advertisement
▼
Scroll to page 2
of 128
PURITAN BENNETT. ©1991 PURITAN-BENNETT CORPORATION COMPANION" 492a AND COMPANION" 590 OXYGEN CONCENTRATORS SERVICE MANUAL PART NUMBER 492779 Copyright© 1991 by Puritan-Bennett Corporation. All rights reserved. The information contained in this manual >. is the sole property of Puritan-Bennett Corporation and may not be reproduced without the permission of the Company. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 PREFACE This manual provides the information needed to service the Puritan-Bennett Companion 492a and 590 Oxygen Concentrators. This information is intended for use by technicians or personnel qualified to repair and service medical equipment. The information is not all-inclusive and may not be applicable to future Companion 492a and 590 models. Technicians should direct any questions regarding Companion concentrator service to Puritan-Bennett Oxygen Concentrator Division: 1-800-248-0890 or 314-739-7070. The information contained in this document, including performance specifications, are subject to change without notice. Puritan-Bennett makes no warranty of any kind with regard to the material in this manual, including but not limited to the implied warranties of merchantabilities and fitness for a particular purpose. Puritan-Bennett shall not be liable for errors contained herein or for incidental or consequential damages in connection with providing this manual or the use of material in this manual. DEFINITION OF STATEMENTS Statements in this manual preceded by the following words are of special significance. WARNING Means there is the possibility of injury or death to yourself or others. CAUTION Means there is the possibility of damage to the unit or other property. NOTE Indicates point of particular interest for more efficient and convenient operation. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 ii a SECTION 1. SECTION 2. SECTION 3. SECTION 4. TITLE TABLE OF CONTENTS INTRODUCTION AND GENERAL INFORMATION ........... 1.1 General Product Description .. ........24224 442420 1.1.1 Optional Accessories and Literature . . . . . . . . . .. 1.2 Performance Specifications . . . . . .............. 1.3 Unpacking and Receiving Inspection . . . . . .. ........ 1.4 Operating Procedures . . . . . . .. .............. 1.4.1 OperationalCheck . . .................. 1.42 Operating Instructions . . . ............... 1.5 Routine Maintenance . .. .. .................. 1.6 Recommended Tools and Test Equipment ........... ТНЕОВУ ОР ОРЕВАТЮМ ........................ 2.1 Overview of Concentrator Operation . . . ..... . _—. ..e. 2.2 Pneumatic System Components . . . . . ............ 2.3 Electrical System Components . . . . . . .. .......... 2.4 Concentrator System Operation . . . . . ............ PERFORMANCE VERIFICATION . . .................. 3.1 3.2 3.3 3.4 Performance Verification . . . . . .. .. ............ 3.1.1 Oxygen Concentration . . . . .............. 3.12 Pneumatic ..... 44041 41 41 01 44 140 3.1.3 Electrical . ...... aerea. 3.1.4 Ventilation System . . ...... .. .. e... 315 Flow ...... aa aa 3.1.6 Sound .....2 222141 LL LL LL LL LL 140 Operational Checklist .. ... . .. .. ............ 3.21 Low Concentration . . ... .............. 3.22 ConcentratorWontRun . . ............... 3.2.3. Improper Operation . . . ................ Troubleshooting Guide . . . . . . . . .............. Balancing Orifices . . . . . . . . eee ea SERVICE AND REPAIR ..... . .. ............... 1. 2. 3. Cabinet . . . . . aa aaa Circuit Breaker .. .. 1... 12111101 LL LL LL 11110 Hour Meter . . . . . . . ea aaa eee 1-10 1-14 1-17 2-1 2-2 2-7 2-16 2-20 3-1 3-2 3-2 3-3 3-8 3-8 3-9 3-10 3-10 3-11 3-11 3-11 3-16 Companion 492a/590 Oxygen Concentrators iii Service Manual 5-15-91 TITLE SECTION 4. SERVICE AND REPAIR (continued) 4. Audio Alarm . ..... _... 2 eee. 5. Flowmeter Flow Control Valve ............ 6. Flowmeter . . . . .. ................. 7. Power Switch . . ................... 8. Control Printed Circuit Board _............. 9. Compressor Platform Assembly . .. ......... 10. Compressor(492a) . ................. 11. Согпргез5ог (590)... . . . ............. 12. Capacitor .. .... ................. 13. Cooling Fan .. .................... 14. Pressure Switch . . . . . ............... 15. Pressure Regulator . . . ............... 16. Solenoid . . ... .. e... eee. 17. Sieve Canister .. .. .. 2.244441 44122220 18. Pilot Маме... . .... .. .... ......... 19. Restrictor Tube Assemblies . . . . . .. .... ... 20. Outlet Gas Filter . . . . . .. .. ........... 21. Shock Mount . . ................... 22. CompressorCup Seals (580) . . .. ......... SECTION 5. TECHNICAL BULLETINS SECTION 6. OXYGEN CONCENTRATION INDICATOR (OCI) 6.1 General product Description ............. 6.2 Theory of Operation .. .... 4444424420 6.3 Performance Verification ............... 6.4 Troubleshooting and Repair .. .. .......... TABLE OF CONTENTS (Continued) 6.4.1 Troubleshooting ............... 6.42 Вера... .................. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 PAGE 4-6 4-6 4-7 4-7 4-9 4-11 4-15 4-19 4-20 4-20 4-21 4-24 4-25 4-28 4-30 4-31 4-32 4-32 FIGURE 1-1 1-2 1-3 1-4 1-5 1-6 1-7 1-8 1-9 1-10 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 2-11 2-12 2-13 2-14 2-15 2-16 2-17 2-18 2-19 2-20 2-21 3-1 3-2 3-3 3-4 3-5 3-6 4-1 4-2 LIST OF ILLUSTRATIONS TITLE Companion 492a Front View . . . .. .... .. Companion 492a Rear View ........... Side Panel View . . . . .............. Oxygen Concentration Testing .......... Oxygen Outlet Adapters ............. Humidifier Adapter Attachment . . . . . . . . .. Humidifier Attachment . . . . . . ........ Oxygen Tubing Attachment to Humidifier . . . . . Oxygen Tubing Attachment to Adapter . . . . . . Side Panel Compartment Filters . . . . . . . .. Pneumatic Diagram . . .............. Electrical Diagram . . . . . ............ Side Panel View . 1... 222112142440 Compressor Inlet Filter .......... .... Compressor Drawer, Front View ......... Compressor Drawer, Bottom View ........ Pilot Air Filter . . . . . . . ............ Pilot Valve Operation . . ............. Sieve Canisters . . ................ Expansion Chamber and Muffler . . . . . . . .. Restrictor Tubes . . . . .............. Product Canister, Regulator, and Vacuum Check Valve Right Cabinet, Interior View . . . . ........ Control Panel . . ................. CoolingAirFlow . . . ............... Cooling Fan . . .................. Filter Compartment . . .. ............ Sieve Canister 1, Concentrating Cycle .. . . . Balance Cycle . . .. ............... Sieve Canister 2, Concentrating Cycle . . . . . . Oxygen Concentration Testing .......... Pneumatic Test Kit Connection/Testing . . . . . . Pneumatic Test Kit Connection/Testing . . . . . . Pressure Switch Testing . . . . ... ...... Outlet Pressure Testing . . . . .......... Flowmeter Testing . . .............. Companion 492a/590 Exterior .......... Left Cabinet Interior . . . . ............ naa aa e ae 1-11 naa aaa ea a 1-12 aaa ae 1-12 ea aa eee 1-13 ae ae . 1-13 aaa en aa 1-16 aa ae a e aaa 2-5 Зе. 2-7 aaa ea e ea 2-3 aa ae 2-9 ae aa aaa 2-11 aaa aa 2-11 aaa ea 2-12 aaa aaa 2-13 aaa e ae 2-13 a a eee 2-15 aaa a 2-15 aan a ea 2-17 ea a ea 2-17 aa ea aaa 2-18 aaa a 2-19 aaa ee 2-20 aaa aa aaa 2-23 aaa ee 2-25 ea a aaa 2-27 aaa aa e a ea 4-35 Век, 4-39 Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 FIGURE 4-3 4-4 4-5 4-6 4-7 4-8 4-9 4-10 6-1 6-2 6-3 LIST OF ILLUSTRATIONS (continued) TITLE PAGE Right Cabinet Interior . . . . . . . . .. . . .............. 4-43 Compressor Platform Assembly . . . . . . . .............. 4-47 Companion 492a Compressor . . . . . . ................ 4-49 Companion 590 Compressor . . . . . .. .............. 4-51 Compressor Cylinder-Head Screw Torque Sequence . . ........ 4-53 Companion 590 Compressor Retaining Plate indexing . . . . . . . .. 4-54 Pressure Regulator . . . . . . . re... eee e ae 4-55 Pilot Valve .... 1.222224 1 4 44 4 4 44 ea ea e ee 4-56 492a Front Panel with OCI . ...... . 2.2... ...—.———— 6-2 OCI Circuit Board Installed ......... . 2... 6-3 OCI System .... 2.240244 44 4 4444 4 4 4 4 14e 144 6-8 Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 vi LIST OF TABLES TABLE TITLE PAGE 1-1 Performance Specifications .. .. .... 4244424444 4 44 46 1-4 1-2 Maintenance Schedule ........ 0202444 444 4 4 4 44 4420 1-14 1-3 Recommended Tools and Test Equipment . . . . . . .......... 1-17 3-1 Troubleshooting Guide . . . . . . . . .. ne nooo 3-12 4-1 Service Procedures .. ..... 2.221122 21441144 1 44440 4-3 6-1 Service Procedures .. ..... 2202022144 4 444 4 111120 6-6 Companion 492a/590 Oxygen Concentrators vii Service Manual 5-15-91 Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 viii (blank) SECTION 1. INTRODUCTION AND GENERAL INFORMATION This section provides introductory information on the Companion 492a and 590 Oxygen Concentrators, including optional equipment, performance specifications, preparation for use, operating procedures, routine maintenance, and recommended tools and test equipment. 1.1 General Product Description The Companion 492a and 590 Oxygen Concentrators provide supplemental oxygen at high concentrations. Each concentrator is equipped with a flowmeter calibrated in .25 liter increments with maximum settings of 4 lpm for the 492a and 5 Ipm for the 590. Boththe 492a and 590 are compact, lightweight units equipped with four wheels anda handle for ease in delivery to the home and in transport by the patient at home. The unit's height of 25.4 inches (64.5 cm) enables the operator to reach the controls with ease (Figure 1-1). In addition, the filter compartment on the side of the unit makes accessible, without disassembly of the unit, the compressor filters, hour meter, battery, and circuit breaker. Always observe the WARNING and CAUTION labels affixed to the front and rear panels as follows: WARNING: This machine produces oxygen. Keep away from heat and open flames. Do not smoke near patient or machine. Do not oil this machine. CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. CAUTION: Do not use in the presence of fluids. CAUTION: Electrical shock hazard. Do not remove cover. Refer service to qualified service personnel. Companion 492a/590 Oxygen Concentrators 1-1 Service Manual 5-15-91 Figure 1-1: Companion 492a Front View Figure 1-2: Companion 492a Rear View NOTE With the exception of flowmeter range and cabinet labeling, the exterior of the Companion 492a and 590 are identical. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-2 r ~~ 1.1.1 Optional Accessories and Literature The following Puritan-Bennett accessories and literature may be used with the Companion 492a and Companion 590. Literature Part Number Companion 432a Patient Instruction Manual 492931" Companion 492a Patient Instruction Manual with O.C.1.** 492778" Companion 590 Patient Instruction Manual 492932" Companion 590 Patient Instruction Manual with O.C.1.** 492827* Companion Concentrator Patient Checklist 799609 Humidifier Adapter lilustration Sheet 799660 Accessory Adapter, Humidifier (standard) 492922* Adapter, Humidifier (long) 492392 Tail Piece 492587 Wing Nut 492588 Humidifier, Bubble (disposable) 001083 Oxygen Concentration Indicator (O.C.1.) (installation required)** 492888 Companion Flowsplitter 493219 Companion Flowsplitter Plug 493220 Companion Flowsplitter Nozzles: 5 LPM 493221 ‚75 СРМ 493222 1.0 LPM 493223 1.5 LPM 493224 2.0 LPM 493225 2.5 LPM 493226 3.0LPM 493227 3.5LPM 493228 В Adapter, Humidifier (Long) Adapter, Humidifier (Standard) Plug | i 4 Humidifier, Bubble (Disposable) Companion Flow Nozzle Tailpiece & Wingnut *Appropriate part supplied with unit **Refer to Section 6 for information conceming ОС! Companion 492a/590 Oxygen Concentrators 1-3 Service Manual 5-15-91 1.2 Performance Specifications The Companion 492a and 590 physical characteristics, environmental and electrical requirements, and pneumatic characteristics are listed in Table 1-1. In most cases the information provided in this table applies to both the 492a and 590. Any differences in characteristics or requirements between the two models are listed separately. TABLE 1-1. PERFORMANCE SPECIFICATIONS PHYSICAL CHARACTERISTICS PARAMETER Cabinet size: Height Width Depth Assembly weight 492a 590 Shipping container size: Height Width Depth Shipping weight 492a 590 Noise Level Cabinet construction POWER switch Audio alarm Oxygen Concentration Indicator (O.C.1.) Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 MEASUREMENT/CONDITION 25.4 in. (64.5 cm) 12.5 in. (31.75 cm) 16.5 in. (41.9 cm) 57 Ib. (25.8 kg) 59 Ib. (26.8 kg) 28.5 in. (72.4 cm) 14.6 in. (37.1 cm) 19.5 in. (49.5 cm) 65 Ib. (29.5 kg) 67 Ib. (30.4 kg) < 50 dBA at 1 meter Molded case, hinges open for easy service equipped with carry handle and 4 casters. Rocker-type standard. When ON, LED illuminates to indicate AC power. When unit is equipped with optional Oxygen Concentration Indicator (O.C.l.), Power Switch is a push-button type. When ON, LED illuminates to indicate AC power. Refer to section 6 for all information pertaining to OCI operation. When POWER switch is ON, sounds to indicate low pressure, which may be caused by cycle failure, or power disconnect. Powered by internal source with battery back-up. Refer to section 6 for information pertaining to OCI operation. ENVIRONMENTAL REQUIREMENTS PARAMETER Storage/shipping temperature Operating temperature Stabilization time Operating Altitude* ELECTRICAL REQUIREMENTS REQUIREMENT -40°F to 158°F (-40°C to 70°C) 50°F to 110°F (10°C to 43°C) 20 minutes minimum Up to 3000 ft. (314 m) above sea level without degrada- tion of performance. PARAMETER Power 492a 590 Consumption 492a 590 Battery Power Cord"* Circuit Breaker 492a 590 Control Printed Circuit Board fuse PNEUMATIC CHARACTERISTICS REQUIREMENT Voltage Frequency Operating Current (nominal) 120 + 10 V AC 60 Hz 2.9 amps 120 + 10 VAC 60 Hz 3.3 amps 350 W average 400 W average 9 V, long-life alkaline Double-insulated 5 amp, time-delay 6 amp, time-delay 63 mA, 250 VAC PARAMETER Flow rate 492a 590 Oxygen concentration* 492a 590 Static delivery pressure System operating pressure MEASUREMENT 0 to 4 Ipm with .25 liter increments 0 to 5 Ipm with .25 liter increments 95 + 3% for 1-3 Ibm 92 + 3% at 4 Ibm 95 + 3% for 1-4 ipm 90 + 3% at 5 Ipm 5 + 0.5 psig Cycles from approximately 14 to 29 psig “Atmospheric pressure at altitudes of 3000 ft. (914 m) above sea level may decrease oxygen concentration levels. **Unit is Canadian Standard Association (CSA) certified as being Double Insulated. Companion 492a/590 Oxygen Concentrators 1-5 Service Manual 5-15-91 1.3 Unpacking and Receiving Inspection Unpack and inspect the Companion 492a/590 as follows: a. по ео Examine shipping container for exterior damage. If container is damaged or contents are suspected of being damaged, contact carrier to request that inspection of damage be made. Photograph damaged container before concentrator is unpacked. Contact the shipping point immediately. Position shipping container so that shipping arrows are pointing up. NOTE If staples were used to seal the shipping carton, take care in opening and removing the staples so that the shipping carton will not be damaged. Open shipping carton top flaps and lay carton on its side. Hold container down with one hand. Grasp handle, and slide concentrator out of container. Stand upright on casters and remove yellow cap from oxygen outlet. Compare packing slip attached to container's exterior with shipment received. If any discrepancies exist, contact Puritan-Bennett immediately. Thoroughly inspect the exterior of the Companion 492a/590 for damage. SAVE ALL PACKING MATERIALS AND SHIPPING CONTAINER FOR REUSE. NOTE Contact Puritan-Bennett for a returned goods authorization (RGA) number if it is necessary to return a unit to the factory. Write the RGA number on the outside of carton before shipping. When required, repack the Companion 492a/590 as follows: ~~ ® ao OT a Disconnect any fitting or accessory at oxygen outlet. Replace oxygen outlet cap (yellow). Ensure that power cord is wrapped around rear holder. Ensure that side filter compartment door is securely in place. Place concentrator inside foam-lined shipping container. Close the two shorter container flaps, then the two longer flaps. Secure container with packing tape. CAUTION Report any problems found during inspection to freight carrier or contact Puritan-Bennett Oxygen Concentrator Division for assistance if necessary. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-6 1.4 Operating Procedures This section provides the operational check and the operating instructions for the Companion 492a/590. 1.4.1 Operational Check Perform the following steps before initial use of concentrator at patient site. NOTE The use of a properly calibrated oxygen analyzer is required. a. Ensure that the following filters are clean and securely in place. (Figure 1-3) Item Location Air inlet filter Side panel Compressor inlet prefilter Prefilter housing in filter compartment Compressor inlet bacteria filter Filter compartment b. Verity that all internal pressure has been vented by tuming the flowmeter knob fully counterclock- wise. The flowmeter ball should indicate zero flow. c. Set POWER switch on top panel to ON before connecting power cord to outlet. If audio alarm does not sound or is weak, replace nine-volt battery in side compartment. Set POWER switch to OFF. d. Connect power cord to AC outlet. NOTE If concentrator has been exposed totemperatures below 40°F (14°C), allow it to reach ambient temperature (approximately 30 minutes) before turning on. e. Set POWER switch to ON and flowmeter to maximum Ipm on scale. The audio alarm will sound for a maximum of one minute, then cease. If alarm does not sound or continues to sound after one minute, see section 3, Performance Verification and Troubleshooting. f. Verify that POWER switch L.E.D. is illuminated. If indicator is not illuminated, see section 3 Performance Verification and Troubleshooting. g. Note hour meter reading. (Figure 1-3) Companion 492a/590 Oxygen Concentrators 1-7 Service Manual 5-15-91 AIR INLET FILTER | COMPRESSOR INLET PREFILTER Figure 1-3: Side Panel View Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-8 COMPRESSOR INLET BACTERIA FILTER HOURMETER 1.4.1 Operational Check (continued) NOTE Allow concentrator to run (stabilize) for approximately 20 minutes before proceeding. п. Connect calibrated oxygen analyzerto concentrator’'s oxygen outlet as shown in Figure 1-4. Follow Oxygen Analyzer manufacturers recommendations for proper connection to concentrator. I After the 20 minute stabilization time, verify that the concentration is 92% + 3% at 4 Ipm for the 492a and 90% + 3% at 5 Ipm for the 590. j- If the unit is equipped with the optional OCI verify that the green (normal) L.E.D. is illuminated. If oxygen concentration is not as specified, see section 3.3, Troubleshooting. K. Ensure that air flow is emitted through cooling air exhaust vents at base of concentrator (side opposite air inlet filter). Perform routine maintenance as outlined in Table 1-2. If concentrator is operated in a dusty environment, the recommended frequency of routine maintenance should be increased. Figure 1-4: Oxygen Concentration Testing Companion 492a/590 Oxygen Concentrators 1-9 Service Manual 5-15-91 1.4.2 Operating Instructions NOTE For additional operating instructions, refer to the appropriate Patient Instruction Manual listed in subsection 1.1.1, Optional Accessories and Literature. Perform the following steps: a. Ensure that air inlet filter (Figure 1-3) is clean. b. Position concentrator near appropriate AC outlet so that power cord can be connected without using extension cord. WARNING Oxygen greatly accelerates combustion. Keep concentrator at least five feet (1.5 m) away from sources of heat, smokers, open flames, or electrical equipment that may spark or become heated during operation. Oil and grease are highly flammable. Do not allow contact with concentrator. CAUTION Keep air inlet filter and cooling air exhaust vents at least six inches (15.2 cm) away from walls or draperies. Obstacles next to unit impede room air from entering concentrator. c. Set flowmeter control knob to setting prescribed by physician. Do not change setting unless ordered by physician. NOTE Refer to 1.1.1., Optional Equipment, for Puritan-Bennett part numbers of accessories used in steps d and e. d. If humidified oxygen has been prescribed by physician, proceed as follows: 1) 2) 3) 4) 5) Remove humidifier adapter (Figure 1-5) from side panel filter compartment. Connect humidifier adapter to a properly filled bubble humidifier (Figure 1-6). Attach the humidifier and adapter to the oxygen outlet spout at unit's front (Figure 1-7). Connect oxygen tubing to humidifier outlet (Figure 1-8). Proceed to step f. e. If humidified oxygen has not been prescribed by physician, proceed as follows: 1) 2) 3) Place wing nut on tail piece to form tubing adapter (Figure 1-5). Connect tubing adapter to oxygen outlet at unit's front. Connect oxygen tubing to tubing adapter (Figure 1-9). Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-10 ‘a 1.4.2 Operating Instructions (continued) f. Connect power cord to AC outlet. NOTE If concentrator has been exposed to temperatures below 40°F (14°C), allow it to reach ambient temperature (approximately 30 minutes) before turning on. g. Set POWER switch to ON. The audio alarm will sound for a maximum of on minute, then cease. NOTE Allow concentrator to run (stabilize) for approximately 20 minutes before using. NOTE If concentrator is accidently tipped over, impact of hitting floor may activate the Printed Circuit Board relay and the audio alarm may sound. Return concentrator to upright position. Set POWER switch to OFF, and after five seconds back to ON. Alarm may sound for a maximum of one minute, then cease. \ HUMIDIFIER ADAPTER у Figure 1-5: Oxygen Outlet Adapters Companion 492a/590 Oxygen Concentrators 1-11 Service Manual 5-15-91 Figure 1-6: Humidifier Adapter Attachment Figure 1-7: Humidifier Attachment Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-12 Figure 1-9: Oxygen Tubing Attachment To Adapter Companion 492a/530 Oxygen Concentrators 1-13 Service Manual 5-15-91 1.5 Routine Maintenance Perform routine maintenance as outlined in Table 1-2. If the concentrator is operated in a dusty environment, the frequency of maintenance should be increased. TABLE 1-2 MAINTENANCE SCHEDULE EBEQUENCY НЕМ ACTION 1. Daily Humidifier bottle Refer to the appropriate manufacturer's operating P/N 001083 manual. 2. Daily Air inlet filter a. Remove from side panel (Figure 1-3). P/N 492672 b. Vacuum or wash with mild detergent and water. Rinse and air dry. d. Return to side panel. O 3. Monthly" Compressor inlet a. Remove filter compartment door of concentrator prefilter (Figure 1-3). P/N 492193 b. Pull up on inlet prefilter housing to disconnect from bacteria filter. (Figure 1-9) C. Remove both compressor inlet prefilters from the prefilter housing (Figure 1-9) and exchange the top filter for a new one. NOTE The bottom inlet prefilter should also be replaced if dirty. d. Reposition the inlet prefilters in the prefilter housing so that the new filter is on the bottom and the filter previously on the bottom is now on top. e. Reconnect inlet prefilter housing to bacteria filter by pushing down. 4. Every Six (6) Compressor inlet a. Remove filter compartment door of concentrator months bacteria filter (Figure 1-3). P/N 492190 b. Pull up on bacteria filter to disconnect from connector. c. Separate prefilter housing from compressor inlet bacteria filter (Figure 1-9). d. Replace bacteria filter. e. Insert prefilter housing onto bacteria filter. f. Reconnect filter assembly to filter connector, and reinstall compartment door. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-14 TABLE 1-2 MAINTENANCE SCHEDULE FREQUENCY 5. Every 1 Year 6. Every 15000 Hrs. 7. As needed 8. As needed 9. As needed ITEM Alarm battery P/N 492297 Outlet gas filter P/N 492141 Cabinet exterior Oxygen connect- ing tubing and cannula Cabinet interior ACTION a. Remove filter compartment door of concentrator (Figure 1-3). b. Unplug connector from battery. с. Remove and discard battery. d. Install new battery and connect plug to battery terminals. e. Reinstall filter compartment door. Refer to outlet gas filter service procedure Section 4. WARNING To prevent a possibility of electrical shock to the operator or damage to the con- centrator, disconnect power cord before cleaning. Wipe with damp cloth or sponge and mild house hold cleaner. Clean or replace. a. Open concentrator cabinet. Refer to the cabinet service procedure in Section 4. CAUTION Ensure that the compressed air supply used to clean the cabinet interior is clean and oil-free. b. Blow or vacuum any dust which may have accumulated inside the concentrator. Leave air inlet and exhaust vents un- obstructed. С. Close concentrator cabinet. “For those units with OCI installed, change the compressor inlet prefilter every 3 months or as necessary. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 | — INLET PREFILTERS INLET PREFILTER HOUSING INLET BACTERIA FILTER Figure 1-10: Side Panel Compartment Filters Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-16 1.6 Recommended Tools and Test Equipment Standard test equipment tools, and materials used to test and service the Companion 492a/590 are listed below. If equipment other than specified is used, the substitute must be equal to or better than that listed. TABLE 1-3. RECOMMENDED TOOLS AND TEST EQUIPMENT TEST EQUIPMENT DESCRIPTION Digital Multimeter Oxygen Analyzer Test Flowmeter Stopwatch Pneumatic Test Kit SPECIAL TOOLS Kit, Pop-Rivet Gun Pliers, Truarc (for internal retaining ring with 1/8-inch hole) Wrench, Torque #0 Phillips screwdriver #2 Phillips screwdriver (min. shaft length of 7 in. recommended) #2 Magnetic Phillips screwdriver SERVICE MATERIALS MODEL/PART MANUFACTURER NUMBER John Fluke Co 8000A Seattle, WA Hudson 6400, cell 5500 Sierra Instruments 820 Top Trak Local Supplier NA Puritan-Bennett P/N 492381 Jensen Tools, Inc Y 339B750 Waldes Kohinoor, Inc 0309 (90” tip) Alcohol, Isopropyl Loctite Swabs, Cotton Pipe Cement Leak Detector Cabinet Touch-Up Paint (4 0z) Torque Controls TS-30 Local Supplier NA Local Supplier NA Local Supplier NA Local Supplier NA Loctite Corp 242,680,262 Hardwood Products, Inc 806 Flouramics Inc. LOX 8 Puritan-Bennett Corp 775272 Puritan-Bennett Corp 492648 Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 1-18 (blank) SECTION 2. THEORY OF OPERATION This section details the operational theory for the Companion 492a/590 Oxygen Concentrators. It includes an overview of the operation, the pneumatic system and components, the electrical system and components, the pressure-swing process, and the safety features. NOTE Numeric values found in this section are nominal values used for descriptive purposes only. Due to atmospheric pressure changes, compressor strength and other factors, pressure characteristics may vary. Companion 492a/590 Oxygen Concentrator 2-1 Service Manual 5-15-91 2.1 OVERVIEW OF CONCENTRATOR OPERATION The Companion Oxygen Concentrators use the pressure swing method of concentrating oxygen from room air. In this method, compressed air is alternately applied to two canisters containing molecular sieve material which adsorbs (attracts) nitrogen from the air and allows oxygen and trace gases to pass through. Molecular sieves belong to a class of compounds known as zeolites. Zeolites are highly porous adsorbents both naturally occurring and synthetically produced. Within each granule of man-made zeolite exists a system of precisely arrayed cavities and pores. These are uniform in size and molecular dimension so that molecules are either readily adsorbed or completely excluded. In an oxygen concentrator, nitrogen-selective zeolite (molecular sieve material) is used to adsorb nitrogen molecules in a pressurized tank. It is by physical force (pressure) that this molecular sieve traps nitrogen and yet allows other gases to flow through the sieve bed. Nitrogen molecules are then desorbed (released) by venting the sieve bed to the atmosphere thereby reducing the pressure and the adsorptive force. High oxygen concentration and long sieve bed life is accomplished by the proper use of pressure swing cycles and efficient purging. However, molecular sieve has a strong affinity for water and will displace any other molecule in favor of H20. When this happens, nitrogen-selective zeolite loses its affinity for nitrogen and can be considered to be contaminated. During normal operation though, room humidity does not affect the sieve's ability to adsorb nitrogen. A thin layer of sieve at the top of the bed is used as a desiccant to wick humidity from the supply gas preventing further contamination during a concentrating cycle. This moisture is atomized back into the atmosphere during the exhaust phase. Regeneration of the sieve bed is then completed by purging the exhausting tank with concentrated, humidity free oxygen. The pneumatic system (Figure 2-1), under the control of the electrical system, produces a source of compressed supply air, switches supply air between two molecular sieve canisters, regulates internal pressures, and controls the final output flow rate of oxygen. The components of this system are the compressor and capacitor, heat exchanger, four canisters, three solenoid valves, five pilot valves, two restrictors, pressure switch, regulator, two check valves, flowmeter, expansion chamber, muffler, and four filters. The electrical system (Figure 2-2) provides the controlling and monitoring functions for the concentrator and distributes the electrical power required for operation. In addition to the compressor and solenoids mentioned above, the components of this system are an electronic printed circuit board (PCB), power switch, audio alarm, hour meter, nine-volt battery, circuit breaker, cooling fan, and an Oxygen Concentration Indicator (OCI) system (optional). Following is a description of each of the components that make up the two concentrator systems. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-2 16-51-65 [епиву\ 221025 (uelq #-2) 6-2 JO1EJJUBJUOS USBAXO 065/e26p иошейшо:) wesbeiq onewnaud :1-2 03N614 HOSSI MOS £ — | 1 HIMNITS LAW MIA NA dil Iva 4137 MIL NIZA 137 | и UIONVHOIG LV3H LO | Ч LIN tiv HIALSINYD YICAHISIY ABAFSS Y 3801 9011536 © ЗАЛУА QICN 3906 3ATVYA LO Md INVYIVa (RS EAT BALSINYD 3A3IS E ZAIVA Q ION 39305 3114 z 1эмлозуо T= JA1VA 4IIHI GINO AON ISSY Эми { 397! 50и20а53ы | | (TWNOLLSO) Y3INOSNYEL 770 LG 1SINVO 1INDOHS e HILIMS HILSNVI INSS3td Mais [ IAA IHD NANIYA BOLYINDIY 3UNSSIUd UIONVHI NOSNYJO LE RCE TO) BLK(A) BLK(A) BRN(C) POWER SWITCH BLK(C) вок) BLK(D) BLK(D) BLK(D) SOLENOID 3 SOLENOID 2 SOLENOID 1 BLK(F) RED(E) 7 BLK(F) RED(G) RED(G) RED(G) + 9 VOLT AUDIBLE PRESSURE BATTERY ALARM SWITCH CIRCUIT BREAKER WwHT(B) =) WHT(B) BLK(B) BLKB) = == WHT(A) COOUNG FAN BLK(B) RE o 90004000 » WHT(A) HOURMETER CONTROL PCB BLK(0) Ci == WHT(C) POWER LED | LEGEND E WIRE REFERENCE VOLTAGE A 120 VAC, HOT B 120 VAC, NEUTRAL С 24 VDC, SUPPLY(+) D 24 VDC, RETUAN(-) E 9 VDE, SUPPLY(+) F 9 VDE, RETUAN(-) + (GS 9/12 VDC, RETURN(.) ( H 9/12 VDC, RETUAN(-) * Supply is 12 VDC when concentrator is connected to AC power. In case of power failure, supply reverts to 9 VOC supplied by the battery. Figure 2-2: Electrical Diagram Companion 492a/590 Oxygen Concentrator 2-5 (2-6 blank) Service Manual 5-15-91 2.2 PNEUMATIC SYSTEM COMPONENTS Air Inlet Fil The air inlet filter (Figure 2-3) is a gross particle filter made of a reusable foam material. It filters large particulate material from air drawn into the concentrator cabinet by the action of the cooling fan. It is secured to the right hand side of the cabinet with velcro strips and may be easily cleaned for servicing. (See Table 1-2 Maintenance Schedule). nl filter The compressor inlet prefilter (Figure 2-3 & 2-4) consists of two felt elements that filter particulate matter and smoke from room air drawn into the compressor. Approximately 60 to 70 liters of air are drawn through the filters each minute. The felt filter elements are individually replaceable. (See Table 1-2 Maintenance Schedule). Inl ria Filter The compressor inlet bacteria filter (Figure 2-3 & 2-4) consists of a bacteria grade element in a sealed housing that filters bacteria and particulate matter greater than .2 microns in diameter from room air drawn into the compressor. (See Table 1-2 Maintenance Schedule). Both the inlet prefilter and bacteria filter are located in the filter compartment accessible from the outside of the right hand side of the cabinet. The combination of the inlet prefilter and the bacteria filter enables the compressor to provide clean air to the molecular sieves canisters. nlet Sil The inlet silencer (Figure 2-6) is a hollow chamber molded into the compressor platform. The inlet silencer muffles the sound of the air being pulled into the compressor's inlet. _- COMPRESSOR INLET PREFILTER AIR INLET FILTER COMPRESSOR INLET BACTERIA FILTER Figure 2-3: Side Panel View Companion 492a/590 Oxygen Concentrator 2-7 Service Manual 5-15-91 Г |_— INLET PREFILTERS INLET PREFILTER HOUSING INLET BACTERIA FILTER Figure 2-4: Compressor Inlet Filters 2.2 PNEUMATIC SYSTEM COMPONENTS (continued) Compressor The compressor (Figure 2-5) is a piston motor-compressor with dual cooling fans and an open chassis design. The two fans draw cooling air in from each end of the compressor and exhaust it out the center to provide equal cooling of the bearings as well as the motor. An automatic thermal cut-out switch in the motor interrupts electrical power to the compressor if it overheats. The function of the compressor is to compress filtered air to an average system working pressure of 18 PSI. This compressed air is routed to the molecular sieve canisters which require air under pressure to produce oxygen. Compressor maintenance can be reduced by proper ventilation and filtering. Capacitor The capacitor (Figure 2-6) is an electrical storage device used to help start the compressor motor and improve compressor performance during operation. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-8 HEAT EXCHANGER COMPRESSOR Figure 2-5: Compressor Drawer, Front View INLET SILENCER CAPACITOR Figure 2-6: Compressor Drawer, Bottom View Companion 492a/590 Oxygen Concentrator 2-9 Service Manual 5-15-91 22 PNEUMATIC SYSTEM COMPONENTS (continued) Heat Exchanger The heat exchanger (Figure 2-5) is a loop of aluminum tubing that routes the compressor outlet gas to the reservoir canister. The heat exchanger is located in the path of the compressor cooling air flow and is designed to dissipate heat through the walls of the tube. This component and the four aluminum canisters (product, sieve (2), and reservoir) create a "heat sink” to allow warm compressed gas to cool by dissipation during the concentrator's operation. В ir Canis! The reservoir canister (Figure 2-10) is a hollow aluminum tank, approximately 2.3 liters in volume, that stores air from the compressor under pressure. The reservoir canister acts as a pneumatic pulsation dampener to moderate the pressure pulses generated by the two compressor pistons. The reservoir canister then supplies the large volume of air required to charge the sieve canister. This source of compressed air is also used to pressurize the five pilot valves under the control of the three solenoids. In addition, this canister plays a major role in the heat dissipation process. Pilot Air Fil The pilot air filter (Figure 2-7) is a small sintered-brass element in a sealed plastic housing that filters particulate matter as small as 25 microns in diameter from the pilot line air coming from the reservoir canister. This filtering action prevents contaminants from accumulating and occluding the small passages in the pilot tubing and solenoid valves. Normally, this filter should not need to be serviced. Solenoid Valves Three miniature 24 VDC solenoid valves (Figure 2-8) are used to control the pressurization of the pilot valves. By system design, if one sieve bed is open to supply air, then the opposite sieve bed is open to exhaust. Therefore, if solenoid valve 1 is activated, pilot pressure is applied to supply pilot valve 1 and exhaust pilot vaive 2. If solenoid valve 2 is activated, pilot pressure is applied to supply pilot valve 2 and exhaust pilot valve 1. Finally, if solenoid valve 3 is activated, pilot pressure is applied to the balance pilot valve. Pilot Valves Five air piloted, diaphragm-poppet valves (Figure 2-8) are used to route gas flow into and out of the two sieve canisters. When pilot pressure is applied to the valve's pilot port, the poppet assembly (a brass stem with two rubber diaphragms at each end) is forced down, allowing gas to flow through the valve. When pilot pressure is removed, pressure against the unbalanced area of the bottom diaphragm moves the poppet assembly against its seat to stop flow through.the valve (Figure 2-9). A spring, used only in the two exhaust pilot valves, keeps the valve closed when the concentrator is turned off to prevent contamination of the molecular sieve material. When a supply pilot valve is actuated, compressed air from the reservoir canister is admitted to a sieve canister to begin the oxygen concentration process. Simultaneously, the corresponding exhaust pilot valve is actuated and adsorbed nitrogen from a sieve canister is exhausted through the valve and into the expansion chamber and muffler and to atmosphere. When the balance pilot valve is activated, the sieve canisters are pneumatically connected to allow the oxygen layer at the bottom of one sieve bed to flow to the opposite bed. This pre-charge of oxygen gas increases the pressure in a bed before it begins its concentrating cycle for greater efficiency in the nitrogen adsorption process. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-10 PILOT AIR FILTER Figure 2-7: Pilot Air Filter EXHAUST PILOT VALVE 1 EXHAUST PILOT VALVE 2 SOLENOID 1 SOLENOID 2 SUPPLY PILOT VALVE 2 SUPPLY PILOT VALVE 1 SOLENOID 3 1 | BALANCE PILOT yd IM AR Figure 2-8: Solenoid and Pilot Valves Companion 492a/590 Oxygen Concentrator 2-11 Service Manual 5-15-91 UNACTUATED ACTUATED E Pilot Pressure (77) sieve Gas * Spring used in Exhaust Pilot Valves only. Figure 2-9: Pilot Valve Operation 2.2 PNEUMATIC SYSTEM COMPONENTS (continued) ¡ev nister The two sieve canisters (Figure 2-10) are aluminum tanks that contain granular molecular sieve material, a nitrogen selective zeolite that strips nitrogen from air by an adsorption process while allowing oxygen and trace gases to pass through the sieve. Nitrogen adsorption occurs when the sieve canister is pressurized from approximately 14 to 20 PSI with compressed room air. A sieve canister that is saturated with adsorbed nitrogen is regenerated (desorbed) by venting the sieve canister pressure to atmospheric pressure. Some of the product oxygen is supplied as a purge gas to increase the efficiency of regenerating the sieve by removing any residual nitrogen molecules remaining in the canister. This oxygen readily combines with the nitrogen and is flushed from the canister by the purge pressure. The sieve canisters must be kept sealed from the atmosphere to prevent contamination of the molecular sieve material by migration of moisture (humidity) into the canister. Because of sieve material's high affinity for water, moisture drawn into the sieve canister without proper purging action will render the sieve useless for future nitrogen adsorption. Expansion Chamber and Muffler The expansion chamber/muffler assembly (Figure 2-11) consists of a paper filter element and a plastic foam-lined chamber that is connected to the exhaust tube coming from the exhaust pilot valves. The assembly effectively muffles the sound of gas exhausting to atmosphere when the sieve canisters are depressurized (desorbed) during the regeneration (exhaust) cycle. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-12 PRODUCT CANISTER SIEVE CANISTER 1 RESERVOIR CANISTER SIEVE CANISTER 2 Figure 2-10: Sieve Canisters EXPANSION ——————— CHAMBER Barri 4 Dh) Ar MUFFLER Figure 2-11: Expansion Chamber & Muffler Companion 492a/590 Oxygen Concentrator 2-13 Service Manual 5-15-91 2.2 PNEUMATIC SYSTEM COMPONENTS (continued) Restrictor Tube Assemblies Two restrictor tubes (Figure 2-12) are located just above the balance pilot valve between the tubing coming from the outlets of the sieve canisters and the product tank. Each restrictor tube assembly contains an orifice drilled to a specific size and a number recorded on the tube for size identification. This number represents the liter flow of a gas through the orifice at a standard pressure. While supply air is entering one sieve bed, the restrictor creates back pressure in this canister and thus determines the rate at which room air can flow through the sieve bed. Meanwhile, the opposite restrictor acts as a metering device allowing a small amount of product gas (oxygen) to pass through this orifice and into the bottom of the exhausting sieve bed for purge. Thus, increasing or decreasing the orifice size will alter the balance of pressure within the sieve beds. This regulation of pressure allows a filling canister to pressurize properly for maximum nitrogen adsorption. The product tank pressure is maintained by the "checking" action of the restrictor tube of the exhausting tank while still permitting a sufficient amount of product oxygen to pass through this orifice and into the exhausting canister for efficient purge of any residual nitrogen gas. Product Canister The product canister (Figure 2-10) is a hollow aluminum tank, approximately one liter volume, that stores product gas (oxygen) from the sieve canisters under pressure. The product canister functions as a reservoir of oxygen to ensure a smooth, steady outlet flow as the unit cycles back and forth between the sieve canisters. It is also responsible for providing sufficient pressure to permit some product oxygen to be used as purge gas through the restrictors. The amount of purge oxygen then is directly dependent upon the product tank pressure. In turn the product tank pressure is affected by both the size of the restrictors and the flowmeter setting. Pressure Switch The electro-pneumatic, normally closed pressure switch (Figure 2-13) is mounted on the product canister and senses the pressure of the product gas (oxygen) in the product canister. Normal product canister pressure keeps the switch contacts open. If the product canister pressure drops below the pressure switch threshold setting of 4.0 PSI, the switch contacts close and the audio alarm will sound. Pressure Regulator The adjustable pressure regulator (Figure 2-13) reduces a nominal oxygen pressure in the product canister to a constant 5 PSI outlet pressure. The concentrator flowmeter requires a constant 5 PSI source pressure to accurately control the flow of oxygen through the flowmeter and to the patient. Vacuum Check Valve The vacuum check valve (Figure 2-13), located on the output side of the regulator, functions as a siphon break when the concentrator cools down after being shut off. The valve prevents a vacuum, created when gas in the product canister cools after the concentrator is shut off, from drawing water in a humidifier bottle into the flowmeter. When the concentrator is operating, the checking action of the valve prevents output flow from bleeding to atmosphere. -Line Ch lv The in-line check valve (Figure 2-14) is installed in the outlet gas tubing between the pressure regulator and outlet gas bacteria filter. The purpose of the check valve is to ensure room air does not migrate back into the sieve canisters through the gas outlet circuit when the concentrator is turned OFF. The check valve is installed in the outlet gas tubing with the black connector toward the outlet of the concentrator. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-14 Figure 2-12: Restrictor Tubes Figure 2-13: Product Canister, Regulator and Vacuum Check Valve Companion 492a/590 Oxygen Concentrator 2-15 Service Manual 5-15-91 22 PNEUMATIC SYSTEM COMPONENTS (continued) Qutlet Gas Filter The outlet gas filter (Figure 2-14) consists of a bacteria-grade element in a sealed housing that filters particulate matter greater than 0.3 microns in diameter from the oxygen flowing to the flowmeter and eventually to the patient. This filter provides clean, bacteria-free oxygen to the patient. (See Table 1-2 Maintenance Schedule). Elowmeter The flowmeter (Figure 2-15) is a back-pressure compensated gas measuring device which controls the amount of product oxygen dispensed to the patient. It consists of a tapered hollow tube inscribed with flow markings, a needle valve, an adjusting knob, an indicator ball, and an inlet and outlet port. With the concentrator operating, the ball will move up in the tube indicating the amount of oxygen delivered to the patient. 2.3 ELECTRICAL SYSTEM COMPONENTS Power Switch The power switch (Figure 2-15) is a double-pole single-throw rocker type with a bezel mounted green L.E.D. indicator. When the switch is in the ON position, one set of contacts routes 24.6 VDC to the coil of a relay mounted on the control printed circuit board. This relay activates the compressor, cooling fan, hour meter, and printed circuit board timing circuits. The second set of power switch contacts routes DC current from the circuit board to the pressure switch and audio alarm. This DC current is provided at 12 volts while the concentrator is running and drops to 9 volts, provided by the nine-volt battery, in case of a power failure. (NOTE: If the power switch is set to ON, the audio alarm will sound when the pressure switch contacts close, regardless of whether the concentrator is plugged in or not). Concentrators equipped with the optional Oxygen Concentration Indicator (OCI) incorporate a push button type power switch. Refer to section 6 for details. С | Printed Circuit Board The control printed circuit board (PCB) (Figure 2-14) contains the electronic logic for controlling the operation of the concentrator. The PCB contains a step-down transformer and a full-wave bridge rectifier that converts AC voltage to 24 VDC for PCB use and for powering the solenoids. A 63 milli-ampere 250 VAC fuse protects the PCB components in the event of an over-current condition. The circuitry includes a timing generator and a detector circuit. The timing generator produces pulses that are used by the alternate/dwell circuit and the pre-charge time circuit. The alternate/dwell circuit uses the timing pulses to activate solenoid valve 1 and solenoid valve 2 alternately for approximately eight seconds each. A dwell period, approximately 0.8 seconds occurring between the activation of the solenoids valves 1 and 2 enables the balance time circuit to activate solenoid valve 3 for the balance cycle. The detector circuit activates a relay if an open condition is detected in a solenoid’s electrical circuit caused, for example, by a disconnected wire or an open solenoid coil. The relay, when activated, removes power from the compressor, cooling fan, and solenoids. The detector circuit thus prevents the concentrator from operating with an electrical failure of the solenoid circuit that could contaminate the molecular sieve material and thereby destroy its nitrogen-adsorbing properties. Since complementary metal-oxide semiconductor (CMOS) integrated circuits are present on the PCB, care must be used when handling the board to prevent static discharge from possibly damaging board components. Always handle PCBs by their edges only and store them in a static-proof bag. Companion 4392a/590 Oxygen Concentrator Service Manual 5-15-91 2-16 CONTROL PRINTED CIRCUIT BOARD IN-LINE CHECK VALVE FLOWMETER —— OUTLET GAS FILTER _ E | Figure 2-14: Right Cabinet, Interior View eo — 8 SWITCH Figure 2-15: Control Panel Companion 492a/590 Oxygen Concentrator 2-17 Service Manual 5-15-91 2.3 ELECTRICAL SYSTEM COMPONENTS (continued) Cooling Fan Cooling-air flow through the concentrator (Figure 2-16) is provided by a fan (Figure 2-17) located beneath the compressor. The fan draws in cooling air through the air inlet filter located on the upper right hand side of the cabinet. The cooling air travels down across the concentrator's four canisters and the compressor and is exhausted out the bottom left side of the cabinet. Both the intake air filter and exhaust vents of the cabinet must be free of obstructions for the cooling system to function properly. ia (439 Ss | — D RY J К J'@ (Gé Figure 2-16: Cooling Air Flow Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-18 COOLING FAN Figure 2-17: Cooling Fan 2.3 ELECTRICAL SYSTEM COMPONENTS (continued) Hour Meter The digital hour meter (Figure 2-18) located in the filter compartment, indicates total concentrator running time. The meter records up to 99,999.99 hours before resetting to zero. ircui ker The AC electrical system is protected from short circuits and power surges by a five-ampere (for Companion 492a) or a six-ampere (for Companion 590) resettable push-button circuit breaker (Figure 2-18) located in the filter compartment. Audio Alarm The audio alarm (Figure 2-18) consists of an electronic module powered by either the control printed circuit board or the nine-volt battery. When the POWER switch is in the ON position, and the pressure switch contacts are closed, the alarm sounds. If the concentrator is operating on AC, the audio alarm is powered by 12 VDC from the circuit board. In the event of an AC power failure, the audible alarm will be powered by the nine-volt battery. Battery A nine-volt long-life alkaline battery (Figure 2-18) located in the side panel filter compartment, provides back-up power to the audio alarm. See Table 1-2. Maintenance Schedule. X n Concentration Indicator (See Section 6) Companion 492a/590 Oxygen Concentrator 2-19 Service Manual 5-15-91 BATTERY — CIRCUIT BREAKER AUDIO ALARM HOURMETER Figure 2-18: Filter Compartment 2.4 CONCENTRATOR SYSTEM OPERATION The Companion 492a/590 concentrates oxygen from room air by a process known as pressure-swing adsorption. This cyclic process involves two molecular sieve canisters which are alternately pressurized to approximately 20 PSI with room air and then vented to atmospheric pressure. The nitrogen-selective zeolite in the molecular sieve canisters adsorbs nitrogen from air under pressure and, in turn, desorbs (releases) nitrogen when the canister is vented to atmospheric pressure. The process is further enhanced by the incorporation of a pressure-balance cycle and by the use of product gas (oxygen) as purge gas to increase the efficiency of regenerating the sieve. A complete system cycle of approximately 17.6 seconds in duration consists of two 8-second concentrating cycles and two 0.8-second balance cycles. Gas flow through the Companion 492a/590 pneumatic system is shown in Figures 2-19 through 2-21. | in Discussion of the complete system cycle may begin with sieve canister 1 concentrating oxygen (Figure 2-19). The compressor pressurizes the reservoir canister, the supply air tubing, and the pilot air tubing with filtered room air. At the appropriate time, the electronic timing circuit on the PCB sends a 24 VDC signal to energize solenoid valve 1 for approximately eight seconds. Then, pilot gas under pressure flows through the solenoid and is applied to the pilot ports of supply pilot valve 1 and exhaust pilot valve 2, simultaneously opening each pilot valve. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 2-20 2.4 CONCENTRATOR SYSTEM OPERATION (continued) Supply air from the reservoir canister passes through open supply pilot valve 1 and into sieve canister 1 where first, water vapor is attracted and held in the top of the canister, then nitrogen is adsorbed and oxygen-rich gas flows to the product canister metered by restrictor 1. Next, some oxygen-rich gas passes through restrictor 2 and into sieve canister 2 once the pressure in this canister has exhausted to atmosphere through exhaust pilot valve 2. This purge oxygen migrates up into sieve canister 2 to increase the efficiency of regenerating the molecular sieve by removing any trace nitrogen in the canister. Oxygen in the product canister, at a pressure higher than that required at the flowmeter, is reduced and maintained at 5 PSI by the pressure regulator. Oxygen then passes through the outlet gas filter, the flowmeter, and finally the oxygen outlet. Balance Cycle 1 After sieve canister 1 has been concentrating for approximately eight seconds, the electronic timing circuit on the PCB simultaneously de-energizes solenoid valve 1 and energizes solenoid valve 3 for approximately 0.8 seconds (Figure 2-20). When solenoid valve 1 is de-energized, the pilot pressure source for supply pilot valve 1 and exhaust pilot valve 2 is cut off and the piloting pressure for the valves is vented to atmosphere through the normally open port of solenoid valve 1. Pressure against the unbalanced area of the bottom diaphragm closes supply pilot vaive 1. An internal spring closes exhaust pilot valve 2. Now solenoid valve 3 is energized and pilot air pressure is applied to the pilot port of the balance pilot valve to open the valve. With the balance pilot valve open and the remaining pilot valves closed, oxygen-rich gas at system pressure from sieve canister 1 is routed directly to sieve canister 2 to increase pressure in this canister in preparation for its concentrating cycle. The placement of restrictor 1 and restrictor 2 downstream of the balance pilot valve ports allows gas from sieve canister 1 under pressure to flow into and pressurize sieve canister 2 during the balance cycle with minimal effect on gas in the product canister. This increase of pressure in sieve canister 2 during the balance cycle helps ensure immediate oxygen flow from this canister at the beginning of its concentrating cycle. Sieve Canister 2 C ina Cyl At the end of balance cycle 1, the electronic timing circuit on the PCB simultaneously de-energizes solenoid valve 3 and energizes solenoid valve 2 for approximately eight seconds (Figure 2-21). When solenoid valve 3 is de-energized, the pilot pressure source for the balance pilot valve is cut off, and the piloting pressure for this valve is vented to atmosphere through the normally open port of solenoid valve 3. Pressure against the unbalanced area of the bottom diaphragm closes the balance pilot valve. When solenoid valve 2 is energized, pilot air pressure is applied to the pilot ports of supply pilot valve 2 and exhaust pilot valve 1 and simultaneously opens each pilot valve. Supply air passes through open supply pilot valve 2 and into sieve canister 2 where water vapor is trapped and nitrogen is adsorbed while oxygen-rich gas flows to the product canister through restrictor 2. Meanwhile, gas is allowed to exhaust out of sieve canister 1 through open exhaust pilot valve 1 and to atmosphere. This exhaust gas drives any water vapor from the top of the sieve bed back into the room. (Maintaining this desiccant layer through proper valving action is critical to long sieve bed life). Then, some oxygen-rich gas also passes through restrictor 1 and into sieve canister 1 once the pressure in sieve canister 1 has exhausted to the atmosphere through exhaust pilot valve 1. This purge oxygen migrates up into sieve canister 1 to increase the efficiency of regenerating the molecular sieve. This cycle duplicates the process of the first concentrating cycle and ensures constant flow of oxygen to the product canister. Balan le 2 After sieve canister 2 has been concentrating for approximately eight seconds, the electronic timing circuit on the PCB simultaneously de-energizes solenoid valve 2 and energizes solenoid valve 3 for approximately 0.8 seconds (Figure 2-20). This cycle then allows the pressure between the two sieve canisters to pre-charge sieve canister 1 in preparation for its concentrating cycle. Upon completion of balance cycle 2, the complete system cycle begins again. Companion 482a/590 Oxygen Concentrator 2-21 Service Manual 5-15-91 £0 à ec. RES E 1 A Ба LL lt = us, 5 i EXHAUST EXHAUST Ч г PROT PILOT 3 VALVE VALVE "e x 2 1 E О E р ЧЕ РЕ, Él q x ( se a Ey О ro Е PE], * Er ry LEY (CE ( A + NE ( + | RN h A Р se | НТ REG TOR " SIEVE NN ee hz CANISTER 3 > SIEVE > CANISTER 1 VACUUM / 5 CHECK VALVE Da cr RINA 7” N y A AA С 3 [Bo mt or € >, + WATER A VAPOR - N 5“ EN? SRE a EXHAUSTED es W o PRESSURE M ADSORBED e SWITCH la e E o ны 3 PRODUCT PLAGE N 2 € CANISTER 02 ha a 8 + О? PASSES ; не + THA 3 ri » £ Wi о | o FFF FF FFF FF FF FF FF FF FFF LF FF FF 553 3° " > x : I by [=] © = © 060 ° оо ° ос ° ос ° со ° ос ° ос ° оо ° со © oo ° оо © oo © oo © © ох + ] = : - | N Я LLINARES EAT TEA EN MIA ENS + | J N - FILTER LEGEND [2 0) SUPALY AR ЛЕЙ PRODUCT OXYGEN +++ PILOT AM | FUNCE OXYGEN [==] sieve exaust FEE LEE E aaa Ea Far RESTRIC TOR TURE ASSEMBLY 2 Tx "+ iT = fast а ad ret "a Va и Fa dal {1 > RESTRICTOR TUBE ASSEMBLY 1 a” L PAL "a = 4 VALVE 3 + Pompe na ia e ..... F Fc ff Fe as - FFF rT TT rN TN YT TN TT TIT NRF TR NANT TT TOT TONY RN NRT OY "- се BALANCE PILOT VALVE LE ELE RESEAYOIA CANISTER AIR INLET FILTER HEAT EXCHANGER N ( COMPRESSOR Figure 2-19: Sieve Canister 1, Concentrating Cycle Companion 492a/590 Oxygen Concentrator 2-23 (2-24 blank) Service Manual 5-15-91 a ET Efe ЛАМ ЗС A A a CXYGEN OUTLET PRESSURE REGULATOR VACUUM CHECK VALVE PRESSURE PRODUCT CANISTER EXHAUST EXHAUST 3 ALOT ALOT 5 VALVE VALVE : 2 1 ПТР р: == E: 10 SUPPLY | SOLENOIO LOT | VALVE2 VALUE 1 AL | ( ree: h NT res LEGEND | = 5) SUPPLY AIR PILOT AIR [1 SIEVE EXHAUST EXT] PRODUCT OXYGEN [+.-..] PRECHARGE OXYGEN PILOT EXHAUST PRESSURE AIR INLET FFE EEE: EEE pe lee ete era FA ; RIA т 1 E E Зы i] oc. = of TRANSDUCER 39 as: (OPTIONAL) Ts у: CTE INTA 1 e: > bel Ea CECI a Sear ааа CAINE : ei se IN-UNE GAS OUTLET 4 CHECK VALVE FILTER x + + ‘3 x ei +; 53 Li уе т AESTRICTOR TUBE RESTA TOA TUBE ASSEMEL Y 7, ASSEMEL Y 2 e 1 E e =; 2 = + = ve ri : re Pa Но = = a Ei Ч “dy 2-25 (2 Figure 2-20: Balance Cycle Companion 492a/590 Oxygen Concentrator 26 blank) Service Manual 5-15-91 EXPANSION CHAMBER PRESSURE REGULATOR SIEVE CANISTER ИТ 2 —- VACUUM CHECK VALVE E Vina : THAPPED arn as E а Tw e: UT, Ta E > Ra tar Ceux rl = ed Lo INLINE CY GEN CHECK VALVE GAS OUTLET FILTER tes a E Cda e NAO ONE = re AAT] Ome e EII ет de + = Loo imi a a >} Ea TTS oy = - ео = — a SAA RUE DR Aan A A У fo vm TERE hy a йо IER, ET = RESTACTORA Tuge ASSEMEL Y a = Ко * x aa ETE El > er a Art + © “Ч. Te BALANCE PILOT VALVE LEGEND 4 Be "| supply AR GEES mooucr OXYGEN +++) PLOTAR pe + Я PURGE OXYGEN ŒUE] seveexuaust MLOT EXHAUST Ei N 4D = a Rs a o ©. Ее gr TE, © o о * > RESERVO o CANISTER o, o o AIR INLET o, FILTER ве od o - зо об INLET o] = PREFILTER o o | | o COMPRESSOR Figure 2-21: Sieve Canister 2, Concentrating Cycle Companion 492a/590 Oxygen Concentrator 2-27 (2-28 blank) Service Manual 5-15-91 2.5 SAFETY FEATURES The Companion 492a/590 has an audio alarm, a solenoid valve shut-down feature, and an optional Oxygen Concentrator Indicator to alert the patient and protect the concentrator in the event of improper operation. The audio alarm will sound due to inadequate system pressure caused by AC power loss, dity compressor inlet filters, or leaks/restrictions in the internal pneumatics. When inadequate system pressure is sensed, an electro-pneumatic pressure switch, mounted on the product canister, activates the aduio alarm. И system pressure drops below 4.0 PSI, the pressure switch contacts close to activate the audio alarm. The solenoid shut-down feature consists of a detector circuit on the PCB that monitors the solenoid's electrical circuits. If an open condition is detected, caused by a disconnected wire or an open solenoid coil, the detector circuit activates a relay which removes power from the compressor, cooling fan, and the solenoids. This feature protects the molecular sieve material from contamination in the event of a solenoid electrical problem. The relay is reset by first correcting the problem and then setting the POWER switch OFF, wait five seconds - then back ON. NOTE Residual pressure in the reservoir canister may prevent the compressor from starting for up to 20 seconds. For units equipped with the optional Oxygen Concentration Indicator (OCI) refer to section 6 for details regarding additional safety features. Companion 492a/590 Oxygen Concentrator 2-29 Service Manual 5-15-91 SECTION 3. PERFORMANCE VERIFICATION AND TROUBLESHOOTING It may be necessary to service or verify performance of either the Companion 492a or 590 Oxygen Concentrator. This section is designed to assist the user in identifying a problem or in checking out the concentrator before application. There are three conditions where this section may be helpful: a) To determine the reason for an operational failure. b) To check overall system operation after the repair or replacement of a component (or system). c) To verify the concentrator is operating within specification. For condition a, refer to subsection 3.2; Operational Checklist for help in troubleshooting and problem analysis. Forconditions b and ¢, subsection 3.1; Performance Verifications will provide checkout procedures for individual systems and key components. Also, refer to any checkout procedures found in Section 4; Service And Repair and Section 1; General Information, pages 1-7 through 1-13. Companion 492a/590 Oxygen Concentrators 3-1 Service Manual 5-15-91 3.1 Performance Verification This subsection is divided into six performance areas: 1) oxygen concentration, 2) pneumatic, 3) electrical, 4) cooling, 5) flow, and 6) sound. Although each may be considered a separate system or function for testing, they all must perform in harmony to achieve the concentrator's end result - high oxygen purity delivered at a continuous pre-determined flow rate. Should the concentrator fail to pass any performance tests, refer to subsection 3.2; Operational Checklist, subsection 3.3; Troubleshooting and/or Section 4; Service and Repair, as needed. WARNING The concentrator is electrically powered. To prevent serious injury or death, observe standard safety procedures when testing the concentrator. The compressor and other internal components are not grounded. To prevent serious electrical shock, do not touch metal components when operating the concentrator with cabinet enclosure opened. Before testing, ensure the following : a) Airinlet filter, inlet prefilters, inlet bacteria filter, and 9 volt battery are installed (Figure 2-3). b) Power cord and plug is not damaged or frayed c) Humidifier bottle and any other delivery apparatus is removed from concentrator outlet. 3.1.1 Oxygen Concentration Oxygen concentration levels may be checked whenever it is necessary to know that the concentrator is producing correct oxygen purity. Due to system design, this area of performance verification may be considered to be the most crucial. Generally, if the concentrator is producing oxygen at or above the minimum specification at the maximum flow rate, there should be a high confidence level that the other systems are operating properly. (If your concentrator is equipped with the optional OCI, refer to Section 6 for all information related to this system). A) Oxygen Test (Figure 3-1) 1) 2) 3) With the concentrator plugged in, turn power switch to ON. Set flowmeter to maximum flowrate (4 LPM or 5 LPM) and allow unit to run for 20 minutes in an open environment. The flowmeter ball should be occasionally checked and adjusted to ensure that it is bisected by the line corresponding to the selected flowrate. Using a properly calibrated oxygen analyzer, follow the manufacturer's recommendations for how to sample the oxygen outflow of an oxygen concentrator. (Generally, it is necessary to allow the oxygen to flow across the sampling device of the analyzer. Do not allow the oxygen to back pressure against the sampling device). Record the oxygen concentration and compare to the performance specifications in Table 1-1 page 1-5. Should the concentration be below anticipated levels, check the following: a) Air inlet, inlet prefilters, and inlet bacteria filters are clean. b) Flowmeter setting is correct (see subsection 3.1.5 Flow). c) Analyzer accuracy - not calibrated, battery weak, or inadequate oxygen to sampling device. If still out of specification, refer to subsection 3.2.1 Low Concentration. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-2 3.1.2 Pneumatic Figure 3-1: Oxygen Concentration Testing Because oxygen concentrators are principally pneumatic devices, verifying the pressure specifications in the Companion concentrator is valuable information in analyzing performance. To do this, the Companion Pneumatic Test Kit P/N 492381 should be used for all tests in this sub-section. A) Leak Test If a concentrator has a leak in a system, it may cause loss of oxygen concentration or possibly sieve contamination. It is important to leak test the unit before proceeding to pressure tests. 1 2. Open the cabinet of the Concentrator. (See service procedure step 1, Section 4). Inspect all hoses, connections, and fittings in the pneumatic system. Check for tightness, Kinks, wear, or damage to parts. Correct any deficiencies. Plug concentrator's power cord into AC outlet, turn power switch ON and allow unit to run for several minutes. Use a leak detector to carefully spray all connections in the concentrator and observe them for a bubbling action indicating a leak. Start with the compressor and work through all of the canisters and valves to the outlet. (The outlet side of each exhaust pilot valve does not need to be leak tight). If a leak is detected, correct by tightening, resealing with pipe cement, or repairing the connection. WARNING Use caution with leak detector. Do not allow to contact electrical connectors or com- ponents. Electrical shock or damage to the concentrator may occur. Companion 492a/590 Oxygen Concentrators 3-3 Service Manual 5-15-91 3.1.2 Pneumatic (continued) B) Pressure Tests There are three pressure tests to perform on the Companion concentrator. Do the following steps to check out the reservoir canister pressure (which indicates sieve canister filling pressure), product canister pressure and outlet gas pressure. Because the pressure in a concentrator can vary with altitude, barometric changes and compressor performance, a range of acceptable pressures are shown for the following tests. With experience you should be able to identify a normal range for your testing location and concentrators. 1. 2. Turn concentrator off after completing the leak test described in step A. Locate 1/4" ID tubing connected betweenthe brass tee, above the restrictor tube assemblies, and the product tank. Remove clamp and product tank tubing from tee fitting. Tee in 1/4" ID tubing from pneumatic test kit (Figure 3-2). Locate and disconnect the small pilot air tubing leading to the brass nipple on the balance solenoid valve. Tee in the 1/16" ID tubing from the pneumatic test kit (Figure 3-3). Set the selector switch on test kit to RSVR (to the left), and turn the concentrator's power switch ON. Set the flowmeter to the maximum LPM indicated on its scale and allow the unit to run for several minutes. (NOTE: If needle on gauge reaches 30 PSI, turn unit OFF and see subsection 3.3; Troubleshooting Guide. Verify pressure in reservoir canister slowly increases, from 14 + 1.5 PSI (drop point) to 20 + 2.0 PSI (fill) and then rapidly climbs to 27 + 2.0 PSI (kick). After the kick, the pressure should fall back to the drop point and repeat. Record all three pressures on two consecutive cycles. Record only the drop point pressures to the tenth of a PSI, (i.e. 14.2). Verify the balance (the difference of pressure) at the drop point on two consecutive cycles is not greater than 0.4 PSI. (i.e. 14.2 and 14.4 PSIG). Refer to subsection 3.4; Balancing Orifices, if difference is greater than .4 PSIG. Set selector switch on test kit to PROD (to the right). Verify the product canister pressure is between 6.5 to 11.0 PSI (Figure 3-2). Although there is not a "balance" of pressure in the product tank, the pressure on two consecutive cycles should not vary by more than 4.0 PSI. NOTE If either of the above tests yield unsatisfactory results, perform Balancing orifices procedures in subsection 3.4. 10. 11. 12. With thumb and forefinger, crimp the short 1/4" ID tube above the brass restrictor tube tee. (Figure 3-4) to restrict the flow of gas to the test kit gauge and product canister. Gradually reduce gauge pressure until the pressure switch in the product canister activates the audio alarm. Verify the alarm sounds at 4.0 PSI + 0.2 PSI. Refer to Section 4; Service And Repair, step 14 if set incorrectly. Set the concentrator's power switch to OFF and disconnect test kit tubing. Reconnect 1/4" ID tubing from product tank to restrictor tube assembly brass tee and secure with new clamp. Reconnect 1/16" 1D tubing to brass nipple at base of balance solenoid valve. Connect tailpiece (P/N 492587) and wingnut (P/N 492588) outlet adapters to outlet of concentrator. Connect 1/4" ID tubing of test kit to adapter and verify the test gauge selector switch is set to PROD (right). Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-4 ~ 3.1.2 Pneumatic (continued) 13. Set the concentrator’s power switch to ON, turn flowmeter to 1 lpm, and let run for several minutes. 14. Place thumb over open end of large tee from test kit (Figure 3-5). 15. Verify concentrator outlet static pressure is 5 + 0.5 PSI. Refer to Section 4, step 15, if out of tolerance. 16. Disconnect the 1/4" ID tubing and outlet adapters. [a ii LL LL LEA E a so ste r a р | nn bof " : 3 E | E E 3 = - Е Ko Me |, N Г | M Let an y 1 = | Е i | ff | M vu CC BR / | i 4 | r y AUX [В y I sa й 1 i |, | “e J 1 | | | | br | i || i fl | VE fi = | 1 |" IE — =! 1 = L SN Figure 3-2: Pneumatic Test Kit Connection/Testing Companion 492a/590 Oxygen Concentrators 3-5 Service Manual 5-15-91 Figure 3-3: Pneumatic Test Kit Connection/Testing Figure 3-4: Pressure Switch Testing Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-6 paria rate emi ainia — hemo осы Figure 3-5: Outlet Pressure Testing 3.1.3 Electrical The electrical systems provide the controlling and monitoring functions for the oxygen concentrator. The ” following tests and observations should be made to verify performance of these components. A) B) Physical Observation With the unit unplugged, open the Companion cabinet and inspect all wires, connectors and terminals for tightness, wear or fraying. Inspect power cord and plug. Repair or replace as necessary. Low Pressure Alarm 1. With the unit unplugged, turn flowmeter control knob counterclockwise to drain any residual pressure from product canister. Remove 9V battery from its holder in the filter compartment. Use a multimeter tester to check battery voltage. Replace if less than 8.0 volts. Connect concentrator's power cord to an appropriate AC outlet and turn power switch to ON. Verify audio alarm activates. Turn the concentrator's power switch OFF. Install 9V battery into holder. Disconnect power cord from AC power source. Turn the concentrator's power switch to ON. /. Verify audio alarm activates. Turn the concentrator's power switch OFF. NOTE When the Companion concentrator is connected to an AC power source, the audio alarm is powered from a 12V signal from the Control PCB. The 9V battery powers the audio alarm only during a loss of power or Control PCB failure. Companion 492a/590 Oxygen Concentrators 3-7 Service Manual 5-15-91 3.1.3 Electrical (continued) C) Control Printed Circuit Board 1. 2. м о с ео Open cabinet Turn flowmeter control knob counterclockwise approximately one revolution and drain any residual pressure from the product canister. With concentrator plugged in, turn the power switch ON. Verify audio alarm sounds until pressure builds in the product canister. Verify that the front panel POWER indicator is illuminated. Verify the cooling fan and the compressor operate. Verify the solenoids cycle by listening for a slight puff of air followed by the exhaust of a sieve canister approximately every eight (8) seconds. 3.1.4 Ventilation (Cooling) System 3.1.5 Flow Remove, inspect, and clean if necessary the air inlet filter. Reinstall the filter. Open cabinet. Inspect air vents at bottom of left half of Companion cabinet. Clean and remove any blockage at vents. Connect concentrator's power cord to AC outlet and set the power switch to ON. Observe the cooling fan and verify its operation. Close cabinet and verify there is a strong push of air coming from the vents at the bottom left of concentrator. Connect concentrator power cord to AC and set power switch to ON. Allow unit to run a minimum of 5 minutes. Verify outlet static pressure is 5 + 0.5 PSI. (See subsection 3.1.2 B, Steps 12 through 15). Connect the test flowmeter (Sierra Top-Trak model 820 mass flowmeter) to the concentrator as shown in Figure 3-6. With the line bisecting the flowmeter ball, verify the flow is + 5% of full scale at each setting. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-8 3.1.6 Sound Figure 3-6: Flowmeter Testing Occasionally, it may be necessary to determine if a concentrator is operating above the minimum acceptable decibel levels. To achieve an accurate measurement requires duplicating the same environmental conditions under which the factory tests are conducted. This may prove to be expensive and difficult. As an alternative, there are comparison tests and observations that may resolve concerns about the noise level of a Companion concentrator. If an unusually loud noise is heard, generally it is detectable by attempting to locate the source of the sound. Open the concentrator cabinet and inspect for the following: 1. a son Inlet bacteria filter properly seated in filter compartment. Compressor inlet hoses properly connected. Shock mounts for tears allowing compressor to drop. Loose compressor bolts. Heat exchanger rubbing. Internal tubing disconnected. Check especially exhaust hose at exhaust pilot valves and expansion chamber. Noisy compressor or cooling fan. Disconnect one terminal from cooling fan during operation to segregate. Reconnect when done. NOTE To resolve problem, adjust, repair, or replace as needed. Referto Section 4 for appropriate component. Companion 492a/590 Oxygen Concentrators 3-9 Service Manual 5-15-91 3.1.6 Sound (continued) If a Companion concentrator is reported "noisy", it may be helpful to compare it to several other Companion units. The sound level (as measured in decibels) is a weighted average of many frequencies. Some users may be more sensitive to certain frequency ranges and, therefore, perceive more sound. Comparison of one unit to another may be valuable in determining excessive noise reported on a concentrator. 3.2 Operational Checklist If a Companion concentrator requires servicing, this section will be helpful in determining the source of the problem. Used in conjunction with the Troubleshooting Guide, (subsection 3.3), this checklist will provide the technician with a logical approach to discovering any equipment malfunctions. There are three categories of concentrator failures 1) Low concentration (subsection 3.2.1), 2) Unit won't run (subsection 3.2.2), or 3) Unit operates improperly (subsection 3.2.3). Refer to the category that best suits your concentrator's problem and follow the checklist to reveal the symptoms that describe the problem. Next, refer to the Troubleshooting Guide by matching your unit's symptoms to those listed to suggest a component or system most suspect. Follow the suggested corrective action to test and/or resolve the problem. (Refer to section 6 for Troubleshooting the OCI system). 3.2.1 Low Concentration 1. Verify oxygen analyzer is calibrated and working properly. Check manufacturer's operating instructions for proper use and maintenance. 2. Measure concentration at maximum setting on flowmeter scale to determine if unit is producing the minimum specified level of oxygen after 20 minutes. (Refer to Table 1-1). 3. Inspect and clean all filters. Open cabinet. Visually inspect all hoses, tubes, and fittings for looseness or kinks. With concentrator running, use a leak detector to spray all pneumatic fittings and observe for a bubbling action which indicates a leak. All fittings with the exception of those coming from the pilot exhaust valves through to the muffler must be air tight. Repair by tightening or removing and resealing the fitting with an approved pipe cement. (Refer to subsection 3.1.2 A). NOTE It is critical that all leaks be eliminated from the concentrator before proceeding. 6. Connect the Pneumatic Test Kit to the concentrator (Refer to subsection 3.1.2 B; Pressure tests). Turn the unit on and allow to cycle several minutes. /. With the selector switch set to RSVR observe the pressures in the reservoir canister. Read subsection 3.1.2 B, steps 5 through 7 and compare results with the expected pressures. 8. Ifthe readings are not normal, refer to the Troubleshooting Guide. (i.e. low, high, imbalance or erratic pressure). Refer to symptom with appropriate description for the observed behavior. 9. Turn selector switch to PROD and repeat observations for product tank. Compare pressures to specifications in subsection 3.1.2 B, step 8 and use Troubleshooting Guide appropriately. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-10 nr 3.2.2 Concentrator Won't Run 1. Verify unit is plugged into proper AC power source. Set power switch to ON position. 2. Green front panel power indicator should illuminate. Audio alarm should activate (open flowmeter and drain product canister first). If circuit breaker pops out refer to Troubleshooting Guide. 3. Test for fan operation by placing hand over vents at bottom left side of cabinet. With concentrator unplugged, inspect all wiring and connections for looseness or discon- nects. 5. Open cabinet and observe compressor and cooling fan. If compressor and cooling fan start to operate and both immediately stop, see Section 4, step 16 for solenoid testing. 6. If cooling fan runs and compressor runs momentarily and then stops, remove the small pilot air tube from the balance solenoid valve to vent pressure from the reservoir canister. Reconnect the pilot air hose and attempt to restart the concentrator. 3.2.3 Improper Operation 1. Noise beyond normal operation is usually detectable by careful observation with cabinet open. 2. A pneumatic leak, especially on the intake tubes and fittings of the compressor may cause excessive noise. Compressor vibration may cause component rubbing which may be another source of noise. 3. Refer to subsection 3.1.6 for discussion on sound. 4. Reported erratic behavior or intermittent shutdown should be tested and observed where possible. Refer to subsection 3.2.2 if appropriate. 3.3 Troubleshooting Guide This section is designed for easy reference to determine the problem with a Companion concentrator. Generally, it is necessary to do some investigation to discover what symptoms the concentrator is exhibiting. Subsection 3.2 Operational Checklist should be consulted before proceeding with troubleshooting in order to get a clear understanding of all of the symptoms a concentrator problem may have. Then match the descriptions in the "Symptom" column that describes the unit's problem as accurately as possible. Next, look at the "Probable Cause” column which will detail several possible reasons (most likely to least likely to occur) to explain the observed behavior of the concentrator. Finally, column three "Corrective Action” suggests possible solutions, tests and/or repairs to resolve a suspected probable cause. This guide is not all-inclusive but it is intended to serve as an outline for solving operational problems. WARNING Some of the following procedures require AC power be applied when testing. To prevent the possibility of serious electrical shock, connect grounding wire to metal components and follow standard safety procedures. Companion 492a/590 Oxygen Concentrators 3-11 Service Manual 5-15-91 SYMPTOM TABLE 3-1 PROBABLE CAUSE CORRECTIVE ACTION 1. No audio alarm when power switch turned ON and AC power connected to con- centrator. AC power disconnected AC power connected, battery removed. a. Residual pressure in product tank b. Loose electrical connection С. Defective audio alarm d. Defective pressure switch e. Defective power switch f. Defective battery g. Control PCB fuse h. Defective control PCB Open flowmeter knob to reduce product tank pressure below 4.0 psi. Inspect all electrical connections, especially pressure switch, power switch, audio alarm, and Control PCB at J2 for good connection. Remove wires from audio alarm terminals. Using wire jumpers and observing both the battery and alarm polarity, connect the 9V battery directly to the alarm terminals. If audio alarm does not activate, replace alarm per the appropriate procedure in Section 4. Remove wires from pressure switch terminals. Using a wire jumper, connect both wires together. If unit alarms, replace pressure switch per the appropriate procedure in Section 4. Test power switch by jumping white and black wires connected to terminals 4 and 5. If alarm sounds replace switch per appropriate service procedure in Section 4. Test with multimeter set on DC. If less than 8.0 volts, replace battery. Test and replace per appropriate service procedure in Section 4. Replace per appropriate procedure in Section 4. 2. Continuous alarm/ power switch on/ pilot lamp not lit/unit not operating. a. No power to unit b. No power at AC outlet с. Circuit breaker activated or defective d. Control PCB fuse e. Defective control PCB f. Loose electrical connection Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-12 Connect power cord to AC outlet. Inspect household circuit breaker or fuse. Reset circuit breaker. If circuit breaker activates, disconnect concentrator from AC power and inspect internal wiring for short circuits. Inspect for locked compressor, shorted capacitor, or faulty circuit breaker. Test and replace per appropriate service procedure in Section 4. Replace per appropriate service procedure in Section 4. Inspect AC power cord, circuit breaker, power switch and Control PCB forgood connections. SYMPTOM TABLE 3-1 PROBABLE CAUSE CORRECTIVE ACTION g. Defective power switch Test power switch by jumping blue and brown wires connected to terminals 1 and 2. If unit starts, replace switch per appropriate service procedure in Section 4. Continuous alarm/ power switch and pilot lamp ON/ com- pressor and cooling fan not running. a. Disconnected wire to one of the three solenoids b. Defective solenoid valve Loose electrical connection Defective PCB e. Defective cooling fan causing compressor thermal cutoff switch activated after concentrator has been operating Reconnect. Set power switch off, then on. Test and replace per appropriate service procedure in Section 4. Inspect internal wiring Replace per appropriate service procedures in Section 4. Test for 120 VAC at fan terminal wires. Replace per appropriate service procedure in Section 4. Continuous alarm/ power switch and pilot lamp ON/ com- pressor runs for a period of time, then stops/ cooling fan still operating. a. Compressor over-heated due to restricted cabinet cooling-air flow b. Defective compressor internal thermal switch Check that unit is operating in an open environment and the air inlet is unobstructed. See also subsection 3.1.4. Allow compressor to cool and retest. Replace compressor per appropriate service procedure in Section 4. Continuous alarm/ power switch and pilot lamp ON/ com- pressor won't start/ cooling fan operates. a. Residual pressure in reservoir canister b. Capacitor (may cause circuit breaker to actuate) Loose electrical connection Faulty compressor Wait 10-15 seconds for valves to cycle and relieve pressure. Check wire terminals at capacitor. Replace capacitor per appropriate service procedure in Section 4. Inspect internal wiring. Replace compressor per appropriate service procedure in Section 4. Concentrator runs - No power switch light a. Loose wires to L.E.D. b. Plugto L.E.D. reversed c. Faulty L.E.D. Check wires to power switch L.E.D. If the plug to the L.E.D. is reversed, the L.E.D. will not illuminate. Refer to Section 4, step 7h ACTION and NOTE statements. Replace lamp. Companion 492a/530 Oxygen Concentrators 3-13 Service Manual 5-15-91 SYMPTOM TABLE 3-1 PROBABLE CAUSE CORRECTIVE ACTION 7. Low reservoir pressure/ compressor and cool- ing fan running/ low concentrations/ con- centrator may alarm. a. Compressor inlet prefilters or bacteria filter dirty b. Leak c. Diaphragm leak in supply, exhaust, or balance pilot valves d. Defective solenoid valve e. Weak or worn compressor f. Outlet pressure too high Replace dirty filters. Leak test entire pneumatic system, per subsection 3.1.2 A. Inspect all 5 pilot air valves per appropriate service procedure in Section 4. Replace as necessary. If air leaks through pilot valve exhaust port and pilot valve has been inspected with no problems found, replace solenoid per appropriate service procedure in Section 4. Replace entire compressor or replace piston cup seals per appropriate service procedure in Section 4. Test per subsection 3.1.2 B steps 12 through 15. If incorrect reset regulator per Section 4, step 15. 8. High reservoir canister pressure/ low con- centrations/ unit may alarm. a. Stuck diaphragm in pilot air valve b. Contaminated sieve beds Inspect solid diaphragms in the supply and exhaust pilot valves per appropriate service procedure in Section 4. Replace per appropriate service procedure in Section 4. 9. Imbalance in drop point pressure/ low con- centration/ unit may alarm. a. Leaks b. Faulty diaphragm in pilot air valve с. Restrictor tube assembly blocked d. Mismatched restrictor tube assembly Companion 492a/530 Oxygen Concentrators Service Manual 5-15-91 3-14 Leak test per subsection 3.1.2 A. Check especially bottoms of sieve beds, and supply and balance pilot valve fittings. Repair as necessary. Remove test and replace, if necessary per appropriate service procedure in Section 4. Remove compression nuts from restrictor tube assemblies and inspect orifice for blockage. Drop point differences greater than 0.4 PSI between cycles. See subsection 3.4; Balancing Orifices. rr TABLE 3-1 SYMPTOM PROBABLE CAUSE CORRECTIVE ACTION 10. Low or erratic product a. Leaks Leak test per subsection 3.1.2 A. canister pressure/ Check especially from restrictor tubes low concentrations/ to flowmeter. Repair as necessary. unit may alarm. b. Regulator dirty, out of adjust- ment or defective c. Faulty diaphragm in balance pilot valve d. Stuck supply or exhaust pilot valve solid diaphragm e. Incorrectly sized restrictors Disassemble, clean, reassemble, and test per appropriate service procedure in Section 4. Replace if necessary. Remove, test and replace, if necessary per appropriate service procedure in Section 4. First disassemble and inspect the supply pilot valve diaphragms and then the exhaust pilot valve diaphragms. Refer to the appropriate service procedure in Section 4. Change restrictors per subsection 3.4, Balancing Orifices. 11. Unit alarms/ cycles/ Defective pressure switch or set- Test or replace per appropriate service oxygen flows and ting out of adjustment procedure in Section 4. concentration within specifications/ pres- sures normal. 12. Low or no oxygen flow a. Flowmeter knob turned oft Set flow to desired level. or pressure/ unit operating/ no alarm. b. Kinked or obstructed tubing Inspect internal tubing from product canister to flowmeter. c. Regulator out of adjustment or Test per subsection 3.1.2 B, steps 12 defective through 15. Replace per appropriate service procedure in Section 4. Unit Alarm d. Disconnected tubing Inspect internal tubing and connections from compressor to flowmeter. 13. Noise. a. Component rubbing Inspect per subsection 3.1.6; steps 3 (Pulsating air) (Grinding sound) b. Tubing disconnected C. Compressor d. Compressor or cooling fan through 6. Inspect per subsection 3.1.6; steps 1 through 5. Piston seal torn or worn out. Replace per appropriate service procedure in Section 4. Inspect per subsection 3.1.6, step 6. Replace as needed per Section 4. Companion 492a/590 Oxygen Concentrators 3-15 Service Manual 5-15-91 3.4 Balancing orifices Restrictor tube assemblies each contain an orifice drilled to a specific size. A number is assigned to each tube that represents the liter flow of gas through the orifice at a standard pressure. It may be necessary to change restrictor tubes in a Companion Concentrator for one of the following reasons: A) During performance verification testing, concentration is out of specification and pneumatic testing reveals an imbalance in the reservoir canister drop point pressures. (Refer to subsection 3.1.2) B) The drop point pressures are within specifications, however the concentration at the maximum flowrate is out of specification. NOTE Before performing the following procedure it is necessary to ensure that all leaks have been eliminated from the concentrator’'s system. 1. Open the concentrator cabinet, connect the pneumatic test kit, and after five minutes of operation, record the drop, fill, and kick pressures of each sieve cycle as described in subsection 3.1.2 B, steps 1 through 7. 2. To determine the balance pressure of the sieve beds, compare the drop pressures of two consecutive pressure cycles. If the pressures differ by more than 0.4 PSI, one or both of the restrictor tubes may need to be replaced to achieve a pressure balance. 3. Determine which drop pressure value corresponds to which restrictor tube (and consequent- ly which sieve bed), by operating the concentrator and observing the test gauge. Lay your finger lightly over the open exhaust port on the end of solenoid valve 2 (refer to Figure 2-8 for solenoid 2 location). With the test gauge connected and the concentrator operating, feel for a slight puff of air from the exhaust port of solenoid 2. This represents the close of the solenoid and thus the end of the fill cycle for sieve canister 2. Immediately observe the gauge and you should note a rapid climb (kick pressure). After this kick pressure the gauge needle should fall to the next drop point. This drop point is the beginning pressure for the right sieve canister (sieve canister 1), which corresponds to the right restrictor tube. 4. Inorderto balance (bring the pressures within .4 PSI at the drop point) the orifices, it will be necessary to change one restrictor tube and install a new one with a larger or smaller orifice. (A larger rating on the tube will result in more flow and lower sieve bed filling pressure. The opposite will occur for a smaller orifice number.) Refer to the example below to appropriately size the restrictors. EXAMPLE: After five minutes of continuous operation, the drop point pressures are 14.9 PSIG and 14.2 PSIG. The restrictors are rated 8.8 (left) and 8.7 (right). Through the testing described above, you have determined that the right restrictor is producing the 14.2 PSIG drop pressure. Preferably, you should bring the lower drop pressure up to the higher drop pressure of the other sieve canister. Thus, to bring the 14.2 PSIG drop pressure up, replace the right restrictor tube with one having less flow (and a smaller rating). Try a restrictor that is .1 or.2 smaller. (Refer to the parts list for Figure 4-2 in Section 4 for part numbers that correspond to the desired orifice size. In this case, an orifice sized 8.6 or 8.3). 5. By trying different orifice sizes you should be able to accomplish a balance. Although the actual pressures may change up or down as the concentrator acclimates, the drop point balance should remain constant. Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-16 r 3.4 Balancing orifices (continued) 6. 7) Occasionally a drop point balance may exist but the oxygen concentration may not be within specification. In this case the restrictor tubes may have orifice sizes that are too restrictive (sieve bed fills too slowly) or too large (sieve bed pressure too low for proper adsorption.) Increase or decrease the orifice sizes as appropriate by observing the reservoir canister pressure specifications as a guide. Whenever restrictor tubes have been changed, operate the concentrator a minimum of 12 hours and recheck both the reservoir canister pressures and the oxygen concentration for compliance with specifications. Companion 492a/590 Oxygen Concentrators 3-17 Service Manual 5-15-91 Companion 492a/590 Oxygen Concentrators Service Manual 5-15-91 3-18 (blank) ‘J 9. SECTION 4. SERVICE AND REPAIR This section describes how to service the individual components of the Companion 492a and 590 Oxygen Concentrators. Included are instructions, where applicable, for removal, disassembly, operational check, cleaning, adjustment, alignment, reassembly, and installation. After removing a component, visually inspect for damage or any other indication that the component is defective. Unless otherwise specified, replace as needed with a new component. Refer to the exploded-view drawings for replacement part numbers. For service and repair procedures relating to the optional Oxygen Concentration Indicator (OC!) refer to section 6. WARNING The concentrator is electrically powered. To prevent serious injury or death, observe standard safety procedures when servicing. Always ensure that power is removed before servicing concentrator. CAUTION After a component has been serviced, the concentrator's overall system operation must be verified per section 3.3, Performance Verifications, before the concentrator is used. CAUTION Ensure that the compressed-air supply used to clean and dry components is clean and oil-free. Companion 492a/590 Oxygen Concentrator 4-1 Service Manual 5-15-91 SECTION 4. SERVICE AND REPAIR (Continued) Service procedures are provided in Table 4-1 as follows: as «= ll - CO N ГО [0 = 05005 05 +5 ША a + NN + OMONYSO UN © N Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 Component Cabinet Circuit Breaker Hour Meter Audio Alarm Flowmeter Flow Control Valve Flowmeter Power Switch Control Printed Circuit Board Compressor Platform Assembly Compressor (492a) Compressor (590) Capacitor Cooling Fan Pressure Switch Pressure Regulator Solenoid Valve Sieve Canister Pilot valve Restrictor Tube Assemblies Outlet Gas Filter Shock Mount Compressor Cup Seals (590) 4-2 TABLE 4-1. SERVICE PROCEDURES (Refer to the appropriate figure(s) for numbers shown in bold print.) COMPONENT ITEM 1. Cabinet (Figure 4-1) OPENING a. Companion 492a/590 b. Airinlet filter 14 C. Screw 16 d. Screw 15 e. Cabinet CLOSING f. Cabinet side panels (Figure 4-3) 4-3 ACTION Remove humidifier, (if installed), from oxygen outlet connector. WARNING Electrical shock hazard. Verify AC power is removed before opening concentrator cabinet. Remove from right side panel. Using a #2 phillips screwdriver, remove from cabinet side panel. Using a #2 phillips screwdriver with a shatt at least seven inches long, loosen screw 15. The cabinet's left and right case hal- ves Will separate as screw is loosened. NOTE An O-ring is installed over screw item 15 on interior of cabinet and is intended to hold screw in place during reassembly of cabinet's case halves. Removal of screw from right case half is not necessary. Pull left and right side panels away from one another to open cabinet. The cabinet will hinge in rear of concentrator. Do not open cabinet more than approximately a 90 degree angle. CAUTION The concentrator will not be stable setting on its casters if left and right side panels are opened more than approximately 90 degrees. Ensure that cabinet screw 37 and o-ring 41 are installed in right side panel. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ACTION Cabinet side panels Close right and left side panels to within approximately two inches of each other. Insert screw 18 through right side panel. Push screw in far enough so that it ex- tends through right side panel tube 22 and into left side panel tube 50 (Figure 4-2). Cabinet side panel Tighten screw 37 first and then screw 18. screws 18 and 37 (Figure 4-3) Air inlet filter Reinstall on right side panel. 2. Circuit Breaker 19 (Figure 4-3) REMOVAL Cabinet Perform OPENING procedure. WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. PCB harness Remove three wire harness connectors connectors from Control PCB 44. Tube Remove from tube clamp 24. Inlet air duct cover 46 Remove five screws 45, and air duct cover 46, from right case interior. Control PCB does not need to be removed from cover. Circuit breaker Disconnect two terminal wires. terminals Filter compartment door Remove. 16 Hex nut 10 Remove from circuit breaker (Figure 4-1). Circuit breaker Remove from cabinet interior. INSTALLATION Replacement circuit Install knurled nut, item 6, (Figure 4-1) (Figure 4-1) breaker onto circuit breaker's threaded shaft so Replacement circuit breaker Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-4 that approximately 1/4 inch of threads are exposed. Install in right side panel so that circuit breaker number is in a vertical and readable position. Secure circuit breaker with hex nut in interior of filter compart- ment. 7 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM ACTION К. Circuit breaker Using needle-nose pliers and old circuit terminals breaker as a guide, carefully bend each terminal at approximately a 45 degree angle toward cabinet hinge. Reconnect wires to terminals. The circuit breaker is not polarity sensitive, either wire may be connected to either terminal. |. Air inlet duct cover, Reverse removal procedures and tube and Control reinstall. PCB wire harness connectors m. Cabinet Perform CLOSING procedures. 3. Hour Meter 21 (Figure 4-3) REMOVAL a. Cabinet Perform OPENING procedure. WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. b. PCB harness Remove three wire harness connectors connectors from Control PCB 44. Tube Remove from tube clamp 24. Air inlet duct cover 46 Remove five screws 45 and cover 46 from right case interior. Control PCB does not need to be removed from cover. e. Hour meter terminals Disconnect two terminal wires. f. Filter compartment door Remove. 16 g. Rivets (Figure 4-1) Remove two rivets 13 securing hour meter with 1/8 inch drill. h. Hour meter Remove from filter compartment interior. INSTALLATION i. Replacement hour Insert into mounting hole from filter com- meter partment so that last digits (red numerals) are toward bottom of concentrator. Secure to cabinet with two 1/8 inch pop- rivets. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 1 + TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM Hour meter terminals ACTION Using needle-nose pliers and old hour- meter as a guide, carefully bend each terminal at approximately a 45 degree and toward cabinet hinge. Reconnect wires to terminals. The hour meter is not polarity sensitive, either wire may be connected to either terminal. k. Air inlet duct cover, Reverse removal procedures and tube and Control reinstall. PCB wire harness connectors |. Cabinet Perform CLOSING procedures. 4. Audio Alarm 25 (Figure 4-3) REMOVAL a. Cabinet Perform OPENING procedure. b. Audio alarm terminals Disconnect two terminal wires. c. Air inlet filter 17 Remove. d. Retaining ring 11 Remove from filter compartment. (Figure 4-1) e. Audio alarm Remove from cabinet interior. INSTALLATION f. Replacement audio Insert into mounting hole from right alarm cabinet interior. g. Retaining ring Attach to replacement audio alarm from filter compartment. h. Audio alarm terminals Reconnect red wire to positive terminal Cabinet and black wire to negative terminal. Perform CLOSING procedures. 5. Flowmeter Flow Control Valve 2 (Figure 4-3) REMOVAL Flowmeter knob 1 Flow control valve Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-6 Remove by pulling away from flowmeter. Remove from flowmeter body with 1/2 socket wrench. CAUTION To prevent flowmeter damage, the flow control valve must be set to a fully counterclockwise position before reassembly. ” TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT Replacement fiow control valve ACTION Install into flowmeter body and carefully tighten with 1/2 inch socket wrench. d. Flowmeter knob Align with flat on flow control valve shaft and install. 6. Flowmeter 3 (Figure 4-3) REMOVAL Cabinet Perform OPENING procedure. Oxygen Tubes 12, 10 Disconnect two tubes from flowmeter inlet and outlet connectors. c. Two flowmeter Remove from flowmeter's threaded con- brackets 8 and nuts nectors. 9 d. Cabinet control panel Remove flowmeter. INSTALLATION e. Replacement flowmeter Install in mounting hole with control knob at top of concentrator. f. Flowmeter brackets Install on replacement flowmeter. and nuts g. Oxygen tubes Reconnect tubing to inlet and outlet flow- meter connectors. Refer to Section 2, (Figure 2-1) for tubing orientation. h. Cabinet Perform CLOSING procedures. 7. Power Switch 5 (Figure 4-3) NOTE For units equipped with optional O.C.I refer to section 6 for Power Switch service procedures. REMOVAL a. Cabinet Perform OPENING procedure. WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. b. PowerL.E.D. 4 Using a needle-nose pliers, unplug the Power switch terminals Power switch body connector from the Power L.E.D.'s leads. Disconnect four terminal wires. Locate four retaining tabs at each corner. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM e. Fourretaining tabs ACTION Press inward with flat-blade screwdriver, one after the other, while pushing switch through control panel toward outside of cabinet. NOTE The Power L.E.D. is mounted in a removable bezel. The bezel and L.E.D. are supplied with replace- ment power switch. Verify correct orientation of bezel before installa- tion. Power L.E.D. should be posi- tioned above switch closest to terminals #2 and #5. Position so that power L.E.D. is towards top of concentrator.Press into mounting hole from outside of control panel until retaining tabs catch. Connect four terminal wires. Referto Sec- tion 2, (Figure 2-2) for wire orientation. Connect to wire harness connector. The longer L.E.D. lead must connect to posi- tion 2 (black wire) on wire connector. Refer to Section 2, (Figure 2-2) for con- nector orientation. NOTE The Power L.E.D. is a polarity sen- sitive device. Improper connection of wire harness connector will not damage L.E.D., but will prevent il- lumination when power is applied. Verify proper operation of the Power L.E.D. after installation. If the L.E.D. does not illuminate, remove the con- nector, rotate it 180 degrees, reinstall the connector, and retest. Perform CLOSING procedure. INSTALLATION f. Replacement power switch g. Power switch terminals h. Power L.E.D. leads i. Cabinet 8. Control Printed a. Cabinet Circuit Board 44 (Figure 4-3) Companion 492a/590 Oxygen Concentrator Service Manual 5-15-81 4-8 Perform OPENING procedure. p TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM ACTION CAUTION The control printed circuit board contains complementary metal- oxide semiconductor (CMOS) in- tegrated circuits (IC's) which are static-sensitive devices. To prevent IC damage, observe standard safety procedures as follows: * Wear grounding wrist strap. * Work on grounded conductive mat. « Handle PCB by edges only. * Store PCB in conductive plas- tic bag. WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. / FUSE TESTING b. Fuse 69 Remove from fuse holder on Control PCB. Verify resistance of approximately 15 ohms using digital multimeter. Reinstall in fuse holder. REMOVAL C. PCB harness Remove three wire harness connectors connectors from Control PCB 44. d. Control PCB Remove five screws 43 from control PCB. Separate Control PCB from cover. INSTALLATION e. Replacement Control Install into cabinet and secure with five PCB SCrews. f. ~~ Three wire harness Reverse removal procedures and connectors reinstall. g. Cabinet Perform CLOSING procedure. 9. Compressor Platform Assembly 49 (Figure 4-3) REMOVAL a. Cabinet Perform OPENING procedure. Companion 492a/590 Oxygen Concentrator 4-9 Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT DISASSEMBLY (Figure 4-4) REASSEMBLY Compressor wire harness connector 25 (Figure 4-4) Compressor platform assembly Inlet tube and clamp 35 Two screws 14 and two clamps 15 and 21 Capacitor 17 terminals Strap 6 and grommet 16 Cooling fan 22 terminals Two screws 33, washer 28, and locknuts 29 Heat exchanger 30 Compressor Compressor platform assembly Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-10 ACTION WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. Disconnect and remove compressor wire harness connector and grommet 26 (Fig- ure 4-4) from air dam 62 (Figure 4-3). Using a magnetic tip #2 phillips screwdriver, remove two screws 50. Dis- connect inlet tube 26 from platform con- nector while sliding platform assembly out of cabinet. Remove from compressor and platform connectors. Remove from platform assembly. WARNING To prevent serious electrical shock, discharge capacitor by shorting ter- minals with insulated - handle screwdriver. Disconnect two terminal wires. Cut strap and pull each capacitor wire, one at a time through the grommet. Disconnect two terminal wires. Use a 5/64-inch 90° Allen wrench and a 5/16-inch socket to remove from com- pressor. Two clamps 27 do not need to be removed from the heat exchanger. Unscrew brass nut 31 and remove heat exchanger 30 from compressor. Support compressor with one hand and remove four cap screws 12 with 7/16-inch socket wrench. Separate compressor from platform. Reverse DISASSEMBLY procedure. ~~ TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM ACTION NOTE Apply a small amount of Loctite 242 to threads of four compressor mounting bolts 12 before reassemb- ly. INSTALLATION m. Compressor platform Reverse REMOVAL procedure. Do not assembly over-tighten two screws 50. 10. Compressor (492a) REMOVAL a. Compressor platform Perform REMOVAL and DISASSEMBLY assembly procedures steps 9a through 9k. DISASSEMBLY b. Twelve cylinder head Remove from cylinder heads 26, 9. (Figure 4-5) screws 1 C. Cylinder heads and Remove both cylinder heads, along with valve plates connector tubes 2, and both valve plates 4. d. Piston sleeves Remove both sleeves 6. г, HOUSING e. Four-blade fan 19 Pry away from motor shaft 15 with WELL A flatblade screwdriver. f. Connecting rod screw 8 Turn motor shaft until screw is visible through lower access port on wire-har- ness side of housing. g. Connecting rod screw Loosen several turns with 5/32-inch allen wrench. п. Set screw 17 Turn eccentric 18 until screw is visible through upper access port. |. Set screw Loosen only 1/4-turn with 1/8-inch allen wrench. В Piston 7 Slip piston connecting rod off of bearing 16, and slide entire eccentric bearing as- sembly 20 off shaft. NOTE Bearing is bonded to connecting rod with Loctite. If necessary, insert screwdriver blade into slit at rod bot- tom, and turn to break rod loose from bearing. ge k. Piston Grasp at top inside housing well, and slip connecting rod off shaft. Companion 492a/590 Oxygen Concentrator 4-11 Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT HOUSING WELL B CLEANING REASSEMBLY — Connecting rod Five-blade fan 11, eccentric bearing assembly, and piston Cylinder head interiors and housing wells Valve plates Eccentric bearing assemblies Compressor body New connecting rod Screw Piston Piston Bearing Bearing Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-12 ACTION Move to one side of shaft, and then lift while moving back into housing above shaft. Remove through top of housing. Replace connecting rod screw. Repeat steps e through | to remove. WARNING To prevent personal injury, use face shield and observe standard safety procedures when using com- pressed-air gun. Blow out or brush away dirt and dust. Wipe off both sides with clean cloth. Wipe oft with clean, cloth. Scrape off ex- cessive Loctite from outside diameter with knife. Stand on end, housing well 8 facing up. CAUTION Keep Loctite away from all plastic parts. Place one drop only of Loctite 242 on threads, and insert, but do not tighten, into side hole on connecting rod. Position above housing well so that con- necting rod screw head is facing wire-har- ness side of housing. Insert into housing, and place connecting rod over shaft end. Wipe any grease or oil off outer race. CAUTION Do not allow Loctite to seep inside bearing or on eccentric. Apply very thin film of Loctite 680 to outer circumference with small pad. Pr TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT у ECCENTRIC- PISTON ALIGNMENT ~~ dad. ab. ac. ad. de. af. ag. ah. a. aj. ak. al. Eccentric bearing assembly Eccentric bearing assembly Eccentric Connecting rod Connecting rod screw Sleeve Sleeve Sleeve Compressor body Sleeve Valve plate Valve plate Two of the six cylinder head screws Motor shaft Motor shaft Set screw 4-13 ACTION Position so that set screw is aligned with flat on shatt. Hold piston steady with one hand, and slip assembly onto shaft, bearingfirst, and into connecting rod. Tum until connecting rod screw is visible through lower access port on wire-har- ness side of housing. Center on eccentric bearing assembly so that bearing face is flush with rod face. CAUTION To prevent bearing failure, do not ovenrtighten connecting rod screw. Tighten to 15 inch-pounds (1.69 new- ton/meters) with torque wrench. Wipe out inner surface. Inspect for scratches and burrs. Position above housing well so that stepped edge is facing out. Install carefully onto piston, taking care not to damage seal around piston head. Set on feet. Position in center of housing well. Position over piston, O-ring 5 side down, so that head of reed valve screw is aligned with notch on piston head. Place on housing well to engage stepped edge of sleeve. Note that six mounting holes are aligned with those in housing. Insert into two valve plate holes on housing’s centerline, and rotate two turns. Hold valve plate in place over piston, and spin shaft. Turn unit eccentric set screw is visible through upper access port. tighten to 30 inch-pounds (3.39 new- ton/meters) with torque wrench. Do not allow eccentric to move on shaft. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM am. Valve plate an. Sleeve ao. O-ring ap. Valve plate aq. Head gasket 3 for cylinder head B 9 ar. Cylinder head B as. Twocylinder head screws at. Eccentric au. Compressor body av. Compressor aw. Cylinder head screws ax. Cylinder head screws ay. Four-blade fan az. Five-blade fan INSTALLATION ba. Compressor platform assembly Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-14 ACTION Remove two cylinder head screws and then plate. Hold, and turn eccentric. Ensure that sleeve is not touching front or back shoulder of well. Inspect for cuts and tears. Ensure proper- ly seated in valve plate groove. Repeat steps ag and ah. Inspect for cuts and tears. Ensure proper- ly seated in cylinder head groove. Place over valve plate, and place con- necting rod in the up position. Insert into holes on housing's center line, and tighten until snug. Turn and check for alignment between piston and valve plate. Stand on end, housing well A facing up. Repeat steps r through at with the follow- ing change: Connector tubes 2 are inserted between cylinder head A and B before head A is connected to housing. Insert remaining into heads, and tighten all 12 to 20 inch-pounds (2.27 new- ton/meters) with torque wrench in se- quence shown in (Figure 4-7). Repeat step aw for 30 inch-pounds (3.41 newton/meters). Press onto housing well A shaft, with flat in fan hole aligned with flat on shaft and with spring clamp facing in. Press onto housing well B shaft, with flat in fan hole aligned with flat on shaft and with spring clamp facing in. CAUTION Ensure that both fans are securely in place. Reverse DISASSEMBLY and REMOVAL procedures steps 9a through 9k. m- TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT 11. Compressor (590) REMOVAL DISASSEMBLY (Figure 4-6) HOUSING WELL A Compressor platform assembly Eight cylinder head screws 29 Cylinder heads and valve plates Piston sleeves Fan À 19 Connecting rod screw 16 Connecting rod screw Set screw 17 Set screw Piston assembly Piston assembly Connecting rod Piston assembly 4-15 ACTION Preform REMOVAL and DISASSEMBLY procedures setup 9a through 9k. Remove from cylinder heads 28 and 10. Remove both cylinder heads, along with connector tubes 1 and both valve plates 3. Remove both sleeves 5. Pry away from motor shaft with flat-blade screwdriver. Label Fan A for correct reas- sembly. Turn motor shaft until screw is visible through lower access port on wire-har- ness side of housing. Loosen several turns with 5/32-inch allen wrench. Turn eccentric 20 until screw is visible through upper access port. Loosen only 1/4-turn with 1/8-inch allen wrench. Slip connecting rod 9 off bearing 21, and slide entire eccentric bearing assembly 18 off shaft 22. NOTE Bearing is bonded to connecting rod with Loctite. If necessary, insert screwdriver blade into slit at rod bot- tom, and turn to break rod loose from bearing. Grasp at top inside housing well, and slip connecting rod off shaft. Move to one side of shaft, and then lift while moving back into housing above shaft. Remove through top of housing. Replace connecting rod screw. Remove four screws 6 on each piston to separate retainer plate 7, cup seal 8, and connecting rod 9. Companion 432a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT HOUSING WELL B CLEANING REASSEMBLY B fan 12, eccentric bearing assembly, and piston Cylinder head interiors and housing wells Valve plates & piston assembly components Eccentric bearing assemblies Compressor body Piston assembly Piston assembly Piston assembly Piston assembly New connection rod screw Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-16 ACTION Repeat steps e through m to remove. WARNING To prevent personal injury, use face shield and observe standard safety procedures when using com- pressed-air gun. Blow out or brush away dirt and dust. CAUTION Use care not to damage intake and exhaust reed valves 23 on valve plates. Wipe off with clean cloth. Wipe off with clean cloth. Scrape off ex- cessive Loctite from outside diameter with knife. Stand on end, housing well B facing up. Install cup seal on stepped edge of retainer plate. Cupped side of cup seal should face retainer plate. Install the cup seal and retainer plate on the connecting rod. Rotate the retaining plate to position the index mark and beveled edges as shown in Figure 4-8. Place one drop of Loctite 242 on threads of the four screws. Install the screws and torque to 20-inch pounds (2.27 new- ton/meters). Repeat steps s through u for the second piston assembly. CAUTION Keep Loctite away from all plastic parts. Place one drop of Loctite 242 on threads, and insert, but do not tighten, into side hole on connecting rod. pP TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT > ECCENTRIC- PISTON ALIGNMENT Y a ITEM X. aa. ab. ac. ad. de. af. ag. ah. a. aj. Piston Piston Bearing Bearing Eccentric bearing assembly Eccentric bearing assembly Eccentric Connecting rod Connecting rod screw Sleeve Sleeve Compressor body Sleeve 4-17 ACTION Position above housing well so that con- necting rod screw head is facing wire-har- ness side of housing. Insert into housing, and place connecting rod over shaft end. Wipe any grease or oil off outer edge of bearing. CAUTION Do not allow Loctite to seep inside bearing or onto eccentric. Apply very thin film of Loctite 680 to outer circumference with small pad. Position so that set screw is aligned with flat on shaft. Hold piston steady with one hand and slip assembly onto shaft, bearing first, and into connecting rod. Turn until connecting rod screw is visible through lower access port on wire-har- ness side of housing. Center on eccentric bearing assembly so that bearing face is tlush with rod face. CAUTION To prevent bearing failure, do not overtighten connecting rod screw. Tighten to 15 inch-pounds (1.69 new- ton/meters) with torque wrench. Wipe out inner surface. Inspect for scratches and burrs. Install carefully onto piston, taking care not to damage cup seal. Set on feet. Position in center of housing well. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ak. al. am. an. do. ap. aq. dar. as. at. au. av. aw. ax. ay. az. ITEM Valve plate Valve plate Four cylinder head screws Motor shaft Motor shaft Set screw Valve plate Sleeve O-ring 6 Valve plate Head gasket 4 for cylinder head B Cylinder head B Four cylinder head screws Eccentric Compressor body Compressor Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 ACTION Position over piston, O-ring side down, so that headof reed valve screw 26 is aligned with notch on piston head and the reed valve on the top of the valve plate is positioned nearest to the harness side of the compressor. Place on housing well to engage edge of sleeve. Note that four mounting holes are aligned with those in housing. Insert into four valve plate holes and rotate two turns. Hold valve plate in place over piston, and spin shaft. Tum until eccentric set screw is visible through upper access port. Tighten to 30 inch-pounds (3.39 new- ton/meters) with torque wrench. Do not allow eccentric to move on shaft. Remove four cylinder head screws and then plate. Hold sleeve, and turn eccentric. Ensure that sleeve is not touching front or back shoulder of well. Inspect for cuts and tears. Ensure proper- ly seated in valve plate groove. Repeat steps ak and al. Inspect for cuts and tears. Ensure proper- ly seated in cylinder head groove. Place over valve plate, and place con- necting rod in the up position. Insert into holes on housing's center line, and tighten until snug. Turn and ensure piston does not contact valve plate. Stand on end, housing well A facing up. Repeat steps aw thru ax for cylinder head A side of compressor with the following change: Connector tubes 1 are inserted between cylinder head A and B before head A is connected to housing. 7 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT INSTALLATION ITEM ba. bb. be. Cylinder head screws Cylinder head screws Fan A Fan B Compressor platform assembly ACTION Tighten all to 20 inch-pounds (2.27 new- ton/meters) with torque wrench in a crisscross pattern. Repeat step ba for 30 inch-pounds (3.41 newton/meters). Press onto housing well A shaft, with flat in fan hole aligned with flat on shaft and with spring clamp facing in. Press onto housing well B shaft, with flat in fan hole aligned with flat on shaft and with spring clamp facing in. CAUTION Ensure that both fans are securely in place. Reverse DISASSEMBLY and REMOVAL procedures. 12. Capacitor 17 rr (Figure 4-4) REMOVAL INSTALLATION Cabinet Compressor platform assembly Capacitor 17 terminals Two straps 18 Replacement capacitor Capacitor terminals 4-19 Perform OPENING procedure. Perform REMOVAL procedure, step 9 a through c. WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. Discharge capacitor by shoring terminals with insulated - handle screwdriver. Disconnect two terminal wires. Cut straps and remove capacitor from compressor platform. Attach to underside of platform assembly with two new straps 18. Reconnect two terminal wires. NOTE The capacitor is not a polarity sensi- tive component. Either wire may be connected to either terminal. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT Compressor platform assembly Cabinet ACTION Perform INSTALLATION procedure. Perform CLOSING procedure. 13. Cooling Fan 22 (Figure 4-4) REMOVAL INSTALLATION Cabinet Cooling fan 22 terminals Two screws 4 Cooling fan Replacement cooling fan Two screws Cooling fan terminals Perform OPENING procedure. WARNING Electrical shock hazard. Verify AC power has been removed before servicing concentrator. Disconnect two terminal wires. Remove from compressor platform as- sembly. Remove by sliding fan toward front of compressor platform assembly. The rear of the cooling fan is retained by clips which are molded into the platform. Position the fan with the flow arrow point- ing down and blades up as shown in Fig- ure 4-4 and slide the rear of the fan into the two clips on the compressor platform. Install into compressor platform assembly and tighten. Reconnect the two terminal wires. Cooling fan Verify correct operation per subsection 3.1.4. NOTE The cooling fan is not a polarity sen- sitive component. Either wire may be connected to either terminal. Cabinet Perform CLOSING procedure. 14. Pressure Switch 37 (Figure 4-2) REMOVAL Cabinet Perform OPENING procedure. Pressure switch Disconnect two terminal wires. terminals Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-20 rr TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM Pressure switch ACTION Using a 7/16-inch open end wrench, un- screw from product canister 34. INSTALLATION d. Replacement pressure Apply a small amount of pipe cement to switch pressure switch thread. Distribute the ce- ment around the entire fitting. Do not allow cement to enter interior of pressure switch. e. Replacement pressure Install into product canister and tighten switch with wrench. Do not overtighten. Position the switch so that the terminals are toward the top of the concentrator. f. Pressure switch Reconnect two terminal wires. terminals SETTING g. Pressure switch The pressure switch is factory preset. To verify current setting perform subsection 3.1.2 B steps 2, 3, 8,9 and 10. If incorrect, turn adjustment screw on face of switch with small screwdriver and retest. If switch will not hold a setting, replace. NOTE The pressure switch is not a polarity sensitive component. Either wire may be connectedto either terminal. h. Cabinet Perform CLOSING procedures. 15. Pressure Regulator 41 (Figure 4-2) REMOVAL Cabinet Perform OPENING procedure. Vacuum check valve Disconnect from pressure regulator con- 38 and tubing 40 nector. c. Oxygen tubing 44 Disconnect from pressure regulator con- nector. d. Pressure regulator 41 Turn counterclockwise at fitting assembly and fitting assembly 39 4-21 39 until reguiator and fitting are removed from product canister 34. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT DISASSEMBLY (Figure 4-9) CLEANING REASSEMBLY Regulator body 10 Regulator body Regulator body Regulator body Bonnet assembly 2 Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-22 ACTION NOTE DISASSEMBLY procedure is provided for regulator cleaning pur- poses only. Internal regulator repair parts are not available. A faulty regulators must be replaced as a complete assembly. Hold firmly in one hand and unscrew bonnet assembly 2. CAUTION To prevent damage to regulator plastic seat 6, remove seat carefully from body with 3/8-inch wide flat blade screwdriver. NOTE During disassembly, removal of fit- ting assembly 11 from regulator body 10 is not necessary. Remove remaining internal parts per Fig- ure 4-9. WARNING To prevent personal injury, use face shield and observe standard safety procedures when using com- pressed-air gun. Apply isopropy! alcohol with cotton swab to cavities. Blow dry with compressed-air gun. CAUTION To prevent damage to regulator plastic seat, install seat carefully into body and hand tighten with 3/8-inch wide blade screwdriver. Replace internal parts per Figure 4-9. Screw onto regulator body. Do not over- tighten. Да TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT INSTALLATION (Figure 4-9) TESTING ADJUSTMENT (Figure 4-9) Pressure regulator Oxygen tubing 44 (Figure 4-2) Vacuum check valve 38 and tube 40 m. AC power cord n. 0. p. q. Г. Front control panel Concentrator flowmeter Test pressure gauge Test pressure gauge Bonnet assembly 2 4-23 ACTION NOTE If a replacement regulator is to be installed, reuse the fitting assembly 11 from the faulty regulator. Apply a small amount of pipe cement to re- placement regulator and fitting as- sembly threads. Distributed the cement around the entire fitting. Do not allow cement to enter interior of regulator. Do not overtighten. Install into the product canister. Use care not to cross thread. Position the regulator as shown in Figure 4-2. Connect to pressure regulator. Connect to pressure regulator. WARNING The following procedure requires AC power be applied to con- centrator. Metal components are not grounded. To prevent possibility of serious electrical shock, connect grounding wire to metal components before applying power to unit. Connect to AC outlet. Push power switch to ON position. Adjust flow to 1 1рт. If installing a new regulator, it is necessary to turn the regulator adjustment knob several turns clockwise to achieve an outlet pressure. NOTE Allow concentrator to run (stabilize) for approximately 20 minutes before proceeding. Connect to oxygen outlet spout 28 or 32 Figure 4-3. Perform subsection 3.1.2 B steps 12 through 15. Verify pressure reading of 5.0 + 0.5 psig. Cut and remove strap 1. Push lock ring 14 up against adjustment knob 15. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT s. Adjustment knob 15 t. Locking ring 14 u. Concentrator v. Cabinet ACTION NOTE To prevent pressure trapping be- tween the pressure regulator and test pressure gauge, make adjust- ments to the regulator in small incre- ments and vent a small amount of oxygen at the pressure gauge be- tween each adjustment. Adjust for test pressure gauge reading of 5.0 + 0.5 psig. Pull down to lock, and install a new strap. Set the power switch to OFF, unplug the AC power cord, and disconnect the test pressure gauge. Perform CLOSING procedure. 16. Solenoid 20, 24 (Figure 4-2) TESTING (Electrical) TESTING (Pneumatic) a. Cabinet b. Solenoid terminals C. Solenoid terminals d. Solenoid terminals e. AC power cord f. Front control panel g. Solenoid valve Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-24 Perform OPENING procedure. Disconnect two terminal wires. Using digital multimeter, verify resistance is approximately 850 ohms. Apply 24 V dc using external power supp- ly. Verify faint clicking sound inside solenoid when 24 V dc is applied. WARNING The following procedure requires AC power be applied to the con- centrator. Metal components are not grounded. To prevent possibility of serious electrical shock, connect ground wire to metal components before applying power to the unit. Connect to AC outlet. Push power switch to ON position. Verify solenoid is correctly activating pilot valve by presence of appropriate supply, exhaust, or balance gas flows. Perform subsection 3.1.2 B steps 4 through 6 for correct cycling. 7 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT REMOVAL INSTALLATION ITEM h. Solenoid valve Front control panel AC power cord Small pilot tubing C, D, orE Solenoid valve Replacement solenoid valve Replacement solenoid valve Replacement solenoid valve terminals Solenoid Valve Cabinet ACTION Leak test solenoid gasket joints and threaded connection with conservative use of liquid leak detector. Push power switch to OFF position Remove from AC outlet. Mark for correct replacement location then remove from solenoid brass connec- tors. Hand loosen and rotate counterclockwise to remove from pilot valve. Apply small amount of pipe cement to threads. Do not allow cement to enter interior of solenoid. Install into pilot valve and hand tighten. Position the solenoid's electrical terminals as shown in Figure 4-2. Connect two terminal wires. NOTE The solenoid valve is not a polarity sensitive component. Either wire may be connectedto either terminal. Leak test threaded connection with con- servative use of liquid leak detector solu- tion. Perform CLOSING procedure. 17. Sieve Canister Assembly 33 (Figure 4-2) REMOVAL a. Cabinet 4-25 NOTE Replacement of sieve canisters is not a normal maintenance proce- dure. Sieve canisters should only be replaced when all other efforts to achieve factory specified oxygen levels have been exhausted. Con- sult your authorized Puritan-Bennett repair facility before proceeding. Perform OPENING procedure Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 COMPONENT TABLE 4-1. SERVICE PROCEDURES (continued) Two straps 17 Two compression nuts 12 (connected to plastic elbows on bottom of sieve canisters) Oxygen a and pilot pressure tubing d Balance pilot valve 19 (including attached fitting and components) Canister insulation 57 Pilot air filter assembly 6 and 1/16-inch pilot tubing 10 Six plastic compression nuts 12, 14 (connected to elbows and tees of four pilot valves 7, 25) Screws 4 and canister clamps 3 Sieve canister 33 Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-26 ACTION CAUTION The following service procedure re- quires removal of fittings or com- ponents which will expose the sieve canister’'s internal sieve material to room air. Care should be taken so that this material is not exposed to room air for long periods of time (i.e overnight). When ever possible plug open ports and fittings to prevent sieve contamination. Do not remove plugs from replacement sieve canisters until ready to install. Do not interrupt the installation process untilcompleted or until all open ports are plugged. Cut and remove. Loosen and disconnect from plastic el- bows. Disconnect from center port of brass tee 21 and brass fitting 11 of solenoid valve 20. Remove from concentrator cabinet and set aside. Remove from interior of concentrator. Remove 1/16-inch pilot air tubing from both brass connectors 11 of solenoids 24. Remove 1/16-inch pilot air tubing both exhaust pilot valves 7. Loosen and disconnect plastic tubing from pilot valves elbows and tee. Remove the two screws and canister clamp for each sieve canister. Remove by pulling top of canister away from cabinet interior and sliding canister toward top of concentrator. r- TABLE 4-1. SERVICE PROCEDURES (continued) MPONENT REPLACEMENT Two brass tees 9 Two plastic elbows 8 Two plastic elbows, and brass tees Replacement sieve canister Two plastic elbows and brass tees 4-27 ACTION NOTE For this procedure it is not neces- sary to separate each sieve canisters solenoid valve, exhaust pilot valve, and supply pilot valve from each other. These components may be removed from each sieve canister as an assembly thus reduc- ing the possibility of creating system leaks. CAUTION When removing solenoid and pilot valves from sieve canister, do not apply force to solenoid body. Remove this assembly as described below. Remove the solenoid, exhaust pilot, and supply pilot valves as one assembly from each sieve canister by turning brass tee with 1/2-inch open-end wrench in counterclockwise direction. Remove elbows connected to bottom of each sieve canister. CAUTION When applying pipe cement, do not allow cement to enter interior of fit- ting. Add a small amount of pipe cement to threads which were removed from the sieve canister of brass tee and plastic elbows. Distribute the cement evenly around the entire threaded area. Remove plugs from threaded connectors. Thread into appropriate port of each sieve canister. Take care not to cross thread or over tighten. Position as shown in Figure 4-2. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM p. Replacement sieve canisters q. Concentrator ACTION Reverse remainder of removal procedure performed above. Hand tighten plastic compression nuts. Refer to Figures 2-1, 2-2, and 4-2, if necessary, for positioning of components and connection of wires and tubes. Perform performance verification proce- dures in subsection 3.1.1 and 3.1.2. 18. Pilot Valve 7,25,19 (Figure 4-2) DISASSEMBLY a. Cabinet b. Solenoid valve 20,24 terminals (Figure 4-10) с. Snapring1 d. Endcap? e. Diaphragm (slotted) 4 f. Snapring1 g. Portcap 9 п. Poppet5 i. Support washer 7 Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-28 Perform OPENING procedure. CAUTION If pilot valve remains disassembled for extended lengths of time, (i.e. longer than necessary to perform service), seal valve openings to prevent contamination of sieve material. If disassembly is to be performed on supp- ly or balance pilot valves, remove terminal wires and pilot air tubing from solenoid valve and remove solenoid from pilot valve. Refer to solenoid valve REMOVAL and INSTALLATION procedures when necessary. Carefully remove from end cap 2 with snap ring pliers (90-degree tip). Remove slowly. Then carefully pull out spring 3 (For exhaust pilot valves only). Peel edge away from valve body, and pull off of poppet 5. Carefully remove from port cap 9. Pull carefully away from valve body. Push poppet, with solid diaphragm 8 at- tached out of pilot valve body 6. Remove from valve body. " TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT CLEANING INSPECTION REASSEMBLY > Y j- Pilot valve body 6 k. Two diaphragms |. Two diaphragms m. Pilot valve body п. Support washer 7 O. Pilot valve body p. Slotted diaphragm 4-29 ACTION WARNING To prevent personal injury, use face shield and observe standard safety procedures when using com- pressed-air gun. Apply isopropyl alcohol with cotton swab to cavities and seating surfaces. Blow dry with compressed-air gun. Apply isopropyl alcohol with soft brush. Blow dry with compressed-air gun. Hold up to light, stretch, and check for cuts, holes, and cracks, especially at seat- ing surfaces. If either is defective, replace both diaphragms with kit 10. Inspect for corrosion, cracks, and burrs. If defective, replace entire component. CAUTION To prevent damage to concentrator, ensure that parts remain clean during reassembly. Place into port cap side of body, with concave end facing inward as shown below. SUPPORT WASHER CD ° (©) Place poppet, with solid diaphragm at- tached, into port cap side, and push in until diaphragm meets support washer. Place into end cap side of body with seat- ing surface facing in. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT ITEM q. Pilot valve body r. Pilot valve body $. Endcap t. Port cap u. Solenoid valves ACTION Hold poppet steady at port cap side and at end cap side. Press seating surface around poppet flange until flange is visible as shown below. — d J J > LA Ty ZF. ml” FLANGE O T Install spring (for only exhaust pilot val- ves) into cavity of poppet on slotted diaphragm side of pilot valve. Install over slotted diaphragm and secure with snap ring. Ensure that snap ring is pressed into valve body groove with ring's sharp edge facing out. Install over solid diaphragm and secure with snap ring. Ensure that snap ring is pressed into valve body groove with ring's sharp edge facing out. Apply a small amount of pipe cement to the threads of the solenoids that were removed. Evenly distribute the cement around the entire solenoid fitting without allowing any cement to enter the interior of solenoid. Reinstall each solenoid and reconnect the appropriate terminal wires. The solenoid is not a polarity sensitive device. Either wire of the appropriate wire pair may be connected to either solenoid terminal. v. Cabinet Perform CLOSING procedures. 19. Restrictor Tube a. Cabinet Perform OPENING procedure. Assemblies 13 (Figure 4-2) Companion 492a/5380 Oxygen Concentrator Service Manual 5-15-91 4-30 CAUTION If restrictor tubes are removed for extended lengths of time, (i.e. longer than necessary to perform service), seal open legs of tees connected to balance pilot valve to prevent con- tamination of sieve material. rr TABLE 4-1. SERVICE PROCEDURES (continued) Oxygen tube a Plastic compression nuts 14 Clamp 22 Tubing inserts 15 plastic compression nuts 14 Tubing inserts, plastic compression nuts, and brass tee Replacement restrictor tube assemblies Concentrator Cabinet ACTION Remove from center leg of brass tee 21. Loosen and remove two plastic compres- sion nuts which are attached to bottom of restrictor tubes, from balance pilot valve’s two plastic tees 16. Loosen and remove each restrictor tube from brass tee 21. Remove two inserts and nuts from restric- tor tubes. Connectto replacement restrictor tube as- semblies. Reinstall into cabinet. Reconnect and hand tighten plastic compression nuts. Reconnect oxygen tubing to center leg of brass tee. Perform pneumatic test procedures listed in subsection 3.1.2 B steps 4 through 7. Refer to subsection 3.4 if necessary. Perform CLOSING procedure. MPONENT INSTALLATION 20. Outlet Gas Filter 13 (Figure 4-3) REMOVAL INSTALLATION Cabinet Oxygen tubing 12, 14 Replacement outlet gas filter Cabinet 4-31 Perform OPENING procedures. Cut flush with inlet and outlet of outlet gas filter. Discard filter. Install between oxygen tubes 12 and 14. If flow arrows are present on replacement filter, point arrow toward oxygen tube which is connected to the flowmeter inlet. Filters without arrows may be installed in either direction. Perform CLOSING procedures. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT 21. Shock Mounts 8 (Figure 4-4) REMOVAL DISASSEMBLY REASSEBMLY REASSEMBLY Compressor platform assembly Sixteen 1/8" rivets 7 Four shock mounts 8 Four replacement shock mounts Rivets Compressor platform assembly ACTION Perform REMOVAL and DISASSEMBLY procedures steps 9a through 9i. Use a 1/8" drill bit to drill out the center of the rivets. Remove and discard. Remove shock mounts from compressor platform and discard. Place shock mounts on compressor plat- form with rivet holes of shock mounts aligned with rivet holes in compressor platform. Secure each shock mount to compressor platform with four 1/8" rivets. Perform REASSEMBLY and INSTALLA- TION procedures steps 9m and 9n. 22. Compressor Cup Seal 8 (Figure 4-6) REMOVAL REASSEMBLY Compressor platform assembly Eight cylinder head screws 31 Cylinder heads and valve plates 3 Piston sleeves 5 Cup seal 8 Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-32 Perform REMOVAL and DISASSEMBLY procedures setup 9a through 9d. NOTE Some compressors may require a #25 torx head screw driver to remove the eight cylinder head screws. Remove from cylinder heads 30 and 10. Remove both along with connector tubes 1. Remove and discard. Remove four retaining plate screws 6 on each piston to separate retainer plate 7, cup seal, and connecting rod 9. Discard cup seals. NOTE An extra connecting rod is required to fit cup seal and retainer plate into piston sieeve. от TABLE 4-1. SERVICE PROCEDURES (continued) COMPONENT COMPONENT ~~ INSTALLATION Y a ITEM f. Piston sleeve Cup seal Retainer plate Retaining plate screw Piston sleeve Cup seal assembly Cup seal assembly Retaining plate screws Valve plates, cylinder heads, connector tubes Cylinder head screws Compressor platform assembly 4-33 ACTION Place over top of extra connecting rod and let drop to the bottom of the rod. Place new cup seal on extra connecting rod top. Make sure that the inside diameter of the cup seal fits over the out- side diameter of the ridge on the top of the connecting rod. Place over the cup seal and connecting rod. Line up screw holes. Install two screws hand tight to secure retainer plate and cup seal to the extra connecting rod. Pull up around the cup seal and the retainer plate to form the cup seal. With piston sleeve in place over the cup seal remove the two screws 6 and remove the complete assembly from the extra connecting rod. Place the piston sieeve, cup seal and retainer plate assembly on the connecting rod 9 in the compressor. Place loctite 242 in the four screw holes of the connecting rod and tighten four screws to twenty inch pounds. NOTE Repeat steps f through m for other piston. Reverse REMOVAL procedures. Install eight screws and tighten to thirty inch pounds in a criss cross pattern. Reverse DISASSEMBLY and REMOVAL procedures 9a through 9d. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-34 (blank) J J (Yuejq 9€-b) GE-t эедигочо) чебАхо 065/826 ЧОШРашох) 6-51-65 ¡ENUEW adas 104813 065/826 uoIUEdulOS :1-p asnfi ITEM a > WE" — O N © 9 10 11 12 13 14 15 16 17 18 19 20 COMPANION 492a/590 EXTERIOR PARTS LIST (Figure 4-1) PART NUMBER 492297 492884 Reference* 492288 492196 492437 Included in item 5 492832 492782 492687 Included in item 5 Included in item 8 492187 492111 492672 492661 492662 492190 492192 492193 492350 DESCRIPTION Battery, 9V Battery Cushion Serial Plate Warning Label Circuit Breaker, 492a, 5 AMP Circuit Breaker, 590, 6 AMP Knurled Nut Battery Connector Assembly Audio Alarm Velcro Strip (5 Required) Hex Nut Locking Ring Hour Meter (120V) 18" Rivet (2 Required) Air Inlet Filter Short Cabinet Screw Long Cabinet Screw Inlet Bacteria Filter Prefilter Housing Inlet Prefilter ( 2 Required) Battery Label "Item shown for descriptive purposes only and is not available for sale as an individual repair part. Companion 492a/590 Oxygen Concentrator 4-37 Service Manual 5-15-91 (HUE1Q Ob-t} 6E-t 1018J]U32U03 uaBAxO 065/e267 LoILEduiOO L6-S1-S [ENUEW aIAJaS зона 19ш9е2 Na :Z-t 840614 ITEM © ©0 NO OV bh WN = — ck - C ok A WwW го Suffix 01 -02 -03 -04 -05 -06 -07 -08 -09 -10 -11 -12 -13 -14 -15 -16 -17 -18 ITEM 14 15 16 17 18 19 20 21 22 23 LEFT CABINET INTERIOR PARTS LIST (Figure 4-2) PART NUMBER 492967 492665 492664 492717 492287" 492760 492634 492875 492666 493176* 492284 Included in item 8 492133 Restrictor Size 6.5 6.6 6.7 6.8 6.9 7.0 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 8.0 8.1 8.2 PART NUMBER Included in item 16 492132 492874 492104 492130 492100 492443 432135 492134 492095 нон DESCRIPTION Cabinet Assembly, Left (Does not include item 53 & 60) Grommet Canister Clamp Screw Brass Fitting Pilot Air Filter Assembly Exhaust Pilot Valve Plastic Eibow w/Insert Brass Tee 146" Tubing Brass Fitting Compression Nut Restrictor Tube Assembly Set (add suffix to part number for correct size) Suffix Restrictor Size -19 8.3 -20 8.4 P C-493930-¢6 -21 8.5 - 905 22 8.6 SA 393 \ -23 8.7 -24 8.8 -25 8.9 -26 9.0 -27 9.1 -28 9.2 -29 9.3 -30 9.4 -31 9.5 -32 9.6 -33 9.7 -34 9.8 -35 9.9 -36 10.0 DESCRIPTION Compression Nut Tubing Insert Plastic Tee w/Insert Plastic Strap, Large 14" |.D. Plastic Tubing Balance Pilot Valve Balance Solenoid Valve Brass Tee Clamp 14" 1.D. Plastic Tubing Companion 492a/590 Oxygen Concentrator 4-41 Service Manual 5-15-91 LEFT CABINET INTERIOR PARTS LIST (Figure 4-2) (continued) ITEM PART NUMBER 24 492442 25 492100 26 492087 27 492890 28 492862 29 492926 30 492890 31 492556 32 493299 33 492990 492991 34 493300 35 492139 36 493177* 37 492633 38 492050 39 492953 40 Included in item 38 41 492621 42 492114 43 492182 44 493174" 45 493178" 46 492873 47 492171 48 492910 49 492172 50 492663 51 492174 52 492309 53 Refer to Figure 4-3 item 7 54 Included in item 60 55 Included in item 60 56 492801 57 Included in item 60 58 Included in item 60 59 Included in item 60 60 492929 * Tubing Sold in 3 foot lengths. ** Tubing Sold in 4 foot lengths. DESCRIPTION Supply Solenoid Valve Supply Pilot Valve 1/4" 1.D. Plastic Tubing Clamp Compressor Exhaust Hose Sleeve Clamp Reservoir Intake Fitting Reservoir Canister Assembly Sieve Canister Assembly, 492a (set of 2) Sieve Canister Assembly, 590 (set of 2) Product Canister Assembly Brass Elbow 1/4" 1.D. Tubing Pressure Switch Vacuum Check Valve Assembly Fitting Assembly Tubing Pressure Regulator Exhaust Muffler Plastic Elbow 316" 1.D. Tubing 34" |.D. Tubing Large Caster Washer 516" Hex Nut Bushing 16" Lockwasher Left Upper Cabinet Tube Caster Small Caster Washer Control Panel Label! (set of two for left and right cabinets) Cabinet Insulation Cabinet Insulation Canister Insulation Cabinet Insulation Cabinet Insulation Cabinet Insulation Left Cabinet Insulation Kit ***For units manufactured August, 1991 and later order P/N 492284. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-42 (yuejq pr-b) Et-t 10)eJJU9ILO USBAXO 06S/2Z6p UOIUEdWON L6-51-5 ¡ENUEW adAIIS ¿OUSlu| ¡UIQ \ubly :E-p 210614 ITEM Y O ua с O © 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 RIGHT CABINET INTERIOR PARTS LIST (Figure 4-3) PART NUMBER 492596 492595 492072 492558 492828 492823 492807 492623 492810 492597 492210 493174** 492071 493174" 492141 493174" 492802 492647 492672 492662 492196 492437 492671 492187 492656 492682 492987 492782 493175*" 492885 492627 492278 492168 492167 492166 492171 492172 492873 492107 492661 DESCRIPTION Flowmeter Knob Flowmeter Valve Assembly Flowmeter, 492a, 4 Ipm Flowmeter, 590, 5 Ipm Green L.E.D. Power Switch (includes item 4) Flowmeter Fitting Label, Control Panel, 492a w/o OCI* Label, Control Panel, 590 w/o OCI* bc, 21334 Flowmeter Bracket Flowmeter Nut Tubing, %16" 1.D. In-line Check Valve Tubing, 316" 1.D. Outlet Gas Filter Tubing, 316" I.D. Filter Door Insulation (Also included in item 68) Filter Compartment Door Air Inlet Filter Long Cabinet Screw Circuit Breaker, 492a, 5 amp Circuit Breaker, 590, 6 amp Wire Grommet Hour Meter, 120VAC Right Upper Cabinet Tube Intake Filter Adapter Small Tube Clamp Audio Alarm Tubing, 38" I.D. Spout Nut Oxygen Outlet Spout, Plastic 14" x 316" Spout Fitting Spout Coupling Spout Washer Oxygen Outlet Spout, Chrome ¥16" Hex Nut 56" Lock Washer Large Caster Washer Cord Clamp Short Cabinet Screw Companion 492a/590 Oxygen Concentrator 4-45 Service Manual 5-15-91 ITEM 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 PART NUMBER 492309 492174 492964 492290 492673 492789 492746 492660 492650 492771 492653 Refer to Figure 4-4 492661 Included in item 53 Included in item 53 492870 Included in item 56 Included in item 56 492840 492839" 492868 Included in item 57 Included in item 58 492800 492774 Included in item 68 Included in item 68 Included in item 68 Included in item 68 Included in item 68 492928 492560 492651 492652 492654 492713 492832 RIGHT CABINET INTERIOR PARTS LIST (Figure 4-3) (Continued) DESCRIPTION Small Caster Washer Caster Cabinet Assembly, Right (Does not include item 7 & 68). O-Ring Cord Wrap Screws Control PCB Screws Control PCB, 120VAC VW /осл A 297) Screw Inlet Air Duct Cover Cord Retainer Cord Wrap Compressor Module Assembly Screw Connector Connector Power Cord Assembly Connector Connector High Voltage Wire Harness Power Switch Wire Harness Low Voltage Wire Harness Connector Connector Air Dam Insulation Air Dam Insulation, Top Middle Insulation, Top Back Insulation, Top Front Insulation, Middle Insulation, Exhaust Tunnel Right Cabinet Insulation Kit Control PCB Fuse (63 MA) Handle Cap Handle Dowel Pins Large Tube Clamp Battery Connector Assembly * If unit is equipped with OCI Refer to Section 6 for appropriate part number. ** Tubing sold in 4 foot lengths. - Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-46 ITEM PART NUMBER 49296] 1 Refer to Figure 4-5 or 4-6 2 492043 3 493175* 4 492108 5 492770 6 492044 8 49 ETC 9 492965 10 492078 11 492079 12 493289 13 492799 14 492789 15 492107 16 492671 17 492989 18 492104 19 Included in item 1 20 Included in item 1 21 492773 22 4921093 23 Included in item 1 24 Included in item 1 25 Included in item 1 26 Included in item 1 27 492942 28 492946 29 492944 30 492966 31 Included in item 31 32 Included in item 31 33 492943 34 49204? 35 492822 * Tubing sold in four foot lengths. in this figure. PER Se Feur 49393 1 (A sera 493930 =) LouEf COMPRESSOR PLATFORM PARTS LIST DESCRIPTION Compressor Assembly Intake Fitting Tube, 36" 1.D. Fan Screw Tyrap Mount Plastic Strap, Small 15" Rivet Shock Mount Platform Assembly (includes items 7 & 8) Cup Washer Lock Washer Compressor Mounting Bolt Insulation Screw Clamp Wire Grommet Capacitor, 120 VAC Plastic Strap, large Connector Wire Harness Clamp Cooling Fan, 120 VAC Connector Wire Harness Connector & Wire Harness Grommet Clamp Flat Washer Locknul Heat Exchanger Assembly (includes items 31 & 32) Brass Nut Ferrule Allenhead Screw Exhaust Fitting Assembly (includes items 31 & 32) Clamp [> An additional set of items 27, 28, 29 and 33 are installed on the rear of the compressor bul are го! = Figure 4-4: Compressor Platform Assembly Companion 492a/590 Oxygen Concentr: 4-47 (4-48 blank) Service Manual 5-1£ O HOUSING WELL B HOUSING WELL A Figure 4-5: Companion 492a Compressor Companion 492a/590 Oxygen Concentrator 4-49 Service Manual 5-15-91 492a COMPRESSOR PARTS LIST (Figure 4-5) ITEM PART NUMBER DESCRIPTION — 492956 492a Compressor Assembly, 120VAC, NEW (includes items 1 through 26) — 492957 492a Compressor Assembly, 120VAC, REMANUFACTURED (includes items 1 thorugh 26) 1 492027 Cylinder Head Screw 2 492028 Connector Tubes (2) with O-Rings (4) 3 492031 Head Gasket 4 492032 Valve Plate 5 492033 Valve Plate O-Ring 6 Included in item 21 Sleeve 7 Included in item 21 Piston/Rod Assembiy 8 492036 Connecting Rod Screw 9 492030 Cylinder Head B 10 492592 Cylinder Head Plug 11 492038 Five-Blade Fan 12 Reference” Cooling Fan Wires/Connector 13 Reference” Compressor Wires/Connectors 14 Reference* Capacitor Wires/Connectors 15 Reference* Motor Shaft 16 Included in item 21 Bearing 17 492039 Set Screw 18 Included in item 21 Eccentric 19 492040 Four-Blade Fan 20 492041 Eccentric Bearing Assembly (includes items 16, 17 & 18) 21 492954 Piston & Sleeve Assembly (includes items 6 & 7) 22 492465 Reed Valve 23 492963 Reed Valve Restraint 24 492563 Reed Valve Keeper Strip 25 492962 Reed Valve Plate Screw 26 492029 Cylinder Head A * Item shown for descriptive purposes only and is not available for sale as an individual repair part. Items 22, 24 and 25 are also used on the underside of each valve plate, item 4. The reed valve restraint, item 23 is not used in this location. Companion 492a/530 Oxygen Concentrator Service Manual 5-15-91 4-50 HOUSING WELL @ (2) (20) © Figure 4-6: Companion 590 Compressor A Companion 492a/590 Oxygen Concentrator 4-51 Service Manual 5-15-91 ITEM (соч ол во о NY N ГО ГО № ГО ГО NY NOY = À eh À AP eh eh mh ed aA O ©0 < OO Oh bb WN = OO © 00 < С) Oh bh Ww ГО — © PART NUMBER 492959 492961 492482 492481 492480 492479 492478 492477 492476 492475 493125** 492471 492592 492038 Reference* Reference" Reference" 492036 492039 493340"* 492470 Included in item 18 Included in item 18 Reference" 492465 492963 492563 492962 492592 492472 492027 590 COMPRESSOR PARTS LIST (Figure 4-6) DESCRIPTION 590 Compressor Assembly, 120 VAC, NEW (includes items 1 through 29) 590 Compressor Assembly, 120 VAC, REMANUFACTURED (includes items 1 through 29) Connector Tubes (2) and O-Rings (4) Head Gasket Valve Plate Valve Plate O-Ring Sleeve Retaining Plate Screw x Retaining Plate <= Cup Seal —> (5 rst C-2942"10 -CO 35 Connecting Rod Cylinder Head B Cylinder Head Plug Fan B Cooling Fan Wires/Connectors Compressor Wire Harness Capacitor Wires/Connectors Connecting Rod Screw Set Screw Eccentric Bearing Assembly (includes items 17, 20 & 21) Fan À Eccentric Bearing Motor Shaft Reed Valve Reed Valve Restraint Reed Valve Keeper Strip Reed Valve Screw Cylinder Head Plug Cylinder Head A Cylinder Head Screw * Item shown for descriptive purposes only and is not available for sale as an individual repair part. ** For concentrators whose last 6 digits are 009055 or lower order part number 492474 for item 9 and item 18 should be part number 492473. Items 23, 25 and 26 are also used on the underside of each valve plate item 3. The reed valve restraint, item 24 is not used in this location. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-52 Figure 4-7: Compressor Cylinder-Head Screw Torque Sequence Companion 492a/590 Oxygen Concentrator 4-53 Service Manual 5-15-91 Figure 4-8: Companion 530 Compressor Retaining Plate Indexing Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-54 PRESSURE REGULATOR PARTS LIST ITEM PART NUMBER 492621 492044 Reference" Reference” Reference” Reference* Reference” Reference” Reference* Reference” Reference” 492953 Included in item 13 492050 Reference" Reference" oO — ©) ло А © го — dl = A À CO RR © Го — © * DESCRIPTION Pressure Regulator, Complete (items 2 through 10) Plastic Strap Bonnet Assembly Spring Plastic Washer Diaphragm Plastic Seat Black Seal Poppet Spring Regulator Body Fitting Assembly Tube Vacuum Check Valve Assembly Lock Ring Adjustment Knob Item shown for descriptive purposes only and is not availalbe for sale as an individual repair part. Figure 4-9: Pressure Regulator Companion 492a/590 Oxygen Concentrator 4-55 Service Manual 5-15-91 ITEM = DO OO NO Сл po WN = * PART NUMBER 492634 492100 492052 Reference* 492935 Included in item 10 492936 Reference” 492055 Included in item 10 Reference* 492952 PILOT VALVE PARTS LIST DESCRIPTION Exhaust Pilot Valve Supply or Balance Pilot Valve Snap Ring End Cap Spring (used only in exhaust pilot valve) Diaphragm (slotted) Poppet Valve Body Support Washer Diaphragm (solid) Port Cap Pilot Valve Diaphragm Kit (items 4 & 8) ltem shown for descriptive purposes only and is not available for sale as an individual repair part. Figure 4-10: Pilot Valve Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 4-56 SECTION 6. OXYGEN CONCENTRATION INDICATOR (OCI™) This section provides information required to test, troubleshoot, and repair the optional Oxygen Concentration Indicator, including additional information in the form of a general product description and theory of operation. Companion 492a/590 Oxygen Concentrator 6-1 Service Manual 5-15-91 6.1 General Product Description The Companion 492a and 590 oxygen concentrators may be equipped with an OCI which may be either factory or field installed. For field installation, a conversion Kit including step by step instructions is available (492888). Changes to the concentrator with OCI installed consists of a different front control panel and addition of an internal circuit board ( Figures 6-1 and 6-2 ). The Oxygen Concentration Indicator (OCI) is a real-time oxygen sampling device utilizing ultrasonic sound wave technology. An internal sensor is located in-line between the concentrator’s product canister and flowmeter to analyze the product gas being delivered to the patient. The OCI system will automatically activate after five minutes of concentrator operation. Upon activation, one of three front panel L.E.D.s will illuminate indicating that the percentage of oxygen in the concentrator's output gas is within the ranges shown below. INDICATORS RANGE (+ 3%) Green Above 85% Y ellow Between 70 and 85% Red Below 70% If the oxygen concentration drops to below 70%, and the red indicator activates, the OCI circuitry will interrupt power to the concentrator's compressor. A low pressure alarm will activate after the gas in the product canister is depleted. NOTE Concentrators with OCI installed require a minimum of 1/4 lpm flow for the system to function properly. A \ © a LT Figure 6-1: 492a Front Panel with OCI Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 6-2 fa ны РО ; OCI PRINTED CIRCUIT BOARD (PCB) O.C.I. FRONT PANEL PRINTED CIRCUIT BOARD (PCB) OXYGEN ——— TUBES GAS SAMPLE CHAMBER Figure 6-2: OCI Circuit Board Installed 6.2 Theory Of Operation The OCI makes use of the well known principle that different gases have different characteristics in their abilities to propagate, or transmit, ultrasonic sound waves. One of the systems components is a gas sample chamber that samples the concentrator gas in a continuous flow. A sound wave is transmitted over a fixed distance from one end of the gas sample chamber to the other end of the chamber. This "time-to-fly" is electronically measured and converted to an electronic signal. The "time-to-fly" is dependent upon the concentrator gas purity, the gas temperature, and the pressure of the concentrator gas. A temperature sensor is located within the concentrator gas stream in order to electronically compensate for temperature effects. The concentrator gas in the sample chamber is maintained at a constant pressure level. The "time-to-fly" is measured and converted to an electronic signal, external factors are compensated for, and the resultant output is processed into useful analytical information (the LEDs and shutdown system) through standard electronic circuitry. Companion 492a/590 Oxygen Concentrator 6-3 Service Manual 5-15-91 6.3 Performance Verification Before attempting to verify the performance of the OCI, it is essential that the Companion Concentrator meets all performance specifications found in Section 1: Introduction and Section 3: Performance Verification and Troubleshooting. If during this process, the OCI is suspect or prevents further concentrator verification (i.e. unit shuts oft) open the cabinet, and with the concentrator unplugged, disconnect the OCI by disconnecting the wire harness header, item 12 from the OCI board at J5 (Figure 6-3). This will disarm the OCI and control panel indicators but will not affect the concentrator's operation. a. Allow the concentrator to stabilize by running a minimum of 20 minutes at the maximum flow rate as indicated on the flowmeter. b. Attach a properly calibrated oxygen analyzer to the outlet spout (Figure 1-4). Verify the concentrators’ performance is within specification (see 1.4.1, step i). c. Verify the green OCI front panel indicator labeled NORMAL is illuminated five minutes after concentrator start-up. d. Rotate the flowmeter knob counter-clockwise two turns for 492a or 1 1/4 turns for 590 to increase the flow and decrease oxygen concentration. e. Observe the oxygen analyzer. Within several minutes the concentration will drop. When the analyzer reads approximately 85%, the yellow OCI front panel indicator should illuminate. f. Allow the concentration to fall until the unit shuts off and the red OCI front panel indicator is illuminated. The analyzer readout should be approximately 70%. NOTE The oxygen analyzer is located several feet downstream from the concentrator. Therefore, the oxygen analyzer is measuring the product gas at a different location and point in time than the OCI. This fact combined with the instantaneous measurement time of the OCI as compared to the relatively long measurement time of common analyzers will cause an apparent disagreement between the OCI and the analyzer indicators. g. Within a few minutes the audio alarm should sound. Depress the power switch to OFF. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 6-4 6.4 Troubleshooting and Repair The OCI installed on a Companion concentrator is a gas monitoring device and does not affect the performance of the concentrator. Should operation of the OCI be suspect, refer to the following sections. 6.4.1 Troubleshooting This section is designed for easy reference to determine the problem with a concentrator equipped with OCI. Refer to the category that matches the symptom(s) you observe. Then refer to the probable cause column which suggest problems in order of most likely to least likely to occur. SYMPTOM 1. Concentrator runs 5 minutes then shuts off PROBABLE CAUSE CORRECTIVE ACTION d. No flow Flowmeter set too high Oxygen below 70% High outlet pressure Defective OCI Set flow at minimum of 1/4 Ipm. Set flowmeter ball at or below maximum setting on flowmeter scale. Troubleshoot concentrator per Section 3. Check outlet pressure per 3.1.2 steps 12 through 15. Reset if necessary per ap- propriate service procedure in section 4. Disconnect wire harness from OCI PCB con- nector J5. Run unit for 5 minutes at maximum flow rate. If Oxygen is above 70% replace OCI board per repair Section 6.4.2. 2. Concentrator runs/ concentration above 85% / no green OCI LED. Concentrator has not run for 5 minutes Loose connection High outlet pressure Detective OCI Run unit for more than 5 minutes. Verify secure wire harness connections be- tween OC! PCB at J5 and front panel PCB at J4. Check outlet pressure per 3.1.2 steps 12 through 15. Reset if necessary per ap- propriate service procedure in section 4. Follow steps a through g, subsection 6.3. If yellow indicator illuminates at step e and the red indicator illuminates (concentrator shuts down) at step f or if no indicators illuminate but concentrator shuts off below 70%, then the front panel PCB may be defective. Replace the front panel PCB as described in repair subsection 6.4.2. If no OCI lights func- tion, and unit does not shut off below 70%, then the OCI PCB is most likely defective. Replace the OCI PCB as described in repair subsection 6.4.2. 3. Concentrator runs/ concentration range does not agree with ОС! indicators (See subsection 6.1) Incorrect outlet pressure Defective OCI PCB board Check outlet pressure per 3.1.2 steps 12 through 15. Reset if necessary per ap- propriate service procedure in section 4. Replace as described in repair subsection 6.4.2. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 6.4.2 Repair The following section describes how to service the individual components of the Oxygen Concentration Indicator (OCI) system. Included are instructions for removal and installation for each major component of the OCi system. After removing a component, visually inspect for damage or any other indication that the component is defective. Unless otherwise specified, replace as needed with a new component. Refer to Figure 6-3 for replacement part numbers. Perform subsection 6.3, Performance Verification, after completing service procedures found in Table 6-1. Service procedures for the OCI system are provided in Table 6-1 as follows. Step Number Component 1 OCI Printed Circuit Board 2 OCI Front Panel Printed Circuit Board TABLE 6-1 SERVICE PROCEDURES (Numbers in bold print correspond to items shown in Figure 6-3) COMPONENT ITEM ACTION 1. OCI Printed Circuit Board (PCB) 16 (Figure 6-3) CAUTION The OC! printed circuit board con- tains complementary metal-oxide semiconductor (CMOS) integrated circuits (IC's) which are static-sensi- tive devices. To prevent IC damage, observe standard safety procedures as follows: « Wear grounding wrist strap. . Work on grounded conduc- tive mat. « Handle PCB by edges only. « Store PCB in conductive plastic bag. REMOVAL a. Cabinet Perform OPENING procedure located in Section 4. b. Harness connector 12 Disconnect from OCI PCB at J5. с. Oxygen tubes Cut and carefully remove the two oxygen tubes where they connect to the gas sam- pling chamber on the OCI PCB (Figure 6-2). d. Screw 15 Remove six screws which attach OCI PCB to air inlet duct cover 18. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 6-6 rr COMPONENT INSTALLATION TABLE 6-1 SERVICE PROCEDURES (continued) ITEM Replacement OCI PCB Cabinet ACTION NOTE The wire hamess connectors are keyed and may only be installed in one direction. Match connector con- figuration to OCI PCB pins before attempting to reinstall the connector. Reverse REMOVAL procedure. Install and tighten six screws in a criss-cross pattern. Perform CLOSING procedures located in Section 4. 2. OCI Front Panel Printed Circuit Board (PCB) 8 REMOVAL INSTALLATION Cabinet Oxygen tubes (not shown) Two flowmeter brack- ets 5 and nuts 6 Flowmeter Two connectors 10,11 Four nylon nuts 9 OCI front panel PCB Replacement OCI front panel PCB Cabinet Perform OPENING procedure located in Section 4. Remove from flowmeter 1 inlet and outlet connectors. Remove from flowmeter threaded con- nectors. Remove from concentrator. Remove from OCI front panel PCB at J4 and J3. Remove using a 5/16-inch socket wrench. Remove from concentrator cabinet inte- rior. NOTE The wire harness connectors 10 and 11 are keyed and may only be in- stalled in one direction. Verify cor- rect location, and match connector configuration to front panel PCB pins before attempting to reinstall the connector. Reverse REMOVAL procedure. After hand tightening four nylon nuts, turn each nut no more than 3/4 of a turn. Verify correct connection of oxygen tubing to flowmeter. Refer to Pneumatic Diagram Figure 2-1 if necessary. Perform CLOSING procedure located in Section 4. Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 Figure 6-3: O.C.I. System Companion 492a/590 Oxygen Concentrator Service Manual 5-15-91 6-8 OXYGEN CONCENTRATION INDICATOR (OCI) SYSTEM PARTS LIST ITEM PART NUMBER DESCRIPTION 1 432072 Flowmeter (432a) 492558 Flowmeter (590) 2 492624 OCI Control Panel Label (4924) 492809 OCI Control Panel Label (590) 3 492438 Nylon Screw 4 492964 Right Concentrator Cabinet 5 492597 Flowmeter Bracket 6 492210 Flowmeter Nut 7 432667 Nylon Spacer 8 492745 Front Panel PCB 9 492439 Nylon Nut 10 Reference’ Connector (15 pin) 11 Reference" Connector (14 pin) 12 Reference" Connector (14 pin) 13 492939 Clamp 14 492741 Screw 15 492789 Screw 16 492744 OCI PCB 17 492746 Control PCB 18 492650 Air Inlet Duct Cover 19 Reference" Connector (11 pin) 20 Reference* Connector (15 pin) 21 492837 Wire Harness (Control PCB to Front Panel PCB) 22 492838 Wire Harness (Front Panel PCB to OCI PCB) 23 492946 Washer J1 Reference’ Control PCB connector (connects to front panel PCB J3) J2 Reference” Control PCB connector J3 Reference” Front panel PCB connector (connects to control PCB J1) J4 Reference” Front panel PCB connector (connects to OCI PCB J5) J5 Reference* OCI PCB connector (connects to front panel PCB J4) J6 Reference* Control PCB connector * Item shown for descriptive purposes only and is not available for sale as an individual repair part. Companion 492a/590 Oxygen Concentrator me “== 6-9 Service Manual 5-15-91 - wh Companion 492a/590 Oxygen Concentrator . ~~Service Manual 5-15-91 6-10 (blank)
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement