Sartorius Master
pro
Series
LP Models
Electronic Precision Balances
Operating Instructions
1
Intended Use
The Master pro
Series from Sartorius offers precision balances with capacities ranging from 1 to 64 kg for measurement of mass. A broad range of special performance features make the Master pro
balances ideal for use as measuring and test equipment in ISO or GLP quality management systems. These features include:
– Fully automatic self-calibrating and adjustment function, isoCAL
(time- and temperature-dependent)
– reproTEST for quick determination of the standard deviation to check the reproducibility of results
– ISO/GLP-compliant recording capability for printouts
– Password-protected menu lock
Master pro
balances meet the highest requirements on the accuracy and reliability of weighing results through the following features:
– Efficient filtering-out of vibration
– Stable and reproducible results
– Excellent readability under any lighting conditions
– Rugged, durable weighing system
Master pro
balances save work and speed up simple routine applications through:
– Ultrafast response times
– Built-in applications
(counting, animal weighing, weighing in percent, etc.)
– Automatic initialization when you switch on the balance
– Easy input of alphanumeric sample, lot and balance IDs
– Flexible, easy-to-use display and control unit
– Connectivity for control through an on-line computer
Contents
Intended Use .............................2
Contents .................................. 2
Warnings and
Safety Precautions ..................... 2
Operating Design ...................... 3
Getting Started ......................... 5
Configuring the Balance
Setting the Language ............. 10
’Info’ Display ........................ 10
Entering User Data (Input) ....... 11
Setting Parameters (Menu) ...... 14
Operating the Balance
Basic Weighing Function ........ 21
Calibration/Adjustment .......... 24
ReproTEST ........................... 30
Application Programs
Toggle between
Weight Units ........................ 31
Counting ............................. 33
Weighing in Percent .............. 36
Net-Total Formulation ............. 40
Animal Weighing .................. 42
Checkweighing .................... 46
Recalculation ........................ 51
Data Output Functions ............ 56
Pin Assignment Chart ............. 68
Cabling Diagram .................. 69
Additional Functions .............. 70
MP8 Interface Emulation ......... 72
Error Codes ............................ 73
Care and Maintenance............. 75
Instructions for Recycling .......... 76
Overview
General Views of the Balances ..................... 77
Description of the Keys ........... 80
Menu Structure ..................... 81
Specifications ....................... 82
Accessories (Options)............. 88
Declaration of Conformity ....... 90
EC Type-Approval Certificates .. 93
Plates and Markings .............. 96
Index .................................. 97
Appendix
Entering the User Password
Warnings and
Safety Precautions
This balance has been constructed in accordance with the European
Directives as well as international regulations and standards for operation of electrical equipment, electromagnetic compatibility, and stipulated safety requirements.
Improper use or handling, however, can result in damage and/or injury.
Read these operating instructions thoroughly before using your balance to prevent damage to the equipment. Keep these instructions in a safe place.
Follow the instructions below to ensure safe and trouble-free operation of your balance:
!
Do not use this balance in a hazardous area/location
!
Make sure that the voltage rating printed on the AC adapter is identical to your local line voltage
– The only way to switch the power off completely is to disconnect the
AC adapter
– Type of protection for the housing:
– LP64001S, LP34001S,
LP16001S, LP34001P and
LP34000 meet IP44 requirements
– Additional balances with a readabilty
≥
1 mg comply with IP54
– AC adapters meet IP20 requirements
– Protect the AC adapter from contact with liquid
– Connect only Sartorius accessories and options, as these are optimally designed for use with your Master pro
balance
When cleaning your balance, make sure that no liquid enters the balance housing; use only a slightly moistened cloth to clean the balance.
Do not open the balance housing.
If the seal is broken, this will result in forfeiture of all claims under the manufacturer’s warranty.
In case you have any problems with your balance: k contact your local Sartorius office, dealer or service center
2
Operating Design
The balances in the Master pro
Series consist of a weighing cell and a display and control unit.
In addition to the choice of power supply (via AC adapter or external rechargeable battery pack), your balance also has an interface port for connecting a printer, computer or universal remote control switch.
The display and control unit and the weighing cell can be set up separately. Operation of Master pro balances follows a uniform
“philosophy” which is described in this manual.
Keys
The functions used most often are assigned to their own specific keys. There are additional keys for assignment of other (multiple) functions, in some cases dependent on the current operating status (so-called “soft keys”).
Each key is described in detail in the chapter entitled “Overview.”
Display
The display is divided into six sections:
Normal Operation
In the operating mode, these four keys function as c, q, w and v keys.
Setup Mode
In the setup mode, these keys take on the function of arrow keys
(x y Y X).
Soft Key Mode
The ‘soft key’ functions depend on the current operating status; the current function is indicated in the text line of the display. In this example, the soft keys are used to access Info, Menu and Input functions.
info--menu--input
– Line for metrological data:
Metrological specifications of the weighing platform.
Max ...g
d=...g
Display during use as a legal measuring instrument:
Max ...g Min ...g e=...g d=...g
– Bar graph: Display in percent of the weight on the balance relative to the maximum capacity and for over/under checkweighing
– Measured value line:
Display of weight readout or alphanumeric input
Important Note Concerning
Verified Balances Approved for
Use as Legal Measuring
Instruments in the EU*:
For verified balances that have a verification scale interval “e” which is greater than the scale interval “d,” the last digit on the display is bordered.
– Weight unit display:
Weight unit, other unit of measure, operating information
The G symbol indicates nonverified weight values.
– Symbol display:
Indicates operating status; application selected
– Text line: Prompts for operator guidance, soft key designations
See the “Data Output” section in the chapter entitled “Operating the Balance” for a detailed description of the information displayed in each of these sections.
* including the Signatories of the Agreement on the European
Economic Area
3
Input
Numeric Input
To enter numbers:
Press the 1 2 … 0 . keys
To store numbers entered:
Press the soft key
To interrupt/cancel numeric input:
Press c
Additional information is displayed in the text line. See the chapter entitled “Configuring the Balance” for a detailed description of all parameter settings.
To save setting and exit menu:
Press M
To interrupt the parameter setting process without saving changes:
Press e
Alphabetic Input
To enter letters:
First press the a key, then press a letter soft key (Y or y) repeatedly until the desired letter or special character is displayed in the text line
To store a word entered:
Press the soft key
To interrupt/cancel alphabetic input:
Press c
Configuring Balance Operating
Parameters
To set parameter options for configuring your balance, you can select the desired parameters from a list. The lists of parameter options comprise a menu, which has three levels.
For configuration functions:
Press the M key; then press the menu
soft key
To move within a menu level:
Press Y or y
To change to another menu level:
Press x or X
To confirm the selected parameter setting in the 3rd menu level:
Press X
“o” indicates the currently set parameter
Data Output
Your Master pro
balance is equipped with a data interface for connecting your choice of the following:
– Printer
– Peripheral device (e.g., computer)
– Universal remote control switch
Printer
You can configure the print functions to meet your individual requirements by selecting the corresponding menu code.
You can have printouts generated automatically, or by pressing p; dependent on or independent of the stability or time parameters; with or without IDs; and as standard or
ISO/GLP-compliant printouts.
ISO: International Organization for Standardization
GLP: Good Laboratory Practice
See the section on “Data Output
Functions” in the chapter entitled
“Operating the Balance” for a detailed description of data output options.
Interface Port
Instead of a printer, you may choose to connect a different peripheral device, e.g. a computer (PC). With an on-line PC you can control both the weighing cell and the display unit of the Master pro
balance.
Request messages are sent via the interface to initiate functions in the weighing cell and in the display unit. Some of the functions generate response messages.
See the section on “Data Output
Functions” in the chapter entitled “Operating the Balance” for a detailed description of the interface port.
Error Codes
If you press a key that has no function, or which is blocked at a certain point in an application program, this error is indicated as follows:
– a double-beep is sounded as an acoustic signal, and
– where necessary, a message is displayed for 2 seconds in the text line, after which the text line returns to the previous display.
The response to an operator error is identical in all models of the
Master pro
series. See the chapter entitled “Error Codes” for a detailed description.
Storing Settings
Storing Parameter Settings
The settings configured are stored in the balance’s non-volatile memory. The most recent parameter settings are active when you switch on the balance.
Saving Parameter Settings
You can assign passwords in order to block access to the
“Menu” and “Input” functions.
4
Getting Started
Getting Started
Warranty
Do not miss out on the benefits of our full warranty. Complete the warranty registration card, indicating the date of installation, and return the card to your Sartorius office or dealer.
Storage and Shipping
Conditions
Allowable storage temperature:
0
°
C …+40
°
C (+32
°
F …+104
°
F)
The packaging has been designed to ensure that the balance will not be damaged even if it is dropped from a height of 80 centimeters (about
31 inches). Do not expose the balance to extreme temperatures, blows, shocks, vibration or moisture.
Unpacking the Balance
● After unpacking the balance, check it immediately for any visible damage as a result of rough handling during shipment.
k If this is the case, proceed as directed in the section on
“Safety Inspection” in the chapter entitled “Care and Maintenance.”
It is a good idea to save the box and all parts of the packaging until you have successfully installed your balance. Only the original packaging provides the best protection for shipment. Before packing your balance, unplug all connected cables to prevent damage. The cardboard strips between the display and control unit and the weighing platform are part of the protective packaging for shipment!
Important Note Concerning
Verified Balances Approved for
Use as Legal Measuring Instruments in the EU*:
Provided that an official seal is required for the verified balance, a control seal is affixed to the balance. This seal will be irreparably damaged if you attempt to remove it. If the seal is broken, the validity of the verification will become void, and you must have your balance re-verified.
* including the Signatories of the Agreement on the European
Economic Area
Equipment Supplied
The equipment supplied includes the components listed below:
LP balances with a readability of 1 mg
– Balance with display and control unit
– AC adapter
– Dust cover
– Shield disk
– Pan support
– Weighing pan
– Glass draft shield cylinder
– Draft shield cover
LP8200S, LP8200P, LP 6200S,
LP 4200S, LP 2200S, LP 820, LP 420,
LP 2200P, LP 5200P
– Balance with display and control unit
– AC adapter
– Dust cover
– Pan draft shield
– Weighing pan
LP 12000S, LP 6200,
LP 4200, LP 2200, LP 12000P
– Balance with display and control unit
– AC adapter
– Dust cover
– Weighing pan
LP balances with a capacity
≥
16 kg
– Balance with display and control unit
– AC adapter
– Weighing pan
Installation Instructions
The Sartorius Master pro
balances are designed to provide reliable weighing results under normal ambient conditions in the laboratory and in industry. When choosing a location to set up your balance, observe the following so that you will be able to work with added speed and accuracy:
– Set up the balance on a stable, even surface
– Avoid placing the balance in close proximity to a heater or otherwise exposing the balance to heat or direct sunlight
– Protect the balance from drafts that come from open windows or doors
– Avoid exposing the balance to extreme vibrations during weighing
– Protect the balance from aggressive chemical vapors
– Do not expose the balance to extreme moisture
Conditioning the Balance
Moisture in the air can condense on the surfaces of a cold balance whenever it is brought into a substantially warmer place. If you transfer the balance to a warmer area, make sure to condition it for about 2 hours at room temperature, leaving it unplugged from AC power.
Afterwards, if you keep the balance connected to AC power, the continuous positive difference in temperature between the inside of the balance and the outside will practically rule out the effects of moisture condensation.
5
6
Setting Up the Balance
Preparing Balances with a Round Glass Draft Shield
● Place the components listed below on the balance in the order given:
– Dust cover
– Shield disk; turn counter-clockwise until it stops and is secured
– Pan support
– Weighing pan
– Glass draft shield cylinder
– Draft shield cover
Preparing Balances with a Rectangular Weighing Pan and a Weighing Capacity
≤
12 kg
● Place the components listed below on the balance in the order given:
– Dust cover
– Pan draft shield (depending on the model)
– Weighing pan
Preparing Balances with a Rectangular Weighing Pan and a Weighing Capacity
≥
16 kg
● Place the weighing pan on the balance
Getting Started
Separate Operation of the Display Unit
● Turn the balance upside down and lay it on a padded surface to avoid damage to the weighing system.
● Use a screwdriver to remove the 2 screws from the display unit retainer
● Remove the display unit
> Cable lengths
– LP balances with a weighing capacity
≤
12 kg: 55 cm
– LP balances with a weighing capacity
≥
16 kg: 80 cm k See the chapter entitled “Accessories” for information on longer cables k If you wish to use a longer cable, it must be installed by an authorized
Sartorius service technician
Options for Mounting the Display Unit for a Balance with a Weighing Capacity
≥
16 kg
The display unit can be mounted as follows:
– on the short side of the weighing cell (factory mounting)
– on the back (long side) of the of the weighing cell
● Turn the weighing cell over
● Remove the fastening screws from the display unit retainer
● Remove the fastening screws from the plate that covers the cable raceway
(channel) and remove the plate
● Fasten the display unit retainer onto the back of the weighing cell with the 2 Allen screws
● Thread the cable through the raceway and replace the cover plate
7
8
Connecting the Balance to AC Power
● Check the voltage rating and the plug design
– If they do not match the rating or standard you use, contact your Sartorius office or dealer
Use only
– Original Sartorius AC adapters
– AC adapters with a registered approval rating from a national testing laboratory k To use a main feeder cable from the ceiling or to mount a CEE plug, you will have to make arrangements inside your facilities to have this equipment installed k See the “Accessories” for information on using an IP65-protected industrial AC adapter or an external rechargeable battery pack with your balance
● Insert the right-angle plug into the jack and tighten the screws
● Then plug the AC adapter into a wall outlet (mains)
Safety Precautions
The AC adapter rated to Class 2 can be plugged into any wall outlet without requiring any additional safety precautions. The ground or earth terminal is connected to the balance housing, which can be additionally grounded, if required. The data interface is also electrically connected to the balance housing (ground).
Information on Radio Frequency Interference
Warning!
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference, when operated in a commercial environment. Operation of this equipment in a residential area is likely to cause interference, in which case the user, at his own expense, will be required to take whatever measures may be required to correct the interference.
Connecting Electronic Peripheral Devices
● Make absolutely sure to unplug the balance from AC power before you connect or disconnect a peripheral device (printer or PC) to or from the interface port.
Warmup Time
To deliver exact results, the balance must warm up for at least 30 minutes after initial connection to AC power or after a relatively long power outage.
Only after this time will the balance have reached the required operating temperature.
Using Verified Balances as Legal Measuring Instruments in the EU*
The balance must warm up for at least 24 hours after initial connection to AC power.
* including the Signatories of the Agreement on the European Economic Area
Getting Started
Fastening an Antitheft Locking Device:
Balances with a Weighing Capacity of 12 kg
To fasten an antitheft locking device, use the lug located on the rear panel of the balance.
● Secure the balance at the place of installation, e.g., with a chain or a lock.
Leveling the Balance
Purpose:
– To compensate for unevenness at the place of installation
– To achieve perfectly horizontal positioning of the balance for consistent reproducibility
Always level the balance again any time after it has been moved.
Leveling Balances with a Weighing Capacity
≤
12 kg
Only the 2 front feet are used for leveling.
● Retract the 2 rear feet
● Turn the 2 front feet as shown in the diagram until the air bubble is centered within the circle of the level indicator
> Several leveling steps are usually required.
●
For weighing heavy samples:
Extend the 2 rear feet until they touch the surface on which the balance rests
Leveling Balances with a Weighing Capacity
≥
16 kg
● Adjust the four leveling feet until the air bubble is centered within the circle of the level indicator
9
Configuring the Balance
Purpose
You can configure your Master pro balance to meet individual requirements by entering user data and setting parameters in the Setup menu. You can also configure the display to show balance-specific information (such as the serial no.).
Setting the Language
Available Features
You can choose from 5 languages for the information display:
1 German
2 English (factory setting)
3 English with U.S. date/time format
4 French
5 Italian
6 Spanish
Setting the Function Switch for Using the Balance in Legal Metrology
To use the balance in legal metrology, the following functions must be activated by setting the switch as described below:
– Display: Verification scale interval e; lower weighing range limit: Min
– External calibration/adjustment:
Blocked
– MP8 interface emulation active
Preparation
● Remove the covering plate from the back of the balance housing
● Move switch 1 in the direction of the arrow
Selecting the Language
● Enter the corresponding number
● Press M
> Switch up: external calibration blocked
Switch down: external calibration accessible
> Note:
Do not move Switch 2
’Info’ Display
Purpose
To have information about the equipment displayed.
Features
You can have the following information displayed:
– Program version number for the display and control unit
– Program version for the weighing cell
– Balance model
– Serial number of the weighing cell
Display Balance Information
● Select the Setup menu:
Press M
● Select information:
Press the info soft key q
> Readout in measured value line:
Version number of the display and control unit (see also “Data Output
Functions,” pages 56–57, in the chapter entitled “Operating the
Balance”) k Select next item of information:
Press Y
> Readout in measured value line:
Next information k Select previous information:
Press y
> Readout in measured value line:
Previous information k Print information: Press p
> Printout (example)
Mod. LP6200S
Ser. no. 60406906
Ver. no. 01-30-13
Software version
(display and control unit)
Ver. no. 00-20-07
Software version
(weighing platform)
● Exit the Setup menu:
Press M
> Balance returns to previous status
10
Configuring the Balance
Entering User Data (Input)
Purpose
To display, input or change user data. You can block access to these data by assigning a password.
Features
You can display, input or change the following user data:
– Workstation number* for the balance: ID (balance ID; max.
20 characters)**
– Weighing series number, to designate a series or lot:
L ID (lot ID; max. 20 characters)**
– Weight set number for calibration/ adjustment: W ID (weight ID; max. 14 characters)**
– Exact weight value for calibration/adjustment of the balance
(see the section on “Calibration/
Adjustment,” starting on page 24 in the chapter entitled “Operating the Balance; in particular refer to page 27)
– Password for access to the Setup menu: Input and Setup:
Menu (max. 8 characters)**
* ** Only in conjunction with
ISO/GLP-compliant printouts
(see the section on “Setting
Parameters” in the chapter entitled “Configuring the
Balance;” under menu code number 8 10 x).
** A decimal point is displayed together with its preceding digit or character; it does, however, count as a separate character.
This also applies when you enter S ID and NUM as well as to data entered via the interface.
*** To delete user password:
Enter a decimal point using the
.
key and confirm
Factory Settings
Password: No designation
If no password has been assigned, anyone can access the “Setup:
Input” and “Setup: Menu” functions without entering a password.
If you assign a password and then forget what the word is, you can use the General Password (see Appendix) to access these menus.
Preparation
Display existing user data
● Select the Setup program:
Press M
> The soft keys Info, Menu and
Input are displayed in the text line
● Select the user data input function:
Press the Input softkey v
> The password prompt is displayed k If access is blocked by a password: enter the password using the alphanumeric input keys
● Display user data:
Press the enter passw.
soft key v
> The last 8 digits of a workstation/ balance number (ID no.), if any ID is assigned, are displayed in the measured value line
Enter/Change Password
● Select the Setup menu:
Press M
> The soft keys Info, Menu and
Input are displayed in the text line
● Select the user data input function:
Press the Input soft key v
If you have already assigned a password:
> The password prompt is displayed k Enter the password k Press the enter passw.
soft key v
● Write down the password here:
Password = .............................
If you no longer remember the password assigned: k Enter the General Password
(see Appendix) k Press the enter passw.
soft key v
> The last 8 digits of a workstation/ balance number (ID no.), if any ID is assigned, are displayed in the measured value line
● Select password setting: Press y
> Password is displayed in the text line
> If a password exists, it is now displayed in the measured value line
● New password: Enter the letters/ numbers for the new password
(8 characters max.)**
The password “none” means that no password is stored.***
●
Confirm input: Press X
● Exit the Setup menu:
Press M
> Restart the application
11
Practical Example
Enter “Workstation 234” as Balance ID; Display and Print Other User Data
Step Key (or instruction) Display/Output
1. Select Setup menu, then Input;
Display balance workstation ID (in this example: no ID number exists) see Preparation
2. Enter the first letter of the balance a workstation ID
13th through 20th digits of ID displayed z(- a -)b ID
3. Set the letter “W” y repeatedly, until the W is in the middle
4. Enter the next letter of the balance a workstation ID
5. Select the letter “o”
Y
repeatedly
6. Repeat steps 4 and 5 with the appropriate letters (display
“longer” values: see “Data Output
Functions” on page 57) a x X
7. Store balance workstation ID
ID soft key (X)
8. Display the 5th through 12th digits x of the balance workstation ID
12
Step Key (or instruction)
9. Display lot number
(In this example: 09-10-96/ABC1)
Y
Configuring the Balance
Display/Output
13th through 20th digits displayed
10. Display 5th through
12th digits of lot no.
x
11. Display other user data
– Weight set no.
– Exact calibration weight
– Password
12. Print user data (example) xyYX p
13. Exit “Setup: Input” M
13
Setting Parameters (Menu)
Purpose
To configure the balance; i.e., adapt the balance to individual requirements by choosing from a list of parameter options in a menu. You can block access to this menu by assigning a password.
Features
The parameter options are divided into the following groups
(1st menu level):
1 Balance functions
2 Application programs
3 Application parameters
4 +/– parameter
(for over/under checkweighing)
5 Interface parameters
6 Print for weighing (print weights)
7 Print for application program
(print app. data)
8 Additional functions
9 Reset menu
Factory Settings
The factory-set configurations are marked with an “o” in the list starting on page 16
Preparation
● Select the Setup menu:
Press M
> The Info, Menu and Input soft keys are displayed in the text line
● Select the parameter menu:
Menu soft key w
> Password prompt is displayed
● Enter password using the alphanumeric input keys
● Confirm password entered: enter passw. soft key v
> Measured value line:
1 (1st menu level)
> Text line: Balance functions k Select the next group: Press Y k Select the next submenu within a group (2nd menu level):
Press X k Select previous group: Press y k Return to next higher menu level:
Press x
Additional Functions
● Exit the menu: Press M
> Restart the application
● Print parameter settings:
– When the 3rd menu level is selected: Press p
> Printout (example)
4 2 Auto print +/-
2 Off
– When the 2nd menu level is selected: Press p
> Printout (example)
4 Parameter +/-
-------------------
4 2 Auto print +/-
2 Off
4 3 +/- ctrl ports
1 Within ctrl r
– When the 1st menu level is displayed: Press p
> Prints all of the menu parameters that are currently set
14
Configuring the Balance
Practical Example
Select the Counting Application Program
Step Key (or instruction)
1. Select Setup menu
M
Display/Output
info--menu--input
2. Select Balance Functions group code (Menu)
3. Select the Application
Prog. group
Menu soft key w
Y
4. Confirm Application Program
(2nd menu level)
X
5. Confirm App. Selection
(3rd menu level shows current setting; in this case: weighing)
X
6. Select the Counting program Y Y
7. Confirm selection of Counting program
X
8. Set other parameters, if desired
9. Save settings and exit menu x y Y X
M
15
Setup Parameters (Overview) o Factory setting
√
User setting
* = setting not applicable in balances verified for use in legal metrology
16
Configuring the Balance
* = setting not applicable in balances verified for use in legal metrology
** = for balances of accuracy class K, only calibration (not adjustment) can be performed with a user-defined weight
17
* = setting not applicable in balances verified for use in legal metrology
18
Configuring the Balance
Setup menu for applications
Level 1
1 to 4
Level 2 see previous pages
Level 3
Factor y setting
Menu item
5 Interface
6 Print for weighing
5 1
5 2 Parity
5 3 No. of stop bits
5 4 Handshake mode
5 5 Communication mode
5 6 Network address
6 1 Print manual/ automatic
5 6 1
5 6 x
5 6 32
6 1 1
6 1 2
6 1 4
6 1 5
6 1 6
5 5 1
5 5 2
5 5 4
5 5 5
5 5 6
5 5 7
5 5 10
5 5 11
5 5 14
5 5 15 see previous page
5 2 2
5 2 3
5 2 4
5 3 1
5 3 2
Space
ο
Odd
Even
ο
1 stop bit
2 stop bits
5 4 1
5 4 3
Software handshake
ο
Hardware handshake,
1 character after CTS
ο
SBI xBPI
YDP01IS
YDP02
YDP03
YDP01IS Label printer
YDP02IS
YDP02IS Label printer
YDP04IS
YDP04IS Label printer
ο
Address 0
Addresses 1 to 30
Address 31
Manual without stability parameter
ο
Manual with stability parameter
Auto print without stability parameter
Auto print at stability
Auto print after weight change**
6 2 Stop auto print
6 3 Time-dependent auto print
6 4 Print -> autotare
7 1 Print application parameters
6 2 1
6 2 2
6 3 1
6 3 2
6 3 4
6 3 7
6 4 1
6 4 2
Use p key
ο
Not possible
1 display update
2 display updates
10 display updates*
100 display updates*
ο
Off
On
ο
Off
All parameters
Only main parameters
7 Print for application
8 Extra functions
7 2 Line format
7 3 Print net total
8 1 Menu
7 1 1
7 1 2
7 1 3
7 2 1
7 2 2
7 3 1
7 3 2
8 1 1
8 1 2
For raw data
ο
For other applications/GLP
ο
Auto print net
Auto print tare
ο
Parameter settings alterable
Parameter settings readable
8 2 to 8 10 see next page
9 Balance menu see next page
* = setting not applicable in balances verified for use in legal metrology
** = Auto print if change in weight >10 d and has stable readout; function enabled when load <5 d
19
* = not in balance models with a weighing capacity
≥
16 kg.
20
Operating the Balance
Operating the Balance
Basic Weighing Function
Purpose
The basic weighing function is always accessible and can be used alone or in combination with an application program (Toggle between Weight Units, Counting,
Weighing in Percent, etc.).
Features
– Taring the balance
– Assigning IDs to weights
– Printing weights
– Printing ID codes for weights
Factory Settings
Tare: After stability (1 5 2)
Print manual/automatic:
Manual after stability (6 1 2)
Line format for printout:
For other applications/GLP (7 2 2)
Alphanumeric input of a weight ID:
Keys unblocked (8 3 1)
Below-Balance Weighing
A port for a below-balance weighing hanger is located on the bottom of the balance.
Balances with a weighing capacity
>12 kg:
● Open cover plate (1) on the bottom of the balance
1
● Attach the sample (e.g., using a suspension wire) to the hook (2).
2 k If necessary, install a shield for protection against drafts
Balances with a weighing capacity
≥
16 kg:
● Use a screwdriver to open the cover plate on the bottom of the balance
Preparation
● Turn on the balance: Press e
> All display segments light up briefly k To change configurations: see the chapter entitled
“Configuring the Balance” k To load factory-set configurations: see “Configuring the Balance,” parameter 9 1 k To tare the balance: Press t
> The U symbol is displayed when the balance is zeroed or tared (only on balances verified for use in legal metrology)
Using Verified Balances
Approved for Use as Legal
Measuring Instruments in the EU*:
The type-approval certificate for verification applies only to nonautomatic weighing instruments; for automatic operation with or without auxiliary measuring devices, you must comply with the regulations of your country applicable to the place of installation of your balance.
$
The temperature range indicated on the verification ID label must not be exceeded during operation
* including the Signatories of the Agreement on the European
Economic Area
● Attach the hook ordered directly from Sartorius
Important Note Concerning
Verified Balances Approved for Use as Legal Measuring
Instruments in the EU*:
The below-balance weighing port may not be opened or used when an approved balance is being operated as a legal measuring instrument.
21
ID for weight value (if desired):
● Select the parameters
“Line format” and “For other app./GLP” from the
Setup menu: Press M
● Select mode:
Press the menu soft key
● Set parameter 7 2 2: See the chapter entitled “Configuring the
Balance”
● Exit the Setup menu:
Press M
Additional Functions
In addition to the functions:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from the weighing application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration
● Press q
> See the section on “Calibration/
Adjustment” for further instructions.
Setup Menu
● Press M
> See the chapter entitled
“Configuring the Balance” for further instructions.
Turning Off the Balance
● Press e
> The balance shuts off
> The display goes blank
Important Note Concerning
Verified Balances of Accuracy
Class k:
To avoid measuring errors, the respective air density must be allowed for. The following formula is used to calculate the mass of the sample: m = nw
1 –
ρ
L/8000 kg m
1 –
ρ
L/
ρ
–3 m = mass of the sample nw = weight readout
ρ
ρ
L = air density during weighing
= density of the sample
Practical Examples
Example W1: Simple Weighing
Step Key (or instruction)
1. If necessary, tare the balance
(U symbol: balance is tared,
– verified balances only) t
Display/Output
2. Enter sample ID
3. Determine sample weight
(Example) see Example W2
Place sample on balance
4. Print weight
22 p
Operating the Balance
Example W2
Enter “ABC123” as a sample ID
Note:
– The sample ID generally applies to one weighing operation only
– The ID is deleted after data output
Step Key (or instruction) Display/Output
Initial status (balance unloaded)
(ID can also be entered while balance is loaded)
1. Enter the letter “A” a
2. Select and enter the letter “B” a
Softkey B (Key Y)
3. Select and enter the letter “C” a Softkey . .
(Key Y Y)
4. Enter the numbers “1,” “2” and “3” 1 2 3
5. Store the ID
(max. 20 characters)
– The next printout will include this ID
Press the SP.ID soft key v
23
Calibration/Adjustment
Purpose
Calibration is the determination of the difference between the weight readout and the true weight
(mass) of a sample. Calibration does not entail making any changes within the balance.
Adjustment is the correction of this difference between the measured value displayed and the true weight (mass) of the sample, or the reduction the difference to an allowable level within the maximum permissible error limits.
Available Features
Your balance can be calibrated externally (menu item 1 9 1 or1 9 3) or internally (1 9 4).
External calibration can be performed
– with the pre-set weight value
(1 9 1), or
– with a user-defined weight value
(1 9 3)
Adjustment can be performed
– automatically following calibration
(1 10 1); or
– if desired, the adjustment operation can be started manually after calibration (1 10 2)
You can also configure whether the calibration mode
– will be activated according to the specific setting (by setting
1 9 1, 1 9 3 or 1 9 4), or
– can be selected by the user after pressing the q key (1 9 12).
You can have the balance automatically display an adjustment prompt after a certain time interval has elapsed since the last calibration/adjustment or when the ambient temperature changes by a defined amount.
You can also configure the balance to perform calibration and adjustment automatically (isoCAL) when the pre-set time and/or temperature limit is reached
(1 15 3 and 1 15 5).
You can have the calibration/ adjustment results documented in a ISO/GLP-compliant printout.
Releasing Access to External
Calibration in Verified Balances of Accuracy Class k
§ Remove the covering plate from the back of the balance housing
§
Move Switch 1 in the direction of the arrow
> Switch down: external calibration accessible
Switch up: external calibration blocked
> Note:
Do not move Switch 2
For service technicians only:
External Calibration in Verified
Balances of Accuracy Class K
– External calibration is blocked when the balance is used in legal metrology
> External calibration can only be released after removing the verification control seal, in which case the validity of the erification becomes void and the balance must be re-verified
– External calibration can now be performed
Factory Settings
Calibration/adjustment mode:
Selection mode (1 9 12)
Calibration/adjustment sequence:
Adjustment automatically follows calibration in a single operation
(1 10 1)
Automatic initiation of calibration/ adjustment (isoCAL function): isoCAL on (1 15 5)
ISO/GLP printout: off (8 10 1)
* including the Signatories of the
Agreement on the European
Economic Area
24
Operating the Balance
Preparation
Configure Parameters for Calibration and Adjustment
Step Key (or instruction)
1. Turn on the balance e
Display/Output
2. Select the Setup menu M
3. Select the Balance Functions menu menu soft key w
4. Set parameters for:
– Calibration key function 1 9
– Calibration/adjustment sequence 1 10
– isoCAL self-calibrating and adjustment function 1 15
See the chapter entitled “Setup”
5. Exit the Setup menu x y Y X
M
info--menu--input
25
Selecting the Calibration/
Adjustment Parameter
The setting 1 9 12 must be selected in the Setup menu.
You can configure the balance so that after selecting a calibration procedure by pressing the q key, you can choose among the following calibration/adjustment modes:
– External calibration/adjustment with the pre-set weight value
(Def. Ext. Adj.)
– External calibration with a user-defined weight value
(user. ext.adj.)
– Internal calibration (int. adjust)
– Reproducibility test (reprotest)
When making your selection, the available modes are displayed in cycles; i.e., after reprotest, def. ext.adj.
is displayed again
Configure External Calibration and Automatic Adjustment of the Balance in Selection Mode
Step Key (or instruction) Display/Output
1. Select the calibration function q
2. Select external calibration/ adjustment mode
(for balances of accuracy class K, only “external adjustment” is possible)
3. Confirm external calibration/ adjustment mode
Press the selec.
soft key three times v
Press the user. ext.adj. soft key q
4. Place the calibration weight on the balance (e.g., 4000.00 g)
Minus sign –: Weight too low
Plus sign +: Weight too high
No plus/minus sign: Weight o.k.
Place weight on balance
This is displayed after calibration:
(on verified balances, the display shows the difference between the weight readout and the actual weight value)
This is displayed after adjustment:
5. Unload the balance Remove weight(s)
26
Operating the Balance
External Calibration/Adjustment* with a User-Defined Weight
First set either 1 9 3 or selection mode (1 9 12) in the menu.
You can define a weight for calibration/adjustment.
External calibration/adjustment must be performed with weights that are traceable to a national standard and that have error limits which are at least 1/3 of the required tolerance of the display accuracy.
The balance has a factory-set weight value (see “Specifications”).
To reset a user-defined calibration/ adjustment weight to the original factory setting, enter the factory-set weight value manually (see
“Specifications”) or set menu code
1 9 1 in the Setup menu.
Define the Calibration Weight
Step
1. Select the Setup menu
Key (or instruction)
M
Display/Output
info--menu--input
2. Access the Input Mode
Input soft key v
3. Select input for calibration weight (currently 3000.00 g)
Y Y Y
4. Enter calibration weight
(e.g., 4000.00 g)
5. Save weight
4 0 0 0 . 0
0
… weight soft key v
6. Exit the Setup menu M
* = for verified balances of accuracy class K, only external calibration is possible
27
Internal Calibration/Adjustment
The menu code setting 1 9 4 must be selected in the Setup menu.
Inside the balance housing is a builtin, motorized calibration weight.
The internal calibration/adjustment sequence is as follows:
● Select the calibration function:
Press q
> The internal calibration weight is applied automatically
> The balance is calibrated
> If the setting for “Calibration automatically followed by adjustment” (1 10 1) is selected in the Setup menu, the balance is now automatically adjusted
> The internal calibration weight is removed
Calibration and Adjustment Sequence
In the Setup menu, you can configure the balance so that:
– calibration is always followed automatically by adjustment
(cal. -> auto adjust, 1 10 1), or
– you have the choice of ending the sequence or starting adjustment after calibration
(cal. -> man. adjust, 1 10 2).
If no deviation is determined during calibration, or the deviation is within the tolerance limits dictated by the degree of accuracy you require, it is not necessary to adjust the balance.
In this case, you can end the calibration/adjustment sequence after calibration. There are 2 soft keys active at this point:
– ext. adjust or int. adjust to start adjustment
– end to end the sequence
28
Operating the Balance isoCAL:
Automatic Calibration and Adjustment
Either 1 15 3 or 1 15 5 must be selected in the Setup menu.
The “isoCAL” display automatically begins flashing if the ambient temperature has changed in relation to the temperature at the time of the last calibration/adjustment, or after a defined time interval has elapsed. The balance is telling you that it wants to self-calibrate and adjust.
This adjustment prompt is activated when:
– The change in temperature or the elapsed time interval is greater than that shown in the table below
– The balance status does not correspond to Setup configurations
– No number or letter input is active
– The load on the pan has not been changed within the last 2 minutes
– The balance has not been operated within the last 2 minutes
– The weight on the pan must be no more than 2% at the most of the maximum capacity of the balance
When these requirements are met, the following symbols are displayed:
– C in the measured value line
– isoCAL in the symbol display
– G in the weight unit display
If the balance is not operated and the load is not changed, internal calibration and adjustment starts after
15 seconds have elapsed.
In the Setup menu, you can configure the balance so that after calibration and adjustment
– the application program must be restarted
(On + reset app. 1 15 3), or
– the application program resumes where it left off
(isoCAL on, 1 15 5)
In the Setup menu, you can also configure the balance so that it displays an adjustment prompt, but does not perform the calibration/ functions automatically
(Only adj. prompt, 1 15 2)
Deactivating the “isoCAL”
Function in Verified Balances:
Automatic calibration is also performed outside of the limited temperature range if you set menu code 1 15 1, “isoCAL function: off” or code 1 15 2, “only at adjustment prompt” in the Setup menu.
To generally deactivate automatic calibration in balances with a weighing capacity
≤
12 kg:
– After the balance has been modified by the Sartorius
Service Center
> Afterwards, the balance can only be used for legal metrology within the temperature range allowed by law
Limited temperature range:
– For balances of accuracy class k: +15
°
C to +25
°
C
(59
°
F to 77
°
F)
– For balances of accuracy class K: +10
°
C to +30
°
C
(50
°
F to 86
°
F)
Extended temperature range:
– 0
°
C to +40
°
C
(32
°
F to 104
°
F)
Fully automatic adjustment is initiated under the following conditions:
Model When the temperature After a time changes by interval of
LP 3200D, LP 1200S
LP8200S, LP8200P,
LP 620S, LP 620P, LP 6200S,
LP 4200S, LP 5200P, LP 220S
1.5 Kelvin
2 Kelvin
4 h
6 h
LP 2200S, LP 2200P,
LP 34001P, LP 34001S, LP 64001S
4 Kelvin 12 h
LP 820, LP 420, LP 16001S,
LP 12000S, LP 12000P,
LP 6200, LP 4200, LP 2200,
LP 34000
4 Kelvin 24 h
These values are also set in the corresponding verified or verifiable balances
(LP models with the -0CE designation).
29
Determination of the
Repeatability (reproTEST)
Definition
Repeatability (reproducibility) is the ability of the balance to display identical readouts when it is loaded several times with the same weight under constant ambient conditions.
The standard deviation for a given number of measurements is used to quantify the repeatability.
Purpose
The “reproTEST” function automatically calculates the repeatability of results (based on 6 individual measurements). In this way, the balance determines one of the most important quantities in relation to the place of installation. The results are displayed with the balance’s accuracy.
Preparation
● Turn on the balance:
Press e
> All display segments light up briefly
● Select reproTEST in the Setup menu: Press M
● Select Menu:
Press w (menu softkey)
● Select either 1 9 11 (reproTEST) or 1 9 12 (selection mode):
See “Configuring the Balance.”
● Exit the Setup menu:
Press M
Check the Reproducibility of the Balance
Step Key (or instruction)
1. If parameter 1 9 11 is set
(reproTEST), proceed with step 4.
2. Access the Selection Mode for calibration/adjustment q
3. Select reproTEST selec. soft key v
Display/Output
4. Start reproTEST
Number of measurements is displayed
6 measurements will now be performed
The standard deviation is displayed q
5. End reproTEST or restart reproTEST
30
End soft key v reprotest soft key
Operating the Balance
Application Programs
Using Verified Balances as Legal
Measuring Instruments in the EU*:
All application programs can be selected on balances used as legal measuring instruments.
Calculated values can be indicated as follows:
Percent = %
Piece count (counting) = pcs
Computed value = o, G
Toggle between Weight Units
Purpose
With this application program you can switch the display of a weight value back and forth between two weight units by pressing a soft key.
Available Features
– Toggling the displayed weight
– Setting the display accuracy
– Other features as for the basic weighing function
Preparation
The following weight units are available in both ranges:
Unit Conversion factor Display
Grams
Kilograms
1.00000000000
0.00100000000
g kg
Carats**
Pounds**
Ounces**
Troy ounces**
Hong Kong taels**
5.00000000000
0.00220462260
0.03527396200
0.03215074700
0.02671725000
ct lb oz ozt tl
Singapore taels**
Taiwanese taels**
Grains**
Pennyweights**
Milligrams**
Parts per pound**
Chinese taels**
Mommes**
Austrian carats**
Tola**
Baht**
Mesghal**
0.02645544638
0.02666666000
15.43235835000
0.64301493100
tl tl
GN dwt
1000.00000000000
mg
1.12876677120
o
0.02645547175
0.26670000000
5.00000000000
0.08573333810
0.06578947436
0.21700000000
o o o o tl o
Factory Settings
Weight unit 1: Grams/g (1 7 2)
Display accuracy 1 (in the 1st range): All digits (1 8 1)
Weight unit 2: Grams/g (3 1 2)
Display accuracy 2 (in the 2nd range): All digits (3 2 1)
* including the Signatories of the Agreement on the European
Economic Area
** = not applicable in balances verified for use in legal metrology
The following levels of display accuracy are available in both ranges:
– All digits
– Fewer for weight change
– Last digit off (reduced by 1 digit)
●
Turn on the balance: Press e
> All segments of the display light up briefly
● Configure the “Toggle between
Weight Units” application in the
Setup menu: Press M
● Select the configuration mode:
Press w (menu soft key)
●
Set the parameter 2 1 2: See the chapter entitled “Configuring the Balance”
● Exit the Setup menu: Press M
Setting Weight Unit 1
● Access the Setup menu: select Menu
●
Set the parameter for weight unit 1 (1 7 2 through 1 7 20):
See “Configuring the Balance”
● Set display accuracy 1
(1 8 1 through 1 8 5) :
See “Configuring the Balance”
Setting Weight Unit 2 k Access Setup: select Menu
●
Set the parameter for weight unit 2 (3 1 2 through 3 1 20):
See “Configuring the Balance”
● Set display accuracy 2 (3 2 1 through 3 2 5):
See “Configuring the Balance“
● Exit the Setup menu: Press M tls tlt
GN dwt mg
/lb tlc mom
K tol bat
MS lb oz ozt tlh
Printout g kg ct
31
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from this application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
● Press e
> The balance shuts off
> The display goes blank
Practical Example
Toggle the Display From Grams [g] (1st Unit) to Troy Ounces [ozt] (2nd Unit)
Step Key (or instruction) Display/Output
(R1: weight unit 1)
1. Toggle to Troy ounces [ozt]
(R2: weight unit 2) ozt soft key v
2. Toggle to grams [g] g soft key v
32
Operating the Balance
Counting Z
Purpose
With the Counting program you can determine the number of parts that each have approximately equal weight.
Available Features
– Reference sample quantity “nRef” loaded from long-term memory when you turn on the balance
– Optional balance configuration in the Setup menu for automatically initializing this application and loading the most recent reference sample quantity
“nRef” and average piece weight
“wRef” when you turn on the balance (automatic initialization; menu code 3 10 1)
– Minimum load checked when the balance is initialized
(factory setting: 10 display increments; can only be changed by the Sartorius FastFactory)
– Reference sample quantity “nRef” entered using the keys
– Average piece weight “wRef” entered using the keys
– Storage parameter (level of accuracy with which the average piece weight “wRef” is stored) for piece count calculation can be configured
– Optional configuration for having the piece count and average piece weight output automatically via the data interface port after initialization or reference sample updating while running the
Counting program (print application parameters)
– Long-term storage of the last reference sample quantity “nRef” entered
– Long-term storage of the last average piece weight “wRef” entered with the corresponding reference sample quantity “nRef,” by configuring automatic initialization
– Toggling between two weight units by pressing w
– Counting program initialized again after using the balance for weighing (after initialization)
Factory Settings
Auto-start application (automatic initialization with reference sample quantity and average piece weight loaded from long-term memory): off (3 10 2)
Counting/percent parameter
(accuracy when storing average piece weights): display accuracy
(3 5 2)
Printout application parameters
(automatic output of application parameters): off (7 1 1)
33
Preparation
To calculate a piece count, the average weight of one piece must be known. This average piece weight can be entered into the
Counting program in one of three ways:
– The last reference sample quantity entered is loaded and displayed when you turn on the balance. Place the same number of parts on the balance and initialize the counting program;
– With automatic initialization switched on, the balance goes into the “counting” mode when you turn it on and loads the last average piece weight and corresponding reference sample quantity that were entered or calculated;
– Enter the average piece weight using the numeric keys and store it.
Reference Sample Updating
When the updat soft key is displayed during counting, this means you can have the average piece weight updated (while the piece count is displayed in the measured value line).
The updat soft key is displayed when:
– the balance has reached stability
– the current piece count is less than double the original piece count
– the current piece count is less than 100
– the internally calculated piece count (e.g., 17.24 pcs) differs from the nearest whole number (here:
17 pcs) by less than 0.3
Reference sample updating can be repeated several times with an approximately doubled piece count.
● To perform reference sample updating: Press v (updat soft key)
● Turn on the balance:
Press e
> All display segments light up briefly
● Select the Counting program in the Setup menu: Press M
● Select menu:
Press the menu soft key (w)
● Set parameter 2 1 4: See
“Configuring the Balance”
● Exit the Setup menu:
Press M
Setting Parameters for the
Counting Application
● Access the Setup menu:
●
Select Menu
Set parameters for:
– Storage accuracy:
3 5 Count/pct. param.
– Automatic initialization:
3 10 auto-start app.
– Automatic output of parameters to interface port:
7 1 Print app. param.
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from this application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
●
Press e
> The balance shuts off
> The display goes blank
34
Operating the Balance
Practical Example
Counting with: Preset Reference Sample Quantity Weighed In
(Parameter settings: Counting program: 2 1 4; Print all parameters: 7 1 2)
Step Key (or instruction) Display/Output
1. Select the counting application in the Setup menu
2. Prepare a container for the parts see “Preparation”
Place the empty container on the balance nref=10 pcs start
3. Tare the balance t nref=10 pcs start
4. Place reference sample quantity Place the displayed number on the balance (example: 10 pcs, of parts in the container each weighing 2.148 g)
5. Initialize the balance
(the number of digits following the decimal point depends on the balance model)
6. If necessary, increase number of parts and update the reference sample
(here: 7 more pieces)
7. Weigh uncounted parts start soft key v
Place parts in container updat soft key v
Place parts to be counted in container nref=10 pcs start
8. If desired, print total piece count
(here: 153 pcs) p nRef + 10 pcs wRef + 2.14800 g
Qnt + 153 pcs
35
Weighing in Percent L
Purpose
This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight.
Available Features
– Reference percentage “pRef” loaded from long-term memory when you turn on the balance
– Optional balance configuration in the Setup menu for automatically initializing this application and loading the most recent reference percentage “pRef” entered with reference weight “w100%” when you turn on the balance
– Minimum load checked when the balance is initialized
(factory setting: 10 display increments; can only be changed by the Sartorius FastFactory)
– Reference percentage “nRef” entered using the numeric keys
– Reference weight “Wxx%” entered using the numeric keys
– Storage parameter (rounding-off factor) for storing the reference weight “W100%” to calculate the percentage can be configured
– Configuration of decimal places displayed with a percentage
– Optional configuration for having the reference weight “Wxx%” and reference percentage automatically output via the data interface port after initialization of the weighing-in-percent program
(print application parameters)
– Long-term storage of the last reference percentage
“pRef” entered
– Long-term storage of the last reference weight “W100%” entered, by configuring automatic initialization (auto-start)
– Toggle between two weight units by pressing w
Factory Settings
Auto-start application (automatic initialization with reference percentage and reference weight loaded from long-term memory): off (3 10 2)
Counting/percent parameter
(accuracy when storing reference weights): display accuracy (3 5 2)
Number of decimal places displayed in “weighing in percent” mode:
2 decimal places (3 6 3)
Printout application parameters
(automatic output of application parameters): off 7 1 1)
36
Operating the Balance
Preparation
To calculate a value in percent, the reference percentage must be known. This value can be entered into the weighing-in-percent program in one of three ways:
– The last reference percentage entered is loaded and displayed when you turn on the balance.
Place the corresponding weight on the balance and initialize the weighing-in-percent program;
– With automatic initialization switched on, the balance goes into the “weighing in percent” mode when you turn it on and loads the last reference percentage entered as well as the corresponding reference weight;
– Enter the reference weight using the numeric keys and store it (W100% soft key).
● Turn on the balance:
Press e
> All display segments light up briefly
● Select the Weighing-in-Percent application in the Setup menu:
Press M
● Select menu:
Press the menu soft key (w)
● Set parameter 2 1 5:
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Setting Parameters for the Weighingin-Percent Application k Access the Setup menu:
●
Select Menu
Set parameters for:
– Storage accuracy:
3 5 Count/pct. param.
– Decimal places displayed:
3 6 decimals f. calc.
– Automatic initialization:
3 10 auto-start app.
– Automatic output of parameters to interface port:
7 1 Print app. param.
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Additional Functions
In addition to functions for:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from this application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
●
Press e
> The balance shuts off
> The display goes blank
37
Practical Examples
Example P1: Weighing in Percent with Reference Weight Taken from Weight on Balance
Step Key (or instruction) Display/Output
1. Select the weighing-in-percent application in the Setup menu
2. Prepare a container for the parts see “Preparation”
Place the empty container on the balance
3. Tare the balance t
4. Place the reference weight on the balance
(here: 1821.48 g =100%)
5. Initialize the balance
Place weight equal to reference weight in the container start soft key v
6. Unload the balance Remove reference weight from the container
7. Determine the percentage of an unknown weight
Place sample to be measured in the container
8. If desired, print percentage
(here: 98.37%) p pRef + 100 %
Wxx% + 1821.48 g
Prc + 98.37 %
38
Operating the Balance
Example P2: Weighing in Percent with Reference Weight Entered Using the Numeric Keys
Step Key (or instruction) Display/Output
1. Select the weighing-in-percent application in the Setup menu
2. Prepare a container for the parts see “Preparation”
Place the empty container on the balance
3. Tare the balance t
4. Enter the reference weight using the numeric keys (here: 120 g)
1 2 0
5. Store the reference weight
W100% soft key q
6. Determine the percentage of an unknown weight
Place sample to be measured in the container
(in this case: 114.78 g)
39
Net-Total Formulation R
Purpose
With this application program you can weigh in different components up to a defined total.
You can also print out the total weight and the individual weights of the components.
Available Features
– Taring
– Weighing different components
(maximum: 99 components) from
“0” to a defined total component weight
– Storing component weights
(“Store xx comp.”), with
– display zeroed after value stored, and
– automatic printout (print application parameters); either
– of the last component weight
(net value) or
– of the total weight (tare value)
– Display of the transaction counter
“XXth” (referring in each case to the next component) in the soft key label display
– Clearing of the component memory when the weighing series is canceled c and printout of the total weight if you have configured the balance for GLP-compliant printouts; otherwise, printout of the net value p
– Toggling between component weight and total weight by pressing w.
– ISO/GLP-compliant printout of the total of the individual component weights (Tot.cp)
Preparation
● Turn on the balance
> All display segments light up briefly
● Select the Net Total application in the Setup menu: Press M
● Select menu:
Press the menu soft key (w)
● Set parameter 2 1 6:
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Setting Parameters for the Net-Total Application k Access the Setup menu:
●
Select Menu
● Set the parameter for automatic printout when component stored
7 3 Print net total
See “Configuring the Balance”
Exit the Setup menu:
Press M
Additional Functions
In addition to functions for:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from this application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
●
Press e
> The balance shuts off
> The display goes blank
Factory Settings
Automatic printout when component value stored: print net total (7 3 1)
40
Operating the Balance
Practical Example
Weighing in Several Components
(Parameter settings: Net-total application program 2 1 6; print all parameters 7 1 2)
Step Key (or instruction) Display/Output
1. Select the net-total application in the Setup menu
2. Prepare a container for the components see “Preparation”
Place the empty container on the balance
3. Tare the balance t
4. Add first component
5. Store component value
Place the first component in the container
1. comp soft key v
Comp1 + 952.48 g
6. Add next component
7. Store component value
8. If desired, add further components
9. Display total weight
Place next component in the container
2. comp soft key v
Repeat steps 6 and 7 as needed w
Comp1 + 952.48 g
Comp2 + 2837.12 g
41
Animal Weighing V
Purpose
Use this program to determine the weights of unstable samples (e.g., live animals) or to determine weights under unstable ambient conditions.
In this program, the balance calculates the weight as the average of a defined number of individual weighing operations. These weighing operations are also known as “subweighing operations.”
Available Features
– Animal weighing started manually or automatically
– Optional balance configuration in the Setup menu for automatically initializing this application when you turn on the balance
– Minimum load threshold for starting animal weighing:
– 100 display increments for automatic start
– 50 display increments for manual start
– Start range: Automatic start of animal weighing operation when three successive subweights lie within a user-defined tolerance range (calm = 2%, normal = 5%, active = 10%)
– Number of weighing operations for calculation of an average mdef can be set before the beginning of each animal weighing operation
– Arithmetic average displayed as a result in the pre-set weight unit
(identified by the G symbol)
– Optional multiplication of the arithmetic average by a user-defined factor mul.
A circle “o” is displayed as weight unit and mul=xxxx is shown in the text line
– Toggling between weighed and calculated results by pressing w
– Automatic output via the interface port:
– Number of weighing operations mDef
– Multiplication factor Mul
– Automatic output of results via the interface port:
– Weighing result x-Net
– Calculated result x-Res
– Stop limit: Unload threshold
(50 display increments)
– Return to weighing mode by unloading the balance; i.e., when the load is below the stop threshold
Factory Settings
Auto-start application (automatic initialization with automatic start of animal weighing): off (3 10 2)
Animal activity (3 subweights lie within a preset range; i.e., animal is calm, normal or active): normal (3 7 2)
Start animal weighing: automatic (3 8 2)
Automatic printout of number of weighing operations used in averaging and of the calculation factors: off (7 1 1)
Print animal weights (automatic printout of weighed or calculated result):
On: animal wt. (3 9 2)
42
Operating the Balance
Preparation
● Turn on the balance:
Press e
> All display segments light up briefly
● Select the animal weighing application in the Setup menu:
Press M
● Select menu:
Press the menu soft key (w)
● Set parameter 2 1 7:
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Setting Parameters for the
Animal Weighing Application k Access the Setup menu:
Select Menu
● Set parameters for:
– Start range:
3 7 Animal activity
– Start animal weighing
3 8 Start animal wgh.
– Printout of results and calculated results:
3 9 Print animal wts.
– Automatic output to interface port:
7 1 Print app. param.
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Additional Functions
In addition to functions for:
– alphanumeric input
(not when automatic start is configured or after animal weighing has been started),
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from this application:
– calibration (not during alphanumeric input or after animal weighing has been started),
– setup (not after animal weighing has been started),
– turning off the balance.
Practical Example
See next page
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
●
Press e
> The balance shuts off
> The display goes blank
43
Practical Example
Determining Animal Weight with Automatic Start of 20 Subweighing Operations for Averaging;
Automatic Printout of the Number of Subweighing Operations and of the Animal Weight
Step Key (or instruction) Display/Output
1. Select the animal weighing application in the Setup menu see “Preparation”
2. Set the following animal weighing parameters in the Setup menu:
– Animal activity: Active Setting 3 7 3 in Setup menu
– Start animal weighing: Automatic Setting 3 8 2 in Setup menu
– Printout: On: anim.wt.*fact.
– Print: All parameters
Setting 3 9 3 in Setup menu
Setting 7 1 2 in Setup menu
3. Prepare a container (cage) Place empty cage on the balance mdef= x10 start
4. Tare the balance t mdef= x10 start
5. Enter number of subweighing operations for averaging
2 0 mul mdef S-id
6. Save number mdef soft key w
7. Weigh the first animal Place 1st animal in cage mdef= 20 start weight value fluctuates due to animal activity mdef= 20 start
8. Start automatic animal weighing start soft key v
M =-20xxxxx---
44
Operating the Balance
Step
The balance delays starting the subweighing operation until three successive subweights lie within the range defined for an
“active” animal
Key (or instruction)
When this criterion is met, the subweighing series begins
Display/Output
M =x20xxxxxxx
xxx19
xxx18
xxxx...
xxxx1
After 20 subweighing operations
(mdef: no. of subweighs)
Mul: Calculation factor xnet arithm. average, net value)
8. Unload the balance
9. If desired, weigh next animal
Remove animal from cage mDef 20
Mul 1 x-Net + 69.72 g x-Res + 69.72 o
Place animal in cage mdef=x20 auto
Next weighing series begins automatically
M =-20xxxxx---
M =x20xxxxxxx
xxx19
xxx18
xxxx...
xxxx1
45
Over/Under Checkweighing O
Purpose
This program is used to check whether a sample corresponds to a pre-set target value or is within a specific tolerance range.
In addition to the display in the measured value line, the results are shown on the bar graph and can also be routed through the interface port via control lines for further electronic processing.
Available Features
– Lower tolerance limit (minimum), target value and upper limit
(maximum) stored in long-term memory
– Optional balance configuration in the Setup menu for automatically initializing this application and loading the values stored in long term memory for the target value and the upper and lower tolerance limits when you turn on the balance
– Enter target value and limits by placing a load on the balance or using the numeric keys
– Control in entering target and tolerance values, so that the upper limit
≥
the target
≥
the lower limit
≥
1 display increment
– Accuracy of a weight readout or keyboard input as target/ tolerance value stored corresponds to the display accuracy
– Optional balance configuration in the Setup menu for automatic output to the interface port
(print application parameters) of target value and tolerance limits when initialization is completed
– Control range for the balance’s data output port lines is 30% to
170% of the target value
– Control range can be set anywhere from 10% to infinity.
Control lines set when stability is reached within the control range; deleted only when you exit this application
– Optional configuration in the Setup menu for activation of control lines independent of weight value (weight within control range, stability reached)
– Toggling the display between weight readout and control
(checkweighing) display by pressing v. If the weight value exceeds tolerances, the measured value line shows the weight while the control display shows “LL” for “too low” or “HH” for “too high.”
– Toggle the text line display between weight value and control display, nominal value “setp” and tolerance values “min” and “max” by pressing the w key.
– Weight displayed on bar graph in relation to upper and lower limits and target value.
– Optional automatic printout of weight when it is within the control range at stability (auto print +/-).
After an automatic printout, the balance is blocked. Before you can generate the next printout, you must unblock the balance by unloading it (weight must be under 30% of the target) or by placing a load on the balance
(bringing the weight up to at least 170% of the target).
– Press c to delete the initialization parameters and end the over/under checkweighing program
Factory Settings
Auto-start application (automatic initialization with target value and tolerance limits loaded from long-term memory): off (3 10 2)
Automatic output of target value and tolerance limits via the interface port (print application parameters): off (7 1 1)
Auto print +/– (automatic printout of a weight when it is within the control range at stability): off (4 2 2)
Control lines +/– on
(activating data output port lines):
Within control range (4 3 1)
46
Operating the Balance
Preparation
The checkweighing program requires a target value for comparison to the current value. This target has a tolerance range, which is defined by absolute weight values: upper and lower limits. These limits can be entered either by storing weights on the balance or via key input.
There are four control lines, called data output port lines, which are activated as follows: (see also the diagram at the right):
– lighter
– equal
– heavier
– set
The control range spans 30% to
170% of the target value. You can configure this parameter in the
Setup menu (4 3 +/- ctrl ports on) to select whether the control lines are:
– always on
– activated within the control range
– activated at stability
– activated at stability within the control range
This makes it possible, for example, to connect a simple indicator for the weighing results
(e.g., three different colors, one each for the weighing results: too light,
O.K., too heavy).
Response of Control Lines
During Checkweighing
Configurations:
– always on
– activated at stability
Configurations:
– activated within control range
– activated at stability within control range
Output port specifications
When not in use, the voltage level is high: >2.4 V/+2 mA
When in active use, the voltage level is low: <0.4 V/–2 mA.
!
The output ports are not protected against short circuits!
47
●
●
Turn on the balance:
Press e
> All display segments light up briefly
Select the over/under checkweighing application in the
Setup menu: Press M
● Select menu:
●
Press the menu soft key (w)
● Set parameter 2 1 8:
See “Configuring the Balance”
Exit the Setup menu:
Press M
Additional Functions
In addition to functions for:
– alphanumeric input
(not during initialization),
– taring
(not during alphanumeric input),
– printing, you can also access the following functions from this application:
– calibration
(not during alphanumeric input or during initialization),
– setup (not during initialization),
– turning off the balance.
Setting Parameters for the Over/
Under Checkweighing Application k Access the Setup menu:
Select Menu
● Set parameters for:
– Automatic initialization when balance switched on:
3 10 auto-start app.
– Automatic printout when weight value is within limits and balance at stability:
4 2 auto print +/-
– Control lines on:
4 3 +/- ctrl ports on
Automatic output of initialization values to interface port:
7 1 Print app. param.
– Signal direction for data output port lines:
8 8 1 input or
8 8 2 output
See “Configuring the Balance”
● Exit the Setup menu:
Press M
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
●
Press e
> The balance shuts off
> The display goes blank
48
Operating the Balance
Practical Example
Checkweighing samples of 170 g, with an allowable tolerance of –5 g and +10 g.
Printout of upper and lower tolerance limits. Weighed values printed out automatically when stability is reached and weight is within the control range.
Step Key (or instruction) Display/Output
1. Select the checkweighing application in Setup
2. Set the following checkweighing parameters in the Setup menu:
– Auto print +/–: On: values in tol.
– Print app. param.: All parameters see “Preparation”
See “Configuring the Balance”
3. Prepare a container for the sample Place empty container on the balance plus/minusxxstart
4. Tare the balance t
plus/minusxxstart
5. Enter initialization values start soft key v
6. Store target value (here: 170 g) Place ideal sample in container
7. Store target value and unload balance
8. Enter value for lower limit
(170 g – 5 g) setp soft key v
Remove ideal sample from balance
1 6 5
49
9. Store lower limit value min soft key v
10. Enter value for upper limit
(170 g + 10 g)
1 8 0
11. Store upper limit value max soft key v
Place sample in container
Setp + 170.00 g
Min + 165.00 g
Max + 180.00 g
12. Weigh sample
(in this case 169.48 g)
If the weight had been too low, the display would have shown the following:
13. In this case, switch to net value display (in this case: 163.28 g) net soft key v
14. Weigh next sample (if any) Place sample in container
N + 169.48 g
50
Operating the Balance
Recalculation
Purpose
With this application program you can compensate for over-poured components in formulation.
If a component is over-poured when weighing in the individual formulation components, the mixture already poured cannot be used in its current composition. To avoid having to discard the materials weighed, you can adjust the proportions of the formulation to compensate for the over-pour.
When you use this application, the recalculation procedure is mainly performed by the balance.
Available Features
– Individual components (up to 99) weighed in with a readout showing from “0” to the desired component weight
– Transaction counter shows the next component expected
– Weighed components are stored, followed by automatic printout and taring
– Additive weighing of components with printout
– Toggle the display between component weight and total formulation weight (additive mode) after first component is stored
– Stored component weight displayed as true net weight for
2 seconds
– Enter a divisor before or during component weighing. For example, if the formulation has a total weight of 100 g, enter the divisor 10 to weigh in a total formulation of 1,000 g.
– If a component is over-poured, you can use the recalculation function to change the amount of this component indicated in the formulation by using plus or minus keys or numeric input. A factor is then calculated by which all components amounts are then adjusted.
– Recalculation factor displayed in the text line, with a warning symbol if the factor is not equal to 1.
– All components displayed with number and the amount (by weight) to be added in follow-on filling. Components displayed in sequence by the balance.
– Display of actual net weight during follow-on filling
– After the amounts of the components already weighed have been corrected, weighing continues according to the adjusted formulation amount.
The readout is recalculated
(updated) according to the divisor.
– You can repeat the over-pour correction procedure as often as necessary, in case other components are over-poured.
– After follow-on (corrective) filling, the total amount differs from that given for the formulation, but the proportion of components in relation to each other is the same.
– You can have the weight printed after each measurement
– Choose whether the current component weight or the tare value is printed after each measurement
– Individual component weights are printed as “Compxx.”
– Press c to exit the application program. The component memory is cleared and the sum of components printed as “S-Comp.”
51
Factory Settings
Print application parameters
(automatic output of application parameters): off (7 1 1)
Line format: for other apps/GLP
(7 2 2)
Preparation
● Turn on the balance: Press e
> All display segments light up
● Select the Recalculation application in the Setup menu: Press M
● Select menu:
Press the menu soft key (w)
● Set parameter 2 1 9: see “Configuring the Scale”
● Exit the Setup menu: Press M
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during alphanumeric input), and
– printing, you can also access the following functions from this application:
– calibration
(not during alphanumeric input)
– setup
– turning off the balance.
Setting Parameters for the
Recalculation Application
$ Access the Setup menu:
Select MENU
● Set parameters for:
– Automatic printout:
7 1 Print app. para.
– Line format:
7 2 Line format see “Configuring the Balance”
● Exit the Setup menu: Press M
Calibration/Adjustment
● Press q
> See “Calibration/Adjustment” for further instructions
Setup (setting parameters)
● Press M
> See “Configuring the Balance” for further instructions
Turning Off the Balance
● Press e
> The balance shuts off
> The display goes blank
52
Operating the Balance
Practical Example
When weighing in formulation components, the second component is over-poured.
(Parameter settings: Recalculation application: 2 1 9; Print all parameters: 7 1 2)
Step Key (or instruction) Display/Output
—————————————————————————————————————————————————
1. Select the recalculation application in the Setup menu see “Preparation”
2. Place container for filling components on the balance
Place empty container on the balance store comp.1
3. Tare t store comp.1
4. Add the first component
5. Store component
6. Add the second component
Weigh the first component into the container store comp.1
Comp1 + 27.08 g
Press the comp.1
soft key (v)
Weigh the second component into the container
recal. add. comp.2
7. Start recalculation, because
12.42 g were poured rather than 12.30 g
8. Either press the minus key to correct the value...
recal soft key (q) minus soft key repeatedly
minus plus comp.2
minus plus comp.2
53
... or enter the desired value 1 2 . 3 0
9. Confirm the new value comp.2 soft key (v) or comp. soft key (w)
Comp1 + 27.08 g
Comp2 + 12.42 g
R.div + 1.00975
comp. sp.id
The true net value is displayed for 2 seconds
store comp.2
Follow-on filling amount for first component is displayed
r.div.=1.00975 comp.1
10. Follow-on filling of 1st component Weigh the first component up to 0 and store comp.1 soft key (v)
Comp1 + 27.08 g
Comp2 + 12.42 g
R.div.+ 1.00975
RCom1 + 27.34 g
The true net value is displayed for 2 seconds
actual net. comp.1
11. Weigh in further components, if called for in the formulation
12. Toggle to the additive mode, if required
Repeat steps 4 and 5 as needed add. soft key (w)
recal. comp. add.6
13. Add further components, as required ...
(here, e.g., up to the total weight of the formulation:1,000 g)
Add components to container
recal. comp. add.6
54
14. ... and store
(here, e.g., the 6th component) add.6 soft key (v)
Operating the Balance
Comp1 + 27.08 g
Comp2 + 12.42 g
R.div.+ 1.00975
RCom1 + 27.34 g
Comp3 + 18.39 g
Comp4 + 31.49 g
Comp5 + 107.50 g
Comp6 + 812.61 g
The true net value (of the 6th component) is displayed for
2 seconds
Then the total weight is displayed actual net. Comp.6
15. End the weighing procedure
Total weight is printed c
recal. comp. add.7
Comp1 + 27.08 g
Comp2 + 12.42 g
R.div.+ 1.00975
RCom1 + 27.34 g
Comp3 + 18.39 g
Comp4 + 31.49 g
Comp5 + 107.50 g
Comp6 + 812.61 g
Tot.cp+ 1009.75 g
Total weight is displayed
Component memory is cleared
recal comp.1
55
56
Data Output Functions
There are 3 options for data output:
– Output to the display/control unit
– Output to a printer (generate a printout)
– Output to a peripheral device (e.g., computer) via the interface port
Output to the Display and Control Unit
The display is divided into 6 sections:
– Line for metrological data
– Bar graph
– Measured value line
– Weight unit display
– Symbol display
– Text line
Max 4200 g
Min
0.5 g e= 0.1 g d= 0.01 g
Line for Metrological Data (when used as a legal measuring instrument)
This line shows:
– Maximum balance capacity (e.g., 4,200 g)
– Minimum balance capacity; the weight must not go below this limit when the balance is used in legal metrology
– Verification scale interval of the balance; irrelevant if the balance is not used in legal metrology (e.g., 0.1 g)
– Readability: Indicates the actual scale interval
(display increment of the balance) (e.g., 0.01 g)
Bar Graph
In the bar graph, weighing results are displayed either
– as a percentage of the maximum balance capacity, or
– in relation to a target value, with tolerance limits indicated.
1234.56
01-30-13
-96-Abc1
Measured Value Line
This line displays:
– The current weight readout
(digits bordered with a broken line are invalid in use for legal metrology)
– Information on the balance and display and control unit
(e.g., version numbers)
– Data entered by the user (e.g., lot number)
Operating the Balance kg pcs
NET P o
Weight Unit Display
This line shows:
– Current weight unit (e.g., kg)
– Designation of other units of measure (e.g., piece count)
– Operating status (e.g., net weight; printing)
– Identification of currently selected menu item isoCAL
L Z V O R
Symbol Display
This line shows:
– Indicators and warnings (e.g., isoCAL)
– Symbol of application program selected reprotest info menu input n n n
Text Line
This line contains:
– Explanatory text for value displayed in measured value line
– Designation of current function of the soft keys (below the arrows)
– Arrows indicating the soft keys designated above
TON 4020
((_____id________
Display of “Longer” Values
You can enter values up to 20 characters in length
(e.g., identification numbers). These are displayed in the measured value line; this line, however, can only show up to 8 characters at a time.
When a 20-character value is displayed, for example, the last 8 characters
(13 through 20) are displayed first (in this example, the ID number is:
“Shelf 5, carton 4020”).
The text line displays the name of the value (“ID”) and an arrow indicating the direction of the characters not displayed.
F 5, CAR
((_____id______))
Display characters 5 through 12: Press x
SHEL
_____id______))
Display characters 1 through 4: Press x
57
Printing a Data Record
Purpose
You can generate a printout of weights, other measured values and identification numbers for documentation purposes. You can format the printout to meet individual requirements.
Available Features
Print manually/automatically:
To print the information contained in the measured value line (weight readout, calculated value, numeric entry, alphabetic entry)
Line format: You can configure a data ID code of up to 6 characters for each of the values printed; this data ID code is printed at the beginning of the line
Sample ID: You can configure an extra line for identification of each weighed or calculated value
Print application parameters:
You can generate a printout of the values configured for initialization of an application before printing the measured results
ISO/GLP-compliant printout:
To print out parameters relating to weighing conditions
Auto print: To have a printout generated automatically when certain conditions are met, e.g., time elapsed, stability reached, etc.
Print net-total: For a printout of a component or total weight when using the “net-total” application
Print animal weights: For an automatic printout of animal weight, or of animal weight plus calculated weight after averaging
Auto print +/–: for automatic printout of a weight when it lies within preset limits at stability
Factory Settings
Print manually/automatically:
Printout generated manually (by pressing p) or automatically, depending on stability parameter:
Manual after stability (6 1 2)
Line format:
A data ID code of up to 6 characters preceding weighed or calculated values:
For other applications/GLP (7 2 2)
Print application parameters:
Print one or more of the initialization values for the current application program: Off (7 1 1)
ISO/GLP-compliant printout/ data record: Documentation of weighing conditions for each series of measurements/each lot:
Off (8 10 1)
Auto print:
Automatic printout of weighed/ calculated values: not a factory setting; see
“Print manually/automatically” (6 1 2)
Not possible to stop by pressing p (6 2 2)
After 1 display update (6 3 1)
Print net-total:
Printout of component weight (net weight) or total weight (tare weight):
Auto print net (7 3 1)
Print animal weights:
Automatic printout of animal weight, or of animal weight plus calculated result: On: animal wt. (3 9 2)
Auto print +/–:
Automatic printout of weights when within tolerance at stability: off (4 2 2)
● Parameter settings:
See “Configuring the Balance”
58
Operating the Balance
Print Manually/Automatically
The printout contains the current value in the measured value display
(weight readout with weight unit; calculated value; numeric/ alphabetic display)
+ 1530.000 g
+ 58.5620 ozt
+ 253 pcs
+ 88.23 %
+ 105.78 o
Line Format
The current value displayed can be printed with a data ID code of up to
6 characters at the beginning of the line. You can use this data ID code, e.g., to designate a weight readout as a net weight (N) or a calculated value as a piece count (Qnt)
ID ABC123DEF456GH
L ID ABC123DEF456GH
W ID ABC123DEF456GH
N + 1530.000 g
Qnt + 253 pcs
Prc + 88.23 %
Nom. + 2000.00 g
Sample ID
You can have each weighed or calculated value that you print preceded by a line of text containing numbers and/or letters. You can either print this ID immediately as an alphanumeric input (press p) or store it as the sample ID (S ID soft key)
S ID ABC123DEF456GH
ABC123DEF456GHI789JK
NUM 12345678
Print Application Parameters
You can generate a printout of one or more of the values configured for initialization of an application as soon as you initialize the balance.
This can include such values as nRef, wRef, pRef, etc.
nRef 10 pcs wRef 1.23456 g pRef 80 %
Wxx% 1200.00 g mDef 10
Mul 0.00347
Setp + 1000.035 g
Min + 981.054 g
Max + 1020.063 g
N1 + 278.11 g
T2 + 1821.48 g
Comp7 + 278.11 g
Tot.cp+ 2117.56 g
Weight in grams
Weight in Troy ounces
Piece count
Percentage
Calculated value
Identification number*
Lot number (weighing series)*
Weight set number*
Net value
Quantity
Percentage
Exact calibration weight
* = only on ISO/GLP-compliant printouts
Sample ID
(with less than 14 characters)
Sample ID
(with more than 14 characters)
Numeric key output when p is pressed
Counting: Reference sample quantity
Counting: Average piece weight
Weighing in percent: Reference percentage
Weighing in percent:
Reference weight
Animal weighing:
Number of subweighs for averaging
Animal weighing: multiplication factor
Over/under checkweighing:
Target weight
Over/under checkweighing:
Lower limit
Over/under checkweighing:
Upper limit
Net-total: net weight
Net-total: tare weight
Net-total: weight of 7th component
Net-total: total weight of components (only for ISO/GLPcompliant records)
59
Auto Print
You can have the weight readout
(or the value displayed in the measured value line) printed automatically; this printout can be generated after a certain number of display updates; you can also configure whether or not the auto-print function is dependent on the stability parameter. The display update frequency depends on both the model of the balance and the current operating state.
N + 1530.00 g
S ID 12345678901234
Stat
Stat L
Stat H
Print Net-Total
When you run the net-total application, you can have the weight of the last component weighed (net value) or the total weight (tare value) printed automatically.
Comp1 + 1821.48 g
Print Animal Weights
When using the animal weighing application, you can have the results printed automatically upon completion of the averaging process.
You can also have both the weight and the calculated result printed.
mDef 10
Mul 0.00347
x-Net + 1530.00 g x-Res + 5.30 o
Auto Print +/–
With the over/under checkweighing application, you can have the result printed automatically if it lies within a defined range (acceptable value).
N + 1530.000 g
Setp + 1000.035 g
Min + 981.054 g
Max + 1020.063 g
N + 1010.147 g
Data Output Functions
Printing a Data Record
Print for use in legal-for-trade applications:
You can configure the scale operating menu to generate a printout that conforms to the regulations for use in legal metrology (last digit marked) on a Sartorius printer:
– YDP01IS: 5 5 4
– YDP02:
5 5 5
– YDP03:
5 5 6
Net weight
Sample ID
Display blank
Display underload
Display overload
First net weight
Number of subweighs for averaging
Multiplication factor
Result of averaging
Calculated result
Net weight
Target weight
Lower limit
Upper limit
Printout of acceptable values
60
Operating the Balance
ISO/GLP-compliant Printout/Record
You can have the parameters pertaining to the ambient weighing conditions printed before (GLP header) and after (GLP footer) the values of a weighing series.
These parameters include:
– Date
– Time at the beginning of a weighing series
– Balance manufacturer
– Balance model
– Model serial number
– Software version
– Lot ID (weighing series no.)
– Time at the conclusion of the weighing series
– Field for operator signature
Operating the Balance with an
ISO/GLP-capable Documentation
Device (Printer)
You can connect a special printer from Sartorius to your balance for printing ISO/GLP-compliant records.
This printer, called “Data Printer”
(order no.: YDP03-0CE), offers the following features:
– ISO/GLP function can be switched on and off
– Date/time
– ID no. for identification of workstation/operator
– Documentation includes balancespecific data
Note:
The time indicated on the printout consists of only hours and minutes.
For ISO/GLP-compliant documentation with a computer, you will need special software.
Contact Sartorius for a detailed description for creating this software.
---------------------
28.06.1997 16:12
SARTORIUS
Mod. LP4200S
Ser. no. 60419914
Ver. no. 01-30-09
ID 12345678901234
---------------------
L ID 12345678901234 nRef 10 pcs wRef 1.35274 g
Qnt + 235 pcs
Qnt + 4721 pcs
S ID 12345678901234
Qnt + 567 pcs
---------------------
28.06.1997 16:13
Name:
---------------------
Dotted line
Date/time
(only with the YDP03-0CE printer)
Balance manufacturer
Balance model
Balance serial number
Software version
(display and control unit)
Balance ID no.
Dotted line
Weighing series no.
Application initialization value
Application initialization value
Counting result
Counting result
ID for counting result
Counting result
Dotted line
Date/time
(only with the YDP03-0CE)
Field for operator signature
Blank line
Dotted line
Blank line
Blank line
61
Interface Description
Purpose
Your Master pro
balance comes equipped with an interface port for connection to a computer or other peripheral device.
You can use an on-line computer to change, start and/or monitor the functions of the balance and the application programs.
The interface port also has four data output port lines for the over/under checkweighing program.
Features
Type of interface: Serial interface
Operating mode: Full duplex
Standard: RS-232
Transmission rates: 150; 300; 600; 1,200; 2,400; 4,800;
9,600; 19,200 baud
Parity:
Data output format of the balance:
Space, odd, even
Character format: 1 start bit, 7-bit ASCII, parity, 1 or 2 stop bits
Handshake: 2-wire interface: via software (XON/XOFF);
4-wire interface: via hardware handshake lines (CTS/DTR)
16 or 22 characters
!
Warning When Using Pre-wired
RS-232 Connecting Cables!
RS-232 cables purchased from other manufacturers often have incorrect pin assignments for use with Sartorius balances. Be sure to check the pin assignment against the chart on the right before connecting the cable, and disconnect any lines marked
“Internally Connected” (e.g., pin 6).
Failure to do so may damage or even completely ruin your balance and/or peripheral device.
Factory settings:
Transmission rate:
Parity:
Stop bits:
Handshake:
1,200 baud
Odd
1 stop bit
Hardware
1 character after CTS
Print manually/automatically:
Stop automatic printing:
Manual with stability
Not possible
Automatic printout, time-dependent: After 1 display update
Line format: For other applications/GLP
Preparation
● See page 68 for the pin assignment chart
(5 1 4)
(5 2 3)
(5 3 1)
(5 4 3)
(6 1 2)
(6 2 2)
(6 3 1)
(7 2 2)
62
Operating the Balance
Line Format (Data Output Format)
You can output the values displayed in the measured value line and the weight unit with or without a data ID code
Example: Without data ID code
+ 253 pcs
Example: With data ID code
Qnt + 253 pcs
Configure this parameter in the Setup menu under item 7 2.
The output with data ID code has
16 characters; without data ID code,
22 characters.
*:
D:
U:
CR:
LF:
Output Format with 16 Characters
Display segments that are not activated are output as spaces.
Characters without a decimal point are output without a decimal point.
The following characters can be output, depending on the characters displayed on the balance:
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * D D D D D D D D * U U U CR LF or or
–
*
.
.
.
.
.
.
.
.
* * * * * * * *
* * * or 0 0 0 0 0 0
Space
Digit or letter
Unit symbol
Carriage return
Line feed
Special Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * – – * * * * * * CR LF or or or
H H
L
C
L
*:
– –:
Space
Weight; all numbers shown in stable readout
H: Overload
H H: Overload during checkweighing
L:
L L:
C:
Underload
Underload during checkweighing
Calibration/adjustment
Error codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * E r r * # # # * * * * CR LF
*: Space
# # #: Error code number
63
Data output example: + 1255.7 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * * * 1 2 5 5 .
7 * g * * CR LF
Position 1:
Position 2:
Positions 3 – 10:
Position 11:
Positions 12 – 14: Unit symbol or space
Position 15: Carriage return
Position 16:
Data Output with an ID Code
When data with an ID code is output, the ID code consisting of 6 characters precedes the data with the 16-character format.
These 6 characters identify the following value.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
I I I I I I + * D D D D D D D D * U U U CR LF
* * * * * –
*
. .
.
.
.
.
.
.
* * * * * * * *
* * *
0 0 0 0 0 0
I: ID code character
1)
*: Space
D: Digit or letter
U: Unit symbol
1)
CR: Carriage return
LF: Line feed
1) depends on balance type; e.g., not all units and characters are available on balances verified for use in legal metrology
Special Codes
Error codes
Plus or minus sign or space
Space
Weight with a decimal point; leading zeros = space
Space
Line feed
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * * * * – – * * * * * * CR LF
H H
*: Space
– –: Weight; all numbers shown in final, stable readout
H: Overload
H H: Overload during checkweighing
L L
C
L: Underload
L L: Underload during checkweighing
C: Calibration/adjustment
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * E r r * # # # * * * * CR LF
*: Space # # #: Error code number
Unit symbol U
1) g k g c t l b o z
No stability parameter
Grams
Kilograms
Carats
Pounds
Ounces o z t Troy ounces t l h Hong Kong taels t l s Singapore taels t l t Taiwanese taels
G N Grains d w t Pennyweights m g Milligrams
/ l b Parts per pound t l c Chinese taels m o m Mommes
K Austrian carats t o l Tola b a t Baht
M S Meshgal
% Percent p c s Pieces (piece count)
ID code characters
1)
Stat
T2
N
Status
Application tare memory
Net weight (T1 = 0)
N1
Qnt
Prc nRef pRef wRef
Wxx%
NUM
Net weight (T1 0)
Quantity
Percentage
Reference sample quantity
Reference percentage
Average piece weight
Reference percentage weight
Numeric input
Compxx
Component no. xx in net-total component mode
Tot.cp
Total weight in net-total component mode mDef
No. of subweighs for animal weighing
Mul
Setp
Min
Multiplication factor for animal weighing x-Net
Result in animal weighing x-Res
Calculated result in animal weighing
Target value for checkweighing
Lower limit for checkweighing
Max
Upper limit for checkweighing
64
Operating the Balance
Data Input Format
You can connect a computer to your balance to send commands via the balance interface port to control balance functions and applications.
The commands sent are control commands and may have different formats; e.g., control commands can have up to 26 characters. Each character must be transmitted according to the settings configured in the Setup menu for data transmission.
Farmat for Control Commands
Format 1: Esc
Format 2: Esc
Format 3: Esc
Format 4: Esc
Esc:
!:
#:
&:
Escape
Number
!
!
!
!
CR
#
#
_
Number or letter
LF
_
&
CR
Command character
CR
(max. 20 &)
LF
LF
_:
CR:
& _ CR
Underline (ASCII: 95)
Carriage return
LF: Line feed max: Depends on command character; i.e., parameter: once the max. length is reached, input received via the interface port is cut off, rather than rejected as for a key input
LF
Format 1
!
Meaning
K Weighing mode 1
L Weighing mode 2
M Weighing mode 3
N Weighing mode 4
O Block keys
P Print
R Unblock keys
S Restart
T Tare t
Z Internal calibration/adjustment
Format 2
!# Meaning f0 Function key w f1 Function key q f2 Function key v f3 Zero (separate zeroing key) f4 Tare (separate zeroing key) s3 Function key c x0 Perform internal calibration x1 Print balance model x2 Print weighing cell serial number x3 Print weighing cell software version x4 Print display and control unit software version x5 Print balance ID number x6 Print weight set number x7 Print lot number
(weighing series ID)
Format 3
(not allowed in the Setup menu)
!# Meaning z5 Input balance ID number z6 Input weight set number z7 Input lot number t
Format 4
!
Meaning
Text input in display
65
Synchronization
During data communication between the balance and an on-line device
(computer), messages consisting of ASCII characters are transmitted via the interface. For error-free data communication, the parameters for baud rate, parity, handshake mode and character format must be the same for both units.
You can set these parameters in the Setup menu so that they match those of the on-line device. You can also define parameters in the balance to make data output dependent on various conditions.
The conditions that can be configured are described under each of the application program descriptions.
If you do not plug a peripheral device into the balance interface port, this will not generate an error message.
Handshake
The balance interface
(Sartorius Balance Interface = SBI) has transmit and receive buffers.
You can define the handshake parameter in the Setup menu:
– Hardware handshake (CTS)
– Software handshake
(XON, XOFF)
Hardware Handshake
With a 4-wire interface,
1 more character can be transmitted after CTS (Clear to Send).
Software Handshake
The software handshake is controlled via XON and XOFF.
When a device is switched on,
XON must be transmitted to enable any connected device to communicate.
When the software handshake is configured in the Setup menu, the hardware handshake becomes active after the software handshake.
The data transmission sequence is as follows:
Transmitting Device:
Once XOFF has been received, it prevents further transmission of characters. When XON in received, it re-enables the transmitting device to send data.
Receiving Device:
XOFF is transmitted after the
26th character has been stored.
To prevent too many control commands from being received at one time, XON is not transmitted until the buffer has transmitted all but
14 characters.
If the device addressed does not understand the control command, the SBI receiving device activates
DTR (Data Terminal Ready) after 6 more characters have been received. The busy signal is deactivated by XON (14 characters).
Balance ––– byte –––> Computer
(trans––– byte –––> (receiving mitting ––– byte –––> device) device) ––– byte –––>
<––– XOFF –––
––– byte –––>
––– byte –––>
…
(Pause)
…
<––– XON –––
––– byte –––>
––– byte –––>
––– byte –––>
––– byte –––>
66
Activating Data Output
You can define the data output parameter so that output is activated either when a print command is received or automatically and synchronous with the balance display or at defined intervals
(see application program descriptions and auto print settings).
Data Output by Print Command
The print command can be transmitted by pressing p or by a software command (Esc P).
Automatic Data Output
In the “auto print” operating mode, data are output to the interface port without a print command. You can choose to have data output automatically at defined print intervals with or without the stability parameter. Whichever parameter you select, the data will be output as the readouts appear on the balance display. The display update frequency depends on both the model of the balance and the current operating state.
If you select the auto print setting, data will be transmitted immediately the moment you turn on the balance.
In the Setup menu, you can configure whether this automatic output can be stopped and started by pressing p.
Operating the Balance
67
Pin Assignment Chart
Female Interface Connector:
25-position D-Submini, DB25S, with screw lock hardware for cable gland
Male Connector Required:
(please use connectors with the same specifications)
25-pin D-Submini, DB25S, with integrated shielded cable clamp assembly (Amp type 826 985-1C) and fastening screws
(Amp type 164 868-1)
Pin Assignment Chart:
Pin 1: Signal Ground
Pin 2: Data Output (TxD)
Pin 3: Data Output (RxD)
Pin 4: Signal Return (TxD/RxD)
Pin 5: Clear to Send (CTS)
Pin 6: Internally Connected
Pin 7: Internal Ground
Pin 8: Internal Ground
Pin 9: Reset _ In**)
Pin 10: – 12 V
Pin 11: + 12 V
Pin 12: Reset _ Out**)
Pin 13: + 5 V
Pin 14: Internal Ground
Pin 15:
Pin 16: p key/ t key/ q key/
UniversalControlkey*) output 1*) output 2*)
“lighter”
Control-
“equal”
Pin 17:
Pin 18:
Pin 19:
Pin 20: Data Terminal Ready (DTR)
Pin 21: Supply Voltage Ground “COM”
Pin 22: Not Connected
Pin 23: Not Connected
Pin 24: Supply Voltage Input + 15 … 25 V
Pin 25: +5 V v key/
Controloutput 3*)
“heavier” c key/
Controloutput 4*)
“set”
*) = See “Additional Functions” for information on changing pin assignments
**) = Hardware restart
68
Operating the Balance
Cabling Diagram
– Diagram for interfacing a computer or different peripheral device to the balance using the RS-232/V24 standard and cables up to 15 m (50 ft.) long
V24
Balance Peripheral Device (PC)
PC
25-pin or 9-pin
2 3
20
3
5
6
7
4
2
8
6
5
Type of cable: AWG 24 specification
69
Additional Functions
Password
You can block access to parameter settings in the Setup menu and to the ID-code input function, as well as to the exact calibration weight, by assigning a password.
Enter the password by selecting
INPUT in the Setup menu.
See “Configuring the Balance” for a detailed description.
Protecting Menu Parameters
In the Setup menu, you can define whether menu parameters are:
– accessible for changes
(alterable, 8 1 1)
– can be read only (readable,
8 1 2)
Acoustic Signal
An acoustic signal is emitted when you press a key. When the key pressed is allowed, the signal is a single beep tone; when it is not allowed, this is signaled by a double beep (key does not initiate a function). In the Setup menu, you can configure whether
– the acoustic signal should sound
(on, 8 2 1), or
– the acoustic signal should not sound (off, 8 2 2)
Blocking the Keys
When you operate the balance via an on-line computer, it is advisable that you block the keys on the display and control unit. In the Setup menu, you can configure whether
– the keys are active
(Keys unblocked, 8 3 1),
– the keys are blocked
(Keys blocked, 8 3 2), except for M and e, or
– the alphanumeric keys are blocked
(alphanum. blocked, 8 3 3).
Power-On Mode
You can configure the balance so that once a power supply is connected,
– the balance is turned off
(Off/on/standby, 8 6 1 or Off/on, 8 6 2), or
– the balance switches on automatically (Auto on, 8 6 4)
You can also set the configurations so that when the balance is switched off after use, it is
– off – not in balances with a weighing capacity
≥
16 kg
(Off/on, 8 6 2), or
– in the standby mode
(Off/on/standby, 8 6 1)
Universal Switch for Remote Control
You can connect an external universal switch to the interface port of your balance (e.g., a foot switch) for remote control of the functions listed below. In the Setup menu, you can configure which function is to be controlled via remote switch:
– Printing (Print key funct.,
8 4 1)
– Taring
(Tare key function, 8 4 2)
– Calibration
(CAL key function, 8 4 3)
– Function key
(F key function, 8 4 4)
– Delete/cancel
(CF key function, 8 4 5)
– Toggling
(Toggle key funct., 8 4 6)
When you turn on the balance, a self-test of the functions is run (test is displayed; the bar graph is shown)
After the self-test has been completed, the weighing range of the balance is displayed (line for metrological data shows different maximum values)
Automatic Shutoff
When parameter 8 6 2 is selected in the Setup menu, you can configure whether
– the balance will shut off automatically after 4 minutes without use
(After 4 min., 8 7 1), or
– automatic shutoff is deactivated
(off, 8 7 2)
Display Backlighting
You can have the display backlighted for improved readability of displayed values. In the Setup menu, you can configure whether the
– display backlighting is on
(on, 8 5 1),
– display backlighting is off
(off, 8 5 2), or
– display backlighting switches off automatically if there is no change in the readout for at least 4 minutes
(Auto off aft. 4 min., 8 5 3)
70
Operating the Balance
Interface Port Input/Output
You can connect a checkweighing display and a remote universal switch to the interface port (factory setting).
When you connect a remote universal switch, you need to change the following parameters.
Undoing All Parameter Changes –
Reset Function
There is a factory setting for each parameter. In the Setup menu, you can configure whether
– menu factory settings will be restored after exiting Setup
(Restore, 9 1 1), or
– menu factory settings will not be restored after exiting Setup
(Do not restore, 9 1 2)
Pin Assignments for the Female Interface Connector
Pin Input Function (8 8 1)
15 p key
16 t key
17 q
key
18 v
key
19 c key
Output Function (8 8 2)
Remove universal switch (see below)
Control output port 1: lighter
Control output port 2: equal
Control output port 3: heavier
Control output port 4: “set”
Remote universal switch
Function p
key t
key q
key v key c
key w
key
Menu setting
8 4 1
8 4 2
8 4 3
8 4 4
8 4 5
8 4 6
See “Pin Assignment Chart” in the chapter entitled “Overview” for detailed information.
Printing an ISO/GLP-compliant
Record
In the Setup menu, you can configure whether
– no ISO/GLP-compliant record will be printed (off, 8 10 1),
– an ISO/GLP-compliant record will be printed after calibration/ adjustment (Only for cal./adj.,
8 10 2), or
– every printout will be an
ISO/GLP-compliant record
(always ON, 8 10 3)
71
MP8 Interface Emulation
Purpose
With the MP8 interface emulation you can connect peripheral devices of the MP8 generation that have separate AC power supplies, such as the 73822... Data Control terminal, a YFC... Flow Rate
Controller, a YDI50Z Data Input dedicated keyboard, etc., to your
LP series balance.
Available Features
– The balance can only be used to determine weights
– The interface communicates exclusively in the MP8 binary protocol.
– Select application programs for use with the MP8 under item 3 in the balance operating menu.
– The Index 2 program for MP8 can be selected under item 4 of the balance operating menu
– The following parameters remain accessible as before:
– Weighing parameters [1-x-x]
– Extra functions [8-x-x]
– Reset function [9-x-x]
(see “Setting Parameters (Menu)” in the chapter entitled “Configuring the Balance”)
– “MP8 interface” not allowed during use in legal metrology.
When the menu access switch is sealed, the MP8 interface will not function.
Factory Settings of the Parameters
(special settings for MP8 functions)
Program selection: MP8: 3-1-1
Program index 2: 1 Ind. 2.1
72
Preparation
● Turn on the balance:
Press e
> All display segments light up briefly
Switch to the MP8 interface:
● Press M
● Select balance operating menu: Menu soft key (press the w key)
● Select and confirm the balance menu: press the y and then the X key
● Confirm factory setting: X key
● Select and confirm MP8 mode [9-1-9]
Press y or Y key, repeatedly if necessary; then press X
● Press M
Parameter Settings for the MP8 Interface
● Press M
● Select the balance operating menu: Menu soft key (press the w key)
● Select and confirm:
– 3 Application prog. 1 app. selection:
1 MP8 3-1-1 or
...
9 MP8 3-1-9 or
10 MP8 3-2-1 or
...
18 MP8 3-2-9 or
19 MP8 3-3-1 or
...
27 MP8 3-3-9
– 4 Program-Index; 2 Ind. 2:
1 Ind. 2.1 or
2 Ind. 2.2 or
3 Ind. 2.3 or
4 Ind. 2.4
– 5 Interface:
1 Baud rate
1 150 Baud or
2 300 Baud or
3 600 Baud or
4 1,200 Baud or
5 2,400 Baud or
6 4,800 Baud or
7 9,600 Baud or
2 Parity
2 Space or
3 Odd or
4 Even
– 6 Print weighing
1 print Man./auto
1 Manual without stability or
2 Manual with stability or
4 Automatic without stability or
5 Automatic at stability
§ Store settings and exit the Setup menu: Press M
Error Codes
Error codes are displayed in the main display or application display for 2 seconds.
The program then returns automatically to the previous mode (e.g., weighing).
Display Cause
No segments appear on the display No AC power is available
The AC adapter is not plugged in h l
or err 54 err 01
> display range err 02 cal.n.possible
err 03 cal./adj. interrupt err 06 int. wt. defective err 07 function blocked err 08 *
<> zero range err 09 *
< 0 not allowed err 10 tare fct. blocked
Solution
Check the AC power supply
Plug in the AC adapter
Automatic shutoff configured in the Setup menu (code 8 7 1)
Press e to turn on the balance or select code 8 7 2 in the Setup menu (“no automatic shutoff”)
Unload the balance The load exceeds the balance’s capacity
The weighing pan is not in place
Data output not compatible with output format
Calibration/adjustment criterion not met, e.g.,
– not tared
– the balance is loaded
Calibration/adjustment could not be completed within a certain time
Built-in calibration weight is defective
Place the weighing pan on the balance
Change the configuration in the
Setup menu
Calibrate only when zero is displayed
Press t to tare
Unload the balance
Allow the balance to warm up again and repeat the adjustment process
Contact your local
Sartorius Service Center
Function not allowed in balances verified for use in legal metrology
The load on the balance is too heavy to zero the readout
Contact your local Sartorius Service
Center for information on having the settings changed
Check whether the
“power-on zero range” is set
Taring is not possible when the gross Zero the balance weight is
≤
zero
The tare key is blocked when there is data in the tare memory
(e.g., when running the net-total application); the tare functions cannot be accessed simultaneously
Press c to clear the tare memory; then you can tare by pressing t
Tare memory not allowed Check the tare value entered err 11 tare 2 blocked err 12 tare > max.
err 17 adj.-wt. >max.
Tare stored in memory greater than weighing range or range limits
Internal adjustment is not possible because the preload is too heavy
Check sample/container
Reduce the preload or change the configuration
* = occurs only when balance is operated via the SBI interface (ESC f3_/f4_)
73
Display err 30 print fct. blocked ref.wt. too light upd. not possible no num. value xxxxx too low xxxxx too high too many characters err 10
All segments displayed continuously err 340
No WP x = x = x = x =
1
:
2
:
3
:
4
: blocked
The special code b remains displayed
The weight readout changes constantly
Cause
Interface port for printer output is blocked
Error in storing reference weight
(with the counting or weighing-inpercent application)
Reference sample updating not possible (with the counting application)
Input wrong
(for any application program), e.g., alphabetic input not allowed
Input text too long
Solution
Contact your local Sartorius
Service Center
Weight too light or there is no sample on the balance
See “Counting” in “Operating the Balance” for reference updating criteria
Follow the instructions for the application programs
Key is stuck
Key pressed when turning on the balance: c
, q, w, v
0
, 3, 4, 9
2
, 5, 6, p, t – right
1
, 7, 8, a, t – left
Either M was pressed when you turned on the balance, or this key is stuck
Operating parameter (EEPROM) is wrong
Weighing cell (platform) defective
Function blocked
None of the keys has been pressed since the balance was turned on
Unstable ambient conditions
Allowable text lengths, incl.
decimal point:
– S ID and L ID: 20 characters max.
– W ID: 14 characters max.
for weights
Contact your local Sartorius
Service Center
Release key
Contact your local Sartorius
Service Center
Contact your local Sartorius
Service Center
None
Press a key
Set up the balance in another area
Too much vibration, or the balance is exposed to a draft
Change Setup configurations to adapt the balance to the ambient conditions
Remove the foreign object
Calibrate/adjust the balance
A foreign object is caught between the pan and the balance housing
The weight readout is obviously wrong The balance has not been calibrated/adjusted
The balance was not tared before weighing
The balance is not level
The dust cover is caught under the weighing pan
If any other errors occur, contact your local Sartorius Service Center!
Tare before weighing
Level the balance
See “Replacing the Dust Cover” in the chapter “Care and Maintenance”
74
Care and Maintenance
Service
Regular servicing by a Sartorius technician will extend the service life of your balance and ensure its continued weighing accuracy.
Sartorius can offer you service contracts, with your choice of regular maintenance intervals ranging from
1 month to 2 years.
Repairs
Repair work must be performed by trained service technicians.
Any attempt by untrained persons to perform repairs may lead to hazards for the user.
Cleaning
!
Make sure that no dust or liquid enters the balance housing
!
Do not use any aggressive cleaning agents
(solvents or similar agents)
● Unplug the AC adapter from the wall outlet (mains supply)
● If you have an interface cable connected to the balance port, unplug it from the port
● Carefully remove any sample residue/spilled powder by using a brush or a hand-held vacuum cleaner
● Clean the balance using a piece of cloth which has been wet with a mild detergent (soap)
●
After cleaning, wipe down the balance with a soft, dry cloth
Replacing the Dust Cover
> Instructions for replacing a damaged dust cover
For LP Series Balances with a Round
Glass Draft Shield
● Remove the following parts from the balance:
– Draft shield cover
– Glass draft shield cylinder
– Weighing pan
– Pan support
– Shield disk: turn clockwise and lift off
– Old dust cover
● Place the new dust cover on the balance and press down on the front and back along the edges until it is seated firmly
● Place the shield disk on the balance and turn it counterclockwise
●
Follow the above instructions in reverse order when placing remaining parts back on the balance.
For LP Series Balances with a Rectangular Weighing Pan and a Weighing Capacity
≤
12 kg
● Remove the following parts from the balance:
– Weighing pan
– Pan draft shield
(depending on balance model)
– Old dust cover
● Place the new dust cover over the balance
● Follow the above instructions in reverse order when placing remaining parts back on the balance.
!
The dust cover must not touch the weighing pan
Safety Inspection
If there is any indication that safe operation of the balance with the AC adapter is no longer warranted:
● Turn off the power and disconnect the equipment from AC power immediately
> Lock the equipment in a secure place to ensure that it cannot be used for the time being
Safe operation of the balance with the AC adapter is no longer ensured when:
– there is visible damage to the AC adapter
– the AC adapter no longer functions properly
– The AC adapter has been stored for a relatively long period under unfavorable conditions
In this case, notify your nearest
Sartorius Service Center or the
International Technical Support Unit based in Goettingen, Germany.
Maintenance and repair work may only be performed by service technicians who are authorized by Sartorius and who
– have access to the required maintenance manuals
– have attended the relevant service training courses
75
Replacement of the Backup Battery
Spare backup battery in the display and control unit (soldered to PCB)
Type: VL 2020; manufacturer:
Panasonic. To be replaced by trained service technicians only.
CAUTION:
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the manufacturer.
Dispose of used batteries according to the manufacturer’s instructions.
Instructions for Recycling
To ensure adequate protection for safe shipment, the Master pro
balance has been packaged to the extent necessary using environmentally friendly materials. After successful installation of the balance, you should return this packaging for recycling.
For information on recycling options, including recycling of old weighing equipment and disposal of used batteries, contact your municipal waste disposal center or local recycling depot.
76
Overview
General Views of the Balances
LP1200S (-0CE), LP620S (-0CE), LP220S (-0CE), LP620P (-0CE), LP3200D
Overview
No.
Designation Order no.
for spare part
1 1 Draft shield cover
1
1
5
6
Shield disk
Leveling foot
1 7 Display and control unit
1 8 Metrological ID label (only on
69 LP0002
1 2 Glass draft shield cylinder
1 3 Weighing pan
69 14290
69 LP0004
1 4 Pan support – LP 3200D: 69 LP0006
– LP 1200S, LP 620, LP 220S: 69 LP0005
69 LP0003
69 B20005 verified models or models acceptable for legal metrological verification)
1 9 Lug for attaching an antitheft locking device
10 AC jack
No.
Designation Order no.
for spare part
11 Tare key
12 Function keys
13 Print key
14 Keys for numeric input
15 Toggle key for alphabetic input
16 On/off key
17 Display
18 Interface port
19 Level indicator
Not shown:
Dust cover
Protective caps and plugs (set)
69 60LP01
69 B20009
77
General Views of the Balances
LP8200S (-0CE), LP8200P (-0CE), LP6200S (-0CE), LP4200S (-0CE), LP2200S (-0CE), LP820 (-0CE), LP420,
LP2200P (-0CE), LP5200P (-0CE), LP12000S (-0CE), LP6200 (-0CE), LP4200, LP2200 (-0CE), LP12000P (-0CE)
1
2
5
6
3
4
7
8
9
No.
Designation Order no.
for spare part
Weighing pan
Pan draft shield
(depending on balance model)
Shock absorber
69 LP0007
69 LP0008
Leveling foot
Display and control unit
69 LP0010
69 B20005
Metrological ID label (only on verified models or models acceptable for legal metrological verification)
Lug for attaching an antitheft locking device
AC jack
Tare key
78
No.
Designation Order no.
for spare part
10 Function keys
11 Print key
12 Keys for numeric input
13 Toggle key for alphabetic input
14 On/off key
15 Display
16 Interface port
17 Level indicator
Not shown:
Dust cover
Protective caps and plugs (set)
69 60LP02
69 B20009
General Views of the Balances
LP64001S, LP34001S, LP16001S, LP34001P, LP34000
Overview
4
5
2
3
No.
1
Designation Order no.
for spare part
Weighing pan: models LP64000S,
LP34000P, LP34
Weighing pan:
69 LC0107 models LP64001S, LP34001S,
LP16001S, LP34001P, LP34000 Available
Leveling foot
Display and control unit on request
69 LC0093
Level indicator
Tare key
No.
Designation
6 Function keys
7 Print key
8 Keys for numeric input
9 Toggle key for alphabetic input
10 On/off key
11 Display
12 Interface port
13 AC jack
14 Metrological ID label (only on verified models or models acceptable for legal metrological verification)
79
Description of the Keys
e
Key On/off switch
Switches the display on/off.
The balance remains in the standby mode.
v Key
Start an Application
Further instructions on running the applications when this key is pressed are contained in the chapter entitled “Operating the Balance;” refer to the section pertaining to the particular program.
M
Key Settings for Configuring the Balance
– Access to the Setup menu
– Stores settings and exits Setup menu
You can select: info
Display basic information about the equipment
(e.g., model name, serial no., software version) menu
Balance operating menu with plain English prompts for adapting the balance to individual requirements input
For entering identifying information (e.g., balance ID) x
For moving around within the parameter submenus for Info, Menu and Input in the Setup menu.
t
y Y X Keys
Keys Tare
Two large keys for initiating the tare function. Ideally situated for both left-handed and right-handed operation. Sets the readout to zero.
With balances that have the
“PolyRange” weighing range structure, the fine range is available when this key is pressed.
c
Key Clear
This key is generally used to interrupt/cancel functions:
– Delete keyboard input and clear memory
– Interrupt calibration/adjustment routines
– Return application program to previous status a Key
Press this key to enter alphabetic characters and/or special characters
(*, /, space, etc.).
1 2
…9 0 Keys
For numeric input
.
Key
Define the decimal point position q
Key Calibration/Adjustment
Press this key to select and start calibration/adjustment functions.
p Key Data Output
Press this key to output data via the interface to a Sartorius
Data Printer or a computer.
w Key Toggle
This key toggles the display readout between a weight and a calculated value (counting, readout in percent, calculated result)
80
Menu Structure
Overview
* = setting cannot be changed in balances used for legal metrology
81
Specifications
General Specifications
AC power source/power requirements
Frequency
Allowable ambient operating temperature
Operating temperature range
Dust and water protection rating according to EN 60529*
AC adapter, 230 or 115 V, +15%… – 20%
48 – 60 Hz
0…+40
°
C (273…313 K, 32
°
F…104
°
F)
+ 10…+ 30
°
C
IP54 (protected against harmful dust deposits and splashes of water)
Adaptation to ambient conditions
Display update
(depends on the filter level selected)
Power consumption
Hours of operation with fully charged
YRB 06 Z external battery pack, approx.
Selectable weight units
By selection of 1 of 4 optimized filter levels
0.1 – 0.4
16 VA: maximum; 9 VA: average
Selectable application programs
Built-in interface
14 h
Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels, Singapore taels,
Taiwanese taels, grains, pennyweights, milligrams, parts per pound, Chinese taels, mommes,
Austrian carats, tola, baht and mesghal
Mass unit conversion, counting, weighing in percent, net-total formulation, animal weighing, over/under checkweighing, recalculation
RS-232 C
Format:
Parity:
Transmission rates:
Handshake:
7-bit ASCII, 1 start bit, 1 or 2 stop bits odd, even or space
150 to 19,200 baud
Software or hardware mode
* = specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories”
Specifications of the Individual Models
Model
Readability g
LP1200S
0.001
Weighing capacity
Tare range (subtractive)
Repeatability (standard deviation)
Linearity
Sensitivity drift within +10…+30
°
C
Response time (average)
External calibration weight
(of at least accuracy class...)
Other allowable external calibration weights
(of at least accuracy class…)
Pan size
Dimensions (W x D x H)
Net weight, approx.
g 1,200 g – 1,200
≤± g 0.001
≤± g 0.002
≤±
/K 2 • 10
–6 s g
1.5
1,000 (E2) g –
LP620S
0.001
620
– 620
0.001
0.002
500 (E2)
300, 400,
600 (E2) mm Ø 130 mm 240x360x147 kg 8.3
6.9
LP220S
0.001
220
– 220
0.001
0.002
200 (E2)
100 (E2)
6.9
LP620P
0.001/0.002/
0.005
120/240/620
– 620
0.001/0.001/
0.003
0.002/0.002/
0.005
LP3200D
0.001/0.01
1,010/3,200
– 3,200
0.001/0.01
0.002/0.01
500 (F1) 1,000 (E2)
200, 300, 400, 2,000, 3,000 (E2)
600 (F1)
6.9
8.4
82
Overview
Model
Readability
Weighing capacity
Other allowable external calibration weights
(of at least accuracy class...) g g
LP8200S
0.01
8,200
Tare range (subtractive) g – 8,200
Repeatability (standard deviation)
Linearity
≤± g 0.01
≤± g 0.02
Sensitivity drift within +10…+30
°
C
≤±
/K 2 • 10
–6
Response time (average)
External calibration weight
(of at least accuracy class...) s g
2
5,000 (E2) g
– 8,200
0.01/0.01/0.03
0.02/0.02/0.05
6,000; 7,000; –
8,000 (E2)
LP8200P
0.01/0.02/0.05
LP6200S
0.01
2,000/4,000/8,200 6,200
2
5,000 (F1)
– 6,200
0.01
0.02
1.5
5,000 (E2)
6,000 (E2)
Pan size
Dimensions (W x D x H)
Net weight, approx.
mm mm kg
218x200
240x360x86
6.5
LP4200S
0.01
4,200
– 4,200
0.01
0.02
1.5
2,000 (E2)
3,000;
4,000 (E2)
Model
Readability
Weighing capacity g g
LP820
0.01
820
Tare range (subtractive)
Repeatability (standard deviation)
Linearity g – 820
≤± g 0.01
≤± g 0.01
Sensitivity drift within +10…+30
°
C
≤±
/K 2 • 10
–6
Response time (average)
External calibration weight
(of at least accuracy class…) s g
1.5
500 (F2)
Other allowable external calibration weights
(of at least accuracy class…)
Pan size g mm
600; 700;
800 (F2)
218x200
Dimensions (W x D x H)
Net weight, approx.
mm kg
240x360x86
6.5
LP420
0.01
420
– 420
0.01
0.01
200 (F2)
300; 400 (F2)
LP2200P LP5200P
0.01/0.02/ 0.01/0.02/
0.05
0.05/0.1
400/800/
2,200
– 2,200
1,200/2,400/
3,800/5,200
– 5,200
0.01/0.01/ 0.01/0.02/
0.03
0.05/0.05
0.02/0.02/ 0.02/0.02/
0.05
0.05/0.1
2,000 (F2)
1,000 (F2)
2,000 (F1)
3,000; 4,000;
5,000 (F1)
LP2200S
0.01
2,200
– 2,200
0.01
0.02
1.5
2,000 (F1)
1,000 (F1)
Model
Readability
Weighing capacity g g
LP12000S
0.1
12,000
Tare range (subtractive)
Repeatability (standard deviation) g
≤± g
– 12,000
0.05
Linearity
≤± g 0.2
Sensitivity drift within +10…+30
°
C
≤±
/K 4 • 10
-6
Response time (average) s 1 kg 5 (F1) External calibration weight
(of at least accuracy class…)
Other allowable external calibration weights
(of at least accuracy class…)
Pan size
Dimensions (W x D x H) kg mm mm
6 to
12 (F1)
218x200
240x360x86
Net weight, approx.
kg 6.5
LP6200
0.1
6,200
– 6,200
0.05
0.1
4 • 10
-6
5 (F2)
4; 6 (F2)
LP4200
0.1
4,200
– 4,200
0.05
0.1
4 • 10
-6
2 (F2)
3; 4 (F2)
LP2200
0.1
2,200
– 2,200
0.05
0.1
2 • 10
-6
2 (F2)
1 (F2)
LP12000P
0.1/0.2/0.5
3,000/6,000/
12,000
– 12,000
0.1/0.1/0.3
0.1/0.2/0.5
4 • 10
-6
5 (F2)
6; 7; 8; 9; 10;
11; 12 (F2)
83
Model
Readability
Weighing capacity
Tare range (subtractive)
Repeatability (standard deviation)
Linearity
Sensitivity drift within
+10…+30
°
C
Response time (average)
External calibration weight
(of at least accuracy class…)
Other allowable external calibration weights
(of at least accuracy class…)
Pan size
Dimensions (W x D x H)
Net weight, approx.
g g
LP16000S
0.1
16,000 g – 16,000
≤± g 0.1
≤± g 0.2
≤±
/K 3 • 10
-6 s kg
1.5
10 (F1) kg 11 to
16 (F1) mm mm kg
307x417
307x538x121
13.8
LP34000P
0.1/0.2/0.5
8,000/16,000/
34,000
– 34,000
0.1/0.2/0.5
0.2/0.2/0.5
2 • 10
-6
1.5
10 (F2)
15; 20; 25;
30 (F2)
307x417
307x538x121
13.8
Model
Readability
Weighing capacity
Tare range (subtractive)
Repeatability (standard deviation)
Linearity
Sensitivity drift within +10 ... +30
°
C
Response time (average)
External calibration weight
(of at least accuracy class...)
Other allowable external calibration weights
(of at least accuracy class...)
Pan size
Dimensions (W x D x H)
Net weight, approx.
g g
LP64001S
0.1
64,000 g – 64,000
≤± g 0.1
≤± g 0.5
≤±
/K 3 • 10-6 s 1.5
kg 10 (F1) kg mm mm kg
LP34001S
0.1
34,000
– 34,000
0.1
0.2
5, 20, 25,
30 (F1)
300 x 400
313x534x120
16.0
2 • 10-6
1.5
10 (F1)
15, 20, 25,
30 (F1)
LP16001S
0.1
16,000
– 16,000
0.05
0.2
LP34001P
0.1/0.2/0.5
LP34000
1
8,000/16,000/ 34,000
34,000
– 34,000 – 34,000
0.05/0.05/0.1
0.5
0.2
0.5
2 • 10-6
1.5
10 (F1)
2 • 10-6
1.5
10 (F2)
11, 12, 13, 15, 20, 25,
14, 15, 16 (F1) 30 (F2)
2 • 10-6
1
10 (F2)
15, 20, 25,
30 (F2)
LP34
1
34,000
– 34,000
0.5
1
2 • 10
-6
1
10 (F2)
15; 20; 25;
30 (F2)
307x417
307x538x121
13.8
84
Overview
Verified Models with EC Type Approval
General Specifications
AC power source/power requirements
Frequency
Dust and water protection rating according to EN 60529
1
)
AC adapter, 230 or 115 V, + 15%... –20%
48 – 60 Hz
IP54 (protected against harmful dust deposits and splashes of water)
Adaptation to ambient conditions
Display update
Power consumption
Hours of operation with fully charged
YRB 06 Z external battery pack, approx.
By selection of 1 of 4 optimized filter levels
0.1 – 0.4 (depends on the filter level selected)
16 VA: maximum; 9 VA: average
14 h
Selectable weight units
Selectable application programs
Grams, kilograms
Mass unit conversion, counting, weighing in percent, net-total formulation, animal weighing, over/under checkweighing, recalculation
RS-232 C Built-in interface
Format:
Parity:
Transmission rates:
Handshake:
1
) specially protected dust-tight and washdown resistant AC adapter; see the section on “Accessories”
7-bit ASCII, 1 start bit, 1 or 2 stop bits odd, even, or space
150 to 19,200 baud
Software or hardware mode
Specifications of the Individual Models
Model
Type
Accuracy class*
Scale interval, d* g
Maximum weighing capacity, Max.*
Verification scale interval, e*
Minimum capacity, Min.*
Tare range (subtractive)
Application range according to CD*
Response time (average)
Allowable operating temperature range
External calibration weight value
(of at least accuracy class...)
Other permissible external calibration weights (of at least accuracy class)
Selectable weight units
Pan size
Dimensions (W x D x H)
Net weight, approx.
g s g g g g mm mm kg
LP1200S-0CE LP620S-0CE isoTEST in conjunction with BD BF k
0.001
K
0.001
LP220S-0CE
K
0.001
1,200
0.01
620
0.01
220
0.01
0.1
0.02
≤
100% of the max. weighing capacity
0.1 – 1,200 0.02 – 620
0.02
0.02 – 220
1.5
0 ... +40
°
C (273 ... 313 K, 32
°
F ... 104
°
F) with the isoCAL function
1
)
1,000 (E2) – –
LP620P-0CE
K
0.001/0.002/
0.005
120/240/620
0.01
0.02
0.02 – 620
–
–
Grams and kilograms
Ø 130
240x360x147
8.3
–
6.9
–
6.9
–
6.9
1
) = With the isoCAL function deactivated, the verified balance can only be used within the limited temperature range (can only be modified by the Sartorius Service Center):
For balances of accuracy class k: +15
°
C to +25
°
C (59
°
F to 77
°
F)
For balances of accuracy class K: +10
°
C to +30
°
C (50
°
F to 86
°
F)
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area
85
Model
Type
Accuracy class*
LP8200S-0CE
K
LP8200P-0CE isoTEST in conjunction with BD BF
K
LP6200S-0CE
K
LP4200S-0CE
K
Scale interval, d* g
Maximum weighing capacity, Max.* g
0.01
8,200
0.01/0.02/
0.05
0.01
2,000/4,000/ 6,200
8,200
0.01
4,200
Verification scale interval, e* g 0.1
Minimum capacity, Min.*
Tare range (subtractive) g 0.5
≤
100% of the max. weighing capacity
Application range according to CD* g 0.5 – 8,200 0.5 – 8,200 0.5 – 6,200 0.5 – 4,200
Response time (average) s 2 2 1.5
Allowable operating temperature range 0 ... +40
°
C (273 ... 313 K, 32
°
F ... 104
°
F) with the isoCAL function
1
)
Selectable weight units
Pan size
Dimensions (W x D x H)
Grams and kilograms mm 218 x 200 mm 240 x 360 x 86
Net weight, approx.
kg 6.5
LP2200S-0CE
K
0.01
2,200
0.5 – 2,200
Model
Type
Accuracy class*
Scale interval, d*
LP820-0CE LP2200P-0CE isoTEST in conjunction with BD BF
K g 0.01
K
Maximum weighing capacity, Max.* g 820
0.01/0.02/
0.05
400/800/
2,200
LP5200P-0CE
K
0.01/0.02/
0.05/0.1
LP12000S-0CE
K
0.1
1,200/2,400/ 12,000
3,800/5,200
Verification scale interval, e* g 0.1
0.1
Minimum capacity, Min.*
Tare range (subtractive) g 0.5
0.5
≤
100% of the max. weighing capacity
Application range according to CD* g 0.5 – 820 0.5 – 2,200
0.1
0.5
1
5
0.5 – 5,200 5 – 12,000
Response time (average) s 1.5
1.5
1.5
1
Allowable operating temperature range 0 ... +40
°
C (273 ... 313 K, 32
°
F ... 104
°
F) with the isoCAL function
1
)
Selectable weight units Grams and kilograms
Pan size
Dimensions (W x D x H)
Net weight, approx.
mm 218 x 200 mm 240 x 360 x 86 kg 6.5
LP6200-0CE
1
5
K
0.1
6,200
5 – 6,200
1
1
) = With the isoCAL function deactivated, the verified balance can only be used within the limited temperature range (can only be modified by the Sartorius Service Center):
For balances of accuracy class k: +15
°
C to +25
°
C (59
°
F to 77
°
F)
For balances of accuracy class K: +10
°
C to +30
°
C (50
°
F to 86
°
F)
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area
86
Overview
Model
Type
Accuracy class*
Scale interval, d*
Maximum weighing capacity, Max.*
Verification scale interval, e*
Minimum capacity, Min.*
Tare range (subtractive)
Application range according to CD*
Response time (average)
Allowable operating temperature range
Selectable weight units
Pan size
Dimensions (W x D x H)
Net weight, approx.
LP2200-0CE LP12000P-0CE isoTEST in conjunction with BD BF
K K
LP16000S-0CE
BB BD 523
K g g g
0.1
2,200
0.1
0.1/0.2/0.5
0.1
3,000/6,000/ 16,000
12,200
1 1
5 g s g 5 5
≤
100% of the max. weighing capacity
5 – 2,200 5 – 12,000
1 1
+10 ... +30
°
C (50
°
F ... 86
°
F)
Grams, kilograms mm 218 x 200 mm 240 x 360 x 86
5 – 16,000
1.5
Grams, kilograms
307 x 417
307 x 538 x 121 kg 6.5
6.5
13.8
LP34000P-0CE
BB BD 523
K
LP34-0CE
BB BD 523
K
0.1/0.2/0.5
1
8,000/16,000/ 34,000
34,000
1 1
5 50
5 – 34,000
1.5
13.8
5 – 34,000
1
Kilograms
13.8
Model
Type
Accuracy class *
Scale interval, d*
Maximum weighing capacity, Max.*
Verification scale interval, e*
Minimum capacity, Min.*
Tare range (subtractive)
Application range acc. to CD *
Response time (average)
Allowable operating temperature range
Selectable weight units
Pan size
Dimensions (W x D x H)
Net weight, approx.
Dust and water protection rating according to EN 60529
2
) g s
LP34001S-0CE LP16001S-0CE isoTEST in conjunction BF BF
K K g g g 0.1
kg 34
0.1
16
LP34001P-0CE
K
0.1/0.2/0.5
8/16/34
LP34000-0CE
K
1
34
1 1
5 5
≤
100% of the max. weighing capacity
5 – 34,000 5 – 16,000
1
5
1
50
5 – 34,000 50 – 34,000
1.5
1.5
1.5
1
0 ... +40
°
C (273 ... 313 K, 32
°
F ... 104
°
F) with the isoCAL function
1
)
Grams and kilograms
Grams and kilograms
Grams and kilograms
Kilograms and grams mm 300 x 400 mm 313 x 534 x 120 kg 16.0
IP44
1
) = With the isoCAL function deactivated, the verified balance can only be used within the limited temperature range (can only be modified by the Sartorius Service Center):
For balances of accuracy class k: +15
°
C to +25
°
C (59
°
F to 77
°
F)
For balances of accuracy class K: +10
°
C to +30
°
C (50
°
F to 86
°
F)
2
) = specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories”
* CD = Council Directive 90/384/EWG for non-automatic weighing instruments used within the European Economic Area
87
88
Accessories (Options)
Product
Data printer with date/time, statistical evaluation data, transaction counter functions and LCD can be used in legal metrology
Order No.
YDP03-0CE
External rechargeable battery pack has a battery-level indicator (LED); can be recharged using the AC adapter
(time it takes to charge the discharged battery pack: 15 hours); see “Specifications” for hours of operation can be used in legal metrology
YRB06Z
Remote display as a second indicator
(weight readout only)
> can be connected via the interface port
– LCD, reflective
– LCD for overhead projectors, transmissive can be used in legal metrology
3-segment checkweighing display conveniently shows whether a sample
(amount filled) is within the tolerance limits can be used in legal metrology
Draft shield chamber for the LP 1200S, LP 3200D, LP 620S,
LP 620P and LP 220S
Density determination kit for the LP 1200S, LP 3200D, LP 620S,
LP 620P and LP 220S
Calibration weights for all LP balances; extensive assortment, optionally available with officially recognized
DKD certificates
YRD02Z
YRD13Z
YRD11Z
YDS01LP
YDK01LP
Information on request
Overview
Product
SartoConnect data transfer software; with RS-232 standard cable; for direct input of weighing data into an application program (e.g., such as Excel)
Standard operating procedure to ensure correct operation and handling of the balance in quality assurance systems
AC adapter, model ING-2 with IP65 protection rating
(dust-tight and washdown-resistant) for 230 V for 120 V
Hanger for below-balance weighing for the LP34001S, LP16001S, LP34001P and LP34000
Universal remote control switch for remote control of one of the following functions
(configured in the balance Setup menu): p , t, q, v, c, w
Foot switch with T-connector
Hand switch with T-connector
T-connector
Cable for connecting the weighing cell to a separate display and control unit (length: 2.70 m) for balances with a weighing capacity
≤
12 kg for balances with a weighing capacity
≥
16 kg
Support arm (for raised display configuration) for balances with a weighing capacity
≤
12 kg for balances with a weighing capacity
≥
16 kg
Weighing bowls:
Made of chrome-nickel steel; without pouring spout; for all models with a weighing capacity > 400 g:
– 3,000 ml capacity
– 1,000 ml capacity
– 500 ml capacity
Carrying case for balances with a weighing capacity
≤
12 kg
Order No.
YSC01L
YSL01E
69 71899
69 71500
69EA0040
YFS01
YHS02
YTC01
YCC01-19M3
Information on request
YDH01LP
YDH02LP
641213
641211
641212
YDB01LP
89
Declarations of Conformity
The C Mark on Sartorius Equipment
In 1985, the Council of the European
Community approved a resolution concerning a new approach to the technical harmonization and standardization of national regulations. The organization for monitoring compliance with the directives and standards concerning the C marking is governed in the individual EU Member States through the implementation of the EC
Directives adopted by the respective national laws. As of December
1993, the scope of validity for all
EC Directives has been extended to the Member States of the
European Union and the Signatories of the Agreement on the European
Economic Area.
Sartorius complies with the EC
Directives and European Standards in order to supply its customers with weighing instruments that feature the latest advanced technology and provide many years of trouble-free service.
The C mark may be affixed only to weighing instruments and associated equipment that comply with the applicable Directive(s):
Council Directive 89/336/EEC
“Electromagnetic Compatibility (EMC)”
Acceptable European Standards:
Limitation of emissions:
EN 50081-1
Residential, commercial and light industry
EN 50081-2
Industrial environment
Defined immunity to interference:
EN 50082-1
Residential, commercial and light industry
EN 50082-2
Industrial environment
Important Note:
The operator shall be responsible for any modifications to Sartorius equipment and for any connections of cables or equipment not supplied by Sartorius and must check and, if necessary, correct these modifications and connections. On request, Sartorius will provide information on the minimum operating specifications (in accordance with the
Standards listed above for defined immunity to interference).
Council Directive 73/23/EEC
“Electrical Equipment Designed for Use within Certain Voltage Limits”
Applicable European Standards:
EN 60950
Safety of information technology equipment including electrical business equipment
EN 61010
Safety requirements for electrical equipment for measurement, control and laboratory use
Part 1: General requirements
If you use electrical equipment in installations and under ambient conditions requiring higher safety standards, you must comply with the provisions as specified in the applicable regulations for installation in your country.
Weighing Instruments for Use in Legal Metrology:
Directive 90/384/EEC “Nonautomatic Weighing Instruments”
This Directive regulates the determination of mass in legal metrology.
For the respective Declaration of
Conformity for weighing instruments that have been verified by Sartorius for use as legal measuring instruments and that have an EC Type-
Approval Certificate, see the page after next.
This Directive also regulates the performance of the EC verification by the manufacturer, provided that an EC Type-Approval Certificate has been issued and the manufacturer has been accredited by an officer or a Notified Body registered at the Commission of the European
Community for performing such verification.
The legal basis allowing Sartorius to perform EC verification is constituted by the EC Council Directive No.
90/384/ EEC on non-automatic weighing instruments that has been in effect since January 1, 1993, in the Internal Market as well as by the
Certificate of Accreditation of the
Sartorius AG Quality Management
System issued by the Metrology
Department of the Regional
Administration Office of Lower
Saxony, Germany (“Niedersächsisches Landesverwaltungsamt
-Eichwesen”) on February 15, 1993.
For information on the C mark on Sartorius equipment and legal regulations currently applicable in your country, and to obtain the names of the persons to contact, please ask your local Sartorius office, dealer or service center.
90
“New Installation” Service
Initial verification is covered in our
“New Installation” service package.
In addition to initial verification, this package provides you with a series of important services which will guarantee you optimal results in working with your weighing instrument:
– Installation
– Startup
– Inspection
– Training
– Initial verification
If you would like Sartorius to perform initial verification of your weighing instrument, contact an authorized service representative.
“EC Verification” –
A Service offered by Sartorius
Our service technicians are authorized to perform the verification* of your weighing instruments that are acceptable for legal metrological verification and can inspect and verify the metrological specifications at the place of installation within the
Member States of the European
Union and the Signatories of the
Agreement on the European
Economic Area.
Subsequent Verifications within the
European Countries
The validity of the verification will become void in accordance with the national regulations of the country in which the weighing instrument is used. For information on verification and legal regulations currently applicable in your country, and to obtain the names of the persons to contact, please contact your local Sartorius office, dealer or service center.
Overview
* in accordance with the accreditation certificate issued to Sartorius AG
91
92
Overview
93
94
Overview
95
96
Overview
Index
Accessories
Acoustic signal
Additional functions
Animal weighing
Antitheft locking device
Application programs
Auto print
Auto print +/–
Automatic calibration and adjustment
(isoCAL)
Automatic shutoff
Balance-specific information
Basic weighing function
Below-balance weighing
Blocking the keys
Cabling diagram
Calibration/adjustment
Care and maintenance
Cleaning
Configuring the balance
Connecting the balance to AC power
Contents
Control lines (checkweighing)
Counting
Data input format
Data output – activating
Data output – automatic
Data output – by print command
Data output format
Data output functions
Description of the keys
Declarations of Conformity
Display – output to
Display backlighting
Display unit – separate
Displays of “longer” values
Dust cover – replacing
Electromagnetic compatibility
Equipment supplied
Error codes
External calibration/adjustment
General password
General views of the balances
Getting started
Handshake
Hardware handshake
Input user data
Installation instructions
Instructions for recycling
Interface description
Interface port input/output
Internal calibration/adjustment
ISO/GLP-compliant printout/record isoCAL
Page
88
70
70
42
9
31
60
60
80
90
56
70
7
57
75
65
67
67
67
63
56
69
24
75
75
10
8
2
47
33
29
70
10
21
21
70
11
5
76
62
71
28
61
29
90
5
73
27
Appendix
77
5
66
66
Language setting
Leveling the balance
Line for metrological data
Line format
Measured value line
Menu items
Menu structure (diagram)
MP8 interface
Net-total formulation
Operating design
Operating the balance
Options
Over/under checkweighing
Password
Pin assignment chart
Power-on mode
Practical use
Print animal weights
Print application parameters
Print manually/automatically
Printing a data record
Printing an ISO/GLP-compliant record
Protecting menu parameters
Radio frequency interference
Reference sample updating
Recalculation
Repairs
Reproducibility determination (reproTEST)
Reset function
Safety inspection
Sample ID
Service
Setting parameters (menu)
Setting up the balance
Setup parameters (overview)
Software handshake
Specifications
Storage and shipping conditions
Remote display
Synchronization
Text line
Type-approval certificate
Toggle between weight units
Undoing all parameter changes –
Reset function
Universal switch for remote control
Unpacking the balance
User data – input
Warmup time
Warnings and safety precautions
Warranty
Weighing in percent
60
59
59
58
70
68
70
2
72
70
8
34
51
70
30
71
6
16
66
82
5
88
66
75
59
75
14
40
3
21
88
46
56
16
81
72
Page
10
9
56
59
57
93
31
8
2
5
36
71
70
5
11
97
Appendix
Entering the General Password
Enter/Change Password
Select the Setup menu: M
> The soft keys info, menu and input are displayed in the text line
Select the user data input function: Press the input soft key (v)
> The password prompt is displayed (ENTER PASSW.)
Enter the General Password
(see below)
Press the ENTER PASS.
soft key (v)
> The last 8 digits of a workstation
ID (balance ID no.), if available, are displayed in the measured value line
Select password setting:
Press y
> If a password exists, it is now displayed in the measured value line
New password: Enter the letters/numbers for the new password (8 characters max.)
To delete the password, enter a decimal point using the
.
key and confirm
Confirm input: Press X
Exit the Setup menu:
Press M
> Restart your application
General Password: 40414243
Sartorius AG b 37070 Goettingen, Germany p Weender Landstrasse 94–108, 37075 Goettingen, Germany t (+49/551) 308-0, f (+49/551) 308-3289
Internet: http://www.sartorius.com
Copyright by Sartorius AG, Goettingen, Germany.
All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG.
The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius AG reserves the right to make changes to the technology, features, specifications, and design of the equipment without notice.
Status: October 2001, Sartorius AG, Goettingen, Germany
Printed in Germany on paper that has been bleached without any use of chlorine · W1A000 · KT
Publication No.: WLP6002-e01107