03 Series
Operator’s Manual
Part Number: 71-0304
Revision: P3
Released: 11/4/14
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating instrument. Improper use of the gas monitor could result in bodily harm or death.
Periodic calibration and maintenance of the gas monitor is essential for proper operation and correct readings. Please calibrate and maintain this instrument regularly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 1 and 3 months, but can be required more often or less often based on your usage.
03 Series Operator’s Manual
Warranty
RKI Instruments, Inc. warrants the 03 Series Single Gas Monitor sold by us to be free from defects in materials, workmanship, and performance for a period of two (2) years from the date of shipment from RKI Instruments,
Inc. This includes the instrument and the original sensor. Replacement parts are warranted for one (1) year from the date of their shipment from
RKI Instruments, Inc. Any parts found defective within their warranty period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items, which by their nature, are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are as follows:
Absorbent cartridges
Filter elements, disks, or sheets
Pump diaphragms and valves
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the instruction manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
T
HIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES
AND REPRESENTATIONS
,
EXPRESSED OR IMPLIED
,
AND ALL OTHER
OBLIGATIONS OR LIABILITIES ON THE PART OF
RKI I
NSTRUMENTS
, I
NC
.
INCLUDING BUT NOT LIMITED TO THE WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE
. I
N NO EVENT SHALL
RKI
I
NSTRUMENTS
, I
NC
.
BE LIABLE FOR INDIRECT
,
INCIDENTAL
,
OR
CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE
OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE
PROPERLY
.
This warranty covers instruments and parts sold to users only by authorized distributors, dealers, and representatives as appointed by RKI
Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor and our warranty is limited to replacement of parts or our complete goods.
03 Series Operator’s Manual Warranty
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Sensor Gasket, Sensor Membrane, and Charcoal Filter . . . . . . . . . . . . . . . . . . . 5
Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Control Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Alarm LED. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Vibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
IrDA Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Printed Circuit Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Protective Rubber Boot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Alligator & Belt Clips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Start-up Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Performing a Fresh Air Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Turning Off the 03 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Measuring Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adjusting the Buzzer Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Display Mode, CO-03 and HS-03 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Display Mode, OX-03 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Data Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
03 Series User Setup Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Calibration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Using Calibration Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Performing a Fresh Air Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Table of Contents 03 Series Operator’s Manual
Performing an Automatic Span Adjustment
(Zero Adjustment for OX-03) in A--CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Performing an Easy Span Adjustment (Zero Adjustment for OX-03) in E--CAL. 39
Performing a Manual Span Adjustment (Zero Adjustment for OX-03) in M--CAL 45
Performing a Bump Test in BUMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Viewing the Instrument’s Firmware Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Setup Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Using Setup Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Setting the Date and Time in DATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Performing a Fresh Air Adjustment in AIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Performing an Automatic Span Adjustment
(Zero Adjustment for OX-03) in A--CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Performing an Easy Span Adjustment (Zero Adjustment for OX-03) in E--CAL. 57
Performing a Manual Span Adjustment (Zero Adjustment for OX-03) in M--CAL 58
Setting the Alarm Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Adjusting the Bump Test Limit Check Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Setting the Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Replacing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Replacing the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Replacing the Sensor Membrane and Charcoal Filter . . . . . . . . . . . . . . . . . . . . 68
Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
WARNING: Understand this manual before operating the 03 Series.
Substitution of components may impair intrinsic safety.
To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be nonhazardous. This unit has not been tested in an oxygen enriched atmosphere (above 21%).
03 Series Operator’s Manual Table of Contents
Introduction
Using an advanced microprocessor controlled detection system, the 03
Series Personal Single Channel Gas Monitor detects the presence of either carbon monoxide (CO), hydrogen sulfide (H
2
S), or oxygen (O
2
). The 03
Series’ compact size and easy-to-use design make it ideally suited for a wide range of applications, including sewage treatment plants, tunnels, hazardous waste sites, petrochemical facilities, oil fields, mines, and chemical plants. The 03 Series is even small enough to be placed conveniently in a pocket. The 03 Series offers the following features:
• Compact design
• Fast, accurate response with backlit digital liquid crystal display (LCD)
• Visual, audible, and vibration alarms
• Microprocessor control for reliability, ease of use, and advanced capabilities
• Resistance to RF (radio frequency) interference
• Datalogging including interval trend data and alarm trend data
• Peak, STEL, and TWA indication for CO-03 & HS-03
• Minimum and maximum indication for OX-03
• Over range alarm
• Gas, battery, sensor failure, and system failure alarms
• Over 3,000 hours operation on one set of alkaline batteries
• Rotatable alligator clip for “hands free” gas monitoring, belt clip optional
• CSA certification for intrinsic safety in Class I, Division I, Groups A, B,
C, and D hazardous atmospheres pending
WARNING: The 03 Series detects oxygen deficiency and elevated levels of oxygen, carbon monoxide, and hydrogen sulfide, all of which can be dangerous or life threatening.
When using the 03 Series, you must follow the instructions and warnings in this manual to assure proper and safe operation of the unit and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 03 Series as described in this manual.
1 • Introduction 03 Series Operator’s Manual
Specifications
Table 1: 03 Series Specifications
CO-03 HS-03 OX-03
Target Gas Carbon Monoxide
(CO)
Detection Range 0 to 500 ppm
Display Increment 1 ppm
Electro Chemical Detection
Principle
Alarm Points Low
High
TWA
STEL
25 ppm
50 ppm
25 ppm
200 ppm
Hydrogen Sulfide
(H
2
S)
0 to 100.0 ppm
0.5 ppm
Electro Chemical
Low
High
TWA
STEL
5.0 ppm
30.0 ppm
1.0 ppm
5.0 ppm
Oxygen
(O
2
)
0 to 40.0% vol.
0.1% vol.
Galvanic Cell
Low 19.5% vol.
(decreasing)
± 5% of reading or
± 5 ppm CO
(whichever is greater)
± 5% of reading or
± 2 ppm H
2
S
(whichever is greater)
High 23.5% vol.
(increasing)
Sampling Method Diffusion
Response Time T90 in 30 seconds
Diffusion
T90 in 30 seconds
Diffusion
T90 in 20 seconds
Accuracy ± 0.5% O
2
Indication
Safety/Regulatory
Digital LCD
Power
C US pending
CSA classified, “C/US”, as Intrinsically Safe. Exia. Class I,
Groups A, B, C, & D. Temperature Code T3.
Two AAA size alkaline batteries standard, Duracell MN2400 or
PC2400, Eveready Energizer E92 or EN92
At 25 °C: Over 3,000 hours, no alarms or backlighting Continuous
Operating Hours
Case
Standard
Accessories
High-impact plastic, dust and weather proof
• Rubber protective boot
• Alligator clip
03 Series Operator’s Manual Specifications • 2
Table 1: 03 Series Specifications
CO-03 HS-03 OX-03
Optional
Accessories
Dimensions and
Weight
Operating Temp. &
Humidity
• Calibration adapter
• Calibration kit
• Belt clip
• Wrist strap
• IrDA/USB cable for downloading data to computer
• Product CD, includes 03 Series Datalogging Program and
03 Series User Setup Program
2.2” (54mm) W x 2.6” (67mm) H x 0.9” (24mm) D;
2.8 oz. (80 g)
-20°C to +50°C, 16 - 95% RH
(non condensing)
-20°C to +50°C,
0 - 95% RH (non condensing)
3 • Specifications 03 Series Operator’s Manual
Description
This section describes the components of the 03 Series.
IrDA Port
Buzzer Opening
Alarm LED
Diffusion Port
Battery Cover Screw
Battery Cover
LCD
CO
AIR
CO-03
POW ER
MODE
Control Buttons
Wrist Strap
Connection
Figure 1: Components of the 03 Series
Case
The 03 Series’ sturdy, high-impact plastic case is blue and consists of a front and rear case. The case is suitable for use in many environmental conditions, indoors and out. The unit is dust proof and weather resistant.
The front case has an LCD (liquid crystal display) that shows various indications. Below the LCD are two black control buttons. The left button is labeled “AIR” and the right button is labeled “POWER MODE”.
The buzzer opening is located in the top left corner of the front case. To the right of the buzzer is the diffusion port for the sensor.
The alarm LED lens is located at the top of the front case.
The battery compartment is located in the rear case. Access to the compartment is accomplished by turning the captive battery cover screw counterclockwise and by removing the battery cover.
There are two spring bars on the rear case. One is on the left and one is on the right. They are the same type of spring bar that is used to retain a watch band and are used to mount the alligator clip or optional belt clip.
A feature in the lower left corner of the rear case is used to install the optional wrist strap.
03 Series Operator’s Manual Description • 4
Sensor Gasket, Sensor Membrane, and Charcoal Filter
The sensor membrane is a paper-like filter disk that is held in place by the sensor gasket and fits into a recessed area on the inside of the front case.
The sensor membrane covers the diffusion port and protects the sensor from dirt and moisture. The sensor membrane should be inspected periodically and replaced if contaminated by dirt or moisture. See
“Replacing the Sensor Membrane and Charcoal Filter” on page 68 for sensor membrane replacement instructions.
The CO-03 includes a charcoal filter disk which is located in the sensor gasket. The charcoal filter disk removes gases from the sampled air that will cause a response on the CO sensor such as H
2
S and certain hydrocarbons. If false or elevated CO readings are noticed, especially in the presence of H
2
S, change the charcoal filter disk. See “Replacing the
Sensor Membrane and Charcoal Filter” on page 68 for replacement instructions.
Sensor
The 03 Series uses either an oxygen, CO, or H
2
S sensor. The sensor is protected by the white sensor membrane which is held in place by the sensor gasket. The sensor membrane allows ambient air to diffuse past it to the sensor. The three sensors used in the three 03 Series models use different detection principles as described below.
Oxygen Sensor
The oxygen sensor is a galvanic type of sensor. A membrane covers the cell and allows gas to diffuse into the cell at a rate proportional to the partial pressure of oxygen. The oxygen reacts in the cell and produces a voltage proportional to the concentration of oxygen. The voltage is measured by the 03 Series’ circuitry, converted to a measurement of gas concentration, and displayed on the LCD.
CO and H
2
S Sensors
The CO and H
2
S sensors are electrochemical sensors that consist of two precious metal electrodes in an acid electrolyte. A gas permeable membrane covers the sensor face and allows gas to diffuse into the electrolyte. The gas reacts in the sensor and produces a current proportional to the concentration of the target gas. The current is amplified by the 03 Series’ circuitry, converted to a measurement of gas concentration, and displayed on the LCD.
LCD
The LCD is visible through the front case. When the 03 Series is in
Measuring Mode, the target gas concentration, battery condition, and alarm indications are displayed on the LCD. Various other items are
5 • Description 03 Series Operator’s Manual
displayed when the LCD is in other modes, such as Calibration Mode.
When either of the two control buttons are pressed, the LCD backlight comes on for 20 seconds.
Control Buttons
Below the LCD are two control buttons: POWER MODE and AIR. The
POWER MODE button turns the 03 Series on and off. The functions performed by the control buttons are summarized in the following table:
Table 2: 03 Series Control Buttons
Button Function
POWER MODE
AIR
• Turns the unit on and off.
• Turns the LCD back light on.
• Resets the alarm circuit (gas alarms).
• Enters Display Mode.
• Enters Calibration Mode with the AIR button.
• Enters Setup Mode with the AIR button.
• Turns the LCD back light on.
• Adjusts LCD reading when the fresh air adjustment is performed.
• Enters Calibration Mode with the POWER MODE button.
• Enters Setup Mode with the POWER MODE button.
• Increases or decreases settings when the unit is in
Calibration Mode or Setup Mode.
Alarm LED
The 03 Series has one red alarm LED. It alerts you to gas, low battery, and sensor failure alarms. The alarm LED is located at the top of the front case beneath a clear plastic lens.
Buzzer
A solid-state electronic buzzer is mounted inside the 03 Series. An opening in the top left corner of the front case allows the buzzer’s sound to emanate from the case. The buzzer sounds for gas alarms, unit malfunctions, and the dead battery alarm. It also serves as an indicator during normal use of the various LCD display options.
Vibrator
A vibrating motor (vibrator) is mounted inside the 03 Series. The vibrator vibrates momentarily during the power-up sequence and for gas alarms.
03 Series Operator’s Manual Description • 6
IrDA Port
An infrared (IR) communications port is located just to the left of the alarm
LED at the top of the front case beneath the clear lens. The data transmitted through the port is in standard IrDA protocol. A computer’s infrared port or an IrDA/USB cable connected to a computer’s USB port can be used to download data saved by the 03 Series to a computer using the 03 Series Data Logger Management Program. See the 03 Series Data
Logger Management Program operator’s manual for data logging and downloading instructions.
Printed Circuit Board
The primary function of the 03 Series’ printed circuit board is to amplify the signal sent to it from the sensor, convert the signal to a meaningful measurement of gas concentration, display the gas concentration on the
LCD, store STEL, TWA, and peak gas readings, and activate the alarm circuit if an alarm point has been reached. It monitors battery level, battery failure, and sensor failure. It also controls various operating modes of the unit.
NOTE: The printed circuit board contains no user serviceable parts.
Batteries
Two AAA-size alkaline batteries run the 03 Series. At 25°C the alkaline batteries last at least 3,000 hours. The battery icon on the LCD shows remaining battery life.
When the 03 Series detects low battery voltage, a low battery warning is activated. When battery voltage is too low for normal operation, the 03
Series sounds a dead battery alarm.
The alkaline batteries can be replaced by removing the battery cover on the rear case. Turn the captive battery cover screw counterclockwise to release the door. See “Replacing the Batteries” on page 64 for detailed battery changing instructions.
WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be nonhazardous.
7 • Description 03 Series Operator’s Manual
Protective Rubber Boot
A protective rubber boot is installed over the 03 Series.
Side Front
Figure 2: Rubber Boot
Alligator and Belt Clips
The 03 Series is available with two types of clips: the alligator clip
(standard) and the belt clip (optional). The alligator clip is shown in Figure 3 below. The alligator clip can be used to attach the 03 Series to clothing or a belt. Teeth in the clip’s jaws prevent the unit from slipping off. The clip can be rotated in 45 degree turns if necessary.
Alligator Clip
Open
Figure 3: Alligator Clip
Closed
03 Series Operator’s Manual Description • 8
The belt clip is shown in Figure 4 below and is used to easily clip the 03
Series on a belt.
Belt Clip
Figure 4: Belt Clip
9 • Description 03 Series Operator’s Manual
Start Up
This section explains how to start up the 03 Series and to get it ready for operation.
NOTE:
The screens shown in this section are for a CO-03. If you are using a different version of the 03 Series, your instrument screen will vary slightly from those shown below.
Start-up Procedure
1. Press and briefly hold down the POWER MODE button. The backlight will turn on and all the display segments will turn on. Release the button when you hear a beep.
2. The vibrator vibrates and the alarm light flashes momentarily.
3. The initial startup screen depends on how Bump Fail Behavior is set.
This parameter along with the Cal. Limit Display and Cal. Limit Check parameters mentioned in Step 5 and the Auto Zero Adjustment parameter mentioned in Step 12 below cannot be set using the 03
Series’ instrument menus, but are set using the 03 Series User Setup
Program. See the 03 Series User Setup Program Operator’s Manual for information regarding changing various instrument parameters that are not available for adjustment in the instrument’s operating modes.
• If Bump Fail Behavior is set to None (factory setting) or if it is set to
Can Not Use
and the most recently performed bump test passed, proceed to Step 5.
• If Bump Fail Behavior is set to Can Not Use and the most recently performed bump test failed, the following screen will appear.
C.AL
FAIL
The instrument cannot be used until a successful bump test or calibration is performed. See “Calibration Mode” on page 31 for instructions.
4. If Cal. Limit Display is set to Off, proceed to Step 6.
03 Series Operator’s Manual Start Up • 10
5. If Cal. Limit Display is set to On (factory setting), the screen that appears next depends on how Cal. Limit Check is set.
• If the unit is due for calibration and Cal. Limit Check is set to
Confirm to use
(factory setting), then the following screen sequence will occur.
C.LIMIT
CAL
YES
C.LIMIT
CAL
MODE
C.LIMIT
CAL
C.LIMIT
CAL
C.LIMIT
CAL
AIR
C.LIMIT
CAL
NO
The alarm LED and buzzer will pulse several times.
To continue the startup sequence without performing a calibration, press and release the AIR button. Continue to Step 6.
To perform a calibration, press and release the POWER MODE button. Depending on the value entered into the One Touch Cal
Time
parameter using the 03 Series User Setup Program, the instrument will display either the A--CAL menu item or the E--CAL menu item in Calibration Mode. See “Calibration Mode” on page 31 for instructions.
WARNING: You must press either the POWER MODE or AIR button to continue. If you do not press a button, the buzzer will continue to beep and the LED will continue to flash for 6 seconds every 5 seconds.
11 • Start Up 03 Series Operator’s Manual
• If the unit is due for calibration and Cal. Limit Check is set to Can’t
use
, then the following screen sequence will occur.
C.LIMIT
FAIL
YES
C.LIMIT
FAIL
MODE
C.LIMIT
FAIL
The alarm LED and buzzer will pulse several times. Press and release the POWER MODE button. Depending on the value entered into the One Touch Cal Time parameter using the 03 Series User
Setup Program, the instrument will display either the A--CAL menu item or the E--CAL menu item in Calibration Mode. The above screen will remain on the display until the unit is either turned off or
Calibration Mode is entered. The 03 Series cannot be used until a complete calibration has been performed either by selecting A--
CAL, E--CAL, or M--CAL in the Calibration Mode menu. See
“Performing an Automatic Span Adjustment (Zero Adjustment for
OX-03) in A--CAL” on page 34, “Performing an Easy Span
Adjustment (Zero Adjustment for OX-03) in E--CAL” on page 39, or
“Performing a Manual Span Adjustment (Zero Adjustment for OX-
03) in M--CAL” on page 45 for calibration instructions.
WARNING: You must perform a successful calibration in order to continue to normal operation. If you do not perform a successful calibration, the screen sequence will continue, the buzzer will continue to beep, and the LED will continue to flash for 6 seconds every 5 seconds and the unit will not enter normal operation.
C.LIMIT
0d
YES
• If a calibration is due and Cal. Limit Check is set to None, the sequence below will occur twice.
C.LIMIT
0d
MODE
C.LIMIT
0d
03 Series Operator’s Manual Start Up • 12
Press and release the POWER MODE button during this time to go to the A--CAL or E--CAL menu item in Calibration Mode depending on the value entered into the One Touch Cal Time parameter of the
03 Series User Setup Program. If no button is pressed, the instrument will continue with the warmup sequence in Step 6 once the sequence shown above has finished.
• If calibration is not due, then the following screen appears for a few seconds indicating when the next calibration is due. “NEXT” and
“CAL” alternate on the bottom of the screen.
14. 2.17
180d
CAL.
14. 2.17
180d
NEXT
NOTE:
The following screens in Step 6 only appear if Bump Test Limit
Display
is set to On using the 03 Series User Setup Program. The standard factory setting for this function is Off.
6. If Bump Test Limit Display is set to Off (factory setting), proceed to
Step 8.
7. If Bump Test Limit Display is set to On using the 03 Series User Setup
Program, the next screen will depend on how Bump Test Limit Check is set in Setup Mode or using the 03 Series User Setup Program.
• If the unit is due for bump testing and Bump Test Limit Check is set to Confirm to use (factory setting), then the following screen sequence will occur.
NOW
teSt
YES
BUMP
teSt
MODE
NOW
teSt
BUMP
teSt
NOW
teSt
AIR
BUMP
teSt
NO
03 Series Operator’s Manual 13 • Start Up
The alarm LED and buzzer will pulse several times.
To continue the startup sequence without performing a bump test, press and release the AIR button. Continue to Step 8.
To perform a bump test, press and release the POWER MODE button. The instrument will display the BUMP menu item in
Calibration Mode. See “Performing a Bump Test in BUMP” on page 48 for instructions.
WARNING:You must press either the POWER MODE or AIR button to continue. If you do not press a button, the screen sequence will continue, the buzzer will continue to beep and the LED will continue to flash for 6 seconds every 5 seconds.
• If the unit is due for bump testing and Bump Test Limit Check is set to Can’t use, then the following screen sequence will occur.
NOW BUMP
FAIL
YES
FAIL
MODE
BUMP
FAIL
NOW
FAIL
The alarm LED and buzzer will pulse several times. Press and release the POWER MODE button to go to the BUMP menu item in
Calibration Mode. The above screens will remain on the display until the unit is either turned off or Calibration Mode is entered. The 03
Series cannot be used until a bump test has been performed. See
“Performing a Bump Test in BUMP” on page 48 for bump test instructions.
WARNING:You must perform a successful bump test in order to continue to normal operation. If you do not perform a successful bump test, the buzzer will continue to beep and the LED will continue to flash for 6 seconds every 5 seconds and the unit will not enter normal operation.
03 Series Operator’s Manual Start Up • 14
BUMP
YES
0d
• If a bump test is due and Bump Test Limit Check is set to None, the sequence below will occur twice.
BUMP
0d
MODE
BUMP
0d
Press and release the POWER MODE button during this time to go to the BUMP menu item in Calibration Mode. If no button is pressed, the instrument will continue with the warmup sequence in Step 8 once the sequence shown above has finished.
NOTE:
If a successful calibration is performed, the next bump test date is reset and starts over even though a bump test was not performed.
• If bump testing is not due, then the following screen appears for a few seconds indicating when the next bump test is due. “NEXT” and
“BUMP” alternate on the bottom of the screen.
14. 2.17
180d
BUMP
14. 2.17
180d
NEXT
8. The Date/Time Screen appears for a few seconds.
2014
1.31
8:20
This screen displays the current date and time.
9. The Battery Voltage Screen appears for a few seconds.
bAtt
3.0
V
15 • Start Up 03 Series Operator’s Manual
The screen displays the current battery voltage.
CAUTION: If the unit gives a low battery warning or dead battery alarm, change the alkaline batteries before using the unit.
10. The display then indicates the following items for about a second each:
• Full scale value
• Warning setpoint (low gas alarm)
• Alarm setpoint (high gas alarm)
• STEL alarm setpoint (CO and H2S only)
• TWA alarm setpoint (CO and H2S only)
11. If the 03 Series experiences a sensor failure during start up, the following screen will appear.
SENSOR
FAIL
The instrument cannot be used if a sensor failure occurs. Replace the failed sensor.
12. If Auto Zero Adjustment is set to On (factory setting is Off), then the
03 Series will perform an automatic fresh air adjustment.
AUTO
- - - -
ZERO
If the fresh air adjustment is successful, the unit will proceed to Normal
Mode. If the sensor fails the air adjustment, the screen will indicate the failure. If a failure occurs, press and release the POWER MODE button to proceed to Normal Mode. Replace the failed sensor as soon as possible.
03 Series Operator’s Manual Start Up • 16
17 • Start Up
WARNING: If the Auto Zero Adjustment feature is turned on, make sure that you start-up the 03 Series in a known fresh air environment, an environment free of combustible or toxic gasses and of normal oxygen content, 20.9%. If you do not start-up the unit in a fresh air environment, the fresh air adjustment will not be accurate.
13. The 03 Series is now operating in Measuring Mode and monitoring for gas. The Normal Operation Screen appears and the instrument beeps once.
CO
0 ppm
The gas concentration of the target gas is displayed along with the battery charge level in the lower left corner. The heart symbol displayed in the upper right corner flashes while the instrument is functioning properly. If it disappears or is steadily on, the unit is experiencing a microprocessor error. The backlight turns off after 20 seconds.
Performing a Fresh Air Adjustment
Before using the 03 Series, set the fresh air reading. Performing this adjustment ensures accurate gas readings in the monitoring environment.
1. Find a fresh air environment of normal oxygen content (20.9%) that is free of toxic or combustible gases.
2. With the unit on and in Measuring Mode, press and hold the AIR button.
The LCD displays “hold” prompting you to hold the AIR button.
AIR
- - - -
HOLD
03 Series Operator’s Manual
3. Release the AIR button when the following screen appears. The unit will set the reading to 0 ppm (20.9% for the OX-03) and return to
Measuring Mode.
ADJ
- - - -
Turning Off the 03 Series
1. Press and hold the POWER MODE button for about five seconds to turn off the unit. The buzzer will pulse while the POWER MODE button is being pressed before the unit turns off.
2. Release the button when the LCD is blank. The unit is off.
NOTE:
If Power Off Password Protection is turned On (factory setting is
Off
) using the 03 Series User Setup Program, a password is required to turn the 03 Series off. When the password screen appears, adjust each digit with the AIR button and press and release the POWER MODE button to move on to the next digit.
Once the password has been entered, the instrument will shut off and the LCD will be blank.
03 Series Operator’s Manual Start Up • 18
Operation
This section describes Measuring Mode, setting the buzzer volume,
Display Mode, and alarm indications.
NOTE:
The screens shown in this section are for a CO-03. If you are using a different version of the 03 Series, your instrument screen will vary slightly from those shown below.
Measuring Mode
In Measuring Mode, the battery level and target gas concentration are displayed on the LCD. The battery icon has three bars visible when the batteries have a full charge. As the battery charge decreases, the bars will gradually disappear, one by one. The heart symbol displayed in the upper right corner flashes while the instrument is functioning properly. If it disappears or is steadily on, the unit is experiencing a microprocessor error.
The target gas concentration is displayed in the middle of the LCD. On the
CO-03, CO is displayed in parts per million (ppm). On the HS-03, H
2
S is displayed in ppm. On the OX-03, oxygen is displayed as volume percent.
CO
0 ppm
Adjusting the Buzzer Volume
The buzzer volume on the 03 Series can be adjusted while in Measuring
Mode if Buzzer Volume Selection is set to On (factory setting is Off). The
Buzzer Volume Selection
setting can be adjusted using the 03 Series
User Setup Program. The default buzzer volume setting is HI. If it is changed to LO, the setting will revert back to HI the next time the instrument is turned on.
1. Make sure the 03 Series is in the Measuring Mode Normal Operation screen.
2. Press and hold the POWER MODE and AIR buttons simultaneously.
NOTE:
To avoid entering Display Mode, press the AIR button slightly sooner than pressing the POWER MODE button.
Release both buttons when the instrument starts simulating an alarm
19 • Operation 03 Series Operator’s Manual
condition after the second beep. The current setting for the buzzer will be displayed and the current volume will be emanating from the buzzer.
b - HI
BEEP
3. Use the AIR button to select LO or HI. The buzzer volume during the simulated alarm condition will change as you select a different volume.
4. Press and release the POWER MODE button to save the setting and return to Normal Operation.
5. If LO was selected, “B--LO“ will appear at the bottom of the Normal
Operation screen.
CO
0
B--LO
ppm
Display Mode, CO-03 and HS-03
This section describes Display Mode for CO-03 and HS-03 instruments only. See “Display Mode, OX-03” on page 22 for a description of the OX-
03’s Display Mode.
You can access Display Mode while in Measuring Mode by using the
POWER MODE button. In Display Mode you can view and reset the peak readings, view the TWA and STEL values, view the detection range full scale, and view the date and time.
STEL is an acronym for short-term exposure limit, and it is the average reading of the target gas during the last 15 minutes. TWA is an acronym for time-weighted average, and it is the average reading for the target gas during the last eight (8) hours. If eight (8) hours has not elapsed since the unit was turned on, the TWA is still calculated over eight hours, with the missing time assigned a zero (0) value for the readings. Similarly, if the unit has not been on for 15 minutes, the missing time is assigned a 0 value and the STEL is calculated over 15 minutes. The peak readings, STEL values, and TWA values are cleared when the unit is turned off.
To enter Display Mode, do the following:
1. Make sure the 03 Series is in the Measuring Mode Normal Operation screen. The 03 Series must be in the Normal Operation Screen for you to access Display Mode.
03 Series Operator’s Manual Operation • 20
2. Press and release the POWER MODE button to enter Display Mode.
The backlight will turn on and the PEAK Screen will appear.
PEAK
15 ppm
The peak reading since the 03 Series was turned on is displayed.
3. If you do not want to clear the peak reading, continue to the next step.
If you want to clear the peak reading, press and hold the AIR button.
After a couple of seconds, the LCD will prompt you to hold the AIR button with the following screen.
CLEAR
- - - -
HOLD
Hold the AIR button until the following screen appears, then release it.
CLEAR
- - - -
The peak reading will be cleared and the unit will return to the PEAK
Screen.
4. Press and release the POWER MODE button again to proceed to the
STEL Screen. The STEL reading will be displayed.
STEL
5 ppm
21 • Operation 03 Series Operator’s Manual
5. Press and release the POWER MODE button again to proceed to the
TWA Screen. The TWA reading will be displayed.
TWA
10 ppm
6. Press and release the POWER MODE button again to proceed to the
Full Scale Screen. The detection range full scale value is displayed. An
“L” or an “A” will appear under the full scale value to indicate whether the alarms are set to latching (L) or auto-resetting (A). This setting can be adjusted using the 03 Series User Setup Program.
F. S.
500 ppm
L
7. Press and release the POWER MODE button again to proceed to the
Date/Time Screen. The date and time of the instrument is displayed.
2014
1.31
8:20
8. Press and release the POWER MODE button once again to return the unit to Measuring Mode.
NOTE:
If you do not press a button for 20 seconds while in Display Mode, the unit will return to Measuring Mode automatically and the backlight will turn off.
Display Mode, OX-03
This section describes Display Mode for the OX-03 only. See “Display
Mode, CO-03 and HS-03” on page 20 for a description of the CO-03 and
HS-03 instruments’ Display Mode.
You can access Display Mode while in Measuring Mode by using the
POWER MODE button. In Display Mode you can view and reset the minimum and maximum readings, view the detection range full scale, and view the date and time.
03 Series Operator’s Manual Operation • 22
To enter Display Mode and view items or reset the peak readings, do the following:
1. Make sure the 03 Series is in the Measuring Mode Normal Operation screen. The 03 Series must be in the Normal Operation Screen for you to access Display Mode.
2. Press and release the POWER MODE button to enter Display Mode.
The backlight will turn on.
3. The Min. Screen will be displayed. This is the minimum oxygen reading taken since the instrument was turned on.
MIN.
20.1
%
4. If you do not want to clear the minimum reading, continue to the next step.
If you want to clear the minimum reading, press and hold the AIR button. After a couple of seconds, the LCD will prompt you to hold the
AIR button with the following screen.
CLEAR
- - - -
HOLD
Hold the AIR button until the following screen appears, then release it.
CLEAR
- - - -
The minimum reading will be cleared and the unit will return to the Min.
Screen.
23 • Operation 03 Series Operator’s Manual
5. Press and release the POWER MODE button to display the Max.
Screen. This is the maximum oxygen reading taken since the instrument was turned on.
MAX.
21.2
%
6. If you do not want to clear the maximum reading, continue to the next step.
If you want to clear the maximum reading, press and hold the AIR button. After a couple of seconds, the LCD will prompt you to hold the
AIR button with the following screen.
CLEAR
- - - -
HOLD
Hold the AIR button until the following screen appears, then release it.
CLEAR
- - - -
The maximum reading will be cleared and the unit will return to the Max.
Screen.
7. Press and release the POWER MODE button again to proceed to the
Full Scale Screen. The detection range full scale value is displayed. An
“L” or an “A” will appear under the full scale value to indicate whether the alarms are set to latching (L) or auto-resetting (A). This setting can be adjusted using the 03 Series User Setup Program.
F. S.
40.0
L
%
03 Series Operator’s Manual Operation • 24
8. Press and release the POWER MODE button again to proceed to the
Date/Time Screen. The date and time of the instrument is displayed.
2014
1.31
8:20
9. Press and release the POWER MODE button once again to return the unit to Measuring Mode.
NOTE:
If you do not press a button for 20 seconds while in Display Mode, the unit will return to Measuring Mode automatically and the backlight will turn off.
Alarms
This section covers alarm indications. It also tells you how to reset the 03
Series after an alarm has occurred and how to respond to an alarm condition.
Alarm Indications
The 03 Series will sound the buzzer, flash its alarm LED, and vibrate when the target gas concentration rises above (falls below for oxygen) the warning level. The 03 Series also sounds the buzzer, flashes its alarm
LED, and vibrates when the alarm level is reached. In addition, the 03
Series has a low battery warning, a dead battery alarm, an over range alarm, a sensor failure alarm, a system failure alarm, and a clock failure alarm. See Table 3 below for a description of each alarm indication.
Table 3: Alarm Types and Indications
Alarm Type
Warning
Concentration of gas rises above (falls below for OX-03) the warning level.
Visual Indications
• Gas reading flashes.
• WARN appears below the gas reading.
• Back light turns on.
• Alarm LED flashes.
Other Indications
• Buzzer sounds alternating between a low and high pitch.
• Vibrator pulses.
25 • Operation 03 Series Operator’s Manual
Table 3: Alarm Types and Indications
Alarm Type Visual Indications Other Indications
Alarm
Concentration of gas rises above the alarm level.
TWA or STEL
(CO-03 & HS-03 Only)
Concentration of CO or H
2
S rises above the
TWA or STEL alarm point.
Over Range
Concentration of gas rises above the measuring limit of the
03 Series. (Or there could be a problem with the unit.)
• Gas reading flashes.
• ALRM appears below the gas reading.
• Back light turns on.
• Alarm LED flashes faster than warning indication.
• Back light turns on.
• Gas reading flashes.
• TWA or STEL appears below the gas reading.
• If the unit is in both
TWA alarm and
STEL alarm, only
STEL
will be displayed.
• Alarm LED flashes.
• Gas reading replaced by blinking brackets.
• Back light turns on.
• ALRM appears below gas reading.
• Buzzer sounds alternating between low and high pitch faster than warning indication.
• Vibrator pulses faster than warning indication.
• Buzzer sounds alternating between a low and high pitch at same rate as warning indication.
• Vibrator pulses at same rate as warning indication.
• Buzzer sounds alternating between low and high pitch at the same rate as alarm indication.
• Vibrator pulses at same rate as alarm indication.
Low Battery Warning
• Last remaining bar on the right in battery icon flashes.
Dead Battery Alarm
• Gas reading replaced by FAIL.
• Battery icon flashes.
• None
• Double pulsing tone
(two pulses in quick succession) occurs once a second.
03 Series Operator’s Manual Operation • 26
System Failure
Clock Failure
Table 3: Alarm Types and Indications
Alarm Type
Sensor Failure
Visual Indications Other Indications
• Gas name replaced by SENSOR.
• Gas reading replaced by FAIL.
• Alarm LED flashes.
• Gas name replaced by SYSTEM.
• Gas reading replaced by FAIL.
• Failure code displays below
FAIL
.
• Gas name replaced by FAIL.
• Gas reading replaced by FAIL.
• Double pulsing tone
(two pulses in quick succession) once a second.
• Double pulsing tone
(two pulses in quick succession) once a second.
• Double pulsing tone
(two pulses in quick succession) once a second.
Resetting Gas Alarms
If the alarms are latching (factory setting), then an alarm indication will continue even if the gas reading causing the alarm decreases below the alarm setpoint (increases above for oxygen Warning) and will continue until the alarm is reset. The gas reading that caused the alarm must decrease below the alarm setpoint (increase above for oxygen Warning) before that alarm can be reset. To reset latching alarms, press and release the
POWER MODE button after the gas reading falls below the alarm setpoint
(rises above for oxygen Warning).
If the alarms are self-resetting, then an alarm condition will automatically reset when the gas reading that caused the alarm decreases below the alarm setpoint (increases above for oxygen Warning).
Responding to Alarms
This section describes response to gas, over range, battery, sensor failure, and system failure alarms.
Responding to Gas Alarms
1. Follow your established procedure for an increasing gas condition or a decreasing oxygen condition.
2. If your unit is set for latching alarms, reset the alarm using the POWER
MODE button once the alarm condition has been cleared.
27 • Operation 03 Series Operator’s Manual
Responding to an Over Range Alarm
WARNING: An over range condition may indicate an extreme toxic gas or oxygen concentration. Confirm the gas concentration with a different 03 Series or with another gas detecting device.
1. Follow your established procedure for an increasing gas condition.
2. Reset the alarm by pressing and releasing the POWER MODE button after the alarm condition has cleared.
3. Calibrate the 03 Series as described in “Calibration Mode” on page 31.
4. If the over range condition continues, replace the sensor as described in “Replacing the Sensor” on page 66.
5. If the over range condition continues after you have replaced the sensor, contact RKI Instruments, Inc. for further instructions.
Responding to Battery Alarms
WARNING: The 03 Series is not operational as a gas monitoring device during a dead battery alarm. Take the 03 Series to a non-hazardous area and change the alkaline batteries as described in “Replacing the Batteries” on page 64.
The 03 Series is fully functional in a low battery warning condition.
However, only a couple of days of operation may remain depending on certain conditions such as alarm occurrences and how often the backlight comes on.
NOTE:
Alarms and the back light feature consume battery power and reduce the amount of operating time remaining.
When a low battery warning occurs, change the batteries as soon as possible. Refer to the instructions in “Replacing the Batteries” on page 64 for more information.
Responding to a Sensor Failure Alarm
1. Try calibrating the 03 Series as described in “Calibration Mode” on page 31.
2. If the sensor failure continues, replace the sensor as described in
“Replacing the Sensor” on page 66.
3. If the gas sensor failure condition continues after you have replaced the gas sensor, contact RKI Instruments, Inc. for further instructions.
03 Series Operator’s Manual Operation • 28
Responding to a System Failure Alarm
1. If a system failure occurs, the system failure screen will display an error code as shown below:
SYSTEM
FAIL
021
2. The error code meanings are shown below:
Table 4: Error Code Explanation
Error Code
000
010
021
Explanation
MPU failure
RAM failure
EEPROM failure
3. The instrument cannot be used if a system failure occurs. Contact RKI
Instruments, Inc. as soon as possible.
Responding to a Clock Failure Alarm
This alarm occurs if the internal unit date has been changed to something unreasonable like 15/34 (month/day).
1. Press and release the POWER MODE button to continue into normal operation.
CAUTION: There will be no datalogging function if you operate the instrument after a clock failure.
2. Attempt to change the date using the DATE menu item in Calibration
Mode. See “Setting the Date and Time” on page 33.
3. If the date cannot be set correctly, contact RKI Instruments, Inc. as soon as possible.
29 • Operation 03 Series Operator’s Manual
Data Logging
The 03 Series features the ability to log data to its internal memory and download it to a computer via the IrDA port on the top of the case. It logs normal operation gas readings, alarm data, calibration data, and bump test data.
To utilize the 03 Series’ downloading capability, you will need the 03 Series
Data Logger Management Program and a computer with an infrared port that conforms to IrDA 1.1 protocol or a USB port that runs one of the following operating systems: Windows XP, Windows Vista, Windows 7 or
Windows 8. If your computer has an infrared port that conforms to IrDA 1.1 protocol, then no additional accessories are needed to download data from the 03 Series. If your computer does not have an infrared port but does have a USB port, a USB/IrDA adapter cable can be used to download data from the 03 Series using the USB port. The 03 Series Data Logger
Management Program is included in the product CD and is also available from RKI Instruments, Inc. The adapter cable is also available from RKI.
See the 03 Series Data Logger Management Program Operator’s Manual for a complete description of the 03 Series Data Logger Management
Program and procedures for downloading data to a computer.
03 Series User Setup Program
There are some instrument operating parameters that are not accessible in either Calibration Mode or Setup Mode such as the calibration frequency, auto zero function, and the alarm logic (latching or self resetting). Many of these parameters either do not typically need to be accessed once the 03
Series is shipped from the factory or may only need to be accessed once because of operator preferences. These parameters can be accessed and updated if necessary using the 03 Series User Setup Program, a computer with the same requirements described above in “Data Logging”, and a
USB/IrDA cable if necessary. The 03 Series User Setup Program is included in the product CD and is also available from RKI Instruments, Inc.
The adapter cable is also available from RKI. See the 03 Series User
Setup Program Operator’s Manual for a complete description of the 03
Series User Setup Program and procedures for accessing and updating instrument operating parameters.
03 Series Operator’s Manual Data Logging • 30
Calibration Mode
This section describes the 03 Series in Calibration Mode. In Calibration
Mode, you can move through a menu of screens to do the following:
• Set the date and time
• Perform a fresh air adjustment (part of a calibration)
• Perform an automatic span adjustment (part of a calibration)
• Perform an easy span adjustment (part of a calibration)
• Perform a manual span adjustment (part of a calibration)
• Perform a bump test (if Bump Test Function is set to On using the
03 Series User Setup Program)
• View the instrument’s firmware version and firmware checksum
NOTE:
You can set the 03 Series to alert you during the startup sequence when calibration or bump testing is due with the 03 Series User
Setup Program. See the 03 Series User Setup Program
Operator’s Manual for information on setting the Cal. Limit
Display
and Bump Test Limit Display parameters.
Calibration Frequency
The optimum frequency of calibration depends heavily on how the 03
Series is used. For example, instruments used daily may need to be calibrated weekly or monthly, while instruments that are used only a few times a year may need to be calibrated before each use. Typical calibration frequencies range from monthly to quarterly. Make sure to develop a calibration schedule tailored to your application.
Using Calibration Mode
WARNING: The 03 Series is not in operation as a gas detector while in Calibration Mode. Although it will respond to gas in parts of AIR CAL, A--CAL., M--CAL., and BUMP, there are no gas alarm indications.
1. Take the 03 Series to a non-hazardous area and turn it off if it is on.
31 • Calibration Mode 03 Series Operator’s Manual
2. Press and hold the AIR button, then press and hold the POWER MODE button. When you hear a beep release the buttons. The DATE Screen is displayed.
DATE
3. Use the AIR button to move forward through Calibration Mode. When you get to the last menu item, the START menu item, continuing will take you back to the beginning of the menu.
4. When you arrive at the item you wish to enter, press and release the
POWER MODE button to enter that item.
5. When you need to adjust the numerical value of a parameter, increase it or decrease it, use the AIR button to change the value. When adjusting a numerical parameter value, it is possible to reverse the direction of adjustment. To switch from increasing to decreasing a value or decreasing to increasing a value, do the following:
• with the parameter flashing on the screen, press and hold the AIR button
• immediately press the POWER MODE button and then release both buttons
• the direction of adjustment when you press the AIR button is now reversed
6. When you are done using the menu items in Calibration Mode, use the
AIR button to scroll through the menu items to the START item.
START
7. At the START screen, press and release the POWER MODE button.
The 03 Series will begin its start-up sequence.
The Calibration Mode menu items are described below in the order in which they appear while moving through Calibration Mode.
03 Series Operator’s Manual Calibration Mode • 32
Setting the Date and Time
Entering the DATE menu item allows you to set the date and time.
DATE
1. When the DATE Screen is displayed, press and release the POWER
MODE button. A screen appears with the year flashing in the upper left, the month and day in the middle, and the time in the lower right.
2014
2.11
3:05
2. Use the AIR button to display the desired year.
3. Press and release the POWER MODE button to save the setting. The month setting flashes.
4. Repeat steps 2 and 3 to enter the month, day, hour and minute setting.
When you save the minute setting, END is displayed before the instrument returns to the DATE Screen.
Performing a Fresh Air Adjustment
Entering the AIR menu item allows you to perform a fresh air adjustment.
AIR
Perform a fresh air adjustment in Calibration Mode when you are performing a calibration before proceeding either to the A--CAL or M--CAL menu item to perform a span adjustment (zero adjustment for OX-03). A fresh air adjustment performed in Calibration Mode is the same as a fresh air adjustment in Normal Mode. The AIR menu item is available in
Calibration Mode for convenience when performing a complete calibration.
WARNING: Calibrate the 03 Series in a non-hazardous environment.
33 • Calibration Mode 03 Series Operator’s Manual
1. Find a fresh air environment, an environment of normal oxygen content
(20.9%) that is free of toxic and combustible gasses.
2. When the AIR screen is displayed, press and release the POWER
MODE button. A screen appears that displays the current gas reading.
AIR
0 ppm
CO
If you want to cancel the fresh air adjustment, press and release the
AIR button to return to the AIR screen. To continue with the fresh air adjustment, proceed to Step 3.
3. Press and hold the AIR button. The LCD prompts you to continue to hold the AIR button.
AIR
- - - -
HOLD
4. Release the AIR button when the following screen appears.
ADJ
- - - -
5. The unit will take a few seconds to adjust the fresh air reading, then indicate “END” in the upper left corner and return to the AIR Screen.
Performing an Automatic Span Adjustment
(Zero Adjustment for OX-03) in A--CAL
The A--CAL menu item only appears in Calibration Mode if the One Touch
Cal Time
parameter in the 03 Series User Setup Program is set to 0
(factory setting). If One Touch Cal Time is set to anything other than 0, E--
CAL will replace A--CAL.
03 Series Operator’s Manual Calibration Mode • 34
Entering the A--CAL menu item allows you to perform an automatic span adjustment (zero adjustment for OX-03).
A--CAL.
Perform a span adjustment (zero adjustment for OX-03) as part of a calibration after performing a fresh air adjustment. Performing a span adjustment (zero adjustment for OX-03) requires the use of a calibration kit. A calibration kit is available from RKI Instruments, Inc. for each 03
Series model (see “Parts List” on page 70). You will need:
• A gas cylinder with an appropriate concentration of the target gas for the CO-03 or HS-03, or a cylinder of 100% nitrogen for the OX-03.
NOTE:
On the OX-03, instead of 100% nitrogen (0% oxygen), it is allowable to use higher than 0% oxygen to set the zero level. RKI
Instruments, Inc. recommends 18% oxygen or lower.
• a fixed-flow regulator with a flow rate of 0.5 LPM (liters per minute)
WARNING: Use a 0.5 LPM (liters per minute) fixed flow regulator when calibrating. Using a different flow rate may adversely affect the accuracy of the calibration.
• non-absorbent tubing
• calibration adapter that will fit over the 03 Series’ sensor
WARNING: Calibrate the 03 Series in a non-hazardous environment.
1. Before performing a span adjustment (zero adjustment for OX-03), perform a fresh air adjustment as described in “Performing a Fresh Air
Adjustment” on page 33.
35 • Calibration Mode 03 Series Operator’s Manual
2. At the A--CAL screen, press and release the POWER MODE button. A screen appears that displays the calibration gas concentration that the
03 Series expects you to use.
A--CAL.
50
CO
ppm
If the displayed concentration matches the calibration cylinder concentration, continue with Step 4.
If the displayed concentration does not match the calibration cylinder concentration, do the following:
• Press and hold the AIR button, then press and momentarily hold the
POWER MODE button.
• Release both buttons as soon as you hear the buzzer sound a beep.
The following screen will display and the gas concentration will be flashing.
A--CAL.
50 ppm
• Use the AIR button to adjust the calibration gas value to the desired value. See Step 5 on page 32 for instructions to increase or decrease the value.
• Press and release the POWER MODE button to accept the displayed value. The previous screen will return.
A--CAL.
50
CO
ppm
03 Series Operator’s Manual Calibration Mode • 36
3. Press and release the POWER MODE button. The LCD will display the current gas readings and “A--CAL.” will flash.
A--CAL.
0 ppm
CO
If you want to cancel the calibration, press and hold the AIR key for 2 seconds. The instrument will return to the A--CAL. Screen. If you hold the AIR key down longer than 2 seconds, the instrument will start scrolling through the Calibration Mode items.
4. Use the sample tubing to connect the calibration adapter to the regulator. Attach the tubing to the calibration adapter on the inlet side as shown below in Figure 5.
To Fixed Flow
Regulator
Calibration Tubing
Calibration
Adapter
Figure 5: Calibration Kit Assembly
5. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
37 • Calibration Mode 03 Series Operator’s Manual
6. Push the calibration adapter onto the 03 Series’ sensor face as shown in Figure 6. The calibration adapter secures itself to the instrument by latching on to two recesses in the instrument’s rear case. The rubber boot (if installed) does not need to be removed to install the adapter.
Calibration
Adapter
AIR
CO-03
POW ER
MO DE
Figure 6: Installing the Calibration Adapter
7. Turn the regulator on/off knob counterclockwise to open it. Calibration gas will begin to flow.
8. Allow the gas to flow for two minutes.
9. Press and release the POWER MODE button.
10. The 03 Series will attempt to make a span adjustment (zero adjustment for OX-03).
11. If the span adjustment (zero adjustment for OX-03) is successful, the
LCD will show the following screen before returning to the A--CAL screen.
PASS
If the Maximum Span setting is turned On (factory setting is Off) using the 03 Series User Setup Program, the LCD will show the maximum reading the sensor could have been calibrated to before returning to the
03 Series Operator’s Manual Calibration Mode • 38
A--CAL screen.
C--MAX
75 ppm
12. If the span adjustment (zero adjustment for OX-03) fails, the LCD will show the following screen, the alarm LED will flash, and the buzzer will sound a double pulsing tone.
CO
ppm
FAIL
Press and release the POWER MODE button to clear the failure indication and return to the A--CAL. screen.
A--CAL.
See “Troubleshooting” on page 63 to investigate the cause of the failure and replace the failed sensor if necessary.
13. Turn the regulator on/off knob clockwise to close it.
14. Remove the calibration adapter from the instrument.
15. Remove the regulator from the calibration gas cylinder.
16. Leave the regulator connected to the calibration adapter for convenience.
17. Store the components of the calibration kit in a safe and convenient place.
Performing an Easy Span Adjustment (Zero Adjustment for
OX-03) in E--CAL
The E--CAL menu item only appears in Calibration Mode if the One Touch
Cal Time
parameter in the 03 Series User Setup Program is set to anything other than 0 (factory setting). When One Touch Cal Time is set to 0, A--
CAL will replace E--CAL. The value entered into the One Touch Cal Time field is the amount of time that the instrument will allow gas to be applied
39 • Calibration Mode 03 Series Operator’s Manual
before attempting to perform a span adjustment.
Entering the E--CAL menu item allows you to perform an easy span adjustment (zero adjustment for OX-03). In an easy span adjustment, the instrument will automatically count down in the E--CAL gas exposure screen when gas is applied and perform a span adjustment when the countdown is over.
E--CAL.
Perform a span adjustment (zero adjustment for OX-03) as part of a calibration after performing a fresh air adjustment. Performing a span adjustment (zero adjustment for OX-03) requires the use of a calibration kit. A calibration kit is available from RKI Instruments, Inc. for each 03
Series model (see “Parts List” on page 70). You will need:
• A gas cylinder with an appropriate concentration of the target gas for the CO-03 or HS-03, or a cylinder of 100% nitrogen for the OX-03.
NOTE:
On the OX-03, instead of 100% nitrogen (0% oxygen), it is allowable to use higher than 0% oxygen to set the zero level. RKI
Instruments, Inc. recommends 18% oxygen or lower.
• a fixed-flow regulator with a flow rate of 0.5 LPM (liters per minute)
WARNING: Use a 0.5 LPM (liters per minute) fixed flow regulator when calibrating. Using a different flow rate may adversely affect the accuracy of the calibration.
• non-absorbent tubing
• calibration adapter that will fit over the 03 Series’ sensor
WARNING: Calibrate the 03 Series in a non-hazardous environment.
1. Before performing a span adjustment (zero adjustment for OX-03), perform a fresh air adjustment as described in “Performing a Fresh Air
Adjustment” on page 33.
03 Series Operator’s Manual Calibration Mode • 40
2. At the E--CAL screen, press and release the POWER MODE button. A screen appears that displays the calibration gas concentration that the
03 Series expects you to use.
E--CAL.
50
CO
ppm
If the displayed concentration matches the calibration cylinder concentration, continue with Step 3.
If the displayed concentration does not match the calibration cylinder concentration, do the following:
• Press and hold the AIR button, then press and momentarily hold the
POWER MODE button.
• Release both buttons as soon as you hear the buzzer sound a beep.
The following screen will display and the gas concentration will be flashing.
E--CAL.
50 ppm
• Use the AIR button to adjust the calibration gas value to the desired value. See Step 5 on page 32 for instructions to increase or decrease the value.
• Press and release the POWER MODE button to accept the displayed value. The previous screen will return.
E--CAL.
50
CO
ppm
41 • Calibration Mode 03 Series Operator’s Manual
3. Press and release the POWER MODE button. The LCD will display the current gas readings and “E--CAL.” will flash.
E--CAL.
0 ppm
G--IN
If you want to cancel the calibration, press and hold the AIR key for 2 seconds. The instrument will return to the E--CAL. Screen. If you hold the AIR key down longer than 2 seconds, the instrument will start scrolling through the Calibration Mode items.
4. Use the sample tubing to connect the calibration adapter to the regulator. Attach the tubing to the calibration adapter on the inlet side as shown below in Figure 7.
To Fixed Flow
Regulator
Calibration Tubing
Calibration
Adapter
Figure 7: Calibration Kit Assembly
5. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
03 Series Operator’s Manual Calibration Mode • 42
6. Push the calibration adapter onto the 03 Series’ sensor face as shown in Figure 8. The calibration adapter secures itself to the instrument by latching on to two recesses in the instrument’s rear case. The rubber boot (if installed) does not need to be removed to install the adapter.
Calibration
Adapter
AIR
CO-03
POW ER
MO DE
Figure 8: Installing the Calibration Adapter
7. Turn the regulator on/off knob counterclockwise to open it. Calibration gas will begin to flow.
8. Once the gas reading reaches 10% of the calibration gas value (HS-03 and CO-03 units) or 18.9% (OX-03 units), “E--CAL” will begin to flash and a countdown will begin. The countdown’s duration is defined by the value of the One Touch Cal Time parameter accessed in the 03 Series
User Setup Program.
E--CAL.
10
60
ppm
If the gas reading never reaches 10% of the calibration gas value (HS-
03 and CO-03 units) or 18.9% (OX-03 units), the countdown will never start and the instrument will remain in the G--IN screen shown in Step 3.
Press and hold the AIR button for 2 seconds to cancel the calibration then continue with Step 12. See “Troubleshooting” on page 63 to investigate the cause of the issue.
9. At the end of the countdown, the 03 Series will attempt to make a span
43 • Calibration Mode 03 Series Operator’s Manual
adjustment (zero adjustment for OX-03).
10. If the span adjustment (zero adjustment for OX-03) is successful, the
LCD will show the following screen before returning to the E--CAL screen.
PASS
If the Maximum Span setting is turned On (factory setting is Off) using the 03 Series User Setup Program, the LCD will show the maximum reading the sensor could have been calibrated to before returning to the
E--CAL screen.
C--MAX
75 ppm
11. If the span adjustment (zero adjustment for OX-03) fails, the LCD will show the following screen, the alarm LED will flash, and the buzzer will sound a double pulsing tone.
CO
FAIL
ppm
Press and release the POWER MODE button to clear the failure indication and return to the E--CAL. screen.
E--CAL.
See “Troubleshooting” on page 63 to investigate the cause of the failure and replace the failed sensor if necessary.
12. Turn the regulator on/off knob clockwise to close it.
13. Remove the calibration adapter from the instrument.
03 Series Operator’s Manual Calibration Mode • 44
14. Remove the regulator from the calibration gas cylinder.
15. Leave the regulator connected to the calibration adapter for convenience.
16. Store the components of the calibration kit in a safe and convenient place.
Performing a Manual Span Adjustment (Zero Adjustment for OX-03) in M--CAL
Entering the M--CAL menu item allows you to perform a manual span adjustment (zero adjustment for OX-03).
M--CAL.
Perform a span adjustment (zero adjustment for OX-03) as part of a calibration after performing a fresh air adjustment. Performing a span adjustment (zero adjustment for OX-03) requires the use of a calibration kit. A calibration kit is available from RKI Instruments, Inc. for each 03
Series model (see “Parts List” on page 70). The procedure below describes a span adjustment (zero adjustment for OX-03). You will need:
• A gas cylinder with an appropriate concentration of the target gas for the CO-03 or HS-03, or a cylinder of 100% nitrogen for the OX-03.
NOTE:
On the OX-03, instead of 100% nitrogen (0% oxygen), it is allowable to use higher than 0% oxygen to set the zero level. RKI
Instruments, Inc. recommends 18% oxygen or lower.
• a fixed-flow regulator with a flow rate of 0.5 LPM (liters per minute)
WARNING: Use a 0.5 LPM (liters per minute) fixed flow regulator when calibrating. Using a different flow rate may adversely affect the accuracy of the calibration.
• non-absorbent tubing
• calibration adapter that will fit over the 03 Series’ sensor
WARNING: Calibrate the 03 Series in a non-hazardous environment.
1. Before performing a span adjustment (zero adjustment for OX-03), perform a fresh air adjustment as described in “Performing a Fresh Air
45 • Calibration Mode 03 Series Operator’s Manual
Adjustment” on page 33.
2. At the M--CAL screen, press and release the POWER MODE button.
The current gas reading will be shown and it will be flashing. The arrows shown at the bottom of the screen indicate the direction of adjustment for the AIR button.
M--CAL.
0 ppm
3. Use the sample tubing to connect the calibration adapter to the regulator. Attach the tubing to the calibration adapter on the inlet side as shown below in Figure 9.
To Fixed Flow
Regulator
Calibration Tubing
Calibration
Adapter
Figure 9: Calibration Kit Assembly
03 Series Operator’s Manual Calibration Mode • 46
4. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
5. Push the calibration adapter onto the 03 Series’ sensor face as shown in Figure 10. The calibration adapter secures itself to the instrument by latching on to two recesses in the instrument’s rear case. The rubber boot (if installed) does not need to be removed to install the adapter.
Calibration
Adapter
AIR
CO-03
POWER
MODE
Figure 10: Installing the Calibration Adapter
6. Turn the regulator on/off knob counterclockwise to open it. Calibration gas will begin to flow.
7. Allow the gas to flow for two minutes.
8. Adjust the display gas reading using the AIR button to match the calibration cylinder’s target gas concentration. See Step 5 on page 32 for instructions to adjust a parameter using the AIR button. If the direction of adjustment is changed, the arrows at the bottom of the screen will change direction.
9. Press and release the POWER MODE button.
10. The 03 Series will make the span adjustment (zero adjustment for OX-
03) and will return to the M--CAL. Screen.
M--CAL.
47 • Calibration Mode 03 Series Operator’s Manual
11. Turn the regulator on/off knob clockwise to close it.
12. Remove the calibration adapter from the instrument.
13. Remove the regulator from the calibration gas cylinder.
14. Leave the regulator connected to the calibration adapter for convenience.
15. Store the components of the calibration kit in a safe and convenient place.
Performing a Bump Test in BUMP
NOTE: Bump Test Function
must be set to On using the 03 Series User
Setup Program in order for BUMP to appear in Calibration Mode.
If Bump Test Function is set to Off, BUMP will not appear. See the 03 Series User Setup Program Operator’s Manual for instructions. The factory setting for Bump Test Function is Off.
Entering the BUMP menu item allows you to perform a bump test to determine if the instrument is responding properly to gas.
BUMP
Performing a bump test requires the use of a calibration kit. A calibration kit is available from RKI Instruments, Inc. for each 03 Series model (see
“Parts List”). You will need:
• A gas cylinder with an appropriate concentration of the target gas for the CO-03 or HS-03, or a cylinder of 100% nitrogen for the OX-03.
NOTE:
On the OX-03, instead of 100% nitrogen (0% oxygen), it is allowable to use higher than 0% oxygen to set the zero level. RKI
Instruments, Inc. recommends 18% oxygen or lower.
• a fixed-flow regulator with a flow rate of 0.5 LPM (liters per minute)
WARNING: Use a 0.5 LPM (liters per minute) fixed flow regulator when bump testing. Using a different flow rate may adversely affect the accuracy of the bump test.
• non-absorbent tubing
03 Series Operator’s Manual Calibration Mode • 48
• calibration adapter that will fit over the 03 Series’ sensor
WARNING: Bump test the 03 Series in a non-hazardous environment.
1. Before performing a bump test, perform a fresh air adjustment as described in “Performing a Fresh Air Adjustment” on page 33.
2. At the BUMP screen, press and release the POWER MODE button. A screen appears that displays the gas concentration that the 03 Series expects you to use.
NOTE:
The bump test gas concentration is the same as the A--CAL gas concentration.
BUMP
50
30
ppm
If the value is not correct, you can change it by changing the A--CAL gas value in Calibration Mode or Setup Mode or by using the 03 Series
Data Logger Management Program or the 03 Series User Setup
Program.
3. Use the sample tubing to connect the calibration adapter to the regulator. Attach the tubing to the calibration adapter on the inlet side as shown below in Figure 11.
To Fixed Flow
Regulator
Calibration Tubing
Calibration
Adapter
49 • Calibration Mode
Figure 11: Calibration Kit Assembly
03 Series Operator’s Manual
4. Confirm that the regulator on/off knob is turned all the way clockwise
(closed) and screw the calibration gas cylinder onto the regulator.
5. Push the calibration adapter onto the 03 Series’ sensor face as shown in Figure 12. The calibration adapter secures itself to the instrument by latching on to two recesses in the instrument’s rear case. The rubber boot (if installed) does not need to be removed to install the adapter.
Calibration
Adapter
AIR
CO-03
POW ER
MO DE
Figure 12: Installing the Calibration Adapter
6. Press and release the POWER MODE button. The LCD will display the current gas reading, “BUMP” and “APPLY” will alternate at the top of the screen, and the instrument will begin a count down whose length depends on how the Bump Test Time(sec) parameter in the 03 Series
User Setup Program is set. The factory setting is 30 seconds.
BUMP
15
30
ppm
APPLY
15
30
ppm
7. Turn the regulator on/off knob counterclockwise to open it. Calibration gas will begin to flow. Do not delay opening the regulator.
8. When the bump test countdown reaches 0, the unit will determine the bump test results and proceed as described in either Step 9 or Step 10 depending on the setting of Calibration After Bump Test Failed. See
03 Series Operator’s Manual Calibration Mode • 50
the 03 Series User Setup Program Operator’s Manual for a complete description of this parameter.
9. When Calibration After Bump Test Failed is set to Off:
• The instrument will display a “P” for pass or an “F” for fail.
BUMP
P
OR
BUMP
F
• To view the bump test gas reading, press the AIR button.
BUMP
45 ppm
• To return to the BUMP screen in Calibration Mode, press POWER
MODE at any time.
BUMP
See “Troubleshooting” on page 63 to investigate the cause of the failure and replace the failed sensor if necessary.
• Turn the regulator on/off knob clockwise to close it.
10. When Calibration After Bump Test Failed is set to On (factory setting):
If the instrument passes the bump test,
• The following screen appears:
BUMP
P
51 • Calibration Mode 03 Series Operator’s Manual
• To view the bump test gas reading, press the AIR button.
BUMP
45 ppm
• To return to the BUMP screen in Calibration Mode, press POWER
MODE at any time.
BUMP
• Turn the regulator on/off knob clockwise to close it.
If the instrument fails the bump test,
• A calibration is immediately and automatically started. The instrument will begin a countdown whose length depends on the
Bump Test Time(sec)
and Calibration Time(sec) After Bump
Test Failed
parameters’ values. These parameters can be changed using the 03 Series User Setup Program. The countdown time for the calibration is the Calibration Time(sec) After Bump Test
Failed
value (factory setting is 90 seconds) minus the Bump Test
Time(sec)
value (factory setting is 30 seconds). Continue to apply the calibration gas.
CAL.
45
60
ppm
APPLY
45
60
ppm
03 Series Operator’s Manual Calibration Mode • 52
• At the end of the calibration, the instrument displays the results from both the bump test and the calibration. The bump test’s result is the first letter and the calibration’s result is the second letter. An “F” represents a failed test and a “P” represents a passed test.
BUMP
FP
OR
BUMP
FF
ppm
• To view the bump test gas readings, press the AIR button.
BUMP
10 ppm
• To view the calibration gas reading, press the AIR button again.
CAL.
45 ppm
• To return to the BUMP screen in Calibration Mode, press POWER
MODE at any time.
BUMP
• Turn the regulator on/off knob clockwise to close it.
11. Remove the calibration adapter from the instrument.
12. Remove the regulator from the calibration gas cylinder.
13. Leave the regulator connected to the calibration adapter for convenience.
14. Store the components of the calibration kit in a safe and convenient place.
53 • Calibration Mode 03 Series Operator’s Manual
Viewing the Instrument’s Firmware Version
The ROM screen shows the firmware version that is loaded in the instrument and the firmware checksum.
1. At the ROM screen, press and release the POWER MODE button.
30010
- - - -
2. After a few seconds, the following screen will be displayed. The top line is the firmware version (30010 in this example) and the bottom line is the firmware checksum (C1AC in this example).
30010
C1AC
3. Press and release the POWER MODE button to return to the ROM screen.
03 Series Operator’s Manual Calibration Mode • 54
Setup Mode
This section describes the 03 Series in Setup Mode. Setup Mode has the same menu items as Calibration Mode with the addition of the ALM--P menu item which allows you to set the alarm points and the
PASS--W menu item which allows you to turn the password feature on or off and set the password. The common menu items are included in Setup
Mode as a convenience to avoid turning off the unit and entering
Calibration Mode if you are updating alarm points and also want to use one of the other menu items. In Setup Mode, you can move through a menu of screens to do the following:
• Set the date and time (same as in Calibration Mode)
• Perform a fresh air adjustment (same as in Calibration Mode)
• Perform an automatic span adjustment (same as in Calibration Mode)
• Perform a manual span adjustment (same as in Calibration Mode)
• Set the alarm points
• Set the Bump Test Limit Check setting
• Turn the password feature on or off and define the password
NOTE:
You can set the 03 Series to alert you during the startup sequence when calibration or bump testing is due with the 03 Series User
Setup Program. See the 03 Series User Setup Program
Operator’s Manual for information on setting the Cal. Limit
Display
and Bump Test Limit Display parameter.
The description of the menu items that are common to Calibration Mode and Setup Mode will refer you to the appropriate pages in the Calibration
Mode section.
Using Setup Mode
WARNING: The 03 Series is not in operation as a gas detector while in Setup Mode. Although it will respond to gas in parts of
AIR CAL, A--CAL, and M--CAL, there are no gas alarm indications.
1. Take the 03 Series to a non-hazardous area and turn it off if it is on.
2. Press and hold the AIR button, then press and hold the POWER MODE button. You will hear a beep after one second. Continue to hold both the
AIR and the POWER MODE buttons.
3. After three seconds you will hear a second beep. Release both buttons
55 • Setup Mode 03 Series Operator’s Manual
when you hear the second beep.
4. If the PASS--W menu item is set to On, the following screen will appear with the first digit flashing prompting you to enter the password.
PASS--W
0 0 0 0
• Use the AIR button to scroll through the numbers 0-9 until you arrive at the desired value, then press and release the POWER MODE button to accept the value.
• Repeat the above procedure for each digit.
• When you have set and accepted the last digit of the password, the first menu item in Setup Mode will appear, the DATE Screen.
DATE
• If you entered the incorrect password, an error screen will appear.
You must use the POWER MODE button to turn the unit off and try again if you wish to enter Setup Mode or turn on the unit and proceed to Measuring Mode.
NOTE:
If you have forgotten your password, contact RKI Instruments, Inc.
5. If the PASS--W menu item is set to OFF, the DATE Screen is displayed.
DATE
6. Use the AIR button to move forward through Setup Mode. When you get to the last menu item, the START menu item, continuing will take you to the beginning of the menu.
7. When you arrive at the item you wish to enter, press and release the
POWER MODE button to enter that item.
03 Series Operator’s Manual Setup Mode • 56
8. When you need to adjust the numerical value of a parameter, increase it or decrease it, use the AIR button to change the value. When adjusting a numerical parameter value, it is possible to reverse the direction of adjustment. To switch from increasing to decreasing a value or decreasing to increasing a value, do the following:
• with the parameter flashing on the screen, press and hold the AIR button
• immediately press the POWER MODE button and then release both buttons
• the direction of adjustment when you press the AIR button is now reversed
9. When you are done using the menu items in Calibration Mode, use the
AIR button to scroll through the menu items to the START item.
START
10. At the START screen, press and release the POWER MODE button.
The 03 Series will begin its start-up sequence.
The Setup Mode menu items are described below in the order in which they appear while moving through Setup Mode.
Setting the Date and Time in DATE
See “Setting the Date and Time” on page 33 for instructions to set the date and time.
Performing a Fresh Air Adjustment in AIR
See “Performing a Fresh Air Adjustment” on page 33 for instructions to perform a fresh air adjustment.
Performing a Span Adjustment (Zero Adjustment for OX-03) in A--CAL
See “Performing an Automatic Span Adjustment (Zero Adjustment for OX-
03) in A--CAL” on page 34 for instructions to perform a span adjustment in
A--CAL.
Performing a Span Adjustment (Zero Adjustment for OX-03) in E--CAL
See “Performing an Easy Span Adjustment (Zero Adjustment for OX-03) in
57 • Setup Mode 03 Series Operator’s Manual
E--CAL” on page 39 for instructions to perform a span adjustment in E--
CAL.
Performing a Span Adjustment (Zero Adjustment for OX-03) in M--CAL
See “Performing a Manual Span Adjustment (Zero Adjustment for OX-03) in M--CAL” on page 45 for instructions to perform a span adjustment in M--
CAL.
Setting the Alarm Points
Entering the ALM--P menu item allows you to set the alarm points.
ALM--P
Table 5 below lists the factory set alarm points for each type of 03 Series.
Table 5: Factory Set Alarm Points
Channel
O
2
CO
H
2
S
Warning
19.5%
Decreasing
25 ppm
5 ppm
Alarm
23.5%
50 ppm
30 ppm
STEL n/a
200 ppm
5 ppm
TWA n/a
25 ppm
1 ppm
All of the alarms are increasing alarms, meaning the alarm condition actuates when the gas level increases above the alarm point, except for the O
2
Warning alarm. The O
2
Warning alarm is a decreasing alarm. The
O
2
Warning condition actuates when the oxygen level decreases below the alarm point.
The Warning setting cannot be adjusted higher than the Alarm setting.
1. At the ALM--P screen, press and release the POWER MODE button.
The following screen will display and the Warning alarm point will be flashing.
WARN
CO
25 ppm
03 Series Operator’s Manual Setup Mode • 58
2. Use the AIR button to adjust the Warning setpoint to the desired value.
3. Press and release the POWER MODE button to save the new Warning setpoint. A screen with the Alarm setpoint flashing is displayed.
ALRM
50 ppm
CO
4. Use the AIR button to adjust the Alarm setpoint to the desired value.
5. Press and release the POWER MODE button to save the new Alarm setpoint.
NOTE:
The following steps apply only to CO-03 and HS-03 instruments.
Users with OX-03 instruments can skip to Step 11.
6. A screen with the STEL setpoint flashing is displayed.
STEL
5 ppm
CO
7. Use the AIR button to adjust the STEL setpoint to the desired value.
8. Press and release the POWER MODE button to save the new STEL setpoint. A screen with the TWA setpoint flashing is displayed.
TWA
1 ppm
CO
9. Use the AIR button to adjust the TWA setpoint to the desired value.
10. Press and release the POWER MODE button to save the new TWA setpoint.
59 • Setup Mode 03 Series Operator’s Manual
11. The display will return to the following screen.
ALM--P
Adjusting the Bump Test Limit Check Setting
You can select how the instrument’s warmup sequence will handle a bump test being due using the BP--CHK. menu item. This menu item will always appear in Setup Mode even if Bump Test Function and/or Bump Limit
Display
are set to Off in the 03 Series User Setup Program.
BP--CHK.
1. At the BP--CHK. screen, press and release the POWER MODE button.
The following screen will display and the current setting will be displayed.
BP--CHK.
CONF
2. Use the AIR button to scroll through the options. CONF is the factory setting.
Table 6: Bump Test Limit Check Options
Operation Selection Description
CANT
CONF (factory setting) The user must press and release the AIR button to acknowledge that a bump test is due or press and release the POWER MODE button to perform a bump test before continuing with the warmup sequence.
NONE
Instrument cannot be used until a successful bump test is performed.
The instrument screen indicates a bump test is due upon startup but continues with the warmup sequence.
03 Series Operator’s Manual Setup Mode • 60
3. When the desired operation is displayed, press and release the
POWER MODE button to select it. The display will return to the BP--
CHK. menu item.
BP--CHK.
Setting the Password
You can password protect entry into the Setup Mode with the password feature. Entering the PASS--W menu item allows you to turn this feature on or off and enter a password if you turn it on.
PASS--W
The factory setting for the password feature is OFF.
1. At the PASSWORD screen, press and release the POWER MODE button. A screen will appear that shows the current password feature setting.
PASS--W
OFF
2. Press and release the AIR button to change the displayed setting.
When the desired setting is displayed, press and release the POWER
MODE button to save the setting.
3. If the password feature was set to OFF, the unit will exit the PASS--W menu item and return to the PASS--W screen.
PASS--W
With the password feature set to OFF, you will be able to enter Setup
61 • Setup Mode 03 Series Operator’s Manual
Mode without entering a password.
4. If the password feature was set to On, the following screen will appear with the first digit flashing prompting you to enter the password you want to use. The factory set password is 0300 but it can be changed.
PASS--W
0 3 0 0
5. Use the AIR button to scroll through the numbers 0 - 9.
6. When the desired number is displayed for the first digit, press and release the POWER MODE button to save the number.
7. The next digit will begin flashing.
8. Repeat Step 5 through Step 7 until you have set all four of the digits.
9. When you save the last digit, the unit will return to the PASSWORD screen.
PASS--W
03 Series Operator’s Manual Setup Mode • 62
Maintenance
This section describes troubleshooting procedures for the 03 Series. It also describes how to change the 03 Series’ batteries, replace the sensor, replace the sensor membrane, and replace the charcoal filter (CO-03 only).
WARNING: RKI Instruments, Inc. recommends that service, calibration, and repair of RKI instruments be performed by personnel properly trained for this work. Replacing sensors and other parts with original equipment does not affect the intrinsic safety of the instrument.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable causes, and recommended actions for problems you may encounter with the 03 Series.
Table 7: Troubleshooting the 03 Series
Symptoms Probable Causes
The LCD is blank.
• The unit may have been turned off.
• The alkaline batteries may need to be replaced.
The LCD shows abnormally high readings but other gas detection instruments do not.
No or low response to a known concentration of gas.
• The unit may need to be recalibrated.
• The sensor may need replacement.
• The unit may need to be recalibrated.
• The sensor may need replacement.
Recommended Action
1. To turn on the unit, press and hold the POWER
MODE button.
2. If the unit does not turn on, replace the alkaline batteries.
3. If the difficulties continue, contact RKI Instruments,
Inc. for further instruction.
1. Recalibrate the unit.
2. Replace the sensor and calibrate the unit.
3. If the difficulties continue, contact RKI Instruments for further instruction.
1. Recalibrate the unit.
2. Replace the sensor and calibrate the unit.
3. If the difficulties continue, contact RKI Instruments for further instruction.
63 • Maintenance 03 Series Operator’s Manual
Symptoms
“FAIL” displays during span adjustment.
Table 7: Troubleshooting the 03 Series
“SYSTEM FAIL” is displayed on the
LCD
Probable Causes
• The calibration value may not match the cylinder gas concentration.
• The sample gas is not reaching the sensor because of a bad connection.
• The calibration cylinder may be out of gas or is outdated.
• The sensor may need replacement.
• A microprocessor failure has occurred.
Recommended Action
1. Check all calibration tubing for leaks or for any bad connections.
2. Make sure the 03 Series has been properly set up for calibration.
3. Verify that the calibration cylinder contains an adequate supply of fresh test sample.
4. If the fail condition continues, replace the sensor.
5. If the difficulties continue, contact RKI Instruments,
Inc. for further instruction.
1. Turn off the unit and turn it on again.
2. If difficulties continue, contact RKI Instruments,
Inc.
Replacing the Batteries
WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be nonhazardous.
Replace the batteries when the battery icon indicates that the unit is in low battery warning. When in low battery warning, only one battery level indication bar is displayed in the battery icon on the LCD, and this icon will be flashing.
03 Series Operator’s Manual Maintenance • 64
To Replace the Batteries
1. Verify that the 03 Series is off.
2. Release the left side of the alligator clip or belt clip (if installed).
Figure 13: Clip Release
3. Rotate the captive battery cover screw counterclockwise to remove the battery cover.
Battery
Cov er
Screw
Battery C over
65 • Maintenance
Figure 14: Removing the Battery Cover
03 Series Operator’s Manual
4. Carefully remove the old alkaline batteries.
NOTE:
To avoid having to reset the date and time, new batteries must be installed within 2 minutes.
5. Carefully install the new AAA alkaline batteries. Follow the battery diagram inside the battery compartment.
6. Reinstall the battery cover.
7. If the instrument has been without batteries for more than 2 minutes, the date and time are reset and need to be set again. When the new batteries are installed, the instrument will turn on automatically and will display the Date/Time Screen. Set the date and time as described in
“Setting the Date and Time” on page 33. Once the date and time have been set, the instrument will begin its warmup sequence. If you do not set the date and time within 30 seconds, the instrument will automatically begin its warmup sequence.
Replacing the Sensor
WARNING: Replace the sensor in a non-hazardous environment.
NOTE:
An alligator or belt clip may be installed on the instrument but is not shown in the figures in this section.
1. Verify that the 03 Series is off.
2. Remove the rubber boot, if installed.
3. With a small Phillips screwdriver, carefully unscrew the four screws that attach the rear case to the front case.
4. Turn the instrument right side up. The screws are not captive and will fall out. Be sure not to lose them.
03 Series Operator’s Manual Maintenance • 66
5. Remove the front case from the rear case.
Front Case Rear Case
Screw, 4x
Rear
Case
Figure 15: Case Separation
6. Carefully remove the old sensor from the sensor socket.
7. Carefully insert the replacement sensor in the socket. Make sure the sensor face with the colored ring is facing up and that the sensor is keyed correctly (HS-03 and CO-03 only).
Tabs
Oxygen
Sensor
H2S
Sensor
CO
Sensor
Oxygen
Sensor Not
Keyed
Figure 16: Replacing the Sensor
Align Slots in H2S and CO Sensor with Tabs in Case
CAUTION: When replacing the sensor, verify that the sensor is properly aligned with its socket before inserting it into the socket. The
CO and H
2
S sensors have alignment slots which match up with alignment tabs in the sockets. Forcing a sensor into its socket may damage the sensor or the socket.
67 • Maintenance 03 Series Operator’s Manual
8. Reinstall the front case to the rear case.
9. Screw in the 4 screws that were removed in Step 3.
10. Reinstall the rubber boot if it is being used.
11. Calibrate the new sensor as described in “Calibration Mode” on page 31.
Replacing the Sensor Membrane and Charcoal Filter
WARNING: Replace the sensor membrane and charcoal filter in a non-hazardous environment.
NOTE:
An alligator or belt clip may be installed on the instrument but is not shown in the figures in this section.
When replacing the sensor filters, it is recommended that you replace all filters at the same time.The HS-03 and OX-03 only have a sensor membrane. The CO-03 has a sensor membrane and a charcoal filter.
1. Verify that the 03 Series is off.
2. Remove the rubber boot, if installed.
3. With a small Phillips screwdriver, carefully unscrew the four screws that attach the rear case to the front case.
4. Turn the instrument right side up. The screws are not captive and will fall out. Be sure not to lose them.
5. Remove the front case from the rear case.
Front Case Rear Case
Screw, 4x
Figure 17: Case Separation
03 Series Operator’s Manual Maintenance • 68
6. Turn the front case upside down so that the sensor gasket is visible.
Front Case
Sensor Membrane
Charcoal Filter
(CO-03 Only)
Sensor Gasket
Charcoal Filter
(CO-03 Only)
Sensor Gasket
Rear Case Front Case,
Inside View
Figure 18: Replacing the Sensor Filters
7. Carefully remove the sensor gasket. If you have a CO-03, the charcoal filter will be installed in the sensor gasket. The sensor membrane will be accessible once the sensor gasket has been removed. Carefully remove the sensor membrane.
8. If you have a CO-03, install a charcoal filter into the recess in the sensor gasket. There are tabs on either side of the recess that will hold the charcoal filter in place. For all models, insert a new sensor membrane in the sensor membrane recess.
9. Reinstall the sensor gasket with the flat side down.
10. Turn the front case right side up and carefully secure it to the rear case using the screws you removed in Step 3.
11. Reinstall the rubber boot if it is being used.
69 • Maintenance 03 Series Operator’s Manual
Parts List
Table 6 lists replacement parts and accessories for the 03 Series.
Table 8: Parts List
Part Number
06-1248RK
07-6031
07-6032
07-6033
10-1105-01
13-0112RK
13-0121
13-0122
13-0204RK
13-1116
20-0325
33-0175
33-7106RK
49-1110RK
81-0064RK-01
81-0064RK-03
81-0078RK-01
81-0078RK-03
81-0151RK-02
81-0151RK-04
81-1050RK
Description
Calibration kit tubing (specify length in feet)
Gasket for battery cover
Gasket between front and rear case
Sensor gasket
Screw, M2X 8mm pan head Phillips with washer,
SUSXM7, for connecting front and rear cases
Wrist strap
Alligator clip
Belt Clip
Spring bars for alligator and belt clips
Captive battery cover screw
Rubber boot
Sensor membrane
Filter disk, charcoal, 10 pack
AAA size alkaline battery
Calibration cylinder, 50 ppm CO in air, 34 liter steel
Calibration cylinder, 50 ppm CO in air, 103 liter steel
Calibration cylinder, 100% nitrogen, 34 liter steel
Calibration cylinder, 100% nitrogen, 103 liter steel
Calibration cylinder, 25 ppm H
2
S in nitrogen, 58 liter aluminum
Calibration cylinder, 25 ppm H
2
S in nitrogen, 34 liter aluminum
Regulator, fixed flow, 0.5 LPM, with gauge and knob, for 17/34 liter steel cylinder
03 Series Operator’s Manual Parts List • 70
Part Number
81-1051RK
81-1146
81-CO03
81-CO03-LV
81-HS03
81-HS03-LV
81-OX03
81-OX03-LV
ES-1821
ES-1827
OS-BM2
Table 8: Parts List
Description
Regulator, fixed flow, 0.5 LPM, with gauge and knob, for 58/34 liter aluminum cylinder, 103 liter steel cylinder
Calibration adapter for 03 Series
Calibration kit for CO-03 containing: one 103-liter steel gas cylinder (50 ppm CO in air), regulator, calibration adapter, case, tubing
Calibration kit for CO-03 containing: one 34-liter steel gas cylinder (50 ppm CO in air), regulator, calibration adapter, case, tubing
Calibration kit for HS-03 containing: one 58-liter aluminum gas cylinder (25 ppm H
2
S in nitrogen), regulator, calibration adapter, case, & tubing
Calibration kit for HS-03 containing: one 34-liter aluminum gas cylinder (25 ppm H
2
S in nitrogen), regulator, calibration adapter, case, & tubing
Calibration kit for OX-03 containing: one 103-liter steel gas cylinder (100% nitrogen), regulator, calibration adapter, case, tubing
Calibration kit for OX-03 containing: one 34-liter steel gas cylinder (100% nitrogen), regulator, calibration adapter, case, tubing
Carbon monoxide sensor
Hydrogen sulfide sensor
Oxygen sensor
71 • Parts List 03 Series Operator’s Manual
advertisement
Related manuals
advertisement
Table of contents
- 6 Introduction
- 7 Specifications
- 9 Description
- 10 Sensor Gasket, Sensor Membrane, and Charcoal Filter
- 10 Sensor
- 11 Control Buttons
- 11 Alarm LED
- 11 Buzzer
- 11 Vibrator
- 12 IrDA Port
- 12 Printed Circuit Boards
- 12 Batteries
- 13 Protective Rubber Boot
- 13 Alligator & Belt Clips
- 15 Start Up
- 15 Start-up Procedure
- 22 Performing a Fresh Air Adjustment
- 23 Turning Off the 03 Series
- 24 Operation
- 24 Measuring Mode
- 24 Adjusting the Buzzer Volume
- 25 Display Mode, CO-03 and HS
- 27 Display Mode, OX
- 30 Alarms
- 35 Data Logging
- 35 03 Series User Setup Program
- 36 Calibration Mode
- 36 Calibration Frequency
- 36 Using Calibration Mode
- 38 Setting the Date and Time
- 38 Performing a Fresh Air Adjustment
- 39 (Zero Adjustment for OX-03) in A--CAL
- 53 Performing a Bump Test in BUMP
- 59 Viewing the Instrument’s Firmware Version
- 60 Setup Mode
- 60 Using Setup Mode
- 62 Setting the Date and Time in DATE
- 62 Performing a Fresh Air Adjustment in AIR
- 62 (Zero Adjustment for OX-03) in A--CAL
- 63 Setting the Alarm Points
- 65 Adjusting the Bump Test Limit Check Setting
- 66 Setting the Password
- 68 Maintenance
- 68 Troubleshooting
- 69 Replacing the Batteries
- 71 Replacing the Sensor
- 73 Replacing the Sensor Membrane and Charcoal Filter
- 75 Parts List