Troubleshooting. ResMed S8 ESCAPE SYSTEM, S8

Add to My manuals
39 Pages

advertisement

Troubleshooting. ResMed S8 ESCAPE SYSTEM, S8 | Manualzz

26

T

ROUBLESHOOTING

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem

No display.

Possible Cause

Power not connected.

Solution

Ensure the appropriate power cord is connected to your

S8 Escape and the power switch

(if available) is on.

Wait for air pressure to build up or change ramp time.

Insufficient air delivered from

S8 Escape.

Ramp Time is in use.

Excessive air delivered from the S8 Escape.

Air filter is dirty.

Air tubing is kinked or punctured.

Air tubing not connected properly.

Mask and headgear not positioned correctly.

Cap(s) missing from access port(s) on mask.

Altitude compensation setting is incorrect.

Pressure required for treatment may have changed.

Altitude compensation setting is incorrect.

Replace air filter.

Straighten or replace tubing.

Check air tubing.

Adjust position of mask and headgear.

Replace cap(s).

Adjust altitude compensation setting to the correct altitude range.

See your clinician to adjust the pressure.

Adjust altitude compensation setting to the correct altitude range.

Problem

Display error message:

Exxxx

Call Service

(where xxxx defines an error) or

MOTOR FAULT

Call Service or

UPLOAD LANGUAGES

Call Service or

DATA LOST

Call Service

The following message is displayed on the

LCD after you try to update settings using the Data Card:

Card Error

Remove Card

Possible Cause

Component failure.

The Data Card is not inserted correctly.

Solution

Call Service.

Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device.

Ensure that the device is switched on and in standby mode before you insert the Data Card.

The following message is displayed on the

LCD after you try to copy data to a Data

Card:

Card Error

Remove Card

The following message is displayed on the

LCD after you try to update the settings using the Data Card:

Settings Invalid

Remove Card

The Data Card is only partly inserted.

You may have removed the Data

Card before settings were copied to the S8 Escape.

The card is not a ResScan Data

Card.

The Data Card is not inserted correctly.

The card is not a ResScan Data

Card.

The identification details on the

Data Card do not match the details on your device.

Ensure that the Data Card is inserted into the slot on the Data

Card module as far as it will go.

Reinsert the Data Card and wait for the

Settings Success

Remove Card message to appear on the LCD.

Only use a ResScan Data Card.

Ensure that the Data Card is inserted with the arrow facing up and pointing toward the device.

Use a ResScan Data Card, as specified in this manual.

Contact your clinician immediately.

T ROUBLESHOOTING 27

28

Problem

The following message is displayed on the

LCD after you try to update the settings using the Data Card:

Settings Error

Remove Card

The following message is NOT displayed on the LCD after you try to update the settings using the Data Card:

Settings Success

Remove Card

Possible Cause

There is a data error on the Data

Card.

The settings were not updated.

Solution

Contact your clinician immediately.

Contact your clinician immediately.