ROUBLESHOOTING
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.
Problem
No display.
Possible Cause
Power not connected.
Solution
Ensure the appropriate power cord is connected to your
S8 Escape and the power switch
(if available) is on.
Wait for air pressure to build up or change ramp time.
Insufficient air delivered from
S8 Escape.
Ramp Time is in use.
Excessive air delivered from the S8 Escape.
Air filter is dirty.
Air tubing is kinked or punctured.
Air tubing not connected properly.
Mask and headgear not positioned correctly.
Cap(s) missing from access port(s) on mask.
Altitude compensation setting is incorrect.
Pressure required for treatment may have changed.
Altitude compensation setting is incorrect.
Replace air filter.
Straighten or replace tubing.
Check air tubing.
Adjust position of mask and headgear.
Replace cap(s).
Adjust altitude compensation setting to the correct altitude range.
See your clinician to adjust the pressure.
Adjust altitude compensation setting to the correct altitude range.
Problem
Display error message:
Exxxx
Call Service
(where xxxx defines an error) or
MOTOR FAULT
Call Service or
UPLOAD LANGUAGES
Call Service or
DATA LOST
Call Service
The following message is displayed on the
LCD after you try to update settings using the Data Card:
Card Error
Remove Card
Possible Cause
Component failure.
The Data Card is not inserted correctly.
Solution
Call Service.
Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device.
Ensure that the device is switched on and in standby mode before you insert the Data Card.
The following message is displayed on the
LCD after you try to copy data to a Data
Card:
Card Error
Remove Card
The following message is displayed on the
LCD after you try to update the settings using the Data Card:
Settings Invalid
Remove Card
The Data Card is only partly inserted.
You may have removed the Data
Card before settings were copied to the S8 Escape.
The card is not a ResScan Data
Card.
The Data Card is not inserted correctly.
The card is not a ResScan Data
Card.
The identification details on the
Data Card do not match the details on your device.
Ensure that the Data Card is inserted into the slot on the Data
Card module as far as it will go.
Reinsert the Data Card and wait for the
Settings Success
Remove Card message to appear on the LCD.
Only use a ResScan Data Card.
Ensure that the Data Card is inserted with the arrow facing up and pointing toward the device.
Use a ResScan Data Card, as specified in this manual.
Contact your clinician immediately.
T ROUBLESHOOTING 27
28
Problem
The following message is displayed on the
LCD after you try to update the settings using the Data Card:
Settings Error
Remove Card
The following message is NOT displayed on the LCD after you try to update the settings using the Data Card:
Settings Success
Remove Card
Possible Cause
There is a data error on the Data
Card.
The settings were not updated.
Solution
Contact your clinician immediately.
Contact your clinician immediately.
advertisement
Related manuals
advertisement
Table of contents
- 1 S8 Escape™ System User's Manual English
- 2 Illustrations
- 4 Contents
- 6 Introduction
- 6 Your Responsibility
- 6 Medical Information
- 6 About the S8 Escape™
- 7 Contraindications
- 7 Warnings
- 8 Cautions
- 9 Adverse Effects
- 10 The S8 Escape System
- 10 S8 Escape
- 10 Front view (A-1)
- 10 Rear view (A-2)
- 10 Power cord (A-3)
- 10 Carry bag (A-4)
- 10 Air tubing
- 10 Masks
- 10 Nasal Masks
- 10 Nasal Pillows Systems
- 11 Full Face Masks
- 11 Humidifiers
- 11 ResScan Data Card Module
- 11 Accessories
- 12 How to Use the S8 Escape
- 12 Set Up the S8 Escape
- 12 Assemble the Mask
- 12 Start Treatment
- 13 Stop Treatment
- 14 How to Attach a Humidifier
- 14 Attach a HumidAire 3i Humidifier
- 14 Storing the HumidAire 3i
- 14 Attach a HumidAire Humidifier
- 15 Attach a ResMed Passover Humidifier
- 16 How to use the Keypad
- 16 LCD Screen
- 16 Keypad Keys
- 16 Backlights
- 17 How to Change the Settings on your S8 Escape
- 17 How to enter the menu screens
- 17 How to scroll through the menu
- 17 How to change a setting option
- 18 How to exit out of a menu
- 18 How to exit out of an item within the menu
- 18 How to Change the Ramp Time
- 18 How to change the Altitude setting
- 18 How to change the language setting
- 18 Other viewable information
- 19 Messages on the S8 Escape LCD
- 20 S8 Modules
- 20 Using a Blank Module
- 20 Using a ResScan Data Card Module
- 20 Attaching the Module
- 20 Removing the Module
- 20 Using the Data Card to Collect data
- 21 1 Insert the Data Card
- 21 2 Copy Data onto the Data Card
- 21 3 Remove the Data Card
- 21 4 Store the Data Card
- 21 5 Mail the Data Card to your Clinician
- 21 Using the Data Card to Update Settings
- 23 Cleaning and Maintenance
- 23 Daily Cleaning
- 23 Weekly Cleaning
- 23 Periodic Cleaning
- 24 Replacing the Air Filter
- 24 Servicing
- 25 Frequently Asked Questions
- 25 How will it feel when I am breathing with the S8 Escape?
- 25 What do I do when I want to get up in the night?
- 25 What do I do if I feel air leaking out of the mask?
- 25 Anything special I should do before putting the mask on?
- 25 My mouth feels dry during treatment-what can I do?
- 25 My nose is irritated during treatment-what can I do?
- 25 Can I travel with the S8 Escape?
- 26 Can I use my S8 Escape if mains (AC) power is not available?
- 27 Troubleshooting
- 30 System Specifications
- 30 S8 Escape
- 30 Accessories
- 30 Data Card
- 31 Data Card Module
- 31 All Devices (where applicable)
- 31 Symbols which appear on the device
- 32 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity
- 36 Limited Warranty
- 38 Index