- Industrial & lab equipment
- Measuring, testing & control
- EUTECH INSTRUMENTS
- pH 2100
- Instruction manual
advertisement
Instruction Manual pH 1100/ pH 2100
Bench pH/Ion Meter
Technology Made Easy ...
68X090818 rev 2.2 Nov 02
CyberScan pH 1100/ 2100
PREFACE
Thank you for selecting the pH 1100/ 2100 benchtop meter. This meter measure pH, millivolts, relative millivolts and temperature. The pH 2100 meter also measures ion concentration.
The instruction manual serves to explain the use of the pH 1100/ 2100 bench meter as a step-by-step operational guide to help you familiarize with the meter’s features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available.
This manual is written to cover as many anticipated applications and uses of the pH 1100/ 2100 Bench meter as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Authorized Distributor or Customer Service Dept. for assistance.
Eutech Instruments/ Oakton Instruments reserve the rights to change, make improvement and modify specifications without prior notice and cannot accept any responsibility for damage or malfunction to the instrument caused by improper use.
Copyright © 2002
Eutech Instruments Pte. Ltd.
Oakton Instruments
All rights reserved.
Revision 2.2, Nov 02.
Instruction Manual
TABLE OF CONTENTS
1 INTRODUCTION
1.1
Introducing the Bench Meter Series
1.2
Keypad
1.3
Rear Panel
1.4
Electrode Holder
2 STARTING UP
2.1
Back Panel Connections
2.2
Powering up and powering down
2.2.1
Powering up
2.2.2
Powering down
3
P
H CALIBRATION & MEASUREMENT
3.1
pH calibration
3.1.1
Automatic Temperature Compensation (ATC)
3.1.2
Starting pH calibration
3.1.3
Standard pH buffer calibration
3.1.4
Custom pH buffer calibration
3.1.5
Calibration error messages
3.2
pH measurement
3.2.1
Automatic temperature compensation
3.2.2
Adjusting manual temperature compensation
3.2.3
Taking pH Measurements
4 mV CALIBRATION AND MEASUREMENT
4.1
mV Calibration
4.1.1
mV Calibration error message
4.2
mV Measurement
5 ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100)
5.1
Ion calibration
5.2
Ion calibration error messages
5.3
Ion measurement
6 MEMORY FUNCTIONS
6.1
Memory input
6.1.1
Manually storing a reading into memory
6.2
Memory recall
6.2.1
Recalling readings in manual recall mode
6.2.2
Recalling readings in automatic recall mode
7 STABILITY INDICATOR
8 ALARM FUNCTIONS
8.1
High and Low measurement alarm
8.2
Calibration due alarm
9 TEMPERATURE CALIBRATION
9.1
Temperature Calibration
9.2
Temperature Calibration Error Messages
10 SETUP MODE
10.1
Setup mode overview
10.1.1
pH/Temperature setup submenus
10.1.2
mV setup submenus
10.1.3
Ion setup submenus (pH 2100 only)
10.1.4
Meter general configuration setup submenus
10.2
pH/ temperature setup mode
10.2.1
Entering pH/temperature setup mode
10.2.2
pH buffer setup program P1.0
10.2.3
pH resolution setup program P1.1
10.2.4
Temperature unit setup program P1.2
10.2.5
pH Measurement Alarm Setup Program P1.3
10.2.6
pH Calibration Due Alarm Setup program P1.4
10.2.7
Stability Setup Program P1.5
10.2.8
View pH calibration data setup program P1.6
4
CyberScan pH 1100/ 2100
18
18
19
18
20
21
22
20
23
23
23
24
24
24
25
14
15
16
16
11
11
12
13
16
17
10
10
10
10
11
6
6
7
8
9
10
25
25
25
26
26
26
31
31
32
33
34
35
29
30
30
30
27
27
28
28
27
Instruction Manual
10.2.9
pH calibration data reset setup program P1.7
10.3
mV setup mode
10.3.1
Entering mV setup mode
10.3.2
mV measurement alarm setup program P2.0
10.3.3
mV calibration due alarm setup program P2.1
10.3.4
Stability setup program P2.2
10.3.5
View mV calibration data setup program P2.3
10.3.6
mV calibration data reset setup program P2.4
10.4
Ion setup mode (pH 2100 meter only)
10.4.1
Entering ion setup mode
10.4.2
Ion unit setup mode P3.0
10.4.3
Ion measurement alarm setup program P3.1
10.4.4
Ion calibration due alarm setup program P3.2
10.4.5
Stability setup program P3.3
10.4.6
Ion mode setup program P3.4
10.4.7
View ion calibration data setup program P3.5
10.4.8
Ion calibration data reset setup program P3.6
10.5
Meter general configuration setup mode
10.5.1
Entering meter general setup mode
10.5.2
Date/time setup mode P4.0
10.5.3
Backlight setup mode P4.1
10.5.4
Data log setup mode P4.2
10.5.5
Memory recall setup mode P4.3
10.5.6
Communication setup mode P4.4
10.5.7
Data transfer setup mode P4.5
10.5.8
Memory clear setup mode P4.6
10.5.9
Meter reset setup mode P4.7
11 ELECTRODE CARE
11.1
Electrode Activation
11.2
Electrode Maintenance
11.3
Storing pH/ORP electrodes
12 RS 232 COMMUNICATION
12.1
Using with printer
12.1.1
Sending data to printer
12.2
Using with computer
13 ADDITIONAL INFORMATION
13.1
pH and Temperature
13.2
pH Buffer Calibration Solution
13.3
Standard pH Buffers
14 SUMMARY OF DEFAULT SETTINGS / OPTIONS
15 TROUBLESHOOTING & ERROR MESSAGES
16 SPECIFICATIONS
17 ACCESSORIES
18 WARRANTY
5
CyberScan pH 1100/ 2100
55
55
55
56
56
56
56
51
52
52
53
49
50
50
50
53
54
54
45
46
47
48
43
43
43
44
39
40
41
42
36
37
37
38
55
57
57
57
57
58
59
60
61
62
Instruction Manual CyberScan pH 1100/ 2100
1 INTRODUCTION
1.1 Introducing the Bench Meter Series
The pH 1100/ 2100 bench meter is microprocessor-based which incorporates new ASIC (Application Specific Integrated
Circuit). It is designed with convenience in mind and offers many advanced, user-friendly features. The meters are capable of storing and recalling up to 100 data sets in its non-volatile memory. In addition, as a space saver, an optional swivel electrode holder can be attached at the either side of bench meter for resting the electrodes and probes during operation.
The pH 1100/ 2100-- bench meter measures pH, millivolts, relative millivolts and temperature. The pH 2100 meter also measures ion concentration. These meters are equipped with a large customized LCD (Liquid Crystal Display) with simultaneous display of the measured values for easy reading. It is most ideal for routine pH/Ion Concentration measurement in indoor applications.
6
Instruction Manual CyberScan pH 1100/ 2100
1.2 Keypad
See Figure 1.
A large membrane keypad with touch feedback makes the meter easy to use. Names and symbols describe the function button controls.
Figure 1: Membrane Keypad
ON/OFF
Power meter On or Off
ENTER
EXIT
CAL / ¾
Confirm selection in all modes of operations;
Scroll through the sub-menus program in the SETUP modes
Exit from current modes of operations;
Exit from calibration mode after 3 rd
point calibration (for pH/Ion)
Enter into calibration modes of the meter;
Select or scroll to the next options available
Select the measurement modes:- pH, Ion Concentration, mV, Relative mV
MODE
HOLD
SETUP
¿
À
/ MI
/ MR
Freeze the displayed value; allow you to print the held reading or store it into memory
Enter SETUP mode of the meter for customization of meter functions as well as view some diagnostic functions (refer to the table for details of SETUP menus)
Store the displayed value into memory;
Increment values or scroll through the next options available
Recall stored values from the memory in the Last-In-First-Out (LIFO) sequence;
Decrement values or scroll through the next options available
Send measured data being measured or data stored in the memory to a peripheral device via the RS232C port of meter
Numeric Keys (0-9 & .)
Enter numeric values and decimal point at appropriate places
- (Minus)
Enter negative sign at appropriate places
7
Instruction Manual CyberScan pH 1100/ 2100
See Figure 2.
The pH 1100/ 2100 meters provide a complete set of input connections for the various accessories commonly used.
Listed in the table below are details of the connections that you can make.
Figure 2 : Rear View of the Meter
Connector
ATC
INPUT
Function
For phono jack connection from the temperature probe for Automatic Temperature Compensation.
The probe should be a 30K Ω thermistor probe.
For connecting sensors with a BNC connector to the meter. The meter accepts any pH, ORP or ISE electrode with a BNC connector. Always make sure that the connector is clean and dry.
REF
REC
GND
RS232
DC
For connection to the pin tip type reference electrode normally used with half cell (mono) type pH or
ISE electrodes.
For connection to the strip chart recorder.
For connection to the ground earth jack (standard pin tip connectors).
For connection to the RS232C serial port.
For connection of the AC adapter power supply (included).
8
Instruction Manual
The electrode holder is included in the same box as the meter.
To attach the electrode holder to the meter:
1. The electrode holder base attached to the bottom of the meter comes in the shipping position. See Figure 3.
2. Use a Phillips screwdriver to remove the screw holding the electrode holder in shipping position. See Figure 3.
3. Slide the electrode base away from the meter until the second screw slot lines up with the original screw hole. Use the screw removed earlier to secure electrode holder base into position. See Figure 3.
4. The electrode holder arm is reversible. If desired, remove screw holding electrode holder base and slide base out of brackets.
Slide base into brackets on opposite side and replace screw.
5. To install the electrode arm, turn meter back to the upright position. See Figure 4.
6. Line up the hole on the base of the electrode holder arm with the peg on the electrode holder base. Slide the hole securely onto the electrode holder base.
See Figure 4.
The electrode arm is now ready to swing into desired position.
Figure 3: Bottom of Meter
Figure 4: Fix electrode holder arm to holder base
9
CyberScan pH 1100/ 2100
Instruction Manual CyberScan pH 1100/ 2100
Attention:
Do not get water on the BNC connector during operation.
Avoid touching the connector with soiled or wet hands.
2.1 Back Panel Connections
Refer back to Figure 2. Make the necessary connections as mentioned in Section 1.3 Rear Panel, pg 8.
2.2 Powering up and powering down
1. When first connected to the power supply, the meter runs a self test, the display flashes Hardware Testing... Status:
Pass.
See Figure 5.
2. It then displays current date and time. See Figure 6. This display will also appear when the meter is turned off while still connected to the power supply. See Figure 6.
Figure 7. To turn the meter on press ON/OFF. The upper display flashes Bench pH/Ion 2100 for the pH 2100,
(Display flashes Bench pH 1100 for the pH 1100 meter.) and then the lower display flashes SW Rev 1.20. See
Figure 7.
4. The display flashes System Initializing, Please wait...
See Figure 8.
5. If this is the first power up since the power was connected, you must indicate desired measurement mode. Use >/CAL to select. The mode selected will flash. See Figure 9.
Hardware Testing. . . .
Status: Pass
Bench pH/Ion 2100
SW Rev 1.20
System Initializing
Please wait. . .
Figure 5
Date Time
Jan 25 2002 09:09:30
Figure 6
Figure 7
Figure 8
Press ENTER to confirm. The meter then enters measurement
pH/Temp mV Ion
Select & Press ENTER
6. After initial startup, the meter will automatically return to the measurement mode previously selected as long as Figure 9 power connection is maintained.
System Shutting Down
Please wait. . .
1. To turn the meter off press ON/OFF. The display flashes
Figure 10
System Shutting Down, Please wait... See Figure 10.
Then, the date and time display will appear. See Figure 6. The date and time will continue to display until the power supply is disconnected.
10
Instruction Manual CyberScan pH 1100/ 2100
3
P
H CALIBRATION & MEASUREMENT
This meter is capable of up to 5 point calibration with selectable buffer sets (USA, NIST, Bf1, Bf2 or any 5 custom buffers of your choice). The meter retains stored pH calibrations even when turned off. For best accuracy, we recommend that you perform at least a 2-point calibration using buffers that bracket (one above and one below) the expected sample range. You can perform a 1-point calibration, but make sure that the buffer value is close to the sample value you are measuring. For custom buffer sets, you must perform at least a 2 point calibration.
Selectable buffer sets:
USA: 1.68, 4.01, 7.00, 10.01, 12.45
NIST: 1.68, 4.01, 6.86, 9.18, 12.45
Bf1: 1.68, 4.01, 6.86, 10.01, 12.45
Bf2: 1.68, 4.01, 7.00, 9.18, 12.45
Any 5 user selected custom buffers
To select which buffer sets you wish to use see Section 10.2.2: Program pH Buffer Setup: P1.0 in the pH/Temp setup mode (page 30).
When you recalibrate your meter, old pH calibration points are replaced on a point by point basis. For example, if you previously calibrated your meter at pH 4.01, 7.00, and 10.01, and you recalibrate at pH 7.00, the meter retains the old calibration data at pH 4.01 and pH 10.01. To view current calibration points, see Section 10.2.8: Program pH Cal Data
P1.6 in the pH/Temp setup mode (page 35).
If you choose to recalibrate to only 1 or 2 pH values, the older calibration values you do not calibrate to will remain stored.
These old stored calibration values may cause accuracy loss when your readings are close to the old values. To clear old calibration data, reset the meter as shown in Section 10.2.9 Program pH Cal Reset: P1.7 in the pH/Temp setup mode
(page 36).
3.1.1 Automatic Temperature Compensation (ATC)
If you will be taking pH measurement using Automatic Temperature
Compensation (ATC), you must perform your pH calibration with
ATC temperature probe attached. Attach the temperature probe to the rear of the meter. The ATC mode annunciator shows on the display. See Figure 11. Insert the probe into the pH buffer solution along with your pH electrode.
If manual temperature compensation is preferred, do not plug a temperature probe into the meter. The MTC mode annunciator shows on the display. See Figure 12. The default manual temperature compensation is 25.0°C. To change the manual temperature compensation default, see Section 9 Temperature
Calibration for instructions (page 26).
Jan 18 2002 08:46:25
Figure 11
Jan 18 2002 08:46:25
Figure 12
11
Instruction Manual CyberScan pH 1100/ 2100
3.1.2 Starting pH calibration
1. Turn the meter ON. If the meter is not in the pH measurement mode, press EXIT and press >/CAL to select pH/Temp. The mode selected will flash. See Figure 13. Press ENTER to confirm selection. The meter is now in pH measurement mode.
See Figure 14.
pH/Temp mV Ion
Select & Press ENTER
2. Rinse the electrode in deionized water or rinse solution. If using the ATC function with a separate temperature probe, rinse the temperature probe as well. Do not wipe the pH electrode or
Figure 13
temperature probe dry. Wiping the probes causes static, and will create calibration and measurement instability.
3. Select pH buffer and pour into a clean container.
4. Dip the electrode and ATC probe into the calibration buffer. The
Jan 18 2002 08:46:25
Figure 14 end of the probes must be completely immersed into the sample. Stir probes gently to create a homogeneous sample.
>/CAL to enter the pH calibration mode. The display flashes Cal Mode Activated, Please wait....
See Figure 15. The meter is now in pH calibration mode.
Cal Mode Activated
Please wait. . .
Figure 15
The meter automatically recognizes the buffers in the set you have selected in the SETUP mode. If you wish to change the buffer set selected, see Section 10.2.2, Program pH Buffer Setup: P1.0 in the pH/Temp setup mode (page 30).
If your meter is set up for standard buffers (USA, NIST, Bf1, Bf2), proceed to Section 3.1.3: Standard pH buffer
calibration (page 13). If your meter is set up for custom buffers, proceed to Section 3.1.4: Custom pH buffer calibration
(page 14).
12
Instruction Manual CyberScan pH 1100/ 2100
3.1.3 Standard pH buffer calibration
1. When entering the standard calibration mode, the first buffer calibration point for standard pH buffers is 7.000 or 6.860
(depending upon which pH buffer set is selected). The upper display shows the current reading, the lower display will show
Auto Buf Scan: 7.000 (or 6.860). The upper display will flash pH until the reading has stabilized. See Figure 16. If the selected buffer value is not within ±1.0 of the measured pH value, the
Auto pH Scan will continue to change until the pH buffer is within ±1.0.
2. When pH reading has stabilized, press ENTER. The lower display flashes *pH Offset Cal Done. See Figure 17. This calibration point is now stored in the meter.
• If you are performing a one-point calibration, press EXIT.
The screen flashes Sys Updating...Please wait....The meter then returns to the pH measurement mode. You may now start taking pH readings.
• If you are performing a multi-point calibration, go to step 3.
3. Rinse the probe with deionized water or a rinse solution, and place it in the next pH buffer. The lower display automatically scans to the next pH buffer solution.
4. Wait for the measured pH value to stabilize. The upper display will flash pH until the reading has stabilized.
5. When pH reading has stabilized, press ENTER. The lower display flashes 2 Point Cal Done * (3 if you have calibrated 3 points,4 if 4 points, etc.). See Figure 18. This calibration point is now stored in the meter. Repeat steps 3-5 for each point you wish to calibrate. When you are finished calibrating, press EXIT.
The screen flashes Sys. Updating... Please wait.... The meter then returns to the pH measurement mode. You may now start taking pH readings.
6. If you perform a complete 5-point pH calibration, when you calibrate the fifth point and press ENTER, the meter automatically returns to measurement mode. You may now start taking pH readings.
NOTE:
7.001pH 25.4 C ATC
Auto Buf Scan: 7.000
10.010
°
Figure 16
Figure 17
2 Point Cal Done
Error: Cal Aborted
*
Figure 18
Figure 19
To exit from pH calibration mode at any point without confirming calibration, DO NOT press ENTER in steps 2, 5 or 6.
Press EXIT instead. If EXIT is press before a 1-point calibration is performed the upper display flashes Error: Cal Aborted.
See Figure 19.
DO NOT REUSE SOLUTIONS AFTER CALIBRATION.
Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements.
13
Instruction Manual CyberScan pH 1100/ 2100
3.1.4 Custom pH buffer calibration
1. When entering the custom calibration mode, the lower display will show Key In Std:. The upper display shows the pH buffer the electrode is dipped into. The lower display will show the same value as the upper display only if the meter has not been calibrated or if the meter has been reset. See Figure 20. If the meter has previously been calibrated, the lower display will toggle to the stored calibration point that is closest to current buffer value.
2. Using the numeric keypad (See Figure 21), enter the value of the first custom pH buffer. If you make a mistake, use the ∆ /MI or ∇ /MR to highlight mistake, then retype.
Figure 20 display will read 1 Point Cal Done *. See Figure 22.
NOTE: In the custom pH buffer calibration mode you must perform at least a 2-point calibration, if you press EXIT before you calibrate at 2 points the screen will flash Error: Cal Aborted
(See Figure 23) and will return to the pH measurement mode without confirming any calibration points.
4. Rinse the probe with deionized water or a rinse solution, and place it in the next custom pH buffer. The lower display will show Key In Std: The upper and lower displays show the pH value.
5. Enter the value of the next custom pH buffer. Press ENTER when done. The display will read 2 Point Cal Done *. If you are performing a 2-point calibration, press EXIT. The screen flashes
Sys Updating...Please wait... The meter then returns to the pH measurement mode. You may now start taking pH readings.
• If you are performing a 3, 4 or 5-point calibration go to step
6.
6. Repeat steps 4 and 5 for each additional custom buffer. Upon confirmation of pH buffer, the display will read 3 Point Cal Done
* (4 if you have calibrated points, etc). When you have finished calibrating desired number of points, press EXIT. The screen flashes Sys. Updating... Please wait...
The meter then returns to the pH measurement mode. If you have entered 5 calibration points, the meter will automatically return to pH measurement mode. You may start taking pH readings.
1
Numeric Keypad
Figure 21
Point Cal Done *
Figure 22
Error: Cal Aborted
Figure 23
NOTE:
To exit from pH calibration mode at any point without confirming calibration, DO NOT press ENTER in steps 3 or 5. Press
EXIT. If EXIT is pressed before a 2-point calibration is performed the upper display flashes Error: Cal Aborted. See
Figure 23.
DO NOT REUSE SOLUTIONS AFTER CALIBRATION.
Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements.
14
Instruction Manual CyberScan pH 1100/ 2100
3.1.5 Calibration error messages
Standard pH Buffer Calibration:
If the first buffer value of 7.000 or 6.860 is off by more than ±1.0, the upper display will flash Error: and the lower display will read
pH Offset Error. See Figure 24. Press EXIT to clear error message and return to pH measurement mode.
If any other pH buffer point is off by more than ±1.0, the upper display will flash Error: and the lower display will read pH Slope
Error. See Figure 25. Press EXIT, to clear error message and return to pH measurement mode.
Custom pH Buffer Calibration:
If the each custom pH buffer value is not within ±1.0 pH of the keyed in value, the upper display will flash Error: and the lower display will read Too much offset @7pH. See Figure 26. Press
EXIT to clear error message and return to pH measurement mode.
If the slope is not within 75% to 105%, the upper display will flash Error: and the lower display will read Slope out: 75% -
105%.
See Figure 27. Press EXIT to clear error message and return to pH measurement mode.
If adjacent calibration points are less than ±1.0 pH apart, the upper display will read Error: Pts too close, and the lower display will read OR slope diff. sign. See Figure 28. Press EXIT to clear error message and return to pH measurement mode.
/measurement
Error :
*
Figure 24
Error :
*
Figure 25
Error :
Too much Offset @7
Figure 26
Error :
Slope out:75% - 105%
Figure 27
Error :Pts too close
Or slope diff. sign
Figure 28
15
Instruction Manual CyberScan pH 1100/ 2100
This meter is capable of taking measurements with automatic or manual temperature compensation. Automatic Temperature
Compensation (ATC) occurs when a temperature sensor is plugged into the meter. If there is no temperature sensor plugged into the meter, the default setting is automatically 25°C
(if the meter has never been manually set for temperature) or the last manually set value. You can manually set the temperature to match your working conditions using a separate thermometer.
3.2.1 Automatic temperature compensation
For automatic temperature compensation, simply plug the temperature probe into the meter (see page 8). The ATC indicator will light on the LCD. See Figure 29.
NOTE: If you are using a temperature probe, the probe must be submersed in the liquid you are measuring.
3.2.2 Adjusting manual temperature compensation
IMPORTANT: For manual compensation, you must disconnect the temperature probe.
1. Turn the meter ON. If the meter is NOT in the pH measurement mode, press EXIT and use >/CAL to select pH/Temp. The mode selected will flash. See Figure 30.
Press ENTER to confirm selection. The meter is now in pH measurement mode. If the temperature probe is disconnected, the upper display will indicate MTC. See
Figure 31.
2. Press >/CAL to enter pH/Temp calibration mode. The display will flash Cal Mode Activated, Please wait... Press
>/CAL key again to enter the temperature calibration mode.
The display will flash Temperature Cal Mode.
(Please remove this gap. The copy should follow as one.)See
Figure 32. The upper display then reads MTC Temp:
25.0°C and the lower display reads Key In Std: 25.0°C. See
Figure 33.
3. Check the temperature of your sample using an accurate thermometer.
4. Key in the temperature value using the numeric keypad. If you make a mistake, use the ∆ /MI or ∇ /MR key to highlight mistake, then re-enter. pH measurement mode. Display will flash Temp Cal
Complete in the upper display and MTC Set Temp:
(temperature that was keyed in). See Figure 34. The display will flash Sys. Updating...Please wait... The meter then returns to pH measurement mode and will now compensate pH reading with the manually set temperature.
NOTE: To exit this program without confirming the manual temperature compensation value, do not press ENTER in step 5.
Press EXIT instead. The display will flash Error: Cal Aborted and meter will return to pH measurement mode.
°
Jan 18 2002 08:46:25
Figure 29
pH/Temp mV Ion
Select & Press ENTER
Figure 30
°
Jan 18 2002 08:46:25
Figure 31
Temperature Cal Mode
Figure 32
MTC Temp : 25.0 C
°
Figure 33
Temp Cal Complete
Figure 34
16
Instruction Manual CyberScan pH 1100/ 2100
3.2.3 Taking pH Measurements
NOTE: Be sure to remove the electrode soaker bottle or protective rubber cap on the electrode before measurement.
1. Rinse the probe with deionized or distilled water before use to remove any impurities adhering to the probe body. If the pH electrode dehydrated, soak it for 30 minutes in electrode storage solution or a 2M to 4M KCl solution.
ON. The upper display shows current pH reading and temperature, and indicates automatic or manual temperature compensation. The lower display shows month, date, year and time. See Figure 35.
3. Dip the probe into the sample. Make sure the sensor or the glass bulb of the electrode is completely immersed into the sample. Stir the probe gently to create a homogenous sample.
4. Allow time for the reading to stabilize. The pH indicator blinks when reading is not stable. See Figure 36. When the readings stabilizes, the pH indicator freezes and no longer blinks.
°
Jan 18 2002 08:46:25
Figure 35
°
Jan 18 2002 08:46:30
Figure 36
17
Instruction Manual CyberScan pH 1100/ 2100
4 mV CALIBRATION AND MEASUREMENT
1. Turn the meter ON. If the meter is not in the mV measurement mode, press EXIT and use >/CAL to select mV. The mode selected will flash. See Figure 37. Press
ENTER to confirm selection. The meter is now in absolute mV measurement mode. See Figure 38.
2. Rinse the electrode in deionized water or rinse solution.
Do not wipe the electrode dry. Wiping the probes causes static, and will create calibration and measurement instability.
3. Select your standard and pour into a clean container.
4. Dip the probe into the standard. The end of the probe must be completely immersed into the sample. Stir the probe gently to create a homogeneous sample.
pH/Temp mV Ion
Select & Press ENTER
Figure 37
Abs. mV : 226.9mV
Jan 20 2002 20:30:05
Figure 38
Cal Mode Activated
Please wait. . .
Figure 39
>/CAL key to enter the mV calibration mode. The display will read Cal Mode Activated, Please wait... See
Figure 39. The meter is now in mV calibration mode. The upper display shows Abs. mV: and lower display shows
Key in Std: See Figure 40.
NOTE: In mV calibration, the value shown in the upper display is the same as the lower display.
6. Key in the mV value to match the standard value using the numeric keypad. If you make a mistake, use the ∆ /MI or ∇ /MR to highlight mistake, then re-enter.
Abs. mV : 226.9mV
Key In Std: 226 . 9 mV mV Cal Completed
Figure 40
mV Offset: -3.7 mV
Figure 41 flashes mV Cal Completed, mV Offset: (shows the mV offset value). See Figure 41. It then flashes Sys
Updating... Please wait.... The meter returns to mV measurement mode. The upper display now reads Rel. mV. See Figure 42.
NOTE: New mV calibrations will override existing the stored mV calibration data. The meter retains stored mV calibrations even when the meter is turned off.
To exit from mV calibration mode at any point without confirming calibration, DO NOT press ENTER in step 7. Press
EXIT.
The upper display then flashes Error: Cal Aborted.
See Figure 43.
4.1.1 mV Calibration error message
Rel. mV : 200.0mV
Jan 20 2002 20:34:45
Error: Cal Aborted
Error :
Figure 42
Figure 43
mV Offset too large
Figure 44
If the offset adjustment of your mV calibration is not within ±150.0 mV; the upper display will flash Error: and the lower display will read mV Offset too large. See Figure 44. Press EXIT to clear error message and return to mV measurement mode.
18
Instruction Manual CyberScan pH 1100/ 2100
1. Turn the meter ON. If the meter is NOT in the mV measurement mode, press EXIT and use >/CAL to select mV.
The mode selected will flash. See Figure 45. Press ENTER to confirm selection.
The meter is now in Rel mV measurement mode. See Figure 46. If you have not performed a mV calibration or have reset the meter, the display will read Abs. mV. See Figure 47.
2. Dip the electrode into the sample. The sensor or glass bulb of the electrode must be completely immersed into the sample.
Stir the electrode gently to create a homogeneous sample.
3. Allow time for the reading to stabilize. The upper display will show the mV reading, and the lower display will show the current month, day and time. See Figure 46.
pH/Temp mV Ion
Select & Press ENTER
Figure 45
Rel. mV : 200.0mV
Jan 20 2002 20:34:45
Figure 46
Abs. mV : 226.9mV
Jan 20 2002 20:30:05
Figure 47
19
Instruction Manual CyberScan pH 1100/ 2100
5 ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100)
The pH 2100 meter reads ion concentration in ppm, ppt, mg/l, g/l, mmol/l and mol/l units. See Ion setup Program P3.0
(page 43) to select ion concentration units.
This instrument is capable of up to 5-point custom ion calibration with a minimum of 2-point calibration. Custom calibration values can be any value between 0.001 and 19998. Once the first calibration value is keyed in, all other calibration will automatically be at least one decade apart.
For example if your first calibration point is at 0.001, all other calibration points will be at least one decade apart from
0.001 (0.001, 0.01, 0.1,1.0, 10.0, 100, 1000, 10000).
For best accuracy calibrate your meter to points with similar concentrations to the solutions you want to test.
1. Connect an Ion Selective Electrode (ISE) to the BNC input connector on the back of the meter.
pH/Temp mV Ion
Select & Press ENTER
Figure 48 mode, press EXIT and use >/CAL to select Ion. The mode selected will flash. See Figure 48. Press ENTER to confirm selection. The meter is now in Ion measurement mode.
3. If the meter has not been calibrated or has been reset, the display will flash Ion Meas and current mV value. See Figure
49. If the meter has been calibrated previously, the display will show the current concentration value plus the absolute mV produced by the electrode.
4. Dip the electrode into the first calibration standard. Start with the calibration standard that has the lowest concentration and move up to higher concentration standards.
Ion Meas 160.3mV
Jan 20 2002 21:54:34
Figure 49
Cal Mode Activated
Please wait. . .
Figure 50
Ion Conc: ppm
Abs. mV : 100.7mV
Figure 51 flash Cal Mode Activated, Please wait... See Figure 50. The display then shows Ion Conc: ppm (or unit selected in ion setup mode) in the upper display and Abs. mV: in the lower display. See Figure 51.
6. The first calibration point must be keyed in by the user. Key in the concentration value of the standard using the numeric keypad. If you make a mistake, use the ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter.
ENTER to confirm the first calibration point. The display will flash 1 Cal Point Done *, then the meter proceeds to the next calibration point at least 1 decade apart from the first calibration point. See Figure 52.
8. Use /MI or ∇ /MR keys to move the cursor to next calibration point. All points will be a decade apart from the first calibration point.
9. Rinse off the electrode with deionized water before placing it in the next calibration standard.
10. Allow probe to stabilize in calibration standard. Press ENTER to confirm. The display flashes 2 Point Cal Done *. See
Figure 53.
The meter proceeds to the next calibration point.
Ion Conc:0.01ppm
1 Point Cal Done *
Ion Conc:0.10ppm
Figure 52
2 Point Cal Done *
Figure 53
20
Instruction Manual CyberScan pH 1100/ 2100
If you are performing a 2-point calibration, press EXIT.
The screen flashes Aver. Slope = (the average slope of the electrode) in the upper display and Sys Updating... in the lower display. See Figure 54. The meter returns to the Ion measurement mode. You may now start taking readings.
If you are performing a 3, 4 or 5-point calibration go to step 11.
11. Use ∆ /MI or ∇ /MR keys to move the cursor to next calibration point.
12. Rinse off the electrode with deionized water before placing it in the next calibration standard.
13. Allow probe to stabilize in calibration standard. Press ENTER to confirm. The display flashes 3 Point Cal Done * (4 for 4 point,
etc.).
See Figure 55. The meter proceeds to the next calibration point.
14. Repeat steps 11-13 until you finish calibrating at desired number of points (up to 5 points). Press EXIT. The screen flashes Aver.Slope = (the average slope of the electrode) in the upper display and Sys Updating... in the lower display. See
Figure 54. The meter returns to the Ion measurement mode.
At the end of a 5-point calibration, the meter automatically flashes Aver. Slope = and Sys Updating, and returns to measurement mode.
NOTES: You may compare the average electrode slope value with the expected slope value for your electrode from your electrode manual to verify electrode operation. If you want to leave the calibration mode prior to confirming two calibration points, press
EXIT key to exit calibration mode. The meter will revert to back to the earlier calibration curve. If you have calibrated to two points, pressing EXIT will confirm the new calibration. New ion calibrations will over-ride the existing stored ion calibration data. The meter retains stored ion calibrations even when the meter is turned off or if the power cord is unplugged or power is cut off to the AC adapter.
5.2 Ion calibration error messages
If you leave the ion calibration mode prior to confirming two calibration points by pressing the EXIT key, the screen will flash
Error: Cal Aborted. See Figure 56. The meter will revert to back to the earlier calibration curve and return to the ion measurement mode.
If the average ion slope is not within 10 to 75 mV per decade, the upper display will flash Error: and the lower display will read Ion
Slope Error. See Figure 57.
Aver. Slope = 49.2mV
Sys Updating. . .
Figure 54
Ion Conc: 1.0ppm
3 Point Cal Done *
Figure 55
Error: Cal Aborted
Figure 56
Error :
Ion Slope Error
Figure 57
21
Instruction Manual CyberScan pH 1100/ 2100
The pH 2100 meter features two different modes of ion measurement: high resolution mode and low resolution mode.
These modes are user selectable in the Ion setup program P3.4.
(page 47). The default is high resolution mode. Low resolution mode should only be used if your ion solution concentration will produce an absolute mV value of ±500 mV.
1. Turn the meter ON. If the meter is NOT in the Ion measurement mode, press EXIT and use >/CAL to select Ion.
The mode selected will flash. See Figure 58. Press ENTER to confirm selection. The meter is now in Ion measurement mode. If the meter has not been calibrated, or has been reset, the display will flash Ion Meas and current mV value. See
Figure 59. If the meter has been calibrated, the display will show the current concentration value plus the absolute mV produced by the electrode.
See Figure 60.
2. Prepare samples as necessary. Sample preparations varies depending on ion type—see your electrode manual for details on the specific electrode that you are using.
3. Dip electrode into the sample. The sensor or glass bulb of the electrode must be completely immersed into the sample.
4. Allow time for the reading to stabilize, then record your measurement.
NOTE: The display in the ion measurement mode shows mV that correlates to the electrode output. Charge in mV will typically be
59.7 mV for monovalent ions (1
+ or 1
- charge) and 29.8 mV for divalent ions (2
+ or 2
- charge).
pH/Temp mV Ion
Select & Press ENTER
Figure 58
Ion Meas 160.3mV
Jan 20 2002 21:54:34
Figure 59
0.013ppm 160.3mV
Jan 21 2002 21:35:34
Figure 60
22
Instruction Manual CyberScan pH 1100/ 2100
This meter can store up to 100 sets of data in several combinations: pH 1100 meter:
• pH, temperature, date and time
• mV, temperature, date and time pH 2100 meter:
• pH, temperature, date and time
• mV, temperature, date and time
• ion, mV, date and time
For example, you can store 32 pH and 68 mV values for the pH 1100 or 50 ion, 25 mV and 25 pH values for the pH 2100.
This meter can be setup to log readings automatically or manually. To set the meter to log automatically, see Meter
Setup Program P4.2, Data Log Setup (page 52). The meter can log readings at intervals from 10 seconds to 23 hours 59 minutes and 59 seconds.
6.1.1 Manually storing a reading into memory
1. During any measurement mode function, press ∆ /MI key to input data into memory.
2. The upper display will read current measurement function and
Data Log. . . . Loc: 9
the lower display will flash Data Log.... Loc: (the memory storage location). See Figure 61. The meter then returns to measurement mode.
NOTE: If the memory is full, the first value stored will be erased to create space for the new value.
Figure 61
23
Instruction Manual CyberScan pH 1100/ 2100
This function recalls the previous readings stored in the memory. You can only access memory recall from the measurement mode. Memory recall is in “Last In First Out” (LIFO) order.
There are two methods of viewing stored data in the memory recall function: manual and automatic. In manual memory recall, the user must manually scroll through the memory points using the ∆ /MI and ∇ /MR keys. In automatic memory recall, the meter automatically scrolls through the memory points. To choose the method of viewing stored data you wish to use, see Meter Setup mode Program P4.3 MemRecall Setup (page 52).
6.2.1 Recalling readings in manual recall mode the /MR key once to retrieve the last reading stored.
The display flashes Mem Recall Activated, Please wait... See
Figure 62.
The display then flashes the memory location last stored in the lower display. See Figure 63. Then the display shows the recorded value (pH/temp, mV or Ion/mV) in the upper display and toggles between memory location and date in the lower display. See Figures 64-65. the /MR key to select the next memory location or the
∆ /MI key to select the previous memory location.
3. Repeat step 2 to review additional stored data sets.
4. To exit Memory Recall, press EXIT and return to the measurement mode.
6.2.2 Recalling readings in automatic recall mode
1. Press the ∇ /MR key once to retrieve the last reading stored.
The display flashes Mem Recall Activated, Please wait... See
Figure 62. The display then flashes the memory location last stored in the lower display. See Figure 63. Then the display shows the recorded value (pH/temp, mV or Ion/mV) in the upper display and date and time in the lower display. See
Figure 64.
2. After 8 seconds, the meter automatically scrolls to the next to memory location. The display flashes the next memory location stored. See Figure 66. The display then shows the recorded value (pH/temp, mV or Ion/mV) in the upper display and date and time in the lower display. See Figure 67.
3. The meter will continuously scroll through all the memory points stored indefinitely. To exit memory recall, press EXIT and return to the measurement mode.
NOTES: If no new readings are stored after memory recall function has been exited, the next time the memory recall function is activated, it will resume at the memory location point where it was last exited. If a new point is stored, the memory recall function will begin with that point. Readings stored in memory are retained even if the unit is turned off. To erase all readings stored in memory, see
Meter Setup Program P4.6, Mem Clear Setup (page 54).
Mem Recall Activated
Please wait. . .
Mem Loc:9
Figure 62
Figure 63
Jan 21 2002 08:52:35
Mem Loc:9 08:52:35
Mem Loc:8
5.695pH 25.3 C ATC
5.695pH 25.3 C ATC
°
Figure 64
Figure 65
Figure 66
Jan 18 2002 08:50:35
Figure 67
24
Instruction Manual CyberScan pH 1100/ 2100
The stability indicator signals when pH, mV or Ion readings are stable.
The pH, mV or Ion annunciator will blink when reading are not stable.
See Figure 68. Once the reading becomes stable the annunciator becomes static and no longer blinks.
You can turn the stability indicator on and off and set the level of stability in pH Setup Section P1.5 (page 34); mV Setup P2.2 (page
40); and Ion Setup P3.3 (page 46).
°
Jan 18 2002 08:46:30
Figure 68
8.1 High and Low measurement alarm
Your meter has high and low measurement alarms that can be individually set for pH, mV and Ion (pH 2100 only). To set high and low measurement alarms, please refer to the setup sections pH Setup P1.3 (page 32); mV Setup P2.0 (page
38); and Ion Setup P3.1 (page 44).
If the measured value exceeds the alarm setting, the meter will “beep”; the upper display flashes measured parameter flashes; and the lower displays Hi-Alarm or Lo-Alarm. See Figures 69 and 70.
The following conditions are necessary to signal an alarm:
• pH: The high alarm must be at least 1.0 pH greater than the low alarm
• mV: The high alarm setting must be at least 100.0 greater than the low alarm
• Ion: The high alarm setting must be at least half of the next decade greater than the low alarm
Lo-Alarm 08:46:25
Figure 69
Hi-Alarm 08:46:25
Figure 70
pH Calibration Due
Calibration Abort
Figure 71
8.2 Calibration due alarm
You can set this meter to signal an alarm when calibration is due.
Individual calibration due alarms can be set for pH, mV and Ion
(pH 2100 only) at a period of 1 to 365 days from the last calibration.
To set calibration due alarm, please refer to the setup sections pH
Setup P1.4 (page 33); mV Setup P2.1 (page 39); and Ion Setup
P3.2 (page 45). The meter will “beep” during power up if calibration is due. The upper display will read pH (mV or Ion)
Calibration Due, the lower display will flash Calibration and read
Abort. See Figure 71.
Use >/CAL to select Calibration or Abort.
• Select
• meter calibration.
Abort to override alarm. Meter reverts to measurement mode.
Press ENTER to confirm selection.
25
Instruction Manual CyberScan pH 1100/ 2100
For this calibration procedure, the ATC probe should be attached to the meter. ATC will appear in the upper display. See
Figure 72.
1. Dip the temperature probe into a solution of known temperature, such as a temperature bath, for a few minutes until the temperature probe stabilizes.
2. If you are not already in pH/temp measurement mode, press
7.546pH 25.3 C ATC
Jan 18 2002 08:46:30
Figure 72
EXIT and use >/CAL to select pH/Temp. The mode selected will flash See Figure 73. Press ENTER to confirm selection.
The meter is now in pH measurement mode.
pH/Temp mV Ion
Select & Press ENTER
Figure 73 will flash Cal Mode Activated, Please wait... Press >/CAL again to enter the temperature calibration mode. The display will flash
Temperature Cal Mode. See Figure 74. Then the upper display will read ATC Temp: (current temperature) and the lower display will read Key In Std: (current temperature). See Figure
75.
4. Key in the known temperature value from step 1 using the numeric keypad. If you make a mistake, use the ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter.
5. Press ENTER to confirm. The display flashes Temp Cal
Complete, ATC Offset: (shows offset value). See Figure 76. The display then flashes Sys Updating, Please wait..., and the meter returns to pH measurement mode.
NOTE: The meter will not allow input values that exceed the allowable limit of ±5°C (±9°F) of the original displayed reading.
9.2 Temperature Calibration Error Messages
Temperature Cal Mode
Figure 74
ATC Temp : 25.0 C
Figure 75
Temp Cal Complete
°
Figure 76
Error: Cal Aborted
If you press EXIT to leave the temperature calibration mode prior to confirming the calibration, the screen will flash Error: Cal Aborted.
See Figure 77. The meter then returns to pH measurement mode. If the temperature value keyed in (temperature offset) is not within
±5.0°C (±9.0°F) of the current temperature value, the upper display will flash Error: and the lower display will read ATC offset too large.
See Figure 78. The meter returns to temperature calibration mode and a new temperature offset value needs to be entered.
Figure 77
Error :
ATC Offset too large
Figure 78
26
Instruction Manual CyberScan pH 1100/ 2100
10 SETUP MODE
The advanced setup mode lets you customize your meter’s preferences and defaults. The pH 2100 benchtop meter features 4 different setup modes (the pH 1100 meter features 3 modes):
1. pH/Temperature
2. mV
3. Ion (pH 2100 only)
4. Meter general configuration
Each setup mode has submenus to configure specific meter parameters.
10.1 Setup mode overview
10.1.1 pH/Temperature setup submenus
Press SETUP to enter setup mode from the pH measurement mode. Use >/CAL to select pH/Temp, and then press
ENTER to enter pH/Temp setup mode. Use the ∆ /MI and ∇ /MR keys to scroll through sub menus. There are 8 sub menus (programs) in the pH/Temperature setup mode: pH Buffer Setup:P1.0
-Selects calibration buffer sets
(USA, NIST, Bf1, Bf2 or
custom) pH Buffer Setup: P1.0
Press ENT. to Select see page 30 pHCalDue Alarm: P1.4
Press ENT. to Select see page 33 pH Resolution Setup:P1.1
-Selects pH resolution
(0.01 or 0.001 pH)
Temperature Unit Setup:P1.2
-Selects temperature
measurement unit (°C or °F) pH Measurement Alarm
Setup P1.3
-Sets low and high pH alarm pH Resol. Setup: P1.1
Press ENT. to Select see page 31
Temp Unit Setup:P1.2
Press ENT. to Select see page 31 pH Meas. Alarm :P1.3
Press ENT. to Select see page 32 pH Calibration Due Alarm
Setup:P1.4
-Sets calibration reminder
alarm, select 1-365 days
from last calibration date
Stability Setup:P1.5
-Sets stability indicator
criteria, off or on (high,
medium or low)
View Calibration Data:P1.6
-Last pH calibration
date/time
-pH buffer set selected
-pH calibration temperature
and temperature
compensation mode
-pH offset value
-pH slopes (from acid to
alkaline) pH Calibration Reset:P1.7
-Clears all pH calibration data
Stability Setup: P1.5
Press ENT. to Select see page 34 pH Cal Data :P1.6
Press ENT. to Select see page 35 pH Cal Reset :P1.7
Press ENT. to Select see page 36
27
Instruction Manual CyberScan pH 1100/ 2100
10.1.2 mV setup submenus
Press SETUP to enter setup mode. Use >/CAL to select mV, and then press ENTER. Use the ∆ /MI and
∇ /MR keys to scroll through submenus There are 5 submenus (programs) in the mV setup mode: mV Measurem ent Alarm
Setup :P2.0
-Sets low and high mV alarms mV Meas. Alarm : P2.0
Press ENT. to Select see page 38 mV Calibration Due Alarm
Setup: P2.1
-Sets calibration reminder
alarm, select 1-365 days from
last calibration date mVCalDue Alarm :P2.1
Press ENT. to Select see page 39
Stability Setup:P2.2
-Sets stability indicator criteria,
off or on (high, medium or low)
Stability Setup:P2.2
Press ENT. to Select see page 40
View mV Calibration
Data:P2.3
-Last mV calibration date/time
mV offset value mV Cal Data : P2.3
Press ENT. to Select see page 41 mV Calibration Reset:P2.4
-Clears all mV calibration data mV Cal Reset : P2.4
Press ENT. to Select see page 42
10.1.3 Ion setup submenus (pH 2100 only)
Press SETUP to enter setup mode from the Ion measurement mode. Use >/CAL to select ion, and then press ENTER. Use the ∆ /MI and ∇ /MR keys to scroll through submenus. There are 7 submenus (programs) in the ion setup mode:
Ion Unit Setup:P3.0
-Selects Ion measurement
unit(ppm, ppt, mg/L, g/L,
mmol/L, mol/L)
Ion Unit Setup :P3.0
Press ENT. to Select see page 43
Ion Measurement Alarm
Setup:P3.1
-Sets low and high Ion
alarms
Ion Meas. Alarm:P3.1
Press ENT. to Select see page 44
Ion Calibration Due Alarm
Setup:P3.2
-Sets calibration reminder
alarm, select 1-365 days
from last calibration date
IonCalDue Alarm:P3.2
Press ENT. to Select see page 45
Stability Setup:P3.3
-Sets stability indicator
criteria, off or on (high,
medium or low)
Stability Setup:P3.3
Press ENT. to Select see page 46
Ion Mode Setup:P3.4
-Selects between high or low
Ion resolution mode (high
has ±500 mV value, low has
±1850 mV value)
Ion Mode Setup:P3.4
Press ENT. to Select see page 47
View Calibration Data:P3.5
-Last Ion calibration
date/time. Calibration points
with absolute mV values
Ion slopes (from low to high
concentration)
Ion Cal Data :P3.5
Press ENT. to Select see page 48
Ion Calibration Reset:P3.6
-Clears all Ion calibration
data
Ion Cal Reset :P3.6
Press ENT. to Select see page 49
28
Instruction Manual CyberScan pH 1100/ 2100
10.1.4 Meter general configuration setup submenus
Press SETUP to enter setup mode. Use >/CAL to select meter, and then press ENTER. Use the ∆ /MI and ∇ /MR keys to scroll through submenus. There are 8 submenus (programs) in the meter setup mode:
Date/Time Setup:P4.0
Sets day, month, year, hour and minute
Date/Time Setup:P4.0
Press ENT. to Select see page 50
Memory Clear Setup:P4.6
Clears data from stored memory
Mem Clear Setup:P4.6
Press ENT. to Select see page 54
Backlight Setup:P4.1
-Turns backlight on or off and
selects amount of time
backlight stays on after button
press
Data Log Setup:P4.2
-Sets automatic or manual
memory recall and sets
datalogging time interval (10
seconds to 23 hours, 59
minutes, 59 seconds).
Memory Recall Setup:P4.3
-Selects automatic or manual
memory recall.
Rs232 Communication
Setup:P4.4
-Sets Baud Rate (4800, 9600 or
19200)
-Sets Stop Bit (One or Two)
-Sets Parity (None, Odd or
Even)
Backlight Setup:P4.1
Press ENT. to Select see page 51
Data Log Setup:P4.2
Press ENT. to Select see page 52
MemRecall Setup:P4.3
Press ENT. to Select see page 52
Tx/Rx Parameter: P4.4
Press ENT.
to Select see page 53
Meter Reset:P4.7
-Resets meter to factory
defaults
-Resets baud rate to 9600
-Resets stop bit to 2
-Resets parity to EVEN
-Clears all pH/mV/Ion
calibration data
-Resets manual temperature
compensation to 25.0°C
-Resets pH/mV/Ion
measurement alarm to OFF
-Resets pH/mV/Ion
calibration due alarm to
OFF
-Resets stable indicator to
ON with low stability setting
-Resets datalogging and
memory recall to manual
-Resets backlight to ON with
2.0 minute duration
-Clears all data from stored
memory
Meter Reset :P4.7
Press ENT. to Select see page 54
Data Transfer Setup:P4.5
-Selects what data gets sent to
computer/printer when PRINT
Data Transfer :P4.5
Press ENT. to Select see page 53
29
Instruction Manual CyberScan pH 1100/ 2100
10.2 pH/ temperature setup mode
10.2.1 Entering pH/temperature setup mode
1. Turn the meter ON. If the meter is not in the pH measurement mode, press EXIT and use >/CAL to select pH/Temp. The mode selected will flash. Press ENTER to confirm. The meter is now in pH measurement mode.
Setup Mode Activated
Please wait. . .
Figure 79
Setup Mode Activated, Please wait... See Figure 79.
3. The meter then enters the pH/Temp Meter Setup screen. Use
>/CAL to select pH/Temp. The mode selected will flash. See
Figure 80. Press ENTER to enter pH/Temp setup mode.
4. The first setup program in pH/Temp mode is pH Buffer Setup:
P1.0. See Figure 81. If you want to skip to another program, use the ∆ /MI and ∇ /MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter returns to the pH/Temp measurement mode.
NOTE: To exit the setup mode at anytime without confirming changes, press EXIT. If you have made any changes to any programs and press EXIT, these changes will be confirmed. To reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
10.2.2 pH buffer setup program P1.0
Setup: pH/Temp Meter
Select & Press ENTER
Figure 80
pH Buffer Setup:P1.0
Press ENT. to Select
Figure 81
Use this program to select calibration buffer sets: USA, NIST, Bf1, Bf2 or custom.
Selectable buffer sets:
USA: 1.68, 4.01, 7.00, 10.01, 12.45
NIST: 1.68, 4.01, 6.86, 9.18, 12.45
Bf1: 1.68, 4.01, 7.01, 9.18, 12.45
Bf2: 1.68, 4.01, 6.86, 10.01, 12.45
Custom: any 5 user selected custom buffers
1. Enter the pH/Temperature setup mode as described above
(10.2.1 Entering pH/temperature setup mode). The first setup mode is pH Buffer Setup: P1.0. See Figure 81.
USA NST Bf1 Bf2 Cust
Select & Press ENTER
Figure 82
Buffer set selected will flash. See Figure 82.
ENTER to confirm selection and return to the setup mode. Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program (P1.7 pH
Calibration Reset). Press EXIT to return to the pH measurement mode.
30
Instruction Manual
10.2.3 pH resolution setup program P1.1
Use this program to select pH resolution (0.01 or 0.001 pH).
1. Enter in 10.2.1 Entering pH/temperature setup mode (page 30).
2. Use /MI key to scroll to program pH Resol. Setup: P1.1. See
Figure 83.
4. Press >/CAL to toggle between 0.01 and 0.001. The resolution selected will flash. See Figure 84. mode. Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
10.2.4 Temperature unit setup program P1.2
Use this program to select temperature measurement unit (°C or
°F).
1. Enter pH/Temperature setup mode as described in steps 1-4 in 10.2.1 Entering pH/temperature setup mode (page 30).
2. Use /MI key to scroll to program Temp Unit Setup: P1.2. See
Figure 85.
3. Press unit selected will flash. See Figure 86.
ENTER to confirm selection and return to the setup mode. Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
Mode
CyberScan pH 1100/ 2100
pH Resol. Setup: P1.1
Press ENT. to Select
Figure 83
0.01 0.001
Select & Press ENTER
Figure 84
Temp Unit Setup:P1.2
Press ENT. to Select
Figure 85
Degree C Degree F
Select & Press ENTER
Figure 86
31
Instruction Manual CyberScan pH 1100/ 2100
10.2.5 pH Measurement Alarm Setup Program P1.3
Use this program to select alarm for low and high pH measurements. pH/Temperature in 10.2.1 Entering pH/temperature setup mode (page 30).
2. Use /MI to scroll to program pH Meas. Alarm: P1.3. See
Figure 87.
ENTER to enter program pH Meas. Alarm: P1.3.
selection will flash. See Figure 88.
ENTER to confirm selection.
• If you selected On continue to step 6
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
pH Set Point. Key in the low pH alarm set point using the numeric keypad. If you make a mistake, use the ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter. See Figure 89. Press
ENTER to confirm keyed in value.
7. The display now reads Key in High Alarm pH Set Point. Key in the high pH alarm set point using the numeric keypad. The high alarm must be set at least 1.0 pH greater than the low alarm. If you make a mistake, use the ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter. See Figure 90. the setup mode.
Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
pH Meas. Alarm :P1.3
Press ENT. to Select pH Alarm On
Figure 87
Off
Select & Press ENTER
Figure 88
Key In Low Alarm pH Set Point:2 . 000
Figure 89
Key In High Alarm pH Set Point:14 . 000
Figure 90
32
Instruction Manual CyberScan pH 1100/ 2100
10.2.6 pH Calibration Due Alarm Setup program P1.4
Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
10.2.1 Entering pH/temperature setup mode (page 30).
2. Use /MI to scroll to program pHCalDue Alarm: P1.4. See
Figure 91. selection will flash. See Figure 92.
• If you selected On, continue to step 6
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
6. If you selected On in step 5, the display reads Key in Cal Due
Period (1-365 days) =. Key number of days using the numeric keypad. If you make a mistake, use the ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter. See Figure 93.
7. Press ENTER to confirm number of days to next calibration, screen flashes Sys Updating, Please wait... and returns to the setup mode. Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press
EXIT to return to the pH measurement mode.
pHCalDue Alarm : P1.4
Press ENT.
to Select
Figure 91
Cal Alarm Due: On Off
Select & Press ENTER
Figure 92
Keyin Cal Due Period
(1-365 days) = 1
Figure 93
33
Instruction Manual CyberScan pH 1100/ 2100
10.2.7 Stability Setup Program P1.5
Use this program to turn on or off the pH stability indicator and to set level of stability (low, medium or high).
Stability definitions
High stability: reading is stabilized quickly, but reading is less accurate.
Medium stability: reading stability is averaged between high and low stability.
Low stability: reading is stabilized slowly, but guarantees high accuracy
1. Enter pH/Temperature setup mode as described in steps 1-
4 in 10.2.1 Entering pH/temperature setup mode (page 30).
∆ /MI to scroll to program Stability Setup: P1.5.
See
Figure 94. and Off. The selection will flash. See Figure 95.
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
6. If you selected On in step 5, the display now reads Degree
of Stability: Low Medium High. Press >/CAL to toggle between Low, Medium and High. The selection will flash.
See Figure 96.
ENTER to confirm selection and return to the setup mode. Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
Mode
Stability Setup : P1.5
Press ENT. to Select
Figure 94
pH Stable Criteria
On Off
Figure 95
Degree of Stability
Low Medium High
Figure 96
34
Instruction Manual CyberScan pH 1100/ 2100
10.2.8 View pH calibration data setup program P1.6
Use this program to view the following:
• Last pH calibration date/time
• pH buffer set selected
• pH calibration temperature and temperature compensation
• pH offset value
• pH slopes (from acid to alkaline)
1. Enter pH/Temperature setup mode as described in steps 1-4 in 10.2.1 Entering pH/temperature setup mode (page 30).
2. Use /MI to scroll to program pH Cal Data: P1.6. See Figure
97. screen in pH Cal Data: P1.6 is the pH Cal Date/Time. See
Figure 98. shows current buffer set and calibration points. See Figure 99.
ENTER to proceed to the third screen. This screen shows calibration temperature and temperature compensation mode. See Figure 100. shows current pH offset value in mV. See Figure 101. shows current acid slope 1 and 2. See Figure 102. shows current acid slope 2 and alkaline slope 1.—See Figure
103. shows current alkaline slope 1 and 2.—See Figure 104.
10. Press ENTER to return to the setup mode. Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
Note: While viewing calibration data, you may press EXIT at any time to return to the setup mode.
pH Cal Data :P1.6
Press ENT. to Select
Figure 97
pH Cal Date / Time
Jan 22 2002 09:01:46
Figure 98
USA: 1.680 4.010
7.000 10.010 12.450
Figure 99
Compensation = ATC
Figure 100
pH Offset = 0.2 mV
Figure 101
Acid Slope1 = 100.0 %
Acid Slope2 = 99.7 %
Figure 102
Acid Slope2 = 99.7 %
Alk. Slope1 = 99.7 %
Figure 103
Alk. Slope1 = 99.7 %
Alk. Slope2 = 98.2 %
Figure 104
35
Instruction Manual CyberScan pH 1100/ 2100
pH Cal Reset :P1.7
Press ENT. to Select
Figure 105
pH Reset: Yes No
Select & Press ENTER
Figure 106
pH being Reset
Please wait. . .
Figure 107
Reset Aborted
Figure 108
10.2.9 pH calibration data reset setup program P1.7
Use this program to erase all pH calibration data.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
Section 10.2.1 Entering pH/temperature setup mode (page 30).
2. Press /MI to scroll to program pH Cal Reset: P1.7. See
Figure 105.
3. Press selection will flash. See Figure 106.
ENTER to confirm selection.
• If you selected Yes, the screen flashes pH being Reset,
Please wait... See Figure 107. The meter then returns to setup mode.
• If you select No, the screen flashes Reset Aborted and then returns to the setup mode. See Figure 108.
This is the last pH setup program. Press ∇ /MR to go to the previous program. Press ∆ /MI to go back to the first pH setup program. Press
EXIT to return to the pH measurement mode.
36
Instruction Manual CyberScan pH 1100/ 2100
10.3 mV setup mode
10.3.1 Entering mV setup mode
1. Turn the meter ON. If the meter is not in the mV measurement mode, press EXIT and use >/CAL to select mV. The mode selected will flash. Press ENTER to confirm selection. Their meter is now in mV measurement mode.
Setup Mode Activated, Please wait... See Figure 109.
Figure 110. Press ENTER to enter mV setup mode. You are now in the mV setup mode.
4. The first setup program in mV mode is mV Meas. Alarm: P2.0.
See Figure 111. If you want to skip to another program, use the ∆ /MI and ∇ /MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter returns to the mV measurement mode.
NOTE: To exit the setup mode at anytime without confirming changes, press EXIT. If you have made any changes to any programs and press EXIT, these changes will be confirmed. To reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
Setup Mode Activated
Please wait. . .
Figure 109
Setup: mV Meter
Select & Press ENTER
Figure 110
mV Meas. Alarm :P2.0
Press ENT. to Select
Figure 111
37
Instruction Manual CyberScan pH 1100/ 2100
10.3.2 mV measurement alarm setup program P2.0
Use this program to select alarm for low and high mV measurements.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37). The first setup mode is mV Meas. Alarm: P2.0. See Figure 112.
2. Press ENTER to enter program mV Meas. Alarm: P2.0.
The selection will flash. See Figure 113.
• If you selected On, continue to step 5.
• If you select Off, pressing ENTER returns to the setup mode.
To continue to the next setup program, press ∆ /MI.
To continue to the previous program, press ∇ /MR.
Press EXIT to return to the mV measurement mode.
5. If you selected On in step 4, the display now reads Key in Low
Alarm mV Set Point. Key in the low mV alarm set point using the numeric keypad, using the “-” key for negative numbers.
See Figure 114. If you make a mistake, use the ∆ /MI or ∇ /MR to highlight mistake, then re-enter.
Press ENTER to confirm value.
6. The display now reads Key in High Alarm mV Set Point. The high alarm setting must be at least 100.0 mV greater than low alarm. Key in the high mV alarm set point using the numeric keypad, using the “-” key for negative numbers. See Figure 115.
If you make a mistake, use the ∆ /MI or ∇ /MR to highlight mistake, then re-enter. Press ENTER to confirm value. flashes Sys Updating, Please wait... and returns to the setup mode. To continue to the next setup program, press ∆ /MI. To continue to the previous program, press ∇ /MR. Press EXIT to return to mV measurement mode.
8. Setup
mV Meas. Alarm :P2.0
Press ENT. to Select
Figure 112
mV Alarm On Off
Select & Press ENTER
Figure 113
Key In Low Alarm mV Set Point:-1550.0
Figure 1 14
Key In High Alarm mV Set Point:1550.0
Figure 115
38
Instruction Manual CyberScan pH 1100/ 2100
10.3.3 mV calibration due alarm setup program P2.1
Use this program to set a reminder when the next calibration is due. Choose from 1 to 365 days from last calibration.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use /MI key to scroll to program mVCalDue Alarm: P2.1.
See Figure 116.
The selection will flash. See Figure 117.
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to mV measurement mode.
6. If you selected On in step 5, the display reads Key in Cal
Due Period (1-365 days) =. Key number of days using the numeric keypad. If you make a mistake, use the ∆ /MI or
∇ /MR keys to highlight mistake, then re-enter.
See Figure 118. screen flashes Sys Updating, Please wait... and returns to the setup mode.
Press ∆ /MI to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to mV measurement mode.
mVCalDue Alarm :P2.1
Press ENT. to Select
Figure 116
Cal Alarm Due On Off
Select & Press ENTER
Figure 117
Keyin Cal Due Period
(1-365 days) = 1
Figure 118
39
Instruction Manual
10.3.4 Stability setup program P2.2
Use this program to turn on or off the mV stability indicator and to set level of stability (low, medium or high).
Stability definitions
High stability: reading is stabilized quickly, but reading is less accurate.
Medium stability: reading stability is averaged between high and low stability.
Low stability: reading is stabilized slowly, but guarantees high accuracy
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use /MI key to scroll to program Stability Setup: P2.2. See
Figure 119.
Off. The selection will flash. See Figure 120.
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the mV measurement mode.
6. If you selected On in step 5, the display now reads Degree of
Stability: Low Medium High. Press >/CAL to toggle between
Low, Medium and High. The selection will flash.
See Figure 121.
ENTER to confirm selection and return to the setup mode. Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the pH measurement mode.
CyberScan pH 1100/ 2100
Stability Setup : P2.2
Press ENT. to Select
Figure 119
mV Stable Criteria
On Off
Figure 120
Degree of Stability
Low Medium High
Figure 121
40
Instruction Manual CyberScan pH 1100/ 2100
10.3.5 View mV calibration data setup program P2.3
Use this program to view the following:
• Last mV calibration date/time
• mV offset value
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use /MI to scroll to program mV Cal Data: P2.3. See Figure
122.
ENTER to enter program mV Cal Data: P2.3. The first screen in mV Cal Data: P2.3. is the mV Cal Date/Time. See
Figure 123. shows current mV offset value. See Figure 124.
Press ∆ /MI to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to mV measurement mode.
Note: While viewing calibration data, you may press EXIT at any time to return to the setup mode.
mV Cal Data :P2.3
Press ENT. to Select
Figure 122
mV Cal Date/ Time
Jan 22 2002 11:55:08
Figure 123
mV Offset = 3.2 mV
Figure 124
41
Instruction Manual CyberScan pH 1100/ 2100
10.3.6 mV calibration data reset setup program P2.4
Use this program to erase all mV calibration data.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use /MI key to scroll to program mV Cal Reset: P2.4.
See Figure 125.
The selection will flash. See Figure 126.
5. Press
• If you selected Yes, the screen flashes mV being Reset,
Please wait... See Figure 127. The meter then returns to setup mode.
• If you select No, the screen flashes Reset Aborted and then returns to the setup mode. See Figure 128. This is the last mV setup program. Press ∇ /MR to go to the previous program. Press ∆ /MI to go back to the first mV setup program. Press EXIT to return to mV measurement mode.
mV Cal Reset :P2.4
Press ENT. to Select
Figure 125
mV Reset: Yes No
Select & Press ENTER
Figure 126
mV being Reset
Please wait. . .
Figure 127
Reset Aborted
Figure 128
42
Instruction Manual CyberScan pH 1100/ 2100
10.4 Ion setup mode (pH 2100 meter only)
10.4.1 Entering ion setup mode mode, press EXIT and use >/CAL key to select ion. The mode selected will flash. Press ENTER to confirm selection. The meter is now in Ion measurement mode.
Setup Mode Activated, Please wait... See Figure 129. selected will flash. Press ENTER to enter Ion setup mode. You are now in the Ion setup mode.
4. The first setup program in Ion mode is Ion Unit Setup: P3.0.
See Figure 131. If you want to skip to another program, use the ∆ /MI and ∇ /MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter returns to the ion measurement mode.
NOTE: To exit the setup mode at anytime without confirming changes, press EXIT. If you have made any changes to any programs and press EXIT, these changes will be confirmed. To reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
10.4.2 Ion unit setup mode P3.0
Use this program to select Ion measurement units: ppm, ppt, mg/L, g/L, mmol/L, or mol/L
1. Enter ion setup mode as described in above in Section 10.4.1
Entering ion setup mode. The first setup mode is Ion Unit
Setup: P3.0 See Figure 132.
>/CAL key to toggle between ppm, ppt, mg/l, g/l, mmol/l or
mol/l. The unit selected will flash. See Figure 133.
ENTER to confirm selection and return to the setup mode. To continue to the next setup program, press ∆ /MI. To continue to the previous program press ∇ /MR. Press EXIT to return to the ion measurement mode.
Setup Mode Activated
Please wait. . .
Figure 129
Setup: Ion Meter
Select & Press ENTER
Figure 130
Ion Unit Setup : P3.0
Press ENT. to Select
Figure 131
Ion Unit Setup :P3.0
Press ENT. to Select
Figure 132
ppm ppt mg/1 g/l mmol/l mol/l
Figure 133
43
Instruction Manual CyberScan pH 1100/ 2100
10.4.3 Ion measurement alarm setup program P3.1
Use this program to select alarm for low and high ion measurements.
1. Enter ion setup mode as described in steps 1-4 in Section 10.4.1
Entering ion setup mode (page 43).
2. Use /MI key to scroll to program Ion Meas. Alarm: P3.1. See
Figure 134. selection will flash. See Figure 135.
4. Press
• If you selected On, continue to step 5.
• If you select Off, pressing ENTER returns to the setup mode.
To continue to the next setup program, press ∆ /MI.
To continue to the previous program, press ∇ /MR.
Press EXIT to return to the ion measurement mode.
5. If you selected On in step 4, the display now reads Key in Low
Alarm Ion Set Point. Key in the low ion alarm set point using the numeric keypad. See Figure 136. If you make a mistake, use the ∆ /MI or ∇ /MR to highlight mistake, then re-enter. Press
ENTER to confirm value.
6. The display now reads Key in High Alarm Ion Set Point. The high alarm setting must be at least one-half decade greater than the low alarm. Key in the high ion alarm set point using the numeric keypad. See Figure 137. If you make a mistake, use the ∆ /MI or ∇ /MR to highlight mistake, then re-enter. Press
ENTER to confirm value. flashes Sys Updating, Please wait... and returns to the setup mode. To continue to the next setup program, press ∆ /MI. To continue to the previous program, press ∇ /MR. Press EXIT to return to mV measurement mode.
Ion Meas. Alarm :P3.1
Press ENT. to Select
Figure 134
Ion Alarm On Off
Select & Press ENTER
Figure 135
Key In Low Alarm
Ion Set Point : 0.001
Figure 136
Key In High Alarm
Ion Set Point:19999
Figure 137
44
Instruction Manual CyberScan pH 1100/ 2100
10.4.4 Ion calibration due alarm setup program P3.2
Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.
1. Enter ion setup mode as described in steps 1-4 in Section 10.4.1
Entering ion setup mode (page 43).
∆ /MI key to scroll to program IonCalDue Alarm: P3.2.
See
Figure 138.
3. Press
The selection will flash. See Figure 139.
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to mV measurement mode.
6. If you selected On in step 5, the display reads Key in Cal Due
Period (1-365 days) = 1. Key number of days using the numeric keypad. If you make a mistake, use ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter. See Figure 140. screen flashes Sys Updating, Please wait… and returns to the setup mode. Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press
EXIT to return to mV measurement mode.
Cal Alarm Due: On Off
Select & Press ENTER
Figure 139
IonCalDue Alarm :P3.2
Press ENT. to Select
Figure 138
Keyin Cal Due Period
(1-365 days) = 1
Figure 140
45
Instruction Manual
10.4.5 Stability setup program P3.3
Use this program to turn on or off the ion stability indicator and to set level of stability (low, medium or high).
Stability definitions
High stability:
Reading is stabilized quickly, but reading is less accurate.
Medium stability:
Reading stability is averaged between high and low stability.
Low stability:
Reading is stabilized slowly, but guarantees high accuracy.
1. Enter ion setup mode as described in steps 1-4 in Section
10.4.1 Entering ion setup mode (page 43).
2. Use /MI key to scroll to program Stability Setup: P3.3 See
Figure 141. and Off. The selection will flash. See Figure 142.
ENTER to confirm selection.
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆ /MI key to continue to the next setup program, or
Press ∇ /MR to go to the previous program. Press EXIT to return to the ion measurement mode.
6. If you selected On in step 5, the display now reads Degree of
Stability: Low Medium High. . Press >/CAL to toggle between
Low, Medium and High. The selection will flash. See Figure
143.
ENTER to confirm selection and return to the setup mode. Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the ion measurement mode.
CyberScan pH 1100/ 2100
Stability Setup : P3.3
Press ENT. to Select
Figure 141
Ion Stable Criteria
On Off
Figure 142
Degree of Stability
Low Medium High
Figure 143
46
Instruction Manual CyberScan pH 1100/ 2100
10.4.6 Ion mode setup program P3.4
Use this program to select high or low ion resolution mode. High resolution has ±500 mV value, low has ±1850 mV value.
1. Enter ion setup mode as described in steps 1-4 in Section
10.4.1 Entering ion setup mode (page 43).
2. Use /MI key to scroll to program Ion Mode Setup: P3.4. See
Figure 144.
3. Press
4. Press >/CAL to toggle between High or Low. The unit selected will flash. See Figure 145. mode.
6. Press /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to the ion measurement mode.
Ion Mode Setup : P3.4
Press ENT. to Select
Figure 144
Resolution: High Low
Select & Press ENTER
Figure 145
47
Instruction Manual CyberScan pH 1100/ 2100
10.4.7 View ion calibration data setup program P3.5
Use this program to view the following:
• Last ion calibration date/time
• Calibration points with absolute mV values
• Ion slope values
1. Enter ion setup mode as described in steps 1-4 in Section
10.4.1 Entering ion setup mode (page 43).
2. Use /MI key to scroll to program Ion Cal Data: P3.5.
See Figure 146. screen in Ion Cal Data: P3.5. is the Ion Cal Date/Time.
See Figure 147.
ENTER to proceed to the second screen, showing 1 st and 2 nd
ion calibration points and corresponding absolute mV value. See Figure 148.
ENTER to proceed to the third screen, showing 2 nd and
3 rd
ion calibration points and corresponding absolute mV value.
See Figure 149.
6. Press rd and
4 th
ion calibration points and corresponding absolute mV value.
See Figure 150.
ENTER to proceed to the fifth screen, showing 4 th
and 5 th ion calibration points and corresponding absolute mV value.
See Figure 151.
Slope 1 and 2 mV value. See Figure 152.
ENTER to proceed to the seventh screen, showing Ion
Slope 2 and 3 mV value. See Figure 153.
10. Press ENTER to proceed to the eighth screen, showing Ion
Slope 3 and 4 mV value. See Figure 154.
Press ∆ /MI key to continue to the next setup program, or Press
∇ /MR to go to the previous program. Press EXIT to return to ion measurement mode.
NOTE: While viewing calibration data, you may press EXIT at any time to return to the setup mode.
Ion Cal Data :P3.5
Press ENT. to Select
Figure 146
Ion Cal Date/ Time
Jan 22 2002 14:55:38
Figure 147
1st :0.001
0.6 mV
2nd:0.010 50.1 mV
Figure 148
2nd:0.010
50.1 mV
Figure 149
3rd:0.100
100.8 mV
4th: 1 .00
149.4 mV
Figure 150
4th: 1 . 00 149.4 mV
5th: 10 . 0 201.0 mV
Figure 151
Ion Slope1 = 49.5 mV
Ion Slope2 = 50.6 mV
Figure 152
Ion Slope2 = 50.6 mV
Ion Slope3 = 48.6 mV
Figure 153
Ion Slope3 = 48.6 mV
Ion Slope4 = 51.5 mV
Figure 154
48
Instruction Manual CyberScan pH 1100/ 2100
10.4.8 Ion calibration data reset setup program P3.6
Use this program to erase all ion calibration data.
1. Enter ion setup mode as described in steps 1-4 in Section
10.4.1 Entering ion setup mode (page 43).
2. Use /MI key to scroll to program Ion Cal Reset: P3.6. See
Figure 155.
No. The selection will flash. See Figure 156.
• If you selected Yes, the screen flashes Ion being
Reset, Please wait... See Figure 157. The meter then returns to setup mode.
• If then returns to the setup mode. See Figure 158.
This is the last ion setup program. Press ∇ /MR to go to the previous program. Press ∆ /MI to go back to the first ion setup program. Press EXIT to return to mV measurement mode.
Mode
Ion Cal Reset :P3.6
Press ENT. to Select
Figure 155
Ion Reset: Yes No
Select & Press ENTER
Figure 156
Ion being Reset
Please wait. . .
Figure 157
Reset Aborted
Figure 158
49
Instruction Manual CyberScan pH 1100/ 2100
10.5 Meter general configuration setup mode
10.5.1 Entering meter general setup mode
1. Turn the meter ON. The meter setup mode can be entered from any measurement mode—pH, mV or ion (pH 2100 meter only).
Setup Mode Activated, Please wait. ..See Figure 159.
See Figure 160, 161 and 162. Press ENTER to enter Meter setup mode. You are now in the Meter setup mode.
4. The first setup program in Ion mode is Date/Time Setup: P4.0.
See Figure 163. If you want to skip to another program, use the ∆ /MI and ∇ /MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter returns to the measurement mode that it was in prior to entering meter setup mode.
NOTE: To exit the setup mode at anytime without confirming changes, press EXIT. If you have made any changes to any programs and press EXIT, these changes will be confirmed. To reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
10.5.2 Date/time setup mode P4.0
Use this program to set the meter date and time.
1. Enter the Meter setup mode as described above in Section
10.5.1 Entering meter general setup mode (Pg 50). The first setup mode is Date/Time Setup: P4.0. See Figure 163.
3. Key in the Date, Month, Year, Hr and Min using the numeric keypad. Use the >/CAL key or the ENTER key to toggle to each field. If you make a mistake keying the value, use the ∆ /MI and
∇ /MR keys to highlight mistake, then re-enter.
See Figure 164.
4. After keying in a value for the last field (Min), press ENTER to confirm. The screen flashes Clock Updating, Please wait. . See
Figure 165. The clock then flashes the date and time.
See Figure 166.
Press ∆ /MI key to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to the measurement mode you were in prior to entering meter setup mode.
Setup Mode Activated
Please wait. . .
Figure 159
Setup:pH/Temp Meter
Select & Press ENTER
Figure 160
Setup: mV Meter
Select & Press ENTER
Figure 161
Setup: Ion Meter
Select & Press ENTER
Figure 162
Date/Time Setup:P4.0
Press ENT. to Select
Clock Updating
Please wait. . .
Figure 163
Date:23 Month:01
Year:02 Hr:8 Min:07
Figure 164
Figure 165
Date Time
Jan 23 2002 08”07:31
Figure 166
50
Instruction Manual CyberScan pH 1100/ 2100
10.5.3 Backlight setup mode P4.1
Use this program to turn backlight on or off and set duration time
(0.1 to 10 minutes).
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Backlight Setup: P4.1.
See Figure 167.
3. Press
The selection will flash. See Figure 168.
BackLight Setup:P4.1
Press ENT. to Select
Figure 167
BackLight On Off
Select & Press ENTER
Figure 168
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Key BL ON Period
( 0 . 1-10 . 0 min) = 2.00
Press ∆ /MI to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
Figure 169
6. If you selected On in step 5, the display now reads Key BL On
Period, (0.1-10.0 min) =2.00. See Figure 169. Key in the backlight on period using the numeric keypad. If you make a mistake, use ∆ /MI or ∇ MR keys to highlight mistake, then reenter .
ENTER to confirm selection. The screen flashes Sys
Updating... Please wait... and then returns to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
51
Instruction Manual CyberScan pH 1100/ 2100
10.5.4 Data log setup mode P4.2
Use this program select automatic or manual data logging.
Automatic data logging can be set at timed intervals from 10 seconds to 23 hours, 59 minutes, 59 seconds.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Data Log Setup: P4.2.
See Figure 170.
4. Press >/CAL key to toggle between Data Log Auto and
Manual. The selection will flash. See Figure 171.
5. Press
• If you selected Auto, continue to step 6.
• If you select Manual, pressing ENTER returns to the setup mode. Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program.
Press EXIT to return to measurement mode.
6. If you selected Auto in step 5, the display now reads Auto
DataLog Period Hr: Min: Sec : Key in the time period
(minimum of 10 seconds to maximum of 23 hours, 59 minutes,
59 seconds) using the numeric keypad. If you make a mistake, use ∆ /MI or ∇ /MR keys to highlight mistake, then re-enter.
See Figure 172.
Updating... Please wait... and then returns to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
10.5.5 Memory recall setup mode P4.3
Use this program to select automatic or manual memory recall. See page 24 for more information on memory recall.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program MemRecall Setup: P4.3.
See Figure 173.
3. Press
Auto and Manual. The selection will flash. See Figure 174.
ENTER to confirm selection and return to the setup mode. Press ∆ /MI key to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
Data Log Setup :P4.2
Press ENT. to Select
Figure 170
Data Log:Auto Manual
Select & Press ENTER
Figure 171
Auto DataLog Period
Hr:1 Min:30 Sec:0
Figure 172
MemRecall Setup:P4.3
Press ENT. to Select
Figure 173
Memory Recall Mode
Auto Manual
Figure 174
52
Instruction Manual CyberScan pH 1100/ 2100
10.5.6 Communication setup mode P4.4
Use this program to set parameters of serial communication.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Tx/Rx Parameter: P4.4.
See Figure 175.
19200. The selection will flash. See Figure 176. Press ENTER to confirm selection and go to the next setup screen.
The selection will flash. See Figure 177. Press ENTER to confirm selection and go to the next setup screen.
Even. The selection will flash. See Figure 178.
ENTER to confirm selection and return to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
10.5.7 Data transfer setup mode P4.5
1. This program lets you select what data is sent to your printer or computer.
2. Enter meter setup mode as described in steps 1-4 in Section
10.5.1Entering meter setup mode (page 50).
3. Use /MI key to scroll to program Data Transfer: P4.5. See
Figure 179.
ENTER to enter program Data Transfer: P4.5. and Memory. .The selection will flash.
See Figure 180.
ENTER to confirm selection and return to the setup mode.
Press ∆ /MI key to continue to the next setup program, or press
∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
Tx/Rx Parameter:P4.4
Press ENT. to Select
Baud Rate
Figure 175
4800 9600 19200
Figure 176
Stop Bit: One Two
Figure 177
Parity Bit:
None Odd Even
Figure 178
Data Transfer :P4.5
Press ENT. to Select
Figure 179
Data send to printer
Current Memory
Figure 180
53
Instruction Manual CyberScan pH 1100/ 2100
10.5.8 Memory clear setup mode P4.6
Clears all pH, mV and Ion (pH2100 only) memory points stored in memory.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Mem Clear Setup: P4.6.
See Figure 181.
3. Press
The selection will flash. See Figure 182.
• If you select Yes, the screen flashes Memory being
Cleared, Please wait... See Figure 183. The meter then returns to the setup mode.
• If you select No, the meter returns to the setup mode.
Press ∆ /MI to continue to the next setup program, or press ∇ /MR to go to the previous program. Press EXIT to return to measurement mode.
10.5.9 Meter reset setup mode P4.7
Resets the meter to factory default settings:
• pH/mV/Ion* calibration data is erased
• Manual temperature compensation is set to 25.0°C
• pH/mV/Ion* measurement alarm is off
• pH/mV/Ion* calibration due alarm is off
• Stability indicator is turned on with low stability setting
• Data logging and memory recall has manual setting
• Backlight is on with 2.0 minute duration
• Memory is cleared
*Ion features on the CyberScan pH2100 meter only
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use /MI key to scroll to program Meter Reset: P4.7.
See Figure 184.
The selection will flash. See Figure 185.
ENTER to confirm selection.
• If you selected Yes, the screen flashes Meter being
Reset, Please wait...See Figure 186. The meter then returns to measurement mode.
• If you select No, the screen flashes Reset Aborted. See
Figure 187. The meter then returns to the measurement mode.
Mem Clear Setup :P4.6
Press ENT. to Select
Figure 181
Memory Clear:Yes No
Select & Press ENTER
Figure 182
Memory being Cleared
Please wait. . .
Figure 183
Meter Reset :P4.7
Press ENT. to Select
Figure 184
Meter Reset: Yes No
Select & Press ENTER
Figure 185
Meter being Reset
Please wait. . .
Figure 186
Reset Aborted
Figure 187
54
Instruction Manual CyberScan pH 1100/ 2100
11 ELECTRODE CARE
11.1 Electrode Activation
DO NOT touch or rub the glass bulb. If you follow the storage and maintenance procedure, you can use your electrode immediately. If the electrode responds sluggishly or drifts, the bulb may be dehydrated. Immerse the electrode in an ideal storage solution such as electrode storage solution or pH 4 buffer solution for 1-2 hours to hydrate it. See Storing
pH/ORP electrodes below.
If this procedure does not hydrate the electrode, reactivate or replace it.
Use 2 or 3 point calibration to test your electrode performance. If you do not get good readings, use a different pH electrode to confirm the meter is working properly. If the results are still not satisfactory, consult your distributor/dealer.
11.2 Electrode Maintenance
pH electrodes are susceptible to dirt, dehydration and contamination. Clean them regularly depending on the extent and condition of use.
11.3 Storing pH/ORP electrodes
For best results, always keep the bulb wet, preferably in pH storage solution. Other pH buffers or tap water are also acceptable. Do not store in de-ionized water. Eutech Instruments/ Oakton Instruments offers a complete range of electrode storage and cleaning solutions.
After Use
1. After each series of measurements, wash the electrode and reference junction in de-ionized water.
2. If using a refillable electrode, close the refilling hole by placing its rubber sleeve or stopper plug into its position.
3. Store the electrode as mentioned above.
Electrode Cleaning
Electrodes that are mechanically intact can often be restored to normal performance by one or a combination of the following options.
Some suggestions for: a. Salt deposits which are normal on all electrodes
Dissolve the deposit by immersing the electrode in tap water for 10 to 15 minutes. Thoroughly rinse with de-ionized water. b. General dirt and protein coatings
Soak the electrode for several hours in the general-purpose electrode cleaning solutions. Rinse in de-ionized or tap water.
Wash the electrode pH bulb in a little dish washing detergent and water. Rinse the electrode tip with de-ionized water. d. Clogged reference junction.
Heat a dilute 1% KCl solution or pH 4.01 buffer solution to 60-80°C. Place the sensing portion of the pH electrode into the heated KCl solution for approximately 30 minutes. Allow the electrode to cool while immersed in some unheated KCl solution or pH 4.01 buffer solution. Rinse with deionized water.
55
Instruction Manual CyberScan pH 1100/ 2100
12 RS 232 COMMUNICATION
This meter will send data to a printer and computer through the RS-232 port located on the back panel of the meter.
See Figure 193. This meter uses a 9 pin serial connection for RS-232 communication.
12.1 Using with printer
If you are going to use your meter to send data to a printer, you must first configure your meter to match the settings for your printer (consult printer manual for proper printer operation). See Meter Setup mode P4.4 Tx/Rx Parameter (page
53) to set baud rate, stop bit and parity for your meter.
12.1.1 Sending data to printer
You can either send current data or data stored in memory to your printer. See Meter Setup mode Program P4.5 Data
Transfer (page 53) to select which data is sent to your printer.
1. Connect your meter to printer (consult printer manual for cable connections). Use a 9 pin serial connection for pH 1100 and pH 2100).
2. To send data to your printer, press the PRINT key on the meter.
12.2 Using with computer
Use OAKTON
®
RS-232 cable 35615-09 to connect your meter to a computer. Refer to enclosed software disc for instructions on how to use software with your meter.
56
Instruction Manual CyberScan pH 1100/ 2100
13 ADDITIONAL INFORMATION
13.1 pH and Temperature
The pH electrode is affected by temperature changes. Automatic Temperature Compensation (ATC) compensates for these temperature changes. Some solutions show an increase while others a decrease in pH with the same temperature change. Record the solution temperature along with the pH value, or the measurement may be meaningless.
Temperature changes also affect the signal the pH electrode sends to the meter and causes a loss of accuracy for the reading. To limit the loss of accuracy during calibration, make the temperature of the pH buffer calibrating solutions and the sample solution the same.
13.2 pH Buffer Calibration Solution
You can use standard buffer solutions to calibrate a pH meter before you measure the pH of a sample. Calibration solutions serve as basis of comparison between measurements. The most common standard buffers are pH 4.01, pH 7.00 and pH 10.01.
For 1-point calibration, use a standard buffer of pH 7.00 or 6.86, or a standard or custom buffer whose pH value is close to that of the sample. Use 2-point calibration when you know the sample is acidic (low pH) or basic (high pH). For example: in acidic samples, use standard buffers such as pH 7.00 and pH 4.01; in basic samples, use standards such as pH 7.00 and pH 10.01.
Use multi-point calibration when the sample pH is completely unknown or for best accuracy over the entire pH range.
Contact your distributor for information on pH buffer and calibration solutions.
13.3 Standard pH Buffers
The following table shows the various pH values at different temperature of the solution during calibration. The table also illustrates why a calibration value may be different from the buffer value at 25°C.
Temperature (°C)
USA Buffer NIST Buffer
1.68 4.01 7.00 10.01 12.45 1.68 4.01 6.86 9.18 12.45
1.67 4.01 7.12 10.32 13.43 1.67 4.01 6.98 9.46 13.43
1.67 4.01 7.09 10.25 13.21 1.67 4.01 6.95 9.40 13.21
1.67 4.01 7.06 10.18 13.00 1.67 4.01 6.92 9.33 13.00
1.67 4.00 7.04 10.12 12.81 1.67 4.00 6.90 9.28 12.81
1.68 4.00 7.02 10.06 12.63 1.68 4.00 6.87 9.23 12.63
1.68 4.01 7.00 10.01 12.45 1.68 4.01 6.86 9.18 12.45
55
60
65
70
75
80
85
90
95
100
0
5
10
15
20
25
30
35
40
45
50
105
57
Instruction Manual CyberScan pH 1100/ 2100
14 SUMMARY OF DEFAULT SETTINGS / OPTIONS
Program
P1.0
P1.1
P1.2
P1.3
P1.4
P1.5
P1.6
P1.7
P3.0
P3.1
P3.2
P3.3
P3.4
P3.5
P3.6
P4.1
P4.2
P4.3
P4.4
P2.0
P2.1
P2.2
P2.3
P2.4
P4.6
P4.7
Setup Menus pH Buffer Setup pH Resolution Setup
Temperature Unit Setup pH Measurement Alarm pH Calibration Due Alarm
Stability Setup pH Cal Data pH Cal Reset mV Measurement Alarm mV Calibration Due Alarm
Stability Setup mV Cal Data mV Cal Reset
Ion Unit Setup
Ion Measurement Alarm
Ion Calibration Due Alarm
Stability Setup
Ion Mode Setup
Ion Cal Data
Ion Cal Reset
Backlight Setup
Data Log Setup
Memory Recall Setup
Tx/Rx Parameter
Memory Clear Setup
Meter Reset
Options pH SETUP Menu
USA, NIST, Bf1, Bf2, Custom
0.01, 0.001 o
C, o
F
On, Off
On, Off
On, Off
-
Yes, No
mV SETUP Menu
On, Off
On, Off
On, Off
-
Yes, No
Ion SETUP Menu ppm, ppt, mg/l, g/l, mmol/l, mol/l
On, Off
On, Off
On, Off
High, Low
-
Yes, No
Meter SETUP Menu
-
On, Off
Auto, Manual
Auto, Manual
Baud Rate: 4800, 9600, 19200, 38400
Stop Bit: 0ne, Two
Parity Bit: None, Odd, Even
Current, Memory
Yes, No
Yes, No
Factory Defaults
USA
0..001 o
C
Off
Off
On (Low)
-
No
Off
Off
On (Low)
-
No ppm
Off
Off
On (Low)
High
-
No
-
On
Manual
Manual
9600
One
Even
Current
No
No
58
CyberScan pH1100/ 2100
15 TROUBLESHOOTING & ERROR MESSAGES
TROUBLESHOOTING
Power ON but no display
Unstable Reading
Slow response
ERROR MESSAGES
LCD Display
Error:
EEPROM Read Error
Error:
EEPROM write Error
Error:
DTR is set too high
Error:
Communication error
Error:
Buffer Overflow
Error: pH Offset Error
Error: pH Slope Error
Error:
Ion Slope Error
(pH 2100 meter only)
Error: mV Offset too large
Error:
Out of Range
Error:
ATC offset too much
Error:
Too much offset@7pH
Error:
Slope out:75%-105%
Error: Pts too close
Or slope diff. sign
AC adapter not connected properly
Electrode submersion in sample is insufficient
Broken electrode
External noise causing instability in reading.
Dirty electrode
Dirty electrode
Solution
Connect AC adapter properly
Place electrode deeper into sample
Replace electrode
Remove or switch off interfering motor
Clean Electrode. Reactivate if necessary
Clean Electrode. Reactivate if necessary
Indicates
Memory write error
Memory checksum error
DTR line is disconnected
Frame/parity errors
Received buffer overflowing pH offset too much
Incorrect pH slope
Incorrect ion slope
Offset too much
Out of Range
Offset of temp. probe too much
Offset too much
Incorrect pH slope
Incorrect slope alignment
Cause
Unit is too old (> 10 years)
Hardware failure
Hardware fault pH offset should not exceed
±1 pH from the display pH slope is out of allowable limit
Average ion slope is not within 10mV-75mV/decade
Value input is not within
±150.0mV from the display
Value input exceeds the specified range
Temp. calibration offset is not within ±5 o C (or ±9 o F) pH offset should not exceed
±1 pH from the display
(custom pH buffer) pH slope is not within 75% to
105% (custom pH buffer)
Slope is not negative or cal.
Points are not ±1 pH apart
(custom pH buffer)
Solution
Turn meter off and return to authorized distributor if necessary
Turn meter off and return to authorized distributor if necessary
Turn meter off and return to authorized distributor if necessary
Turn meter off and return to authorized distributor if necessary
Turn meter off and return to authorized distributor if necessary
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Recalibrate meter
Re-enter value
Re-enter value
Re-enter value
Re-enter value
Re-enter value
59
CyberScan pH1100/ 2100
16 SPECIFICATIONS
pH
Range -1.000 to 14.000 pH
Resolution 0.001 / 0.01 pH (user-selectable)
Accuracy ± 0.002 / ±0.01 pH
Calibration 1 to 5 points (standard buffers); first point must be pH 7.000, 6.860 (offset)
Min. 2 points and Max. 5 points for custom entry pH Buffer Set Options 1 - USA: pH 1.68, 4.01, 7.00, 10.01, 12.45
2- NIST: pH 1.68, 4.01, 6.86, 9.18, 12.45
3- Bf1: pH 1.68, 4.01, 6.86, 10.01, 12.45
4- Bf2: pH 1.68, 4.01, 7.00, 9.18, 12.45
5 - Custom entry buffer calibration mV/ Relative mV
Range - 1850.0 to 1850.0 mV
Temperature
Accuracy ± 0.2 mV
Calibration mV offset up to ± 150 mV
Range -5.0 to 105.0 o C; 23.0 to 221.0 o F
C / o
F
Accuracy ± 0.3 o C / 0.5 o F
Calibration Manual Temperature compensation: adjust from -0.5 to 105.0°C/ 23.0 to 221.0°F
Automatic Temperature compensation (ATC) offset ±5.0°C/ ±9.0°F pH 2100 meter only
Ion
Range 0.001 to 19999
Resolution (High resolution mode:
± 500.0 mV)
0.001: from 0.001 to 0.999
0.01: from 1.00 to 9.99
0.1: from 10.0 to 99.9
1: from 100 to 19999
Resolution (Low resolution mode:
± 1850.0 mV)
0.001: from 0.001 to 0.099
0.01: from 0.10 to 0.99
0.1: from 1.0 to 9.9
1: from 10 to 19999
Accuracy ± 0.5% of full scale for monovalent ions
1% of full scale for divalent ions
Calibration 2 to 5 points, First point keyed in. the rest are selected using the ¿MI and ÀMR keys
Minimum Average Slope 10 mV/ decade
Maximum Average Slope 75 mV/ decade
Inputs BNC, Phono (ATC), Reference, Ground
Output RS232C
Recorder Output ± 1800 mV, 1 mV/mV unit
Power Requirements AC/DC 9V Adapter (110 VAC/220 VAC)
Display Custom Dual LCD (1 x 4 digit, 1 x 3.5 digit, 24 annunicators)
Dimensions (L x W x H) 230 x 180 x 63 mm (meter only); 395 x 260 x 90 mm (boxed)
Weight 1.4lbs/ 850 gm (unit); 1400 gm (boxed)
60
Instruction Manual CyberScan pH 1100/ 2100
17 ACCESSORIES
Consult your Authorized Distributors for these items and other range of specialized pH electrodes or Ion Selective
Electrodes.
Item Description Eutech Instruments
Order Code
Oakton Instruments
Order Code
EC-FG73504-01B 35801-00 Glass-body combination pH electrode (refillable) with 1m cable length and BNC connector
Epoxy-body combination double junction pH electrode with 1m cable length and BNC connector
Epoxy-body combination double junction ORP electrode with 1m cable length and BNC connector
Temperature probe (for ATC)
Meter to computer communication 1 m cable length (9-pin male to 25-pin female)
Meter to computer communication 1 m cable length (25-pin male to 25pin male)
Serial Impact Dot-matrix Micro-printer; paper-roll portable printer with 25pin female connector, with a roll of paper and 120VAC power adapter provided
Serial Impact Dot-matrix Micro-printer; paper-roll portable printer with 25pin female connector, with a roll of paper and 220VAC power adapter provided
110/120V AC/DC 9V 500mA power adapter (50/60 Hz) – 2 pin type
220/240V AC/DC 9V 500mA power adapter (50/60 Hz) – 2 pin type pH 4.01 calibration buffer (480 ml) (1 pint) pH 7.01 calibration buffer (480 ml) (1 pint) pH 10.01 calibration buffer (480 ml) (1 pint)
Protein Cleaning Solution for pH electrode (480 ml) (1 pint)
Storage Solution for pH electrode (480 ml) (1 pint) pH sachet assortment pack; 5 each of pH4.01, 7.00, 10.01 and deionized water (20 x 20 ml per box) pH 4.01 buffer sachets (20 x 20 ml per box) pH 7.00 buffer sachets (20 x 20 ml per box) pH 10.01 buffer sachets (20 x 20 ml per box)
Deionised water rinse sachets (20 x 20 ml per box)
EC-PH5-TEMB01P 35613-05
ECCA01M09F25 35615-09
ECCA01M09M25 35622-59
EC-MICROPRNTR01 35622-00
EC-MICROPRNTR02 35622-05
EC-110-ADA
EC-220-ADA
EC-BU-4BT
EC-BU-7BT
EC-BU-10BT
35621-50
35621-55
00654-00
00654-04
00654-08
EC-DPC-BT
EC-RE-005
00653-06
00653-04
EC-AST-PK 35653-04
EC-BU-4BS
EC-BU-7BS
EC-BU-10BS
EC-RIN-WT
35653-01
35653-02
35653-03
35653-00
61
Instruction Manual CyberScan pH 1100/ 2100
18 WARRANTY
Eutech Instruments/ Oakton Instruments supplies this bench meter with a 3-year warranty and 6-month warranty for against manufacturing defects from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warranty period, please return, freight prepaid, and correction will be made without charge.
Out of warranty items will be repaired on a charge basis.
Exclusions to the Warranty
The warranty shall not apply to defects resulting from:
• Improper or inadequate maintenance by customer;
• Unauthorized modification or misuse;
• Operation outside of the environmental specifications of the products.
Return of Items
Authorization must be obtained from your Authorized Distributor or Customer Service Dept. before returning items for any reason. When applying for authorization, please include data regarding reason the items are to be returned.
Packing the item for repair should be done using the original packaging or material, with information about any fault identified.
Shipment damage as a result of inadequate packaging is your or your distributor’s responsibility, whoever applicable.
Note: Eutech Instruments/ Oakton Instruments reserves the rights to make improvements in design, construction, and appearance of products without notice.
62
NOTES
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below:
Oakton Instruments
P.O Box 5136,
Vernon Hills, IL60061, USA
Fax: (1) 847-247-2984
www.4oakton.com
www.oaktoninstruments.com
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-16/24 Singapore 139949
Tel: (65) 6778 6876
Fax: (65) 6773 0836
E-mail:
[email protected]
Web-site: www.eutechinst.com
Distributed by:
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project