapy)
240 V AC
100 V AC
0.11 A
0.25 A in standby mode (Standby)
240 V AC
100 V AC
0.035 A
0.022 A
Classification as per DIN EN 60601-1-11:
Protection class against elec. shock Protection class II
Degree of protection against elec. shock Type BF
Protection against harmful ingress of water and solid particles
Classification as per IEC 60601-1:
Operating mode
IP21
Continuous operation
13 Appendix
SPECIFICATION
Applied part
Average sound pressure level in operation as per ISO 80601-2-70
THERAPY DEVICE
Respiratory mask
Approx. 26.5 dB(A) at 10 cmH
2
O (corresponds to a sound power level of 34.5 dB(A)
Average sound pressure level in operation as per ISO 80601-2-70 with respiratory air humidifier
Approx. 27.5 dB(A) at 10 cmH
2
O (corresponds to a sound power level of 35.5 dB(A)
Sound pressure level of alarm message At least 58 db(A)
All device types
Disconnection, high leak (optional)
Alarms (optional)
Alarm output
CPAP operating pressure range
AcSV pressure range
BiLevel pressure range
Pressure accuracy
P Lim max error)
(maximum pressure in case of prisma30ST, prismaLAB
Apnea, low minute volume, low tidal volume
Optical and acoustic
4 cmH
2
4 cmH
2
O to 20 cmH
O to 30 cmH
2
O
2
O
4 cmH
< 20 cmH
2
≥
2
O to 30 cmH
20 cmH
2
2
O
O: ± 0.6 cmH
2
O: ± 0.8 cmH
2
O
O
< 40 cmH
2
O
Target volume in AcSV mode
It is not possible to set a target volume for the
AcSV mode. The pressure control stabilizes the volume at the respective current level.
Automatic backup frequency in AcSV and autoS/T modes
The automatic backup frequency is continuously adapted between 10 bpm and 20 bpm, depending on the filtered spontaneous rate and the relative respiratory minute volume of the patient.
0; 5-45 min min. 4 hPa softSTART can be adjusted softSTART pressure prisma25S-C
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Available modes
4 cmH
4 cmH
2
2
O to 25 cmH
2
O
20% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
CPAP, S
O to 25 cmH
2
O
EN
|
53
54
|
EN
SPECIFICATION prisma25S
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Available modes prisma25ST
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Trigger
- Pressure rise rate
- Backup frequency
- Available modes prisma30ST
- Inspiratory positive airway pressure (IPAP)
- Expiratory positive airway pressure (EPAP)
- Relative inspiration duration Ti/Tset
- Ti
- Trigger inspiration
- Trigger expiration
- Pressure rise rate
- Pressure drop rate
- Backup frequency
- Target volume
- Pressure adjustment
- Available modes
THERAPY DEVICE
4 cmH
2
O to 25 cmH
20% to 67%
2
O
4 cmH
2
O to 25 cmH
2
O
20% to 67%
Auto, can be set to 3 levels
Can be set to 3 levels
CPAP, APAP, S, autoS
4 cmH
4 cmH
2
Auto, can be set to 3 levels
Can be set to 3 levels
Auto, 0 bpm to 35 bpm
CPAP, APAP, S, autoS, autoS/T, S/T, T
4 cmH
2
2
O to 25 cmH
O to 25 cmH
2
O
O to 30 cmH
2
2
O
O
4 cmH
2
O to 25 cmH
2
O
20% to 67%
500 ms to 4000 ms
Auto, can be set to 3 levels
Auto, can be set to 3 levels
Can be set to 4 levels
Can be set to 3 levels
Auto, 0 bpm to 35 bpm
300 ml to 2000 ml
Can be set to 3 levels
CPAP, APAP, autoS/T, S, S/T, T, aPCV
13 Appendix
13 Appendix
SPECIFICATION
Peak flow as per
ISO 80601-2-70
CPAP and APAP modes
Test pressures:
4 cmH
8 cmH
2
O
2
O
12 cmH
2
16 cmH
2
20 cmH
2
O
O
O
AcSV mode, BiLevel
Test pressures:
4 cmH
2
17 cmH
O
10.5 cmH
2
23.5 cmH
2
O
O
25 cmH
2
30.0 cmH
2
O
O
2
O
Warming of respiratory air
THERAPY DEVICE
Pressure measured at the patient connection opening at a flow of 40 l/min
Average flow at the patient connection opening
4.0 cmH
2
8.0 cmH
2
O
O
11.9 cmH
2
15.9 cmH
2
19.9 cmH
2
O
O
O
235 l/min
230 l/min
220 l/min
215 l/min
210 l/min
4.0 cmH
2
10.4 cmH
17.0 cmH
O
2
O
2
23.5 cmH
2
25 cmH
2
O
30.0 cmH
O
O
2
O
235 l/min
225 l/min
215 l/min
200 l/min
195 l/min
190 l/min
Max. +3°C
EN
|
55
56
|
EN
SPECIFICATION
Stability of the dynamic pressure
(short-term accuracy) at 10 breaths/ min as per ISO 17510-1:2007 when using the 19 mm hose.
7 cmH
2
O
10 cmH
2
O
13.5 cmH
20 cmH
2
O
2
O
Stability of the dynamic pressure
(short-term accuracy) at 15 breaths/ min as per ISO 17510-1:2007 when using the 19 mm hose.
7 cmH
2
10 cmH
O
2
O
13.5 cmH
20 cmH
2
O
2
O
Stability of the dynamic pressure
(short-term accuracy) at 20 breaths/ min as per ISO 17510-1:2007 when using the 19 mm hose.
7 cmH
2
O
10 cmH
2
O
13.5 cmH
20 cmH
2
O
2
O
THERAPY DEVICE
Δ p < 0.24 cmH
2
O
Δ p < 0.28 cmH
2
O
Δ p < 0.3 cmH
2
O
Δ p < 0.4 cmH
2
O
Δ p < 0.24 cmH
2
O
Δ p < 0.32 cmH
2
O
Δ p < 0.4 cmH
2
O
Δ p < 0.48 cmH
2
O
Δ p < 0.4 cmH
2
O
Δ p < 0.32 cmH
2
O
Δ p < 0.46 cmH
2
O
Δ p < 0.56 cmH
2
O
13 Appendix
13 Appendix
SPECIFICATION
Stability of the dynamic pressure
(short-term accuracy) as per
ISO 80601-2-70 in CPAP and APAP modes
- when using the 19 mm hose
4 cmH
8 cmH
2
O
2
O
12 cmH
2
16 cmH
2
20 cmH
2
O
O
O
THERAPY DEVICE
Δ p < 0.68 cmH
2
O
Δ p < 0.58 cmH
2
O
Δ p < 0.52 cmH
2
O
Δ p < 0.44 cmH
2
O
Δ p < 0.64 cmH
2
O
- when using the 15 mm hose, bacteria filter, and oxygen safety valve
4 cmH
8 cmH
2
O
2
O
12 cmH
2
16 cmH
2
20 cmH
2
O
O
O
Stability of the dynamic pressure
(short-term accuracy) as per
ISO 80601-2-70 in modes with
2 pressure levels at 10 bpm inspiratory at 15 bpm inspiratory at 20 bpm inspiratory at 10 bpm expiratory at 15 bpm expiratory at 20 bpm expiratory
Stability of the static pressure (longterm accuracy) as per ISO 80601-2-70
- when using the 19 mm hose
- when using the 15 mm hose, bacteria filter, and oxygen safety valve
Pressure drop via the oxygen valve at 90 l/min at 60 l/min at 30 l/min
Recommended maximum additional oxygen flow
Accuracy of volume measurement at 20°C
Δ p < 1.06 cmH
2
O
Δ p < 1 cmH
2
O
Δ p < 1.08 cmH
2
O
Δ p < 1.02 cmH
2
O
Δ p < 0.96 cmH
2
O
Δ p = 0.8 cmH
2
O
Δ p = 1.4 cmH
2
O
Δ p = 2.4 cmH
2
O
Δ p = 0.6 cmH
2
O
Δ p = 0.6 cmH
2
O
Δ p = 0.6 cmH
2
O
Δ p = 0.15 cmH
2
O
Δ p = 0.19 cmH
2
O
0.5 cmH
2
O
0.25 cmH
2
0 cmH
2
O
O
15 l/min
±20%
EN
|
57
58
|
EN
SPECIFICATION
Filter and smoothing techniques
Pollen filter down to 1 μ m down to 0.3 μ m
Service life of pollen filter
13 Appendix
THERAPY DEVICE
• Target volume that can be set:
In the "slow" level, the device checks after every 8 breaths if the target volume has been reached and changes the pressure by 0.5 cmH
2
O. If the pressure reaches a corridor around the target volume, the device switches to exact regulation.
In the "medium" level, the device checks after every 5 breaths if the target volume has been reached and changes the pressure by
1.0
cmH
2
O. If the pressure reaches a corridor around the target volume, the device switches to exact regulation.
In the "fast" level, the device checks after every breath if the target volume has been reached and changes the pressure by
1.5
cmH
2
O. If the pressure reaches a corridor around the target volume, the device switches to exact regulation.
• Alarms:
The "low minute volume" and "low tidal volume" alarms are triggered if at least three of the last five breaths were below the alarm limit. The alarms are reset automatically as soon as the corresponding alarm limit is exceeded again with at least three of the five breaths.
If a target volume is activated, the "low tidal volume" alarm is only triggered once IPAPmax or PDIFFmax has also been attained.
The "Apnea" alarm is triggered if apnea is identified which is longer than the set alarm limit. The alarm is reset automatically as soon as the end of the apnea is identified.
Filter class E10
≥ 99.5%
≥ 85% approx. 250 hours
13 Appendix
SPECIFICATION
SD card
THERAPY DEVICE
Memory sizes of 256 MB to 8 GB can be used, interface compatible with SD physical layer version 2.0
TOLERANCES FOR MEASUREMENTS
Pressure:
Flow:
± 0.75% of measurement or ± 0.1 cmH
± 4 l/min
2
O
Temperature: ± 1.5°C
Sound pressure level and sound power level
± 2 dB(A)
The right to make design modifications is reserved.
All flow and volume values are determined under STPD conditions.
All the parts of the therapy device are free from latex.
The WM 100 TD therapy devices use the following open source software:
FreeRTOS.org
This device’s software contains code which is subject to the GPL. You will receive the source code and the GPL upon request.
13.1.2 Technical data of power supply unit
SPECIFICATION
Input voltage/maximum current
Input frequency
Output voltage/maximum current
POWER SUPPLY UNIT
100 V - 240 V AC, 3 A - 1.5 A
50 Hz - 60 Hz
37 V DC, 2.5 A
EN
|
59
60
|
EN
13 Appendix
13.1.3 Pressure volume curve
pV curve at RV=0.5 l and f=20/min
0.5
0.4
0.3
0.2
0.1
0
0.00
5.00
13.1.4 Pneumatic system diagram
10.00
Pressure [cmH
2
O]
15.00
20.00
Inlet for am bient air
Air filter Fan
Pressure sensor for patient pressure
Optional humidifier
O
2
pressure source
Flow regulator
Optional O
2 safety valve
Optional bacteria filter
Respiration hose,
1.8m
Patient mask
Nasal or full face mask
Exhalation system
13 Appendix
13.2 Emission of electromagnetic interference
GUIDELINES AND MANUFACTURER DECLARATION - EMISSION OF
ELECTROMAGNETIC INTERFERENCE
The device can be used in both static and mobile operation, in both domestic and appropriate hospital environments.
In a domestic environment, the device may cause radio interference, possibly making it necessary to take suitable remedial measures, such as realigning the device, for example.
MEASUREMENTS OF INTERFERENCE EMISSION COMPLIANCE
HF emissions to CISPR 11
HF emissions to CISPR 11
Emission of oscillations IEC 61000-3-2
Emission of voltage fluctuations/flicker to IEC 61000-3-3
Group 1
Class B
Class A
Complies
13.3 Electromagnetic interference immunity
GUIDELINES AND MANUFACTURER DECLARATION - ELECTROMAGNETIC
INTERFERENCE IMMUNITY
The device can be used in both static and mobile operation, in both domestic and appropriate hospital environments.
In a domestic environment, the device may cause radio interference, possibly making it necessary to take suitable remedial measures, such as realigning the device, for example.
INTERFERENCE
IMMUNITY TESTS
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
Discharge of static electricity (ESD) to
IEC 61000-4-2
± 8 kV contact discharge
± 15 kV air discharge
± 8 kV contact discharge
± 15 kV air discharge
ELECTROMAGNETIC ENVI-
RONMENT GUIDELINE
Floors should be made of wood or concrete or have ceramic tiles laid on them. If the floor has a synthetic material laid on it, relative
Electrical fast transients/bursts to IEC 61000-4-4
Surge immunity to
IEC 61000-4:-5
± 2 kV for power supply cables
± 1 kV for input and output cables
Connection
Source impedance: 2 Ω ,
18 μ F: 0.5 kV,
1 kV
Number of surges:
± 2 kV for power supply cables
± 1 kV for input and output cables
Connection
Source impedance: 2 Ω ,
18 μ F: 0.5 kV,
1 kV
Number of surges:
The quality of the supply voltage should correspond to that of a typical business or hospital environment.
The quality of the supply voltage should correspond to that of a typical business or hospital environment.
EN
|
61
62
|
EN
13 Appendix
GUIDELINES AND MANUFACTURER DECLARATION - ELECTROMAGNETIC
INTERFERENCE IMMUNITY
The device can be used in both static and mobile operation, in both domestic and appropriate hospital environments.
In a domestic environment, the device may cause radio interference, possibly making it necessary to take suitable remedial measures, such as realigning the device, for example.
INTERFERENCE
IMMUNITY TESTS
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVI-
RONMENT GUIDELINE
Voltage dips, short interruptions and voltage variations in supply voltage to
IEC 61000-4-11
Magnetic field at power frequency
(50/60 Hz) to
IEC 61000-4-8
Number of voltage drops: 3 drop levels/ duration:
30% / 500 ms
30 A/m
Duration: 30 s per axis
Axes: x axis, y axis, z axis
Number of voltage drops: 3 drop levels/duration:
30% / 500 ms
60% / 100 ms
30 A/m
Duration: 30 s per axis
Axes: x axis, y axis, z axis
The quality of the supply voltage should correspond to that of a typical business or hospital environment. If the user of the device requires
Magnetic fields at power supply frequency should correspond to the values typical of those found in business and hospital
13 Appendix
13.4 Electromagnetic interference immunity for ME equipment and ME systems
GUIDELINES AND MANUFACTURER DECLARATION - ELECTROMAGNETIC
INTERFERENCE IMMUNITY
The device can be used in both static and mobile operation, in both domestic and appropriate hospital environments.
In a domestic environment, the device may cause radio interference, possibly making it necessary to take suitable remedial measures, such as realigning the device, for example.
INTERFERENCE
IMMUNITY TESTS
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVI-
RONMENT GUIDELINE
Portable and mobile radio equipment should not be used at a distance from the device, including its cables, of less than the recommended safety distance calculated in accordance with the equation applicable to the transmission frequency.
Recommended safety distance:
Conducted HF interference to
IEC 61000-4:-6
Radiated HF interference to
IEC 61000-4:-3
Magnetic field at power frequency
(50/60 Hz) to
IEC 61000-4-8
10 V effective value
150 kHz to
80 MHz within
ISM bands
10 V/m
80 MHz to
2.7 GHz
80% AM at 2 Hz
30 A/m
10 V
10 V/m
30 A/m
1.7 m
1.7 m for 80 MHz to
800 MHz
3.25 m for 800 MHz to
2.7 GHz
Magnetic fields at power supply frequency should correspond to the values typical of those found in business and hospital environments.
EN
|
63
13 Appendix
13.5 Labels and symbols
13.5.1 Labels on the therapy device
1 2
64
|
EN
9 8 7 6 5
NO.
SYMBOL DESCRIPTION
TYPE PLATE ON THE RIGHT SIDE OF THE THERAPY DEVICE
SN Serial number of the therapy device
Year of manufacture
LABELS AND SYMBOLS ON THE THERAPY DEVICE
Consult instructions for use
Device inlet: inlet for room air at ambient temperature
Follow the instructions for use.
Slot for SD card
USB port
On/Off: indicates the on/off button
3
4
13 Appendix
NO.
SYMBOL DESCRIPTION
Device output: outlet for room air at 4 cmH
(depending on type of device)
2
O to 30 cmH
2
O
TYPE PLATE ON THE UNDERSIDE OF THE THERAPY DEVICE
TYPE: Type designation of the therapy device
37V 37 V DC
IP21
Degree of protection against solid foreign bodies. The device is protected against dripping water.
Degree of protection against electric shock: protection class II device
Do not dispose of device in household waste
Suitable for use in airplanes. Complies with RTCA/DO-160G chapter 21, Category M.
Type BF applied part
Manufacturer
CE mark (confirms that the product complies with the applicable
European directives).
13.5.2 Labels on the type plate of the power supply unit
SYMBOL DESCRIPTION
AC voltage
DC voltage
China RoHS label (confirms that the product does not emit toxic substances for the number of years indicated)
Only intended for indoor use.
EN
|
65
13 Appendix
SYMBOL DESCRIPTION
Degree of protection against electric shock: protection class II device.
Do not dispose of device in household waste.
IP21
CE mark (confirms that the product complies with the applicable
European directives).
IP protection class: Degree of protection against solid foreign bodies.
The device is protected against dripping water.
13.5.3 Labels on the therapy device packaging
SYMBOL
°C
DESCRIPTION
Permitted temperature for transport and storage: -25°C to
+70°C
25
93 % Permitted humidity for transport and storage: 15% to 93% relative humidity
15 %
13.5.4 Labels on the respiratory hose packaging
SYMBOL DESCRIPTION
For use on one patient only!
66
|
EN
13.6 Scope of supply
13.6.1 Standard scope of supply
A current list of the products included in delivery is available on the manufacturer's website or from your authorized dealer.
The following parts are supplied as standard:
PART
Basic device
Respiration hose
Power supply unit
ARTICLE NUMBER
Varies according to version of device
WM 24445
WM 29657
13 Appendix
PART
Power supply cable
Set, 2 air filters
Transport bag
Domed sticker with logo
SD card
Instructions for use
ARTICLE NUMBER
WM 24133
WM 29928
Varies according to version of device
WM 29899
WM 29794
WM 67841
13.6.2 Accessories
Accessories can be ordered separately, if required. A current list of the accessories is available on the manufacturer's website or from your authorized dealer.
13.6.3 Spare parts
Spare parts can be ordered separately, if required. A current list of the spare parts is available on the manufacturer's website or from your authorized dealer.
13.7 Warranty
Starting from the date of purchase, Löwenstein Medical offers the customer a limited manufacturer’s warranty on a new original Löwenstein Medical product or spare parts installed by Heinen + Löwenstein in accordance with applicable warranty terms and conditions for the particular product and the warranty periods listed below. The warranty conditions can be downloaded from the manufacturer's website. We can also send you the warranty conditions on request.
In the event of a claim under warranty, please contact your specialist dealer.
PRODUCT
Devices including accessories
(with the exception of: masks)
Masks, incl. accessories, batteries (unless otherwise stated in the technical documentation), sensors, hose systems
Disposable products
WARRANTY PERIODS
2 years
6 months
None
13.8 Declaration of conformity
The manufacturer, Löwenstein Medical Technology GmbH + Co. KG,
Kronsaalsweg 40, 22525 Hamburg, Germany, hereby declares that the product complies with the relevant regulations of Directive 93/42/EEC governing medical devices. The complete text of the Declaration of Conformity is available on the manufacturer's website.
EN
|
67
Löwenstein Medical
Technology GmbH + Co. KG
Kronsaalsweg 40
22525 Hamburg, Germany
T: +49 40 54702-0
F: +49 40 54702-461 www.loewensteinmedical.de
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 4 1.1 Intended use
- 4 1.2 Function
- 5 1.3 User qualifications
- 5 1.4 Indications
- 6 1.5 Contraindications
- 6 1.6 Side effects
- 7 2.1 Safety information
- 10 2.2 General information
- 11 2.3 Warnings in this document
- 12 3.1 Therapy device overview
- 13 3.2 Display
- 18 3.3 Accessories
- 19 4.1 Setting up the therapy device
- 19 4.2 Connecting up the power supply
- 23 5.1 Navigating the menu
- 23 5.2 Switching on the therapy device
- 25 5.3 Switching off the therapy device
- 25 5.4 Starting the therapy
- 26 5.5 Ending the therapy
- 27 5.6 Performing a mask test
- 28 5.7 Switching softSTART on/off
- 29 5.8 Setting the respiratory air humidifier
- 31 5.9 Setting the alarm
- 32 5.10 Viewing therapy data and device information
- 35 5.11 Using the SD card
- 38 6.1 Setting comfort parameters
- 39 6.2 Setting accessories parameters
- 40 6.3 Setting time parameters
- 40 6.4 Setting device parameters
- 42 7.1 General information
- 42 7.2 Cleaning intervals
- 43 7.3 Hygienic preparation of the therapy device
- 45 7.4 Hygienic preparation of the respiration hose
- 46 8.1 Intervals
- 46 8.2 Checking the therapy device
- 47 9.1 Alarms
- 49 9.2 Faults in the therapy device
- 50 9.3 Display messages
- 52 13.1 Technical Data
- 61 13.2 Emission of electromagnetic interference
- 61 13.3 Electromagnetic interference immunity
- 63 ME systems
- 64 13.5 Labels and symbols
- 66 13.6 Scope of supply
- 67 13.7 Warranty
- 67 13.8 Declaration of conformity