Troubleshooting. Medtronic Paradigm 522

Add to my manuals
176 Pages

advertisement

Troubleshooting. Medtronic Paradigm 522 | Manualzz

Chapter 9

Troubleshooting

This Troubleshooting chapter is designed to help you understand the messages that your pump is giving you when you get an alarm or an alert message. The procedures in the beginning of this chapter are to be used when you get specific alarms or cover a likely condition that might happen. A list of alarms is provided at the end of this chapter.

NOTE -

It is recommended that you read your warranty statement included with your pump for information on what is covered during your warranty period.

My pump has a no delivery alarm...

When a “No Delivery” alarm occurs, it means the pump is working correctly but it has detected that something is preventing insulin from being delivered. Your pump is not broken. Do the following steps:

1 Check your blood glucose and take an injection if needed.

2 Make sure that there is insulin in your reservoir and the tubing is not kinked. If these are all right, go to step 5 .

3 If necessary, unkink tubing. Clear the alarm by pressing ESC and ACT . A screen will appear with two choices: Resume and Rewind . Select Resume .

4 If the reservoir is empty, clear the alarm by pressing ESC and ACT . Select Rewind and change your reservoir and infusion set per the instructions in chapter 4, “Starting on insulin.”

5 Continue troubleshooting by disconnecting at the quick-disconnect, and set a 10 unit Fixed Prime.

6 Does insulin come out of the needle at the quick-disconnect? a.

If yes, change your entire infusion set per the instructions in chapter 4, “Starting on insulin.”

If NO insulin comes out of the needle at the quick-disconnect, or you receive another No

Delivery alarm, call the 24 Hour HelpLine.

b.

Your pump remembers the last fixed prime you deliver, so make sure to set your fixed prime amount back to your usual setting. To do this, deliver another fixed prime in your normal amount (the amount specified in your infusion set instructions).

7 Monitor your blood glucose closely.

Troubleshooting 121

8 If you followed these steps and you are still receiving a No Delivery alarm, call the 24 Hour HelpLine.

What happens if I leave the battery out too long?

If you leave the battery out too long (more than five minutes), you may receive a BATT OUT LIMIT alarm message when you install the new battery. Do the following steps:

1 Set your pump clock to the correct time, date, and year.

2 Check to make sure that all your settings, such as basal rate, are set as desired. If need be, reapply your last saved settings to the pump by using the Restore Settings option under User Settings in the

UTILITIES MENU as described in chapter 8 (you can only use this option if you have previously saved your pump settings).

3 Check the ALARM HISTORY screen and the STATUS screen for any alarms and/or alerts that may still need attention.

If you leave the battery out for more than three or four days, you may receive an A17 alarm and several

A47 alarms when you install a new battery. Do the following steps:

1 Clear all A17 and A47 alarms.

2 Set your pump clock to the correct time, date and year.

3 Check to make sure that all your settings, such as basal rate are set as desired. If need be, reapply your last saved settings to the pump by using the Restore Settings option under User Settings in the

UTILITIES MENU as described in chapter 8 (you can only use this option if you have previously saved your pump settings).

4 Check the ALARM HISTORY screen and the STATUS screen for any alarms and/or alerts.

Why doesn’t my pump battery last very long?

A short battery life does not mean that something is wrong with your pump. Battery life in your pump is variable and based on the conditions below:

The brand of battery you use (we recommend Energizer).

The storage and /or handling of the battery before use (avoid high or low temperatures).

The usage of pump in cold temperatures; this may shorten the battery life.

The usage of your pump, such as how often the buttons are pushed, the number of alerts/alarms and set changes.

The amount of insulin the pump is delivering.

➠ The usage of some features. The backlight, vibrate, remote control and/or meter options decrease battery life.

122 Chapter 9

What is a CHECK SETTINGS alarm?

This alarm occurs after an E alarm or after you clear your pump. It is advising you to make sure that all your settings are correct. A CHECK SETTINGS alarm occurs after any of these actions:

All user settings were cleared (set back to their defaults) because there was an E-error alarm.

The “Clear Settings” function was performed.

After you rewind when practicing without a reservoir when your first get your pump. In this case, it is just a reminder to make sure all your settings are programmed before you begin using the pump with insulin.

My screen appears distorted

The screen may appear distorted or have a “rainbow” appearance if you are wearing polarized sunglasses, are in bright sunlight, or in extreme high or low temperatures. If your screen appears distorted:

Take off your sunglasses.

Move into the shade.

➠ Make sure your pump is not in direct heat (i.e., next to a heater) or cold (worn on the outside of your clothing on a very cold day).

Do not return the pump: this is a normal property of this type of screen on any device.

I cannot get out of the priming loop

1 Is there a filled reservoir in the pump?

➠ If no, place a filled reservoir or shipping cap in the pump.

➠ If yes, make sure you are disconnected from the pump.

2 Hold the ACT button until the second set of beeps and the numbers appear on the screen.

If yes, your pump is okay, go to step 4 in the section “Manual prime” in chapter 4 to finish the

manual prime.

➠ If you did not hear a second set of beeps or numbers did not appear on the screen, change your infusion set and repeat this step.

3 If you still do not hear the beeps and see the numbers count up on the screen, call the

24 Hour HelpLine.

Troubleshooting 123

The pump is asking me to rewind

This is normal after any of the following:

1 Any E-alarms

2 Clear Settings function

3 No Delivery Alarm (during the Prime sequence)

My bolus stopped

The Bolus Stopped error can occur if the battery cap is loose or the pump was bumped or dropped during a bolus. It can also happen if the pump receives a static shock. As a safety measure, the pump stops the bolus when this happens.

1 If you dropped your pump, visually inspect it to make sure that it is not damaged in any way.

2 Review your bolus history and reprogram the remaining bolus, if needed.

My pump buttons are not acting right during a bolus

If the , or buttons are pressed and held down while a bolus is being delivered, the screen will freeze on that amount. Once the button is released, the units will ramp up to the amount delivered so far. Pressing and holding down the button will not stop the delivery of a bolus.

My pump won’t display my BG reading from my meter

1 Make sure that you are using the correct linked meter. Your pump will communicate with this meter only.

2 Make sure your meter is on (set to “snd”) and working correctly.

3 Make sure the meter option in your pump is set to On and you have set the meter ID number correctly on the pump.

4 Make sure your pump does not have a Low Battery alert condition.

5 Make sure the meter is within 4 feet (1.2 meters) of the pump without anything in between such as another person, a wall, etc.

6 Make sure there is no RF (radio frequency) interference from other electronic devices that could prevent communication. These devices can include some cell phones, cordless phones, televisions, computers, radios, other Paradigm pumps meters and pump remote controls. To restore communication, simply move away from these other types of devices, or turn them off.

7 Your pump will not show another reading. Make sure the pump is idle and the HOME screen is blank.

124 Chapter 9

8 If your pump still does not receive your BG reading from the linked meter, use the up/down buttons to manually enter your BG (in the ENTER BG screen).

I dropped my pump

Take care to protect your pump from being dropped.

1 Check that all connections are still tightly in place.

2 Check the LCD, keypad and pump case for cracks or damage.

3 Check infusion set, including the tubing connector and tubing for cracks or damage.

4 Review the status screen, basal rates and other pump settings.

5 Perform the Selftest procedure located in the UTILITIES MENU.

6 Call the Medtronic MiniMed 24 Hour HelpLine for assistance.

I submerged my pump in water

Your pump is designed to resist accidental contact with water. Do not submerge in water during bathing, swimming, or other water activities.

1 Pat the outside of the case until dry.

2 Open the reservoir compartment and check the compartment and reservoir for water. If wet, dry it completely within ten (10) minutes of exposure to water. Exposure to liquids, including water or insulin can corrode the mechanism.

3 Dry the reservoir completely—do NOT place a wet reservoir in the pump.

4 Do not use hot air to dry your pump. This may damage your pump's internal electronics.

5 Check the battery compartment and the battery—if wet, let them dry completely before using the pump.

6 Perform a selftest.

I cannot get to the User Settings menu

If you do not hold down while pressing ACT when you have User

Settings highlighted, you will see the following screen:

1 Go to the UTILITIES screen and highlight User Settings. Then, hold down

and press ACT.

2

See “User settings” on page 117 for information about the menu options.

This feature is not normally accessible. To access, consult your

User Guide.

Troubleshooting 125

Alert conditions

Your pump has a sophisticated network of safety checks and systems. If it detects an unusual condition that requires your immediate attention, it will beep or vibrate periodically to alert you. The pump will go into

Special mode (an open circle will display), and the backlight will illuminate.

➠ Dual Wave or Square Wave bolus delivery

➠ Pattern A or B is active

➠ Temp basal delivery

➠ Block feature is on

➠ Low reservoir

➠ Low battery

➠ Why are alerts important?

Your pump monitors activities and will alert you if a Special mode is active. Some alerts are a normal part of pump therapy, such as an active temporary basal. There are alerts that indicate a condition that is outside normal pump activity. For example, your pump notifies you with an alert when you need to replace the reservoir (LOW RESERVOIR) or replace your pump battery (LOW BATTERY).

What to do

When your pump beeps or vibrates notifying you that an alert condition exists:

1 Read and follow the instructions on the screen. Press ESC , ACT to silence an alert.

2 Check the STATUS screen to determine what caused the alert.

3 If the condition is due to a low battery, replace the battery.

4 If the condition is due to a low reservoir, monitor the reservoir volume frequently and change the reservoir when appropriate. Make sure you have a new reservoir, infusion set and vial of insulin with you.

Low reservoir alert

You can program the pump to sound an alert when either a specified number of units remains or a specified amount of time remains before the reservoir will be empty.

3:36P

LOW

RESERVOIR

ESC, ACT to clear

126 Chapter 9

Low Battery alert

If you get this alert, DO NOT go to sleep without replacing the battery.

The backlight, Remote control and meter functions are disabled during a

Low Battery condition. If the alert type is set to “vibrate,” the pump will change to the audio alert “beep-medium.” Clear ( ESC , ACT ) this alert before you replace your battery.

10:09A

LOW

BATTERY

Replace battery now

Use 1 AAA

ESC, ACT to clear

Alarms

Your pump has a sophisticated network of safety checks and systems. If the safety network detects anything unusual, your pump notifies you of conditions that require your immediate attention. The backlight illuminates the pump screen and the alarm/alert message displays on the screen.

NOTE -

The STATUS screen shows any alarms and alerts that are active.

➠ Why are alarms important?

Your pump monitors activities and notifies you if there is an unusual pump status or your attention is required. When an attention alarm is active, INSULIN DELIVERY IS STOPPED and immediate operator interaction is required.

If the vibrate mode is on, all alarms and alerts start as vibrations and then change to beeps. For your safety, if there is no response after ten

(10) minutes, the beeps change to a siren. The siren continues every minute until the alarm is cleared.

when a solid circle appears, follow the instructions on the screen.

Troubleshooting 127

What to do

When an alarm is triggered, the pump goes into Attention mode and an alarm message shows on the screen. The pump then defaults to the HOME screen. Do these steps when you get an alarm:

1 View the alarm:

From the HOME screen, press any button to see the alarm message.

(any button)

2 Read all of the alarm text. There are instructions on how to fix the alarm condition. (Press to read more text, if available.)

ALARM

MESSAGE

/

3 Clear the alarm:

Press ESC then ACT after you read the alarm instructions.

4 The HOME screen appears.

5 Follow the instructions that appeared with the alarm to fix the alarm condition.

6 Check your settings (i.e., time/date, basal, etc.) to make sure they are correct.

Alarm description and instructions show here.

ESC, ACT

128 Chapter 9

Alarm conditions

Alarms put the pump in “Attention” mode.

A (Alarm)

This alarm shows an “A” followed by two numbers. A-alarms cause all insulin delivery to stop. Your pump settings are retained. If this alarm repeats often, call the Medtronic MiniMed 24 Hour HelpLine for assistance.

9:42A

Pump reset

Settings preserved

ESC, ACT to clear

Axx

XX indicates the alarm number

Auto off

Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature, and so insulin delivery has been stopped.

11:17A

Delivery stopped

No buttons pushed during time limit

ESC, ACT to resume delivery

AUTO

OFF

Batt out limit

Occurs if the battery has been out of the pump for more than five minutes.

Verify that the pump time and date are correct. If not correct, go to the

UTILITIES MENU and reset the date and time.

12:00A

BATT

OUT LIMIT

Batt change too slow

ESC, ACT to clear

Bolus stopped

If this alarm occurs, it is very important to check bolus history to review how much of the bolus was actually delivered. If necessary, reprogram a bolus with the amount not yet delivered.

8:35A

BOLUS

STOPPED

Loose battery cap?

Pump dropped or bumped?

Check bolus history,

Reprogram bolus if required

ESC, ACT to clear

Button error

This error occurs if a button has been continually pressed for more than 3 minutes.

10:05A

BUTTON

ERROR

Button pressed for more than

3 min

ESC, ACT to clear

Troubleshooting 129

Check settings

When this alarm is active, you should check and/or reprogram your pump settings, including the time/date.

8:35A

CHECK

SETTINGS

Delivery stopped

Reprogram settings

ESC, ACT to clear

E (Error)

After receiving this alarm, note the error number then call the Medtronic

MiniMed 24 Hour HelpLine for assistance. An error alarm will show an “E” followed by two numbers. E-alarms cause all insulin delivery to stop, the pump resets, and all your settings are cleared.

Empty reservoir

There is no insulin in the reservoir. Change the reservoir immediately.

5:30A

Settings cleared

Reprogram settings

Call help-line for assistance

ESC, ACT to clear

Exx

XX indicates the error number

3:36P

EMPTY

RESERVOIR

Delivery stopped

Change reservoir

ESC, ACT to clear

Failed batt test

The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not used. If the battery does not have enough voltage, this alarm will occur. The pump will not function and the battery must be replaced. (Always make sure that you install a NEW battery into the pump.)

Is priming complete?

If you manually prime your pump with more than 30u insulin, this message appears. Press ESC , ACT to clear the message. If manual prime is complete, press ESC . If the manual prime is not complete, press and hold ACT until manual prime is complete.

8:35A

FAILED

BATT TEST

Delivery stopped

Replace battery now

Use 1 AAA

ESC, ACT to clear

IS PRIMING

COMPLETE?

Disconnect from body

See user guide for instruction

ESC, ACT to clear

130 Chapter 9

Max delivery

This alarm alerts you when you have taken more insulin than expected based on maximum bolus and maximum basal rates.

3:25P

MAX

DELIVERY

Exceeded 1 hour max delivery

Check BG

ESC, ACT to clear

Motor error

Insulin delivery has stopped. This alarm occurs if your pump detects a motor error.

3:25P

MOTOR

ERROR

Delivery stopped

Disconnect set

ESC, ACT to clear

No delivery

Insulin delivery has stopped. This alarm occurs if your pump detects a blockage.

No reservoir

The reservoir is not inserted correctly or no reservoir has been inserted.

11:17A

NO

DELIVERY

Delivery stopped

Change entire set

Check BG

See user guide to troubleshoot

ESC, ACT to clear

4:36P

NO

RESERVOIR

Delivery Stopped

Change reservoir set

ESC, ACT to clear

Off no power

The battery is dead. Replace battery immediately. Follow the directions on the screen. Check to make sure that the time is correct on the screen. Reset the time if necessary.

04:42

Use 1 AAA

ESC, ACT to clear

OFF

NO POWER

0% battery life

Delivery stopped

Replace battery now

Troubleshooting 131

Reset

The Reset alarm triggers when pump settings are cleared because of one of these reasons:

➠ Pump settings were cleared (Clear Settings function) and settings have not been reprogrammed.

➠ A download attempt from the PC is incomplete. (The download function is applicable to the optional software feature. Refer to the software user guide for more information.)

Weak Battery

The pump tests the voltage of each battery installed. If the battery voltage is less than full strength, this alarm may occur. The pump will operate normally, but the battery life will be shorter than expected.

Always make sure to install a NEW battery in the pump.

8:35A

RESET

Settings cleared by user

Reprogram settings

ESC, ACT to clear

11:17A

WEAK

BATTERY

Shorter battery life expected

ESC, ACT to clear

132 Chapter 9

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement

Table of contents