advertisement
Warranty Information
Guidance and manufacture’s declaration – electromagnetic immunity
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test Compliance level level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz/60Hz) magnetic field IEC
61000-4-8
±6 kV contact
±8 kV air
3A/m
±6 kV contact
±8 kV air
3A/m
Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
41
Warranty Information
Immunity test
Radiated RF
IEC 61000-4-3
IEC 60601 test Compliance level level
3A/m
80 MHz to 2.5
GHz
3A/m
Electromagnetic environment - guidance survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Digital Blood Pressure
Monitor is used exceeds the applicable RF compliance level above, the Digital Blood
Pressure Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
Digital Blood Pressure Monitor.
43
Warranty Information
Guidance and manufacture’s declaration – electromagnetic immunity
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test Compliance level level
Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-3
3A/m
80 MHz to 2.5
GHz
3A/m
Portable and mobile RF communications equipment should be used no closer to any part of the Digital Blood Pressure Monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
42
Warranty Information
Recommended separation distances between portable and mobile RF communications equipment and the Digital Blood Pressure Monitor.
Pressure Monitor is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Digital Blood Pressure
Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Digital
Blood Pressure Monitor as recommended below, according to the maximum output power of the communications equipment.
output power of transmitter
Separation distance according to frequency of transmitter
(m)
80 MHz to 800 MHz
80 MHz to 800 MHz
0.01
0.1
1
0.117
0.369
1.167
0.233
0.738
2.333
10
100
3.689
11.667
7.379
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
44
advertisement
Related manuals
advertisement
Table of contents
- - Introduction
- - Symbols Use
- - Important Safety Notes
- - Konw Your Unit
- - IHB Introduction
- - Before Taking A Measurement
- - Battery lnstallation/Replacement
- - 1. To set year
- - 2. To set date and time
- - 3. To set positioning sensor
- - 4. To set displaying of the date and time
- - 5. To set unit
- - 7. To set user
- 1 Proper Way of Measurement
- 2 Take a Measurement
- 5 To read the average value
- 6 To read the measurement value
- 7 To delete the memory value
- 8 Blood Circulation
- 9 Health and Blood Pressure
- 10 Classification of Blood Pressure
- 12 Error Indicators
- 9 Eliminate Fault
- 9 Care
- 9 Maintenance
- 10 Statement