ERGO-FIT Cardio Line 400 Owner's Manual


Add to my manuals
116 Pages

advertisement

ERGO-FIT Cardio Line 400 Owner's Manual | Manualzz

O w n e r ‘ s m a n u a l

CAR-

CARDIO LINE 400

Please read this manual carefully before use and keep it in a safe place for future reference

Further information on ERGO-FIT products available from:

ERGO-FIT GmbH & Co. KG

Blocksbergstraße 165

66955 Pirmasens

Phone: +49 (6331) 2461-0

Fax.: +49 (6331) 2461-55

Email: [email protected]

http: www.ergo-fit.de

Development and production of all devices of the MED series according to the

European Medical Device Directive 93/42/EWG. They thus show the CE marking and the number of the notified body.

© 2011 by ERGO-FIT GmbH & Co. KG. All rights reserved.

This owner's manual has been created with great care. Please inform us of any detail that does not correspond to your training tool to allow for the quickest possible remedy of any possible discrepancy.

This document is protected by copyright. All rights originating from this, particularly the right of copying, duplication, distribution as well as translation and reprint, even copies in extract, are reserved. No part of this manual may be reproduced, manifold or distributed for any kind of purposes or in any kind of form, by any means, (print, photocopy, microfilm, or any other information-saving mechanisms) without prior written permission given by the company

ERGOFIT GmbH & Co. KG.

Trade mark:

ERGO-FIT and the ERGO-FIT Logo are registered trademarks of ERGO-FIT GmbH & Co. KG. All other trademarks mentioned or shown in the following text are trademarks of the respective owners and are recognised as protected.

Technical and optical modifications as well as misprints reserved.

Version: CARDIO LINE 400/400 MED-20110809-en

Printed in Germany

Dear customer, we are glad that you have decided to purchase an ERGO-FIT training tool. You are now the owner of a sophisticated and exclusive training system that combines highest technical standards with practice oriented ease of use.

You will find important information on the operation and use of your training machine in this owner’s manual. We recommend that you read this owner’s manual carefully before training in order to become familiar with your training device quickly and to understand its correct and safe use.

Should you have any questions that are not answered in this manual, please contact us. The ERGO-FIT team is always there for you!

CARDIO LINE 400/400 MED

Table of Contents

7

8

9

10

A

1

2

3

4

5

6

General information ................................................................................... 1

Safety information ..................................................................................... 5

Short manual ............................................................................................. 11

Destination of the product ......................................................................... 13

Transport and set up ................................................................................. 17

Start up ...................................................................................................... 25

Operation .................................................................................................. 31

Training ..................................................................................................... 45

Maintenance .............................................................................................. 49

Troubleshooting ........................................................................................ 55

Appendix ................................................................................................... 61

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

1.1

1.2

1.3

1.4

Chapter 1: General information

ERGO-FIT Cardio equipment at a glance ........................................... 2

General informaion on this manual ...................................................... 3

Scope of delivery .................................................................................. 3

Service ................................................................................................ 4

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1

General information

1 General information

1.1 Cardio equipment at a glance

ERGO-FIT's CARDIO LINE 400 offers you a cardio machine range dedicated to train the cardiovascular system. It allows for training of smaller muscle groups. ERGO-FIT cardio machines offer you best training possibilities, regardless of your age, gender or state of fitness.

Among others, the optimal load regulation and the precise training control are highlights of the whole product line. In addition, the quiet operation, the ease of use and the customisation demonstrate ERGO-FIT's focus: A high technical standard, optimal training possibilities and precise training control, combined with customerfriendly ease of use.

However, technology alone is not all that is required to develop outstanding training machines. They also need to meet biomechanical and sports medical requirements.

Priority is given to the human being. Thus, a sophisticated training and testing system can only be developed by combining technical electronic expertise with the latest advances in sports medicine and coaching science. ERGO-FIT clearly met this target.

Our CARDIO LINE is divided in:

CARDIO LINE 400: training machines especially designed for use in a gym

CARDIO LINE 400 MED: training machines especially designed for medical purposes

The lifetime of the equipment is 6 years.

Advantages and benefits

Regular training on these machines reduces the risk of cardio-vascular diseases and increases physical capacities in an optimal way - even at an advanced age. Your workout machine represents an indispensable tool in injury prevention and rehabilitation. You will feel fit, more agile, more attractive, and more balanced.

2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

1.2 General information on this manual

This manual provides you with helpful information, regardless of if you are already familiar or have no experience with ERGO-FIT training machines.

It is structured in such a way that you can find the desired information in the table of contents easily and thematically. In addition, a short manual has been produced for those users who are already familiar with ERGO-FIT training machines. However, if you belong to this user group and wish to read the short manual only, you should review the safety information first.

The manual will give you many hints and tips, which will familiarise you with your cardio machine's features and allow you to become an experienced user very quickly.

You should always keep this manual easily accessible. This saves you from having to make unnecessary and time-consuming queries and enables you to fix any possible error rapidly.

1.3 Scope of delivery

Please check if the delivery is complete and inform our sales department immediately of any missing parts (phone: +49 (6331) 2461-0).

Please ensure that the following parts are included in your delivery:

1. The correct type of training machine (model, series)

2. One mains connection cable for each unit (CYCLE 407/457 MED; CYCLE

400/450's mains cable is firmly connected with the machine)

3. The manual in paper copy and the digitalised version on a multimedia CD-

ROM

Please note that accessories (e.g. POLAR-sender) are not included in the delivery and need to be ordered separately.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

3

General information

1.4 Service

Our service comprises problem analysis, technical support, and information services.

In case of technical questions and service orders, please call us at:

Head office:

Service and spare parts:

Phone: +49 (6331) 2461-0

Fax: +49 (6331) 2461-55

Phone: +49 (6331) 2461-22

Phone: +49 (6331) 2461-23

Phone: +49 (6331) 2461-27

Phone: +49 (6331) 2461-29

4

Technical and optical modifications as well as misprints reserved - © 2009 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

2.1

2.2

2.3

2.4

2.5

2.6

Chapter 2: Safety information

What you need to know when using your training machine ................. 6

Operational safety - What needs to be done ........................................ 8

Caution: Risk of electric shock ............................................................. 8

Choosing the right place of installation ................................................. 9

What needs to be considered in case of repair .................................... 9

What should be avoided ....................................................................... 9

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5

Safety information

2 Safety information

Before starting to use your new ERGO-FIT equipment, please read the following chapter carefully and respect all safety regulations.

Please keep this manual in a safe place in order to be able to provide it to future owners if you sell your training machine.

For evidence of ownership, please fill in the following data:

Model/product line:

Serial number:

Date of purchase:

__________________________________

__________________________________

__________________________________

You will also need these data in case of guarantee claims.

The following symbols designate important information:

Caution!

It is absolutely necessary to observe this warning in order to avert any danger to your life or health.

Danger!

It is absolutely necessary to observe this warning in order to avoid any material damage.

Danger!

Danger: It is imperative to switch off and disconnect the training machine.

Hint!

Important information and hints are displayed here to improve operations.

6

2.1 What you need to know when using your training machine

⊗ Please read the owner's manual carefully before starting to use your new ERGO-

FIT equipment.

⊗ Familiarise yourself with the machine before you start it (only CARDIO LINE 400

MED, MPBetriebV).

⊗ Only use the machine after a proper function test (get more information on this in chapter 7.3). For your own safety and before every use, please check the

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED machine for damage (loose screws, worn parts, cords etc). If the machine is damaged, do not use it until it is repaired.

⊗ Before using the machine please check that the saddle is firmly seated by fixing the horizontal and vertical saddle adjustment (see chapter 6.3). Otherwise there is a risk of injury

⊗ In order to reduce any risk during training, please put on sportswear and appropriate footwear.

⊗ In case of nausea, dizziness, pain in the chest, limbs or joints, stop training immediately and see a doctor.

⊗ If you have a cardiac pace maker or a health condition, please see your doctor before using the training machine. If this is the case, you should discuss the training programme with him. In addition, you should only train in MANUAL mode.

⊗ Your training machine is not a toy! Never leave children unattended with the machine. Children cannot always assess possible danger. Parents or other supervisors should always be aware of their responsibility because children dispose of a natural play instinct and eagerness to experiment that may produce situations and behaviour the training machine is not designed for.

⊗ The machine should only be used after consulting a doctor and / or a supervisor.

The machine must not be used without the presence of a supervisor (only

CARDIO LINE 400 MED).

⊗ Make sure that persons who stand close to the machine are not hit by moving parts.

⊗ Danger! The user's maximum body weight is 180 kg. Make sure not to overload the machine.

⊗ Do not place any beverages or food on your training machine.

⊗ Do not stand on the casing of the machine and do not lean on the control panel.

⊗ Once per week, inspect the seat, the handlebar, the pedals, and the casing for damage. If the machine is damaged, get it repaired immediately.

⊗ Tighten the pedals and the pedal arms after 3 to 5 operating hours. Otherwise, there is a risk of damage, which can be a risk of injury to the user (see chapter

9.1).

⊗ Make sure to tighten the seat clamp after adjusting the seat. Otherwise, there is a risk of injury.

⊗ Warm up thoroughly before training. Start the training slowly and gradually increase the intensity until you reach the desired degree of difficulty within your range of control.

⊗ Do not lean over the handlebar and do not shift your body weight over the side

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

7

Safety information of the machine. There is a danger of falling over.

⊗ Do not take your feet off the pedals during training.

⊗ Do not jump from the training machine during your workout.

⊗ Please review the additional safety and operational instructions in this manual.

All safety instructions in this manual are based on many years of experience and selfconception

2.2 Operational safety - what needs to be done

⊗ After delivery, make sure that the machine has not been damaged during transport. In case of doubt, contact our customer service and do not start the machine.

⊗ Slots and openings on the machine serve as ventilation. Do not cover these openings, because this can cause the elements to overheat.

⊗ Check if the power connection cords are damaged before starting the machine.

⊗ Switch off the machine after training and disconnect it from the grid.

⊗ Set up the training machine so that there is sufficient open space around it (at least 1 m) to avoid any danger for the user or others. The user will not collide and in addition, this prevents interferences in heart rate measuring.

⊗ The provisions EN 60601-1-1:2001 (medical electrical equipment) are valid for machines of the series CARDIO LINE 400 MED.

2.3 Caution: Risk of electric shock

⊗ Only connect your machine to an appropriate mains supply (see Chapter

5.4.1).

⊗ Make sure to have the right climate conditions (see Chapter 5.3).

⊗ Do not use damaged power cords.

⊗ To unplug the cord, grasp the plug and not the cord. Do not unplug by pulling on the cord.

⊗ The machine must not be opened before it is switched off and unplugged.

⊗ If liquid gets inside the machine, disconnect the plug immediately and call the customer service.

⊗ Do not put any objects through the ventilation slots into the machine. This may cause a short circuit.

8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

2.4 Choosing the right place of installation

⊗ The machine can be set up on any level and stable floor. Make sure that it stands firmly on the floor.

⊗ Never put wood, cardboard or similar materials underneath the machine to compensate for surface unevenness. This increases the risk of an accident.

2.5 What needs to be considered in case of repair

⊗ Electric parts may only be replaced by original equipment.

⊗ Repairs must be carried out by a qualified technician only. If you do not have the necessary qualifications, contact the ERGO-FIT Service Centre.

2.6 What should be avoided

⊗ Only use the machine for the purposes it is designed for. If you use the ERGOFIT machine improperly, you will be charged for all damages resulting from this. Any guarantee claim will be excluded!

⊗ Never use the machine in any way other than for those purposes described in this manual. Improper use can damage the machine and be hazardous to your health.

⊗ You should never exercise on a damaged machine.

⊗ The machine must not be used without electric current.

⊗ When exercising, you should never try to exceed your current performance level. This may seriously damage your health.

⊗ You should never prop up your body on the machine and make improper movements. This increases the risk of falling.

A summary of the most important safety precautions can be found in the appendix of this manual. You should remove this summary and display it near the machine where it is clearly visible. All users of the machine must familiarise themselves with the dangers and safety regulations. The manufacturer will not be liable for personal injury or material damage.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

9

Safety information

1 0

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Chapter 3: Short manual

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

11

Short manual

3 Short manual

After delivery of your cardio machine, please check first if the serial number (see type label) is identical with the one indicated on the delivery note and if all components listed in chapter 1.3 ("Scope of delivery") are included in the delivery (not available on

CARDIO LINE 400 MED).

After the machine has been connected to the mains and switched on, the software version is shown on the display. Thereafter the main menu will appear.

The control panel contains the PLUS, MINUS, START and STOP buttons in this order.

The big monochrome display is illuminated and shows the elapsed training time

(min:s), your present heart rate, the performance in Watts, the speed range (1/min), the speed (km/h, not available on CYCLE 400/407 MED), the distance (m or km) as well as calories burnt.

When you switch on your cardio machine, the main menu will always appear first on the display. The choice "MANUAL" is selected automatically. Press the START-button to confirm this. You are now in the manual mode.

In this mode, you can work out for any amount of time and choose any strain level that pleases you by pressing the PLUS and MINUS buttons. The workout parameters will be displayed for the full training session.

If you would like to end the training session, please press the STOP-button. The training parameters remain on the display. Press the STOP-button again to return to the main menu. The display will also return to the main menu automatically after 2 minutes if no action is carried out.

Danger! For workouts in other modes, please read the detailed instructions.

1 2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Chapter 4: Destination of the product

4.1

4.1.1

4.1.2

4.2

4.2.1

4.2.2

4.2.3

4.2.4

Product lines ......................................................................................... 14

CARDIO LINE 400 ............................................................................... 14

CARDIO LINE 400 MED ...................................................................... 14

Models .................................................................................................. 14

CYCLE 400 .......................................................................................... 14

CYCLE 407 MED.................................................................................. 15

CYCLE 450 .......................................................................................... 15

CYCLE 457 MED ................................................................................. 15

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1 3

Destination of the product

4 Destination of the product

4.1 Product lines

In order to meet our customers' desires and to implement technical innovations,

ERGOFIT developed the following product lines:

4.1.1 CARDIO LINE 400

The machines of this product line are stationary exercise machines primarily dedicated to train the cardiovascular system. They are especially designed for home use (EN 957

HA). They allow for management, control, and documentation of the workout.

4.1.2 CARDIO LINE 400 MED

The machines of this product line are stationary exercise machines that are designed for precise performance diagnostics in addition to training the cardiovascular system, diagnostics and therapy of cardio-vascular diseases and performance diagnostics in competitive sports. These product lines fulfil medical requirements and therefore need to allow for precise measuring technology besides pure fitness training. Explicit assessment of the results provides optimisation of health training and allow for its continuous documentation. To ensure measuring accuracy, metrological controls need to be carried out on these machines at regular intervals (every second year, in compliance with

MPBetreibV).

4.2 Models

All models of the CARDIO LINE 400/400 MED are exercise bike ergometers designed for cardiovascular training. A cyclic kicking movement produces the strain. The training strain is controllable to provide optimal cardiovascular system load. The CYCLE is equipped with various features in order to attain an optimised adaptation to the customer's individual needs.

4.2.1 CYCLE 400

The CYCLE 400 is equipped with a manually controllable programme, a heart rate controlled cardio exercise programme, and 5 predefined profiles as well as the option to exercise in gearshift mode. An external connection (RS 232) is standard equipment on the CYCLE 400.

1 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

4.2.2 CYCLE 407 MED

The CYCLE 407 MED is equipped with a manually controllable programme, a heart rate controlled cardio exercise programme, and 2 predefined WHO-profiles as well as the option to do countdown training in manual or heart rate controlled mode.

4.2.3 CYCLE 450

The CYCLE 450 is equipped with a manually controllable programme, a heart rate controlled cardio exercise programme, several predefined profiles as well as individual user profiles, 3 test programmes and the option to exercise in gearshift mode. An external connection (RS 232) is standard equipment on the CYCLE 450

4.2.4 CYCLE 457 MED

The CYCLE 457 MED is equipped with a manually controllable programme, a heart rate controlled cardio exercise programme, several predefined profiles as well as individual user profiles, 3 test programmes, 2 WHO-profiles that can be modified, the option to do countdown training in manual or heart rate controlled mode. Furthermore, the machine can be controlled by ECG equipment. An external connection (RS 232) is standard equipment on the CYCLE 457 MED.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1 5

Destination of the product

1 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

Chapter 5: Transport and set up

5.1

5.2

5.3

5.4

5.4.1

5.4.2

5.5

5.5.1

5.6

5.6.1

5.6.2

5.6.3

5.6.4

Transport .............................................................................................. 18

Set up and the right place for set up .................................................... 18

Ambient temperature ............................................................................ 20

Power connection ................................................................................. 20

Power supply ........................................................................................ 21

Cabling ................................................................................................. 21

Potential compensation ........................................................................ 21

Position of the potential compensation clamp ...................................... 21

Components ......................................................................................... 22

CYCLE 400 .......................................................................................... 22

CYCLE 450 .......................................................................................... 22

CYCLE 407 MED ................................................................................. 23

CYCLE 457 MED ................................................................................. 23

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

1 7

Transport and set up

5 Transport and set up

5.1 Transport

In order to avoid damage, ERGO-FIT machines are transported by ERGO-FIT GmbH

& Co. KG directly or by an authorised freight forwarding company. After delivery, packaging will be collected and disposed of professionally. If ERGO-FIT machines are delivered by a freight forwarder, the customer must dispose of the packaging himself or can send it back to ERGO-FIT GmbH & Co. KG. (The customer is responsible for the transportation costs).

The machines do not have any shipping locks!

To place the CYCLE at the desired location, please consider the following aspects:

1. Position yourself so that you can see the display.

2. Grasp the seat with both hands and slightly lift the machine. You can now roll it easily to the desired position.

5.2 Set up and the right place for set up

Place the mains cable between the casing and the foot frame to protect it against damage during transport.

Move the machine to its future location and mount the foot caps.

1 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Danger! Mount the foot caps first before laying the cable underneath the foot frame:

Lay the cable underneath the foot frame to the rear end of the machine. Make sure

not to pinch the cable between the casing and the foot frame or between the casing and the floor!

CORRECT WRONG

5.2 Set up and the right place for set-up

⊗ Make sure that the surface underneath the machine is level and horizontal.

⊗ The adjustable feet underneath the machine are designed to compensate for minor unevenness. Adjust these feet to attain a stable position on the floor and prevent tilting of the machine.

⊗ Please make sure the distance between each machine is at least one metre, or you might experience errors of the POLAR heart rate receiver.

⊗ High-frequency and magnetic disturbance signals (e.g. radios, TV sets, mobile phones etc.) in close proximity to your cardio machine area may also interfere with the pulse transmission.

⊗ In rare cases, strong electromagnetic fields in some locations may interfere with the transmission of the heart rate (e.g. high voltage circuit or tramway contact wires).

⊗ In case of interference or if you suspect there is interference with the heart rate transmission, do not, under any circumstances, use the machine for a cardio controlled workout (in CARDIO mode).

Please note that the adjustable feet of the CYCLE are equipped with plastic protection caps. In exceptional cases (e.g. when using strong cleaning agents), this may leave marks or stains on the floor.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1 9

5.3 Ambient temperature

⊗ Your ERGO-FIT exercise machine may be used at an ambient temperature of

+10°C to +40°C, a relative humidity of 30 to 70% (non-condensing) and an atmospheric pressure of 700 hPa to 1060 hPa without causing any problems.

⊗ When switched off, the ERGO-FIT machine may be stored at a temperature between -30°C to +50°C.

5.4 Power connection

1. Carry out a visual inspection of the mains connection cable and the connectors

(power input module on CARDIO LINE 400 MED only) before starting the machine. Damaged power cords and connectors need to be replaced immediately.

2. Damaged power cords and connectors need to be replaced immediately. Take the mains connection cable and plug it into the appropriate power input module

(see figure, CARDIO LINE 400 MED only). Plug the other end of the cable into the mains socket.

3. Switch on your exercise machine by pushing the button located on the power input module (CARDIO LINE 400 MED, I = switch on, O= switch off) or by pushing the foot actuated switch on the power cord (CARDIO LINE 400).

Power input module with mains plug

2 0

4. After your exercise machine has been connected to mains and switched on, it automatically carries out an operating check. During this operating check, you will be able to read the software version of the unit on the display. Thereafter the main menu will appear.

5. Stand on the side of the control panel (view onto the display) and check if the

display works. If this is not the case, make sure you followed the steps above correctly. In addition, verify if there is electricity in the mains socket.

5.4.1 Power supply

Use your exercise machine only on earthed (grounded) mains sockets with 230 V~/50-

60 Hz (see chapter A). If you have any doubts about the power supply at the setup location, ask your electricity supplier. Only use customary 10 ampere automatic circuit breakers (type B tripping characteristic). In the rare event that these automatic circuit breakers should switch off when you switch on your machine, the circuit needs to be fused with 10 A lead fuses or with a different type of tripping characteristic. In case of doubt, ask your electrician.

Before connecting your ERGO-FIT exercise machine to your power supply system, compare the data on the type label (next to power input module) on the allowable voltage and grid frequency with your local data.

Always connect your machine directly to the mains socket. Never use extender cables or multi-outlet power strips.

5.4.2 Cabling

If you have connected more than one ERGO-FIT machine to one main switch, please switch each machine on and off individually.

⊗ Install the cable in such a way that no one can step on it or trip over it.

⊗ Do not place any objects on the cable as it might become damaged.

5.5 Potential compensation

To avoid disturbances, a potential compensation cable may be attached to the machine. This potential equalisation is not standard equipment. If required, ask the service to install this feature. In this case, communicate with the service department or the field staff. As some machines are already prepared for the potential compensation feature, on-site installation is possible.

5.5.1 Position of the potential compensation clamp

The potential equalisation clamp is located underneath the front wheel cover.

Never connect the potential compensation line to the water or gas supply line or any other kind of pipe. Always use the appropriate potential compensation connection only.

2 1

5.6 Components

5.6.1 CYCLE 400

8

9

7

10

1

6

5

3

2

4

1 Seat

2 horizontal seat adjust

ment

3 Seat height adjustment

4 Mains cable outlet

5 Pedals

6 Adjustable feet

7 Control panel

8 Display

9 Handlebar (equiped with electrodes for hand heart rate)

10 Book support

5.6.2

7

CYCLE 407 MED

8

9

10

1

5

3

2

4

1 Seat

2 horizontal seat adjust

ment

3 Seat height adjustment

4 Power input modul

5 Pedals

6 Adjustable feet

7 Control panel

8 Display

9 Handlebar

10 Book support

2 2

6

5.6.3 CYCLE 450

8

9

7

10

1

5

3

6

2

4

1 Seat

2 horizontal seat adjust

ment

3 Seat height adjustment

4 Mains cable outlet

5 Pedals

6 Adjustable feet

7 Control panel

8 Display

9 Handlebar (equiped with electrodes for hand heart rate)

10 Book support

5.6.4

7

CYCLE 457 MED

8

9

10

1

5

3

6

2

4

1 Seat

2 horizontal seat adjust

ment

3 Seat height adjustment

4 Power input modul

5 Pedals

6 Adjustable feet

7 Control panel

8 Display

9 Handlebar

10 Book support

2 3

2 4

CARDIO LINE 400/400 MED

Chapter 6: Start-up

6.1

6.2

6.3

6.4

6.4.1

6.4.2

Switch on .............................................................................................. 26

Switch off .............................................................................................. 26

The right training technique .................................................................. 26

The control panel .................................................................................. 27

The buttons .......................................................................................... 29

The display ........................................................................................... 30

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

2 5

Start-up

6 Start-up

6.1 Switch on

⊗ Before switching on your exercise machine (CARDIO LINE 400 MED), make sure the mains plug is connected to the mains socket.

If you have connected more than one machine to one main switch, please switch each machine on and off individually, or technical disturbances may occur if all machines are switched on at the same time.

⊗ Now switch on your machine by pushing the switch located on the power input module. Push the switch into the I-position. The machine is turned off when the switch is in the 0-position (only on CARDIO LINE 400 MED). The CARDIO LINE

400 is switched on by pushing the foot actuated switch.

⊗ The illumination of the display shows you immediately if the machine is switched on.

6.2 Switch off

⊗ Switch off your machine by pushing the switch located on the power input module.

To do so, push the switch into the 0-position (CARDIO LINE 400 MED) or by pushing the foot actuated switch (CARDIO LINE 400).

Make sure that the switch on/switch off intervals are longer than 30 s. Otherwise disturbances might occur.

6.3 The right training technique

The optimal load intensity of the user is controlled by the heart rate controlled training according to the ERGO-FIT system. However, you should also take into account biomechanical considerations when training. This chapter provides relevant considerations for the use of exercise bike ergometers.

When exercising with your training machine, make sure not to step on its casing.

Stand on the designated surfaces only.

1. Gauge the proper seat height first. The height of the seat is essential for your sitting comfort and especially for the radial runout. To gauge the right seat height, sit on the seat and put one heel on a pedal. In the lowest pedal position, you should be able to extend your leg completely

2.a) CYCLE 400/407: To adjust the seat, stand next to the exercise machine and turn the knob on the seat tube anticlockwise. Now pull the knob out of the punched matrix located on the seat tube until you can move the seat tube. To

2 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED lock the seat at the desired height, let the knob snap into the seat tube again.

Lock the seat tube by turning the knob clockwisev.

b) CYCLE 450/457: To adjust the seat height, stand next to the exercise machine and pull the adjustment lever located on the seat tube upwards. The integrated pneumatic spring will automatically lift the seat. To lower the seat, loosen the adjustment lever and push the seat downwards into the desired position with the help of your body weight. To lock the seat at the desired height, bring the adjustment lever back to its initial position.

3. To adjust the seat horizontally, loosen the clamp lever underneath the seat by turning it anticlockwise. Now the seat is continuously adjustable. Lock the seat by turning the clamp lever clockwise.

4. During CARDIO training, make sure the number of rotations is higher than 50 rpm

(performance > 25 W); otherwise, the strain of the machine becomes too great.

Please look at the speed range (the up arrow on CYCLE 450/457 or the right arrow on CYCLE 400/407 means you should pedal faster whereas the down arrow on CYCLE 450/457 or the left arrow on CYLCE 400/407 asks you to pedal more slowly). The higher the number of rotations, the lower the strain on the joints.

5. Maintain the training position described here for the whole workout.

6.4 The control panel

One of ERGO-FIT's main goals is to produce particularly user-friendly exercise machines. For this reason, all the CARDIO LINE 400/400 MED machines are equipped with a user guidance system that is simple and easy to understand. In addition, all control panels in a series are designed homogenously to provide you with easy use and comfort. Consequently, once you know how to operate one machine, you will know how to operate all the others of the same series.

There are two different types of displays: A small monochrome display (CYCLE

400/407 MED) and a large monochrome display (CYCLE 450/457 MED).

The control panel consists of a display and buttons. Before you take a closer look at the control panel, you should note the following aspects:

1. Do not prop up your body on the control panel or the display to avoid damage.

2. Do not exert pressure on the display.

3. Only press the buttons lightly. The button press is confirmed by a beep.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

2 7

Cockpit CYCLE 400/407 MED (kleine monochrome Anzeige):

Inbetriebnahme

2 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Cockpit CYCLE 450-457 MED (große monochrome Anzeige):

6.4.1 The buttons

The following buttons are installed on the control panel:

⊗ PLUS-button

⊗ MINUS-button

⊗ START-button

⊗ STOP-button

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

2 9

Inbetriebnahme

The button functions are described below:

Button Function

Increase the strain or change parameters.

Decrease the strain or change parameters.

Change the training mode or confirm set or changed parameters.

Stop functions or the machine.

6.4.2 The display

The machines of the CARDIO LINE 400/400 MED are equipped with a large (CYCLE

450/457 MED) or a small monochrome display (CYLCE 400/407 MED).

In the section below you will find - depending on the model - information on the displays, measuring units and their meanings.

Training parameters

Model Display Meaning

DIST.

Distance covered

1/MIN Number of revolutions per minute

KM/H* Speed

CYCLE

WATT Power produced currently

KCAL Calories burnt (average)

TIME Training time

PULSE Current heart rate with POLAR-sender or hand heart rate(only CYCLE 400/450)

* Parameter “KM/H”only CYCLE 450/457 MED

Unit km

1/min km/h

Watt kcal

00:00 (min:s)

1/min

3 0

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Chapter 7: Operation

7.1.6

7.1.7

7.1.8

7.2

7.3

7.4.

7.1

7.1.1

7.1.2

7.1.3

7.1.4

7.1.5

7.4.1

7.4.2

7.4.3

7.4.4

Modes of operation ............................................................................... 32

MANUAL ............................................................................................... 32

PROFILES ............................................................................................ 33

CARDIO ............................................................................................... 36

GEARSHIFT ......................................................................................... 38

WHO-PROFILES .................................................................................. 38

COUNTDOWN ..................................................................................... 39

TEST .................................................................................................... 40

ECG CHOICE ................................................................................... 41

Behavior after operation ....................................................................... 42

Function control .................................................................................... 42

Heart rate messurement ....................................................................... 43

POLAR band and POLAR sender ........................................................ 43

Operating range of the POLAR sender ................................................ 43

Battery of the POLAR sender ............................................................... 44

Potential interferences .......................................................................... 44

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

3 1

Operation

7 Operation

7.1 Modes of operation

When you switch on your training machine, the main menu and the programme selection will always appear first on the display. The choice "MANUAL" is selected automatically. Push the PLUS and MINUS buttons to change the training mode.

Confirm this with START. The following modes can be chosen (depending on the model):

MANUAL

PROFILES

CARDIO

GEARSHIFT

WHO-PROFILES

COUNTDOWN

TEST

ECG-CHOICE

CYCLE 400 CYCLE 407 MED CYCLE 450 CYCLE 457 MED x x x x x x x x x x x x x x x x x x x x

To return from the submenu to the main menu, just press the STOP button once or repeatedly.

Danger! If you have a cardiac pace maker, you should exercise in MANUAL mode only!

7.1.1 MANUAL

In this mode, you can work out for any amount of time and choose any strain level that pleases you.

1. Push the PLUS/MINUS buttons until the programme MANUAL is selected.

Confirm your choice with the START button.

2. You are now entering training mode. You may change the strain intensity in steps of 5 Watts by pushing the PLUS and MINUS buttons. If you want to change the strain intensity substantially, press and hold the PLUS or MINUS button.

3. Stop the training by pressing the STOP button. The training parameters remain on the display. By pushing the STOP-button one more time, you will return to the main menu. The display will also return to the main menu automatically after 2 minutes if no action is carried out.

3 2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

In the MANUAL mode, you can choose any minimum/maximum strain you like. The lower/upper strain limit is the following:

Model

CYCLE 400

Power Range

15-400 W *

Steps

5 W

Rotational Speed

20-120 rpm

CYCLE 407 MED

CYCLE 450

15-400 W *

15-400 W *

5 W

5 W

20-120 rpm

20-120 rpm

CYCLE 457 MED 15-400 W * 5 W 20-120 rpm

* Default setting is 25 W; the strain can be reduced to a minimum of 15 W by pushing the MINUS button.

CYCLE 407/457 MED: Please note that this model does not comply with the standard

DIN VDE 0750-238 when used with a strain level below 25 W!

7.1.2 PROFILES

You can choose between performance profiles, gradient profiles (only on CYCLE 450) and heart rate profiles (only on CYCLE 450 and CYCLE 457 MED), depending on the model.

Performance profiles (CYCLE 400, CYLCE 450, CYCLE 457 MED):

Choose between five predefined profiles (P1 - P5) and four individual user profiles

(only on CYCLE 450 and CYCLE 457 MED). The profiles differ in their sequences of alternating strain levels, comparable with a hilly landscape.

The 5 predefined profiles (P1 - P5) are:

Profile 1:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Profile steps

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

3 3

Operation

Profile 2:

Profile 3:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Profile steps

Profile 4:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Profile steps

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Profile steps

3 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Profile 5:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Profile steps

Gradient profiles (CYCLE 450):

Choose between four individual user profiles.

Heart rate profiles (CYLCE 450, CYCLE 457 MED):

Choose between four individual user profiles.

How to choose a predefined user profile (P1-P5):

1. Push the PLUS/MINUS buttons until the programme PROFILES is selected.

Confirm your choice with the START button.

2. In the submenu "PROFILES", choose a profile with the PLUS and MINUS buttons. Confirm your choice with the START button. On CYCLE 450 and

CYCLE 457 MED, push the PLUS and MINUS buttons to choose the type of profile first. Confirm this with START.

3. Now you need to determine the maximum strain intensity. Push the PLUS and

MINUS buttons to set the strain and confirm with START.

4. Now choose the maximum training time. You can choose different training durations of 10 to 60 minutes by pressing the PLUS and MINUS buttons (default time: 10 minutes). Confirm your choice again with the START button.

5. You are now entering training mode. You can change strain intensity within the limits of the pre-assigned strain intensity range by pushing the PLUS and MINUS buttons at any time.

6. The timing will also be displayed in the profile (in italic letters). The training will stop automatically after the chosen duration has elapsed. You can also stop the training by pressing the STOP button at any moment. On CYCLE 450 und

CYCLE 457 MED, the profile distance is repeated until the workout is stopped by pushing the STOP button. In all cases, the training parameters (dist., kcal, km etc.) remain on the display. Pushing the STOP button again will return you

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

3 5

Operation to the main menu.

How to create a new user profile (A-D):

1. Push the PLUS/MINUS buttons until the programme PROFILES is selected.

Confirm your choice with the START button.

2. On CYCLE 450 and CYCLE 457 MED, push the PLUS and MINUS buttons to choose the type of profile first (strain, heart rate, gearshift). Confirm this with

START.

3. If you choose "Gearshift", your weight will be asked. Push the PLUS and MINUS buttons to change the preset weight to your body weight and confirm this with

START.

4. Now push the PLUS/MINUS buttons until the desired user profile (A-D) is selected. Push the STOP button to open the user profile editor.

5. In the user profile editor, you can set the strain intensity per strain level. Push the PLUS and MINUS buttons to set the strain (default: begin profile step 2).

Press the START and STOP buttons for choosing the profile step. Press the

STOP button at profile step 1 and then press the START button or press the

START button at profile step 20 to enter the menu "Training time”.

6. Now choose the maximum training time. You can choose different training durations of 10 to 60 minutes by pressing the PLUS and MINUS buttons (default time: 20 minutes). Confirm your choice with the START button.

7. If you chose the profile "Heart rate", the cardio parameters will be displayed.

Furthermore, you can change the initial strain by pressing the PLUS and MINUS buttons. Confirm your choice with the START button.

8. You are now entering training mode. You can change strain intensity within the limits of the pre-assigned strain intensity range by pushing the PLUS and MINUS buttons at any time. The timing will also be displayed in the profile (in italic letters).

9. On CYCLE 400, the training will end automatically when the chosen duration has elapsed. You can also stop the training by pressing the STOP button at any moment. On CYCLE 450 und CYCLE 457 MED, the profile distance is repeated until the workout is stopped by pushing the STOP button.

7.1.3 CARDIO

For a CARDIO workout (heart rate controlled training) on the machines of the CARDIO

LINE 400 MED, you need a chest band with a POLAR sender. This is the only way to measure your heart rate and control the strain automatically. On models of the CARDIO

LINE 400 HOME, the heart rate can also be assessed through the hand pulse. The

CARDIO mode allows for strain control according to the actual heart rate so that the

3 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED set training heart rate will be maintained during a workout sequence

Users who have a cardiac pace maker or take heart drugs (e.g. beta-blockers) should not exercise in Cardio mode!

Cardio training

You need to set the following parameters before training:

PULSMAX = heart rate upper limit during training

PULSMIN = heart rate lower limit during training

STARTING STRAIN = initial strain at the beginning of the workout

How to proceed when training:

1. Push the PLUS/MINUS buttons until the programme CARDIO is selected.

Confirm your choice with the START button.

2. You are in the "CARDIO" submenu. Set the heart rate upper limit first. Modify the default value (100/min) with the PLUS and MINUS buttons (the heart rate lower limit (default: 90/min) will change synchronously). After choosing your individual heart rate upper limit, confirm with START.

3. Now you can modify your heart rate lower limit with the PLUS and MINUS buttons (default: difference of ten beats from the heart rate upper limit. However, a difference of less than 10 beats cannot be chosen because of training relevance). Reconfirm your choice with START.

4. Now set the starting strain for your exercise. Again, use the PLUS and MINUS buttons and confirm with START for this purpose.

5. The display now shows the CARDIO mode. Your current heart rate is assessed.

The training starts with the first strain level.

6. Stop the exercise by pressing the STOP button. The training parameters (Watt,

1/min etc.) remain on the display. On the CYCLE 450/457 MED model, the gradient of the heart rate will be shown also. The training sequence is marked by two horizontal lines. On CYCLE 457 MED, the average speed is displayed by pressing the START button. Pushing the STOP button again will return you to the main menu.

7.1.4 GEARSHIFT

In this mode, you can exercise for any amount of time. The integrated gearshift allows you to adjust the strain and thus, to simulate realistic cycling.

1. Push the PLUS/MINUS buttons until the programme GEARSHIFT is selected.

Confirm your choice with the START button.

2. Press the PLUS and MINUS buttons to enter your body weight and press the

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

3 7

Operation

START button.

3. You are now entering training mode. In this mode, press the PLUS and MINUS buttons to change the gears at any time and thus, to modify the strain. The change will be displayed for 5 seconds. Furthermore, the gradient can be displayed by pressing the START button. You can modify the gradient settings by pressing the PLUS and MINUS buttons. By pushing the START-button one more time, you will return to the gear selection.

4. Stop the training by pressing the STOP button. The training parameters remain on the display. By pushing the STOP-button one more time, you will return to the main menu. The display will also return to the main menu automatically after 2 minutes if no action is carried out.

If the operational range of the brake is exceeded, an exclamation mark will be displayed next to the performance display. This indicates that the displayed performance does not correspond to the performance on the crank handle. Please modify the speed or the gear to return to the correct performance.

7.1.5 WHO-PROFILES

The WHO profiles are step profiles defined by the World Health Organization (profiles with stepping strain increase). On CYCLE 407 MED, you can choose between the predefined profiles WHO1 and WHO2; on CYCLE 457 MED, the user profile can be set individually

WHO1: 25 W initial strain, increase of 25 W every 2 minutes, recovery strain of 25 W

WHO2: 50 W initial strain, increase of 50 W every 3 minutes, recovery strain of 50 W

Please proceed as follows (CYCLE 407 MED):

1. Push the PLUS/MINUS buttons until the programme WHO-PROFILES is selected. Confirm your choice with the START button.

2. In the submenu, press the PLUS or MINUS button to choose a WHO profile and confirm this with START.

3. You are now in the "WHO PROFILE" training mode. Press the STOP button to stop the strain phase and to proceed to the recovery phase. By pushing the

STOP-button one more time, you will return to the main menu.

Please proceed as follows (CYCLE 457 MED):

Start WHO profile (description with data protocol P01; without the data protocol P01, the sequence resting ECG nr. 4 as well as the remote start sequence will not apply 10 seconds before strain change).

3 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

You need to set the following parameters before training:

Initial strain: Strain of the first strain level [W]

Time:

Strain step:

Duration of each strain step [min]

Intensity of each strain step [W]

Recovery: Strain in recovery phase [W]

How to create a new user profile (A-D):

1. Push the PLUS/MINUS buttons until the programme WHO-PROFILES is selected. Confirm your choice with the START button.

2. Now push the PLUS/MINUS buttons until the desired user profile (A-D) is selected. Push the STOP button to open the user profile editor.

3. Press the PLUS and MINUS buttons to modify each of the parameters. Press

STOP to change to the next parameter. After setting the parameters, confirm this with START. The modified parameters will be saved for other applications.

4. You are now in the training mode. Press the STOP button to stop the strain phase and to proceed to the recovery phase. 10 seconds before the strain step ends, the remote start sequence will be output through the interface.

5. By pushing the STOP-button one more time, you will return to the main menu.

The automatic strain increase in the preset as well as in the individual WHO profiles can be interrupted during operation by pressing the START button. The displayed information changes from ACTIVE to HOLD. In the HOLD mode, the performance can be modified manually with the PLUS and MINUS buttons. If you want to reactivate the automatic strain increase, press the START button again. The display changes from

HOLD to ACTIVE. In both modes, you can change from the strain phase to the recovery phase by pressing the STOP button. The display changes to PASSIVE. This allows the doctor to adjust the strain for his patient more precisely than before.

7.1.6 COUNTDOWN

In diesem Modus können Sie nach einer vorher festgelegten Dauer trainieren und dabei die Belastung selbst festlegen. Das Countdown-Training unterscheidet die

Programme Manuell- und Cardio-Countdown.

Manual countdown:

1. Push the PLUS/MINUS buttons until the programme COUNTDOWN is selected.

Confirm your choice with the START button.

2. Choose the MANUAL programme with the PLUS and MINUS buttons and confirm this with START.

3. Reduce or increase the time to the desired training duration by pressing the

PLUS and MINUS buttons. Confirm this with START.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

3 9

Operation

4. You are now entering training mode. You may change the strain intensity by pushing the PLUS and MINUS buttons at any time.

5. The training will stop automatically after the chosen duration has elapsed.

However, you can stop the exercise at any time by pressing the STOP button.

The training parameters remain on the display. By pushing the STOP button one more time, you will return to the main menu. The display will also return to the main menu automatically after 2 minutes if no action is carried out.

Cardio Countdown:

1. Push the PLUS/MINUS buttons until the programme COUNTDOWN is selected.

Confirm your choice with the START button.

2. Choose the CARDIO programme with the PLUS and MINUS buttons and confirm this with START.

3. Reduce or increase the time to the desired training duration by pressing the

PLUS and MINUS buttons. Confirm this with START.

4. Now set the starting strain for your exercise. Again, use the PLUS and MINUS buttons and confirm with START for this purpose.

5. You are now entering training mode. You may change the strain intensity by pushing the PLUS and MINUS buttons at any time.

6. The training will stop automatically after the chosen duration has elapsed. The training parameters remain on the display. By pushing the STOP-button one more time, you will return to the main menu. The display will also return to the main menu automatically after 2 minutes if no action is carried out.

4 0

7.1.7 TEST

These tests are step tests. The data resulting from these tests are recommendations only and not suitable for diagnostic purposes!

To carry out a test, a chest band with a POLAR sender is needed. Otherwise, the heart rate cannot be assessed. An assessment of the hand pulse may only be carried out for monitoring purposes!

In the training mode TEST, you can choose from three different test programmes:

Depending on your age, actual fitness or health conditions, you need to choose one of the preset heart rates (130, 150, and 170).

A PWC test is a submaximal test with a preset heart rate:

PWC 130 = heart rate: upper limit of 130, untrained test person of advanced age

PWC 150 = heart rate upper limit of 150, untrained younger test person

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

PWC 170 = heart rate upper limit of 170, trained test person

On PWC 130, the initial strain is 25 W; on PWC 150 and PWC 170, the initial strain is

50 W each. The strain increases by 25 W (PWC 130) and by 50 W (PWC 150, PWC

170) every two minutes. An assessment will only be carried out after the set heart rate has been reached and the set strain step has been ended.

The test result will be a Watts/kg value. With this value and the PWC classification table (see Chapter A8 "Test evaluation"), you can assess your actual fitness level.

How to carry out a test:

1. Push the PLUS/MINUS buttons until the programme TEST is selected. Confirm your choice with the START button.

2. You are now within the "TEST" submenu. Choose the desired test programme with the PLUS and MINUS buttons and confirm this with START.

3. Now choose your gender with the PLUS and MINUS buttons and again confirm your choice with START (M= male; F= Female).

4. Press the PLUS and MINUS buttons to enter your age and press the START button.

5. Now enter your body weight. For this purpose, press the PLUS and MINUS buttons (default weight: 70 kg). Confirm this with START.

6. The display now shows the RESTING PULSE ASSESSMENT mode. Your current resting heart rate will be assessed. This takes about 15-20 seconds.

Your resting heart rate is displayed.

7. If the assessment is successful, the test will start after a few seconds. The display changes to operating mode and starts with the first strain step.

8. As soon as your heart rate exceeds the set upper limit (130, 150, 170), the test ends automatically after completion of the strain level.

9. The training parameters remain on the display after the test is completed. By pushing the STOP button one more time, you will return to the main menu.

7.1.8 ECG Choice

An external connection (RS 232, serial interface) is standard equipment on the CYCLE

457 MED. The interface cable is not included in the delivery.

Danger! Some ECG devices cannot be connected with the CYCLE 457 MED. Call

0049-633-24610 for more information.

Control of the ergometer by external devices:

Use the interface cable to connect the bike ergometer with the external device (ECG,

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

4 1

Operation

PC etc.). Then the correct data protocol needs to be set:

Choosing data protocols:

1. Switch off and on your machine for a restart.

2. Simultaneously press the PLUS and MINUS buttons to open the menu of the serial data protocols.

3. Choose the desired protocol (P01, P02 …) with the PLUS and MINUS buttons and confirm this with START. The protocol will be stored for further applications.

See more details on each data protocol in the annexed sheet "Data protocols”.

If the initialisation is received through the interface, the ergometer automatically changes to the operating mode "external control" ("ECG no." and the protocol number will be shown in the middle of the display). It is not necessary to select the menu item.

In addition, the recognised control characters will be displayed for checking purposes.

Unknown commands will be shown as " / ". Printing is not supported during external control.

If disturbances occur between the ECG device and the bike ergometer, potential compensation can be installed (see Chapter 5.5).

7.2 Behaviour after operation

No special notices need to be observed at the end of the training on a CYCLE model of the CARDIO LINE 400/400 MED. Just stop the training. There is no further risk due to free rolling.

4 2

7.3 Function control

Function test: How to proceed:

Brake function

⊗ Switch on the exercise machine. The illumination of the display shows you immediately if the machine is switched on.

⊗ Choose the MANUAL mode and increase the set performance (see Chapter

7.1.1).

⊗ Move the crank handle at the lowest speed range (observe the arrows!). The strain increases. Increase the speed range to its maximum. The strain will decrease. If this is the case, it may be assumed that the operation independent from the rotational speed works fine.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Heart rate

⊗ Check the function of the heart rate measurement (see chapter 5.2). As you know, heart rate measurement requires the POLAR sender and chest band (use the POLAR sender kit as described in Chapter 7.6).

⊗ Check the function of the heart rate measurement by grasping the hand heart rate electrodes with both hands. The heart rate value is displayed (only on

CARDIO LINE 400 HOME). If no heart rate value is displayed, wet your palms and check again.

Other functions

⊗ Make sure that the seat can be adjusted easily.

7.4 Heart rate measurement

For optimal training results, we recommend controlled training by heart rate measurement. In this regard, the POLAR sender allows for constant displaying and monitoring of the heart rate.

7.4.1 POLAR-Gurt und POLAR-Sender

Wet the electrodes (the two rippled rectangular areas on the bottom side) carefully before use. To provide optimal skin contact, you may use a contact gel as it is used in

ECG measurements. Adjust the length of the band to have a close but comfortable fit against the skin underneath the chest muscle. The band should not loosen during exercise. Please also make sure that the belt is fastened with the correct side up - the

POLAR logo needs to be readable correctly when looking at it. You should also make sure that the two band electrodes are not buckled.

A pulse measurement is only possible if the chest band fits correctly. Otherwise, the

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

4 3

Operation display shows an "E" on the position of the pulse indication (incorrect or no pulse transmission). In this case, please check again the correct fit of the chest band.

The sender (the electrodes in particular) should be cleaned and dried after use with warm water and a mild soap for hygienic reasons. Never brush the electrodes! Do not use alcohol for cleaning!

7.4.2 Operating range of the POLAR sender

The operating range of the sender is approx. 80cm. If you have more than one

ERGOFIT exercise machine, make sure to keep a distance of at least 100 cm between them to avoid interference of the senders.

7.4.3 Battery of the POLAR sender

If the heart rate transmission only works within a short distance between the sender and the receiver or not at all after a prolonged operation, it is possible that the sender's battery is depleted (the battery usually lasts for approx. 2500 hours). In this case, please send the sender and the depleted battery to the following address:

POLAR Electro GmbH Deutschland

Am Seegraben 1

D-64572 Büttelborn/Klein-Gerau

Your sender will be returned to you with a new battery at a charge. Do not try to change the battery yourself!

7.4.5 Potential interference

⊗ Screens, electric motors

⊗ High voltage lines (including tramway contact wires)

⊗ Fluorescent tubes in close proximity

⊗ Central heating radiators

⊗ Other electric devices (e.g. mobile phone)

The problem can usually be addressed by moving the machine a few metres away from the source of interference. In some cases, it is sufficient to change the position of the machine by a few degrees.

If the heart rate is displayed irregularly despite faultless technical conditions, check your heart rate manually. In case of doubt, you should see your doctor.

4 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

8.1

8.2

8.3

8.4

8.5

8.6

Chapter 8: Training

The effect of training ............................................................................. 46

Cardiovascular training ......................................................................... 46

Strain parameters ................................................................................. 47

Training routine - aspects to be considered ......................................... 47

Weight reduction: the benefits .............................................................. 47

Training tips .......................................................................................... 48

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

4 5

Training

8 Training

8.1 The effect of training

The demands of today's lifestyle in modern societies are not enough to remain in good conditional shape. Cardiovascular diseases are still the most common cause of death.

This fact should make cardiovascular training a top priority.

All training activities during which the heart rate increases for 15 to 20 minutes or longer are called "aerobic".

8.2 Cardiovascular training

To get the most benefit out of your training, you should be familiar with some training principles.

Your fitness depends to a great extent on your body's ability to provide your muscles with oxygen. Oxygen is the key to the energy stored in the muscles.

Let us take a closer look at some of the factors crucial to this process: The heart and its role as a complex pump is responsible for the smooth blood flow in the body.

Regular aerobic training will increase the heart's stroke volume, i.e. it will transport more blood through your body with every beat. Accordingly, the heart works more efficiently not only during training, but also at rest.

When oxygen enters the lungs, it will be mixed with blood in tiny "air sacs", the socalled alveoli. Regular aerobic training will improve the efficiency of the alveoli and thus, more oxygen will enter the bloodstream and be transported to the muscles.

Haemoglobin is the substance of the blood that absorbs the oxygen. Regular aerobic training will increase the part of the haemoglobin in the blood, which in turn allows your muscles to be better supplied with oxygen.

It is a fact that heart diseases are much less common in people who train regularly.

Overall, it may be said that regular training improves the amount of oxygen supplied to the muscles and the risk of heart disease decreases. For this reason, ERGO-FIT cardio exercise machines are used in both gyms and rehabilitation centres.

4 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

8.3 Strain parameters

The intensity of your training programme should be suitable for your heart rate. It can only be assessed by a targeted performance evaluation. Our cardio training machines allow you to check your heart rate constantly even during training.

If you are a training beginner, it is advisable to train in the lower part of your aerobic training zone until your fitness begins to improve.

8.4 Training routine - aspects to be considered

If you exercise for the first time or restart training after a longer break, your training routine should include the following stages (example):

1. Warm up: Exercise for five minutes and with little strain to prepare your body optimally for training.

2. Stretching: Get off the exercise machine and stretch the muscle groups you are going to target during training.

3. Main sequence: Now you are well prepared for the aerobic phase, which should last a minimum of 15 to 20 minutes. Your target is to maintain your heart rate continuously at the correct strain level.

8.5 Weight reduction: the benefits

Most beginners primarily wish to reduce body weight, the adipose tissue. Regular training stimulates the metabolism, which leads to a higher burning of calories, both during training and at rest.

Most beginners get the most aerobic benefit from a strain level of 70 % of the maximum heart rate. Increasing fitness requires an adaptation of training intensity. However, the assumption "The harder the training the greater the progress in fitness" is wrong.

If a certain strain limit is exceeded, the benefits from aerobic training drop dramatically because the body can no longer provide the muscles with oxygen and instead produces large amounts of lactic acid, and this will make us stop training very fast.

A strain level just below the anaerobic threshold enables us to exercise substantially longer. This way, we will burn a lot of fat and simultaneously strengthen our aerobic system optimally.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

4 7

Training

8.6 Training tips

During training on our exercise machines, your heart rate should not exceed a certain upper limit. Training in the aerobic range requires us to train below this limit. It can be set according to a doctor's specifications, tests in performance diagnostics or general rules of thumb.

The heart rate upper limit is calculated using the following rule of thumb:

180 - age = Heart rate upper limit

As an example, a 50-year-old man would have a maximum heart rate of 130 beats per minute and a 30-year-old person of 150 bpm etc.

4 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

9.1

9.2

Chapter 9: Maintenance

Care and maintenance ......................................................................... 50

Cleaning ............................................................................................... 51

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

4 9

Maintenance

9 Maintenance

All exercise machines used for commercial purposes need to undergo regular a safety inspection/maintenance test.

Regular, thorough care and appropriate maintenance in particular contribute to preserve the value and lifetime of your exercise machine. For this reason, we recommend regular preventive maintenance. Once per week, inspect the casing, seat and its guiding, handlebar and pedals for damage. In case of a damage, have it repaired immediately. These regular inspections are essential for compliance in case of guarantee claims. In case of a malfunction, ERGO-FIT's technicians and engineers are pleased to assist you.

Before switching on the machines, always make sure that mains cable, mains plug, mains socket and power connection units are free from defects.

Immediate maintenance works are to be carried out if:

⊗ the machine has undergone extreme mechanical stress (impact, damaged cable, or inappropriate tensile strain),

⊗ liquid has entered the machine,

⊗ cables, plug-in connectors or casings are damaged,

⊗ covers have dropped.

Maintenance of the machine may be carried out by ERGO-FIT's customer service department. A maintenance contract may be concluded also.

9.1 Care and maintenance

During manufacturing of its training machines, ERGO-FIT makes every effort to reduce future maintenance.

In the following chapters, some of the maintenance and verifications are described.

You should carry out these tasks regularly.

Switch off and unplug the machine before any maintenance is carried out or the machine is opened.

When carrying out maintenance, consider the following:

⊗ This machine requires very little maintenance.

⊗ Moving parts need no further oiling or greasing.

5 0

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

⊗ Once per week, inspect the seat, the handlebar, the pedals, and the casing for damage. If the machine is damaged, have it repaired immediately.

Treadle

⊗ As screws tend to loosen over time, you should check the treadles and pedals after 3 to 5 operational hours for the first time, then every month.

⊗ If a treadle loosens, screw it on again tightly immediately. To do this, remove the black cap from the crankshaft and retighten the screw underneath with a 14mm socket wrench. The pedal can be retightened to the crankshaft with a 15mm open-end wrench.

9.2 Cleaning

Sweat, dust particles and dirt will damage your exercise machine. This can be observed after only a few weeks. Metal and aluminium parts of your machine may alter its surfaces in combination with sweat. For this reason, you should clean your machine daily.

To clean your exercise machine we recommend disinfectant "Ecolab P3-steril" or

"Scarabig". This product is available at:

Ecolab Deutschland GmbH

Reisholzer Werftstraße 38-42

Postfach 13 04 06

40554 Düsseldorf www.ecolab.com

SCARAPHARM chem.-pharm. Produkte GmbH

Wachmannstraße 86

D-28209 Bremen www.scarapharm.de

During cleaning, consider the following aspects:

Make sure to disconnect the mains plug before opening the control panel or the motor cover!

⊗ Clean your machine with a damp cloth, mild cleaning agent or soap suds only and dry it with a soft cloth.

⊗ Avoid any oiling or greasing of the machine's outside.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5 1

Maintenance

5 2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

10.1

Chapter 10: Troubleshooting

Identifying the cause of failure .............................................................. 54

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5 3

Troubleshooting

10 Troubleshooting

Despite the high quality of our ERGO-FIT products, occasional disturbances might occur in rare cases. The target of this chapter is to inform you of the possible causes of these disturbances and to give you the possibility to avoid them. If you assume a technical defect, the machine must no longer be used for safety reasons. If you cleared a defect yourself, it would be very helpful for us if you communicate this defect to us immediately. This allows us to record the failure in the model's documentation file, and this in turn allows for quality improvement.

For safety reasons, unplug the machine before any work is carried out or the machine is opened!

10.1 Identifying the cause of failure

Malfunctions often have a simple cause. However, defective components are sometimes at fault. This chapter provides you with guidelines to resolve possible problems. If the recommendations listed are not successful, please contact our service department immediately. Our service team will be pleased to help you.

Please proceed as follows in case of failure:

The machine does not work (no audible sound when switching on, blank display

⊗ The fuse box is located on the machine's power input module. Pull out the fuse box and check if a fuse is defective (CARDIO LINE 400 MED).

⊗ Did you use an extension cable or a multi-outlet power strip? Always connect your machine directly to the mains socket.

⊗ Check the mains socket. Plug in a different electric device to check the mains socket.

⊗ Pull the mains plug out of the mains socket and visually inspect the mains cable.

An error message appears on the display

⊗ Write down this message immediately after the appearance of the error message.

⊗ Assess the frequency of this error. If yes: when and how often does the error occur?

⊗ Check if more electric devices were connected in parallel. If so, which?

⊗ Check if a button has been pressed when the error message was displayed.

5 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

⊗ If you were not present when the error message was displayed, ask the user about the exact progress of events.

⊗ If an error message appears on the display, please ask our service team

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5 5

Troubleshooting

5 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Chapter A: Appendix

A.6

A.7

A.8 zu A.2

A.2.1

A.2.2

A.2.3

A.2.4

A.2.4

A.9

A.10

A.1

A.2

A.3

A.3.1

A.3.2

A.4

A.5

A.5.1

A.5.2

A.5.3

Customer service ................................................................................. 56

Spare parts ........................................................................................... 56

Technical specifications ........................................................................ 56

CARDIO LINE 400 ............................................................................... 57

CARDIO LINE 400 MED ...................................................................... 58

Electromagnetic interference imunity ................................................... 59

Safety regulations ................................................................................. 62

Safety instructiones .............................................................................. 62

Mark of conformity ................................................................................ 63

Pictographs .......................................................................................... 65

Error margins ........................................................................................ 66

Warranty clauses .................................................................................. 67

Test evaluation ..................................................................................... 69

Spare parts ........................................................................................... 70

CYCLE 400 .......................................................................................... 71

CYCLE 450 ......................................................................................... 77

CYCLE 407 MED ................................................................................ 83

CYCLE 457 MED ................................................................................. 89

Brake unit ............................................................................................. 95

Entry in the registry of medical devices ................................................ 97

Conformities ........................................................................................ 101

Safety information ............................................................................... 103

Please note:

This owner's manual contains information on multiple gym machines.

There may be variations in detail according to the type of machine!

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5 5

Appendix

A Appendix

A.1 Customer service

If you cannot correct a malfunction yourself, please get in touch with our customer service.

Service: Phone:

Fax:

+49 (6331) 2461-22

+49 (6331) 2461-23

+49 (6331) 2461-29

+49 (6331) 2461-55

Email: [email protected]

Repairs of ERGO-FIT machines are carried out by highly qualified and competent service technicians. Only original spare parts are used for repairs.

A.2 Spare parts

Spare parts and current exploded views are available from ERGO-FIT's service department:

Service: Phone:

Fax:

Email:

+49 (6331) 2461-22

+49 (6331) 2461-23

+49 (6331) 2461-29

+49 (6331) 2461-55 [email protected]

When ordering, please provide the following information:

⊗ Model

⊗ Serial number

⊗ Name of the spare part

⊗ Number of the spare part

Please refer to the drawings and tables on page 69 ff. to find the correct number and name of the spare part!!

A.3 Technical specifications

5 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.3.1 CARDIO LINE 400

This chapter provides the technical specifications for your cardio exercise machine.

The specifications are listed in separate charts for each model of the CARDIO LINE

400 series.

Model

Supply voltage

48-60 Hz

Safety standard

Device standard

Protection class

Tested for use in

CYCLE 400

220 - 240 V~

DIN EN 60335-1

EN 957 HA

II

Home use

Accuracy

Brake system

Moment of internia

Dimensions in cm

(L/W/H)

10%, EN 957 HA

Eddy current brake

11 +/- 2 kg.m²

105/50/130

Weight

Speed range

Power range

Steps approx. 45 kg

20 - 120 I/min

15 -400 W

5 W

Trainig programmes MANUAL, PROFILES,

CARDIO, GEARSHIFT

Test programmes

Display parameters

POLAR heart rate measuring

Heart rate dependent strain control

Control

Max. body weight

Interfaces

-

MANUAL, PROFILES,

CARDIO, GEARSHIFT,

TEST

PWC 130, PWC 150, PWC

170

WATT, PULSE, TIME,

KM/H, DIST., I/MIN, KCAL

WATT, PULSE, TIME,

KM/H, DIST., I/MIN, KCAL

1 channel, acuracy of ECG 1 channel, acuracy of ECG with POLAR sender or hand pulse depending on training mode

180 kg

RS 232

CYCLE 450

220 - 240 V ~

DIN EN 60335-1

EN 957 HA

II

Home use

10%, EN 957 HA

Eddy current brake

11 +/- 2 kg.m²

105/50/130 approx. 45 kg

20 - 120 I/min

15 -400 W

5 W with POLAR sender or hand pulse depending on training mode

180 kg

RS 232

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5 7

Appendix

A.3.2 CARDIO LINE 400 MED

This chapter provides the technical specifications for your cardio exercise machine.

The specifications are listed in separate charts for each model of the CARDIO LINE

400 MED series.

Model

Supply voltage 48-

60 Hz

Safety standard

Device standard

Protection class

Tested for use in

Accuracy

Brake system

Moment of internia

Dimensions in cm

(L/W/H)

Weight

Speed range

Power range

Steps

Trainig programmes

Test programmes

Display parameters

POLAR heart rate measuring

Heart rate dependent strain control

Control

Max. body weight

Interfaces

-

I

CYCLE 407 MED

220 - 240 V~

DIN EN 60601-1

DIN VDE 750-238

Medical therapy

5%, DIN VDE 750-238

Eddy current brake

11 +/- 2 kg.m²

105/50/130 approx. 55 kg

20 - 120 I/min

15 -400 W

5 W

MANUAL, PROFILES,

CARDIO, COUNTDOWN

WATT, PULSE, TIME,

KM/H, DIST., I/MIN, KCAL

1 channel, acuracy of

ECG with POLAR sender depending on training mode

180 kg

RS 232

I

CYCLE 457 MED

220 - 240 V ~

DIN EN 60601-1

DIN VDE 750-238

Medical therapy

5%, DIN VDE 750-238

Eddy current brake

11 +/- 2 kg.m²

105/50/130 approx. 55 kg

20 - 120 I/min

15 -400 W

5 W

MANUAL, PROFILES,

WHO-PROFILES, TEST,

CARDIO, COUNTDOWN,

ECG SELECT

PWC 130, PWC 150,

PWC 170

WATT, PULSE, TIME,

KM/H, DIST., I/MIN, KCAL

1 channel, acuracy of

ECG with POLAR sender depending on training mode

180 kg

RS 232

5 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.4 Electromagnetic transmission and interference immunity

Electromagnetic transmission

Products made by ERGO-FIT are designed for use in the environments listed below.

Please make sure to use the product in the appropriate environment.

Measurement of emitteed interference

HF emission in compliance with

CISPR11

Compliance Electromagnetic enivronment

Group 1 The product uses HF energy for internal function only. For this reason, HF emission is very low and therefore, interference with adjacent electronic devices is unlikely

HF emission in compliance with

CISPR11

Emission of harmonics in compliance with

IEC 61000-3-2

Emission of harmonics in compliance with

IEC 61000-3-3

Class B not applicable not applicable

Electromagnetic interference immunity for non-life support devices

Products made by ERGO-FIT are designed for use in the environments listed below.

Please make sure to use the product in the appropriate environment.

Interference immunity test IEC 60601 test level

Conducted HF disturbance in compliance with IEC 61000-

4-6

3 Veff 150 kHz upto 80 MHz

Conducted HF disturbance in compliance with IEC 61000-

4-6

3 V/m 80 MHz upto 2.5 GHz

Compliance level

Electromagnetic enivronment

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

5 9

Appendix

Electromagnetic interference immunity CARDIO LINE 400/400 MED

Products made by ERGO-FIT are designed for use in the environments listed below.

Please make sure to use the product in the appropriate environment.

Inerference immunity test

Discharge of static electricity in comliance with IEC 61000-4-2

IEC 60601 test level

+/- 6 kV contact discharge

+/- 8 kV contact discharge

Electrical fast transient bursts in compliance with IEC 61000-4-4

+/- 2 kV for mains cable

Surge voltages in compliance with IEC

61000-4-5

+/- 2 kV for lead in and lead out

+/- 1 kV opposedmode voltage

+/- 2 kV common voltage

Voltage dips, shortterm loss and fluctuations of the supply voltage in compliance with IEC

61000-4-11

Magnetic field induced by supply frequency

(50/60 Hz) in compliance with

IEC 61000-4-8

Compliance level

6 kV

8 kV

2 kV

2 kV

Electromagnetic environment

Floors should be made of wood, concrete or laid with ceramic tiles.

If the floor is laid with synthetic materials, the relative humidity must be greater than 30 %.

The quality of the supply voltage should conform to a typical business or hospital environment.

The quality of the supply voltage should conform to a typical business or hospital environment

The quality of the supply voltage should conform to a typical business or hospital environment.

The magnetic fields induced by the supply frequency should conform to typical values as they can be found in a typical business or hospital environment.

6 0

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Recommended protective distances between portable as well as mobile HF telecommunication devices and the ERGO-FIT product

Products made by ERGO-FIT are designed for operation in an electromagnetic environment where HF disturbances are controlled. To avoid electromagnetic interference, make sure to maintain the minimum distance between portable as well as mobile HF telecommunication devices and the ERGO-FIT product - depending on the output power - as indicated below.

0,01

0,1

1

10

100

Power output of the sender / W

Protective distance depending on transmission frequency

/ m

150 kHz to 80 MHz d = 3,5/V1*sqrtP

150 kHz to 80 MHz d = 3,5/E1*sqrtP

150 kHz to 80 MHz d = 7/E1*sqrtP

0,12

0,37

0,12

0,37

0,23

0,74

1,17

3,69

11,67

1,17

3,69

11,67

2,33

7,38

23,33

For senders whose maximum nominal output is not listed in the table shown above, the recommended protective distance d (in metres) may be calculated by using the equation belonging to the respective column, whereas P is the maximum nominal output of the sender (in Watts) indicated by the manufacturer of the sender.

Please note:

⊗ With 80 MHz and 800 MHz, the higher frequency range is applied.

⊗ These guidelines may not be applicable in all cases. The propagation of electromagnetic parameters is influenced by absorption and reflection of buildings, objects and persons.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

6 1

Appendix

A.5 Safety regulations

6 2

A.5.1 Safety instructions

To protect the user, the Verband Deutscher Elektrotechniker e.V. (VDE, German electrical engineering technician association) published special instructions for electromedical devices and rooms used for medical purposes.

According to these instructions, devices connected to the grid must be equipped not only with a reliable insulation of live parts but also with an additional protective measure to protect the user against the transfer of the supply voltage to touchable metal parts.

For this purpose, VDE subdivides so-called protection classes.

Of the protection classes allowed for electromedical devices, in most cases protection class I (protective measures with protective earth conductor) as well as protection class II (protective measures without protective earth conductor but double insulation) are used:

In protection class I devices, metal casing parts are connected with the protective earth conductor of the grid through its earthing contact. In case of insulation failure, the upstream circuit breaker will close the circuit.

The CYCLE models of the CARDIO LINE 400 are classified as protection class II devices. The CYCLE models of the CARDIO LINE 400 MED are classified as protection class I devices.

In the user's environment, parts of non-medical electric devices, which are touchable after removing covers, connection devices etc. during routine maintenance, must operate with a tension not exceeding 25 Volts of alternating current or 60 Volts of direct current. In addition, the tension of the power supply must be produced by a separate power source as described in IEC 601-1.

In this case, such a part of the device and the user must not be touched simultaneously.

The use of electromedical devices is restricted to safety-relevant innocuousness taking account of the state of the art, health and safety regulations and accident prevention. Protective measures must be taken to avoid both direct and indirect contact. Covers, coatings, insulation of energized parts in combination with protective measures using protective earth conductors (in compliance with protection class I), melting fuses as well as the observation of distances between devices are all part of this.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

The distance to be maintained is 1.5 metres as a practical value. With this distance, two training devices cannot be connected conductively by a person and it is unlikely that users will receive an electric shock while training.

The instructions given in this chapter refer to the safety model as it is recognised in

Germany. These instructions may vary in other countries.

A.5.2 Mark of conformity

The CE mark attached to the cardio machine (CYCLE 407/457 MED only) refers to the EMC conformity (electromagnetic conformity in compliance with EMC Directive

2004/108/EC, Medical Device Directive 93/42/EEC as well as Low Voltage Directive

2006/95/EC). The CE mark attached to the cardio machine (CYCLE 400/450 without

MED) refers to the EMC Directive 2004/108/EC and the Low Voltage Directive

2006/95/EC. The tests are based on interference emission and interference immunity parameters.

The ERGO-FIT exercise machines of the CARDIO LINE 400 MED are manufactured in accordance with the strictest safety and quality standards and are designed for commercial use.

The CYCLE models of the CARDIO LINE 400 series are manufactured in compliance with the guideline 2004/108/EEC (EMC), the requirements of the standards EN 957 and EN 60335-1 and are designed for home use.

The CYCLE models of the CARDIO LINE 400 MED series are manufactured in compliance with the guidelines 93/42/EEC (MDD) and 2004/108/EC (EMV) as well as the requirements of the standards EN 957, DIN VDE 750-238, EN 60601-1 and EN

60601-1-2.

Furthermore, all CYCLE models of the CARDIO LINE 400/400 MED are subjected to the standard DIN 62304.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

6 3

Appendix

The type label on the device contains the information as shown in the drawing below

(example taken from CYCLE 407 MED):

Mains supply values

Fuse model

Type B

CE mark (for medical devices with notified body CE 0297)

6 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.5.4 Pictographs

The pictographs used on ERGO-FIT devices comply with the standards IEC 417 and

IEC 878.

The following pictographs are used:

Alternating current

Earthing terminal

Earthing

Potential compensation

Devives of protection class II

Danger! See accompanying documents

OFF (Power supply, connection with the grid)

ON (Power supply, connection with the grid)

Device of type B

Device of type BF

Dangerous electric tension

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

6 5

Appendix

A.6 Error margins

The following error margins are valid for the CYCLE models of the CARDIO LINE 400

MED series in compliance with the standard DIN VDE 750-238:

1. The read-out error of the wattage p must not exceed ± 5% of the displayed value. However, the read-out error is not required to fall below ± 3W.

2. The read-out error of the rotational speed n is set to a maximum of ± 2 min-1 above 40 min-1.

3. The measuring device to assess the output calculated on the base of deceleration torque and the rotational speed of the treadle ergometer must not exceed an error margin of 1 %..

The characteristic curve family of the deceleration torque control's work capacity can be assessed from the following figure:

0 100 200

Power

300 400

6 6

The work capacity is shown on the display as follows:

Devices in compliance with

As shown right to the rotational speed

(CYLCE 450/457)

As shown right/left to the rotational speed

(CYCLE 400/407) none

Points

Arrows

VDE 750-238

5% range

EN 957

10% range

10% range 10% range

Deviation > 10% Deviation > 10%

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.7 Warranty clauses

Warranty of 2 years

(cf. general business conditions)

The seller shall be liable, to the exclusion of other liability, for deficiencies of the delivery, of which the absence of expressly affirmed properties is part, as follows:

1. All parts that are found to be unusable or restricted in use in consequence of a circumstance dated before the transfer of risk - notably because of faulty design, bad manufacturing material or faulty workmanship - shall be repaired or replaced during a period of 24 months after delivery. The decision if the fault will be repaired or should be replaced will be at the seller's discretion. The supplier shall only be liable for deficiencies of drawings and materials delivered or chosen by the seller if he would have been able to recognise the deficiency when applying professional accuracy unless the seller notified the purchaser of the recognised deficiency immediately.

2. The purchaser's right to claim ends in all cases after 24 months after transfer of the object.

3. No guarantee shall be assumed for damage as a result of inappropriate or improper use, faulty assembly or faulty startup carried out by the purchaser or a third party, natural wear, faulty or careless handling, inappropriate equipment or exchange work material, faulty construction works, chemical, electrochemical or electric influences, except in the event that they can be traced to the fault of the seller. The purchaser will assume all risks in connection with the delivery, even in case of free freight ex-factory.

4. Initially, the seller has the right to two rectifications or replacements. Should these fail, the purchaser has the right of abatement or rescission within the scope of valid legislation. The seller will have a period of six weeks as from communicating the notice of defect for rectification.

5. Improper modifications or repairs carried out on behalf of the purchaser or a third party without prior permission of the seller will abrogate warranty claims.

6. If goods are exported, warranty will be restricted to the availability of loose spare parts ex-factory within the warranty period. Packing costs, freight charge and labour will be at the expense of the purchaser. In case the purchaser demands on-site repair by a technician of the factory or another service address, the purchaser will assume the costs of travelling expenses and labour.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

6 7

Appendix

7. All merchandise that has not been produced by the seller is subject to legal provisions.

Wear parts are excluded from warranty specifications, e.g.:

⊗ pedal string

⊗ handle bar tube

⊗ seat

⊗ toothed belt

⊗ chains

⊗ lever to fasten the handle bar

⊗ pedals

⊗ seat, tread, handle bars

⊗ fuses

The warranty is voided if maintenance instructions are not observed!

6 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.8 Test evaluation

The classification in fitness level 9 corresponds to a "very good" fitness. An Olympic champion would have a fitness level of 18.

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

6 9

Appendix

7 0

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

zu A.2 Spare parts

A.2.1 CYCLE 400

Fastening torque 20 Nm secured with Loctite 243

Fastening torque 20 Nm secured with Loctite 243

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

7 1

Appendix

36

37

38

39

33

40

41

40

44

14

1

43 glued

35

32

42 45

34

Secured with Loctite 243

3

16

15

15

47

47

16

14

2

7 2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

7 3

Appendix

7 4

Item Part name

1 Foot frame (welded)

2 Neck tube ( welded)

3

4

5

6

Brake unit compl. (see separate spare part list)

Left crank handle

Right crank handle

Cab clips

7

8

Screw with captive washer DIN 933 M8x1x18 (10.9)

Pedals (1 pair)

8.1

Pedal strings long (1 pair)

10 Cable grommet

11 Roller axle

12 Mounting cab D8

13 Polyamide wheel

14 Hexagon svrew DIN 933 M8x20

15 Hexagon nut DIN 958 M8 self-locking

16 Hexagon screw DIN 931 M8x60

17 Left side cover

18 Right side cover

19 Access cover part 1

20 Access cover part 2

21 Foot

22 Bottle holder

23 Tread forming screw DIN 7500 M4x12

24 Edge protection (55mm long))

25 Book support (welded))

26 Handlebar for hand pulse

26.1

Lamella plug D32

26.2

Grommet

26.3

Hand pulse 1,25 Inches (bored)

26.4

Tesa-Moll foamed plastic (100mm long))

26.5

Hand pulse cable, left side

26.6

Hand pulse cable, right side

26.7

Hand pulse printed circuit 400 series

27 Pulse cable

28 Power supply unit, unregulated, euro plug + foot switch

29 Rear cover

30 Label 400 right side

31 Label 400 left side

32 Pad D34,5x7

33 Pad D43x7

34 Headless setscrew DIN 915 M6x10

35 Cylindrical pin DIN 913 M6x10

36 Seat, wide

37 Seat adjustment clamp

38 Setscrew DIN 913 M8x25

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

340.066.KT

340.068.KT

19020

10476

16100

027287F

340071G

16241

92009

340073F

99019

525729

525730

525734E

525611

63024

340.062.KT

40216

40215

340.047.KT

340.018.KT

11072

11065

18030

340058F

11063

Spare part no.

340023F

340024F

340013G

15253

15254

15241

15242

15080

15172

16194

340046F

14020

22014

10077

12038

10055

340.063.KT

340.064.KT

340.065.KT

CARDIO LINE 400/400 MED

Item. Part name

39 Star knob DIN 6336 M8

40 Lamella plug 30x30x2

41 Seat adjustment (welded)

42 Seat tube cover

43 Hexagon nut DIN 439 M6

44 Seat tube ( welded)

45 Handle with plunger (M16x1,5)

46 Thread forming screw DIN 7500 M8x20

47 Fender washer DIN 9021 D8,4

48 Spring lock washer DIN 127 D8

49 EJOT DELTAPT screw D3x25

50 Cheese head screw DIN 912 M8x16 (12.9)

51 Hexagon nut DIN 934 M8

52 Retaining washer (Schnorr) D8

53 Cap M8 SW13

54 Control panel Cycle 400

54.1

Control panel small display

54.2

Touch sensitive membrane keyboard

54.3

Cover RAL 7021

54.4

LCD 400/407

54.5

Printed circuit 400

54.6

LCD connector cable (30cm)

55 Control panel bottom part with RS232

55.1

EJOT DELTAPT screw D2,5x7

55.2

Locking screws 1 set

55.3

Cable RS 232

55.4

Card reader (treadmill) cable

55.5

Printed circuit RS 232

55.6

Control panel bottom part A Bike

55.7

Screw base

55.8

EJOT DELTAPT screw D4x8

56 2-pole double pin case

57 3-pole blade terminal 6,3

58 Cable harness

59 12V-conduction

60 Eprom 3000/3000 S

61 Pan head screw M4x12

62 Thread forming screw DIN 7500 M5x20

63 Hexagon nut DIN 934 M6

64 Self-tapping threaded bushing M5

65 Countersunk DIN 965 M5x50

66 Raised cheese head screw DIN 7985 M6x16

67 Sheet metal screw 3,9x19

68 Sheet metal screw 3,9x19

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

Spare part no.

24041

16039

340.057.KT

340.049.KT

12000

340.051.KT

24054

10475

13021

13004

10461

10331

12009

13161

16144

525727G

027.235.KT

027240E

027.269.KT

525733E

525722E

61034

525738E

10462

75002

525528

525.474

525.483

234139G

77001

10465

87107

87065

525728

61098

525732

10208

10483

12007

12075

10492

10491

10489

10488

7 5

Appendix

7 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.2.2 CYCLE 450

Fastening torque 20Nm secured with Loctite 243

Fastening torque 20Nm secured with Loctite 243

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

7 7

Appendix

36

41

37

46

45

39

38

49

33

45

15

16

43 glued

42

35

52

50

51

40

53

32 44

Secured with Loctite 243

34

Secured with Loctite 243

3

48

47

14

15

56

16

47

56

15

14

1

2

7 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

7 9

Appendix

8 0

4

5

6

7

Item Part name

1 Foot frame (welded)

2

3

Neck tube (welded)

Brake unit compl. (see separate spare part list)

Left crank handle

Right crank handle

Cab clips

Screw with captive washer DIN 933 M8x1x18 (10.9)

8 Pedals (1 pair)

8.1

Pedal strings long (1 pair)

10 Cable grommet

11 Roller axle

12 Mounting cab D8

13 Polyamide wheel

14 Hexagon screw DIN 933 M8x20

15 Hexagon nut DIN 958 M8 self-locking

16 Hexagon screw DIN 931 M8x60

17 Left side cover

18 Right side cover

19 Access cover part 1

20 Access cover part 2

21 Foot

22 Bottle holder

23 Tread forming screw DIN 7500 M4x12

24 Edge protection (55mm long)

25 Book support (welded

26 Handlebar for hand pulse

26.1

Lamella plug D32

26.2

Grommet

26.3

Hand pulse 1,25 Inches (bored)

26.4

Plasticised PVC tube (40mm long))

26.5

Hand pulse cable, left side

26.6

Hand pulse cable, right side

26.7

Hand pulse printed circuit 400 series

27 Pulse cable

28 Power supply unit, unregulated, euro plug + foot switch

29 Rear cover

30 Label 400 right side

31 Label 400 left side

32 Pad D34,5x7

33 Pad D43x7

34 Headless setscrew DIN 915 M6x10

35 Cylindrical pin DIN 913 M6x10

36 Seat, wide

37 Seat adjustment clamp

38 Headless setscrew DIN 913 M8x25

39 Star knob DIN 6336 M8

40 Lamella plug 30x30x2

41 Seat adjustment (welded)

42 Seat tube cover

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

027287F

340071G

16241

92009

340073F

99019

525729

525730

525734E

525611

63024

340.062.KT

40218

40217

340.047.KT

340.018.KT

11072

11065

18030

340058F

11063

24041

16039

340.057.KT

340.049.KT

Spare part no.

340023F

340024F

340013G

15253

15254

15241

15242

15080

15172

16194

340046F

14020

22014

10077

12038

10055

340.063.KT

340.064.KT

340.065.KT

340.066.KT

340.068.KT

19020

10476

16100

CARDIO LINE 400/400 MED

Item Part name

43 Hexagon nut DIN 439 M6

44 Headless setscrew 913 M16x1,5x16

45 Bush Rd. D24x

46 Hexagon screw DIN 931 M8x50

47 Bush Rd. D24x

48 Hexagon screw DIN 931 M8x70

49 Seat tube compl. (welded)

50 Handgrip for operating lever

51 Release lever D8x95

52 Squared release head 15x15x42,5

53 Pneumativ spring HY6 300

54 Cover cap (shortened)

55 Thread forming screwDIN 7500 M8x20

56 Fender washer DIN 9021 D8,4

57 Spring lock washer DIN 127 D8

58 EJOT DELTAPT screw D3x25

59 Cheese head screw DIN 912 M8x16 (12.9)

60 Hexagon nut DIN 934 M8

61 Retaining washer (Schnorr) D8

62 Cap M8 SW13

63 Control panel Cycle 450

63.1

Control panel upper part

63.2

Touch sensitive membrane keyboard

63.3

Cover RAL 7021

63.4

LCD module

63.5

Printed circuit

63.6

LCD cable (CARDIO 3000)

64 Control panel bottom part withRS 232

64.1

EJOT DELTAPT screw D2,5x7

64.2

Locking screws, 1 set

64.3

Cable RS 232

64.4

Card reader (treadmill) cable

64.5

Printed circuit RS 232

64.6

Control panel bottom part A Bike

64.7

Screw base

64.8

EJOT DELTAPT screw D4x8

65 2-pole double pin case

66 3-pole blade terminal 6,3

67 Cable harness

68 12 V-conduction

69 Eprom 3000/3000 S

70 Pan head screw M4x12

71 Thread forming screw DIN 7500 M5x20

72 Hexagon nut DIN 934 M6

73 Self-tapping threaded bushing M5

74 Countersunk head screw DIN 965 M5x50

75 Raised cheese head screw DIN 7985 M6x16

76 Sheet metal screw 3,9x19

77 Sheet metal screw 3,9x19

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

525726G

027.226.KT

027264E

027.269.KT

53042

525723E

61107

525738E

10462

75002

525528

525.474

525.483

234139G

7701

10465

87107

87065

Spare part no.

12000

11071

340061F

10136

340060F

10057

340.080.KT

24010

234073F

234.099.KT

23061

16014

10475

13021

13004

10461

10331

12009

13161

16144

525728

61098

525732

10208

10483

12007

12075

10492

10491

10489

10488

8 1

Appendix

8 2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.2.3 CYCLE 407 MED

Fastening torque 20Nm secured with Loctite 243

Fastening torque 20Nm secured with Loctite 243

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

8 3

Appendix

43

44

45

46

40

50 glued

42

39

47

48

47

51

49

52 secured with Loctite 243

41

3

14

16

15

54

16

54

14

15

1

2

8 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

8 5

Appendix

8 6

Item Part name

1 Foot frame (welded))

2 Neck tube ( welded))

3

4

5

6

Brake unit compl. (see separate spare part list))

Left crank handle

Right crank handle

Cab clips

7

8

Screw with captive washer DIN 933 M8x1x18 (10.9)

Pedals (1 pair)

8.1

Pedal strings long (1 pair)

10 Cable grommet

11 Roller axle

12 Mounting cab D8

13 Polyamide wheel

14 Hexagon svrew DIN 933 M8x20

15 Hexagon nut DIN 958 M8 self-locking

16 Hexagon screw DIN 931 M8x60

17 Left side cover

18 Right side cover

19 Access cover part 1

20 Access cover part 2

21 Foot

22 Bottle holder

23 Tread forming screwDIN 7500 M4x12

24 Edge protection (55mm long)

25 Book support (welded))

26 Label 407 right side

27 Label 407 left side

28 Serrated lock washer DIN 6798 D4,3

29 Double tab connector

30 Raised cheese head screw DIN 7985 M4x10

31 Earth wire

32 Handlebar with tube (without hand pulse)

32.1

Pulse sleeve

32.2

Lamella plug D35

32.3

Lamella plug D32

32.4

Heart rate receiver printed circuit

32.5

Pulse cable

32.6

Foamed plastic pad (0,25)

32.7

Grommet

33 Countersunk head screw DIN 965 M3x12

34 Hexagon nut DIN 934 M3

35 Power supply unit, unregulated

36 Power input module

37 Fine-wire fuse 1A, slow-acting

38 Power-cord 2,5m, unwound, grey

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

340.066.KT

340.068.KT

19020

10476

16100

027287F

40220

40219

13013

49019

10100

61101

340072G

340078F

16077

16241

525737

525611

18019

92009

10103

12004

63019

87100

91005

61037

Spare part no.

340023F

340024F

340013G

15253

15254

15241

15242

15080

15172

16194

340046F

14020

22014

10077

12038

10055

340.063.KT

340.064.KT

340.065.KT

CARDIO LINE 400/400 MED

Item Part name

38 Rear cover, with cutout

39 Pad D34,5x7

40 Pad D43x7

41 Headless setscrew DIN 915 M6x10

42 Cylindrical pin DIN 913 M6x10

43 Seat, wide

44 Seat adjustment clamp

45 Headless setscrew DIN 913 M8x25

46 Star knob DIN 6336 M8

47 Lamella plug 30x30x2

48 Seat adjustment (welded)

49 Seat tube cover

50 Hexagon nut DIN 439 M6

51 Seat tube (welded)

52 Handle with plunger (M16x1,5)

53 Thread forming screwDIN 7500 M8x20

54 Fender washer DIN 9021 D8,4

55 Spring lock washer DIN 127 D8

56 EJOT DELTAPT screw D3x25

57 Cheese head screw DIN 912 M8x16 (12.9)

58 Hexagon nut DIN 934 M8

59 Retaining washer (Schnorr) D8

60 Cap M8 SW13

61 Control panel bottom part

62 Control panel Cycle 400

62.1

Control panel, small display

62.2

Touch sensitive membrane keyboard

62.3

Cover RAL 7021

62.4

LCD 400/407

62.5

Printed circuit 400

62.6

LCD connector cable (30cm)

63 2-pole double pin case

64 3-pole blade terminal 6,3

65 Cable harness

66 12V-conduction

67 Eprom 3000/3000 S

68 Pan head screw M4x12

69 Thread forming screwDIN 7500 M5x20

70 Hexagon nut DIN 934 M6

71 Self-tapping threaded bushing M5

72 Countersunk head screw DIN 965 M5x50

73 Raised cheese head screw DIN 7985 M6x16

74 Sheet metal screw 3,9x19

75 Sheet metal screw 3,9x19

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

Spare part no.

340070F

340.047.KT

340.018.KT

11072

11065

18030

340058F

11063

24041

16039

340.057.KT

340.049.KT

12000

340.051.KT

24054

10475

13021

13004

10461

10331

12009

13161

16144

027.225.KT

525727G

027.235.KT

027240E

027.269.KT

525733E

525722E

61034

87107

87065

525728

61098

525732

10208

10483

12007

12075

10492

10491

10489

10488

8 7

Appendix

8 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.2.4 CYCLE 457 MED

Fastening torque 20Nm secured with Loctite 243

Fastening torque 20Nm secured with Loctite 243

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

8 9

Appendix

43

48

44

53

52

45

46

56

40

52

15

59

58

57

60

61 shortened

47

50 glued

42

49

39

51 secured with Loctite 243 secured with Loctite 243

41

3

15

55

54

63

14

1

14

16

54

63

15

2

9 0

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

9 1

Appendix

9 2

Item Part name

1 Foot frame (welded))

2 Neck tube ( welded))

3

4

Brake unit compl. (see separate spare part list)

Left crank handle

5

6

7

Right crank handle

Cab clips

Screw with captive washer DIN 933 M8x1x18 (10.9)

8 Pedals (1 pair))

8.1

Pedal strings long (1 pair)

10 Cable grommet

11 Roller axle

12 Mounting cab D8

13 Polyamide wheel

14 Hexagon svrew DIN 933 M8x20

15 Hexagon nut DIN 958 M8 self-locking

16 Hexagon svrew DIN 931 M8x60

17 Left side cover

18 Right side cover

19 Access cover part 1

20 Access cover part 2

21 Foot

22 Bottle holder

23 Tread forming screwDIN 7500 M4x12

24 Edge protection (55mm long)

25 Book support (welded)

26 Label 457 right side

27 Label 457 left side

28 Serrated lock washer DIN 6798 D4,3

29 Double tab connector

30 Raised cheese head screw DIN 7985 M4x10

31 Earth wire

32 Handlebar with tube (without hand pulse))

32.1

Pulse sleeve

32.2

Lamella plug D35

32.3

Lamella plug D32

32.4

Heart rate receiver printed circuit

32.5

Pulse cable

32.6

Foamed plastic pad (0,25)

32.7

Grommet

33 Countersunk head screw DIN 965 M3x12

34 Hexagon nut DIN 934 M3

35 Power supply unit, unregulated

35.1

Power input module

36 Fine-wire fuse 1A, slow-acting

37 Power-cord 2,5m, unwound, grey

38 Rear cover, with cutout

39 Pad D34,5x7

40 Pad D43x7

41 Headless setscrew DIN 915 M6x10

42 Cylindrical pin DIN 913 M6x10

43 Seat, wide

44 Seat adjustment clamp

45 Headless setscrew DIN 913 M8x25

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

340.066.KT

340.068.KT

19020

10476

16100

027287F

40222

40221

13013

49019

10100

61101

340072G

340078F

16077

16241

525737

525611

18019

92009

Spare part no.

340023F

340024F

340013G

15253

15254

15241

15242

15080

15172

16194

340046F

14020

22014

10077

12038

10055

340.063.KT

340.064.KT

340.065.KT

10103

12004

63019

87100

91005

61037

340070F

340.047.KT

340.018.KT

11072

11065

18030

340058F

11063

CARDIO LINE 400/400 MED

Item Part name

46 Star knob DIN 6336 M8

47 Lamella plug 30x30x2

48 Seat adjustment (welded))

49 Seat tube cover

50 Hexagon nut DIN 439 M6

51 Headless setscrew DIN 913 M16x1,5x16

52 Bush Rd. D24x

53 Hexagon screw DIN 931 M8x50

54 Bush Rd. D24x

55 Hexagon screw DIN 931 M8x70

56 Seat tube ( welded)

57 Handgrip for operating lever

58 Release lever D8x95

59 Squared release head 15x15x42,5

60 Pneumativ spring HY6 300

61 Cover cap (shortened)

62 Thread forming screwDIN 7500 M8x20

63 Fender washer DIN 9021 D8,4

64 Spring lock washer DIN 127 D8

65 EJOT DELTAPT screw D3x25

66 Cheese head screw DIN 912 M8x16 (12.9)

67 Hexagon nut DIN 934 M8

68 Retaining washer (Schnorr) D8

69 Cap M8 SW13

70 Control panel Cycle 450

70.1

Control panel upper part

70.2

Touch sensitive membrane keyboard

70.3

Cover RAL 7021

70.4

LCD module

70.5

Printed circuit 450

70.6

LCD cable (Cardio 3000)

71 Control panel bottom part with RS232

71.1

EJOT DELTAPT screw D2,5x7

71.2

Locking screws, 1 set

71.3

Cable RS 232

71.4

Card reader (treadmill) cable

71.5

Printed circuit RS 232

71.6

Control panel bottom part A Bike

71.7

Screw base

71.8

EJOT DELTAPT screw D4x8

72 2-pole double pin case

73 3-pole blade terminal 6,3

74 Cable harness

75 12V-conduction

76 Eprom 3000/3000 S

77 Pan head screw M4x12

78 Thread forming screw DIN 7500 M5x20

79 Hexagon nut DIN 934 M6

80 Self-tapping threaded bushing

81 Countersunk head screw DIN 965 M5x50

82 Raised cheese head screw DIN 7985 M6x16

83 Sheet metal screw DIN 7981 3,9x16 (black, galvanised)

84 Sheet metal screw DIN 7981 3,9x38 (black, galvanised)

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

525738E

10462

75002

525528

525.474

525.483

234139G

77001

10465

87107

87065

525728

61098

525732

12208

16144

525726G

027.226.KT

027264E

027.269.KT

53042

525723E

61107

16014

10475

13021

13004

10461

10331

12009

13161

10483

12007

12075

10492

10491

10489

10488

Spart part no.

24041

16039

340.057.KT

340.049.KT

12000

11071

340061F

10136

340060F

10057

340.080.KT

24010

234073F

234.099.KT

23061

9 3

Appendix

9 4

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

A.2.5 Brake unit secured with Loctite 243 for disassembly only

Fastening torque 3,3Nm secured with Loctite 243

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

9 5

Item part name

3.1

Support frame (welded)

3.2

Flywheel shaft rd.D24,5x138,5

3.3

Freewheel flange D60x23

3.4

Flywheel

3.5

I-retaining ringDIN 472 D42,1x1,75

3.6

Ball bearing compensating washer for 6004

3.7

Deep-groove ball bearing RK 6004 2Z 20x42x12

3.8

A-retaining ring DIN 471 D20x1,2

3.9

Freewheel HF2016 20x26x16

3.10

Rd. 18x30 (cut off)

3.11

POM thrust washer D10,5x24x1,5

3.12

Hexagon nut DIN 985 M10 self-locking

3.13

Coil with core

3.14

Hexagon screw DIN 933 M8x80

3.15

Hexagon screw DIN 933 M8x110

3.16

Hexagon nut DIN 985 M8 self-locking

3.17

Cheese head screw DIN 912 M4x25

3.18

Washer DIN 125 D10,5

3.19

Hexagon screw DIN 931 M10x60

3.20

Poly-V-belt 1244 PJ9

3.21

Tension spring Z115V-37X

3.22

Ball bearing compensating washerD12xD27x0,3

3.23

Pedal shaft, compl. (welded))

3.24

A-retaining ringDIN 471 D20x1,2

3.25

Belt tensioner (mounted))

3.26

Fl. 30x6x225

3.27

Deep-groove ball bearing 6300 2RS 10x35x11

3.28

ASK plain bearing Zl 110N D10xD12x6

3.29

Hexagon screw DIN 933 M10x35

3.30

Hexagon screw DIN 933 M6x20

3.31

Hexagon nut DIN 934 M6

3.32

Shim ringDIN 988 10x16x1,0

3.33

Rev counter 0,9m

3.34

Magnet Secolit 215

3.35

Printed circuit MAG IF

3.36

Pan head screw M4x12

3.37

Spacer ring D8,4xD4,3x5

3.38

Shim ring DIN 988 D28xD20x2

3.39

Cheese head screw DIN 912 M6x30

9 6

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

Appendix

340034F

14003

340026G

340020F

15257

16278

10133

10473

12038

10487

13011

10137

15226

23025

13055

Spare part no.

340003F

340044F

340015F

340.010.KT

14007

13032

15003

14003

15113

340028F

13141

12062

140099G

10427

10075

12007

13176

60015

19010

525.087

10208

16044

13180

10264

CARDIO LINE 400/400 MED

A.9 Entry in the Registry of medical devices

In compliance with Art. 11 para. 7 and Art. 7 of the regulation on the erection, operation, and use of medical devices

("MPBetreibV") as of June 29, 1998 (BGBt 1. p. 1762), the person who carries out metrological controls must immediately record the measured values, the measuring method, as well as other evaluation results into the registry of medical devices. As during metrological control of your medical device the registry of medical devices was not available, we ask you to use the following data for your documentation.

Operator:

Company:

Contact person:

__________________________________________

__________________________________________

Street: __________________________________________

Post code, place: __________________________________________

Manufacturer::

ERGO-FIT GmbH & Co. KG, Blocksbergstraße 165, D-66955 Pirmasens

Device identification

Device designation: ___________________________________________

Model: ___________________________________________

Serial number: ___________________________________________

__________________________________________________________

Measuring method and evaluation:

 Leitfaden zu messtechnischen Kontrollen (LMK)

 AAppendix 15 or appendix 23 of calibration regulations (EQ 15 or EQ 23)

 Remarks: _______________________________________

Applied perpendicular: ___________________________

Entry of measured results: see following page(s)

 metrological control o.k.; annual designation of sealing:

metrological control not o.k.; old sealing oliterated

_______________________________

Signature

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

9 7

Appendix

Measuring method and evaluation:

 Leitfaden zu messtechnischen Kontrollen (LMK)

AAppendix 15 or appendix 23 of calibration regulations (EQ 15 or EQ 23)

 Remarks: _______________________________________

Applied perpendicular: ___________________________

Entry of measured results: see following page(s)

 metrological control o.k.; annual designation of sealing:

metrological control not o.k.; old sealing oliterated

_______________________________

Signature

______________________________________________________________

Measuring method and evaluation:

 Leitfaden zu messtechnischen Kontrollen (LMK)

AAppendix 15 or appendix 23 of calibration regulations (EQ 15 or EQ 23)

 Remarks: _______________________________________

Applied perpendicular: ___________________________

Entry of measured results: see following page(s)

 metrological control o.k.; annual designation of sealing:

metrological control not o.k.; old sealing oliterated

_______________________________

Signature

9 8

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

Registration of medical devices add-in card

Operator: _____________________________________________

_____________________________________________

_____________________________________________

1.

2.

Designation of the medical device:

_________________________________________________________

Functional test and introduction:

Functional test carried out on:________________ by:___________________________________

Introduction carried out on:________________ by:___________________________________

Introduced person: ____________________________________

____________________________________

____________________________________

____________________________________

3. Metrological inspection: at least every two years

Next inspection: _______________________________________

by (person's name: _______________________________________

4. Safety inspection/maintenance test: at least every 12 months

Next inspection: _______________________________________

by (person's name): _______________________________________

5. Date, type and consequence of the defect and repeated identical operating

fault: ____________________________________________________

_________________________________________________________

_________________________________________________________

6. Reports of incidents to authorities and manufacturer:

_________________________________________________________

_________________________________________________________

CARDIO LINE 400/400 MED

A.10 Conformities

EC Declaration of conformity

This is to certify that the products:

CYCLE 400

CYCLE 450

fulfil the requirements laid down in the following standards and guidelines

• 2004/108/EC

EMC (electromagnetic compatibility) Directive

• 2006/95/EG

Low voltage directive

• EN 957,

Stationary training equipment

• DIN EN 60335-1, protection class II

Household and similar electrical appliances - Safety – Part 1: General require-

ments

This declaration applies to all machines supplied from 01.01.2011 to 31.12.2011

and has been issued with full responsibility to the producer

ERGO-FIT GmbH & Co. KG

Blocksbergstraße 165

66955 Pirmasens

by:

Michael Resch

(Managing Director)

Pirmasens, 08.02.2011

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1 0 1

Appendix

EG-Konformitätserklärung

EC Declaration of conformity:

CYCLE 407 MED

CYCLE 457 MED

fulfil the requirements laid down in the following standards and guidelines

• 2004/108/EC

EMC (electromagnetic compatibility) Directive

• MPG, MDD 93/42/EEC, class of models IIa

Medical Device Directive

• EN 957

Stationary training equipment

• DIN EN 60601-1, class of protection I, degree of protection type B

Medical electrical equipment – Part 1: General requirements for safety and essential performance

• DIN EN 60601-1-2

Medical electrical equipment – Part 1: General requirements for safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests

• DIN EN 60601-1-6

Medical electrical equipment – Part 1: General requirements for safety and essential performance – Collateral standard: Usability

• DIN VDE 0750-238

Medical electrical equipment – Part 238 Particular requirements for the safety of crank ergometers

• DIN EN 62304

Medical device software - Software life-cycle processes

Notified body:

DQS GmbH

August-Schanz-Straße 21

60433 Frankfurt am Main

Kennnummer 0297

This declaration applies to all machines supplied from 01.01.2011 to 31.12.2011

and has been issued with full responsibility to the producer

ERGO-FIT GmbH & Co. KG

Blocksbergstraße 165

66955 Pirmasens

by:

Michael Resch

(Managing Director)

Pirmasens, 08.02.2011

1 0 2

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

CARDIO LINE 400/400 MED

Safety indications

CARDIO LINE 400

⊗ Do not start the machine before carefully reading the owner's manual.

⊗ Check the power connection before training.

⊗ Never use the device without current and only use after a proper functional test.

⊗ Switch off the machine after training and disconnect it from the grid.

⊗ Wear only appropriate sportswear and footwear during training.

⊗ Never lean on the control panel or the casing and make improper movements.

⊗ Never start training with the maximum strain. Increase its intensity slowly.

⊗ Check if the seat is secured in position before getting on the machine.

⊗ Do not lean over the handlebar and do not shift your body weight over the side of the machine. There is a danger of falling over.

⊗ Children must not use the machine unsupervised and must stay away from the machine and moving parts.

⊗ In case of nausea or dizziness, you should stop training immediately, inform the fitness coach and / or see a doctor.

⊗ If you have a cardiac pace maker or a health condition, see your doctor before using the training machine.

⊗ Do not jump from the training machine during your workout.

⊗ Do not take your feet off the pedals during training.

⊗ Once per week, check the moving parts and the casing for damage. If the machine is damaged, have it repaired immediately.

⊗ The ventilation slots must not be covered. This can cause the machine to overheat.

⊗ Only use the machine for the purposes it is designed for.

⊗ Please consider further safety and operational notices in the manual.

These safety precautions must be displayed where they are visible from the strength exercise machine! All users of the machine must inform themselves of the dangers and safety regulations.

The manufacturer will not be liable for personal injury or material damage

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1 0 3

CARDIO LINE 400/400 MED

Safety Precautions

CARDIO LINE 400 MED

⊗ Do not start the machine before carefully reading the owner's manual.

⊗ Check the power connection before training.

⊗ Never use the device without current and use only after a proper functional test.

⊗ The machine is only to be used after consulting a doctor and / or a supervisor.

The machine must not be used without the presence of a supervisor.

⊗ Switch off the machine after training and disconnect it from the grid.

⊗ Wear only appropriate sportswear and footwear during training.

⊗ Never lean on the control panel or the casing and make improper movements.

⊗ Never start training with the maximum strain. Increase its intensity slowly.

⊗ Check if the seat is secured in position before getting on the machine.

⊗ Do not lean over the handlebar and do not shift your body weight over the side of the machine. There is a danger of falling over.

⊗ Children must not use the machine unsupervised and must stay away from the machine and moving parts.

⊗ In case of nausea or dizziness, you should stop training immediately, inform the fitness coach and / or see a doctor.

⊗ If you have a cardiac pace maker or a health condition, see your doctor before using the training machine.

⊗ Do not jump from the training machine during your workout.

⊗ Do not take your feet off the pedals during training.

⊗ Once per week, check the moving parts and the casing for damage. If the machine is damaged, have it repaired immediately.

⊗ The ventilation slots must not be covered. This can cause the machine to overheat.

⊗ Only use the machine for the purposes it is designed for.

⊗ Please consider further safety and operational notices in the manual.

These safety precautions must be displayed where they are visible from the strength exercise machine! All users of the machine must inform themselves of the dangers and safety regulations.

The manufacturer will not be liable for personal injury or material damage

Technical and optical modifications as well as misprints reserved - © 2011 by ERGO-FIT GmbH & Co. KG

1 0 5

ERGO-FIT GmbH & Co. KG

Blocksbergstraße 165 – D-66955 Pirmasens

Phone: 06331 2461-0 – Fax: 06331 2461-55 [email protected] – www.ergo-fit.de