Edge Ultrasound System Service manual

Add to My manuals
92 Pages

advertisement

Edge Ultrasound System Service manual | Manualzz

Service Manual

i

ii

S onoSite, Inc.

21919 30th Drive SE

Bothell, WA 98021-3904

USA

Telephone: 1-888-482-9449 or 1-425-951-1200

Fax: 1-425-951-1201

SonoSite Ltd

Alexander House

40A Wilbury Way

Hitchin

Herts SG4 OAP

UK

T: +44-1462-444800

F: +44-1462-444801

Caution: Federal (United States) law restricts this device to sale by or on the order of a physician.

Edge, SiteLink, SonoCalc, SonoHD2, SonoMB, SonoMBe, SonoSite, and the SonoSite logo are registered (in some jurisdictions) and unregistered trademarks owned by SonoSite, Inc.

DICOM is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications of medical information.

Non-SonoSite product names may be trademarks or registered trademarks of their respective owners.

The SonoSite ultrasound system(s) referenced in this document may be covered by one or more of the following U.S. patents:

55722412, 5817024, 5893363, 6135961, 6203498, 6364839, 6371918, 6383139, 6416475, 6447451, 6471651, 6569101, 6648826, 6575908, 6604630,

6817982, 6835177, 6962566, 7169108, 7449640, 7534211, 7549961, 7588541, 7591786, 7604596, 7643040, 7686766, 7694814, 7727153, 7740586,

7804970, 7809400, 7819807, 7841575, 7849250, 7867168, 7883276, 7978461, 7996688, 8052606, 8066642, 8088071, 8090065, 8128050, 8137278,

8147408, 8164011, D456509, D461895, D509900, D538432, D544962, D558351, D559390, D591423, D592750, D592760, D625014, D625015, D657361, and by the following counterpart foreign patents: AU727381, AU730822, CA2372152, CA2372158, CA2373065, CA 2371711, CN ZL 97113678.5, CN ZL

98106133.8, CN ZL 98108973.9, CN ZL 200830007734.8, DE60021552.0, DE60029777.2, DE60034670.6, DE69730563.5, DE69830539.6, DE69831698.3,

DE60200423816.3-08, DE 1589878, DE 1791086, ES2229318, ES2251057, ES2337140, FR0815793, FR0875203, FR0881492, FR1175713, FR1180970,

FR1589878, GB0875203, GB0881492, GB1175713, GB1180970, GB1180971, GB1589878, GB 1791086, IT0815793, IT0881492, IT1175713, JP 4696150,

KR528102, KR532359, NO326202, NO326814, NZ542968, RCD000897368-0001. Patents pending.

P15644-02 06/2012

Copyright 2012 by SonoSite, Inc.

All rights reserved.

iii

iv

Table of Contents

Introduction ...................................................................................... 1

Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Contact Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Conventions, symbols, and terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Labeling symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Specifications.................................................................................... 5

Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Environmental limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Battery specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Compatible accessories and peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Safety................................................................................................. 9

Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Electrical safety classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Equipment safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Battery safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Clinical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Hazardous materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Electrostatic discharge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Separation distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Guidance and manufacturer’s declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Electrical safety standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

EMC standards classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Acoustic standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Biocompatibility standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Airborne equipment standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

DICOM standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

HIPAA standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

System Overview .............................................................................. 21

About the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Description of Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Additional System Feature Performances. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Front End Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

PW Doppler Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

CW Doppler Processing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Back End Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Control Subsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Power Supply and Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

ECG Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

IMT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

v

vi

Troubleshooting............................................................................... 35

System and Subsystem Diagnosis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

System Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Failure (Assert) Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

Verifying a System Assert Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Replacement Procedures ................................................................. 39

Display Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Display Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Display Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Test the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Control Panel Subassembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Control Panel Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Control Panel Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Main System Disassembly for Repair and/or Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

System Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Speaker Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Required Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Power Supply PCBA Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Required Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

SD Card Daughter-card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

USB Extension PCB Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Required Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Main PCBA Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Middle Machined Enclosure Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Bottom Enclosure Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Maintenance ..................................................................................... 53

Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Cleaning and disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Cleaning and disinfecting the ultrasound system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Cleaning and disinfecting transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

Cleaning and disinfecting the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

Cleaning the footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

Cleaning and disinfecting ECG cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Performance Testing ........................................................................ 59

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Recommend Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Setting Up Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Basic Operational Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

2D Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

2D Performance / Image Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Axial Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Lateral Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Penetration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Additional Performance Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Color Doppler (Color) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Color Power Doppler (CPD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

M Mode Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Tissue Harmonic Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Pulsed Wave (PW) Doppler Imaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Continuous Wave (CW) Doppler Imaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Image Quality Verification Test/Livescan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Battery Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Video Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Replacement Parts............................................................................ 67

Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Main PCBA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Miscellaneous Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

Transducer Nest Frame Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Ordering Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Service Event Reporting................................................................... 75

Service Event Report Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

Service Event Report Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

Returning Products to SonoSite. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

Shipping Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

vii

viii

Chapter 1: Introduction

Before servicing the Edge Ultrasound System, please read this manual.

The ultrasound system has multiple configurations and feature sets. All are described in this service manual but not every option may apply to your system. System features depend on your system configuration, transducer, and exam type.

Refer to the Edge Ultrasound System User Guide for additional information regarding safety, system controls, operation, capabilities, and specifications.

This chapter also defines labeling symbols, specifications, and standards.

Audience

The intended audience of this manual is properly trained field and in-house service personnel.

Contact Information

Questions and comments are encouraged. SonoSite is interested in your feedback regarding the service manual.

If you encounter difficulty with the system, use the information in this manual to help correct the problem. If the problem is not covered here, contact SonoSite Technical Support as follows:

Technical Support (USA, Canada)

Technical Support fax:

Technical Support e-mail:

SonoSite website:

International Technical Support:

European Service Center

Japan Service Center

1-877-657-8118

1-425-951-6700 [email protected]

www.sonosite.com (Select Resources > Support & Service)

Contact your local representative or call (USA) +425-951-1330

+44-(0)1462-444-800 e-mail: [email protected]

+81-3-5304-5337

Conventions, symbols, and terms

The user guide follows these conventions:

• A WARNING describes precautions necessary to prevent injury or loss of life.

• A Caution describes precautions necessary to protect the products.

• Numbered steps in procedures must be performed in order.

• Items in bulleted lists do not require performance in sequence.

Chapter 1: Introduction 1

Labeling symbols

The following symbols are used on the products, packaging, and containers.

Table 1: Labeling Symbols

Symbol Definition

Alternating Current (AC)

Class 1 device indicating manufacturer’s declaration of conformance with Annex

VII of 93/42/EEC

Class 1 device requiring verification by the Notified Body of sterilization or measurement features, or to a Class IIa, IIb, or III device requiring verification or auditing by the Notified Body to applicable Annex(es) of 93/42/EEC

Attention, see the user guide

Follow instructions for use.

LOT

REF

Device complies with relevant Australian regulations for electronic devices.

Batch code, date code, or lot code type of control number

Biological risk

Device complies with relevant Brazilian regulations for electro-medical devices.

Canadian Standards Association. The “C” and “US” indicators next to this mark signify that the product has been evaluated to the applicable CSA and ANSI/UL

Standards, for use in Canada and the US, respectively.

Catalog number

Collect separately from other household waste (see European Commission

Directive 93/86/EEC). Refer to local regulations for disposal.

Corrugated recycle

Dangerous voltage

Date of manufacture

2 Chapter 1: Introduction

Table 1: Labeling Symbols (Continued)

Symbol Definition

Manufacturer

Direct Current (DC)

Do not get wet.

Do not stack over 2 high.

Do not stack over 5 high.

Do not stack over 10 high.

Electrostatic sensitive devices

Device complies with relevant FCC regulations for electronic devices.

Fragile

GEL Gel

Sterilized using irradiation

STERILE R

STERILE EO

Sterilized using ethylene oxide

Hot

Device emits a static (DC) magnetic field.

Non-ionizing radiation

Paper recycle

Chapter 1: Introduction 3

Table 1: Labeling Symbols (Continued)

Symbol

SN

Definition

Serial number type of control number

Temperature limitation

Atmospheric pressure limitation

Humidity limitation

Submersible. Protected against the effects of temporary immersion.

Water-Tight Equipment. Protected against the effects of extended immersion.

Handle transducer with care.

Follow manufacturer’s instructions for disinfecting time.

Disinfect transducer.

Type BF patient applied part

(B = body, F = floating applied part)

Defibrillator proof type CF patient applied part

Underwriter’s Laboratories labeling

Pollution Control Logo. (Applies to all parts/products listed in the China RoHS disclosure table. May not appear on the exterior of some parts/products because of space limitations.)

WARNING:

Connect Only

Accessories and

Peripherals

Recommended by

SonoSite

China Compulsory Certificate mark (“CCC Mark”). A compulsory safety mark for compliance to Chinese national standards for many products sold in the People’s

Republic of China.

WARNING: Connect Only

Accessories and Peripherals

Recommended by SonoSite

4 Chapter 1: Introduction

Chapter 2: Specifications

This chapter contains information regarding system specifications and accessory compatibility. The information applies to the ultrasound system, transducers, accessories, and peripherals.

Specifications

Dimensions

System

Length: 13 in. (33 cm)

Width: 12.4 in. (31.5 cm)

Height: 2.5 in. (6.3 cm)

Display

Length: 9.7 in. (24.6 cm)

Height: 7.3 in. (18.5 cm)

Diagonal: 12.1 in. (30.7 cm)

Environmental limits

Note: The temperature, pressure, and humidity limits apply only to the ultrasound system, transducers, and battery.

Operating (system, battery, and transducer)

10–40°C (50–104°F), 15–95% R.H.

700 to 1060hPa (0.7 to 1.05 ATM)

Mode of Operation:

Continuous 35°C or below

Non-Continuous above 35°C (30 minutes on /30 minutes off)

Shipping and storage (system and transducer)

-35–65°C (-31–149°F), 15–95% R.H.

500 to 1060hPa (0.5 to 1.05 ATM)

Shipping and storage (battery)

-20–60°C (-4–140°F), 15–95% R.H. ( For storage longer than 30 days, store at or below room temperature.)

500 to 1060hPa (0.5 to 1.05 ATM)

Electrical specifications

Power Supply Input: 100-240 VAC, 50/60 Hz, 2.0 A Max @ 100 VAC

Power Supply Output #1: 15 VDC, 5.0 A Max

Power Supply Output #2: 12 VDC, 2.3 A Max

Combined output not exceeding 75 watts.

Chapter 2: Specifications 5

Battery specifications

The battery is comprised of six lithium-ion cells plus electronics, a temperature sensor, and battery contacts.

Run time is up to two hours, depending on imaging mode and display brightness. This chapter contains electrical, and clinical safety information required by regulatory agencies. The information applies to the ultrasound system, transducer, accessories, and peripherals.

Compatible accessories and peripherals

SonoSite has tested the Edge ultrasound system with the following accessories and peripherals and has demonstrated compliance to the requirements of IEC60601-1-2:2007.

You may use these SonoSite accessories and third-party peripherals with the Edge ultrasound system.

WARNING:

WARNING:

Use of the accessories with medical systems other than the Edge ultrasound system may result in increased emissions or decreased immunity of the medical system.

Use of accessories other than those specified may result in increased emissions or decreased immunity of the ultrasound system.

Accessories and peripherals compatible with Edge ultrasound system

Description

C8 transducer

C11x transducer

C60x transducer

D2x transducer

HFL38x transducer

HFL50x transducer

ICTx transducer

L25x transducer

L38xi transducer

L52x transducer (Vet)

P10x transducer

P21x Transducer

SLAx transducer

TEEx Transducer

Bar code scanner

Battery for PowerPack

Battery Pack

PowerPack

Black & white printer

Part Number

P12742

V00033

P07696

P07698

P07699

P05183

P14166

P13123

P08010

P07678

P07680

P05165

P07682

P07693

P07690

P07691

P15051

P13122

P13745

Maximum Cable Length

6.0 ft/1.8 m

6.5 ft/2.0 m

6.0 ft/1.8 m

6.0 ft/1.8 m

6.0 ft/1.8 m

6.0 ft/1.8 m

6.0 ft/1.8 m

8.0 ft/2.4 m

6.0 ft/1.8 m

7.9 ft/2.4 m

6.5 ft/2.0 m

6.5 ft/2.0 m

8.0 ft/2.4 m

7.5 ft/2.3 m

4.8 ft/1.5 m

6 Chapter 2: Specifications

Accessories and peripherals compatible with Edge ultrasound system (Continued)

3.3 ft/1 m Black & white printer power cable

Black & white printer USB cable

Color printer

Color printer power cable

Color printer video cable

ECG lead wires

ECG module

Edge Dock

Edge Stand

Footswitch

P13983

P14202

P08501

P15078

P15800

P14689

Petite mouse

Power cord (system)

P14451

P00848 (USA)

Power Supply/Battery Charger P09823

PowerPark P12822

Triple Transducer Connect

USB wireless adapter

P15922

P12046

10.8 ft/3.3 m

3.3 ft/1 m

6.0 ft/ 1.8 m

24 in/ 0.6 m

5.8 ft/1.8 m

9.8 ft/3.0 m

6 ft /1.8 m

10 ft/3 m

6.8 ft/ 2 m

Chapter 2: Specifications 7

8 Chapter 2: Specifications

Chapter 3: Safety

This chapter contains electrical and clinical safety information required by regulatory agencies. The information applies to the ultrasound system, transducers, accessories, and peripherals.

Electrical safety

This system meets EN60601-1, Class I/internally-powered equipment requirements and Type BF and Type CF isolated patient-applied parts safety requirements.

This system complies with the applicable medical equipment requirements published in the Canadian Standards

Association (CSA), European Norm Harmonized Standards, and Underwriters Laboratories (UL) safety standards. See

“Standards” on page 19 .

For maximum safety observe the following warnings and cautions.

WARNING: To avoid the risk of injury, do not operate the system in the presence of flammable gasses or anesthetics. Explosion can result.

WARNING: To avoid the risk of electrical shock or injury, do not open the system enclosures. All internal adjustments and replacements, except battery replacement, must be made by a qualified technician.

WARNING: To avoid the risk of electrical shock:

• This equipment must be connected only to a supply mains with protective earth.

• Use only properly grounded equipment. Shock hazards exist if the power supply is not properly grounded. Grounding reliability can be achieved only when equipment is connected to a receptacle marked “Hospital Only” or “Hospital Grade” or equivalent. The grounding wire must not be removed or defeated.

• When using the system in an environment where the integrity of the protective earth conductor arrangement is in doubt, operate the system on battery power only and disconnect the power supply.

• Do not let the bar code scanner or external mouse touch the patient.

• Do not touch any of the following:

• The power supply and the patient at the same time

• The ungrounded signal input/output connectors on the back of the ultrasound system

• The system battery contacts (inside the battery compartment)

• The system transducer connector when the transducer or Triple Transducer

Connect (TTC) is disconnected

• The system transducer connector on the TTC if no transducers are connected

• Do not connect the system power supply or docking system to a multiple portable socket outlet (MPSO) or extension cord.

• Before using the transducer, inspect the transducer face, housing, and cable. Do not use the transducer if the transducer or cable is damaged.

• Always disconnect the power supply from the system before cleaning the system.

• Do not use any transducer that has been immersed beyond the specified cleaning or disinfection level. See

Chapter 7, “Maintenance”

• Use only accessories and peripherals recommended by SonoSite, including the power supply. Connection of accessories and peripherals not recommended by SonoSite could result in electrical shock. Contact SonoSite or your local representative for a list of accessories and peripherals available from or recommended by SonoSite.

Chapter 3: Safety 9

WARNING: To avoid the risk of electrical shock and fire hazard:

• Inspect the power supply, AC power cords, cables, and plugs on a regular basis.

Ensure that they are not damaged.

• The power cord set that connects the power supply of the ultrasound system or the stand to mains power must only be used with the power supply or docking system, and cannot be used to connect other devices to mains power.

WARNING: To prevent injury to the operator/bystander, the transducer must be removed from patient contact before the application of a high-voltage defibrillation pulse.

WARNING: To avoid possible electrical shock or electromagnetic interference, verify proper operation and compliance with relevant safety standards for all equipment before clinical use. Connecting additional equipment to the ultrasound system constitutes configuring a medical system. SonoSite recommends verifying that the system, all combinations of equipment, and accessories connected to the ultrasound system comply with JACHO installation requirements and/or safety standards such as

AAMI-ES1, NFPA 99 OR IEC Standard 60601-1-1 and electromagnetic compatibility standard IEC 60601-1-2 (Electromagnetic compatibility), and are certified according to

IEC Standard 60950 (Information Technology Equipment (ITE)).

Caution:

Caution:

Do not use the system if an error message appears on the image display: note the error code; call SonoSite or your local representative; turn off the system by pressing and holding the power key until the system powers down.

To avoid increasing the system and transducer connector temperature, do not block the airflow to the ventilation holes on the side of the system.

10 Chapter 3: Safety

Electrical safety classification

Class I equipment The ultrasound system is classified as Class I equipment when powered from the external power supply or mounted on the stand because the external power supply is a Class 1 protectively earthed power supply.

The stand has no protective earth. Ground bond testing is not applicable to the ultrasound system or the stand.

Note: AC powered peripherals that may be used with the system are Class

I and are individually protectively earthed. Ground bond testing may be conducted on each AC powered peripheral.

Internally powered equipment Ultrasound system not connected to the power supply (battery only)

Type BF applied parts

Type CF applied parts

IPX-7 (watertight equipment)

IPX-8 (watertight equipment)

Non AP/APG

Ultrasound transducers

ECG module/ECG leads

Ultrasound transducers

Footswitch

Ultrasound system power supply, docking system, and peripherals. Equipment is not suitable for use in the presence of flammable anaesthetics.

Equipment safety

To protect your ultrasound system, transducers, and accessories, follow these precautions.

Caution:

Caution:

Caution:

Caution:

Caution:

Caution:

Excessive bending or twisting of cables can cause a failure or intermittent operation.

Improper cleaning or disinfecting of any part of the system can cause permanent damage. For cleaning and disinfecting instructions, see

Chapter 7, “Maintenance.”

Do not submerge the transducer connector in solution. The cable is not liquid-tight beyond the transducer connector/cable interface.

Do not use solvents such as thinner or benzene, or abrasive cleaners on any part of the system.

Remove the battery from the system if the system is not likely to be used for some time.

Do not spill liquid on the system.

Chapter 3: Safety 11

Battery safety

To prevent the battery from bursting, igniting, or emitting fumes and causing personal injury or equipment damage, observe the following precautions.

WARNING: The battery has a safety device. Do not disassemble or alter the battery.

WARNING: Charge the batteries only when the ambient temperature is between 0° and 40°C (32° and 104°F).

WARNING: Do not short-circuit the battery by directly connecting the positive and negative terminals with metal objects.

WARNING: Do not touch battery contacts.

WARNING: Do not heat the battery or discard it in a fire.

WARNING: Do not expose the battery to temperatures over 60°C (140°F). Keep it away from fire and other heat sources.

WARNING: Do not charge the battery near a heat source, such as a fire or heater.

WARNING: Do not leave the battery in direct sunlight.

WARNING: Do not pierce the battery with a sharp object, hit it, or step on it.

WARNING: Do not use a damaged battery.

WARNING: Do not solder a battery.

WARNING: The polarity of the battery terminals are fixed and cannot be switched or reversed. Do not force the battery into the system.

WARNING: Do not connect the battery to an electrical power outlet.

WARNING: Do not continue recharging the battery if it does not recharge after two successive six hour charging cycles.

WARNING: Do not ship a damaged battery without instructions from SonoSite Technical Support.

(See “Technical Support (USA, Canada)” on page 1

.)

WARNING: If the battery leaks or emits an odor, remove it from all possible flammable sources.

WARNING: Periodically check to make sure that the battery charges fully. If the battery fails to charge fully, replace it.

Caution: To avoid the battery becoming damaged and causing equipment damage, observe the following precautions:

• Do not immerse the battery in water or allow it to get wet.

• Do not put the battery into a microwave oven or pressurized container.

• If the battery emits an odor or heat, is deformed or discolored, or in any way appears abnormal during use, recharging or storage, immediately remove it and stop using it.

If you have any questions about the battery, consult SonoSite or your local representative.

• Store the battery between -20°C (-4°F) and 60°C (140°F).

• Use only SonoSite batteries.

• Do not use or charge the battery with non-SonoSite equipment. Only charge the battery with the system.

12 Chapter 3: Safety

Clinical safety

WARNING: Non-medical (commercial) grade peripheral monitors have not been verified or validated by SonoSite as being suitable for diagnosis.

WARNING: To avoid the risk of a burn hazard, do not use the transducer with high frequency surgical equipment. Such a hazard may occur in the event of a defect in the high frequency surgical neutral electrode connection.

WARNING: Do not use the system if it exhibits erratic or inconsistent behavior. Discontinuities in the scanning sequence are indicative of a hardware failure that must be corrected before use.

WARNING: Some transducer sheaths contain natural rubber latex and talc, which can cause allergic reactions in some individuals. Refer to 21 CFR 801.437, User labeling for devices that contain natural rubber.

WARNING: Perform ultrasound procedures prudently. Use the ALARA (as low as reasonably achievable) principle and follow the prudent use information concerning MI and TI.

WARNING: SonoSite does not currently recommend a specific brand of acoustic standoff. If an acoustic standoff is used, it must have a minimum attentuation of .3dB/cm/MHz.

WARNING: Some SonoSite transducers are approved for intraoperative applications if a market-cleared sheath is used.

WARNING: To avoid injury or reduce the risk of infection to the patient, observe the following:

• Follow Universal Precautions when inserting and maintaining a medical device for interventional and intraoperative procedures.

• Appropriate training in interventional and intraoperative procedures as dictated by current relevant medical practices as well as in proper operation of the ultrasound system and transducer is required. During vascular access, the potential exists for serious complications including without limitation the following: pneumothorax, arterial puncture, guidewire misplacement, and risks normally associated with local or general anesthesia, surgery, and post-operative recovery.

WARNING: To avoid device damage or patient injury, do not use the P10x, P17x, or P21x needle guide bracket on patients with pacemakers or medical electronic implants. The needle guide bracket for the P10x, P17x, and P21x transducers contains a magnet that is used to ensure the bracket is correctly oriented on the transducer. The magnetic field in direct proximity to the pacemaker or medical electronic implant may have an adverse effect.

Hazardous materials

WARNING: Products and accessories may contain hazardous materials. Ensure that products and accessories are disposed of in an environmentally responsible manner and meet federal and local regulations for disposing hazardous materials.

WARNING: The liquid crystal display (LCD) contains mercury. Dispose of the LCD properly in accordance with local regulations.

Chapter 3: Safety 13

Electromagnetic compatibility

The ultrasound system has been tested and found to comply with the electromagnetic compatibility (EMC) limits for medical devices to IEC 60601-1-2:2001. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.

WARNING: The Edge ultrasound system should not be used adjacent to or stacked with other equipment. If such use occurs, verify that the Edge ultrasound system operates normally in that configuration.

Caution: Medical electrical equipment requires special precautions regarding EMC and must be installed and operated according to these instructions. Portable and mobile RF communications equipment can affect the ultrasound system. Electromagnetic interference (EMI) from other equipment or interference sources could result in performance disruption of the ultrasound system. Evidence of disruption may include image degradation or distortion, erratic readings, equipment ceasing to operate, or other incorrect functioning. If this occurs, survey the site to determine the source of disruption, and take the following actions to eliminate the source(s).

Caution:

Turn equipment in the vicinity off and on to isolate disruptive equipment.

Relocate or re-orient interfering equipment.

Increase distance between interfering equipment and your ultrasound system.

Manage use of frequencies close to ultrasound system frequencies.

Remove devices that are highly susceptible to EMI.

Lower power from internal sources within facility control (such as paging systems).

Label devices susceptible to EMI.

Educate clinical staff to recognize potential EMI-related problems.

Eliminate or reduce EMI with technical solutions (such as shielding).

Restrict use of personal communicators (cell phones, computers) in areas with devices susceptible to EMI.

Share relevant EMI information with others, particularly when evaluating new equipment purchases which may generate EMI.

Purchase medical devices that comply with IEC 60601-1-2 EMC Standards.

To avoid the risk of increased electromagnetic emissions or decreased immunity, use only accessories and peripherals recommended by SonoSite. Connection of accessories and peripherals not recommended by SonoSite to the ultrasound system may result in malfunction of the ultrasound system or other medical electrical devices in the area.

Contact SonoSite or your local representative for a list of accessories and peripherals available from or recommended by SonoSite. See the SonoSite accessories user guide.

Electrostatic discharge

Caution: Electrostatic discharge (ESD), or static shock, is a naturally occurring phenomenon.

ESD is common in conditions of low humidity, which can be caused by heating or air conditioning. ESD is a discharge of the electrical energy from a charged body to a lesser or non-charged body. The degree of discharge can be significant enough to cause damage to a transducer or an ultrasound system. The following precautions can help reduce ESD: anti-static spray on carpets, anti-static spray on linoleum, and anti-static mats.

14 Chapter 3: Safety

Separation distance

Recommended separation distances between portable and mobile RF communications equipment and the

Edge ultrasound system

The Edge ultrasound system is intended for use in an electromagnetic environment in which radiated radio frequency (RF) disturbances are controlled. The customer or the user of the Edge ultrasound system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Edge ultrasound system as recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter m

Rated maximum output power of transmitter

Watts

150 kHz to 80 MHz d=1.2 P

80 MHz to 800 MHz d=1.2

P

800 MHz to 2.5 GHz d=2.3 P

0.01

0.1

1

10

0.12

0.38

1.2

3.8

0.12

0.38

1.2

3.8

0.23

0.73

2.3

7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Chapter 3: Safety 15

Guidance and manufacturer’s declaration

WARNING: Other equipment, even equipment that complies with CISPR emission requirements, can interfere with the Edge ultrasound system.

The Edge ultrasound system contains an IEEE 802.11 transmitter that utilizes the ISM frequency band from 2.412 to

2.4835 GHz and implements two methods of transmission:

• IEEE 802.11b with Complementary Code Keying (CCK), Differential Quaternary Phase Shift Keying (DQPSK), and Differential Binary Phase Shift Keying (DBPSK) at 16 dB

• IEEE 802.11g with Orthogonal Frequency Division Multiplexing (OFDM) at 13 dBm

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

The Edge ultrasound system is intended for use in the electromagnetic environment specified below.

The customer or the user of the Edge ultrasound system should assure that it is used in such an environment.

Emissions Test

RF emissions

ClSPR 11

RF emissions

ClSPR 11

Harmonic emissions

IEC 61000-3-2

Voltage fluctuations/flicker emissions

IEC 61000-3-3

Compliance Electromagnetic Environment

Group 1

Class A

Class A

The Edge ultrasound system uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The Edge ultrasound system is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network which supplies buildings used for domestic purposes.±

Complies

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The Edge ultrasound system is intended for use in the electromagnetic environment specified below.

The customer or the user of the Edge ultrasound system should assure that it is used in such an environment.

Immunity Test IEC 60601 Test Level Compliance Level

Electromagnetic

Environment

Electrostatic

Discharge

(ESD)

IEC 61000-4-2

Electrical fast

Transient burst

IEC 61000-4-4

±6.0KV contact

±8.0KV air

±2KV for power supply lines

±1KV for input/output lines

±6.0KV contact

±8.0KV air

±2KV for power supply lines

±1KV for input/output lines

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least

30%.

Mains power quality should be that of a typical commercial or hospital environment.

16 Chapter 3: Safety

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (Continued)

The Edge ultrasound system is intended for use in the electromagnetic environment specified below.

The customer or the user of the Edge ultrasound system should assure that it is used in such an environment.

Immunity Test IEC 60601 Test Level Compliance Level

Electromagnetic

Environment

Surge

IEC 61000-4-5

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC 61000-4-11

Power

Frequency

Magnetic Field

IEC 61000-4-8

Conducted RF

IEC 61000-4-6

±1KV line(s) to line(s)

±2KV line(s) to earth

>5% U

T

(>95% dip in U

T

) for

0.5 cycle

40% U

T

(60% dip in U

T

) for 5 cycles

70% U

T

(30% dip in U

T

) for 25 cycles

>5% U

T

(>95% dip in U

T

) for

5s

3 A/m

3 Vrms

150 kHz to 80 MHz

±1KV line(s) to line(s)

±2KV line(s) to earth

>5% U

T

(>95% dip in U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

) for

5 cycles

70% U

T

(30% dip in U

T

) for

25 cycles

>5% U

T

(>95% dip in U

T

) for 5s

3 A/m

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment. If the user of the Edge ultrasound system requires continued operation during power mains interruptions, it is recommended that the Edge ultrasound system be powered from an uninterruptible power supply or a battery.

3 Vrms

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Portable and mobile RF communications equipment should be used no closer to any part of the Edge ultrasound system including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended Separation

Distance

d = 1.2

P

Chapter 3: Safety 17

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (Continued)

The Edge ultrasound system is intended for use in the electromagnetic environment specified below.

The customer or the user of the Edge ultrasound system should assure that it is used in such an environment.

Immunity Test IEC 60601 Test Level Compliance Level

Electromagnetic

Environment

Radiated RF

IEC 61000-4-3

3 Vim

80 MHz to 2.5 GHz

3 V/m

Radiated RF

IEC 61000-4-3

(continued)

d = 1.2

P

80 MHz to 800 MHz

d = 2.3

P

800 MHz to 2,5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic Site survey a , should be less than the compliance level in each frequency range b .

Interference may occur in the vicinity of equipment marked with the following symbol:

Note: U

T

is the AC mains voltage prior to application of the test level.

At 80 MHz and 800 MHz, the higher frequency range applies.

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a. Field strengths from fixed transmitters such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the SonoSite ultrasound system is used exceeds the applicable RF compliance level above, the SonoSite ultrasound system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the SonoSite ultrasound system.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

FCC Caution: Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:

• This device may not cause harmful interference.

• This device must accept any interference received, including interference that may cause undesired operation.

18 Chapter 3: Safety

Immunity testing requirements

The Edge ultrasound system complies with the essential performance requirements specified in IEC 60601-1-2 and

IEC 60601-2-37. Results of immunity testing show that the Edge ultrasound system meets these requirements and is free from the following:

• Noise on a waveform or artifacts or distortion in an image or error of a displayed numerical value that cannot be attributed to a physiological effect and that may alter the diagnosis

• Display of incorrect numerical values associated with the diagnosis to be performed

• Display of incorrect safety related indications

• Production of unintended or excessive ultrasound output

• Production of unintended or excessive transducer assembly surface temperature

• Production of unintended or uncontrolled motion of transducer assemblies intended for intra-corporeal use

Standards

Electrical safety standards

AAMI/ANSI ES 60601-1:2005, Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and

Essential Performance.

CAN/CSA C22.2, No. 60601-1, Canadian Standards Association, Medical Electrical Equipment—Part 1. General

Requirements for Safety.

CAN/CSA C22.2, No. 60601-1:08, Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and

Essential Performance.

IEC 60601-1:1988, International Electrotechnical Commission, Medical Electrical Equipment—Part 1. General

Requirements for Safety.

IEC 60601-1:2005, Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential

Performance.

IEC 60601-1-1:2000, Medical Electrical Equipment—Part 1-1. General Requirements for Safety-Section 1-1.

Collateral Standard. Safety Requirements for Medical Electrical Systems.

IEC 60601-2-37:2001, International Electrotechnical Commission, Particular Requirements for the Safety of

Ultrasonic Medical Diagnostic and Monitoring Equipment.

IEC 60601-2-37:2007, Medical Electrical Equipment—Part 2-37: Particular Requirements for the Basic Safety and

Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment.

IEC 61157, International Electrotechnical Commission, Standard Means for the Reporting of the Acoustic Output of

Medical Diagnostic Ultrasonic Equipment.

EMC standards classification

CISPR 11, International Electrotechnical Commission, International Special Committee on Radio Interference.

Industrial, Scientific, and Medical (ISM) Equipment—Radio-Frequency Disturbance Characteristics—Limits and

Methods of Measurement. Classification for the ultrasound system, docking system, accessories, and peripherals when configured together: Group 1, Class A.

IEC 60601-1-2:2007, Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral

Standard: Electromagnetic Compatibility—Requirements and Tests.

Acoustic standards

NEMA UD 2-2004, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.

NEMA UD 3-2004, Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on

Diagnostic Ultrasound Equipment, American Institute of Ultrasound in Medicine.

Chapter 3: Safety 19

Biocompatibility standards

AAMI/ANSI/ISO 10993-1, Biological evaluation of medical devices—Part 1: Evaluation and testing (2009).

AAMI/ANSI/ISO 10993-5, Biological evaluation of medical devices—Part 5: Tests for In Vitro cytotoxicity (2009).

AAMI/ANSI/ISO 10993-10, Biological evaluation of medical devices—Part 10: Tests for irritation and delayed-type hypersensitivity (2002).

AAMI/ANSI/ISO 10993-11, Biological evaluation of medical devices—Part 11: Tests for systemic toxicity (2006).

AAMI/ANSI/ISO 10993-12, Biological evaluation of medical devices—Part 12: Sample preparation and reference materials (2007).

Airborne equipment standards

RTCA DO-160E, Radio Technical Commission for Aeronautics, Environmental Conditions and Test Procedures for

Airborne Equipment, Section 21.0 Emission of Radio Frequency Energy, Category B. 118.

DICOM standard

NEMA PS 3.15, Digital Imaging and Communications in Medicine (DICOM)—Part 15: Security and System

Management Profiles.

HIPAA standard

Health Insurance and Portability and Accountability Act, Pub.L. No. 104-191.

45 CFR 160, General Administrative Requirements.

45 CFR 164, Security and Privacy.

20 Chapter 3: Safety

Chapter 4: System Overview

About the System

The SonoSite Edge high-resolution ultrasound system is a portable, full featured, general purpose, software controlled, diagnostic ultrasound system using all digital architecture. The system is used to acquire and display high-resolution, real-time ultrasound data in 2D, M Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW)

Doppler, Color Power Doppler (CPD), and color Doppler (Color) or in a combination of these modes.

The system has an electrocardiography (ECG) display feature and supports a 3-lead ECG cable assembly to collect data for M Mode and Doppler measurements. The system provides measurement capabilities for anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The system has a PW and CW Doppler audio output feature, cine review, image zoom, labeling, biopsy, measurements and calculations, image storage and review, printing, and recording capabilities.

The system includes the optional ability to measure the intima-media thickness (IMT) of the carotid artery using digital ultrasound images. The IMT measurement of the carotid artery may be used adjunctively with other medical data obtained by a physician to help assess the cardiovascular health of a patient.

The system includes optional Digital Imaging and Communications (DICOM) capabilities as well as general computer communication capabilities to provide the acceptance, transfer, display, storage, and digital processing of ultrasound images and loops. Security support is also provided to facilitate HIPAA compliance.

The system/transducer is capable of exceeding a TI or an MI of 1.0 in certain operating modes or mode combinations. The system displays the current output level in terms of one of two bioeffects indices (“Mechanical

Index [MI]” and “Thermal Index [TI]”) in accordance with the AIUM/NEMA Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment.

Chapter 4: System Overview 21

Theory of Operation

The Edge ultrasound system has seven (7) major functional groups:

• Transducer

• Acquisition Subsystem

• Processing Subsystem

• Display Subsystem

• Control Subsystem

• User Interface Subsystem

• Power Subsystem

is a system block diagram that shows the relationship of the functional groups.

Transducer

External video to monitor, printer

Acquisition subsystem

RF Bus

Processing subsystem

AQ Bus

Display subsystem Video

Control Bus

User interface

Control subsystem

IrDA

Serial Bus

Power subsystem

Display power

Battery pack assembly

Logic power

Pulser voltage

Power adapter

External power

Figure 4.1 SonoSite Edge High-Resolution Ultrasound System Block Diagram

The Transducer elements convert the pulser voltage to acoustic energy during the transmit portion of the ultrasound acquisition cycle. The elements convert the acoustic echo to voltage in the receive portion of the acquisition. The voltage developed on the transducer elements is sensed by the acquisition subsystem. The system transducers have 64 to 192 elements.

The Acquisition Subsystem consists of the beamformer and interface to the transducer. The beamformer controls the timing of the transmit pulses to focus the acoustic beam. The beamformer amplifies the low-level received echos and controls the receive focusing. The system beamformer transmits on up to 128 elements and receives on

64 elements.

The Processing Subsystem includes capabilities for interfacing with the beamformer and performing high speed processing. The processing subsystem demodulates, filters, detects, and compresses the signal supplied by the beamformer into display information.

22 Chapter 4: System Overview

The Display Subsystem converts the detected ultrasound data into picture elements (pixels). The software user interface graphics are combined with the ultrasound information and converted to a video stream. The external video port supports NTSC and PAL format.

The Control Subsystem consists of the central processing unit, program and video memory, permanent image storage and retrieval memory, external communication interface ports, and connection to the user interface keys.

The control software includes the acoustic power and intensity software subsystem, power group monitors, and a beamformer monitor. This software guarantees a level of patient safety by ensuring the system is operating within acoustic power and intensity limits.

The User Interface Subsystem represents the software interface and form factor. The software interface is the interaction between the user and the screen layout components. The form factor is the type of physical buttons, location, and grouping of the buttons and the device size, shape, and weight. Dedicated controls are for high usage activities and grouped according to the user workflow.

The Power Subsystem provides the system power and protects the hardware from destructive and/or unsafe conditions by detecting failures in the system through hardware and software monitors. Detection of a fault results in disabling of the pulser supply, and signaling of an error to the Control Group. The power subsystem includes the battery pack and battery charging electronics.

Description of Operating Modes

2D Mode

M Mode

2D mode is a two dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes. In 2D, a two-dimensional cross-section of a 3-dimensional soft tissue structure such as the heart is displayed in real time. Ultrasound echoes of different intensities are mapped to different gray scale or color values in the display. The outline of the

2D cross-section may be a rectangle, parallelogram, trapezoid, sector, or a full circle, depending on the particular transducer used. 2D mode can be used in combination with any other modes.

M Mode is also known as “T-M mode” or “time-motion” mode. It is used primarily for cardiac measurements such as valve timing and septal wall thickness when accurate timing information is required.

Ultrasound echoes of different intensities are mapped to different gray scale values in a scrolling display. M Mode displays time motion information of the ultrasound data derived from a stationary beam. Depth is arranged along the vertical axis with time along the horizontal axis. M Mode can be used alone but is normally used in conjunction with a 2D image for spatial reference. The 2D image has a graphical line (M-line) superimposed on the

2D image indicating where the M Mode beam is located.

Chapter 4: System Overview 23

Color

Doppler

(Color)

Color Power

Doppler

(CPD)

Continuous

Wave (CW)

Doppler

In color Doppler, a real-time, two-dimensional cross-section of blood flow is displayed. The

2D cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a full circle, depending on the particular transducer used.

The 2D cross-section is presented as a full color display, with various colors being used to represent the velocity, both positive and negative, of the blood flow echoes. Often, to provide spatial orientation, the full color blood flow cross-section is overlaid on top of the gray scale cross-section of soft tissue structure (2D echo). For each pixel in the overlay, the decision of whether to display VCD, gray scale (echo) information or a blended combination is based on the relative strength of echoes from the soft-tissue structures and from the red blood cells.

A high pass filter (wall filter) is used to remove the signals from stationary or slowly moving structures. Tissue motion is discriminated from blood flow by assuming that blood is moving faster than the surrounding tissue, although additional parameters may also be used to enhance the discrimination. The remaining signal after wall filtering may be averaged over time (persistence) to present a steady state image of blood flow distribution. Variance information may also be displayed to provide information when large variance is observed in the velocity information.

In CPD, a real-time two-dimensional cross-section of blood flow is displayed. The 2D cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a full circle, depending on the particular transducer used.

The 2D cross-section is presented as a full color display, with various colors being used to represent the power in blood flow echoes. Often, to provide spatial orientation, the full color blood flow cross-section is overlaid on top of the gray scale cross-section of soft tissue structure (2D echo). For each pixel in the overlay, the decision of whether to display CPD, gray scale (echo) information or a blended combination is based on the relative strength of echoes from the soft-tissue structures and from the red blood cells.

A high pass filter (wall filter) is used to remove the signals from stationary or slowly moving structures. Tissue motion is discriminated from blood flow by assuming that blood is moving faster than the surrounding tissue, although additional parameters may also be used to enhance the discrimination. The power in the remaining signal after wall filtering may be averaged over time (persistence) to present a steady state image of blood flow distribution.

CW provides a real-time representation of blood flow and is displayed as a velocity-versus-time sweeping output. Velocity (or frequency) is presented as the vertical axis with time along the horizontal axis. The magnitude of the detected signal is represented as different gray scale values.

CW Doppler mode provides the clinician with the ability to obtain blood flow velocities focused about a user specified focal region. A continuous transmit waveform of ultrasound energy with a known frequency is transmitted and focused by the system; on the receive side, the transducer receive echoes are continuously amplified, focused about the focal region and converted to a base band quadrature signal. The signal is analyzed by a quadrature phase detector that establishes two receive channels to allow detection of flow direction. These two channels are then analyzed by a fast complex Fourier transform (FFT) circuit to establish the spectrum of frequencies present in the echoes. The data are displayed as spectrum frequencies with respect to time.

CW can be used alone but is normally used in conjunction with a 2D image for spatial reference. The 2D image has a graphical line (D-line) superimposed on the 2D image indicating where the M-mode beam is located.

24 Chapter 4: System Overview

Pulsed Wave

(PW) Doppler

PW provides a real-time representation of blood flow and is displayed as a velocity-versus-time sweeping output. Velocity (or frequency) is presented as the vertical axis with time along the horizontal axis. The magnitude of the detected signal is represented as different gray scale values. The ultrasound data is derived from a single area, the sample volume, on a stationary beam.

PW Doppler mode provides the clinician with the ability to obtain blood flow velocities about a spatial sample volume. A burst of ultrasound with a known spectrum is transmitted by the system; on the receive side, the transducer receive echoes are amplified and range gated at the appropriate depth. The signal is analyzed by a quadrature phase detector that establishes two receive channels to allow detection of flow direction. These two channels are then analyzed by a fast complex Fourier transform (FFT) circuit to establish the spectrum of frequencies present in the echoes. The data are displayed as spectrum frequencies with respect to time.

PW can be used alone but is normally used in conjunction with a 2D image for spatial reference. The 2D image has a graphical line (D-line) superimposed on the 2D image indicating where the M-mode beam is located. The sample volume position (depth) and size are also indicated on the D-Line.

Additional System Feature Performances

Broadband Imaging

Tissue Specific

Imaging

Biopsy Guidance

Measurement and

Calculation

Capabilities

This ultrasound acquisition system uses high resolution broadband technology in the transmit pulsers, transducer, and receivers. The receive path can capture and process signals over a wide spectrum, from below 2.0 MHz to beyond 10 MHz. For each application, the transmit pulse is designed to produce an appropriate bandwidth. For example, in 2D grayscale imaging, a wide band pulse is used to support good axial resolution. For Doppler modes, a narrower band pulse is used, which improves the spectral resolution of the detected Doppler signal.

In addition to transmit pulse control, programmable digital signal processing is used in the receive path to further refine the bandwidth used to produce the final image.

Digital filters are applied to the digitized received signal to limit and shape the spectral bandwidth used to generate the displayed output.

In this feature, parameters for signal and image processing are optimized to maximize the image quality or to obtain the best compromise of resolution and penetration for different specific clinical applications. These parameters include: the order of received filters, the bandwidth, the dynamic range, the compression curve, the gain setting and parameters for compounding frequency band, etc. For example, different system parameter setups are used for abdominal or peritoneal scanning. This feature is for ease of use for the operator by automatically setting up system control parameters rather than manually adjusting settings for best performance.

The system can display a pair of biopsy guidelines that represent the anticipated path of the biopsy needle. The image of an anatomical target, biopsy guidelines, a scan plane marker, and a biopsy needle are displayed to assist in guiding the biopsy needle to the target. The system also provides needle guidance for vascular access procedures. For additional information, see the biopsy user guides.

The system offers a variety of measurements and calculations, specific to exam type and transducer. A list of them, and author references, are in the system user guide.

Measurement accuracy is also discussed.

Chapter 4: System Overview 25

Continuous Wave

Doppler Audio

Output

Pulsed Wave Doppler

Audio Output

Electrocardiograph

(ECG) Display

The system provides for audio output of the CW velocity information. This can be presented as stereo information, with flow moving towards the transducer on one channel and flow away on the other, or as a mono output with the single audio output representing the summation of the flow directions.

The system provides for audio output of the PW velocity information. This can be presented as stereo information, with flow moving towards the transducer on one channel and flow away on the other, or as a mono output with the single audio output representing the summation of the flow directions.

ECG is provided to measure the electrical signal generated by the heart. A three lead interface: Right Arm (RA), Left Arm (LA) and Left Leg (LL), is provided on the system.

The ECG signal is displayed as an amplitude-versus-time sweeping output.

Amplitude is presented on the vertical axis with time along the horizontal axis.

Front End Overview

The Front End is designed to support various imaging modalities such as 2D, M-Mode, Spectral Doppler and Color

Doppler. From the Front End's perspective, all modes can be grouped into a few basic types: Single mode, simultaneous modes and triggered modes. All these modes are built from similar, basic transmit and receive sequences controlled within the Front End. A generic top level block diagram of a typical Front End is in the figure below.

Transducer Receive Section

T/R Sw

TGC Amplifier

A/D Delay X

Weight

Pulser Delay

Beamformed

Data

Waveform

T/R Sw

Pulser

TGC Amplifier

Delay

A/D

Waveform

Figure 4.2 Front End Block Diagram

Transmit Section

Delay X

Weight

Control

The transmit section consists of a waveform generator, delay block, and high power high voltage driver to excite the transducer element. Multiple elements are driven with delays determined by the time of flight in the medium from the elements to the point in space where the beam is to be focused. The longer the time of flight is to the focal point the smaller the delay is for a given transmit element to allow all to arrive at the focal point at the same time.

26 Chapter 4: System Overview

The number of elements driven is determined by element sensitivity off axis and depth of field considerations. The waveform is selected to drive the transducer at a certain center frequency, bandwidth, and power and is optimized for the given mode.

The receive section consists of a transmit/receive switch to protect the receiver from the transmit voltage, a variable gain receiver to amplify and condition the return echoes, an A/D to digitize the data, a delay block to focus the return signals and a weight block to scale the return echoes for each channel. All the signals are then summed together to generate the beam-formed receive data. The analog gain varies with depth to compensate for signal attenuation through the medium. The delays and weights are independent for each channel. The delay and weight for the receive channel can typically be changed dynamically to keep the receive beam in continuous focus. The delay is simply set by the time of flight in the medium from the point of interest to the element, which starts at skin-line and proceeds to the deepest depth of interest.

The control section drives the data to the various data path elements on a line by line basis, controls the timing of the transmit and receive sections and controls the tagged information and timing of the data to the rest of the system.

Chapter 4: System Overview 27

PW Doppler Processing

Doppler processing includes both audio processing which presents Doppler signal in the form of stereo audio and spectral processing which generates data for display of Doppler spectrum in the form of a scrolling spectrogram.

Doppler power spectrum is estimated performing Discrete Fourier Transforms on short, overlapped segments of wall filtered Doppler signal. Doppler audio data is generated from wall filtered data by phase shifting the in-phase component.PW Doppler Processing Function Block Diagram

Temporal

Averaging

Compress

Peak &

Mean

Baseline shift

Display

Interpolate

Back

End

RF

QBP

Wall

Filter

Post

Gain

Resample Window FFT |.| indicates IQ pairs

I

Hilbert

Phase Shifter

Q

Figure 4.3 PW Doppler Processing Function Block Diagram

Delay

+

+

-

+

+

+

Audio Output

Audio

Gain

Audio

Gain

2x16 bits @ PRF rate

28 Chapter 4: System Overview

CW Doppler Processing

CW Doppler data will be presented to the signal processor as complex (I/Q) data from the analog front end of the external DSP. The 16-bit data will be presented as consecutive samples at a data rate varying from 1.5 kHz to 64 kHz for the complex pair. Most of CW processing is similar to that of PW except for the QBP function. In place of

QBP will be a low pass decimating filter that operates on incoming I/Q data.

The Doppler Processing block must allow storage of 128 undetected I/Q pairs in to allow the system to measure and correct for phase mismatch. Measuring and correcting will need to be accomplished in system software.

CW Asic

CW Asic

Amplifier

Amplifier

Other Channel with same Delay

Freq

Delay

Freq

LO Gen

LO Gen

LO I

LO Q

LO I

LO Q

Master Clk

I

Q

I

Q

QSum

I Sum

Mux

Mux

Amplifier

Other Channel with same Delay

Amplifier

Freq

Master Clk

Freq

Delay

Master Clk

LO Gen

LO I

LO Q

LO I

LO Q

I

I

Q

Q

Figure 4.4 CW Doppler Processing Function Block Diagram

Q

I-V

I

Q

I

PCB

PCB

Q

I

Q

A/D A/D

Q

I

A/D

A/D

I

Q

I

DSP DSP

Q Q

I

I

Processing

Chapter 4: System Overview 29

Back End Overview

The Back End subsystem is responsible for the conversion of raw acquisition data into a raster image ready for display. The Back End subsystem also contains the video data path that supports generation of video comprising of the ultrasound image as well as graphics annotation. Video generation of both standard composite interlaced video and progressive scan video is supported. Most functionality is within the ASIC but the memory resources for acquisition memory, and display memory are found in external memory components. The conversion from PC type video to TV type video is also performed externally.

Control is received initially from the CPU to setup each functional block and afterward the hardware is completely data driven. This control takes the form of programming setup registers inside the blocks and setting up scan conversion tables. Each block provides temporary storage as required to buffer data and keep their respective processing pipeline full and operating. Also note that the block diagrams show only the data path, but each block is responsible for generating any necessary memory addresses for their respective input data stream.

The Edge Back End subsystem is shown in the figure below.

Figure 4.5 Back End Subsystem Block Diagram

The Back End Subsystem performs processing encompassing three main data domains, acquisition data, raster data, and video data. Support for acquisition data includes the input buffer, flash suppression, frame average, and external ACQ memory. Cine buffer management is performed by the acquisition controller. Conversion from acquisition data to raster data is performed by the graphics overlay, scan conversion engine, sweeping engine, and 3D engine. Raster data is stored in an external DISPLAY memory. Also supporting raster operations is the graphics support block that provides acceleration hardware for pixel operations from the CPU and graphics overlay engine. Video data is processed as progressive scan and supplied externally on a digital bus. In addition, interlaced video is supplied in both composite and S-video formats. The progressive video path includes buffers, priority logic, and LUTs. External video in signals are input and multiplexed onto the external video out path to allow for external sources to display information on connected displays, VCRs, or printers.

30 Chapter 4: System Overview

Control Subsystem

The Edge Control Subsystem is shown in the figure below.

INFRASTRUCTURE

SYSPWR,

SYSCLKS,

SYSRST

EXPANSION

MASS

STORAGE

ETHERNET FAST

ETHERNET

SYSTEM

FLASH

SYSTEM

RAM

PCI BUS

PCI BRIDGE

DUAL UART,

USB

2 U/I, DEBUG

USB

SYSTEM BUS

DATA

ADDRESS

CONTROL

SYSTEM

TIMERS DMA

FLOATING

POINT

IIC

CONTROLLER

SH

CONTROL

SH MUX

CONTROL cache

INTERRUPTS

JTAG

PROCESSOR

8 GPIO

SCANHEAD

IIC BUS

ULTRASOUND

DATA PATH

VIDEO

AUDIO

ECG

DATA

CONTROL

Figure 4.6 Control Subsystem Block Diagram

The core control subsystem contains the processor, the system bus, the system memory resources of FLASH and

RAM, the interrupt logic, system timers, a DMA engine, and a floating point unit. Support for the ultrasound subsystem consists of a scanhead interface, scanhead mux control.

Communication interfaces consists of an Ethernet interface, USB port, two general purpose serial bus interfaces, and the IIC bus. The EDGE control architecture is an open architecture. It supports functionality extension through the incorporation of the PCI bridge to the PCI bus. Functionality may be added by adding to the PCI Bus.

Chapter 4: System Overview 31

Power Supply and Control

The Edge Power Supply and Control System consists of an easily replaced rechargeable battery pack; an On/Off

Key; a standby power regulator; digital, analog, display and transducer power supplies; a power monitor and a power control system. Operating current is drawn from the battery or an external AC/DC Adapter which also contains circuitry for charging the battery. A fan and provision for a temperature sensor are also included.

Figure 4.7 Power Supply and Control Subsystem Block Diagram

32 Chapter 4: System Overview

Battery Pack (VBAT)

A rechargeable lithium-ion battery pack will be used to operate the unit in battery mode. The pack will include a capacity monitoring circuit and any required pack protection circuitry. A one-wire, bidirectional, serial interface

(BDATA) will be used to read and write the pack data.

Battery Charger

The charge circuitry is in the external AC/DC Adapter as shown in the following block diagram.

Line

Neutral

Ground

POWER SUPPLY / BATTERY CHARGER

Voltage

Source

Output Cable

+15VDC

5 5

Universal

AC

Input

0.1

UF

10

UF

R2

Current Limited

Voltage Source

Charge

Control

Current

Sense

Voltage

Sense

2K

.C57

.01uF

Ferrite /

Common

Mode

Choke

3 turns on

Fair Rite

PN:

263162510

2

20K

VCHG

RATE/SYNC

VSENSE

PWR_RETURN

Cable Shield

Ferrite

2 2

VBAT+

3

4

3

4

1 1

VBAT-

SYSTEM

System

Power

Supply

Figure 4.8 Battery Charging Subsystem Block Diagram

Chapter 4: System Overview 33

ECG Module

The ECG module allows a representation of the heart electrical activity to be displayed in real time with ultrasound images acquired and displayed on the system video display.

The ECG module interfaces to the patient through three (3) ECG leads: Right Arm ECG lead (RA), Left Arm ECG lead

(LA), and Left Leg ECG lead (LL). The ECG received signal from the ECG electrodes are isolated, amplified, and filtered by the ECG module before it is sent to the system for further processing and display.

The ECG module and cable are an integrated assembly. The module receives power from the system. Patient isolation is provided by the ECG module, allowing the connection and signals to the system to be system-ground referenced. The isolation between the patient and the system meets the requirements of IEC 601-1 for Type CF equipment.

The ECG function accepts input from an external serial A/D and performs gain, filter, DC Offset and trigger functions. The resultant data is output at either the 200Hz sample rate or decimates the data by 2 or 4 and outputs the data into acquisition memory. The data is assumed to be signed. The ECG trigger function is implemented by a simple edge sensitive trigger along with SW monitoring the ECG data and triggering the FE after a user defined delay from the detected R wave. An interrupt is provided that will interrupt the processor after a set delay from the detected level and slope. A simple block diagram of the HW is shown below.

Figure 4.9 ECG Block Diagram

DICOM

The system features Digital Imaging and Communications (DICOM) capability to provide the acceptance, transfer, display, storage, and digital processing of single ultrasound images as well as loops of ultrasound images.

IMT

The system includes the ability to measure the intima-media thickness (IMT) of the carotid artery using digital ultrasound images. The intima is that region of the arterial wall from and including the endothelial surface at the lumen to the luminal margin of the media. The media layer extends from the intima to the adventitia of the vessel wall. The adventitia is normally quite echogenic on ultrasound images when compared to the media. The IMT measurement of the carotid artery may be used adjunctively with other medical data obtained by a physician to help assess the cardiovascular health of a patient.

34 Chapter 4: System Overview

Chapter 5: Troubleshooting

This chapter contains information to help you correct problems with system operation.

System and Subsystem Diagnosis

This section covers basic diagnostic and troubleshooting procedures you may follow if the system does not operate properly. To diagnose system failures, consult the referenced diagnostic figures that follow or SonoSite

Technical Support.

Table 5.1: Troubleshooting Subassemblies and Diagnostic Figures

Subassemblies

DICOM

Display

Battery

Control Panel

Diagnostic Figures or Table

Table 5.2

TBA

TBA

TBA

System Repair

The system is repairable through subassembly replacement or through replacement of parts as recommended by

SonoSite. Component level repair of Printed Circuit Board Assemblies is performed only at the SonoSite repair facility. Replacement of board level components by unauthorized service facilities voids the SonoSite warranty.

Test Equipment

Test equipment is not required for this troubleshooting section. Troubleshooting test aids include an external monitor and a spare battery.

Chapter 5: Troubleshooting 35

Failure (Assert) Codes

The system displays an “assert screen” for hardware and software issues related to Main PCBA failures. Main PCBA failures typically result in “assert codes” that are output to the display. If an assert screen appears, note the assert information and contact SonoSite Technical Support to clarify the failure. Figure 5.1

shows an assert screen. The assert information required is the information listed on the “C:” line and the “D:” line.

Record all information on the “C:” line and

“D:” line and contact Technical Support

Figure 5.1 Assert Screen

Verifying a System Assert Code

System asserts are caused by hardware and/or software faults. Hardware asserts typically require main PCBA replacement. Software asserts can be reset and the system may recover. A simple method to identify the cause of the assert is identified here:

Assert Cause 1 Record the assert code.

2 Press and release the Power button to power the system down.

3 Press the Power button again to power on the system.

• If the system powers on normally, it has recovered from the fault (software assert) and you may use the system.

• If the assert condition remains, corrective action must be taken; usually replacement of the main PCBA is required. Contact SonoSite Technical Support for assistance and to obtain repair parts.

If the Power button is not functional, all sources of power must be removed to allow the system to power down. I.e., disconnect AC power and remove the battery.

36 Chapter 5: Troubleshooting

DICOM

Table 5.2: DICOM Troubleshooting

Error Message

Socket communication failed

Tiller Error Code

TSOCKET_CONNECT_FAILURE

Archiver transaction failed

Printer transaction failed

DICOM network communication failed

Internal failure detected

TDICARCH_OPEN_FAILURE

TDICPRNT_OPEN_FAILURE

TDNETWORK_OPEN_FAILURE

TDNETWORK_READ_FAILURE

Cause

Invalid network configuration.

Wrong port number.

Application is not running.

Printer is offline.

Wrong Capture

Type selected

Wrong Image settings

Device does not recognize Edge, rejects association

Invalid DICOM

Attribute

Troubleshooting

Using Ping, verify that the

Printer/Archiver is connected.

• If Ping fails, check the devices IP address, Edge IP address,

Subnet mask, and Gateway IP address.

• If Ping is OK, use Verify to check if device is available.

If Verify fails: a) Check the Printer/Archiver’s

Port configuration on the Edge.

b) Ensure that the Printer is online and the Archiver’s application is running.

Verify that the Archiver supports the selected Capture Type setting, e.g., US Image, SC Image or

US-Ret Image.

Verify that the printer supports the selected Image settings. E.g,.

Color (RGB) or Grayscale

(Monochrome)

Verify that Edge AE Title or IP address is correctly configured on the Printer/Archiver.

Note: Some devices require that the

Imaging modality (Edge) be recognized in order to accept images. This requires configuration on the device.

Check Edge Printer DICOM settings for correctness (e.g., film size, format)

Chapter 5: Troubleshooting 37

38 Chapter 5: Troubleshooting

Chapter 6: Replacement Procedures

Caution:

Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first. Initial ESD symptoms may be slightly degraded performance or image quality.

Caution:

All fasteners should be torque to 5.5-inch pounds except where noted such as the Control Panel Subassembly which is should be torque to 3.0-inch pounds.

Display Replacement

Required Parts

One of the following:

• P15637 Service Assembly, Display, EDGE

• P15638 Warranty Service Assembly, Display, EDGE

• V15637 Vet Service Assembly, Display, EDGE

• V15638 Vet Warranty Service Assembly, Display, EDGE

Required Tools

• #1Phillips screwdriver

• #2 Phillips screwdriver

• Torque screwdriver, 2.0-10.0 inch pounds (0.23-1.1 Newton meter)

• An anti-static mat

• A wrist grounding strap

Display Removal

(Please be advised that the EDGE Display is now a complete assembly comprised of the top cover, display frame, wiring harnesses, hinges, and bezel. Do not attempt to replace the LCD as this is not possible as a field replacement.)

Display Removal

1 Lay the EDGE System on its top exposing the bottom section and remove the system battery.

2 Remove the eight screws from the Bottom Enclosure of the system with a #2 Phillips screw driver per

Figure 6.1

.

Chapter 6: Replacement Procedures 39

Screws (8x)

Figure 6.1 System Bottom

3 After removing the EDGE Bottom Enclosure, remove the two screws located in 

Figure 6.2

.

Don’t forget these two screws to release the top enclosure. (Refer to both

arrow lines in the figure to the left).

Figure 6.2 Main Frame Bottom

Caution:

Use caution when removing the Top Enclosure area under the Display. This area is retained with an adhesive strip and using too much force will break the Top Enclosure requiring replacement.

Carefully slide a small flat bladed screwdriver from one side to the other to release the tape.

4 Turn the system over, fully open the display, and remove the EDGE Top Enclosure per

Figure 6.3

.

40 Chapter 6: Replacement Procedures

Caution:

Adhesive backing under this area.

Lift front area first.

Figure 6.3 System Control Panel View

5 Remove the 10 perimeter screws holding the Control Panel, carefully lift from under the Control Panel and

slide forward. Refer to Figure 6.4

.

Screws

(10x)

Front Edge

Figure 6.4 Control Panel Removal

6 Remove the 2 screws holding the Hinge Cover and remove per

Figure 6.5

.

Chapter 6: Replacement Procedures 41

Hinge Cover screws (2x)

Figure 6.5 Hinge Cover

Caution:

When lifting the front edge of the Control Panel away from the Main Frame, lift under the Control Panel board. Do not lift by the rubber membrane. This can easily become detached from the Control Panel board and cannot be refastened!

7 Expose the underside of the Control Panel by lifting from under the front edge and hinge via the top edge toward the LCD display.

8 Remove the J6 “LCD OUT”, J8 “BL OUT”, J5 “LCD IN” and J7 “BL IN” cables by gently pulling away from their sockets.

9 Remove the J4 “SPI” and J9 “GAIN” ribbon cables after unlocking the dark gray retainer on both sockets as shown in

Figure 6.6

.

J9

J5

Caution:

J4 & J9 require the gray lock tab to be lifted at the end closest to the ribbon cable.

..

Use care as the lock tab hinge points are fragile.

J4

J6

J7 J8

Figure 6.6 Control Panel Bottom

10 Tilt the system back onto the top cover allowing the LCD to lie on its back.

11 Remove the 4 screws from the LCD display hinges as shown in

Figure 6.7

.

42 Chapter 6: Replacement Procedures

Screws (4x)

Figure 6.7 Display Screws

Display Replacement

Display Replacement

1 Set the new Display Assembly in place.

2 Install the four hinge screws that hold the Display Assembly in place. Torque the screws to 5.5-inch pounds.

3 Place the Control Panel onto the EDGE frame.

4 Connect both Display Assembly cables (J6 “LCD OUT” & J8 “BL OUT”) to the Control Panel and (J5 “LCD IN” &

J7 “BL IN”) from the Main PCBA to the Control Panel.

5 Connect both ribbon cables (J4 “SPI” & J9 “GAIN”). Lock the J4 and J9 cable retainer in the closed position.

6 Reinstall the ten screws that hold the Control Panel in place. Torque the screws to 3.0-inch pounds.

7

Install the Hinge Cover and two screws removed per Figure 6.5

. Torque the screws to 3.0-inch pounds.

8 Install the EDGE Top Enclosure by sliding and snap down into place.

9 Close the Display Assembly and turn the system over.

10 Install the two screws removed from

Figure 6.2

. Torque the screws to 5.5-inch pounds.

11 Install the eight screws on the bottom of the system per Figure 6.1

. Torque the screws to 5.5-inch pounds.

Test the Display

Test Display

1 Replace the battery or attach an external power supply.

2 Press the Power key to apply power to the system.

3 Verify the display operates correctly.

Chapter 6: Replacement Procedures 43

Control Panel Subassembly Replacement

Required Parts

One of the following:

Service Assembly Control Panel

P15618-01 Control Panel, EDGE, English

P15621-01 Control Panel, EDGE, French

P15622-01 Control Panel, EDGE, German

P15623-01 Control Panel, EDGE, Italian

P15624-01 Control Panel, EDGE, Spanish

P15625-01 Control Panel, EDGE, Russian

P15626-01 Control Panel, EDGE, Portuguese

Warranty Service Assembly Control Panel

P15630-01 Control Panel, EDGE, English

P15631-01 Control Panel, EDGE, French

P15632-01 Control Panel, EDGE, German

P15633-01 Control Panel, EDGE, Italian

P15634-01 Control Panel, EDGE, Spanish

P15635-01 Control Panel, EDGE, Russian

P15636-01 Control Panel, EDGE, Portuguese

Required Tools

• #1 Phillips screwdriver

• #2 Phillips screwdriver

• Torque screwdriver, 2.0-10.0 inch pounds (0.23-1.1 Newton meter)

• An anti-static mat

• A wrist grounding strap

Caution:

Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first. Initial ESD symptoms may be slightly degraded performance or image quality.

Control Panel Removal

Control Panel Removal

1 Lay the EDGE System on its top exposing the bottom section and remove the system battery.

2

Remove the 8 screws from the Bottom Enclosure of the system using a #2 Phillips screwdriver per Figure 6.1

.

3

Remove the 2 screws in the EDGE Main Frame securing the Top Enclosure per Figure 6.2

.

Use caution when removing the Top Enclosure area under the Display. This area is retained with an adhesive strip and using too much force will break the Top Enclosure requiring replacement. Carefully slide a small flat bladed screwdriver from one side to the other to release the tape.

Caution:

When lifting the front edge of the Control Panel away from the Main Frame, lift under the Control Panel board. Do not lift by the rubber membrane. This can easily become detached from the Control Panel board and cannot be refastened!

4

Turn the system over, fully open the Display Assembly, and remove the EDGE Top Enclosure per Figure 6.3

.

5 Remove the ten perimeter screws holding the Control Panel, carefully lift, and slide forward. Refer to

Figure 6.4

.

6 Expose the underside of the Control Panel by lifting from the front EDGE and hinge via the top EDGE toward the LCD display.

7 Remove J6 “LCD OUT”, J8 “BL OUT”, J5 “LCD IN”, and J7 “BL IN” cables by gently pulling away from their sockets.

8 Remove the J4 “SPI” and J9 “GAIN” ribbon cables after unlocking the dark gray retainer on both sockets as per

Figure 6.6

.

9 Lift off the loose and un-tethered Control Panel assembly from the EDGE frame.

44 Chapter 6: Replacement Procedures

Control Panel Replacement

Control Panel Replacement

1

Place the new Control Panel onto the EDGE frame and line-up the holes as per Figure 6.8

.

Line up these holes with the holes in the core before tightening the screws otherwise top enclosure is difficult to fit.

Figure 6.8 Control Panel Alignment

2 Expose the underside of the Control Panel by lifting from the front edge and hinge via the top edge toward the LCD display.

3 Connect the J4 “SPI” and J9 “GAIN” ribbon cables and lock the dark gray retainer on both sockets as shown in

Figure 6.6

.

4 Connect the J6 “LCD OUT”, J8 “BL OUT”, J5 “LCD IN”, and J7 “BL IN” cables.

5

Install the ten screws removed per Figure 6.4

. Torque the screws to 3.0-inch pounds.

6 Install the EDGE Top Enclosure and push into place.

7 Close the display and turn the system over.

8

Install the 2 screws in the EDGE Main Frame securing the Top Enclosure per Figure 6.2

. Torque the screws to

5.5-inch pounds.

9 Install the eight screws on the Bottom Enclosure of the system per

Figure 6.1

. Torque the screws to 5.5-inch

pounds.

Main System Disassembly for Repair and/or Replacement

Required Parts

• Main PCBA

• P15615 Service Assembly Main PCBA, EDGE

• P15617 Warranty Service Assembly Main PCBA, EDGE

• V15615 Vet Service Assembly Main PCBA, EDGE

• V15617 Vet Warranty Service Assembly Main PCBA, EDGE

Note: The replacement Main PCBA does not include the Transducer Nest Frame Assembly. These parts must be transferred from the original Main PCBA. If new parts are required, please order the following components to complete the Transducer Nest Frame Assembly.

Chapter 6: Replacement Procedures 45

• Nest Frame Assembly, M-Turbo (Compatible with EDGE system):

• P07750 Nest Frame Assembly

• P00364 Connector, Interposer (8x)

• P03833 Shield, Perimeter, Short (2x)

• P03834 Shield, Perimeter, Long (2x)

• P00924 Screw, Shoulder, Thrust Plate (4x)

• P00353 Wear Plate

• P00646 Spring, Thrust Plate, .047 wire (4x)

• P08200 Socket Head Cap Screw, M2.5-.45x10mm (4x)

Required Tools

• #1 Phillips screwdriver

• #2 Phillips screwdriver

• Torque screwdriver, 2.0-10.0 inch pounds (0.23-1.1 Newton meter)

• 2 mm allen key

• An anti-static mat

• A wrist grounding strap

Caution:

Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first. Initial ESD symptoms may be slightly degraded performance or image quality.

System Disassembly

System Disassembly

1 Remove the battery.

2 Remove the Bottom Enclosure from the system following the removal procedures in “Display Removal” on page 39.

3 Remove the Top Enclosure and Control Panel from the system following the removal procedures in “Display

Removal” on page 40.

4 Follow the cable removal from the procedure in “Display Removal” on page 42. This exposes all of the

replacement parts for the main system per Figure 6.9

.

Major System Components

Main PCBA

USB Extension

CB Assembly

Fan

Nest Frame Assembly

SD Card

Daughter-card under shield)

Power Supply

(under shield)

Speaker

Speaker

Figure 6.9 System Components

46 Chapter 6: Replacement Procedures

Speaker Replacement

Required Part

• P03872 Speaker Assembly

Caution:

Use caution when removing the left speaker connector to prevent damage to the Main PCB A components around the connector.

Speaker Replacement

1 Press on the connector release and pull the connector out of the receptacle.

2 Gently pry off the retaining clip with a flat bladed pry tool. See

Figure 6.10

.

3 Replace the speakers by reversing steps 1-2.

Connector

Speaker

Retaining Clip

Figure 6.10 Speaker Replacement

Chapter 6: Replacement Procedures 47

Power Supply PCBA Replacement

Required Part

• P08850 Service Assembly, Power Supply, M-Turbo (Compatible with the EDGE system)

Power Supply Removal

1 Remove the 7 long screws and 2 short screws holding the Power Supply Shield. Please note the location of

the 2 short screws. See Figure 6.11

.

Power Supply shield

Long Screws (7x)

Short Screws (2x)

Figure 6.11 Power Supply Shield

2 Gently lift the Power Supply PCBA away from the Main PCBA as shown in

Figure 6.12

. (Connector in upper left

corner/removing the Power Supply PCB without the shield is difficult)

3 Install the new Power Supply PCBA by reversing steps 1-2.

Power Supply connector underneath mates Power

Supply PCB to the Main

PCBA.

Power Supply PCBA

Figure 6.12 Power Supply PCBA

48 Chapter 6: Replacement Procedures

SD Card Daughter-card

Required Parts

• P10168 SD Card, 2GB (4 required)

• P07442 SD Card Daughter-card

SD Card Daughter-card Removal

1 Remove the 1 long screw holding from the SD Card Daughter-card. See

Figure 6.13

.

USB Extension PCB

SD Card

Daughter-card alignment pins

Long Screw (1x)

Note: Kapton tape is used to retain SD

Card in place.

Figure 6.13 SD Card Daughter-card

2 Gently lift the SD Card Daughter-card straight up away from the Main PCBA.

SD Card Daughter-card Replacement

1 Install the SD Card Daughter-Card onto the Main PCBA frame using the alignment holes/pins on the card and

frame. See Figure 6.13

.

Caution:

Improper installation of the SD Card Daughter-card will cause all or part of the internal image storage memory to not be recognized by the system.

2 Install the 1 long screw ensuring the proper location. See

Figure 6.13

.

3 Install the 7 long screws and 2 short screws holding the Power Supply Shield. Please note the location of the

2 short screws. See Figure 6.11

.

Chapter 6: Replacement Procedures 49

USB Extension PCB Assembly Replacement

Required Part

• P14727 USB Extension PCB Assembly

USB Extension PCB Assembly Removal

Note: Please proceed to the Main PCBA Replacement section as all the steps detailed to replace the Main PCBA will require the removal of the USB Extension PCB Assembly.

Main PCBA Replacement

Main PCBA Removal

1 Remove the Power Supply PCBA and SD Card Daughter-card as described in the previous steps.

2 Disconnect the power cable to the Fan Assembly and remove the 2 screws mounting the Fan Assembly. Refer

to Figure 6.14

.

Screws (2x)

Small Hole Large Hole (key)

Figure 6.14 Fan Assembly Screws

3 Remove the 5 screws holding the Main PCBA and the 2 screws holding the USB Extension PCB onto the EDGE

Base. Refer to

Figure 6.15

.

4 Disconnect the speaker wires from the Main PCBA.

Caution:

Use caution when removing the left speaker connector to prevent damage to the Main PCBA components around the connector.

50 Chapter 6: Replacement Procedures

Screws (5x)

Screws (2x)

Figure 6.15 Main PCBA Screws

5 Turn the system over.

6

Remove the 4 Socket Head Cap Screws as shown in Figure 6.16

. This releases the Nest Frame and will allow the

Main PCBA to be removed.

7 As you remove the nest frame assembly from the PCBA, tilt the PCBA and enclosure to almost vertical to avoid spilling the Interposer Connectors from the assembly.

Note: The Nest Frame is keyed onto the Main PCBA via two smaller and two larger hole sizes. Installing incorrectly (180 degrees out) will result with the EDGE system unable to recognize any connected transducers.

8 Lift on the edge of the Main PCBA closest to the system handle at the front right edge. Pull up from the side where the Fan Assembly mounts.

9 Disconnect the USB Extension PCB Assembly from the Main PCBA.

Screws (4x)

Figure 6.16 Nest Frame Top Screws

2.5mm Socket Head Cap Screw

Chapter 6: Replacement Procedures 51

Main PCBA Replacement

Replace the Main PCBA by following the reverse of the removal procedure. Do not tighten all the screws until everything is in place.

1 Connect the USB Extension PCB Assembly to the Main PCBA.

2 Replace the Main PCBA.

3 Reinstall the Fan Assembly.

4 Reinstall the Nest Frame Assembly. The Nest Frame Socket Head Cap Screws should be torque to 4.5-inch pounds.

Caution:

The Nest Frame is keyed onto the Main PCBA via two smaller and two larger hole sizes. Installing incorrectly (180 degrees out) will result with the EDGE system unable to recognize any connected transducers. Refer to

Figure 6.14

.

5 Reconnect the speaker wires.

6 Reinstall the Power Supply PCBA.

7 Reinstall the SD Card Daughter-card.

8 Reinstall the Power Supply Shield.

9 Tighten all screws to their specified torque of 5.5-inch pounds.

10 Reinstall the Control Panel.

11 Reinstall the Top Enclosure Assembly.

12 Reinstall the Bottom Enclosure.

Middle Machined Enclosure Replacement

Required Parts

• P15304-04 Enclosure, Middle, Machined

• P15984-01 Gasket, EMI/ESD (2 Required)

• P07885-01 Thermal Pad, 1.00 x 1.00 x 0.100” (4 Required)

• P07886-01 Thermal Pad, 0.35 x 0.35 x 0.100” (4 Required)

Middle Machined Enclosure Replacement

Note: The EMI/ESD Gaskets and the Thermal Pads may be reused from the old assembly but advise having new items to ensure that these will contact the necessary areas properly.

1

Refer to P15304 for a view of the Middle Machined Enclosure on page 73

.

2 Follow all the above procedures to remove all PCB and PCBA assemblies as well as all the hardware fastened from the original Middle Machined Enclosure.

3 Transfer all the components to the new Middle Machined Enclosure.

4 Finish assembly by reversing step 2.

Bottom Enclosure Replacement

Required Parts

Note: Contact SonoSite Technical Support if it is necessary to replace the bottom enclosure. Ordering the bottom enclosure requires special handling due to the serial number label

52 Chapter 6: Replacement Procedures

Chapter 7: Maintenance

This chapter contains information to help you properly care for the system, transducers, and accessories.

Periodic Maintenance

No periodic or preventive maintenance is required for the system, transducers, or accessories other than cleaning and

disinfecting the transducer after every use. (See “Cleaning and disinfecting transducers” on page 55 .) There are no

internal adjustments or alignments required and there are no internal components that require periodic testing, calibration, adjustment, or alignment. Performance tests are described in

Chapter 8, “Performance Testing”

of this manual. Performing maintenance procedures not described in this manual may void the product warranty.

Local regulations may require electrical safety testing.

Contact SonoSite Technical Support for any maintenance questions. (See “Technical Support (USA, Canada)” on page 1 .)

Cleaning and disinfecting

Use the recommendations in this section when cleaning or disinfecting the ultrasound system, transducer, and accessories. Use the cleaning recommendations in the peripheral manufacturer’s instructions when cleaning or disinfecting peripherals.

For recommended cleaners and disinfectants, see the disinfectant list available on the CD included with the ultrasound system and on www.sonosite.com.

WARNING:

WARNING:

WARNING:

Caution:

Disinfectants and cleaning methods listed are recommended by SonoSite for compatibility with product materials, not for biological effectiveness. Refer to the disinfectant label instructions for guidance on disinfection efficacy and appropriate clinical uses.

The level of disinfection required for a device is dictated by the type of tissue it contacts during use. To avoid infection, ensure that the disinfectant type and the solution strength and duration are appropriate for the equipment. For information, see the disinfectant label instructions and the recommendations of the

Association for Professionals in Infection Control and Epidemiology (APIC) and the FDA.

To prevent contamination, the use of sterile transducer sheaths and sterile coupling gel is recommended for clinical applications of an invasive or surgical nature. Do not apply the transducer sheath and gel until you are ready to perform the procedure.

Some transducer sheaths contain natural rubber latex and talc, which can cause allergic reactions in some individuals. Refer to 21 CFR 801.437, User labeling for devices that contain natural rubber.

Chapter 7: Maintenance 53

Cleaning and disinfecting the ultrasound system

The exterior surface of the ultrasound system and the accessories can be cleaned and disinfected using a recommended cleaner or disinfectant.

WARNING: To avoid electrical shock, before cleaning, disconnect the system from the power supply or remove from the mini-dock or docking system.

WARNING: To avoid infection always use protective eyewear and gloves when performing cleaning and disinfecting procedures.

WARNING: To avoid infection, ensure that the solution expiration date has not passed.

Caution:

Caution:

Caution:

Caution:

Caution:

Do not spray cleaners or disinfectant directly on the system surfaces. Doing so may cause solution to leak into the system, damaging the system and voiding the warranty.

Do not use strong solvents such as thinner or benzene, or abrasive cleansers, since these will damage the exterior surfaces.

Use only recommended cleaners or disinfectants on system surfaces. Immersion-type disinfectants are not approved for use on system surfaces.

When you clean the system, ensure that the solution does not get inside the system controls or the battery compartment.

Do not scratch the LCD screen.

To clean the LCD screen

Dampen a clean, non-abrasive, cotton cloth with an ethanolic-based cleaner, and wipe the screen clean. Apply the cleaner to the cloth rather than the surface of the screen.

To clean and disinfect system surfaces

1 Turn off the system.

2 Disconnect the system from the power supply, or remove it from the mini-dock or docking system.

3 Clean the exterior surfaces using a soft cloth lightly dampened in a mild soap or detergent cleaning solution to remove any particulate matter or body fluids.

Apply the solution to the cloth rather than the surface.

4 Mix the disinfectant solution compatible with the system, following disinfectant label instructions for solution strengths and disinfectant contact duration.

5 Wipe surfaces with the disinfectant solution.

6 Air dry or towel dry with a clean cloth.

54 Chapter 7: Maintenance

Cleaning and disinfecting transducers

To disinfect the transducer and its cable, use the immersion method or the wipe method.

WARNING: To avoid electrical shock, before cleaning, disconnect the transducer from the system.

WARNING: To avoid injury, always use protective eyewear and gloves when performing cleaning and disinfecting procedures.

WARNING: To avoid infection, ensure that the solution expiration date has not passed.

Caution:

Caution:

Caution:

Caution:

Caution:

Caution:

Transducers must be cleaned after every use. Cleaning transducers is necessary prior to effective disinfection. Ensure that you follow the manufacturer's instructions when using disinfectants.

Do not use a surgeon's brush when cleaning transducers. Even the use of soft brushes can damage a transducer. Use a soft cloth.

Using a non-recommended cleaning or disinfection solution, incorrect solution strength, or immersing a transducer deeper or for a longer period of time than recommended can damage or discolor the transducer and void the transducer warranty.

Do not allow cleaning solution or disinfectant into the transducer connector.

Do not allow disinfectant to contact metal surfaces. Use a soft cloth lightly dampened in a mild soap or compatible cleaning solution to remove any disinfectant that remains on metal surfaces.

Attempting to disinfect a transducer or transducer cable using a method other than the one included here can damage the transducer and void the warranty.

To clean and disinfect a transducer (wipe method)

1 Disconnect the transducer from the system.

2 Remove any transducer sheath.

3 Clean the surface using a soft cloth lightly dampened in a mild soap or detergent cleaning solution to remove any particulate matter or body fluids.

Apply the solution to the cloth rather than the surface.

4 Rinse with water or wipe with water-dampened cloth; then wipe with a dry cloth.

5 Mix the disinfectant solution compatible with the transducer, following disinfectant label instructions for solution strengths and disinfectant contact duration.

6 Wipe surfaces with the disinfectant solution.

7 Air dry.

8 Examine the transducer and cable for damage such as cracks, splitting, or fluid leaks.

If damage is evident, discontinue use of the transducer, and contact SonoSite or your local representative.

Chapter 7: Maintenance 55

To clean and disinfect a transducer (immersion method)

1 Disconnect the transducer from the system.

2 Remove any transducer sheath.

3 Clean the surface using a soft cloth lightly dampened in a mild soap or compatible cleaning solution to remove any particulate matter or body fluids.

Apply the solution to the cloth rather than the surface.

4 Rinse with water or wipe with water-dampened cloth, and then wipe with a dry cloth.

5 Mix the disinfectant solution compatible with the transducer, following disinfectant label instructions for solution strengths and disinfectant contact duration.

6 Immerse the transducer into the disinfection solution not more than 12-18 inches (31-46 cm) from the point where the cable enters the connector.

Follow the instructions on the disinfectant label for the duration of the transducer immersion.

7 Using the instructions on the disinfectant label, rinse to the point of the previous immersion, and then air dry or towel dry with a clean cloth.

8 Examine the transducer and cable for damage such as cracks, splitting, or fluid leaks.

If damage is evident, discontinue use of the transducer, and contact SonoSite or your local representative.

Cleaning and disinfecting the battery

Caution: To avoid damaging the battery, do not allow cleaning solution or disinfectant to come in contact with the battery terminals.

To clean and disinfect a battery (wipe method)

1 Remove the battery from the system.

2 Clean the surface using a soft cloth lightly dampened in a mild soap or detergent cleaning solution.

Apply the solution to the cloth rather than the surface.

3 Wipe the surfaces with the disinfection solution. Sani-Cloth HB, Sani-Cloth Wipes, or 70% isopropyl alcohol is recommended.

4 Air dry.

Cleaning the footswitch

Caution: To avoid damaging the footswitch, do not sterilize. It is not intended for use in a sterile environment.

To clean the footswitch

1 Dampen a non-abrasive cloth with one of the following products:

• Isopropyl alcohol

• Soap and water

• Cidex

• Sodium Hypochlorite 5.25% (Bleach) diluted 10:1

2 Wring out cloth until slightly wet and then gently rub soiled area until clean.

56 Chapter 7: Maintenance

Cleaning and disinfecting ECG cables

Caution: To avoid damaging the ECG cable, do not sterilize.

To clean and disinfect the ECG cable (wipe method)

1 Remove the cable from the system.

2 Clean the surface using a soft cloth lightly dampened in a mild soap or detergent cleaning solution.

Apply the solution to the cloth rather than the surface.

3 Wipe the surfaces with any of the following products:

• Bleach (sodium hypochlorite)

• Cidex disinfectants

• Green soap

4 Air dry or towel dry with a clean cloth.

Chapter 7: Maintenance 57

58 Chapter 7: Maintenance

Chapter 8: Performance Testing

Overview

WARNING: Critical Test Function — A failure of the system functions tested in this section could affect safety or effectiveness of the system adversely. While performing the steps in this section, verify that the images on the system display and on the external monitor are acceptable.

To obtain 2D images, SonoSite recommends using the Gammex 403GS Soft Tissue Phantom or the Gammex 413A

Multipurpose Phantom. A .7db/cm phantom is recommend but not required.

Some features and capabilities are optional and therefore may be unavailable to test.

Recommend Test Equipment

• SonoSite ultrasound system under test

• C60x/5-2 MHz transducer

• P21x/5-1 MHz transducer

• Gammex 403 GS Multipurpose Phantom, 413A Soft Tissue Phantom, or equivalent.

• Acoustic gel

Setting Up Performance Tests

Set up

Performance

Tests

1 Attach the C60x/5-2 MHz transducer to the system.

2 Select Gen for optimization and OB for exam type.

3 Couple the transducer to the phantom, adjusting gain settings and transducer for a proper phantom image (e.g., pins are high-level echoes positioned in straight lines; cysts are sonolucent, edges are sharp, and graphite particles of the phantom are mid-grays).

Chapter 8: Performance Testing 59

Basic Operational Tests

Basic System

Operation

Tests

1 Verify that the correct transducer name appears in the upper right corner of the system display.

2 Verify proper date and time.

3 Verify that the scan plane orientation mark in the image located near the skinline corresponds to element #1 on the transducer. To test, put your finger on the probe and run it across the transducer face. Your finger touching the transducer face should appear at the orientation mark on the display image format.

4 Verify that all of the keyboard keys are functional. Verify that all controls operate smoothly over their full range and that the system responds properly.

5 Verify that as the Gain controls are increased and decreased, there is a corresponding increase and decrease in echo intensity.

6 Capture a Cineloop buffer. Exercise the Cineloop controls and verify proper operation.

7 Close the lid and verify the unit goes into sleep mode. Open the lid and verify the unit returns to normal operation.

8 Verify the airflow from the vent on the left side of the system is blowing out.

2D Performance Tests

2D Performance / Image Quality

Test 2D

Performance and Image

Quality

1 Use a C60x/5-2 MHz transducer in 2D mode.

2 Adjust the position of the C60x/5-2 MHz transducer on the phantom.

3 With the array pointing down and the orientation mark to the operator’s left, element #1 corresponds with the left side of the array.

4 Use the 2D system controls to obtain a clear image that shows both the horizontal and vertical rows of pins.

5 Verify that the ultrasound image appears uniform in both the axial and lateral direction, with no dropouts or intensity variations.

6 Verify that the cystic structure at the focal zone is clearly differentiated from the surrounding tissue and is echo-free, while solid tissue with numerous echo sources, appears solid.

7 Press the Freeze key and then save the image. Press the Freeze key again to return to live imaging.

60 Chapter 8: Performance Testing

Axial Measurement Accuracy

Note: Measurements must be performed while the image is frozen.

Set Up Axial

Measurement

Accuracy

1 Acquire the image.

2 Press the Freeze key.

3 Press the Caliper key. The caliper appears on the image display. (See the Edge Ultrasound

System User Guide, if necessary, for caliper operation.)

4 Use the touchpad to position one of the calipers.

5 Press the Select key to fix the caliper and enable the other caliper.

6 Use the touchpad to move the other caliper. The results update as you move the caliper, and the measurement is complete when you finish moving the calipers. (Press the Select key to alternate the active caliper, and adjust the measurement with the touchpad.)

Test Axial

Measurement

Accuracy

1 Measure the distance, center to center, of any two pins that are 5-12 cm apart vertically.

2

Verify that the distance measured is within the tolerance listed in Table 8.1

.

Lateral Measurement Accuracy

Set Up Lateral

Measurement

Accuracy

Perform

“Set Up Axial Measurement Accuracy” on page 61

.

Test Lateral

Measurement

Accuracy

1 Measure the distance, center to center, of any two pins that are 4-10 cm apart horizontally.

2

Verify that the distance measured is within the tolerance listed in Table 8.1

.

3 Press the Freeze key to return the system to live 2D mode.

Table 8.1: System Measurement Accuracy

Measurements

Axial Distance

Lateral Distance

Tolerance

+/- 2%

+/- 2%

Chapter 8: Performance Testing 61

Penetration

The penetration measurement is an integral part of the quality assurance program. Penetration is defined as the deepest depth at which an ultrasound system can provide adequate image quality of small anatomical structures.

Penetration measurements should be performed and the results retained for comparison to future measurements.

Penetration measurements should remain fairly consistent over time assuming use of the same system settings and scanhead. Degradation of the penetration measurement in excess of 1cm may indicate a transducer or system electronics issue.

Loss of measured penetration may also be caused by degradation (dessication) of the ultrasound phantom.

Ultrasound phantoms used for penetration measurements must also be part of a quality assurance program to maintain their integrity. Follow all of the phantom manufacturer recommendations for use, storage, and maintenance of the phantom.

Test

Penetration

1 Use the same scanhead and system settings as previous measurements if possible.

2 Adjust the system controls to obtain a clear image that shows the limits of echo penetration.

3 Press the Freeze key and then save the image.

4 Measure from the center of the skinline to the deepest vertical position—where the scatter echoes start to break up and tissue definition is lost.

5 Record and retain the results for future reference. Scanhead type and system settings (exam type, depth, resolution mode, etc.) should also be recorded to ensure proper comparison with future tests.

6 Press the Freeze key again to return to live imaging.

Additional Performance Tests

Color Doppler (Color)

Test Color 1 Connect any transducer.

2 Press the Color key. “Color” should be annotated in the top left corner of the display.

3 A Region of Interest (ROI) box is displayed on top of the grayscale image. Use the touchpad to move the Color ROI. Verify that the ROI moves to the new position on the display.

4 Adjust the Depth control for minimum depth in the image.

5 Adjust the Gain control so that color speckles just appear inside the ROI box.

6 Gently tap the face of the transducer and observe that the ROI box fills with color information.

7 Press the Freeze key and then save the image. Press the Freeze key again to return to live imaging.

62 Chapter 8: Performance Testing

Color Power Doppler (CPD)

Test CPD

M Mode Imaging

1 Connect any transducer.

2 Press the Color key. A Region of Interest (ROI) box is displayed on top of the grayscale image.

3 Press the Color softkey to switch to CPD. “CPD” should be annotated in the top left corner of the display.

4 Adjust the Depth control for minimum depth in the image.

5 Adjust the Gain control so that color speckles just appear inside the ROI box.

6 Gently tap the face of the transducer and observe that the ROI box fills with color information.

Test M Mode

Imaging

1 Attach a C60x transducer and acquire an image.

2 Press the M Mode key for the M Mode sample line.

3 Position the M Mode sample line over the image using the touchpad.

4 Press the M Mode key again to turn on M Mode.

5 Select the desired sweep speed from the on-screen menu (slow, med, or fast). The on-screen menu will show the selected sweep speed.

6 Press the Freeze key to freeze the image. Save the image. Press the Freeze key again to return to live imaging.

7 Press the 2D key to return to 2D imaging.

Tissue Harmonic Imaging

Test THI

Imaging

1 Attach the C60x transducer and acquire an image.

2 Set the depth to maximum and note the depth at which echo information is lost.

3 Press the THI key on the control panel so it displays THI on the display. Tissue Harmonic

Imaging in now active.

4 Observe a decrease in dot size and a significant loss in penetration due to the higher frequency. Image resolution increases.

5 Press the Freeze key and then save the image. Press the Freeze key again to return to live imaging.

6 Press the THI key again to turn off Tissue Harmonic Imaging.

Chapter 8: Performance Testing 63

Pulsed Wave (PW) Doppler Imaging

Test PW

Doppler

Imaging

1 Attach the P21x transducer.

2 Press the Doppler key for the Doppler sample gate.

3 Press the Doppler key again for the Doppler spectral trace.

4 Place a large drop of ultrasound gel on the transducer lens.

5 Adjust the Gain control as necessary and then gently tap the top of the gel and observe a reflection on the spectral trace and the sound from the speakers.

6 Press the Freeze key and then save the image. Press the Freeze key again to return to live imaging.

7 Press the 2D key to return to 2D imaging.

Continuous Wave (CW) Doppler Imaging

Test CW

Doppler

Imaging

1 Attach the P21x transducer.

2 Press the Patient key.

3 Select the Cardiac exam type.

4 Press the Done softkey.

5 Press the Doppler key for the Doppler sample gate.

6 Press the PW softkey to switch to CW Mode.

7 Press the Doppler key again for the Doppler spectral trace.

8 Place a large drop of ultrasound gel on the transducer lens.

9 Adjust the Gain control as necessary and then gently tap the top of the gel and observe a reflection on the spectral trace and the sound from the speakers.

10 Press the Freeze key and then save the image. Press the Freeze key again to return to live imaging.

11 Press the 2D key to return to 2D imaging.

Image Quality Verification Test/Livescan

• Products with replaced subassemblies, or products that have been otherwise disassembled, must undergo an

Image Quality Verification Test/Livescan.

• The Image Quality Verification Test/Livescan should be performed after successfully completing all applicable performance tests listed prior in this chapter.

• The test is completed before returning the system to service.

• A certified sonographer must perform the test.

• The Livescan test performed is at the discretion of the Sonographer and will represent their acceptance of a successful service event.

• Review all saved images and verify that the images are displayed properly.

64 Chapter 8: Performance Testing

Printer

The printer test is an optional test that requires a video printer and minidock to be connected to the system under test. Skip this test if a printer and minidock are not available.

Test Printer

Operation

1 Connect the minidock to the system under test.

2 Connect the video output of the minidock to the printer.

3 Verify proper printer type is configured in the system Setups page.

4 Press the print button and verify that the printer begins to print an image. After the image begins to emerge from the printer, press the print button again. The printer should ignore the second print command.

5 Verify the proper content of the printed image.

Battery Charging

Test Battery

Charging

Operation

1 Remove the system from the docking system and insert a battery into the system.

2 Press the Power key to turn the system on. Allow the battery to discharge. The battery indicator icon on the display, below the Transducer Type indicator, will extinguish from left to right as the battery discharges.

Note: The Power and Sleep delays in the Setup page should be selected to “Off” to properly perform this test. The battery may take 1–2 hours to discharge.

3 Reattach the system to the Docking System and attach the AC power cord to the power connector.

4 Note that the battery indicator indicates that the battery is charging. The sections of the battery indicator will light sequentially from left to right as the battery charges.

Video Output

The video output test is an optional test that requires a minidock and external video monitor to be connected to the system under test. Skip this test if a minidock and external monitor are not available.

Test Video

Output

1 Connect the minidock to the system under test.

2 Connect the video output of the minidock to an external video monitor.

3 Turn on the system power and verify that the video on the external monitor matches the video on the system display.

If the video does not appear similar, or there is no display on the external monitor, see

Chapter 5, “Troubleshooting” for troubleshooting procedures.

Chapter 8: Performance Testing 65

66 Chapter 8: Performance Testing

Appendix A: Replacement Parts

The following tables contain all the field-replaceable parts for the EDGE ultrasound system. Quantities are one unless otherwise noted.

Display

Table A.1: Display

Part Number Description

P15637

P15638

V15637

V15638

Service Assembly, Display, Edge

Warranty Service Assembly, Display, Edge

Vet Service Assembly, Display, Edge

Vet Warranty Service Assembly, Display, Edge

Appendix A: Replacement Parts 67

Control Panel

P15618

P15621

P15622

P15623

P15624

P15625

P15626

Table A.2: Control Panel

Part Number Description

P15630

P15631

P15632

P15633

P15634

P15635

P15636

Warranty Service Assembly Control Panel, English, Edge

Warranty Service Assembly Control Panel, French, Edge

Warranty Service Assembly Control Panel, German, Edge

Warranty Service Assembly Control Panel, Italian, Edge

Warranty Service Assembly Control Panel, Spanish, Edge

Warranty Service Assembly Control Panel, Russian, Edge

Warranty Service Assembly Control Panel, Portuguese, Edge

Service Assembly Control Panel, English, Edge

Service Assembly Control Panel, French, Edge

Service Assembly Control Panel, German, Edge

Service Assembly Control Panel, Italian, Edge

Service Assembly Control Panel, Spanish, Edge

Service Assembly Control Panel, Russian, Edge

Service Assembly Control Panel, Portuguese, Edge

68 Appendix A: Replacement Parts

Main PCBA

1

4

5

3 2

Table A.3: Main PCBA

Item

1

2

3

4

5

Not shown

Part Number Description

P15615

P15617

V15615

V15617

P08850

P09541

P10168

P07442

P09542

Service Assembly Main PCBA, Edge

Warranty Service Assembly Main PCBA, Edge

Vet Service Assembly Main PCBA, Edge

Vet Warranty Service Assembly Main PCBA, Edge

Note: This part does not include the transducer nest frame assembly. Those parts must be ordered separately if needed to complete the replacement of the

Main PCBA.

Service Assembly, Power Supply, M-Turbo (compatible with Edge system)

Power Supply Shield

2GB SD Card (requires 4)

SD Card Daughter-card

Power Supply Shield Cover. Attaches to Item 3 Power Supply Shield

Appendix A: Replacement Parts 69

Miscellaneous Parts

Table A.4: Miscellaneous Parts

Part Number Description

P02308

Cables (images are not to scale)

FFC, 12 Position Jumper, 0.5 Pitch, 3” Length (3” Flat Flex Cable)

P14729

Note: This cable is used in two locations on the Control Panel PCB.

Cable Assy, Main to Control Panel PCB, Video

P14747 Cable, Main To Control Panel PCB, Backlight

P15410 FFC, 12 Position, 0.5MM Pitch, Opposite Side (8” Flat Flex Cable)

70 Appendix A: Replacement Parts

Table A.4: Miscellaneous Parts

Part Number Description

Other Electrical Parts

P02312 Fan Assembly

P03872 Speaker Assembly

P14727 USB Extension PCB Assy

P14742 Compression Connector Insulator, EDGE (Used with P15723)

P15273 Connector, Compression 166 Wires/Inch, 33.350 (Used with P14742)

P15984 Gasket, EMI/ESD (Used with P15304 - requires 2)

Appendix A: Replacement Parts 71

Table A.4: Miscellaneous Parts

Part Number Description

See Note

Mechanical/Cosmetic Parts

Bottom Enclosure

P09542

Note: Contact SonoSite Technical Support if it is necessary to replace the bottom enclosure. Ordering the bottom enclosure requires special handling due to the serial number label.

Power Supply Shield

P14750 Foot (4x)

72 Appendix A: Replacement Parts

Table A.4: Miscellaneous Parts

Part Number Description

P15305 Hinge Cover

P16156 Service Assembly, EDGE Top Enclosure

P15304 Enclosure, Middle, Machined, EDGE

P07885 Thermal Pad, 1.00 x 1.00 x 0.100” (Used with P15304 & P15615 - requires 4)

P07886 Thermal Pad, 0.35 x 0.35 x 0.100” (Used with P15304 & P15615 - requires 4)

Appendix A: Replacement Parts 73

Transducer Nest Frame Assembly

7

1

2 6

5

4

3

Figure A.1 Nest Frame Parts

Table A.5: Nest Frame Assembly

Find Number Part Number Description

3

4

1

2

5

6 not shown

7

P07750

P00364

P03833

P03834

P00924

P00353

P00646

P08200

Nest Frame Assembly

Connector, Interposer (8x)

Shield, Perimeter, Short (2x)

Shield, Perimeter, Long (2x)

Screw, Shoulder, Thrust Plate (4x)

Wear Plate

Spring, Thrust Plate, .047 wire (4x)

Socket Head Cap Screw, M2.5-.45x10mm (4x)

Ordering Replacement Parts

To order parts, contact SonoSite Technical Support as indicated in

“Contact Information” on page 1 .

74 Appendix A: Replacement Parts

Appendix B: Service Event Reporting

The Service Event Report provides information about product failures to the manufacturer and to authorized service facilities, which provide approved warranty services for SonoSite products. For all repairs completed, complete the form and email a copy of it to [email protected] or mail to the following address:

SonoSite, Inc.

Technical Support

21919 30th Drive SE

Bothell, Washington 98021

USA

To contact SonoSite Technical Support, see

“Contact Information” on page 1.

Appendix B: Service Event Reporting 75

Service Event Report Form

Service Event Report

Instructions on reverse

Choose a Part Status

Service Type

Enter complete

(check one)

† Out of Box Failure

† Warranty Service

† Out of Warranty Service contact information

†

†

†

†

†

Parts Status

(check one)

No parts necessary for this repair. Service

Event Report for your information.

I need parts for this repair (list the parts below and attach Purchase Order)

I need parts to replenish my stock (list the parts used below and attach Purchase Order)

Will not replenish stock. Please give me a

RMA for the return of the faulty parts.

No parts necessary. Please issue a RMA for repair at SonoSite.

For SonoSite Use O

Service Request

Order Number

RMA Number

Work Order

Service Provider

Name:

Company:

Address:

Phone Number:

E-mail address:

Device Description

Ref Number:

Enter product information for the system being repaired

Name:

ARM/SHDB Version: Configuration:

Problem Found

Service Performed

Enter DETAILED problem description here.

Put the system configuration here.

You can find this on the System

Information Page

Performed By:

Parts Removed

Part Name

Date:

Enter details for parts being removed from the system.

Parts Installed

Part Name

Part Number Serial Number Lot Number

Enter details for parts being installed into the system.

Part Number

Rev Replace

Serial Number Lot Number Rev Replace

Tests Performed (attach test data)

Test: Test:

76 Appendix B: Service Event Reporting

Service Event Report Instructions

Instructions for completing the Service Event Report

Sections highlighted in yellow must be completed for SonoSite to accept the Service Event Report. If additional information is required for certain circumstances you will be advised.

Forward the completed form to:

Email: [email protected]

Fax: +1-425-951-6700

Service Type x Out of Box Failure: the item has arrived from SonoSite with failures. x Warranty Service: the item has failed after arrival and is covered by either the included warranty or a valid extended warranty. x Out of Warranty Service: the item has failed and it is no longer covered by a warranty.

Parts Status x Check One.

Service Provider x Name: the name of the technician performing the work. x Provider Reference: a unique number used by the Provider to track Service Event Reports. Any format is acceptable. x Company: the name of the Distributor or authorized repair facility. x Address: the address replacement parts will be shipped to. x Date Reported: the date the failure was reported to SonoSite. x Phone Number: the phone number to contact the service technician. x Fax Number: the fax number to contact the service technician. x Email Address: the email address to contact the service technician.

Device Description: x Name: the description of the failed product. x Ref Number: the reference number from the part number label of the failed product. x Serial Number: the serial number from the part number label of the failed product. x Lot Number: if applicable, the Lot Number from the device identification label. x ARM/SHDB Version: the software level of the failed device. Typically found on the system information screen. x Configuration: for configurable devices, the optional features enabled.

Event Description x A description of the problem in the words of the user. Typically what the user reports to the repair facility.

Diagnosis x A description of what the repair technician found. Include a list of the suspect parts.

Service Performed x A description of the work performed to repair the system. Typically only completed if it is repaired from stock repair parts.

Parts Removed x Part Name: the name of the failed/suspect part to be replaced. x Part Number: the part number of the failed/suspect part. x Serial Number: the serial number from the failed/suspect part. x Lot Number: the lot number if applicable. x Rev: the revision of the failed/suspect part if available. x Replaced By: the person replacing the part.

Parts Installed x The same information as the Parts Removed except from the parts installed if work has already been performed. If you are waiting for parts to be ordered, leave this section blank.

Tests Performed x The results of any testing performed, if testing has already been performed.

Appendix B: Service Event Reporting 77

Returning Products to SonoSite

You will be asked to provide the following information:

• Contact name and phone number

• Product name

• Serial number

• Description of the problem

Shipping Instructions

Please contact SonoSite to get a return material authorization number (RMA). Contact SonoSite before returning any product.

The shipping address for all returned products is:

SonoSite, Inc.

Attn: Technical Support RMA ___________________

21919 30th Drive SE

Bothell, Washington 98021

USA

78 Appendix B: Service Event Reporting

INDEX

Numerics

2D performance tests

axial measurement accuracy 61

image quality

61

lateral measurement accuracy

61

penetration 62

A

ALARA

13

assert code

36

assistance, customer 1

B

battery clean

56

safety

12

specifications 6

battery charging test

65

biological safety

13

bottom enclosure replacement 52

C

cables

clean and disinfect ECG 57

cautions, definition 1

clean battery

56

ECG cable

57

footswitch

56

LCD screen

54 system 54

transducers 55

cleaners, recommended

53

control panel subassembly replacement

44

D

disinfect battery

56

ECG cable

57

system 54

transducers 55

disinfectants, recommended 53

display removal 39 display replacement 39, 43

E

electrical safety

9

specifications 5

electromagnetic compatibility 14

equipment safety

11

error message 10

F

failure reporting

75

79

H

humidity limits 5

I

image quality verification test

64

review

65

L

LCD screen clean

54

M

main PCBA failures

36

main system disassembly

45

middle machined enclosure 52

P

performance tests

2D 60

battery

65

CPD 63

CW 64

M-Mode 63

overview

59

printer

65

PW

64

THI

63

Velocity Color 62

video output 65

periodic maintenance 53

power supply PCBA replacement 48

pressure limits

5

printer test

65

product failures 75

R

replacement parts list

67

ordering

74

return material authorization number (RMA) 78

returning products

78

S

safety battery

12

biological

13

electrical

9

electromagnetic compatibility 14

equipment 11

SD card daughter-card replacement

49

service event report

75

shipping instructions

78

shipping specifications 5

speaker replacement 47

80

storage specifications

equipment 5

subassembly replacement

35

system clean and disinfect

54

measurement accuracy

61

overview

21

T

temperature limits 5

theory of operation 22

transducer clean and disinfect

55

disinfect

55

U

USB extension PCB replacement

50

user guide, conventions used

1

V

video output tests 65

W

warnings, definition

1

81

82

P15644-02

*P15644-02*