Mettler Toledo LC-P45 Operating Instructions Manual
Add to my manuals
36 Pages
Mettler Toledo LC-P45 is a versatile dot matrix printer for normal paper that provides advanced print functions, enabling you to supplement weight data with date, time, lot identification, balance identification, and current sample number. It can generate statistics of weight series, totalize weight values, multiply or divide weight by a factor, check balance accuracy with a test weight, and calibrate the balance with a calibration weight.
advertisement
Operating instructions
METTLER TOLEDO
LC-P45 Printer
11.01.2000 15:28:33
ID 71-00.6-34
1 206.55 g
2 3165.92 g
3 45.73 g
4 2.54 g
Sum 3420.74 g
---------- END ---------
7
Def
4
ID
1
F
LC-P45 Printer
Result
2
Code
5
3
Menu
8
1 2
3
6
31
9
Menu
0
Clear
METTLER TOLEDO
2
Overview
What are the capabilities of the LC-P45 Printer → Page 3
Inserting the batteries and paper, attachment, printing → Page 5
Setting date and time → Page 6
Selecting applications → Page 7
Configuring the printer in the menu → Page 10
Entering values of variables (e.g. lot identification ID) → Page 12
Selecting printer function → Page 13
Printing out weight values → Page15
Generating statistics of a weight series → Page16
Totalizing weight values → Page 17
Multiplying (dividing) the weight by a factor → Page 18
Recording the moisture determination with the LP16/LJ16 IR Dryer → Page 19
Pipette test for calibrating volumetric dosing apparatus → Page 19
Recording the processes with M-Pac application software packages → Page 21
Checking a balance with a test weight → Page 22
Calibrating (adjusting) a balance with a calibration weight → Page 23
Changing paper and ribbon → Page 24
Maintenance, Service and Cleaning → Page 25
Interpreting the error messages → Page 26
Accessories for GA45 Printer (paper, ribbon, etc.) → Page 28
Attachment to METTLER TOLEDO instruments, additional hardware, available functions → Page 30
1. Introduction
1.1
What are the capabilities of the LC-P45 Printer?
The LC-P45 is a versatile dot matrix printer with advanced print functions for normal paper. It can supplement the weight value with date, time and other data. It has two LocalCAN interfaces and one RS-232-C interface and is primarily suitable for attachment to METTLER TOLEDO balances and scales equipped with the standard software, but will also print out the records of the METTLER TOLEDO application software
Pacs (M-Pacs) and of the LP16 and LJ16 METTLER TOLEDO Infrared Dryers.
The various print functions of the LC-P45 can be selected in the dialog mode (English, German, French, Spanish or Italian). The settings are stored in a fail-safe manner thanks to battery buffering.
• Print weight
• Insertion of - Date and time
- Lot identification (ID)
- Balance identification
- Current sample No.
The LC-P45 prints out the weight displayed on the balance directly.
For users complying with quality management regulations such as GLP or GMP or for those who wish for a detailed record of a weighing. You can supplement the weighing result by automatically printing out any of the adjacent data.
- Sample code
• Generate statistics
14:53:32
7.4.5-22.03
11.01.2000
ID
3468.08 g
Pages 7/8
• Totalization
The sample code, a comment for a more detailed definition of the weighing sample, is entered directly for every sample.
Without the need to purchase peripherals or software, the LC-P45 automatically generates statistics of the weight values of your sample series and calculates the most important statistical parameters.
The LC-P45 can also totalize weight values, thus calculating and documenting running totals.
This function is especially useful when weighing in formulations.
3
• Multiplication/division
• Check balance and calibrate
• LP16/LJ16 Infrared Dryer
• Pipette check
4
This function allows weight values to be multiplied or divided by any desired factors, Typical examples are calculating prices based on a price per unit weight or converting weights into volumes.
A check on the accuracy of a balance by means of a test weight is fully documented.
Likewise, adjustment of the balance by means of a calibration weight.
Date, time and lot identification (ID) are automatically added to the record.
This function enables pipettes and other volumetric measuring and dosing apparatus to be calibrated by a simple method which is also completely documented.
2. Startup procedure
2.1
Cautionary notes
For reasons of safety, the housing of the printer must not be opened by the user. If technical faults appear or if the printing unit or power line fuse has to be changed, please notify your responsible METTLER TOLEDO service.
• The LC-P45 Printer may be used only indoors and on no account in hazardous areas.
• The electronics of the LC-P45 Printer are protected against the ingress of splashing water and dust. However, as the printer housing is not tight the printer should not be used in the presence of liquids.
• Comply with cleaning instructions (see page 25), protect power plug against wet conditions.
• Never open the printer housing. Should you ever have problems with your printer, please contact your responsible METTLER TOLEDO dealer.
Def
7
5
Menu
8
9
0
Def
7
5
Menu
8
9
0 balance
2.2
Inserting the batteries
Two batteries are provided and supply the clock and the memories for the variable settings of the printer when it is not connected to the power supply.
Please note: To avoid loss of any printer settings you may have made, attach the LC-P45
Printer to the power supply when changing the batteries.
• Lift up paper cover and pull paper backward out of the printing unit
• Open battery compartment and insert batteries
Inserting paper
• Feed paper through the slot in the printing unit in the direction of the arrow
• Switch on printer (toggle switch at right rear, the number of the software version will be printed out automatically.)
• Press and hold until sufficient paper has been fed in
Attaching printer, printing
• Plug the cable supplied into the LC-P45 and the I/O data interface of a balance or instrument. The printer is now ready for operation!
• Print out the weight value of the balance
If no printout appears: Set both devices to the factory setting! (The interface parameters are then matched. Green pilot lamp flashes, see page 26; available functions, see page 30)
8
Menu
0
Menu set the LC-P45 to the factory setting
5
7
Def
1
31
4
7
Def
1
3
3
....
3
....
5
6
2.3
Setting date and time
Setting or changing date and time to local time.
Date (EURO or US version)
• Start date definition
The input process can be chek– ked in the balance display
Correction of wrong entry
Clear
• Entry of the date, e.g.
23.01.97 (EURO format, day.month.year) or
1-23-97 (US format, month-day-year)
• Concluding definition
Please note: Separation of the numbers with EURO by " . ", with US by "-"
Time (24-hour format)
• Start time definition
• Entry of the current time, e.g. 13.46.05 (hours.minutes.seconds)
• Concluding definition
Note:
3 prints out date and time directly at all times
3. Selecting applications
Printing out weight values
1
Page 15
3468.08 g
Print out weight by itself
1
11.01.2000
14:53:32
3468.08 g
Date + time + weight (ID set to 0)
ID 04-83-18.5
3468.08 g
2
Lot identification (ID) + weight
11.01.2000
ID 04-93-18.5
+ weight
3468.08 g
Date + lot identification
3
11.01.2000
14:53:32
ID 04-93-18.5
3468.08 g
4
Date + time + lot identification
+ weight
11.01.2000
14:53:32
ID 04-83-18.5
STANDARD V10.50.00
TYPE : PM6100
INR : 911375
3468.08 g 5
Complete documentation following
GLP (Good Laboratory Practice)
14
15
16
17
0.49 g
55.61 g
64.03 g
123.97 g
Current sample number
+ weight
Code 8806-4.29
Weight + sample code (comment on weighing)
1500.08 g
Generating statistics of weight values
Page 16 Totalization
ID 52-88.6.31
1 99.5 g
2 101.5 g
3 103.5 g
4 100.5 g
5 102.4 g
6 104.5 g x 101.98 g s 1.87 g srel 1.83 % min. 99.5 g max. 104.5 g dif. 5.0 g
--------- END --------
11.01.2000 15:28:33
ID 71-00.6-34
3
4
1
2
206.55 g
3165.92 g
45.73 g
2.54 g
Sum 3420.74 g
--------- END --------
• Simple formula weighing
Multiplying
(dividing)
Page 17
Page 18
Factor 3.5
2.51 g
* 8.80
• Determine production homogeneity
• Demonstrate reproducibility of an experiment
• Calculating prices
• Determining volume
• Determining density
• Weighing in any weight units
• Calculating weight of area or length (kg/m 2 or g/m)
7
LP16/LJ16
IR Dryer
Page 19
11.01.2000
15:28:33
ID 71-00.6-34
Time 5.0 Min
Temp 160 C
Mode Timer
0..-100%
0.0M 11.010 g
1.0M -0.20 %
2.0M -1.36 %
3.0M -2.53 %
4.0M -3.18 %
5.0M -3.41 %
Time eff.
5.0 Min
-3.41 %
-0.377 g
--------- END --------
• Documentation of a drying operation
8
Page 19 Pipette check
11.01.2000
ID
SNR:1118000002
11:15:50
1128489
Factor
Rounding
1
2
1.0032
0.0001
0.09844 g
98.7550 µl
0.09870 g
99.0158 µl
0.09876 g
3
4 n
99.0760 µl
0.09882 g
99.1362 µl
4 x s srel min.
max.
dif.
98.99575 µl
0.16786 µl
0.17
%
98.7550 µl
99.1362 µl
0.3812 µl
--------- END --------
• Verification of volumetric dosing apparatus (pipettes, volumetric flasks, burettes, etc.)
• Generating statistics and calculating the most important statistical parameters
METTLER M-Pacs Page 21 application software
----------------------
% FORMULA
----------------------
11.01.2000
13:29
*Target: 2500.00 g
*******
1*Comp
Weight
Weight
20.00 %
19.91 %
497.84 g
2*Comp
Weight
Weight
Total
7.50 %
7.56 %
188.94 g
686.80 g
Total 27.47 %
----------------------
Total
Total
2501.10 g
100.04 %
----------------------
• Documentation of the process sequence, e.g. of a formula weighing operation
Checking balance Page 22 with test weight *
---- BALANCE TEST ----
11.01.2000
14:31:57
METTLER TOLEDO
Balance
Type: PG503DR-S
SNR: 1113000000
Target: 200.000
Actual: 200.005
Diff: 0.005
Internal test done
Signature:
......................
--------- END --------
Calibrating
(adjusting) balance
Page 23 with calibration weight *
-BALANCE CALIBRATION -
11.01.2000
16:51:57
METTLER TOLEDO
Balance
Type: PG503DR-S
SNR: 1113000000
Int. calibration done
Signature:
......................
--------- END --------
• Documentation of the check following GLP and ISO 9001
• Documentation of the calibration following GLP and ISO 9001
* The printed records of the functions “Checking” and “Adjusting” appear in the respective national language only on connection to PR/SR balances. With all other balance models, English is used as the standard language.
9
10
4. Operation
1
F
7
Def
4
ID
8
Menu
5
Code
2
Result
9
6
1 2 3
3
0
Menu
31
Clear
8
Menu
9 ...
...
0
Menu
4.1
Using the keypad
Purpose of the keypad
1.
Configuring the printer
2.
Entry of variable values, e.g. lot identification (ID)
3.
Selection of the printer function
Input mode
The printer functions are set in the dialog mode. The entries appear in the balance display and the LC-P45 prints them out as confirmation.
4.2
Configuring the printer in the menu
What is set and how?
The menu is used to configure the operational setting of the LC-P45:
• Set LC-P45 to factory setting
• Activate checking and calibration of the balance
• Select record header
• Select language of the dialog and record texts
• Select interface parameters
These three keys are used to enter the menu, choose the setting and store the selected setting.
Note: Variables are inputted directly. It is not necessary to enter the menu.
--------- Menu ------
AKTIVATE:
Factory setting
Verification
Calibration
RECORD HEADER:
8
Menu
8
Menu
2 ID
3 Date/ID
4 Date/Time/ID
5 Date/Time/ID/Bal.
LANGUAGE/SPRACHE:
*English
Deutsch
Français
Español 1)
Italiano 1)
BAUD RATE: 2)
*2400
4800
9600
1200
PARITY: 2)
*even/7 bit odd/7 bit no/7 bit no/8 bit
HANDSHAKE: 2)
*off xon/xoff hardware
9
0
Menu
Clear
8
Menu
Configuring the printer
:
0
Menu
8
Menu
8
Menu
9
Entry into menu
Jump to next menu item, e.g. RECORD HEADER, LANGUAGE/SPRACHE ...
Selection of the setting within a menu item, e.g. English, Deutsch ...
( ... record header layout, see page 7) 4 urther settings if required
Quitting the menu with simultaneous storage of all selected settings
Quitting the menu without storage of the selected settings (abort)
Clear
Printing out the current menu settings
8
Menu
Printout of a list of the current settings.
Clear
Example of a setting
Entry into menu and changing the
LC-P45 LANGUAGE/SPRACHE setting from English to Spanish or Italiano
Notes:
8
Menu
8
Menu
8
Menu
9 9 9 0
Menu
• * These settings are the factory settings; * marks the current setting after a change
• The menu settings are stored in a fail-safe manner (battery)
• 1)
• 2)
Either Español or Italiano
I/O via LocalCAN interface: no setting necessary.
11
Variable value
Time
Date
Lot identification (ID)
Sample code (comment on individual weighing)
Sample number start value
Factor for multiplication (division)
Rounding in multiplication
Number of line feeds after printout of weight or result
12
4.3
Entering values of variables
Key sequence
3
Def
Numeric values
Range of values
Hours.minutes.seconds
(0.0.0 ... 23.59.59)
7
Def
31
Numeric values
EURO:
US:
Day.month.year (1.1.00 ... 31.12.99)
Month-day-year (1-1-00 ... 12-31-99)
7
Def
4
ID
5
Code
Numeric values
Numeric values
Number with max. 20 digits (0 ... 9 plus „.“ and „-“)
(0 removes the ID from the record)
Number with max. 10 digits (0 ... 9 plus „.“ and „-“)
[is not stored!]
7
Def
6
1 2 3
Numeric values
1 ... 9 999 (0 removes the sample number from the record)
7
Def
Numeric values
0.000 001 ... 9 999 999
7
Def
7
Def
2
Result
Numeric values
Numeric values
0.05 : 2 decimal places, the 2nd rounded to 5
1.00 : 2 decimal places, both are = 0
0.001: 3 decimal places, the 3rd to the nearest 1
0 ... 3
Notes: • The input process can be checked in the balance display; correction of wrong entry with
• The inputted values are stored in a fail-safe manner (exception: sample code)
Clear
7
Def
1
F
Selection of function
* 1 = Print
2 = Statistics
3 = Totalization
4 = Multiplication
5 = IR Drying LP16/
LJ16
6 = Pipette check or or
1
2
3
4.4
Selecting printer function
• Entry into selection
On entry, a list of the available functions with their selection numbers are automatically printed out.
• Printing of weight values
• Generation of statistics of weight values
• Totalization of weight values
The input process can be checked in the balance display.
Correction of wrong entry
Clear
13
or or or
4
5
6
14
• Multiplication (division) of weight values by a factor
• Record of drying operation of
LP16 and LJ16 IR Dryer
The selection of the printer function is stored in a fail-safe manner (battery)
• Pipette test for calibrating volumetric dosing apparatus
5. Printer functions
8
Menu
8
Menu
9
...
0
Menu
31
8
Menu
7
Def
7
Def
8
Menu
4
ID
6
1 2 3
9
...
0
Menu
...
...
or
1
F
5.1
Printing out weight values
Printing with manual insertion of date and time
• Select "No record header" in the menu item RECORD HEADER (page 11)
• Print date + time
Set date and time page 6
• Print weight
Printing with automatic insertion of date, time, lot identification (ID) and sample number, depending on requirements
• Select the record header according to requirements in the menu (e.g.date/time/ID; page 11)
• Enter according to requirements (page 12):
– Lot identification (ID)
– Sample number start value
• Print record header and weight automatically
• Print record header and current sample No. + weight automatically
15
Clear
Resets the sample No. to the start value at any time.
2 7
Def
1
F
8
Menu
8
Menu
7
Def
4
ID
9
...
0
Menu
...
:
:
1
F
:
2
Result
:
Clear
16
5.2
Generating statistics of a weight series
• Select "Statistics" function (page 13)
• Select the record header in the menu (page11)
• If required, enter lot identification (ID)
(page 12)
• Load 1st sample and transfer weight value to the statistics x =
1 n n
∑
i = 1 x i s =
1 n - 1
∑
x i
- x 2 s rel
= s
• 100% x min.
= smallest value max.
= largest value dif.
= largest - smallest value max. number of samples = 999
• Continue the sample series
• If desired, print out intermediate statistics (the sample counter is not reset)
• Continue the sample series
• Conclusion of the sample series with automatic printout of the overall statistics (the sample counter is reset, the statistics memory cleared)
3 7
Def
8
Menu
7
Def
8
Menu
1
F
4
ID
9
...
0
Menu
...
:
:
1
F
:
2
Result
:
Clear
5.3
Totalizing weight values
• Select function "Totalization" (page 13)
• Select record header in the menu (page 11)
• If required, enter lot identification (ID) (page 12)
• Load 1st sample and transfer weight value to totalization memory max. number of samples = 9 999 max. total = 99 999 999 g
• Continue the sample series
• If need be, interrupt the sample series for automatic printout of the intermediate total
(the sample counter is not reset)
• Continue the sample series
• Completion of the sample series with automatic printout of the total
(the sample counter is reset, the totalization memory cleared)
17
4 7
Def
7
Def
7
Def
8
Menu
7
Def
8
Menu
4
ID
6
1 2 3
7
Def
2
Result
1
F
9
...
0
Menu
...
...
...
...
1
F or
2
Result
18
5.4
Multiplying (dividing) the weight by a factor
• Select function "Multiplication" (page 14)
• Select the record header in the menu (page 11)
• According to requirements, enter (page 12):
– Lot identification (ID)
– Sample number start value
– Factor for multiplication (division)
– Rounding in multiplication (division)
• Loading the weighing sample
• Automatic multiplication of the weight value by the preselected factor and printout of the result together with multiplier and weight
• Automatic multiplication by the preselected factor and printout of the result alone
5 7
Def
8
Menu
8
Menu
7
Def
4
ID
1
F
9
...
0
Menu
...
• Start drying
6 7
Def
8
Menu
8
Menu
7
Def
4
ID
1
F
9
...
0
Menu
...
5.5
Recording the moisture determination with LP16/LJ16 IR Dryer
The LC-P45 automatically inserts date, time and lot identification.
• Select function "IR Drying LP16/LJ16" (page 14)
• Select "Date/Time/ID" in menu item RECORD HEADER (page 11)
• If required, enter lot identification (ID) (page 12)
• Start drying operation according to instructions in the LP16/LJ16 operating instructions.
The LC-P45 prints out the record.
5.6
Pipette test for calibrating volumetric dosing apparatus
• Select function "Pipette check" (page 14)
• Select the record header in the menu (page 11)
• Enter pipette identification (ID) (page 12)
19
7
Def
7
Def
6
1 2 3
...
...
7
Def
2
Result
Vo lu
M ate
S
Y
S
E
M
S
®
...
1
F
:
: :
2
Result
:
Clear
20
– When necessary enter sample number start value (page 12)
• Enter the factor z to calculate the volume from the mass (page 12). The factor z for water is given in our application brochure "Good Pipetting Practice", order number 11795198, available free of charge from your METTLER TOLEDO dealer. This factor is essentially a function of the water temperature and atmospheric pressure
• Enter the rounding off convention for the calculated result (page 12)
• Dosing the liquid and inclusion of the first volume in the statistics. The dosed volume is calculated automatically from the measured weight by multiplying it with the factor z. The weight and the calculated volume in µl, (V = m x z) are printed out
• Continue the sample series
• If desired, print out intermediate statistics (the sample counter is not reset) (page 16)
• Continue the sample series
• Conclusion of the sample series with automatic printout of the overall statistics (the sample counter is reset, the statistics memory cleared) (page 16)
8
Menu
8
Menu
9
...
0
Menu
• Work with M-Pac
5.7
Recording the processes with M-Pac application software packages
The LC-P45 prints out the records of the M-Pacs, e.g. CountPac or LabPac.
• Select "No record header" in menu item RECORD HEADER (page 11)
• The LC-P45 is ready for work with M-Pacs.
It prints out their records.
Note: and
1
F are inactive.
21
------lO aD
1000.02
g
22
5.8
Checking a balance with a test weight
• Selection and starting the check on the balance (page 11)
8
Menu
9
ACTIVATE:
Factory setting
Verification
Calibration
0
Menu
*
* Minimum weight 900 in the last 3 places of the display.
Exemple: 9.38
g
0.945
kg the menu section ACTIVATE is automatically printed out
The prompt for weight loading/removal is shown on the balance display. At the same time, the following record is printed out:
---- BALANCE TEST ----
11.01.2000 14:31:57
METTLER TOLEDO
Balance
Type: PG503DR-S
SNR: 1113000000
Target: 200.000
Actual: 200.001
Diff: 0.001
Internal test done
Signature:
.......................
--------- END ---------
– date and time of verification
– balance identification
– target value of the checking procedure
– displayed weight value of the automatically tared balance
– difference between target and actual value
– signature of the tester
---- BALANCE TEST ----
11.01.2000 14:31:57
METTLER TOLEDO
Balance
Type: AB204
SNR: 1113000000
Weight ID: ...........
Target: ...........
Actual: 199.9800 g
Diff: ...........
External test done
Signature:
.......................
--------- END ---------
– date and time of verification
– balance identification
– identification of the test weight
– target value of the checking procedure
– displayed weight value of the automatically tared balance
– difference between target and actual value
– signature of the tester
8
Menu
9 9
ACTIVATE:
Factory setting
Verification
Calibration
0
Menu
-------
500.0
g
-------
0.0
g
-------
0.0
g
5.9
Calibrating (adjusting) a balance with a calibration weight
• Selection and starting the calibration procedure (page 11) the menu section ACTIVATE is automatically printed out
The prompt for weight loading/removal is shown, as usual, on the balance display. With balances with an internal calibration weight, the automatic calibration routine is started. At the same time, the following record is printed out:
Adjusting with external adjustment weight Adjusting with internal adjustment weight
-BALANCE CALIBRATION--
11.01.2000 14:31:57
METTLER TOLEDO
Balance
Type: PG503DR-S
SNR: 1113000000
Weight ID: ...........
Weight: 500.000 g
Ext. calibration done
Signature:
.......................
--------- END --------date and time of verification balance identification
-BALANCE CALIBRATION--
11.01.2000 14:31:57
METTLER TOLEDO
Balance
Type: PG503DR-S
SNR: 1113000000 identification number of the test weight target value of the adjustment procedure signature of the tester
Int. calibration done
Signature:
.......................
--------- END --------date and time of verification balance identification signature of the tester
23
6. Maintenance
4
1
3
Def
7
4
METTLER
F
ID
1
Result
2
Printer
Code
5
3
Menu
8
1
2
3
6
31
9
Menu
0
Clear
2
7
Def
4
METTLER
F
ID
1
Resu lt
2
Printer
Code
5
3
Menu
8
1
2
3
6
31
9
Menu
0
Clear
5
24
6.1
Changing paper and ribbon
Changing paper
• Insertion of a new roll of paper is described on page 5, commercially available, normal paper is used. Order number, see page 28.
Changing ribbon
Order number, see page 28.
• 1 Pull paper out of printing unit
• 2...4 Remove ribbon cover
• 5 Remove ribbon cartridge
• Insert new ribbon (if need be, take up slack with tensioning wheel) and refit ribbon cover
• Insert paper and feed through slit in the printing unit as far as it will go
• Press and hold until sufficient paper has been fed in
6.2
Maintenance, Service and Cleaning
Under normal ambient conditions, the LC-P45 Printer requires practically no maintenance thanks to its rugged construction. The METTLER TOLEDO service department is available should servicing be required. Please ask your responsible METTLER TOLEDO dealer for details.
Cleaning
As the printer housing is made of top grade, resistant materials, all commercially available cleaning agents may be used.
25
26
7. Malfunctions
7.1
Interpreting the error messages
Malfunction/Message Possible cause
Printout unreadable • Ribbon worn/entangled
• Service life of printing unit at an end
Green pilot lamp does not light up • No power, fuse defective
Green pilot lamp flashes slowly
Green pilot lamp flashes rapidly
---- Interruption ----
- Transmission error -
------- ABORT --------
• Balance waiting for stability/ cable not attached
• Transmission error
• Connection cable not attached or older balance (unidir. interface)
• Interface parameters of LC-P45 and balance do not match
• Calibration has been aborted
Rectification
• Change/tension ribbon
• Replace printing unit
• Switch on power switch, have fuse replaced by
METTLER TOLEDO
• Check handshake, initiate print command with PRINT key of balance
• Set interface parameters
• Repeat calibration
------ Timeout -------
- Battery discharged -
------ Overflow ------
---- Input error -----
----- Unit error -----
ES
EL
• Entry incomplete, menu not ended, balance has not sent any weight value in last 30 s
• Pause before completion of function must not be longer than 30 s, move balance to more stable location
• Batteries for clock and settings discharged
• Number of samples for multiplication exceeded
• Replace batteries; message repeated approx. every 2 hours
• Range of values exceeded • Repeat entry
• Different units in totalization, statistics or multiplication
• Syntax error (error message of balance), balance does not understand LC-P45 instruction
• Logistical error (error message of balance), balance can not execute LC-P45 instruction
27
• Switch on LC-P45
+
• Switch off LC-P45
28
7.2
Printer test
The LC-P45 is provided with a self-test which automatically prints out its character set. This allows easier recognition of a possible malfunction of the printing unit.
• Switch on LC-P45 and at the same time press and hold the feed key for a few seconds
• Terminate the printout by switching off the LC-P45
8. Appendix
8.1
Accessories for LC-P45 Printer
Accessory Type/Order number Standard equipment
Paper roll,set of 5
Ribbon cartridge, black, minimum order quantity 2
Batteries, 2 x UM3/AA (1.5 V)
Connection cable LC-P45-balance/scale, LocalCAN, 1 m
Extension cable/connection cable LocalCAN, 2 m
Extension cable/connection cable LocalCAN, 5 m
T-socket for devices with only one LocalCAN socket
Connection cable LC-P45-balance, RS-232-C, 1 m
72456
65975 commercially available
LC-LC1 cable, 229154
Connection cable LC-P45-balance, RS-232-C (MiniMettler), 1.5 m
Spare printing unit (may be changed only by METTLER TOLEDO service)
Power line fuse 230 V (slow-blow 100 mA,250 V) + 115 V (slow-blow 200 mA,250 V) –
(may be changed only by METTLER TOLEDO service)
229029
–
2 rolls
1 piece
2 pieces
1 piece
LC-LC2 cable, 229115
LC-LC5 cable, 229116
–
–
LC-LCT, 229118 –
RS9 (m)–RS9 (f) cable, 11101051 1 piece
–
–
–
1
6
LocalCAN RS-232-C
9
5
In/Ou t
Hand-
Shake
GND
+ 5 V/
20 mA
8.2
I/O interfaces
The LC-P45 has a LocalCAN and an RS-232-C interface for the attachment of METTLER
TOLEDO instruments and peripherals.
One or other can be used, but if both are connected only the LocalCAN is active.
LocalCAN interface
• Bus interface with 2 4-pin looped sockets for the attachment of balances/scales/ instruments and peripherals.
• No transmission parameters need be adapted.
RS-232-C interface
• 9-pin male connector (pin assignment as IBM laptop)
• Matching to other device (transmission parameters) in the menu; see page 11
• The handshake lines are active when "hardware" has been selected in menu item HAND-
SHAKE
• The standard cable 229029 has no handshake lines
LC-P45
Connection example
A scale and a PC are attached to the LC-P45 via the LocalCAN bus.
29
30
8.3
Attachment to METTLER TOLEDO instruments, additional hardware, available functions (software version 2.40)
Attachment to LC-P45 Available functions
Attachment to balance or instrument
AB, PB, SB, CB, GB, B *
AB, AB-S, PB, PB-S, SB, **
CB, GB, B
AG, PG
PG-S with RS232C
PG-S with LocalCAN Option
PR
AT/MT/UMT
PM/SM/AM
PJ/AJ Option 018/019
BB
M3/UM3 with Option 03
AE with Option 011
AE with Option 012
PE with Option 016
PE with Option 017
Remarks
9
RS
15
3
3
1
1
1
1
13
9
9
9
RS
13
14
13
❍
❏
❏
❏
❏
❏
❏
❏
❏
❍
❏
❍
❍
❏
❏
Additional hardware
2
4
5
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
Attachment to LC-P45 Available functions
Attachment to balance or instrument
LP16/LJ16
PM-Pacs
ID1
ID3 with Option 082
ID5 with Opion 082/089
IF with 504949
DT
DL18/21/25/35
FP62
Remarks
1
1
10
10
10
13
13
13
13
❏
12,13
❏
❏
❏
❏
❏
❏
❏
❏
Additional hardware
6
8
7
7
6
6
6
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
31
32
Legend and general remarks to the attachment of the LC-P45 to METTLER TOLEDO instruments
Legend
Connection cable to LC-P45
1 229029 is standard
2
3
4
5
Order number: 42931
59759 included in option
Order number: 229029
Order number: 47936
6
7
8
9
Order number: 504376
Order number: 17842
Produce cable
➞ page 29
LC-LC1 cable is standard
RS RS9 (m) – RS9 (w) cable is standard
❏ =
LocalCAN
❍ = RS-232-C/CL
CL/RS-Interface converter
CL249, Order number: 59759
Continuation legend
10 Interface mode:
DIALOG MODE (standard setting).
If the terminal is in break condition, it has to be switched off and on again.
The LC-P45 can print maximally 24 characters a line, additional characters will be discarded.
12 Printout of Pac record only
13 Check of key entries via printout only, the balance display is inactive.
7 3
Example: Setting the time. After the set time is printed out:
Def
-> 13:55
14 Documentation of calibration after balance software version 10.45.
15 For the AB, PB, SB, GB, CB and B balances and scales earlier than 02.97, the interface "LC-B Option" is required.
Adapter cable and hand switch
Order number: 47473
Order number: 42500
* Balance models earlier than 02.97
** Balance models after 02.97
General remarks
1
• With or
F
only stable weight values can be printed out; to print out unstable values, set the balance to "send all" and give the print command with its
PRINT key.
• Do not set the balance to "send all" for statistics, totalization and multiplication.
• If the PRINT key of the balance is pressed, the weight will be printed out without record header.
8.4
Technical data
Printer
Printing unit
Printing speed
Ribbon cartridge
Paper roll dot matrix printer 5x9 dots, 24-character line length, Epson/IBM character table No. 4 faster than 1 line per second exchangeable, black normal paper, 58 x dia. 51 mm, integrated in housing, commercial size
Printer functions see page 3
Clock date representation EURO or US format, leap year taken into account
Interface
Transmission parameters LocalCAN
Transmission parameters RS-232-C
Handshake
Printer buffer
LocalCAN and RS-232-C bidirectional fixed adjustment transmission rate 1200, 2400*, 4800, 9600 baud, parity even*, odd, no character length 7 bit*, 8 bit
Bidirectional handshake: off*, xon/xoff, hardware
128 characters
* = factory setting
33
34
Designed test conditions
Power supply voltage, Frequency
Current consumption
Height
Temperature range
Humidity
Installation category
Pollution degree
Dimensions / weight
115V or 230V, admissible fluctuations -20%+15%, 50/60Hz
140mA or 70mA
4000m
5...40°C max. relative humidity 80% for temperatures to max. 31°C linear decreasing to 50% at 40°C
II
2
W x D x H = 157 x 210 x 85 mm / net 1,5 kg (incl. paper roll)
Printed on 100 % chlorine-free Paper, for the sake of our environment.
To protect your METTLER TOLEDO product's future:
METTLER TOLEDO service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come.
Please send for details about our attractive terms of service.
Thank you.
© Mettler-Toledo GmbH 2001 705108C Printed in Switzerland 0102/2.12
Mettler-Toledo GmbH, Laboratory & Weighing Technologies, CH-8606 Greifensee, Switzerland
Phone +41-1-944 22 11, Fax +41-1-944 30 60, Internet: http://www.mt.com
P705108
Subject to technical changes and to the availability of the accessories supplied with the instruments.
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
Frequently Answers and Questions
What are the capabilities of the LC-P45 Printer?
How do I set the date and time on the printer?
How do I print out the weight value of a balance?
How do I generate statistics of a weight series?
How do I perform totalization?
advertisement