Omega LDP6-CDS Owner Manual

Add to My manuals
8 Pages

advertisement

Omega LDP6-CDS Owner Manual | Manualzz

User’s Guide

Shop online at

omega.com e-mail: [email protected]

For latest product manuals: omegamanual.info

6-C DSS

LP0683X

OMEGAnet

®

Online Service omega.com

Internet e-mail [email protected]

Servicing North America:

U.S.A.:

ISO 9001 Certified

Canada:

One Omega Drive, P.O. Box 4047

Stamford, CT 06907-0047

TEL: (203) 359-1660 e-mail: [email protected]

976 Bergar

Laval (Quebec) H7L 5A1, Canada

TEL: (514) 856-6928 e-mail: [email protected]

FAX: (203) 359-7700

FAX: (514) 856-6886

For immediate technical or application assistance:

U.S.A. and Canada:

Sales Service: 1-800-826-6342/1-800-TC-OMEGA ®

Customer Service: 1-800-622-2378/1-800-622-BEST

®

Engineering Service: 1-800-872-9436/1-800-USA-WHEN ®

Mexico:

En Español: (001) 203-359-7803

FAX: (001) 203-359-7807 e-mail: [email protected]

[email protected]

Servicing Europe:

Czech Republic:

Frystatska 184, 733 01 Karviná, Czech Republic

TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114

Toll Free: 0800-1-66342 e-mail: [email protected]

Germany/Austria:

Daimlerstrasse 26, D-75392 Deckenpfronn, Germany

TEL: +49 (0)7056 9398-0

Toll Free in Germany: 0800 639 7678 e-mail: [email protected]

FAX: +49 (0)7056 9398-29

United Kingdom:

One Omega Drive, River Bend Technology Centre

ISO 9002 Certified Northbank, Irlam, Manchester

M44 5BD United Kingdom

TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622

Toll Free in United Kingdom: 0800-488-488 e-mail: [email protected]

It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New

Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.

The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.

WARNING : These products are not designed for use in, and should not be used for, human applications.

DESCRIPTION

This bulletin serves as a guide for the installation, configuration and operation of LDP6-CDS Setpoint cards. The setpoint cards are available as dual relay, quad relay, quad sourcing transistor, or quad sinking transistor outputs. Only one setpoint card can be installed at a time.

The meter can be fitted with up to three option cards. The slot bays of the option cards are dedicated to a particular card function. The option card functions are: serial communications, analog output and setpoint output. Only one card from each function category can be installed.

INSTALLING AN OPTION CARD

Caution: The option and main circuit cards contain static sensitive components. Before handling the cards, discharge static charges from your body by touching a grounded bare metal object. Ideally, handle the cards at a static controlled clean workstation. Also, only handle the cards by the edges. Dirt, oil or other contaminants that may contact the cards can adversely affect circuit operation.

Warning: Exposed line voltage exists on the circuit boards.

Remove all power to the meter AND load circuits before accessing the unit.

1. Remove the main assembly from the rear of the case. Squeeze the finger holds on the rear cover, or use a small screwdriver to depress the side latches to release it from the case. It is not necessary to separate the rear cover from the main circuit card.

2. Locate the option card connector for the type of option card to be installed. Hold the unit by the rear connector, not the display board, when installing an option card.

3. Install the option card by aligning the option card connector with the slot in the rear cover. The cards are keyed by position with different main board connector locations. Be sure the connector is fully engaged and the tab on the option card rests in the alignment slot on the display board.

4. Slide the assembly back into the case. Be sure the rear cover latches fully into the case.

5. Apply the option card label to the bottom side of the meter. Do not cover

the vents on the top surface of the meter. The surface of the case must be clean for the label to adhere properly. Apply the label to the area designated by the large case label.

TOP VIEW

REAR TERMINAL CONNECTIONS

3

SPECIFICATIONS

Setpoint Output Cards : Four types of field installable cards

Response Time: 200 msec. max. to within 99% of final readout value

(digital filter and internal zero correction disabled) 700 msec. max.

(digital filter disabled, internal zero correction enabled)

DP63700 only: 1 sec. max. to within 99% of final readout value (digital filter disabled)

DP63500 only: 200 msec. typ.; 700 msec max. (digital filter disabled)

DP63700 Isolation For All Four Cards:

Isolation To Sensor Common: 1400 Vrms for 1 min.

Working Voltage: 125 V

Isolation To User Input Common: 500 Vrms for 1 min.

Working Voltage: 50 V

Dual Relay Card: LDP6-CDS10

Type: Two FORM-C relays

Isolation To Sensor & User Input Commons: 2000 Vrms for 1 min.

Working Voltage: 250 V

Contact Rating:

One Relay Energized: 5 amps @ 120/240 VAC or 28 VDC (resistive load), 1/8 HP @120 VAC, inductive load

Total Current With Both Relays Energized not to exceed 5 amps

Life expectancy: 100 K cycles min. at full load rating. External RC snubber extends relay life for operation with inductive loads

Quad Relay Card: LDP6-CDS20

Type: Four FORM-A relays

Isolation To Sensor & User Input Commons: 2300 Vrms for 1 min.

Working Voltage: 250 V

Contact Rating:

One Relay Energized: 3 amps @ 250 VAC or 30 VDC (resistive load),

1/10 HP @120 VAC, inductive load

Total Current With All Four Relays Energized not to exceed 4 amps

Life Expectancy: 100 K cycles min. at full load rating. External RC snubber extends relay life for operation with inductive loads

Quad Sinking Open Collector: LDP6-CDS30

Type: Four isolated sinking NPN transistors.

Isolation To Sensor & User Input Commons: 500 Vrms for 1 min.

Working Voltage: 50 V. Not Isolated from all other commons.

Rating: 100 mA max @ V

SAT

= 0.7 V max. V

MAX

Quad Sourcing Open Collector: LDP6-CDS40

= 30 V

Type: Four isolated sourcing PNP transistors.

Isolation To Sensor & User Input Commons: 500 Vrms for 1 min.

Working Voltage: 50 V. Not Isolated from all other commons.

Rating: Internal supply: 24 VDC ± 10% , 30 mA max. total all four

External supply: 30 VDC max., 100 mA max each output

Sourcing Output Logic Card

Sinking Output Logic Card

Quad Sourcing Open Collector Output Card Supply Select

For Quad Sourcing Plug-in Card (LDP6-CDS40), set the jumper for internal or external supply operation before applying power.

MODULE 6 - Setpoint (Alarm) Parameters (

PARAMETER MENU

)

Indicates Program Mode Alternating Display.

Parameter

«

ª

Selection/Value

Factory Settings are shown.

4

ª

SETPOINT SELECT

« NO SP-1

SP-3

SP-2

SP-4

Enter the setpoint (alarm output) to be programmed. The in the following parameters will reflect the chosen setpoint number. After the chosen setpoint is completely programmed, the display will return to . Repeat step for each setpoint to be programmed. The chosen at will return to . The number of setpoints available is setpoint output card dependent.

ª

SETPOINT ACTION

« OFF dE-HI

Ab-HI

dE-LO

Ab-LO bANd

AU-HI totLo

AU-LO totHI

Enter the action for the selected setpoint (alarm output). See

Setpoint Alarm Figures for a visual detail of each action.

=

=

=

=

=

=

=

Setpoint always off, (returns to SPSEL NO)

Absolute high, with balanced hysteresis

Absolute low, with balanced hysteresis

Absolute high, with unbalanced hysteresis

Absolute low, with unbalanced hysteresis

Deviation high, with unbalanced hysteresis *

Deviation low, with unbalanced hysteresis *

= Outside band, with unbalanced hysteresis *

= Lower Totalizer absolute high, unbalance hysteresis**

= Upper Totalizer absolute high, unbalance hysteresis**

* Deviation and band action setpoints are relative to the value of setpoint 1.

It is not possible to configure setpoint 1 as deviation or band actions. It is possible to use setpoint 1 for an absolute action, while its value is being used for deviation or band.

** The lower Totalizer action allows setpoints to function off of the lower 5 digits of the Totalizer. The upper Totalizer action allows setpoints to function off of the upper 4 digits of the Totalizer. To obtain absolute low alarms for the Totalizer, program the logic as reverse.

or output

ª

ª

«

SETPOINT VALUE

-19999 to

99999

Enter desired setpoint alarm value. These setpoint values can also be entered in the Display Mode during Program Lockout when the setpoint is programmed as in Parameter

Module 3. When a setpoint is programmed as deviation or band acting, the associated output tracks as it is changed.

The value entered is the offset, or difference from .

HYSTERESIS VALUE

« 1 to

65000

Enter desired hysteresis value. See Setpoint Alarm Figures for visual explanation of how setpoint alarm actions (balance and unbalance) are affected by the hysteresis. When the setpoint is a control output, usually balance hysteresis is used. For alarm applications, usually unbalanced hysteresis is used. For unbalanced hysteresis modes, the hysteresis functions on the low side for high acting setpoints and functions on the high side for low acting setpoints.

Note: Hysteresis eliminates output chatter at the switch point, while time delay can be used to prevent false triggering during process transient events.

Setpoint Alarm Figures

With reverse output logic , the below alarm states are opposite.

Absolute High Acting (Balanced Hys) = Ab-HI Absolute Low Acting (Unbalanced Hys) = AU-Lo

Band Outside Acting = bANd

Absolute Low Acting (Balanced Hys) = Ab-Lo

Deviation High Acting (SP > 0) = dE-HI Deviation High Acting (SP < 0) = dE-HI

Absolute High Acting (Unbalanced Hys) =

This is also for Totalizer alarms: ,

AU-HI

Deviation Low Acting (SP > 0) = dE-Lo

5

Deviation Low Acting (SP < 0)= dE-Lo

ª

ª

ª

ª

ª

«

ON TIME DELAY

0.0 to

3275.0

Sec

Enter the time value in seconds that the alarm is delayed from turning on after the trigger point is reached. A value of

0.0 allows the meter to update the alarm status per the response time listed in the Specifications. When the output logic is , this becomes off time delay. Any time accumulated at power-off resets during power-up.

OFF TIME DELAY

«

0.0

to

3275.0

Sec

Enter the time value in seconds that the alarm is delayed from turning off after the trigger point is reached. A value of

0.0 allows the meter to update the alarm status per the response time listed in the Specifications. When the output logic is , this becomes on time delay. Any time accumulated at power-off resets during power-up.

OUTPUT LOGIC

«

nor rEv

Enter the output logic of the alarm output. The logic leaves the output operation as normal. The logic reverses the output logic. In , the alarm states in the

Setpoint Alarm Figures are reversed.

RESET ACTION

«

Auto LAtC1 LAtC2

Enter the reset action of the alarm output.

= Automatic action; This action allows the alarm output to automatically reset off at the trigger points per the

Setpoint Action shown in Setpoint Alarm Figures. The “on” alarm may be manually reset (off) immediately by a front panel function key or user input.The alarm remains reset off until the trigger point is crossed again.

= Latch with immediate reset action; This action latches the alarm output on at the trigger point per the

Setpoint Action shown in Setpoint Alarm Figures. Latch means that the alarm output can only be turned off by front panel function key or user input manual reset, serial reset command or meter power cycle. When the user input or function key is activated (momentary or maintained), the corresponding “on” alarm output is reset immediately and remains off until the trigger point is crossed again.

(Previously latched alarms will be off if power up Display

Value is lower than setpoint value.)

= Latch with delay reset action; This action latches the alarm output on at the trigger point per the Setpoint

Action shown in Setpoint Alarm Figures. Latch means that the alarm output can only be turned off by front panel function key or user input manual reset, serial reset command or meter power cycle. When the user input or function key is activated (momentary or maintained), the meter delays the event until the corresponding “on” alarm output crosses the trigger off point. (Previously latched alarms are off if power up Display Value is lower than setpoint value. During a power cycle, the meter erases a previous Latch 2 reset if it is not activated at power up.)

STANDBY OPERATION

« no YES

When , the alarm is disabled (after a power up) until the trigger point is crossed. Once the alarm is on, the alarm operates normally per the Setpoint Action and Reset Mode.

ª

ª

«

SETPOINT ANNUNCIATORS

OFF nor rEv FLASH

The mode disables display setpoint annunciators. The mode displays the corresponding setpoint annunciators of “on” alarm outputs. The mode displays the corresponding setpoint annunciators of “off” alarms outputs. The mode flashes the corresponding setpoint annunciators of “on” alarm outputs.

PROBE BURN-OUT ACTION (DP63500 ONLY)

«

ON OFF

Enter the probe burn-out action. In the event of a temperature probe failure, the alarm output can be programmed to go on or off.

Setpoint Alarm Reset Actions

Alternate Setpoints

An Alternate list of setpoint values can be stored and recalled as needed. The

Alternate list allows an additional set of setpoint values. (The setpoint numbers nor rear terminal numbers will change in the Alternate list.) The Alternate list can only be activated through a function key or user input programmed for in Module 2. When the Alternate list is selected, the Main list is stored and becomes inactive. When changing between Main and Alternate, the alarm state of Auto Reset Action alarms will always follow their new value. Latched “on” alarms will always stay latched during the transition and can only be reset with a user input or function key. Only during the function key or user input transition does the display indicate which list is being used.

6

WARRANTY/DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 25 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal two (2) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.

If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service

Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge.

OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.

OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by

OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR

REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF

TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY

AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF

LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of

OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic

Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS/INQUIRIES

Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE

RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)

NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING

DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.

The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.

FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:

1. Purchase Order number under which the product was PURCHASED,

2. Model and serial number of the product under warranty, and

3. Repair instructions and/or specific problems relative to the product.

FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:

1. Purchase Order number to cover the COST of the repair,

2. Model and serial number of the product, and

3. Repair instructions and/or specific problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.

OMEGA is a registered trademark of OMEGA ENGINEERING, INC.

© Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.

Where Do I Find Everything I Need for

Process Measurement and Control?

OMEGA…Of Course!

Shop online at omega.com

TEMPERATURE

]

]

]

]

]

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers

PRESSURE, STRAIN AND FORCE

]

]

]

]

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories

FLOW/LEVEL

]

]

]

]

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers

pH/CONDUCTIVITY

]

]

]

] pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment

DATA ACQUISITION

]

]

]

]

]

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters

HEATERS

]

]

]

]

]

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters