Woodpecker PT 3 Instruction Manual

Add to My manuals
16 Pages

advertisement

Woodpecker PT 3 Instruction Manual | Manualzz

Content

1. The installation and components of equipment ............................................... 1

2. Installation and adjustment .............................................................................. 4

3. Maintenance and sterilization ........................................................................... 4

4. Precaution ........................................................................................................ 7

5. After service ..................................................................................................... 9

6. Environmental protection ................................................................................. 9

7. Manufacturer’s right ........................................................................................ 9

8. Symbol instruction ........................................................................................... 9

9. European Authorised Representative .............................................................. 1 0

10. EMC-Declaration of conformity ................................................................... 1 0

11. Statement ...................................................................................................... 1 3

1. The installation and components of equipment

1.1 Instruction

Guilin Woodpecker Medical Instrument Co., Ltd. is a professional manufacturer in researching, developing and producing ultrasonic scalers. The product is mainly used for teeth cleaning and also an indisensable equipment for teeth disease prevention and treatment.

The ultrasonic periodontal treatment device PT 3 has scaling, perio functions with the following features: a) Circular vibration orbit, scaling and polishing together.

b) Small vibrating amplitude, enjoy the painless treatment.

c) Titanium scaling tips, no injury to cementum. ding d) Clinical solutions are applicable for automatic water supply mode, inclu-

Hydrogen peroxide, sodium hypochlorite, chlorhexidine e)LED Handpiece, LED mode and non-led mode available, better visibility.

f)Automatic frequency tracking ensures that the machine always works on the best frequency and more steadily.

g) Digital control, easy operation and more efficient for scaling.

h)The silicon-cover can be autoclaved to high temperature 134 pressure 0.22Mpa.

°C and high

1.2 Components

1.2.1 The components of machine are listed in the packing list.

1.2.2 Product performance and structural composition

Periodontal treatment device PT 3 is composed of electro circuit, water way, handpiece, periodontal treatment device tip and foot pedal.

1.2.3 Scope of application

Periodontal treatment device PT 3 is used for the dental calculus elimination.

1.3 The main technical specifications

1.3.1 Technical specifications of PT 3 a) Main unit input: 220-240V~ 50Hz/60Hz 150mA b) Output primary tip vibration excursion: ≤60μm c) Output half-excursion force: <2N d) Output tip vibration frequency: 28kHz±3kHz e) Output power: 3W to 20W f) Main unit fuse: T0.5AL 250V g) Water pressure: 0.01MPa to 0.5MPa

h) Weight of main unit: 2.0kg

i) Operating mode: Continuous operation j) Type of protection against electric shock: class II equipment k) Degree of protection against electric shock: Type BF applied part

1

l) Applied part of the equipment: handpiece and tip m) Degree of protection against harmful ingress of water: Ordinary equipment, the foot switch is drip-proof equipment (IPX1) n) Degree of safety of application in the presence of a Flammable Anesthetic

Mixture with air, Oxygen or Nitrous Oxide: Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.

1.4 Instruction of the main components

Instruction and component sketch map

1.4.1 Front of the main unit sketch map

Picture 1

1.4.2 Back of the main unit sketch map

Picture 2

2

1.4.3 Sketch map for connection of detachable handpiece. are showed in picture 3.

1.4.4 Instruction of using wrench picture 3

Picture 4

3

2. Installation and adjustment

2.1 Operation

2.1.1 Open the packing box, make sure that all the parts and accessories are complete according to the packing list, take the main unit out of the box, and put it on the the stable plane facing to the operator.

2.1.2 Turn the water control knob to the max according to the picture direction,

Do not screw it over tight in case of damage. [ note 1]

2.1.3 Insert the plug of the foot switch to its socket. (see picture 2)

2.1.4 Fill the box with water or connect one end of the water pipe to the water entrance, and the other end to the clean water source. (see picture 2)

2.1.5 Turn off the power switch and plug the power cord to socket

2.1.6 Turn on the power switch, the power indicator lighted and the machine is ready for work. Touch panel is applied to this machine, water supply mode or power can be adjusted by directly touching the water supply mode identification or power indicator on the touch panel.

2.1.7 In the condition of normal working, the frequency of the tips is very high, light touch and a certain to-and-fro motion will eliminate the tartar without obvious heating, overexetion and overatay are forbidden.

2.1.8 Vibrating intensity: Adjust the vibrating intensity according to your need, usually adjust to the middle grade, and adjust the vibrating during the clinical treatment according to the patient's sensitivity and the rigidity of the tartar.

2.1.9 Step on the foot switch, the tip begins to vibrate, and the LED lamp on the top of the handpiece shines. Release the foot switch, the LED lamp keep shining for 10 seconds.

2.1.10 Select different led mode according to clinical situation.

2.1.11 Water volume adjustment: Step on the foot switch, and the tip begins to vibrate, then turn the water control switch to fine spray to cool down the handpie ce and clean the teeth.

2.1.12 The handpiece can be handled in the same gesture as a pen in hand.

2.1.13 Make sure not to make the end of the tip touch the teeth vertically, and not use too much pressure when the tip touch the surface of the teeth, in order not to hurt the teeth and the tip.

2.1.14 After finishing operation, keep the machine working for 30 seconds with the water supply to clean the handpiece and the tip.

2.1.15 Unscrew the scaling tip and sterilize it.

Note: Don't screw the scaling tips when stepping on the foot switch, and the machine is working.

3. Maintenance and sterilization

3.1 All the scaling tips ,Handpiece,wrench can be autoclaved.

4

3.1.1 Autoclaved under high temperature, pressure, time:

134 °C , 2.0bar~2.3bar (0.20MPa~0.23MPa), 4min.

3.2 The tip and wrench can be cleaned by ultrasonic cleaner.

3.3 Troubles shotting

Fault Possible

The plug is in loose or

The scaling tip doesn’t vibrate wrong contact.

and no water flowing out when stepping on the foot

The foot switch is in loose contact.

switch.

The fuse is broken.

Solutions

Connect the power plug well.

Connect the switch well.

Change a new fuse.

The scaling tip is in loose contact.

The scaling tip doesn’t vibrate, but there is waterflowing out The connector plug of the when stepping on the foot switch.

handpiece with the circuit board is in loose contact.

Screw it tightly (see picture 4)

Contact with the local distributor or manufacturer.

Malfunction of the handpiece.

Contact with the local distributor or manufacturer.

The water control switch is Turn on the switch [note 1].

off.

The scaling tip vibrates but there is no spray come out when stepping on the foot switch.

There is water flow out when turn off the power.

The handpiece generates heat.

There is impurity in the solenoid valve.

Contact with the local distributor or manufacturer.

The water pipe is jammed. Clean water pipe by multi-

There is impurity in the solenoid valve.

The amount of spouting water is too little.

The potentiometer is broken.

function syringe [note 2].

Contact with the local distributor or manufacturer.

Turn the water control switch to a higher grade [note 1].

Change a new one.

5

Fault Possible Solutions

The water control knob is a low grade.

The amount of spouting water The water pressure is not is too little.

Turn the knob to a high grade

[note 1] .

Enhance the water pressure.

enough.

The water pipe is jammed. Clean the water pipe by multi-function syringe [note 2].

The vibration of the tip becomes weak.

The tip hasn't been screwed Screw the scaling tip tightly tightly or becomes loose because of vibration.

(picture 4).

The tip is damaged.[note 3] Change a new one.

The tip is damaged

[ note 3 ].

The vibrating intensity control The potentiometer is knob is seized up.

damaged.

Change a new one.

Contact with the local distributor or our company.

If the troubles still can't be solved, please contact with the local distributors or manufacturer.

3.4 Notice

[Note 1] The water control knob can adjust the water volume according to the image.

[Note 2] To clean the water pipe with the multi-function syringe of the dental unit ( see picture 4):

Picture 4 a) Snip the water pipe at a distance of 10cm to 20cm from the water entrance.

b) Turn on the power switch, get through to the power.

c) Connect the multi-function syringe of the dental unit to the water pipe.

6

d) Screw off the scaling tip or pull out the handpiece.

e) Step on the foot switch.

f) Turn on the switch of the multi-function syringe, press the air or water into the water pipe to clean and eliminate the impurity.

[Note 3 ] If the scaling tip has been screwed on tightly and there is fine spray too, the following phenomena show that the scaling tip is damaged: a) The vibrating intensity and the pulverization degree become weak obviously.

b) During operating, there is some buzz when the scaling tip is working.

3.5 Cleaning mode

It is suggested to wash and disinfect the water pipe after scaling everyday.

The “Cleaning mode” can wash and disinfect the water pipe, reducing the crystal substance and bacterias in water pipe.

Operation:

1.Put distilled water or mineral substance into water tank.

2.Press the automatically water supply button and outside water button at the same time (1s) to start the “Cleaning mode” after buzzer beeps. The button of automatic water supply mode would flashes and others buttons will go out.

3.Link the connector and the handpiece with the drainage device.

4.Step the pedal, the device would start self-cleaning. After that the pedal could be loosen.

5.After cleaning for 30s, the device would stop self-cleaning. Or u can stop by step the pedal again or press the automatic water supply button.

6.After cleaning, press the automatically water supply button and outside water button at the same time (1s) to exit the “Cleaning mode” after buzzer beeps.

4. Precaution

4.1 Usage notice

4.1.1 Keep the periodontal treatment device clean before and after operation.

4.1.2 The scaling tip, wrench and handpiece must be sterilized before each treatment.

4.1.3 Don't screw the scaling tip when stepping on the foot switch.

4.1.4 The scaling tip must be fastened. There must be fine spray coming out from the tip when operating.

4.1.5 Change a new one when the tip is damaged or worn excessively.

4.1.6 Don't twist or rub the tip.

4.1.7 While periodontal treatment device works, the heat of scaling tip will become higher if there is no water flowing out, please keep the water flow smoo thly.

4.1.8 Don’t use impure water source, please use pure water source, and normal brine can be used.

7

4.1.9 If use the water source without hydraulic pressure, the water surface should be one meter higher than the head of the patient.

4.1.10 Don't knock or rub the handpiece.

4.1.11 Please put the power plug into the socket and make it easy to pull out, to make sure it can be pulled out in emergency.

4.1.12 When using the equipment, please keep the water get through smoothly, otherwise patient's tooth surface would be injured by overheat in the handpiece.

4.1.13 After operating, turn off electrical source, and then pull out the plug.

4.1.14 As a professional manufacturer of medical instruments, we are only responsible for the safety on the following conditions:

I. The maintenance, repair and modification are made by the manufacturer or the authorized dealer.

II. The changed components are original of "WOODPECKER" and operated correctly according to instruction manual.

4.1.15 The screw thread of the scaling tips produced by other manufacturers maybe coarse, rusty and collapsed, which will damage the screw thread of the handpiece irretrievably. Please use "WOODPECKER" brand scaling tip.

4.1.16 Please select a suitable power when using different type of tips.

4.2 Contraindication

4.2.1 The patient who has hemophilia is not allowed to use this equipment.

4.2.2 The patient or doctor who with heart pacemaker is forbidden to use this equipment.

4.2.3 The heart disease patient, pregnant woman and children should be cautious to use the equipment.

4.3 Storage and maintenance

4.3.1 The equipment should be handled carefully and lightly. Be sure that it is far from the vibration, and is installed or kept in a cool, dry and ventilated place.

4.3.2 Don't store the machine together with the articles that are combustible poisonous, caustic, or explosive.

4.3.3 This equipment should be stored in a room where the relative humidity is

10% ~ 93%, atmospheric pressure is 70kPa to106kPa, and the temperature is -20 ℃

~ +55 ℃ .

4.3.4 Please turn off the power switch and pull out the power plug when the equipment is not used. If the machine is not used for a long time, please make it get through to the power and water once per month for five minutes.

4.4 Transportation

4.4.1 Excessive impact and shake should be prevented in transportation. Lay it carefully and lightly and don't invert it.

4.4.2 Don't put it together with dangerous goods during transportation.

4.4.3 Avoid solarization and getting wet in rain or snow during transportation.

4.5 Working condition

8

4.5.1 Environment temperature: +5 °C

4.5.2 Relative humidity: 30% ~ 75% to +40 °C

4.5.3 Atmosphere pressure: 70kPa to 106kPa

4.5.4 A temperature of the water at the inlet : not higher than +25 °C

5. After service

We offer one year free repair to the equipment according to the warranty card.

The repair of the equipment should be carried out by professional technician. We are not responsible for any irretrievable damage caused by the not professional person.

6. Environmental protection

Please dispose according to the local laws.

7. Manufacturer’s right

We reserve the rights to change the design of the equipment, the technique, fit tings, the instruction manual and the content of the original packing list at any time without notice. If there are some differences between blueprint and real equipment, take the real equipment as the norm.

8. Symbol instruction

Trademark

Anti-drip device

Alternating current

Date of manufacture

Ordinary equipment

Foot switch interface

Manufacturer

Class II equipment Type BF applied part serial number

Adjustment for the water flow

Water entrance pressure

0.01MPa - 0.5MPa

Handle with care

Used indoor only

Appliance compliance WEEE directive

Keep dry

Recovery

9

CE marked product

Can be autoclaved

Switch of automatic water supply

Earth (ground)

Mode of outside-water system

Mode of LED

Adjustment for the water flow, in Mode of auto-water system

Humidity limitation for storage

Atmospheric pressure for storage

Temperature limitation for storage

Authorised Representative in the EUROPEAN COMMUNITY

9. European Authorised Representative

10. EMC-Declaration of conformity

10

11

12

The device has been tested and homologated in accordance with EN 60601-

1-2 for EMC. This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electro magnetic environment.

11. Statement

All rights of modifying the product are reserved to the manufacturer without further notice. The pictures are only for reference. The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by WOO-

DPECKER, any copy or fake product must take legal responsibilities.

13

ZMN/WI-09-205 V1.7-20180913