advertisement
GR 1551-C Sound Level Meter_0001.djvu
INSTRUCTION MANUAL
Type 1551-C
Sound-Level
Meter
a
GENERAL RADIO
I
GR 1551-C Sound Level Meter_0002.djvu
CONTENTS
Specifications
Section 1 Introduction.
Section 2 Operating Procedure ,
Section 3 Calibration . • .
Section 4 Service and Maintenance .
Section 5 Type 1262-B Power Supply .
Handbook of Noise Measurement
This 282-page book, by Dr. A . P. G. Peterson and Ervin E. Gross. Jr •• of the Gener a l Radio Engineering Staff, covers thoroughly the subject of noise and vibration measurement. Copies are available from Ge neral Radio at $1.00 each. postpaid in the United States and Canada.
1
3
9
11
18
Type 1551-C
Sound- Leve I
Meter
Q
Copyright 1968 by
GEN_~~~h ~~.91~J~NY
West M assach u setts USA 01781
Form 1551.0110.0
10 1276
December, 1968
GR 1551-C Sound Level Meter_0003.djvu
SPECIFICATIONS
Sound-Level Range: From 24 to 150 dB (re 20 ;.tN/m2).
Frequency Characteristics: Four response characteristics, A, B,
C, or 20 kHz, as selected by panel switch. The A-, B-, and Cweighting positions are in accordance with USA Standard S1.4·
1961 and IEC Publication 123, 1961. Frequency response for the
20-kHz position is flat from 20 Hz to 20 kHz, so that complete use can be made of very wide-band microphones such as the
1551-P1 Condenser Microphone Systems.
Microphone: GR Type 1560-P5. Accessory condenser microphone is available.
Sound·Level Indication: Sound level is indicated by the sum of the meter and attenuator readings. The clearly marked, open-scale meter covers a span of 16 dB with calibration from -6 to +10 dB.
The attenuator is calibrated in 10-dB steps from 30 to 140 dB above 20 ~tNfm2.
Calibration Accuracy: When amplifier sensitivity has been standardized, the absolute accuracy of sound-level measurements at
500 Hz is within ± 1 dB and at all frequencies is in accordance with the USA Standard.
Panel adjustment is provided for standardizing amplifier gain with internal calibration circuit.
Absolute acoustic sensitivity is factory calibrated at 500 Hz.
Microphone response and sensitivity are measured in a free field from 20 Hz to 15 kHz by comparison with a WE 640AA laboratorystandard microphone with calibration traceable to the National
Bureau of Standards. Complete electrical frequency-response measurements are made on each instrument.
The 1562-A Sound-Level Calibrator or the 1559-B Microphone
Reciprocity Calibrator can be used for making periodic over-all acoustic checks.
Output: 1.4 V behind 7000 0 (panel meter at full scale). The output can be used to drive analyzers, recorders, oscilloscopes, and headphones. Harmonic distortion (panel meter at full scale) <1 %.
Input Impedance: 25 MO in parailel with 50 pF.
Meter: Rms response, and fast and slow meter speeds in accordance with USA Sl.4-1961 and IEC 123, 1961.
Environmental Effects
Temperature and Humidity: Microphone is not damaged at temperatures from -30 to +95°C and relative humidities from 0 to
100%. When standardized by its interna I calibration system or a
1562 Sound-Level Calibrator, the instrument will operate within catalog specifications (for panel-meter indications above 0 dB) over the temperature range of 0 to 60°C and the relative humidity range of 0 to 90%.
Magnetic Fields: When exposed to a 60-Hz, 1-oersted (80 A/m) field, the sound-level meter will indicate 60 dB (C weighting) when oriented for maximum sensitivity to the magnetic field.
Electrostatic Fields: Aluminum case provides sufficient shielding, so that normally encountered electrostatic fields have no effect.
Vibration: Case is fitted with soft rubber teet and amplifier is resiliently mounted for vibration isolation. When the instrument is set on its feet on a shake table and vibrated at 10 mils pk-pk displacement over the frequency range of 10 to 55 Hz, the unwanted signals generated do not exceed an equivalent C-weighted sound-pressure level of 45 dB when motion is vertical, 60 dB when motion· is lengthwise, or 40 dB when motion is sidewise.
GR 1551-C Sound Level Meter_0004.djvu
SPECIFICATIONS (Con't)
GENERAL
Power Supply: Two 1'12 V size D flashlight cells and one 671h-V battery (Burg ess XX45 or equiva l en t) are supplied. An ac power supply, the Type 1262-B, is available.
Accessories Supplied: TelephOne plug.
Acces s ories Available : 1551 -P2 Leathe r Case (pe rm its operation of instrument without removal from case), 1562 Sound-Level Calibrator, 1560 ·P95 Adaptor Cable for connecting output to 1521-B
Graphic Level Recorder .
Mounting: Aluminum cabinet.
Dimensions (width x height x depth): 71/4 x 9114 x 6'.1* ln. (185 x
235 x 160 mm).
Weight, Net, 7¥4 lb (3.6 kg); shipping, 16 lb (7.5 kg), batte ries incl. Ad d 2 lb for leather case.
For a mo r e detailed description, r efe r to Gene ral Radio Experimenter , Au gus t 1 96 1 .
U.S. Patent No. 3,012,197
Figure 1 .
Ty pe 1551-C Sound-Level Meter.
I nse t s how s Type 1262-B P ower
Supply attoched.
GR 1551-C Sound Level Meter_0005.djvu
TABLE OF CONTENTS
Section 1. INTRODUCTION
1.1 Purpose . . .
1.2 Description .
1.3 Power Supply.
1.4 Carrying Case
Section 2. OPERATING PROCEDURE
2.1 Preliminary Checks . . . . . .
2.2 Operating Procedure. . . . . .
2.3 Selection of Weighting Network .
2.4 Sound Analysis and Recording
2.5 Extension Cable . . . .
2.6 Wide-Range Microphones ·
2.7 Use of Headphones . .
2.8 Background Noise . .
2.9 Microphonics . . . . .
2ol0 Environmental Effects.
2o11 Vibration Pickup . .
Section 3. CALIBRATION
3.1 General . . . . . .
3.2 Observer Influence .
3.3 Preferred Angle of Incidence .
3.4 Auxiliary Calibration . . . .
3.5 Meter Circuit. . . . . . . .
Section 4. SERVICE AND MAINTENANCE
4. 1 Genera I . . . . . . . . o . . . .
4.2 Removal of Instrument from Cabinet .
4.3 Tube Failure. . . o
4.4 Tube Replacement . . .
4.5 Battery Replacement . .
4.6 Replacement of Bias Cell
4.7 Soldering of Connections to Cable Wires.
4.8 Microphone Sensitivity Adjustment
4.9 Internal Noise . . . . . . . .
4.10 Amplifier Gain . . . . . . . o .
4.11 Test Voltages and Resistances .
Section 5. TYPE 1262-B POWER SUPPLY
12
12
12
13
13
11
11
11
11
12
12
9
9
10
H)
10
3
4
4
5
6
6
7
7
7
7
9
1
1
3
3
GR 1551-C Sound Level Meter_0006.djvu
Section 1
INTRODUCTION
1.1 PURPOSE. The Type 1551-C Sound-Level Meter (Figure 1) is the basic instrument of the General Radio sound-measuring system. This instrument conforms to all requirements of the ASA American Standard
Specification for General Purpose Sound- Level Meters. An accurate, portable, low-priced meter, it indicates the sound-pressure level at its microphone in terms of a standard reference level (0.0002
~ar).
In addition to its primary use as a Sound- Level Meter, the Type
1551-C can be used as a highly sensitive ac voltmeter. With the microphone sensitiv·ity adjustment (paragraph 4.8) set at -55.5 and the attenuator switch set at 30, the meter has a full-scale sensitivity of 30 microvolts. Voltages to be measuredcan be applied to the microphone connector through a Cannon three-terminal plug.
1.2 DESCRIPTION.
1.2.1 GENERAL. The major components of the Sound-Level Meter are a nondirectional microphone, a calibrated attenuator, an amplifier, an indicating meter, and weighting networks to modify the amplifier frequency response.
1.2.2 MICROPHONE. The Type 1560-P5 Microphone supplied with the
Type 1551 Sound-Level Meter is a new piezoelectric ceramic microphone developed expressly for sound-level meter use. Several features of the Type 1560-P5 make it an excellent microphone for sound-level measurements:
10.0002 f.Lbor "' 20f.LN/m2; cps = Hz.
GR 1551-C Sound Level Meter_0007.djvu
-2
-
+4
+2
~ft
T
I
50
(CORRESPONDING ANGLE OF INCIDENCE
FOR PLANE WAVES; 70°)
4
20c1s I 00 200 500 lk<;!S
FREQUENCY
2
/
-
~I\
\
10 20
~
Figure 2.
Frequency response of Type 1560-
PS Microphone to sounds of random incidence.
1. a frequency characteristic (see Figure 2) carefully controlled in manufacture to give flat response to sounds of random incidence from
20 cps to 12 kc,
2. rugged, dependable design, capable of withstanding wide elimatic changes (from -30° to 200° F in temperature, from 0 to 100% relative humidity),
3. low temperature coefficient of sensitivity, with minimal change in output voltage from 0°to 200° F,
4. low temperature coefficient of internal impedance; the cable correction is not affected by temperature at the microphone.
The nominal internal impedance of the microphone is 380 pf.
1.2.3 CONTROLS AND CONNECTORS. The controls and connectors on the panel of the Sound-Level Meter are listed below.
2
TABLE 1. CONTROLS AND CONNECTORS
Name Type Function
Attenuator 12 -position selector switch Selects meter range.
CAL Thumb- set adjustment Electrical internal calibration or gain adjustment.
METER-
BATTERIES
5-position selector switch Selects fast or slow meter response; provides battery checks.
WEIGHTING 5-position selector switch Selects weighting.
OUTPUT
On-Off
None
Phone jack
Switch built into microphone arm
6-terminal Jones receptacle
Output connection.
Turns instrument on and off (with battery power only).
Connection to Type
1262-B Power Supply.
GR 1551-C Sound Level Meter_0008.djvu
1.3 POWER SUPPLY.
1.3.1 BATTERIES. Two TypeD flashlight batteries supply the filament
(A) power, and one 67-1/2 -volt portable-radio-type B battery supplies the plate (B) power for the instrument. Contact is made to each battery by spring terminals, and all batteries are secured by a screw -down bracket.
1.3.2 AC POWER SUPPLY: The Type 1262-B Power Supply(refer to section 5) can be used to operate the Sound- Level Meter from a 115-volt ac line. When the Typel262-B is used, its own power switch, ratherthan the microphone switch on the Type 1551-C, controls application of power to the Sound- Level Meter.
1.4 CARRYING CASE. Available for use with the Sound-Level Meter is a leather carrying case with shoulder straps. A hole in the back of this case matches the tapped hole in the back of the Sound-Level Meter so that the instrument can be secured in the case for field use. The cover flaps of the case are equipped with a luggage-type fastener to keep them closed over the panel and with snap buttons to hold them open when the· instrument is in use.
Section
OPERATING PROCEDURE
2.1 PRELIMINARY CHECKS.
2.1.1 BATTERY CHECK. To check batteries, set the METER-BATTER-
IES switch to FIL-l, turn the instrument on by raising the microphone, and check that the meter pointer is within the area marked BAT on the meter face. Turn the METER-BATTERIES switch to FIL-2 and then to
PL, similarly checking that the meter pointer falls in the white BAT area for each position.
Filament batteries should last 6 to 7 days at 8 hours a day, or 30 to 35 days at 2 hours a day. Plate batteries should last 12 to 14 days at
8 hours a day, or 60 to 70 days at 2 hours a day.
2.1. 2 CALIBRATION CHECK. Before using the Sound- Level Meter, make the following simple calibration check. (This check calibrates the amplifier of the Sound-Level Meter, but does not include a check on the microphone.) a. Turn the instrument on. b. Set the attenuator switch to 130 CAL (i.e., with the pointer on the knob toward the window).
3
GR 1551-C Sound Level Meter_0009.djvu
c. Set the WEIGHTING switch to CAL, and check that t e meter pointer falls within the white area marked CAL on the meter face. If not, set it there by adjusting the CAL thumbset control.
2 .1.3 ACOUSTICAL CALIBRATION. The following over -all acoustical calibration can be made with the 1562 Sound-Level Calibrator. This calibration checks the microphone sensitivity as well as the amplifier sensitivity.
Perform the calibration check as described in paragraph 2 .1.2.
Then set the WEIGHTING switch to FLAT. Thrn on the calibrator and set its frequency to 500 cps. Place the calibrator over the microphone. The sound-level meter should read 114±0.5 dB (refer to paragraph 2.2). Switch the calibrator to each of its other frequencies. If the reading of the sound-level meter at 1000 cps is not between 113 and 115 dB, and/or if the readings at the various frequencies vary by more than ± 1 dB, it is possible that the microphone has been damaged; it should then be checked independently.
When the microphone is known to be in good condition, the gain of the sound-level meter can be calibrated by setting the meter reading to 114 dB at 1000 cps by means of the CAL control.
The above calibration will be made free of charge for any GR sound-level meter delivered to one of the General Radio offices listed elsewhere in this manual.
2.2 OPERATING PROCEDURE. a. Turn the instrument on by raising the microphone (battery power) or by turning on the Type 1262-B (ac power). b. Adjust the WEIGHTING switch for the desired frequency response. Refer to paragraph 2. 3. c. Adjust the attenuator switch for an on-scale deflection of the
'ndicating meter. The sound level in decibels is the algebraic sum of he readings of attenuator dial and meter. d. Set the METER-BATTERIES switch to FAST or SLOW depending on the type of meter response desired. The fast response is suitable for most noise measurements and for measuring minimum and maximum values of fluctuating sounds. The SLOW position provides a high degree of damping for measuring the average value of fluctuating sounds.
2. 3 SELECTION OF WEIGHTING NETWORK. Many early noise criteria specified weighted sound levels, using this rule of thumb: A weighting for sound levels from 24 to 55 db, B for levels from 55 to 85 db, and C for levels from 85 to 140 db. (The appropriate range was selected after a preliminary C-weighted measurement.) More recent opinions favor selection of weighting network on the basis of the type of noise measurement; for instance, A weighting is often preferred for speech -interference measurements, while B is recommended for surveys of traffic noise.
In the absence of specific weighting requirements, it is usually helpful to take measurements on all three weighting networks. For a full discussion of weighting networks, refer to the General Radio Handbook of Noise
Measurement. Response curves for the various weighting networks, equalized for microphone response, are given in Figure 3.
GR 1551-C Sound Level Meter_0010.djvu
+10 z
0
-
w~ n:~ 0
~2
~0
...J
~
-1020
"""""'
f--.-+ oVER-ALL FREE-FIELD FREQUENCY RESPONSE r-c WEIGHTING (TO SOUND OF FANDOM INCIDENCE)
50 100 200 500 1000 2000
FREQUENCY IN CYCLES PER SECOND
5000 10,000 20,000
11~51-27'(
+5
0
-5
"'
-10
8
-15 c
~
-20
!2
~ -25
1/)
"'
"'
_,.
~
-35
-40
A
_,
__.
~
c ~
~'f" /
/
BAN~
I
/s ~
ELECTRICAL FREQUENCY RESPONSE
I
/ FOR THE ASA WEIGHTING CHARACTERISTICS
/
j
A
/
/
-45
50 100 200 500 1000 2000
FREQUENCY IN CYCLES PER SECOND
5000 10,000 20,000
~
Figure 3. Typical Acoustical and Electrical
Response Curves for Type 1551-C.
If theSound-Level Meter istobe used withananalyzeror recorder, set the WEIGHTING switch to 20 kc or to C. When an extended-range microphone is used in place of the microphone supplied with the instrument, use the 20 kc position.
2.4 SOUND ANALYSIS AND RECORDING. The low-distortion outputamplifier circuit of the Sound-Level Meter was designed especially for use with the Type 1558-A Octave-Band Analyzer and Type 1564-A Sound and
Vibration Analyzer. Other analyzers, such astheTypel900-A WaveAnalyzer, can also be used. Connect the input of the analyzer or recorder to the OUTPUT phone jack on the Sound-Level Meter. For greatest accuracy, the impedance connected to the output circuit of the Sound-Level
Meter should be about 20,000 to 25,000 ohms. The output impedance is
7000 ohms, and output voltage is 1.1 V into .20,000 ohms, 1.5 V opencircuit. Any load impedance can be connected across the output terminals.
The output system of the Sound-Level Meter can be used to drive a Type 1521 Graphic Level Recorder to obtain permanent records of
5
GR 1551-C Sound Level Meter_0011.djvu
sound measurements, or to operate a magnetic tape recorder to o}:)tain field records that may be studied and analyzed later in the laboratory.
When recording or analyzing, set the attenuator switch so that the meter reads +6 db or more,if possible, to utilize the maximum dynamic range. If the meter reading falls below +6 db, it is sometimes desirable to set the attenuator switch one step lower, even if the meter pointer goes off scale as a result.
The internal calibration system of the Sound-Level Meter makes available, at the OUT jack, the 1000-cycle signal useful for setting the level of an analyzer or recorder to match that of the Sound-Level Meter.
2e5 EXTENSION CABLE. The Type 1560-P73* 25-foot Extension Cable can be used to permit operation of the microphone at some distance from the meter. The extension cable introduces about 7 db loss. For best measurements this cable insertion loss should be determined by calibration with a Type 1562 Calibrator.
2.6 WIDE- RANGE MICROPHONES. The over-all frequency response characteristic of the Type 1551-C,with the WEIGHTING switch set at 20 kc, is essentially flat from 20 cps to 20 kc. (See Figure 4.) This permits use of such wide-range microphones as those used in the Types
1551-PIL and 1551-P1H Condenser Microphone Systems.
The Type 1551-P1L uses the Altec 21-BR-150 Condenser Microphone, useful over the sound-pressure range from 50 to 150 db. The
Type 1551-P1H uses the Altec Type 21-BR-180 Microphone, useful from about 70 to 170 db. (When using the Type 1551-P1L, set the meter to 121 rather than 120 db in the calibration described in paragraph 2.1.3. When using the Type 1551-P1H, set the meter to 101 db and thereafter add 20db to meter readings.) These microphones have a good frequency response from 20 cps to 18 kc.
*Also supplied as part of the Type 1560-P34 Tripod and 25-ft Cable Assembly.
(WEIGHTING SWITCH SET AT 20 Kc)
I v
/_.,..-
20 50
I
~
"\
\
100 200 500 1000 2000
FREQUENCY IN CYCLES PER SECOND
5000 10,000 20,000 50,000
Figure 4.
Frequency Response of Type 1551-C Amplifier with WEIGHTING switch at 20 kc.
GR 1551-C Sound Level Meter_0012.djvu
Also useful with the Type 1551-C over the range from 80 to 200 db is the Massa Type M-141F Standard Microphone, a piezoelectric microphone available from the Massa Laboratories Inc, Hingham, Mass. Its response is reasonably uniform up to 30 kc.
2.7 USE OF HEADPHONES. If desired, a set of headphonescanbeplugged into the OUTPUT jack to monitor the sound being measured.
2.8 BACKGROUND NOISE. When possible, sound measurements should be made with negligible background noise - at least 10 db below the level being measured. However, this is not always possible, and Figure 5 is convenient in determination of errors caused by background noise.
2.9 MICROPHONICS. When high-intensity sound fieldsare being measured, it is good practice to use the microphone on the end of a cable (refer to paragraph 2.5) and to keep the Sound-Level Meter well removed from the sound field. This is especially true if the high-level noise contains frequencies of 1000 cps or higher. The following quick check will determine whether the tubes in the Type 1551-C are beingexcited microphonically: remove the microphone head, set the WEIGHTING switch to
A, and check that there is no meter reading with the attenuator switch set at 70 db or higher.
At sound levels below 140 db no special vibration precautions should be necessary. The amplifier is resiliently mounted, vacuum tubes are resiliently mounted on the amplifier board, the amplifier cover is special- · ly treated to reduce airborne sound transmission, and the instrument sits on soft rubber-feet.
When the Sound- Level Meter, with microphone replaced by an equivalent electrical impedance, is subjected to pure-tone sounds at a soundpressure level of 100 db re 0.0002 ~r and 8000 cps, the output level in each octave band is at least 30 db below that for full-scale meter reading with the attenuator set for 100 db fullscale meter reading.
The Type 1551-C, with its microphone mounted on the panel microphone post, can measure sound pressure levels up to 140 db within the tolerances allowed in ASA Sl.4, 1961. When the microphone is mounted on a cable, the maximum safe sound pressure level for the microphone is 150 db.
Withan auxiliary high-level microphone separated from the Sound-
Level Meter by a long cable to protect meter and operator from intense sound fields, levels up to 200 db can be measured.
2.10 ENVIRONMENTAL EFFECTS.
2.10.1 TEMPERATURE AND HUMIDITY. The Type 1560-P5 Microphone is notdamaged at temperatures from -30°to +95° C and relative humidities from 0 to 100%. When standardized by its internal calibration systern or a Type 1562 Sound- Level Calibrator, the Type 1551-C will operate within catalog specifications (for panel-meter indications above 0 db) over the temperature range from 0 to 60 C and the relative- humidity range from 0 to 80%.
7
GR 1551-C Sound Level Meter_0013.djvu
8
~ z
7
\
\
\
:I!;
~ fil 4
~
~
....
ID
2 g
I
0:: i
0 I
"'
3
""
4
""""
~
5 6 7
-
8
r---
--
9 10
~
DB Dl FFERENCE BETWEEN TOTAL NOISE AND BACKGROUND ALONE
Figure 5.
Effect of Background
Noise on Measurement.
+10
+8
+6
!l'l+4
Figure 6. 1 +2 w
Frequency Response as a~
0
Function of Incidence, ~ o::-2
Type 1560-PS Microphone.
-4
-6
20 fPERbENDICJLAR
I I
~-PARALLEL
I
I
PERPENDICULARJ
/
INCIDENCE
RANDOM INCIDENCJ y
1~-
1/
/
/
\
'
PARALLEL INCIDENCE
'
'
100
I
I
1000
FREQUENCY IN CYCLES PER SECOND lopoo
~
The internal impedance of the microphone is capacitive, 475 pf at
25° C, varying from 445 to 510 pf from 0 to 50° C.
2.10.2 VIBRATION. The Type 1551-C case is fitted with soft rubber feet and the amplifier compartment is resiliently mounted for vibration isolation. When set on its rubber feet on a shake table and vibrated over the frequency range of 10 cps to 55 cps, the motion transmitted to the amplifier generates unwanted signals. The magnitude of these signals is listed in the table below. peak-to-peak d i sp I acement
(inches)
0.030
0.010 direction of vibration
X y
Z maximum meter reading(db)
65
45
70
60
55
40
GR 1551-C Sound Level Meter_0014.djvu
2.10.3 MAGNETIC FIELDS. When exposed to a time-varying magnetic field of 1 oersted, the Type 1551-C will indicate 60 db when oriented for maximum sensitivity to the magnetic field.
2.10.4 ELECTROSTATIC FIELDS. The Type 1551-C is completely enclosed in an aluminum case so that the instrument is not affected by the electrostatic fields which are normally encountered.
2.11 VIBRATION PICKUP. It is often possible to measure the vibration amplitude, velocity, or acceleration of radiating surfaces. Useful vibration measurements can often be obtained where high ambient sound levels make acoustical measurements impossible. For vibration frequencies between 20 and 2000 cps, the Type 1551-C will operate as a vibration meter when used with a Type 1560-PllB Vibration Pickup System. With a Type
1560- Pll S2 Vibration Pickup System (refer to the General Radio Expertmenter, Volume 36, Number 11, November 1962), the Type 1551-C can be used for vibration measurements between 20 and 10,000 cps.
Refer to the General Radio Handbook of Noise Measurement for further information on vibration measurements.
Section 3
CALIBRATION
3.1 GENERAL. The calibration (acoustical and electrical) of the Type
1551-C Sound-Level Meter is carried out with pure tones. The Sound-
Level Meter calibration is adjusted for random incidence sounds with the microphone mounted on the panel microphone post.
Microphone calibrations are made at 0 degrees (perpendicular incidence) in a free field with the microphone at the end of a cable. All microphone calibrations supplied by General Radio are at 0 degrees incidence. A correction curve is supplied to convert from 0 degrees to random-incidence response.
3.2 OBSERVER INFLUENCE. When measurements are made in reverberant or semireverberant fields, the observer and Sound- Level Meter case have a very small effect on the microphone response. Under freefield conditions and with a single sound source, both the observer and the case can have a large effect on microphone response (refer to the General
Radio Handbook of Noise Measurement). These effects can be minimized if the instrument is held in front of the observer, with the observer oriented so that the sound passes in front of him at right angles to the axis of the Sound-Level Meter and its microphone.
For greatest accuracy it is recommended that the microphone be mounted on a tripod and connected to the Sound- Level Meter by an. ex tention cable to keep the observer and the meter case out of the sound field being measured.
9
GR 1551-C Sound Level Meter_0015.djvu
3.3 PREFERREDANGLEOFINCIDENCE. When measurementsarema e on sounds in reverberant fields, the angle of incidence of sounds reaching the microphone is indeterminant. In this case there is no preferred angle of incidence between the microphone and the sound source. When measurements are made on a source in a free field, an angle of incidence of
70degrees between the axis of the microphone and the sound source will approximate random response.
3.4 AUXILIARY CALffiRATION. The Type 1551-C has an internal calibrating system for quick checks on the electrical circuit of the instrument. This system connects the output of the amplifier to the input through an appropriate attenuator, filter, and limiter to produce a predetermined oscillation amplitude when the amplifier gain is correct. This system checks the electrical operation only and does not check the microphone.
To make an acoustical check on the calibration of the instrument, the
Type 1562 Acoustic Calibrator is recommended. -This is a closed coupler with a driving loudspeaker that produces aknown sound pressure level at the microphone.
3.5 METER CIRCUIT. The metercircuitusedin the Sound-Level Meter,. while not a true rms indicator, indicates much more closely to the rms value than have meters in earlier instruments. The meter circuit comhines an average-reading circuit with a peak-reading circuit to create a
"quasi-rms" indicating circuit. Since the rms value of most waveforms falls above the average value but below the peak value, a porti.on of the peak indication added to the average value of a given waveform should result in an rms indication. The meter circuit is designed to approximate rms values for various types of waveforms.
Table 2 below shows the db difference between the indication of a true rms meter and that of the Type 1551-C Sound-Level Meter for several test signals. In the column headed "db Fluctuation . . . . . etc" the signal consists of two tones, one at 1000 cps set for a convenient meter indication. The other tone is within a few cycles of 3000 cps and has 30 percent of the amplitude of the first. A true rms meter will showno fluctuation with this type of signal. The column headed "For Two-Signal Addition"l"efersto the two-signaltestoutlined in ASA Standard Z24.3-1944.
With pulses of constant height but varying length, the Type 1551-C
Sound-Level Meter indicates the rms- value within ±1 db until the pulse duration becomes as short as 1/25 that of a square wave.
TABLE 2.
Type
Meter
1551-C db Fluctuation with Phase Changes at 30% 3rd Harmonic
0.45
Difference in Meter Indication and RMS
Decibels
For
Two-Signal
Addition
For
Square
Waves
For
Noise
+.05 +0.1 -0.25
10
GR 1551-C Sound Level Meter_0016.djvu
Section
SERVICE AND MAINTENANCE
4.1 GENERAL. Wewarrant that each newinstrument manufactured and sold by us is free from defects in material and workmanship, and that, properly used, it will perform in full accordance with applicable specifications for a period of two years after original shipment. Any instrument or component that is found within the two-year period not to meet these standards after examination by our factory, Sales Engineering Office, or authorized repair agency personnel, will be repaired, or, at our option, replaced without charge, except for tubes or batteries that have given normal service.
The two-year warranty stated above attests the quality of materials and workmanship in our products. When difficulties do occur, our service engineers will assist in any way possible. If the difficulty cannot be eliminated by use of the following service instructions, please write or phone our Service Department (see rear cover), giving full information of the tr-ouble and of steps taken to remedy it. Be sure to mention the serial and type numbers of the instrument.
Before returning an instrument to General Radio for service, please write to our Service Department or nearest Sales Engineering
Office, requesting a Returned Material Tag. Use of this tag will ensure proper handling and identification. For instruments not covered by the warranty, a purchase order should be forwarded to avoid unnecessary delay.
4.2 REMOVAL OF INSTRUMENT FROM CABINET. To remove the instrument from its cabinet, remove the two large black screws from the bottom of the cabinet and lift the cover off.
4.3 TUBE F AlLURE. As tubes age, one or more of them may become open -circuited. Check the filament resistance with an ohmmeter between pins Band 5 of the tube base (see Figure 8). Make sure that the ohmmeter current through the filaments is less than 10 rna.
4.4 TUBE REPLACEMENT. Foraccess to tubes,removethetwoPhillipshead screws that attach the cover of the amplifier compartment, and lift off the cover. The tubes are held in place between rubber pads on the amplifier shelf and on the inside of the cover.
When replacing a tube, cut the leads on the new tube to between
5/16 and 3/8 inch before inserting the tube in the socket. Install the tube so that the red dot on the tube base is on the same side as that on the socket.
11
GR 1551-C Sound Level Meter_0017.djvu
4.5 BATTERY REPLACEMENT. To replace any of the batteries, remove the black screw that secures the holding bracket. (See Figure 9.)
After the bracket is removed, the batteries can be withdrawn and replaced. Observe proper polarity as marked, when replacing batteries.
4.6 REPLACEMENT OF BIAS CELL. For access to the bias cell~ remove the two studs and washers from the ends of the amplifier etchedcircuit board. Pivot the amplifier shelf upward to expose the bias cell.
To remove the bias cell, first pull the contacting clip sideways and away from the bias cell The bias cell can then be unscrewed and replaced.
When returning the amplifier shelf to its original position, be careful that all wires are free and that none will be pinched between shield and chassis.
4.7 SOLDERING OF CONNECTIONS TO CABLE WIRES. The plastic insulation on the wires in cables and on many interconnecting leads melts easily, and extreme caution is advised in soldering connections. If the wire is held with pliers near the spot to be soldered, heat will be carried away from the insulation, helping to avoid melted insulation.
4.8 MICROPHONE SENSITIVITY ADJUSTMENT. An internal sensitivity control is adjusted in the General Radio laboratory to match the characteristics of the microphone sent with each meter. If a microphone with a different sensitivity is substituted for that furnished, set this adjustment (see Figure 9) to indicate the new microphone sensitivity. Also, a check of acoustical calibration should be made by the method outlined in paragraph 2.1.3.
4.9 INTERNAL NOISE. Internal noise should not normally affect readings with the attenuator switch set at 40 or higher. If desired, the following procedure can be used to determine how much noise is generated in the mstrument. Remove the microphone, and connect a 380-pf capacitor across terminals 1 and 3 of the microphone socket. This capacitor and its connecting leads must be completely shielded in a metal can connected to either the metal part of the microphone socket or to terminal
1, which is grounded. The connection is identical to that shown in Figure 7, except that a short circuit replaces the oscillator.
SOUND-LEVEL
METER
INPUT CONNECTOR
USE C-WEIGHTING
NETWORK lt551-28!
Figure 7. Circuit for Calibrating Amplifier Gain.
Be certain that there is no external pickup of any stray fields. The meter on the instrument should not read above -6 with the attenuator
12
GR 1551-C Sound Level Meter_0018.djvu
switch set at 30, the WEIGHTING switch at C, and the METER BATTER-
IES switch at FAST.
Internal noise can be caused by the bias cell, B50, or by the batteries. Tube noise is generally caused only by V50 or V52, which can be interchanged with V51 or V53 or replaced if less noise is required. It may be necessary to try several tubes to achieve a satisfactory reduction in noise. Typical noise levels arc given in the following table.
TYPICAL INTERNAL SIGNAL-TO-NOISE RATIOS
C WEIGHTING
(db BELOW FULL SCALE· SET FOR -60 db MICROPHONE)
1551-C
Attenuator
Setting db
All Pass
20 cps·
20 kc
NOISE LEVELS· db
OCTAVE BAND· cps
31.5 63 125 250 500 1000 2000 4000 8000 16000
30
40
50
60
70
80
90
100
110
120
130
140
17
27
37
46
52
54
55
55
54
55
55
55
25
35
45
55
64
73
78
80
79
80
79
80
26
36
46
56
64
28
38
48
58
67
73 74
77
77
78
78
76
76
78
78
76
77
77 78
30
40
50
60
68
74
75
76
73
76
75
76
30
41
51
61
68
73
73
74
71
74
73
74
67
71
71
71
30
40
50
60
69
71
70
71
28
38
48
58
65
68
68
68
67
68
68
68
26
36
46
56
63
65
66
66
65
66
66
66
27
37
47
56
62
63
63
63
63
63
63
63
29
39
. 48
57
59
60
60
60
60
60
60
60
4.10 AMPLIFIER GAIN. With the Sound-Level Meter set for normal operation with C weighting, apply a 1.0-volt, 400- or 1000-cycle signal to the microphone socket through a shielded 380-pf capacitor, as shown in Figure 7. The meter reading should agree with that given below for the sensitivity of the microphone used, (Microphone sensitivity ls listed at the front of this manual, as well as on a label inside the instrument.)
1560-P3 (981Ul:S}
Microphone Sensitivity db re 1 volt/ubar
-62
-61
-60
-59
-58
-57
Type 1551-C reading
136
135
134
133
132
131
4.11 TEST VOLTAGES AND RESISTANCES. The table on the following page lists de voltages and resistances between tube and transistor pins and ground, as measured with a vacuum-tube voltmeter. For voltage measurements, set the attenuator to 130 CAL, the WEIGHTING switch to
A, and the METER-BATTERIES switch to FAST. For resistance measurements, remove batteries and ground terminals 2, 3, and 4 of socket 802.
13
GR 1551-C Sound Level Meter_0019.djvu
14
TABLE OF VOLTAGES AND RESISTANCES.
TUBE
(TYPE)
PIN
FUNC-
TION
VOLTS DC RES
BTRY 1262-B TOGND
V50
(CK512AX)
V51
(CK512AX)
V52
(CK512AX)
V53
(CK512AX)
V54
(CK6418)
V55
(CK6418)
TRI
(2Nl372)
RX51
S02
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
1
2
3
4
5
K
4 p s c
G c p 15.0 15.5 890 k s
15.0 15.5 890 k c
0.71 0.7 16
G c
-1.40
0
-1.41
0
*
0
16.0
14.1
1.45
0
0.72
16.1
14.2
1.41
0
0.71
320 k
940 k
0
1M
16 p s c
G c
9.4
9.4
0.76
-0.48
0
9.1
9.1
0.76
-0.48
0 p 21.5 20.5 s
14.9 14.3 c
1.44 1.4
G c
0
0.76
0
0.73
850 k
850 k
16
18M
0
340 k
1.21 M
0
18M
16 p s c
G c p 17.2 16.3 123 k s
17.6 17.0 400 k c
1.48 1.42 0
G c
-0.35
0
-0.35
0
18M
0
14.5
21.5
1.47
-0.22
0
13.9
20.7
1.37
-0.22
0
100 k
270 k
0
18M
0
E
B c
14.3
14.2
0
40.2
65.0
13.0
12.9
0
39.7
59
30 k
-
-
96 k
0
*Do not attempt to measure (bias cell in circuit).
GR 1551-C Sound Level Meter_0020.djvu
Ref. No. Description
PARTS LIST
GR
Part No.
Fed. Mfg.
Code
Mfg. Part No.
Fed. Stock
No.
CAPACITORS
Cl Mica, 330 pF ±2% 500 V
C2
C3
1 c. 4 pF
Trimmer, 7 - 45 pF
Mica, 30 pF ±2% 500 V C4 cs
C7
C59
C60
C61
C62
C63
C64
C65
C66
C67
C68
C52
C53
C54
C55
C56
C57
C58
C69
C70
C71
C8
C9
C10
C11
C12
C13
CIS
C16
C19
C20
C21
Mica, 90.0 pF ±2% 500 V
Plastic, 0.0033 j..LF ±2% 200 V
Mica, 178 pF ±2% 500 V
Mica, 178 pF ±2% 500 V
Plastic, 0.0332 j..LF ±2% 100 V
Mica, 220 pF ±2% 500 V
Wax, 0.4 7 1-LF ±10% 100 V
Electrolytic, 60 1-LF 25 V
Ceramic, 12 pF ±5% 500 V (NPO)
Mica, 35 pF ±2% 500 V
Mica, 0.001 f.LF ±5% 500 V
Mica, 0.001 f.LF ±5% 500 V
Mica, 0.0012 j..LF ±5% 500 V
C22 Mica, 0.001 f.LF ±5% 500 V
C23 Plastic, 0.00158 j..LF ±1% 200 V cso
Wax, 0.1 1-LF ±10% 100 V
CSI Wax, 1 1-LF ±10% 100 V
Electrolytic, 50 f.LF 50 V
Wax, 0.22 1-LF ±10% 100 V
Wax, 0.1 1-LF ±10% 100 V
Wax, 0.22 1-LF ±10% 100 V
Wax, 0.22 f.LF ±10% 100 V
Electrolytic, 30 j..LF 100 V
Wax, 0.01 1-LF ±10% 100 V
Wax, 0.1 1-LF ±10% 100 V
Wax, 0.1 1-LF ±10% 100 V
Wax, 0.01 1-LF ±10% 100 V
Wax, 0.22 1-LF ±10% 100 V
Wax, 0.47 1-LF ±10% 100 V
Plastic, 0.0606 j..LF ±2% 100 V
Wax, 1 1-LF ±10% 100 V
Wax, 0,04 7 1-LF ±10% 100 V
Wax, 0.4 7 1-LF ±10% 100 V
Wax, 0.4 7 1-LF ±IO% 100 V
Wax, 0.4 7 1-LF ±10% 100 V
Electrolytic, 19 j..LF 150 V
Ceramic, 3.3 pF ±10% 500 V
4690-3200 00656 CM20E, 300 pF ±2%
1551-0420 24655 1551-0420
4910-0100 72982 T8AN300, 7-45 pF
4690-(1400 00656 CM20E, 30 pF ±2%
5905-858-4049
5910-799-9275
4690-1201 00656 CM20E, 90.9 pF ±2%
4860-7350 84411 663UW, 0.0033 1-LF ±2%
4690-2217 00656 CM20E, 178 pF ±2%
4690-2217 00656 CM20E, 178 pF ±2%
4860-7868 84411 663UW, 0.0332 f.LF ±2%
4690-2700
·5010-3600 80183 78P4 74183
4690-(1600
5910-448-5765
4450-2900 56289 D17872
4400-3200
5910-799-9280
00656 Type CL-1, 12 pF ±5% 5910-705-6442
00656 CM20E, 35 pF ±2% 5910-731-1750
4580-(1100
00656 CM20E221G
14655 SA, 0.001 1-LF ±5%
4580-0100 14655 SA, 0.001 f.LF ±5%
5910-931-4153
5910-052-7016
5910-052-7016
4580-(1200 14655 SA, 0.002 1-LF ±5%
4580-(1100 14655 SA, 0.001 1-LF ±5%
4860-7319 84411 663UW, 0.00158 f.LF ±1%
5010-2700 56289 78Pl049183
5010-3700 80183 78P1057183
4450-2200 80183 50 1-LF +100-10%
5010-3300 80183 78P2241183
5010-2700 56289 78P1049183
5910-052-7016
5910-448-5758
5910-:615-5-255
5910-799-9283
5910-448-5777
5910-448-5758
5910-448-5777
5910-448-5777
5910-708-5191
5010-3300 80183 78P2241183
5010-3300 80183 78P2241183
4450-1900 37942 9594 7G7823Cl0X3
5010-1600 80183 78Pl031S3
5010-2700 56289 78P1049183
5010-2700 56289 78P1049183
5010-1600 80183 78P103183
5010-3300
5010-3600
80183 78P2241183
80183 78P474183
4860-8213
5010-3700
84411 663UW, 0.0606 f.LF ±2%
80183 78P1057183
5010-2100 56289 78P4 73183
5010-3600 80183 78P474183
5010-3600 80183 78P4741S3
5010-3600 80183 78P474183
4450-(1200 37942 D33104
4400-(1400 78488 GA, 3.3 pF ±10%
5910-448-5788
5910-448-5758
5910-448-5758
5910-448-5788
5910-448-5777
5910-448-5765
5910-615-5255
5910-448-5765
5910-448-5765
5910-448-5765
5910-829-3313
5910-708-5197
RESISTORS
R1 Film, 22.'J MQ ±1% 1 W
R2
R3
R4
Film, 2.17 MQ ±1% 1/2 W
Film, 333 kQ ±I% 1/8 W
RS
R6
R7
R8
R9
RIO
*R18
R20
R21
R23
R24
R25
R26
R27
R28
R29
R30
R31
R32
R33
R34
R35
R36
R37
*R38
R39
R40
Rll
Rl2
R13
R14
RIS
R16
R17
Film, 1.0 MQ ±1% 1/8 W
Film, 659 kQ ±1% 1/8 W
Film, 73,2 kQ ±1% 1/8 W
Film, 500 kQ ±1% 1/8 W
Film, 159 kQ ±1% 1/8 W
Film, 108 kQ ±1% 1/8 W
Film, 50 kQ ±1% 1/8 W
Composition, 3 kQ ±5% 1/2 W
Film, 16.9 kQ ±1% 1/2 W
Film, 16.9 kn ±1% 1/2
·w
6182-5229
6450-4217
6250-3333
6250-4100
6250-3659
6250-2732
6250-3500
6250-3159
6250-3108
6250-2500
6250-3154
6250-2159
6250-3169
03888 PT1000, 22.9 MQ ±1%
75042 CEC, 2.17 MQ ±1%
75042 CEA, 333 kQ ±1%
75042 CEA, 1 MQ ±1%
75042 CEA, 659 kQ ±1%
75042 CEA, 73.2 kQ ±1%
75042 CEA, 500 kQ ±1%
75042 CEA, 159 kQ ±I%
75042 CEA, 108 kQ ±I%
75042 CEA, 50 kQ ±I%
75042 CEA, 154 kQ ±1%
75042 CEA, 15.9 kQ ±1%
75042 CEA, 169 kQ ±1%
Film, 154 kQ ±1% 1/8 W
Film, 15.9 kQ ±1% 1/8 W
Film, 169 kQ ±1% 1/8 W
Film, 7.32 kQ ±1% 1/8 W
Film, 63.3 kQ ±1% 1/8 W
Film, 216 kQ ±1% 1/8 W
Film, 100 kQ ±I% 1/8 W
Composition, 100 kQ ±5% 1/2 W
Composition, 110 kQ ±5% 1/2 W
Composition, 75 kQ ±5% 1/2 W
Composition, 620 kQ ±5% 1/2 W
Composition, 3 kQ ±5% 1/2 W
Composition, 11 kQ ±5% 1/2 W
Composition, 3 kQ ±5% 1/2 W
6250-1732
6250-2633
6250-3216
6250-3100
6100-4105
6100-4115
75042
75042
75042
75042
CEA, 7.32 kQ ±1%
CEA, 63.3 kQ ±I%
CEA, 216 kQ ±I%
CEA, 100 kQ ±I%
01121 RC20GF104J
01121 RC20GF114J
6100-3755 01121 RC20GF753J
6100-4625
6100-2305
01121 RC20GF624J
01121 RC20GF302J
6100-3115 01121 RC20GF113J
6100-2305 01121 RC20GF302J
6100-2305 01121 RC20GF302J
6250-2169 75042 CEA, 16.9 kQ ±1%
6250-2169 75042 CEA, 16.9 kQ ±1%
Composition, 820 kQ ±5% 1/2 W 6100-4825 01121 RC20GF824J
Composition, 20 kQ ±5% 1/2 W
Composition, 43 kQ ±5% 1/2 W
Composition, 330 kQ ±5% 1/2 W
6100-3205 01121 RC20GF203J
6100-3435 01121 RC20GF433J
Composition, 24 kQ ±5% 1/2 W 6100-3245 01121 RC20GF243J
Potentiometer, composition 50 kQ ±10% 6010-1400 01121 JU, 50 kQ ±10%
6100-4335 01121 RC20GF334J
Composition, 24 kQ ±5% 1/2 W 6100-3245 01121 RC20GF243J
Composition, 51 kQ ±5% 1/2 W 6100-3515 01121 RC20GF513J
Potentiometer, composition 100 kQ ±10% 6010-1700 01121 JU, 100 kQ ±1%
Potentiometer, composition 100 kQ ±20% 6040-1000
Comp9sition, 1.0 MQ ±5% 1/2 W 6100-5105
01121
01121
FWC, 100 kQ ±20%
RC20GF105J
5905-646-5678
5905-655-1982
5905-783-8446
5905-8 92-6921
5905-578-0994
5905-577-6743
5905-195-6761
5905-279-1867
5905-2 79-3495
5905-221-5841
5905-279-1751
5905-279-2667
5905-279-1751
5905-279-1751
5905-806-8487
5905-806-8487
5905-221-5848
5905-192-0649
5905-279-3498
5905-279-1878
5905-055-3046
5905-279-2519
5905-279-1878
5905-279-3496
5905-797-1054
5905-958 -7949
15905-192-0390
15
GR 1551-C Sound Level Meter_0021.djvu
PARTS LIST {Cont)
R69
R70
R71
R72
R73
R74
R75
R76
R77
Ref. No. Description
RESISTORS (Conti
*R50 Composition, 430 kQ ±5% 1/2 W
R51
R52
Composition, 110 kQ ±5% 1/2 W
Film, 1 MQ ±1% 1/8 W
R53
R54
R55
Composition, 910 kQ ±5% 1/2 W
Composition, 33 kQ ±5% 1/2 W
Film, 290 kQ ±I% 1/8 W
*R56
*R57
R58
R59
*R60
R61
R62
R63
R64
*R65
Composition, 18 MQ ±5% 1/2 W
Composition, 750 kr.l ±5% 1/2 W
Composition, 100 kr.l ±5% 1/2 W
Composition, 18 MQ ±5% 1/2 W
Composition, 680 kr.l ±5% 1/2 W
Composition, 1.2 MQ ±5% 1/2 W
Composition, 330 kr.l ±5% 1/2 W
Composition, 18 MQ ±5% 1/2 W
Composition, 390 kr.l ±5% 1/2 W
Composition, 110 kr.l ±5% 1/2 W
R66
*R67
R68
GR Fed. Mfg.
Part No. Code
6100-4435 01121
6100-4115 01121
6250-4100 75042
6100-4915 01121
6100-3335 01121
6250-3290 75042
6100-6185 01121
6100-4755 01121
6100-4105 01121
6100-6185 01121
6100-4685 01121
6100-5125 01121
6100-4335 01121
6100-6185 01121
6100-4395 01121
6100-4115 01121
Potentiometer, composition 2.5 kr.l ±10% 6040-1700 01121
Composition, 300 n
±5% 1/2 W
Composition, 10 kQ ±5% 1/2 W
6100-1305 01121
6100-3105 01121
Composition, 220 kQ ±5% 1/2 W
Composition, 18 MQ ±5% 1/2 W
Composition, 2.4 MQ ±5% 1/2 W
Composition, 300 kQ ±5% 1/2 W
Composition, 100 kQ ±5% 1/2 W
Composition, 120 kQ ±5% 1/2 W
Composition, 470 kQ ±5% 1/2 W
Composition, 10 kr.l ±5% 1/2 W
Composition, 20 kr.l ±5% 1/2 W
6100-4225 01121
6100-6185 01121
6100-5245 01121
6100-4305 01121
6100-4105 01121
6100-4125 01121
6100-4475 01121
6100-3105 01121
6100-3205 01121
MISCELLANEOUS
Bl
B2
B3
B50
01
02
03
04
05
Battery, 1 -1/2 V Type D cell
Battery, 1-1/2 V TypeD cell
Battery, 67-1/2 V
Battery, 1-1/2 V
Diode, type 1N34A (S)
Diode, type 1N34A (S)
Diode, type 1N34A (S)
Diode, type 1N34 A (S)
Diode, type 1N645
06
Jl
Ml
PLl
Ql
Diode, type 1N645
Jack
Meter, 100 fla de
Plug
Transistor, type 2Nl372
RX51 Diode, type 1N967B
S1
S2
Switch
Switch
S3
S4
SOl
S02
V50
Switch
Switch,DPST
Socket
Socket
V51
V52
V53
V54
V55
Tube
Tube
Tube
Tube
Tube
Tube
8410-D200 77542
8410-D200 77542
8410-2300 09823
4020-DlOO 90303
6082-1003 82389
6082-1003 40931
6082-1003 71785
6082-1003 58854
6082-1016 58854
6082-1016 58854
4260-1500 58854
5730-1050 24446
24446
8210-1372 86800
6083-1020 76854
7890-1600 76854
7890-1600 24655
7890-1530 24655
1551-D450 24655
4230-2840 71785
4230-3500 01295
8370-1400 94144
8370-1400 94144
8370-1400 94144
8370-1400 94144
8380-6418 94144
8380-6418 94144
*Must be Allen-Bradley.
Mfg. Part No.
2LP
2LP
XX45
BC4
#111
MEDS-105
P-302-AB
IN34A(S)
IN34A(S)
IN34A(S)
IN34A(S)
IN645
IN645
IN976B
179901-H4C
179901-H4C
7890-1530
1551-0450
4230-2840
S-306-AB
2Nl372
CK512AX
CK512AX
CK512AX
CK512AX
CK6418
CK6418
RC20GF434J
RC20GF114J
CEA, lMO ±1%
RC20GF914J
RC20GF333J
CEA, 290 kr.l ±1%
RC20GF186J
RC20GF754J
RC20GF104J
RC20GF186J
RC20GF684J
RC20GF125J
RC20GF434J
RC20GF186J
RC20GF394J
RC20GF114J
FWC, 2500 Q ±10%
RC20GF301J
RC20GF103J
RC20GF224J
RC20GF186J
RC20GF245J
RC20GF304J
RC20GF104J
RC20GF124J
RC20GF474J
RC20GF103J
RC20GF203J
Fed. Stock
No.
5905-279-2518
5905-279-1867
5905-646-5678
5905-221-5853
5905-171-1998
5905-279-2503
5905-279-1759
5905-195-6761
5905-279-2503
5905-171-2000
5905-190-8874
5905-279-2518
5905-279-2503
5905-279-2517
5905-279-1867
5905-279-5481
5905-185-8510
5905-192-0667
5905-279-2503
5905-279-2512
5905-185-6859
5905-195-6761
5905-192-3981
5905-279-2515
5905-185-8510
5905-192-0649
6625-708-5186
5961-170-4430
5961-170-4430
5961-170-4430
5961-170-4430
5961-944-8222
5961-944-8222
5960-854-8469
5930-708-5199
5930-708-5198
5960-581-9593
5960-581-9593
5960-581-9593
5960-581-9593
5960-537-3967
5960-537-3967
Fig. 8
Ref.
4
5
6
Name
Microphone
Knob Asm.
Knob Asm.
Thumbset
Support
Knob Asm.
8
(Hidden)
9
Meter Cover*
Foot
Handle
MECHANICAL PARTS LIST
Description
GR
Part No.
Fed. Mfg.
Code Mfg. Part No.
Piezoelectric -Ceramic
Knob, attenuator; includes bushing 4140-0300
Knob, WEIGHTING; includes bushing 4140-0300
CAL control
Black rubber
Knob, METER/Batteries;
1560-9605
5530-1200
5500-0400
5540-2500
5260-1600 includes bushing 4140-0300 5500-0400
ME3-701 Cover; Dark Gray Mask 5720-3712
Block rubber 5250-1902
Leather strap handle 1551-0460
*When ordering, please specify manufactured by Weston or Honeywell.
24655 1560-9605
24655 5530-1200
24655 5500-0400
24655 5540-2500
24655 5260-1600
24655 5500-0400
24655 5720-3712
24655 5250-1902
24655 1551-0460
Fed. Stk.
No.
5355-926-5196
5355-051-6594
5355-051-6594
16
GR 1551-C Sound Level Meter_0022.djvu
GR 1551-C Sound Level Meter_0023.djvu
---·
7
6
Figur e 8. External mechanica l po rt s of the
Sound-Level Meter.
GR 1551-C Sound Level Meter_0024.djvu
en
()
r:Y
o"'
06
Oo/
( )
~':!
MARKING FOR
S-1 DIAL 8 KNDB
S-2
ENGRAVING
S-3
ENGRAVING
V50
V51
V52
V53
RAYTHEON
S S I
REO
007
V54
V55
JRANSISTORS
()/
Rotary switch sections are shown as vi•w•d from th• panel end of th• shaft. Th• first digit of th• contact number r•f•rs to th• section. Th• s•ction n•orest the pan•l is 1, th• n•xt section bock is 2, •tc. Th• n•xt two digits ref•r to th• contact. Contoct 01 is the First position clockwi .. from o strut screw (usu-
•lly th• scr•w oboveth•locating k•y), ond the oth•r contocts or• numb•r•d uqu•ntially (02, 03, 04, etc), proc•eding clockwiu around th• sect;on. A suffix
F or R indicatu !hot th• contact is on th• front or rear of th• uction, r•spectiv•ly.
NOTE.-
R£SISTORS 1/2 WATT UNLESS OTHERWISE SPE
RESISTANCE IN OHMS UNLESS OTHERWISE
K=IOOO OHMS M=IMEGOHM
CAPA Cl TANCE VALUES ONE 8 OVER IN MICRO
LESS THAN ONE IN MICROFARADS UNU
OTHERWISE NO TED.
@ SCREW DRIVER AO./USTMENT
Q
PANEL CONTROL
GR 1551-C Sound Level Meter_0025.djvu
5
R31
20K
R32
43K
R33
24K
Cl9
001
S-2
ENGRAVING
S-3
ENGRAVING
CABI.E
"an"
/( C52:..J!T+
50pf
50v
R54
33K
NOTE:
RESISTORS 1/2 WATT UNLESS OTHERWISE SPECIFIED
RESISTANCE IN OHMS UNLESS OTHERWISE SPECIFIED
K, 1000 OHMS M, I MEGOHM
CAPACITANCE VALUES ONE 8 OVER IN MICRO' MICROFARAOS,
LESS THAN ONE IN MICROFARADS UNLESS
OTHERWISE NOTED.
@ SCREW DRIVER ADJUSTMENT
Q PANEL CONTROL
R68
IOK
'----4
R72
300K
Figure 9. Schemati
Type 1551-C Sound-L
GR 1551-C Sound Level Meter_0026.djvu
C/9
001
O(V~~ OFF
Wv---
R31
20'1< """
6
'
<JI' 54
R33
24K
R32
43K
~
<I 2 ..
,
.......
·
~
--
81 82 ,:!!!:83
WH-8H~PPi·#L -~ H~G>'
1
"'"
-r-~1Cf<f1 ~ 02
-~=·~
~==~~~~==~~~~~-~~~~~~~~~~~~-~-1
CABLE
"n~
I
I[
I I
1
[I
I
1
11 vr (f'lc;..
~<J
" /
I
·~a
·.
2~
I
I'
I[ llt06F
-~~
. l1
112
F-'6'11.9V\Kr-'-
,t~~
R29
1 o~t"-f'
R30
_ _
04
JJ!T • • -:-"::o:!or_.._~
1ou r1.
1~2F
J...
·WH-8L-8K
_:::.~~
. .
L
~
R/6 I
~
-~ ·
"
•o•"
216K 1 t/8w t/8w 1
~
R/7 lOOK
I/8W
I
I
~I
CABLE"wH".:.-1
I
I
11
11
:: li /
I[
I I
I
I f---g--:'1!.~-----='-
£1~~~
~¥ j~l
0,. •
,....
~ aR-
H.l---!oa:"-:" .
C/1
220·
J05R
6%~
, I _ - - - - - -
R21
7.51<
""'fruR if+ /
I
~
'
,-
1 -
J;.
Jl,-
3!.-
-ll
[I
,IJj
I!
\
CABU"OR' \ \
rit \\
l.
ct6 ,..
.
I
&&ik
a
0"
20>R il
~ ~~
~ ~I /Y!K
.~~I
i
[[
I
II
11
II lTI CAsali
'"Ci
Afl li
I[
I[
_jdJ
20...
II
II lj
I
II
---·1
.
I
C
I/78
~
8 our
~: 3 /f lK~
~ 3 --'W'r---'JJ,j?
.~pzf -~R26
K ..
~I
~};
4 l3K
J.(
C/3 l60sd
.-K i/
WH-811-!IJI
4 _ ·
I
. ' .
/04R lOTH
l&-+
·
j
I
I
WH•IfD
w ·
~-
53
WH-vr
~
.
WH-OR
II-
! I
WH GH·BR ""E="
--- -- - - -- --
~-
-llm---l
~-c;--
-- -- -;;- _-_-
--1 c:;* J~~* :::~r:/
f%7--=-------,
C60
.I 11 ~
V53 R62 1\.01
r:~3~K
RIH
'cs.if::: '
./ R61
!.2M
18M
~- ,1~;.
20 GY
~24M tCb~
'
'Ci2
8
"
~
,-;rf
2
.~:.:~;;gO
18
J -
5 5
' -;
C67
.47 R64
390~
R67
300 u
R74
120K
R68
IOK
'-----<
R72
300K
R75
~
R76
-
!OK
470K ~
R77
20K
Figure 9. Schematic Diagram,
Type 1551-C Sound-Level Meter.
17
GR 1551-C Sound Level Meter_0027.djvu
S.•:tior
TYPE 1262-11 f'OWE~ SUPPlY
GR 1551-C Sound Level Meter_0028.djvu
Figure 10.
Interior View, T ype 1551
Sound-Level Meter.
18
F i gure 11. Componen t Layout on Etched Boord .
Complete board i s PIN 1551-2201.
NOTE
T he nu m be r shown on t h e foil side of the b oa rd i s n ot t h e part number for the complete assem bl y. The assembly n u mber is given in the cap ti on .
R77
R75
I
BLOCK
GR 1551-C Sound Level Meter_0029.djvu
Figure 10 . erior View, Type 1551-C
Sou n d-Level Meter .
BLOCK DIAGRAM
Figure 12.
Elementary Block l
GR 1551-C Sound Level Meter_0030.djvu
MICROPHONE
SENSITIVITY
ADJUSTMENT
Figure 12
~~:~:: .
Elementary Blo c k Diag r am.
R50
C51
R51
5.1 Il' movet andre saryfi the tw• cover to th e of the
Th e s : the reo
5 .
2 0
210-2
Sound·
Sound· is batt
BATT1 meter
5.3 81
5.4 N depenc
Supply nents : weight orB w voltag weldeJ listed
GR 1551-C Sound Level Meter_0031.djvu
Section 5
TYPE 1262-B POWER SUPPLY
5.1 INSTALLATION. Before using the Type 1262-B Power Supply, remove the batteries from the Sound-Level Meter(refer to paragraph 4.5), and replace the Sound- Level Meter in its case.
To attach the power supply to the Sound-Level Meter, it is necessary first to remove the cover from the power supply. To do this,loosen the two 10-32 screws at the ends of the power supply. Then slide the cover off, i.e., away from the engraved panel of the pCNler supply.
Two 1/4-28 binder-head screws are used to attach the power supply to the Sound-Level Meter. Insert these screws through holes in the back of the power supply and mating threaded holes in the Sound-Level Meter.
The six-terminal male connector in the power supply should slip into the receptacle on the Sound-Level Meter.
5.2 OPERATION. Connect the Type 1262-B input plug to a 105-125 (or
210-250) -volt, 50- 60-cycle power source. Turn the Power Supply and
Sound-Level Meter on by means of the power supply on-off switch. (The
Sound-Level Meter's on-off switch functions only when the instrument is battery-operated.) To check for proper operation, setthe METER-
BATTERIES switch to PL and FIL, checking that for each position the meter pointer falls within the white BAT area on the meter face.
5.3 SERVICE AND MAINTENANCE. Refer to pa\ragraph 4.1.
5.4 NOISE LEVEL. The internal noise level of the Sound- Level Meter depends on the quality of the power line supplying the Type 1262-B Power
Supply. With a good line (i.e.~ no transient humps or frequency components below 50 cps), the noise level will be 28 db or less for the 20-kc weighting position, below 26 db with C weighting, and below 24 db with A or B weighting.
If the Sound-Level Meter is operated in an area where the line voltage fluctuates abruptly and repeatedly (e.g. near devices such as spot welders), the internal noise level will appear to be higher than thevalues lusted above.
GR 1551-C Sound Level Meter_0032.djvu
1262-B P"AitTS liST
ClR Fo<!Mfg_
---~ ~-----
•J4jft
---i---.-.-
f~-
GR 1551-C Sound Level Meter_0033.djvu
Ref. No.
CAPACITORS
CIA
C1B
C2A
C2B
C2C
C3A
C3B
C3C
C4A
C4B
Description
Electrolytic, 300 flF 35 V
1200
Electrolytic, 2400 flF 15 V
1200
200
Electrolytic block, 100 flF 100 V
100
Electrolytic, 300 flF 35 V
1262-B PARTS LIST
GR Fed, Mfg.
Part No. Code Mfg. Part No.
4450-2400 37942 2021149S4Cl0Xl
4450-4200 37942 20-21339-99-6
4460-1500 80183 DFP, 200/100/100 flF
4450-2400 37942 2021149S4Cl0Xl
PILOT LAMP
P1 115V 8390-0600 24446 NE-51
RESISTORS
Rl
R2
R3
R4
R5
R6
R7
R8
R9
RIO
Rll
Composition, 100 kn ±5% 1/2 W
Composition, 200 n ±5% 1/2 W
Composition, 130 n ±5% 1/2 W
Composition, 43 n ±5% 1/2 W
Composition, 47 n ±5% 1/2 W
Composition, 180 n ±5% 1/2 W
Composition, 1.5 kn ±5% 1/2 W
Composition, 1.5 kn ±5% 1/2 W
Composition, 51 kfl ±5% 1/2 W
Composition, 220 n ±5% 1/2 W
Composition, 430 n ±5% 1/2 W
RECTIFIERS
RXl
RX2
RX3
RX4
RX5
SWITCH
Sl Toggle
6100-4105 01121 RC20GF104J
6100-1205 01121 RC20GF201J
6100-1135 01121 RC20GF131J
6100-0435 01121 RC20GF430J
6100-0475 01121 RC20GF470J
6100-1185 01121 RC20GF181J
6100-2155 01121 RC20GF152J
6100-2155 01121 RC20GF152J
6100-3515 01121 RC20GF513J
6100-1225 01121 RC20GF221J
6100-1435 01121 RC20GF431J
6080-2800 77638 2J9
6080-2700 77638 6YI
6080-2900 77638 6J2
6080-2500 77638 IYI
6080-2600 77638 3YI
7910-1300 04009 83053-SA
TRANSFORMER
T1 0746-4310 24655 0746-4310
Fed. Stock
No.
5910-822-2691
5910-822-2692
5910-822-2691
62 40-223-9100
5905-195-6761
5905-279-2674
5905-252-5436
5905-279-1887
5905-252-4018
5905-279-3514
5905-841-7461
5905-841-7461
5905-279-3496
5905-279-3513
5905-279-3512
5930-909-3510
GR 1551-C Sound Level Meter_0034.djvu
RESISTORS 1/2 WATT UNLESS OTHER W ISE
~~F~a
ReSISTANCE IN OHMS UNLESS OTHERWISE
SPECIFIED
K1000 OHMS
CAPACITANCE VALUES I N NICRCFARA OS
UNLESS OTHEHWISE SPECIFIED
T·l
~ ~ --.~ ---.----~~-~ ~---------
INPUT
115r. OR 2J0fl.
50fiOc .
R3
IJD
112
200
R4
4J
C · 18
1 3 00
0
I~;
FOR 230v. INPUT CONNECT #2 T O I>:J
F i gu re 14 . Lay ou t of C om pone n t s on Et ched Boord .
F i g u r e 1 3 . S ch e ma ti c D iagra m , Typ e
R£CTIFIE~
I
·•
FI LTE R f
"
~ f
: ReCTIFIER idh *~ j
~
RECTIFIER j
~
GR 1551-C Sound Level Meter_0035.djvu
A4
"lC-48
1 3 0 0
~~+ RX-5
A3 y~
R6
/80 l>
R4
43
C-18
300
AI
0
R5
47
C-2A
1200
=;;:~j;gg
A6
:~
RX-3
~
..-IUD
__j+
I
R8
1.5K c
;;:
C-38
IOU
";;: C-3C
100
R!J
51K
A2
GY
VT c
C-2C
' 1200
RO
~.; l
-,
: c Diagram, Type 1262-B Power Supply.
';TIF!ER
••
FILTER
RECTIFIER
VOLTAGE
LIMITER
Figure 15. Elementary Schematic Diagram.
, . - - - - • • PL-2,1FI
~
19
GR 1551-C Sound Level Meter_0036.djvu
FEDERAL MANUFACTURER'S CODE
From Federal Supply Code for Manufacturers Cataloging Handbooks H4-1
(NMDe to Code) and H4-2 (Code to Name) as supplemented through August, 1968.
Code
01121
01295
07261
07263
07387
07595
07828
07910
079B3
07999
09408
09823
12498
12672
12954
13327
14655
14674
14936
15238
17771
19396
19644
19701
21335
22753
23342
24446
24454
24455
24655
26806
38443
40931
42498
43991
Manufacturer
Jones Mfg. Co, Ch1cago, IllinOIS
Walsco Electronics Corp, L.A .. Calif.
Schweber Electronics, Westburg, L.l., N.Y.
Aerovox Corp, New Bedford, Mass.
Alden Products Co, Brockton, MaliS.
Allen-Bradley, Co, Milwaukee, Wise.
Te)(as Instruments, Inc, Dallas, Texas
FerroxcubeCorp, Saugentes, N.Y. 12477
Fenwal Lab Inc, Monon Grove, Ill.
Amphenol Electron Corp, BroadvteW, Ill.
Fastex, Des Plaines, Ill. 60016
G. E. Sem1con Prod, Syracuse, N Y 13201
Grayburne, Yonkers, N.Y. 10701
Pyroftlm Resistor Co, Cedar Knolls, N.J.
Claire ... Corp, New York, N.Y. 10001
Arrow· Hart & Hegeman, Hartford, Conn.
06106
Motorola, Phoenix, Ar1z, 85008
Engr'd Electronics, Santa Ana, Calif. 92702
Barber-Colman Co, Rockford, Ill. 61101
Wakefteld Eng, Inc, Wakefield, MaS5. 01880
D1g1tron Co, Pasadena, Calif.
Eagle S1gnal ( E.W. Bliss Co), Baraboo, Wtsc.
Avnet Corp, Culver Ctty, Calif, 90230
Fairchild Camera, Mountatn Vtew, Calif.
Btncher Corp, No. Los Angeles, Calif.
Amer Semtcond, Arlington Hts, Ill. 60004
Bodine Corp, Bridgeport, Conn. 06605
BOdtne Electnc Co, Ch1cago, Ill. 60618
Cant Device Corp, Hawthorne, Calif.
State Labs Inc, N.Y., N.Y. 10003
Borg lnst .. Delavan, W1sc. 53115
Vemaltne Prod Co, Franklin Lakes, N.J.
G.E. Semiconductor, Buffalo, N.Y.
Star•Trontcs Inc, Georgetown, Mass. 01830
Burgess Battery Co, Freeport, Ill.
Burndy Corp, Norwalk, Conn, 06852
C.T.S. of Berne, Inc, Berne, Ind. 46711
Chandler Evans Corp, W. Hartford, Conn.
National Semiconductor, Danbury, Conn,
Crystalomcs, Cambridge, Mass. 02140
RCA, Woodbrtdge, N.J.
Clarostat Mfg Co, Jnc, Dover, N.H. 03820
Dtckson Electromcs, Scottsdale, Ariz.
Solltron Devices, Tappan, N.Y. 10983
ITT Semicondlctor-s, W.Palm Beach, Fla.
Corneii·Dubllier Electric Co, Newark, N.J.
Corning Glass Works, Cormng, N.Y.
General Instrument Corp, HICksville, N.Y.
ITT, Semiconductor Div, Lawrence, Mass.
Cutlet-Hammer Inc, Milwaukee, Wtsc, 53233
Spruce Ptne Mica Co, Spruce Ptne, N.C.
Singer Co, Diehl D1v, Somerville, N.J.
Jillnots Tool Works, Pakton Dtv, Chicago, Ill.
LAC ElectrOniCS, Horseheads, N.Y.
Electra Mfg Co, Independence, Kansas 67301
Fafnir Bearjng Co, New Briton, Conn,
UID Electronics Corp, Hollywood, Fla.
Avnet Electronics Corp, Franklin Park, Ill.
G.E., Schenectady, N.Y. 12305
G.E .. Electronics Camp, Syracuse, N.Y.
G. E. !Lamp Div), Nela Park, Cleveland, Ohto
General Red to Co, W. Concord, Mass. 01781
American Zentet Inc, Costa Mesa, Calif.
Hayman Mfg Co, Kemlworth, N.J.
Hoffman Electrontcs Corp, El Monte, Calif.
I.B.M, Armonk, New York
Jensen Mfg. Co, Chicago, Ill. 60638
G. E. Camp, Owensboro, Ky. 42301
Constanta Co, Mont, 19, Que.
P.R. Mallory & Co Inc, Indianapolis, Ind.
Marlln·Rockwetl Corp, Jamestown, N.Y.
Honeywell Inc, MinneapOliS, M,nn, 55408
Muter Co, Chicago, Ill. 60638
National Co, Inc, Melrose, Mass. 02176
Norma-Hoffman, Stanford, Conn. 06904
Code
54715
56289
63743
65083
70485
70563
71294
71400
71744
71785
71823
72619
72699
72765
73138
73445
73559
73899
74193
74861
74970
75042
75382
75915
76005
76149
76487
76545
76684
77166
77263
77042
77630
77638
78189
78277
78488
79963
80030
B0048
80131
80183
Manufacturer
RCA, New York, N.Y.10020
Raytheon Mfg Co, Waltham, Mass. 02154
Sangamo Electric Co, Springfield, Ill. 62705
Shallcross Mfg Co, Selma, N.C.
Shure Brothers, Inc, Evanston, Ill.
Sprague Electric Co, N. Adams, Mass.
Thomas and Betts Co, Elizabeth, N.J. 07207
TRW Inc, !Accessories Div), Cleveland, Oh10
Tornngton Mfg Co, Torrington, Conn.
Union Carbide Corp, New York, N.Y. 10017
Untted-Carr Fastener Corp, Boston, Mass.
Vtctoreen Instrument Co, Inc, Cleveland, 0.
Ward Leonard Electric Co, Mt. Vernon, N.Y.
Westmghouse (Lamp Dtv), Bloomfield, N.J.
Weston Instruments, Newark, N.J.
Atlantic·lndta Rubber, Chicago, Ill. 60607
Ampente Co, Union City, N.J. 07087
Belden Mfg Co, Chtcago, Ill. 60644
Bronson, Homer D, Co, Beacon Fall5, Conn.
Canfteld, H.O. Co, Clifton Forge, va. 24422
Bussman (McGraw Edison), St. Louis, Mo.
ITT Cannon Elec, L.A., Calif. 90031
Centralab, Inc, Milwaukee, Wise, 53212
Continental Carbon Co, Jnc, New York, N.Y.
Cote Coil Co Inc, Prov1dence, R.I.
Chicago Mtntature Lamp Works, Chicago, Ill.
Cinch Mfg Co, Chtcago, Ill. 60624
Darnell Corp, Ltd, Downey, Calif. 90241
Electro Motive Mfg Co, Wilmington, Conn,
Nytrontcs Inc, Berkeley Heights, N.J. 07922
Dialtght Co, Brooklyn, N.Y. 11237
Generallnstr Corp, Newark, N.J. 07104
Drake Mfg Co, Ch1cago, Ill. 60656
Hugh H. Eby Inc, Philadelphia, Penn. 19144
Elastic Stop Nut Corp, Union, N.J. 07083
Ene Technological Products Inc, Erie, Penn,
Beckman Inc, Fullerton, Calif, 92634
Amperex Electronics Co, Hicksvdle, N.Y.
Carltng Electric Co, W.Hartford, Conn.
Elco Reststor Co, New York, N.Y.
JFD Electron,cs Corp, Brooklyn, N.Y.
Heinemann Electnc Co, Trenton, N.J.
Industrial Condenser Corp, Chicago, Ill,
E. F. Johnson Co, Waseca, Minn. 56093 lAC Jnc, Philadelphta, Penn, 19108
Kulka Electric Corp, Mt. Vernon, N.Y.
Lafayette Industrial Electronics, Jam\ca, N.Y. ltnden and Co, Provtdence, R.I.
Litteltuse, Inc, Des Pla1nes, Ill. 60016
Lord Mfg Co, Erie, Penn. 16512
Mallory Electric Corp, Detroit, Mtch, 48204
James Millen Mfg Co, Malden, Mass, 02148
Mueller Electric Co, Cleveland, Ohio 44114
National Tube Co, Pittsburg, Penn,
Oak Mfg Co, Crystal Lake, Ill.
Patton MacGuyer Co, Providence, R.I.
Pass-Seymour, Syracuse, N.Y.
Pierce Robens Rubber Co, Trenton, N.J.
POSitive Lockwasher Co, Newark, N.J.
Ray•Q-Vac Co, Madison, Wtsc.
TRW, Electromc Camp, Camden, N.J. 08103
General Instruments Corp, Brooklyn, N.Y.
Shakeproof (Ill. Tool Works), Elgtn, Ill. 60120
Sigma Instruments Inc, S.Bratntree, Mass,
Stackpole Carbon Co, St. Mary5, Penn.
Ttnnerman Products, Inc, Cleveland, OhiO
RCA, Rec Tube & Semlcond, Harrison, N.J.
Wiremold Co, Hartford, Conn, 06110
Zierick Mfg Co, New Rochelle, N. Y,
Prestole Fastener, Toledo, Oh10
Vtckers Inc, St. Louis, Mo.
ElectroniC Industries Assoc. Washtngton, D.C.
Sprague Products Co, No. Adams, Mass.
Motorola Inc, Franklin Park, IlL 60131
Standard Dtl Co, Lafayette, Ind.
Bourns Inc, R1verside, Calif, 92506
91637
31662
91929
92519
92678
93332
93916
94144
94154
95076
95121
95146
95238
Code
80431
80583
81143
81349
81350
81831
81840
82273
82389
83033
83058
83186
84411
84835
86687
86800
88419
88627
89482
89665
96341
Manufacturer
A1r Fdter Corp, M1lwauku, Wise. 53218
Hammarlund Co, Inc, New York, N.Y.
Beckman Instruments, Inc, Fullenon, Calif. lnternattonat lnsturment, Orange, Conn,
GrayhLII Inc, LaGrange, tit. 60525 lsolantite Mfg Corp, St1rling, N.J. 07980
Military Sp~Kifications
Joint Army· Navy Specifications
Columbus Electronics Corp, Yonkers, N, y.
F11tron Co, Flushing, L.l., N.Y. 11354
Ledex Inc, Dayton, Ohio 45402
Barry-Wrigt1t Corp, Watenown, Mass.
SYivan1a Elec Prod, Emporium, Penn. lndtana Pattern & Model Works, LaPort, Ind.
SWttchcraft Inc, Chicago, Ill. 60630
Metals & Controls Inc, Attleboro, Mass.
Mtlwaukee Reststor Co, Milwaukee, W1sc.
Metssner Mfg, (Maguire lnd) Mt. Carmel, Ill.
Carr Fanener Co, Cambridge, Mass.
Vtctory Engineering, Spnngfield, N.J. 07081
Bearing SPecialty Co, San FranctSCO, CaHf,
Solar Electnc Corp, Warren, Penn.
Un1on Carbtde CorP, New York, N.Y. 10017
Nat1onal Electrontcs Inc, Geneva, Ill.
TAW Capac1tor Dtv, Ogellala, Nebr.
Lehigh Metal Prods, Cambridge, Mass. 02140
TA Mtg Corp, Los Angeles, Calif.
Prectslon Metal Prods, Stoneham, Mass. 02180
RCA \Elect, Camp & Dev), Harrison, N.J.
REC Corp, New Rochelle, N.Y. 10801
Cant Electronics Corp, Brooklyn, N.Y. 11222
Cutler-Hammer Inc, ltncoln, IlL
Gould Nat. Battenes Inc, Trenton, N.J.
Corneii·Dubdler, FuQuay·Varina, N.C.
K & G Mfg Co, New York, N.Y.
HOitzer-Cabot Corp, Boston, Mass.
United Transformer Co, Chicago, Ill.
Mallory Capacitor Co, lndtanapohs, Ind.
Westinghouse E!ectnc Corp, Boston, Mass.
Hardware Products Co, Reading, Penn. 19602
Conttnental Wire Corp, York, Penn. 17405
ITT (Cannon Electric Inc), Salem, Mass.
Johanson Mfg Co, Boonton, N.J. 07005
Au get Inc, Attleboro, Mass. 02703
Chand;er Co, Wethersfield, Conn. 06109
Dale Electromcs Inc, Columbus, Nebr.
Elco Corp, Willow Grove, Penn.
General Instruments, Inc, Dallas, Texas
Honeywell Inc, Freeport, IJJ.
Electra lnsul Corp, Woodstde, l.l., N.Y.
E.G.&G., Boston, Mass.
~~~~~~ap~~~c~c:ocd~: ~nee~ ~~~~~~-~~;s0 013
Raytheon Co, Components Dtv, Quincy, Mass.
Tung Sol Electric Inc, Newark, N.J.
Garde Mfg Co, Cumberland, R.I.
Quality Components Inc, St. Mary's, Penn.
Alec Electronics Mfg Co, Lawrence, Mass.
Continental Connector Corp, Woodside, N.Y.
V+tramon, Inc, Bridgeport, Conn,
Methode Mfg Co, Ch1cago, Ill.
General Electric Co, Schenectady, N.Y.
Anaconda A mer 8ra55 Co, Torrtngton, Conn.
H1-Q D1v, at Aerovox CorP, Orlean, N.Y.
Texas Instruments Inc, Dallas, Texas 75209
Thordarson-Meiuner, Mt. Carmel, ill,
Microwave As5octates Inc, Burltngton, Mass.
Amphenol Corp, Jonesville, Wise, 53545
Military Standards
Seaiectro Corp, Mamaroneck, N.Y. 10544
Compar Inc, Burltngame, Caltf,
North Hills ElectroniCS tnc, Glen Cove, N.Y.
Transttron Electromcs Corp, Melrose, Mass.
Varian, Palo Alto, Calif. 94303
Atlee Corp, Winchester, Mass. 01890
Delevan EJectromcs Corp, E. Aurora, N.Y.
9/68
GR 1551-C Sound Level Meter_0037.djvu
et7 :ae•-••oo
GENERAL RADIO
et7 •••-7•oo
. t
. I
- J
;
DISTRICT OFFICES
I
METROPOLITAN
NEW YORK*
'~
\
'
·
CHICAGO*
9440 W . Fosler A ...
845 IJooad AY en u e
R i dg e fi e ld, New Ieney 076S7
Tele p hone N .
r . 2J2 964-2722
Ch i cago, Ill i nois 60656 lJ 2 992 0800
N.J
. 20J 94l lJ 4 0
CLEVELAND
SYRACUSE
P ic lra r d iJ u i ld i ng foal Molloy Rood
S7racuse . New Y o r l J 3 2 fl
Jeleplton e 31.5 4.54
9323
5579 ,._,I Rood
Cfe Y eland, Ohia 4 4 J 29
T el ephen e 2J6 886 0JSO
NEW ENGLAND*
22 lalrer AY en ue
Werf Concord, ltl a uochurellr OJ71JJ
Telephon e 6r 7 646 0550
LOS ANGELES*
JOOO North S e wa r d Sl rH I los Angel e s, California 90038
2J3
FRANCISCO
626 San Anton i o Rood
PHILADELPHIA
Fo r t Waslt i nglon lndurrriol Par ..
Fori Waslt i n.,on , P e nnsyiYania l9034
T e l.,hane 2 J'I 646-8030
WASHifiGTON*
AND BALTIMORE
RoclcYille Pilce flochille , Maryland 208.52
T el e phone 30J 946 J600
ORLANDO
J J 3 fasl Coloni a l OriYe
Orlando, Florida 32801
Telephene 305 425 4671
DALLAS*
2600 Slemmonr Fre e way, Su ite 2JO •
Dallas, T u as 75207
Telepltone 2J4 637 2240
TORONTO*
99 flnool Porlway
To ro nto J .5
, Onta r i o, Canada
Telepltone 4J6 247 2J7l
• Rept~lr s ervic e s are GYailabl e al thes e dist rl cl a lll cn .
I 255 lai r d IJoul e Yard
Town of Mount Re,.al
, Quebec, Canada
Telephone 5J 4 7373 673
OTTAW A T e l ephon e
6U ~ 33-4237
General Radio Company (Overseas} , 8008 Zurich , Switzerland
G e neral Radio Company (U.K.J Limited, Bourne End, Buckinghamshire, E11gland
Repres e ntatives in Principal Overseas Countries
Prin ~ ;,. U SA
-~ -~ ~----~ ~ -
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement