Vega VEGATOR 256C Controller for conductive probes Product Information
Add to my manuals
20 Pages
Vega VEGATOR 256C is a signal conditioning instrument for level detection in conjunction with conductive probes series EL. It provides many useful features:
- You can adjust the measuring system to the conductivity with VEGATOR.
- VEGATOR 256C can be mounted in various ways: carrier rail mounting, wall mounting.
- It has 1 sensor input and 1 relay output.
- It has a control lamp for indication of the relay status.
- VEGATOR 256C is suitable for areas with Ex approval.
advertisement
Signal conditioning instruments for level switches
Conductive
VEGATOR 256C
VEGATOR 532
VEGATOR 631
Product Information
Content
Content
1 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Type overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 Mounting instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4 Connecting to power supply
4.1
Preparing the connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.2
Wiring plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5 Operation
5.1
Operating system - VEGATOR 256C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.2
Adjustment elements - VEGATOR 256C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.3
Operating system - VEGATOR 532. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.4
Adjustment elements - VEGATOR 532 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.5
Operating system - VEGATOR 631. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.6
Adjustment elements - VEGATOR 631 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
7 Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
8 Product code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2
Take note of safety instructions for Ex applications
Please note the Ex specific safety information which you will find on our homepage www.vega.com\services\downloads and which come with the appropriate instrument with Ex approval. In hazardous areas you should take note of the appropriate regulations, conformity and type approval certificates of the sensors and power supply units. Each VEGATOR with Ex approval is an associated, intrinsically safe instrument and must not be installed in hazardous areas.
Conductive – Signal conditioning instruments for level switches
Product description
1 Product description
VEGATOR
VEGATOR signal conditioning instruments power the connected sensor and output level-dependent switching signals via integrated relay outputs.
VEGATOR 256C, 532 and 632 signal conditioning instruments are suitable for level detection in conjunction with conductive probes series EL.
Functional principle
An alternating voltage is connected to two electrodes with the conductive measurement. The medium is in contact with the electrodes and conducts the current according to the resistor. E.g.
reaching of a certain level in a vessel is detected by the sensor and transferred to the VEGATOR signal conditioning instrument for further processing. The measuring system can be adapted to the conductivity with VEGATOR.
Area of application
The instruments are mainly used for level detection or pump control.
The different signal conditioning instruments have various mounting options.
l
Carrier rail mounting - VEGATOR 256C, 631 l l
Wall mounting - VEGATOR 256C, 631
19"-carrier - VEGATOR 532 l
Single housing (type 505) - VEGATOR 532
Conductive – Signal conditioning instruments for level switches 3
2 Type overview
VEGATOR 256C
Type overview
Applications: Single level detection
Functions:
Sensor input:
Outputs:
Adjustment
1 sensor input
1 relay output
Indication on the instrument: Control lamp for indication of the relay status.
VEGATOR 532 VEGATOR 631
Applications:
Functions:
Sensor input:
Outputs:
Double level detection, double pump control
Adjustment
2 sensor inputs
2 relay outputs, 2 transistor outputs
Indication on the instrument: 2 control lamps for indication of the relay conditions, 2 fault signal lamps
Level detection, pump control
Adjustment
1 sensor input
1 relay output, 1 transistor output
1 control lamp for indication of the relay condition, 1 fault signal lamp
4 Conductive – Signal conditioning instruments for level switches
Mounting instructions
3 Mounting instructions
3 .1
VEGATOR 256C
VEGATOR 256C can be mounted in the following ways: l
Carrier rail 35 x 7.5 according to EN 50022 l
Wall mounting
You can either mount the signal conditioning instrument directly on the wall with 2 screws or plug it onto a carrier rail.
l l l l
Plug connection 2,8 x 0.8 mm
Termi-Point standard connection 1,6 x 0.8 mm
Soldering connection
Screw terminals 2 x 0.5 mm²
You can find further information on mounting in the operating instructions manual of the carrier.
Carrier rail mounting
Place the signal conditioning instrument onto the carrier rail
(35 x 7,5 according to EN 50022) from below and press the instrument against the carrier rail until it snaps in.
Wall mounting
Fasten the instrument directly to the wall by means of two screws
(max. ø 3 mm/0.12 in).
When you are mounting the signal conditioning instrument with Ex approval in a carrier, you have to use a
VEGA Ex module.
In Ex applications, a protection of IP 20 must be maintained. Cover the gaps or free module from the front with appropriate blind covers.
Keep a distance of at least 2 TE (10 mm/0.4 in) to the module cards of other manufacturers.
If you want to mount VEGATOR 532 in the complete left position in the carrier, you have to mount a blind cover with at least 4 TE (20 mm/0.8 in) in front of the module of the signal conditioning instrument.
1
22 mm
( 55 /
64
")
7,5 mm
( 19 /
64
") on
Fig. 1: Drilling template VEGATOR 256C
Fig. 2: Distance to the carrier side
1 Blind cover
3 .2
VEGATOR 532
VEGATOR 532 can be mounted in the following ways: l l
Mounting in single housing type 505 Ex
Mounting in carrier rail BGT596 (Ex)
Mounting in single housing type 505 Ex
The socket of the single housing type 505 Ex can either be screwed directly to the mounting plate or can be plugged onto a carrier rail TS35 x 7.5 according to EN 50022 or TS32 according to EN 50035.
You can find further information on mounting in the operating instructions manual of the single housing type 505 Ex.
Instrument coding
All series 500 signal conditioning instruments are provided with different gaps dependent on type and version (mechanical coding).
The module is provided with coded pins that can be inserted to prevent accidental interchanging of the various instrument types.
3 .3
VEGATOR 631
VEGATOR 532 in Ex version is an auxiliary, intrinsically safe instrument and must not be installed in hazardous areas.
Installation location
Each series 600 VEGATOR consists of the actual signal conditioning instrument as well as a plug-in socket for carrier rail mounting. Because it has protection class IP 30 or IP 20, the instrument is intended to be used in switching cabinets.
VEGATOR 631 in Ex version is an auxiliary, intrinsically safe instrument and must not be installed in hazardous areas.
Mounting in carrier rail BGT596 (Ex)
Mount the respective module (standard or Ex version) to your carrier BGT596 or BGT596 Ex.
The female multipoint connector is available in the following connection versions: l
Wire-Wrap standard connection 1 x 1 mm
The Ex separating chamber must be plugged in before starting with the setup of the Ex versions of VEGATOR 631. The instrument must not be opened.
Conductive – Signal conditioning instruments for level switches 5
Mounting instructions
Mounting
The plug-in socket is constructed for carrier rail mounting according to EN 50022. Power supply is connected to terminals 17 and
18. For neighbouring series 600 signal conditioning instruments, it is possible to continue connection L1 and N directly via the supplied bridges.
Instrument coding
All series 600 signal conditioning instruments are provided with different gaps dependent on type and version (mechanical coding).
The plug-in socket is provided with coded pins that can be inserted to prevent accidental interchanging of the various instrument types.
6 Conductive – Signal conditioning instruments for level switches
Connecting to power supply
4 Connecting to power supply
4 .1
Preparing the connection
4 .2
Wiring plan
Note safety instructions
Always keep in mind the following safety instructions: l l
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should be installed
Take note of safety instructions for Ex applications
In hazardous areas you should take note of the appropriate regulations, conformity and type approval certificates of the sensors and power supply units.
VEGATOR 256C
Level detection
3
4
2
1 min.
max.
1
2
3
0
L1
N
5
6
7
4
10
8
Select power supply
The power supply can be 20 … 253 V AC, 50/60 Hz or
20 … 253 V DC.
Fig. 3: Level detection - VEGATOR 256C
1 Power supply
2 Relay output
3 Mass
4 Max.
Selecting connection cable
Power supply of VEGATOR is connected with standard cable according to the national installation standards.
Standard two-wire cable without screening can be used to connect sensors. If electromagnetic interference is expected, screened cable must be used.
Pump control (min./max. control)
2
3
1
Cable screening and grounding
Connect the cable screen on both ends to ground potential. In the sensor, the screen must be connected directly to the internal ground terminal. The ground terminal outside on the housing must be connected to the potential equalisation.
If potential equalisation currents are expected, the screen connection on VEGATOR must be made via a ceramic capacitor (e.
g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus suppressed, but the protective effect against high frequency interference signals remains.
1
3
2
6
Select connection cable for Ex applications
Take note of the corresponding installation regulations for Ex applications. In particular, make sure that no potential equalisation currents flow over the cable screen.
In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation.
5
4
3
2
1 min.
max.
1
2
3
0
L1
N
6
7
8
4
5
10
Fig. 4: Pumpen control - VEGATOR 256C
1 Power supply
2 Relay output
3 Mass
4 Max.
5 Min.
6 Probe, e.g. EL3
Note:
Mutiple rod probes connected to several signal conditioning instruments or to a multiple channel instrument need a ground rod to keep the signal conditioning instruments from influencing each other.
If there are several VEGATOR, it is absolutely necessary that they be connected identically, i.e. the first supply line to all no. 7 terminals and the second supply line to all no.
8 terminals. Exchanging no. 7 and no. 8 or connecting to different phases is not permitted.
Conductive – Signal conditioning instruments for level switches 7
Connecting to power supply
VEGATOR 532
Double level detection
1 2 3
11
7
2 d b z
+ -
5
L (+)
N (-)
6
10
12
1
2
3
6
8
9
10
28
30
32
16
18
20
+
22
-
24
+
-
Fig. 5: Double level detection - VEGATOR 532
1 Relay output 1
2 Relay output 2
3 Transistor output 2
4 Sensor input - Channel 2
5 Voltage supply
6 Transistor output 1
7 Sensor input - Channel 1
8 Mass
9 Max.
10 Min.
11 Probe, e.g. EL3
Pump control (min./max. control)
1 2 3
11
7
1
2
3
2 d b z
+ -
5
L (+)
N (-)
6
8
9
10
6
28
30
32
10
12
16
18
20
+
22
-
24
+
-
Fig. 6: Pump control (min./max. control) - VEGATOR 532
1 Relay output 1
2 Relay output 2
3 Transistor output 2
4 Sensor input - Channel 2
5 Voltage supply
6 Transistor output 1
7 Sensor input - Channel 1
8 Mass
9 Max.
10 Min.
11 Probe, e.g. EL3
1
2
3
4
1
2
3
4
Note:
Mutiple rod probes connected to several signal conditioning instruments or to a multiple channel instrument need a ground rod to keep the signal conditioning instruments from influencing each other.
Double pump control (min./max.)
1
2
3
4
5
9
5
4
L (+)
N (-)
2
1
3
5
4
6
7
8
2 d b z
+ -
6
10
12
28
30
32
16
18
20
+
22
-
24
+
-
7
8
1
2
3
Fig. 7: Double pump control - VEGATOR 532
1 Relay output 1
2 Relay output 2
3 Transistor output 2
4 Transistor output 1
5 Voltage supply
6 Mass
7 Max.
8 Min.
9 Probe, e.g. EL5
VEGATOR 532 Ex with housing type 505
The terminal designation for the voltage supply and the relay and transistor outputs corresponds to that of the multipoint connector.
Only the connection of the probes must be carried out according to the following illustration.
1 d32 d30 d28 d32 d30 d28
2 d32 d30 d28 1 2 3
1
2
3 1
2
3
Fig. 8: Deviating terminal assignment - Housing type 505
1 Double level detection
2 Pump control (min./max. control)
8 Conductive – Signal conditioning instruments for level switches
Connecting to power supply
VEGATOR 631
Level detection
1 2
5
4
1 2 3 4
6
5 6
9 10
L
(+)
N
(–)
7 8
12 13 14
1
2
3
Fig. 9: Level detection - VEGATOR 631
1 Relay output
2 Power supply
3 Transistor output
4 Mass
5 Max.
6 Probe, e.g. EL1
Pump control (min./max. control)
1
2
3
6
5
4
1 2 3 4
7
1 2 3
5 6
9 10
7 8
12 13 14
L
(+)
N
(–) 1
2
3
Fig. 10: Pumpen control - VEGATOR 631
1 Relay output
2 Power supply
3 Transistor output
4 Mass
5 Max.
6 Min.
7 Probe, e.g. EL3
Note:
Mutiple rod probes connected to several signal conditioning instruments or to a multiple channel instrument need a ground rod to keep the signal conditioning instruments from influencing each other.
Pump control (min./max.) with overfill protection
1
2 3
4
7
6
5
4
1 2 3 4 1 2 3 4
A B
8
2
1
3
4
5 6
9 10
7 8
12 1314
L
(+)
N
(–)
5 6
9 10
L N
(+) (–)
7 8
12 1314
1
2
3
Fig. 11: Pump control with overfill protection - VEGATOR 631
1 Relay output
2 Power supply
3 Transistor output
4 Max. overfill protection
5 Min.
6 Max.
7 Mass
8 Probe, e.g. EL3
A Overfill protection
B Min./max. control
Conductive – Signal conditioning instruments for level switches 9
Operation
5 Operation
5 .1
Operating system - VEGATOR 256C min.
max.
1
2
3
0
L1
N
5
6
7
8
4
10
Fig. 12: Indicating and adjustment elements
1 Signal lamp - Relay output
2 Potentiometer for switching point adjustment
5 .2
Adjustment elements - VEGATOR 256C
Control lamp
The yellow relay control lamp (LED) shows the switching condition of the relay.
Generally the relay control lamp shows the activated (energized) condition of the relay.
A dark relay control lamp means that the relay is deenergised.
Potentiometer for switching point adjustment
A potentiometer for switching point adaptation is located on the front plate of the signal conditioning instrument. With this potentiometer you can adapt the measuring system to the conductivity of the product.
5 .3
Operating system - VEGATOR 532
1
1
2
2
6
BR1
BR2
5 9 8 7 4
Fig. 13: Indicating and adjustment elements - Circuit board
1 Ex fuse T50 mA/253 V
2 Mains fuse T1 A/253 V
3 Wiring plan
4 Connection plug board
5 Soldering bridge for fault message adjustment - Channel 1
6 Soldering bridge for fault message adjustment - Channel 2
7 Selection switch (DIL switch) mode AB - Channel 1
8 Selection switch (DIL switch) mode AB - Channel 2
9 Transparent cover (lead-sealable)
3
5
6
7
8
3
4
1
2
9
5
0 10
1
5
0 on
VEGA
10
2
Fig. 14: Indicating and adjustment elements - Front plate
1 Fixing screw (lead-sealable)
2 Potentiometer for switching point adjustment - Channel 1
3 Control lamp - Relay output (LED) - Channel 1
4 Control lamp - Fault message (LED) - Channel 1
5 Potentiometer for switching point adjustment - Channel 2
6 Control lamp - Relay output (LED) - Channel 2
7 Control lamp - Fault message (LED) - Channel 2
8 Control lamp - power supply (LED)
9 Fixing screw
5 .4
Adjustment elements - VEGATOR 532
Control lamps
Control lamps (LED) in the front plate indicate operation, switching status and fault signal.
l
Green l
-
-
Red
Operating control lamp
Mains voltage on, instrument is operating
-
Failure lamp l
-
Fault on the sensor circuit due to sensor failure or line break
If the fail safe relay is deenergized, the red failure lamp will
light
Yellow
-
-
Relay control lamp
The yellow relay control lamp reacts depending on the set
mode (A/B)
Generally the relay control lamp shows the activated
(energized) condition of the relay
-
A dark relay control lamp means that the relay is deenergised (transistor blocks)
Potentiometer for switching point adjustment
Two potentiometers for switching point adaptation are located on the front plate of the signal conditioning instrument. With this potentiometer you can adapt the measuring system to the conductivity of the medium separately for each channel.
Use a small screwdriver to carry out adjustments on the potentiometer.
DIL switch - Mode
One slide switch per channel is located on the circuit board of the signal conditioning instrument. Set the requested mode before inserting VEGATOR because the switch will no longer be accessible in assembled condition.
l
A - Max. detection or overfill protection l
B - Min. detection or dry run detection
10 Conductive – Signal conditioning instruments for level switches
Operation
Selection of the mode
You can set mode A or B by means of the selection switch.
Mode A
Preferrably as overflow protection, compulsory as overfill protection.
Fig. 15: Mode A - Overfill protection
1
2 3
Means with covered max. electrode: l l l
Relay (channel 1) denergizes, connection d10/b10 connected through relay
Transistor output (channel 1) blocks
Control lamp output (channel 1) extinguishes
Means with uncovered max. electrode (level detection) or min.
electrode (pump control): l
Relay (channel 1) energizes, connection d10/z10 connected l l through relay
Transistor output (channel 1) conducts
Control lamp output (channel 1) lights
Fault signal adjustment, bridge
To monitor the electrodes and their circuits, a resistor of 220 kΩ must be mounted between connection 1 and connection 2 in the connection housing of the probe, i.e. with single point control, the measuring and earth electrodes are monitored, and with double point control, the max. and earth electrodes are monitored.
If you want to use both channels for two separate measurements
(no two-point control), the measuring electrodes are monitored, e.
g. ground and max. electrode of channel 1 (connection d30 and d32) and ground and min. electrode of channel 2 (connection z30 and z32).
By doing this, a failure is triggered separately for each channel. In case of failure: l the channel-specific control lamp fault signal lights l l the relay output of the concerned channel deenergises the transistor output of the concerned channel is blocked l
The functions of the inteferred channel remain
For probes with 220 kΩ resistor, a soldering bridge must be closed on the circuit board of VEGATOR.
l l
Channel 1 - Soldering bridge BR1
Channel 2 - Soldering bridge BR2
1
2
BR1
BR2
Mode B
Preferrably as dry run protection system.
Fig. 17: Fault message adjustment - Soldering bridges
1 Soldering bridge for fault message adjustment of channel 1 (BR1)
2 Soldering bridge for fault message adjustment of channel 2 (BR2)
1
2 3
Note:
The line monitoring and the fault signal are inactive with this bridge.
If a measuring system is used as part of an overfill protection system, the bridge on the signal conditioning instrument must not be closed. With Ex probes, the resistor in the probe housing is already present.
Fig. 16: Mode B - Dry run protection
Means with covered max. electrode: l
Relay (channel 1) energizes, connection d10/z10 connected l l through relay
Transistor output (channel 1) conducts
Control lamp output (channel 1) lights
Means with uncovered max. electrode (single point control) or min. electrode (double point control): l
Relay (channel 1) denergizes, connection d10/b10 connected through relay l l
Transistor output (channel 1) blocks
Control lamp output (channel 1) extinguishes
Conductive – Signal conditioning instruments for level switches 11
Operation
5 .5
Operating system - VEGATOR 631
5
1 2 3 4 1 2 3 4
6
7
2
3
A
B
C
1
!
5
0 10 on
631Ex
7 8 5 6
9 10 11 12 13 14
3
4
1
2
7
8
9
N
L1
-
+
5 6 7 8
9 10 11 12 13 14
8
9
10
11
12
Fig. 18: Indicating and adjustment elements
1 Control lamp - Fault signal (LED red)
2 Potentiometer for switching point adjustment
3 Control lamp - Output (LED yellow)
4 Control lamp - Power supply (LED green)
5 Bridge for adjusting the fault signal
6 Terminals for probe
7 Function coding Ex version
8 Instrument coding
9 Sockets for bridges
10 Transistor output
11 Relay output
12 Power supply sec t
12
6
2
A off
B
1 l l l
Green
-
Operating control lamp
-
Red
Mains voltage on, instrument is operating
-
Failure lamp
Fault on the sensor circuit due to sensor failure or line
break
If the fail safe relay is deenergized, the red failure lamp will
light
Yellow
-
Relay control lamp
-
The yellow relay control lamp reacts depending on the set mode (A/B)
-
-
Generally the relay control lamp shows the activated
(energized) condition of the relay
A dark relay control lamp means that the relay is deenergised (transistor blocks)
Potentiometer for switching point adjustment
A potentiometer for switching point adaptation is located on the front plate of the signal conditioning instrument. With this potentiometer you can adapt the measuring system to the conductivity of the product.
DIL switch block
On top to the side (covered when mounted) there is a DIL switch block with four switches. The individual switches are assigned as follows: l
1 - A/B mode
-
A - Max. detection or overfill protection l l l
-
B - Min. detection or dry run detection
2 - Integration time 2 s
3 - Integration time 6 s
4 - Integration time 12 s
With switch 1 you can adjust the mode (A - overfill protection or B dry run protection).
Sensor
1 2 3 4 out max: 36V
60mA
+ -
5 6
CE a
: -20É+60¡C
IP 30
I nsp.
max.250VAC
125VA
2A
12 13 14 power supply
20É250 VAC
L
20É72VDC
1...8 VA
N
+ -
9 10
1 2 3 4
2
Fig. 19: Indicating and adjustment elements
1 DIL switch block
2 Type label
3 Transparent cover
3
5 .6
Adjustment elements - VEGATOR 631
Control lamps
Control lamps (LED) in the front plate indicate operation, switching status and fault signal.
Fig. 20: DIL switch block
In the example (see previous illustration), mode A (max. detection or overfill protection) is selected (switch 1). The integration time is set to 8 seconds (switch 2, 3 and 4).
You can adjust the integration time accordingly with switch 2, 3 and 4. The time periods of the activated time switches add up.
The set time applies to the switch on as well as the switch off delay.
Fault signal adjustment, bridge
To implement line monitoring, you have to mount a resistor of
220 kΩ between terminals 1 and 2 in the connection housing of the probe. This means that with single point control, the measuring and ground electrode are monitored and with two-point control, the max. and ground electrode.
12 Conductive – Signal conditioning instruments for level switches
Operation
When a fault message is generated, the switching output is simultaneously activated.
1
1
2
1 3
2
3
1
2 3
1 2 3 4
Fig. 21: Measuring system with line monitoring
1 Resistor (220 kΩ)
If a fault message is not wanted, a bridge must be provided on the signal conditioning instrument instead of the resistor in the connection housing of the probe.
Note:
The line monitoring and the fault signal are inactive with this bridge.
Fig. 23: Mode A - Overfill protection
Means with covered max. electrode: l
Relay deenergises, connection 12 - 13 is connected through l l relay
Transistor output blocks
Control lamp - Output extinguishes
Means with uncovered max. electrode (level detection) or min.
electrode (pump control): l
Relay energises, connection 12 - 14 is connected through relay l l
Transistor output is conductive
Control lamp - Output lights
1
1
2
3
2
3
1 2 3 4
1
Mode B
Preferrably as dry run protection system.
Fig. 22: Measuring system without line monitoring
1 Bridge between terminal 3 and 4
1
2 3
If a measuring system is used as part of an overfill protection system, the bridge on the signal conditioning instrument must not be closed. With Ex probes, the resistor in the probe housing is already present.
You can find the electrical connection of VEGATOR in the operating instructions manual of the corresponding signal conditioning instrument.
Line monitoring
Fig. 24: Mode B - Dry run protection
Means with covered max. electrode: l l l
Relay energises, connection 12 - 14 is connected through relay
Transistor output is conductive
Control lamp - Output lights
With Ex versions, this resistor of 220 kΩ is already integrated Ex factory in the connection housing of the probe.
The Ex measuring system (max. and ground connection cable of the probe to the signal conditioning instrument) is generally monitored for line break.
Means with uncovered max. electrode (level detection) or min.
electrode (pump control): l
Relay deenergises, connection 12 - 13 is connected through relay l l
Transistor output blocks
Control lamp - Output extinguishes
Mode selection switch
You can set mode A or B by means of the selection switch.
Mode A
Preferrably as overflow protection, compulsory as overfill protection.
Conductive – Signal conditioning instruments for level switches 13
Technical data
6 Technical data
General data
VEGATOR 256C
Series
Weight
Housing material
VEGATOR 532
Series
Weight
VEGATOR 631
Series
Weight
Housing material
Socket material
Voltage supply
VEGATOR 256C
Supply voltage
- Standard
- optional
Max. power consumption
VEGATOR 532
Supply voltage
Max. power consumption
Fuse
- Supply area
- Switching o ff possibility
VEGATOR 631
Supply voltage
Max. power consumption
Fuse
- Supply area
- Switching o ff possibility
Sensor input
VEGATOR 256C
Quantity
Response resistor
Meas. circuit
Switching hysteresis
VEGATOR 532
Quantity
Response resistor
Parallel resistor for fault monitoring
Meas. circuit
Permissible line capacitance
Switching hysteresis
VEGATOR 631
Quantity
Response resistor
Parallel resistor for fault monitoring
Meas. circuit
14
Instrument for wall mounting or mounting on carrier rail 35 x 7.5 or 35 x 15 according to EN 50022
170 g (6 oz)
Noryl SE100, Lexan 920A
19" module card, multipoint connector according to DIN 41612, including transparent cover (lockable)
150 g (5.3 oz)
Module unit with plug-in socket for mounting on carrier rail 35 x 7.5 or 35 x 5 according to EN 50022
170 g (6 oz)
Noryl SE100, Lexan 920A
Noryl SE100, Noryl SE1 GFN3
200 … 253 V AC, 50/60 Hz
24 V, 42 V, 48 V, 100 … 130 V AC (+10 %, -15 %)
1 VA
20 … 53 V AC, 50/60 Hz, 20 … 72 V DC
2 W, 3 VA
T 1 A, 253 V min. 35 A at 253 V AC or 125 V DC
20 … 253 V AC, 50/60 Hz, 20 … 72 V DC
1.5 W (1 … 9 VA)
T 315 mA, 253 V min. 35 A at 253 V AC or 125 V DC
1 x level detection or 1 x pump control (min./max.)
1 … 200 kΩ adjustable approx. 12 V eff., max. 1 mA
20 %
2 x level detection or 2 x pump control (min./max.)
1 … 200 kΩ adjustable
220 kΩ max. 5 V eff., max. 1 mA
1 x 100 nF or 2 x 70 nF with min./max. control
15 %
1 x level detection or 1 x pump control (min./max.)
1 … 200 kΩ adjustable
220 kΩ max. 5 V eff., max. 1 mA
Conductive – Signal conditioning instruments for level switches
Permissible line capacitance
Switching hysteresis
Relay output
VEGATOR 256C
Quantity
Mode
Integration time
Contact
Contact material
Turn-on voltage
Switching current
Breaking capacity
VEGATOR 532
Quantity
Mode
Integration time
Contact
Contact material
Turn-on voltage
Switching current
Breaking capacity
VEGATOR 631
Quantity
Mode
Integration time
Contact
Contact material
Turn-on voltage
Switching current
Breaking capacity
Transistor output
VEGATOR 256C no transistor output
VEGATOR 532
Number, function
Galvanic separation
Maximum values
- U B
- I B
Transistor voltage loss (U
CE
)
Inverse current (I
0
)
VEGATOR 631
Number, function
Galvanic separation
Maximum values
- U B
- I B
Transistor voltage loss (U
CE
)
Inverse current (I
0
)
Conductive – Signal conditioning instruments for level switches
Technical data
1 x 100 nF or 2 x 70 nF with min./max. control
15 %
1 (1 x level detection)
Max. detection or overfill protection
500 ms
Changeover contact (spdt)
AgNi 0.15 hard gold-plated
≥ 10 mV DC, ≤ 253 V AC, 253 V DC
≥ 10 µA DC, ≤ 5 A AC, 1 A DC
≤ 750 VA, ≤ 54 W
2
A/B switch (A - max. detection or overfill protection, B - min. detection or dry run protection) adjustable separately for each channel
500 ms
1 spdt for each output
AgNi, hard gold-plated
≥ 10 mV DC, ≤ 253 V AC, 253 V DC
≥ 10 µA DC, ≤ 3 A AC, 1 A DC
≤ 750 VA, ≤ 54 W DC
1
A/B switch (A - max. detection or overfill protection, B - min. detection or dry run protection)
500 ms
1 spdt for each output
AgNi, hard gold-plated
≥ 10 mV DC, ≤ 253 V AC, 253 V DC
≥ 10 µA DC, ≤ 3 A AC, 1 A DC
≤ 750 VA, ≤ 54 W DC
1 output, synchronously switching with the relay
Floating
36 V DC
≤ 60 mA approx. 1.5 V at I
B
≤ 10 µA
60 mA
1 output, synchronously switching with the relay
Floating
36 V DC
≤ 60 mA approx. 1.5 V at I
B
60 mA
≤ 10 µA
15
Technical data
Adjustment elements
VEGATOR 256C
Control lamp
Potentiometer
VEGATOR 532
DIL switch
Potentiometer
Control lamps in the front plate
- Status indication operating voltage
- Status indication fault signal
- Status indication switching point control
VEGATOR 631
DIL switch block
Potentiometer
Control lamps in the front plate
- Status indication operating voltage
- Status indication fault signal
- Status indication switching point control
Ambient conditions
VEGATOR 256C
Ambient temperature
Storage and transport temperature
VEGATOR 532
Ambient temperature
Storage and transport temperature
VEGATOR 631
Ambient temperature
Storage and transport temperature
Electromechanical data
VEGATOR 256C
Spring-loaded terminals
VEGATOR 532
Electrical connection
- Carrier BGT596 Ex
- Housing type 505 Ex
VEGATOR 631
Screw terminals to the indication of the relay switching status to the adaptation of the product conductivity for preadjustment of the mode for switching point adjustment
Signal lamp green (LED)
Signal lamp red (LED)
Signal lamp yellow (LED) for preadjustment of the integration time and mode for switching point adjustment
Signal lamp green (LED)
Signal lamp red (LED)
Signal lamp yellow (LED)
-20 … +50 °C (-4 … +122 °F)
-40 … +70 °C (-40 … +158 °F)
-20 … +60 °C (-4 … +140 °F)
-40 … +70 °C (-40 … +158 °F)
-20 … +60 °C (-4 … +140 °F)
-40 … +70 °C (-40 … +158 °F) for wire cross-section up to 1.5 mm² (AWG 16)
33-pole multipoint connector, series F (d, b, z) with coding holes
Screw terminal for wire cross-section up to 1.5 mm² (AWG 16) for wire cross-section up to 1.5 mm² (AWG 16)
Electrical protective measures
VEGATOR 256C
Protection
Protection class
VEGATOR 532
Protection
- Signal conditioning instrument - not mounted
- mounted into BGT596 Ex - front side (completely equipped)
- mounted into BGT596 Ex - upper and lower side
- mounted into BGT596 Ex - wiring side
- mounted into housing type 505 Ex
Overvoltage category
IP 20
II
IP 00
IP 30
IP 20
IP 00
IP 30
II
16 Conductive – Signal conditioning instruments for level switches
Protection class
VEGATOR 631
Protection
- Signal conditioning instrument
- Plug-in socket
Overvoltage category
Protection class
Electrical separating measures
Approvals 1 )
VEGATOR 532
ATEX
Others
VEGATOR 631
ATEX
Others
Technical data
II
IP 30
IP 20
II
II reliable separation (VDE 0106, part 1) between power supply, sensor input, level relay and transistor output
ATEX II (1) GD [EEx ia] IIC
WHG
ATEX II (1) G [EEx ia] IIC
WHG
Ship approval
1)
Deviating data in Ex applications: see separate safety instructions.
Conductive – Signal conditioning instruments for level switches 17
7 Dimensions
VEGATOR 256C
66 mm (1 19 /
32
")
Fig. 25: VEGATOR 256C
VEGATOR 532 min.
max.
0
L1
N
5
6
7
8
10
4
1
2
3
37 mm
(1 29 /
64
")
1
9 mm ( 23 /
64
")
5,5 mm ( 7 /
32
")
Fig. 26: VEGATOR 532
1 Male multipoint connector
VEGATOR 631
162 mm (6 3 /
8
")
1
! on
VEGA
25,4mm
(1")
2
134mm (5 9 /
32
")
Fig. 27: VEGATOR 631
1 Transparent cover
2 Carrier rail 35 x 7.5 or 35 x 15 according to EN 50022
3 Ex separating chamber on
5 6 7 8
9 10 12 13 14
36mm (1 27 /
64
")
3
Dimensions
18 Conductive – Signal conditioning instruments for level switches
8 Product code
VEGATOR 256C
TOR256C.X
Operating voltage
E 24VAC
D 42VAC
C 48VAC
B 100...130VAC
A 200...250VAC
VEGATOR 532
TOR532
Approval
EX.A ATEX II (1) GD [EEx ia] IIC + WHG
VEGATOR 631
TOR631
Approval
.X Without
EX.A ATEX II (1) G [EEx ia] IIC + WHG
.M Ship approval
Plug-in socket
K Inclusive plug-in socket
Product code
Conductive – Signal conditioning instruments for level switches 19
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
Phone +49 7836 50-0
Fax +49 7836 50-201
E-Mail: [email protected]
www.vega.com
Subject to change without prior notice
You can find at www.vega.com downloads of the following
l
operating instructions manuals
l
menu schematics
l
software
l
certificates
l
approvals and much, much more
33065 -EN-071204
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Product description
- 4 Type overview
- 5 Mounting instructions
- 7 Preparing the connection
- 7 Wiring plan
- 10 Operating system - VEGATOR 256C
- 10 Adjustment elements - VEGATOR 256C
- 10 Operating system - VEGATOR
- 10 Adjustment elements - VEGATOR
- 12 Operating system - VEGATOR
- 12 Adjustment elements - VEGATOR
- 14 Technical data
- 18 Dimensions
- 19 Product code