4.22. Gas time. Suunto EON Steel Black

Add to My manuals
71 Pages

advertisement

4.22. Gas time. Suunto EON Steel Black | Manualzz

Suunto EON Steel Black

• tx50/10, MOD 22m (pO

2

1.6) oxygen, MOD 6m

In the screenshot below, the menu has three gases and tx18/45 is selected as the active gas.

Even though only one gas is active, the decompression algorithm calculates ascent time

(during dive) using all these three gases.

To select another active gas before diving:

1. While in the Gases menu, press the middle button to see gas options.

2. Scroll with the upper or lower buttons to highlight Select.

3. Press the middle button again to confirm.

If you are diving with only one gas, ensure you have only that one gas in the Gases menu.

Otherwise, Suunto EON Steel Black expects you to use all gases in the list and notifies you to change gases during the dive.

NOTE: When you select CCR mode, gas mixtures are divided to open-circuit and closed-

circuit gases. See 4.28. Rebreather diving.

4.22. Gas time

Gas time refers to remaining air (gas) left with current gas mixture, measured in minutes. The time is based on tank pressure value and your current breathing rate.

Gas time is also highly dependent on your current depth. For example, all other factors being the same, including breathing rate, tank pressure and tank size, depth affects gas time as follows:

At 10 m (33 ft, surrounding pressure 2 bar), gas time is 40 minutes.

At 30 m (99 ft, surrounding pressure 4 bar), gas time is 20 minutes.

At 70 m (230 ft, surrounding pressure 8 bar), gas time is 10 minutes.

Gas time information is visible in Air/Nitrox and Trimix dive modes by default. If you have not paired a Suunto Tank POD, the gas time field shows n/a. If you have paired a POD but there is no data being received, the field shows –. This may be because the POD is not in range, the tank is closed, or the POD battery is low.

NOTE: Set tank size under the Gases menu to get the gas consumption and the gas time calculations right.

NOTE: Gas time reading shows zero and turns red if tank pressure goes below 35 bar

(500 psi).

35

advertisement

Related manuals

advertisement

Table of contents