- Home
- Domestic appliances
- Climate control
- Environment thermometers
- EUTECH INSTRUMENTS
- 60010-20
- Instruction manual
advertisement
Thermocouple
Thermometer Models
60010-00 (Type J)
60010-10 (Type K)
60010-20 (Type T)
Technology Made Easy ...
68X309905 Rev 1 04/04
TABLE OF CONTENTS
1.
INTRODUCTION 4
2.
SAFETY PRECAUTIONS 6
3.
SPECIFICATIONS 7
4.
BATTERY INSTALLATION AND
5.
REPLACEMENT 10
SELF TEST 12
6.
CONNECTING A THERMOCOUPLE
12
7.
SELECTING TEMPERATURE
SCALE 16
8.
HOLD FUNCTON 17
9.
CALIBRATION 17
10.
CANCELING A FIELD CALIBRATION
18
11.
FIELD CALIBRATION LOCKOUT
AND RE-ENABLE
12.
LOCKOUT PROCEDURE
13.
RE-ENABLE PROCEDURE
19
19
20
14.
MAINTENANCE 20
15.
CLEANING 21
16.
BATTERIES 21
17.
ACCESSORIES 22
18.
SERVICE 24
19.
WARRANTY 25
20.
PRODUCT RETURN 25
1. INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements. The instrument is designed for easy operation and includes the following features:
• Operator selection of Celsius or
Fahrenheit scale
• Resolution of 0.1° from –99.9 to
299.9°
• Four-digit LCD
• Hold feature for temporarily retaining a reading
• Field calibration capability
4
• Low battery warning
• Built-in tilt stand for easy hands-free operation
• Two blade female ANSI miniconnector input
• Operates with a wide selection of probes 2
5
WARNING: THIS INSTRUMENT IS
DESIGNED TO ACCEPT LOW
LEVEL SIGNALS SUPPLIED BY
STANDARD THERMOCOUPLES.
UNDER NO CIRCUMSTANCES
SHOULD THE INPUT VOLTAGE
EXCEED THE SPECIFIED 50V RMS.
CAUTION: DO NOT USE OR STORE
THIS INSTRUMENT IN MICROWAVE
OVENS OR ANY ABNORMALLY HOT
OR COLD AREAS.
CAUTION: WEAK BATTERIES
SHOULD NOT BE LEFT IN THE
INSTRUMENT. DEAD BATTERIES
CAN LEAK AND CAUSE DAMAGE TO
UNIT.
DANGER: VOLTAGES PRESENT AT
THE THERMOCOUPLE MAY ALSO
BE PRESENT AT THE BATTERY
TERMINALS. ALWAYS DISCONNECT
THE THERMOCOUPLE WHEN
CHANGING BATTERIES.
WARNING: TO PREVENT IGNITION
OF A HAZARDOUS ATMOSPHERE,
BATTERIES MUST ONLY BE
CHANGED IN AN AREA KNOWN TO
BE NON-HAZARDOUS.
6
3. SPECIFICATIONS
THERMOCOUPLE PROBES
Type J
Type K
Type T
–200°C to 1000°C
(–328°F to 1832°F )
–250°C to 1372°C
(–418°F to 2501°F )
–250°C to 400°C
(–418°F to 752°F )
Out of range display: - - - -
Resolution
0.1°/1° autoranging, 0.1° from –99.9° to
299.9°, 1° outside this range.
Accuracy
> –99.9°: +(0.2% of reading +0.5°C)
+(0.2% of reading +0.9°F)
< –99.9°: +(0.25% of reading, +1°C)
+(0.25% of reading +2°F)
Display
4-digit LCD with 0.5 in high numerals.
Display update rate
0.6 sec per update.
Input
One thermocouple with ANSI connector.
7
Input Protection
50V rms
Battery
Two AA, 1.5V alkaline ANSI-L40, IEC-
LR6.
Battery Life: 750 hours continuous, typical.
Low battery indication
Battery Symbol on when 8 to 10 hours of battery life remains.
Battery symbol blinking: replace battery
Operating Conditions
Stated accuracy:
18°C to 28°C (64°F to 82°F)
Useful range:
0°C to 40°C (32°F to 104°F)
Storage:
–40°C to 65°C (–40°F to 149°F)
Humidity (non-condensing):
10% to 90% 4
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H
(1.2 in x 3.3 in x 6.2 in)
Weight with batteries
227 grams (8 ounces)
8
Ingress protection
Meets IEC-529 IP-54 for dust and water resistant enclosures.
Intrinsic safety
This product is energy limited for intrinsically safe operation in hydrogen atmospheres per Class I, Division 1,
Groups A, B, C and D hazardous(classified) locations for UL per UL913 and CSA per C22.2 No. 0-
M1982 and No. 157-M1987. Maximum surface temperature: 135°C (T4); UL file No. E182612 (1997).
Electromagnetic Compatibility
(For CE Mark) EN61326-1/1A: 1998
(EU EMC Directive)
9
INSTALLATION
REPLACEMENT
If battery indicator turns on, battery life is approximately 8 to 10 hours. The battery indicator will start blinking with less than 1 hour of life remaining.
Selected settings are stored in memory and will remain in memory even after power is turned off, or while batteries are being replaced.
1. Before changing battery, turn instrument off and disconnect thermocouple.
2. Loosen screw and lift battery cover off back of case.
3. Remove the two AA batteries.
4. Insert two new batteries observing polarity.
5. Install cover and tighten screw.
10
CAUTION: WEAK BATTERIES
SHOULD NOT BE LEFT IN THE
INSTRUMENT. DEAD BATTERIES
CAN LEAK AND CAUSE DAMAGE TO
UNIT.
DANGER: VOLTAGES PRESENT AT
THE THERMOCOUPLE MAY ALSO
BE PRESENT AT THE BATTERY
TERMINALS. ALWAYS DISCONNECT
THE THERMOCOUPLE WHEN
CHANGING BATTERIES.
WARNING: TO PREVENT IGNITION
OF A HAZARDOUS ATMOSPHERE,
BATTERIES MUST ONLY BE
CHANGED IN AN AREA KNOWN TO
BE NON-HAZARDOUS.
11
Press the ON/OFF key. The thermometer performs a self-test and all display digits and indicators, as shown below, should remain on for approximately one second.
THERMOCOUPLE
Use the correct thermocouple type for your instrument. Using an incorrect thermocouple type will result in erroneous readings. Thermocouples are colour coded by type using the
North American ANSI Color Code as follows:
TYPE COLOR
J Black
K Yellow
T Blue
Thermocouple connectors have one wide blade and one narrow blade. Do not force connector in backwards.
Connect thermocouple to receptacle at top of instrument as shown in the following illustration.
12
Thermocouple wiring polarity must be correct. If readings decrease as the temperature increases, the thermocouple wires may be reversed.
The red wire is negative for thermocouple wires manufactured in
North America.
13
14
15
If a thermocouple is not connected or if the thermocouple is defective, the display will indicate OPEn.
Thermocouples are sensitive at the tip or measuring junction. When taking measurements, allow time for the reading to stabilize. Multiplying the time constant of the probe by 5 will give you the approximate time required.
SCALE
Select °C or °F by pressing the °C/°F key. Each time the key is pressed the temperature scale will switch. Switching between °C and °F can be done at any time during operation.
Each time you turn the instrument on, it will power up with the same settings that were set when the unit was last turned off.
16
Press the HOLD key to retain the reading on the display. Press the
HOLD key again for normal operation.
9. CALIBRATION
The thermometer is factory calibrated and does not require calibration before use. The CAL function allows single point calibration of the thermometer, at
0°C (32°F) to compensate for thermocouple off-set error. It is not necessary to perform a field calibration to obtain specified meter accuracy. Use the field calibration feature to improve
. thermometer/probe accuracy or to compensate for thermocouple drift.
1. Pack sensing end of probe in a container tightly packed with crushed ice and filled with distilled water. Allow temperature to stabilize.
2. Press and hold the CAL key for 5 seconds to enter the calibration mode, the CAL annunciator on the display starts blinking. Release the CALkey.
3. If the measured temperature is from
_10°C to10°C (14 to 50°F), when the
17
temperature reading is stable, press the CAL key.
The CAL indicator will stop blinking and the reading will be set to 0°C
(32°F). The CAL indicator will remain turned on, indicating a field calibration is active.
If “Err” is displayed, either the displayed reading is outside the above limits or the batteries are weak.
10. CANCELING A FIELD
CALIBRATION
1. Turn the thermometer off.
2. Hold the CAL key down while pressing the ON/OFF key.
The field calibration is canceled and the thermometer reverts to the default factory calibration. The CAL annunciator is now turned off.
18
11. FIELD CALIBRATION LOCKOUT
AND RE-ENABLE
The calibration lockout feature prevents any field calibration changes. The lockout remains in effect until a lockout re-enable has been performed. Use the following procedures to lockout or re-enable the field calibration operation.
12. LOCKOUT PROCEDURE
1. Turn the thermometer off.
2. Simultaneously press and hold the
CAL and the °C/°F keys down and momentarily press the ON/OFF key.
Continue to hold the CAL and °C/°F keys for at least 5 seconds.
19
13. RE-ENABLE PROCEDURE
1. Turn the thermometer off.
2. Simultaneously press and hold the
HOLD and the CAL keys down and momentarily press the ON/OFF key.
Continue to hold the HOLD and
CAL keys until the display blanks.
14. MAINTENANCE
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and changing of the batteries.
20
15. CLEANING
WARNING:
TO PREVENT IGNITION OFA
HAZARDOUS ATMOSPHERE BY
ELECTROSTATIC
DISCHARGE,CLEAN WITH DAMP
CLOTH.
Do not clean with abrasives or solvents. Use mild detergents, never immerse nor use excessive fluid.
16. BATTERIES
If there is no display when the thermometer is turned on, check condition of the two AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary, refer to the BATTERY INSTALLATION
AND REPLACEMENT section for replacement procedure.
21
17. ACCESSORIES
Replacement Meter and Meter
Accessories
Type J
Type K
Type T
Soft carrying case
Replacement AA batteries, pack of 4
Certificate of Calibration
60010-00
60010-10
60010-20
08520-05
09376-01
Call for part number
Traceable to
NIST
22
General purpose probe (immersion
Into liquids), type J
Penetration probe
(meat, semi-soft
Materials), type J
Surface probe
(direct contact on
Hot surfaces), type J
Clip-on probe
(surface contacts-
Electronics), type J
General purpose probe (immersion
Into liquids), type K
Penetration probe
(meat, semi-soft
Materials), type K
Surface probe
(direct contact on
Hot surfaces), type K
Clip-on probe
(surface contacts-
Electronics), type K
EC-TPGLPJ-
01M/ 08517-55
EC-TPPENJ-
01M/ 08517-65
EC-TPSURJ-
01M/ 08517-60
EC-TPCLPJ-
01M/ 08469-00
EC-TPGLPK-
01M/ 08516-55
EC-TPPENK-
01M/ 08516-65
EC-TPSURK-
01M/ 08516-60
EC-TPCLPK-
01M / 08469-02
23
18. SERVICE
WARNING
SUBSTITUTION OF COMPONENTS
MAY IMPAIR INTRINSIC SAFETY.
There are no internal adjustments or user replaceable parts.
If “Err” followed by the numbers 1 through 9 is displayed (see example below) return unit for service. Note that
“Err” alone may be displayed during improper field calibration.
Note: Serial number label is located inside battery compartment.
24
19. WARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifications for a period of three years. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
20. PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period.
When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.
25
NOTES
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the
Manufacturer or authorized seller.
Trademarks bearing the ® symbol in this publication are registered in the
U.S. and in other countries.
For more information on Eutech
Instruments/ OAKTON Instruments’ products, contact your nearest distributor or visit our web site listed below:
Eutech Instruments Pte Ltd
Blk 55 Ayer Rajah Crescent
#04-16 Singapore 139949
Tel: (65) 6778 6876
Fax: (65) 6773 0836
Website: www.eutechinst.com
Email: [email protected]
OAKTON Instruments
P.O. Box 5136
Vernon Hills, IL 60061, USA
Tel (in U.S.): 888-462-5866
Tel (outside U.S.) 1-847-549-7600
Website: www.4oakton.com
E-mail: [email protected]
Fax: (1) 847-247-2984
Distributed by:
advertisement
Related manuals
advertisement