Safety MLC M Series
4.2.5 Safety Instructions for Cleaning and Disinfection of the Device
WARNING
Risk of electrical shock to the user caused by presence of potentially conductive liquids:
Prior to any cleaning and disinfection procedure, the power cable must be disconnected from the power supply at all times.
WARNING
Risk of unintended operation for touchscreen-equipped devices:
Before cleaning and / or disinfecting your device, turn off the touchscreen via the Touchscreen
Enable / Disabl e button (23) (see S ection Error! Reference source not found.
Operational elements).
WARNING
Risk of electrical shock to the user and device damage caused by inappropriate processing procedures:
The device is not suited for any kind of mechanical and / or thermal decontamination procedures such as autoclaving, steam sterilization, etc.
WARNING
Risk of electrical shock to the user and device damage caused by excessive amounts of liquids:
Never apply excessive amounts of liquids to your device. Do not immerse it in liquids of any substance.
Be conservative with all cleaning and disinfection solutions used to clean the device and never apply liquids directly to openings such as I/O interface ports.
CAUTION
Risk of device damage caused by usage of abrasive materials:
Your ADLINK device provides a surface finish that is sufficiently protected for the intended purpose.
However, do not use any abrasive materials for cleaning processes, e.g. brushes, sponges, or other objects with hard and / or rough characteristics. Doing so may lead to damage to the front side, the touch surface, or the housing surface. Never use e.g. sponges made of metal grafts or similar, or sharp/hard objects like knives, pencils etc. for cleaning purposes. Thoroughly follow all instructions provided in this chapter.
advertisement
Related manuals
advertisement
Table of contents
- 4 Contents
- 6 List of Figures
- 7 Revision History
- 8 General Information
- 8 About this Manual
- 9 Trademarks
- 9 Warranty Information
- 9 Support, Problems and Failure Analysis
- 9 Recommended action on problems or failures
- 9 Procedure for returning products
- 10 Drivers
- 10 Conventions Used in this Manual
- 10 Conventions for signal words and symbols
- 10 Conventions for different user groups
- 11 Safety
- 11 Intended Purpose
- 11 Intended purpose of MLC M used as stand-alone device
- 13 Safety and Warning Instructions
- 13 General Electrical Hazards
- 15 Specific Hazards Caused by Electrical Leakage in Medical Devices
- 15 Other Hazards
- 19 Hazards Regarding Repair and Maintenance
- 20 Safety Instructions for Cleaning and Disinfection of the Device
- 21 Product Identification Label
- 22 User Information
- 22 General Description
- 23 Model Variants
- 23 Device Models with Optional DC Power Supply
- 24 Device Overview
- 24 Interfaces
- 26 Operational Elements
- 27 Front panel buttons
- 29 Cleaning and Disinfection
- 30 Safety instructions for cleaning and disinfection of the device
- 31 Device Options and Accessories
- 32 Administrator Information
- 32 Functional Block Diagram
- 33 Technical Data of Main System Components
- 34 Detailed System Components
- 35 Optional Built-In Accessories
- 35 WLAN / Bluetooth module (optional)
- 35 RFID Option
- 36 Maintenance and Service