Operation Method
Operation Mode and Function List
All the operation mode of this unit is as follows;
No. Name
1.
Fixed Temperature
Operation
Description
Pressing the FIXED TEMP key enters into the fixed temperature operation setting mode.
Pressing it again enters into the temperature setting mode.
The "▲▼" are used to set temperature.
Pressing the START/STOP key starts or stops operation.
Page
18
2.
Quick Auto Stop
Operation
3. Auto Stop Operation
4. Auto Start Operation
This operation is used to specify the period up to automatic stop during operation.
The period up to operation stop can be set by pressing the
TIMER key during fixed temperature operation.
The "▲▼" are used to set the time.
Pressing the START key starts the quick auto stop operation, activates the timer function and stops the operation automatically after specified period.
This operation is used to specify the automatic stop time in the fixed temperature operation.
Pressing the TIMER key displays "AStp".
The setting temperature "SV" can be set by pressing the
ENTER key.
The operation time "tim" can be set by pressing it again.
Pressing the START/STOP key starts the auto stop operation.
This operation is used to specify the period up to automatic start after power on.
Pressing the TIMER key displays "AStr".
The setting temperature "SV" can be set by pressing the
ENTER key.
The operation time "tim" can be set by pressing it again.
Pressing the START/STOP key starts the auto start operation.
This operation is used to change the temperature according to the setting temperature and time.
Pressing the PROGRAM key displays "PrG1".
Press it again to select the program mode.
Press the ENTER key to select the pattern "PA t".Press the
ENTER key to display "End". Input the number of patterns to be used.
Input the temperature and time of patterns "SV-n" and "t-n" respectively.
19
20
22
24
NOTE) This unit is impossible to be changed the mode during the operation. If the mode requires to be changed, stop the operation.
14
Operation Method
Operation Mode and Function List
The operation function of this unit is as follows;
No. Name
Auto overheating
prevention function
Description
This function is set to be automatically activated
(auto reset) when the temperature exceeds the setting temperature by 12℃.
Page
1.
3.
Overheating
prevention
function
Overheating
prevention device
2. Calibration offset function
Overheating prevention
temperature calibration
function
4. Recovery at power failure
5. Setting value locking
Though the device shares power source, display, and key input with the controller, it has independent temperature measurement circuit,
CPU, sensor and output circuit. Overheating prevention temperature can be set using the operation panel.
The unit stops operation when the device is activated. The unit starts operation again when the POWER switch is pressed again (manual reset).
17
This calibration offset function is for calibrating the difference occurred between the required in- furnace temperature and control temperature
(sensor temperature) of the controller. This unit can be calibrated toward either plus side or minus side of the whole temperature range.
The temperature of overheating prevention device is automatically corrected when the temperature of controller is collected.
The unit starts operation with the same condition as just before power failure if it occurs during operation. Press the START/STOP key to start the unit again.
This function locks the established operation status.
It can be set and cancelled with the SUBMENU key.
31
-
-
31
15
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 Cautions in Using with Safety
- 5 Explanation
- 6 Table of Illustrated Symbols
- 7 Fundamental Matters of "WARNING!" and "CAUTION!
- 8 Before Using this unit
- 8 Requirements for Installation
- 13 Description and Function of Each Part
- 13 Main Unit
- 16 Control Panel
- 17 Characters of the Controller
- 18 Operation Method
- 18 Operation Mode and Function List
- 20 Operation Mode, Function Setting Key, and Characters
- 21 Setting of Overheating Prevention Device
- 22 Fixed Temperature Operation
- 23 Quick Auto Stop Operation
- 24 Auto Stop Operation
- 26 Auto Start Operation
- 28 Program Operation
- 35 Other Functions
- 36 Handling Precautions
- 38 Maintenance Method
- 38 Daily Inspection and Maintenance
- 39 Long storage and disposal
- 39 When not using this unit for long term / When disposing
- 40 In the Event of Failure
- 40 Safety Device and Error Code
- 41 Trouble Shooting
- 42 After Service and Warranty
- 43 Specification
- 44 Wiring Diagram
- 48 Replacement Parts Table
- 49 Reference
- 49 List of Dangerous Substances