Service manual | Bunn AFPO-2 SL Water Dispenser User Manual

BUNN
®
1
2
FILL
RE
A INR
E
M
W
PO
AFPO-2
AFPO-2 SL
AFPO-3
AFPO-3 SL
R
TE
WAP U T
IN
1
OR
AV T
FL TPU
OU
2
1
OR
AV
FL PUT
IN
2
R
TE T
WA P U
T
OU
OR
AV T
FL TPU
OU
R
TE T
WA P U
T
OU
1
RE
3
2
FILL
IN
MA ER
W
PO
R
TE
WAP U T
IN
1
1
OR
AV
FL PUT
IN
OR
AV T
FL TPU
OU
R
TE T
WA P U
T
OU
2
2
3
OR
AV T
FL TPU
OU
3
R
TE T
WA P U
T
OU
OR
AV T
FL TPU
OU
R
TE T
WA P U
T
OU
OPERATING & SERVICE MANUAL
BUNN-O-MATIC CORPORATION
POST OFFICE BOX 3227
SPRINGFIELD, ILLINOIS 62708-3227
PHONE: (217) 529-6601 FAX: (217) 529-6644
28791.0000E
4/01 ©1998 Bunn-O-Matic Corporation
www.bunnomatic.com
INTRODUCTION
This equipment supplies a liquid concentrate mixed with water at various ratios to the hoppers on the CDS-2 or
CDS-3 dispensers. It is for indoor use only on a sturdy counter or shelf.
WARRANTY
Bunn-O-Matic Corp. (“Bunn”) warrants the equipment manufactured by it to be commercially free from defects
in material and workmanship existing at the time of manufacture and appearing within one year from the date of
installation. In addition:
1.) Bunn warrants electronic circuit and/or control boards to be commercially free from defects in material and
workmanship for two years from the date of installation.
2.) Bunn warrants the compressor on refrigeration equipment to be commercially free from defects in material
and workmanship for two years from the date of installation.
3.) Bunn warrants that the grinding burrs on coffee grinding equipment will grind coffee to meet original factory
screen sieve analysis for three years from date of installation or for 30,000 pounds of coffee, whichever comes first.
This warranty does not apply to any equipment, component or part that was not manufactured by Bunn or that,
in Bunn’s judgement, has been affected by misuse, neglect, alteration, improper installation or operation, improper
maintenance or repair, damage or casualty.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY OTHER WARRANTY, WRITTEN OR
ORAL, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF EITHER
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The agents, dealers or employees of Bunn are
not authorized to make modifications to this warranty or to make additional warranties that are binding on Bunn.
Accordingly, statements by such individuals, whether oral or written, do not constitute warranties and should not
be relied upon.
The Buyer shall give Bunn prompt notice of any claim to be made under this warranty by telephone at (217)
529-6601 or by writing to Post Office Box 3227, Springfield, Illinois, 62708-3227. If requested by Bunn, the Buyer
shall ship the defective equipment prepaid to an authorized Bunn service location. If Bunn determines, in its sole
discretion, that the equipment does not conform to the warranty, Bunn shall repair the equipment with no charge
for parts during the warranty period and no charge for labor by a Bunn Authorized Service Representative during
the warranty period. If Bunn determines that repair is not feasible, Bunn shall, at its sole option, replace the
equipment or refund the purchase price for the equipment.
THE BUYER’S REMEDY AGAINST BUNN FOR THE BREACH OF ANY OBLIGATION ARISING OUT OF THE SALE
OF THIS EQUIPMENT, WHETHER DERIVED FROM WARRANTY OR OTHERWISE, SHALL BE LIMITED, AS
SPECIFIED HEREIN, TO REPAIR OR, AT BUNN’S SOLE OPTION, REPLACEMENT OR REFUND.
In no event shall Bunn be liable for any other damage or loss, including, but not limited to, lost profits, lost sales,
loss of use of equipment, claims of Buyer’s customers, cost of capital, cost of down time, cost of substitute
equipment, facilities or services, or any other special, incidental or consequential damages.
USER NOTICES
Carefully read and follow all notices on the equipment and in this manual. They were written for your protection.
All notices are to be kept in good condition. Replace any unreadable or damaged labels.
00656.0000
00986.0002
2
28791 051599
ELECTRICAL REQUIREMENTS
CAUTION - The dispenser must be disconnected from the power source until specified in Initial Set-Up.
The 120 volt version of this dispenser has an attached cordset and requires 2-wire, grounded service rated 120
volts ac, 15 amp, single phase, 60 Hz. The mating connector must be a NEMA 5-15R.
The 230 volt CE version requires 2 wire, grounded service rated 230V ac, single phase, 50 Hz.
The 200 volt or 240 volt versions require 2 wire, grounded service rated 200 volts ac or 240 volts ac, single
phase, 50 Hz.
(Refer to the dispenser’s data plate for exact voltage requirement.)
Electrical Hook-Up
CAUTION – Improper electrical installation will damage electronic components.
1. An electrician must provide electrical service as specified.
2. Using a voltmeter, check the voltage and color coding of each conductor at the electrical source.
3. Connect the dispenser to the power source.
PLUMBING REQUIREMENTS
This dispenser must be connected to a cold water system with operating pressure between 20 and 90 (138
and 620 kPa) psi from a 1⁄2" or larger supply line. A shut-off valve should be installed in the line before the
dispenser. Install a regulator in the line when pressure is greater than 90 psi (620 kPa) to reduce it to 50 psi (345
kPa). The water inlet fitting is 1⁄4" flare.
NOTE - Bunn-O-Matic recommends 1⁄4" copper tubing for installations of less than 25 feet and 3⁄8" for more than
25 feet from the 1⁄2" water supply line. At least 18 inches of an FDA approved flexible beverage tubing, such as
reinforced braided polyethylene or silicone, before the brewer will facilitate movement to clean the countertop.
Tubing is supplied by Bunn-O-Matic (part number 00326-0000). Bunn-O-Matic does not recommend the use of
a saddle valve to install the brewer. The size and shape of the hole made in the supply line by this type of device
may restrict water flow
This equipment must be installed to comply with the Basic Plumbing Code of the Building
Officials and Code Administrators International, Inc. (BOCA) and the Food Service Sanitation Manual of the Food and Drug Administration (FDA).
INITIAL SET-UP
NOTE: This refill unit should not be used to initially fill the hopper. Running the motor for extended periods will
trip the thermal switch. The thermal switch will reset, but requires up to 30 minutes of cooling time.
NOTE: Hoses and product container connectors are not supplied. Fitting, clamps and labels are supplied. Make
hose assemblies as required. Refer to PIPING DIAGRAM on page 22.
1. Assemble hoppers with probe housing assemblies to CDS-2, CDS-3 or modify existing hoppers using template
provided.
2. Make sure probe housing assembly is pushed down into slots on back of hoppers.
3. Install cable assembly to the lower right front of the auto refill box and the probe housing assemblies on the
hoppers. The connectors on the cable pushes in and cap rotates 1/4 turn to lock into position.
4. Run output tubes from auto refill box to the quick disconnects in the hoppers.
NOTE: Make sure numbers on the labels match up. Cable#1 with Hose#1, Cable#2 with Hose#2,....etc.
3
28791 040501
INITIAL SET-UP (cont.)
5. Connect one end of input hoses to the auto refill box and the other end to the product container. Product
container connectors (BIB) are not supplied. Install your own interface.
6. Connect water line to auto refill box. A minimum of 20 psi (138 kPa) is required.
NOTE: Make sure all connections are made before connecting the auto refill box to the power source.
7. Each product container has an ON/OFF refill switch located on the front of the auto refill box. A second switch
for each hopper is at the probe housing assembly. These are 3 position switches, high level, low level and a
center “OFF” position.
8. Place the main ON/OFF switch and ON/OFF refill switches in the “OFF” position.
9. Make sure all hoses are connected. Place the main ON/OFF switch in the “ON” position.
10. Place one of the ON/OFF refill switches in the “ON” position, after a four second delay product will begin to flow
through the hoses into the hopper. Be sure the switch selected is filling the correct hopper, if not, check the
labels. See NOTE in step #4.
11. To test refill circuit short out probes on one of the hoppers with a screwdriver, so that hopper should shut off.
Remove the screwdriver and the hopper will start to fill again after a four second delay. If hopper does not start
to fill, check labels, refer to NOTE in step #4.
12. To change mix ratio refer to MIX RATIO ADJUSTMENTS.
CLEANING
RECOMMENDED WEEKLY CLEANING:
This should be done in conjunction with the recommended weekly cleaning of your CDS-2 or CDS-3 machine. The
CDS Hoppers must be empty before starting.
1. Prepare a 1 gallon cleaning solution consisting of 1 gallon of hot water and a sanitizing cleaner which contains
3-5% chlorine based sanitizer. Mix per manufacturers instructions
2. Remove connectors from each concentrate container and place them directly into the cleaning solution.
3. Energize each refill station and allow the cleaning solution to pump through the system into the hoppers. The
cleaning procedure should remove the color stains from the tubing. If not, prepare another gallon of cleaning
solution and repeat.
4. After the cleaning solution has been pumped into the hoppers, rinse the cleaning solution container with hot
water.
5. Fill the cleaning solution container with hot water. Energize each refill station and rinse thoroughly. The amount
of clean hot rinse water should equal the amount of cleaning solution pumped through the system.
6. Empty the hoppers of your CDS machine and follow the recommended cleaning instructions.
NOTE: The hoses should be checked monthly for deterioration, cracks and possible discoloration from some
concentrates. Replace the tubing when necessary.
MIX RATIO ADJUSTMENTS
1.
2.
3.
4.
Remove the top access panel.
Connect the test hose to the “WATER OUTPUT” to be adjusted.
Connect the provided hose (or other suitable water hose) to the “WATER INPUT”.
Ensure water regulator is set at 20 psi(137.9 kPa). Press the “TEST OUTPUT” switch and while water is flowing
from the test hose, check that the gauge reads 20.
5. Place a one-cup container under the test hose. Press the “TEST OUTPUT” switch, and carefully measure and
note the elapsed time from when water first flows form the test hose to when one cup of water has been
dispensed. Hoses and tubing should be primed and free of substantial air spaces and/or bubbles prior to
calibration.
4
28791 040501
MIX RATIO ADJUSTMENTS(cont.)
6. Compare results with chart or formula below.
7. If the time observed is too short or long, turn the needle valve clockwise or counterclockwise, respectively, a
little at a time. Repeat from the preceding step until correct.
Mix ratio (water/flavor)
1+1
2+1
3+1
4+1(see below)
5+1
6+1
Time for FLAVOR OUTPUT (1/2 cup)
assume 20 sec. or obtain time per note below
assume 20 sec. or obtain time per note below
assume 20 sec. or obtain time per note below
assume 20 sec. or obtain time per note below
assume 20 sec. or obtain time per note below
assume 20 sec. or obtain time per note below
Time for WATER OUTPUT (1 cup)
40 sec. or 2 times flavor output time
20 sec. or equal to flavor output time
13.3 sec. or 0.67 times flavor output time
10 sec. or 0.5 times flavor output time
8 sec. or 0.4 times flavor output time
6.7 sec. or 0.33 times flavor output time
NOTE: For a typical adjustment, the time for flavor output in the center column above is assumed to be 20
seconds. A more specific flavor pump output time may be obtained prior to adjustment of the needle valve.
To do so, connect the flavor to be tested to the “FLAVOR INPUT”, and the test hose to the corresponding
“FLAVOR OUTPUT” (make no connection to the “WATER INPUT”). With a one-cup container under the test
hose press the corresponding “TEST OUTPUT” switch, and carefully measure and note the elapsed time from
when the product first flows from the test hose to when one-half cup of the product has been dispensed. This
value may then be used in the formula below. Following flavor pump output measurement, disconnect the
flavor line from the “FLAVOR INPUT” and purge remaining product from the line.
Formula for setting mix ratio: To obtain the time for dispensing 1 cup of water, divide the time it
takes to dispense one-half cup of product by the first number in the mix ratio and then multiply the
result by two.
Example: W=2(F/X) Where W is the time it should take to dispense one cup of water, it took F
seconds (20 here) to dispense 1/2 cup of flavor concentrate, we desire a mix ratio X of 4+1 (X will
be the first number in the mix ratio, 4).
Therefore:
W = 2/(20/4)
W = 2/5
W = 10 seconds
So if it took 20 seconds to dispense one-half cup of flavor concentrate, and a mix ratio of 4 to 1
is required, the needle valve should be adjusted to dispense 1 cup of water in 10 seconds.
NOTE: Hoses and tubing should be primed and free of substantial air spaces and/or bubbles prior
to calibration.
1
RE
3
FIL
2
L
IN
MA R
WE
PO
R
TE
WA UT
INP
1
1
R
VO
FLA UT
INP
OR
AV UT
FL TP
OU
2
R
TE
WATP UT
OU
2
3
OR
AV UT
FL TP
OU
3
R
TE
WATP UT
OU
OR
AV UT
FL TP
OU
R
TE
WATP UT
OU
P1521
NEEDLE VALVE
5
28791 040501
OPERATING CONTROLS
C
B
1
A
2
L
FIL
RE
3
IN
MA ER
W
PO
R
TE
WAP U T
IN
1
OR
AV T
FL TPU
OU
1
OR
AV
FL PUT
IN
P1949
2
A. MAIN POWER ON/OFF
The main power ON/OFF switch is located on the right of the front panel just above the power cord.
B. REFILL CIRCUIT ON/OFF
The refill circuit ON/OFF switches are located on the right of the front panel just above the main power switch.
These allow each circuit to operate independently.
C. TEST/OPERATE
The TEST/OPERATE switch is located inside the autofill box just below the top access cover. For setting flow
rates and mix ratios for flavors, place the TEST/OPERATE switch in the "TEST" position. This allows the operator to run the water and flavor lines for each circuit
depending on the requirements for the flavor recipe.
D
D. PROBE CONTROL BOX LOW/OFF/HIGH
On models equipped with a dual level probe, there is a
LOW/OFF/HIGH switch located at the probe control box
on the hopper. The "OFF" position allows the operator
to disable the probe for that hopper only. THE "LOW"
and "HIGH" positions allow the operator to select the
level of product desired for that hopper.
P1951
6
28791 040501
TROUBLESHOOTING
A troubleshooting guide is provided to suggest probable causes and remedies for the most likely problems
encountered. If the problem remains after exhausting the troubleshooting steps, contact the Bunn-O-Matic
Technical Service Department.
•
•
•
•
•
•
•
Inspection, testing, and repair of electrical equipment should be performed only by qualified service personnel.
All electronic components have 120 volt ac and low voltage dc potential on their terminals. Shorting of
terminals or the application of external voltages may result in board failure.
Intermittent operation of electronic circuit boards is unlikely. Board failure will normally be permanent. If an
intermittent condition is encountered, the cause will likely be a switch contact or a loose connection at a
terminal or crimp.
Solenoid removal requires interrupting the water supply to the valve. Damage may result if solenoids are
energized for more than ten minutes without a supply of water.
The use of two wrenches is recommended whenever plumbing fittings are tightened or loosened. This will help
to avoid twists and kinks in the tubing.
Make certain that all plumbing connections are sealed and electrical connections tight and isolated.
Keep away from combustibles.
WARNING – •
•
•
•
Exercise extreme caution when servicing electrical equipment.
Unplug the dispenser when servicing, except when electrical tests are specified.
Follow recommended service procedures
Replace all protective shields or safety notices
PROBLEM
PROBABLE CAUSE
REMEDY
Product will not dispense
1. No water
Water lines and valves to the dispenser must be open.
2. No power or incorrect voltage to (A1) Check the outlet for 120 volts
on two wire 120 volt dispensers.
the dispenser
(A2) Check the outlet for 200 volts
or 240 volts ac for two wire 200
volt or 240 volt dispensers.
(A3) Check the outlet for 230 volts
on two wire 230 volt dispensers.
(B) Check circuit breakers or fuses.
3. Signal cable unplugged
Signal cable must be plugged in.
4. Hopper is full
Product must not be touching
probes to fill.
5. Fuse and fuse holder
Refer to Service - Fuse and Fuse
Holder for test procedures. See
page 11
7
28791 040501
TROUBLESHOOTING (cont.)
PROBLEM
PROBABLE CAUSE
REMEDY
Product will not dispense (cont.)
6. Main ON/OFF switch
Refer to Service - Main ON/OFF
Switch for testing procedure. See
page 12
7. Probe box LOW/OFF/HIGH Refer to Service - Probe system. See
switches
page 13
Incorrect mix ratio
8. Circuit board
Refer to Service - Circuit Board for
testing procedures. See page 10
9. Vacuum switch
Refer to Service - Vacuum switch for
testing procedures. See page 21
10. Vacuum/Supply pump
Refer to Service - Vacuum/Supply
pump for testing procedures. See
page 19
11. Pump (flavor) start switches
Refer to Service - Pump (flavor) start
switches for test procedures. See
page 14
1. No water
Water lines and valves to the dispenser must be open.
2. Needle valve
Refer to Setup - Mix Ratio Adjustments for adjustment. See page 4
3. Solenoids
Refer to Service - Solenoids for test
procedures. See page 16
Motor thermal cutout (motor stops 1. Motor overheat
running)
The thermal switch will reset, but
requires up to 30 minutes of cooling
time. See NOTE below
NOTE: This refill unit should not be used to initially fill the hopper. Running the motor for extended periods will
trip the thermal switch. The thermal switch will reset, but requires up to 30 minutes of cooling time.
8
28791 040501
SERVICE
This section provides procedures for testing and
replacing various major components used in this
dispenser should service become necessary. Refer to
Troubleshooting for assistance in determining the
cause of any problem.
1
WARNING - Inspection, testing, and repair of electrical equipment should be performed only by qualified
service personnel. The dispenser should be unplugged when servicing, except when electrical tests
are required and the test procedure specifically states
to plug in the dispenser.
RE
3
2
FILL
N
M AIER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV T
FL PU
O UT
2
ERT
WATPU
O UT
2
3
OR
AV T
FL PU
O UT
3
ERT
WATPU
O UT
OR
AV T
FL PU
O UT
ERT
WATPU
O UT
COMPONENT ACCESS
WARNING - Disconnect the dispenser from the power
source before the removal of any panel or the replacement of any component.
FIG. 1 COMPONENT ACCESS
P1521
All components are accessible by removing the
eight #8-32 screws (five on the front and three on the
rear) securing the main housing to the main mounting panel and the removal of the four #8-32 nuts securing the pump/solenoid switch panel to the solenoid mounting bracket. Refer to Fig. 1
Refer to the contents listing for component location.
Contents
Circuit Board ......................................................... 10
Fuse & Fuse Holder .............................................. 11
Main ON/OFF Switch ............................................. 12
Probe LOW/OFF/HIGH Switches ........................... 13
Pump (Flavor) Start Switches ............................... 14
Refill ON/OFF Switches ........................................ 15
Solenoid ............................................................... 16
Solenoid (Water) Start Switches ........................... 17
Test/Operate Switch.............................................. 18
Vacuum/Supply Pump Assembly .......................... 19
Vacuum Switches ................................................. 21
Piping Diagram ..................................................... 22
Wiring Diagrams ................................................... 23
9
28791 040501
SERVICE
CIRCUIT BOARD
c) 230 volts ac for two wire 230 volt models.
4. Disconnect the dispenser from the power source.
If voltage is present as described, replace the circuit
board.
If voltage is not present as described, refer to the
Wiring Diagrams and check main wiring harness.
FFO
Removal and Replacement:
1. Remove the four #8-32 keps nuts securing the
switch mounting bracket. Set aside with the wires
attached.
2. Remove the circuit board shield and set aside for
reassembly.
3. Disconnect the ten pin plug from the main wiring
harness to the circuit board.
4. Remove the six #4-40 screws and spacers securing the circuit board to the solenoid mounting
bracket. Remove circuit board and discard.
5. Install new circuit board using #4-40 screws and
spacers to secure the circuit board to the solenoid
mounting bracket. The spacers must be between
the solenoid mounting bracket and the circuit
board.
6. Place circuit board shield over the circuit board.
7. Install switch panel assembly and secure with four
#8-32 keps nuts.
8. Refer to Fig. 3 for circuit board ten pin connector.
NO
FFO
NO
FFO
NO
FFO
NO
FIG. 2 CIRCUIT BOARD
P1520
Location:
The circuit board is located inside the autofill box,
mounted on the right side of the solenoid mounting
bracket under the switch panel.
J4
J1
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect the ten pin plug on the main wiring
harness from the connector on the circuit board.
3. Connect the dispenser to the power source. With
a voltmeter, check the voltage across the black
wire(#3) and the white wire (#6). The indication
must be:
a) 120 volts ac for two wire 120 volt models.
b) 200 to 240 volts ac for two wire 200 or 240 volt
models.
J2
J3
P1528
FIG. 3 CIRCUIT BOARD CONNECTOR LOCATION
10
28791 040501
SERVICE
FUSE AND FUSE HOLDER
Removal and Replacement:
Fuse:
1. Remove the cap from the fuse holder.
2. Remove fuse from the fuse holder and inspect. If
blown, discard.
3. Install new 10 amp fuse in the fuse holder.
4. Reinstall fuse holder cap.
1
RE
3
2
FILL
Fuse Holder:
1. Disconnect the dispenser from the power source.
2. Disconnect the wires from the rear of the fuse
holder.
3. Remove the nut securing the fuse holder to the
front of the main mounting panel.
4. Push the fuse holder through the hole in the panel.
5. Install new fuse holder and fuse through the hole
in the panel.
6. Secure the fuse holder to the panel.
7. Refer to Fig. 5 and reconnect the wires.
N
M AIER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV T
FL PU
O UT
2
ER
T
WATPU
O UT
2
3
OR
AV T
FL PU
O UT
3
ER
T
WATPU
O UT
OR
AV T
FL PU
O UT
ER
T
WATPU
O UT
FIG. 4 FUSE HOLDER
P1516
WHI/GRN from Main
Harness/ON/OFF
Switch
BLK from Main
Location:
The fuse holder is located on the lower right front
of the autofill box just to the left of the power cord.
Wiring Harness
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Remove cap and fuse from fuse holder.
3. Remove fuse from the cap.
4. Check for continuity through the fuse.
WHI/GRN from Main
BLK to Main Board
Harness/Control
If continuity is present, reinstall the fuse, the fuse is
operating properly.
If continuity is not present, replace the fuse.
ON/OFF Switch
FIG. 5 FUSE HOLDER TERMINALS
11
P1331
28791 040501
SERVICE (cont.)
5. Check for continuity across the switch terminals
with the switch in the “ON” position. Continuity
must not be present when the switch is in the
“OFF” position.
MAIN ON/OFF SWITCH
If continuity is present as described, reconnect the
wires to the switch terminals.
If continuity is not present as described, replace the
switch.
1
RE
3
2
FILL
N
M AIER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV T
FL PU
O UT
2
ER
T
WATPU
O UT
2
Removal and Replacement:
1. Remove the wires from the switch terminals.
2. Compress the clips inside the autofill box and
gently push the switch through the opening
3. Push the new switch into the opening and spread
the clips to hold switch in the autofill box.
4. Refer to Fig. 7 when reconnecting the wires.
3
OR
AV T
FL PU
O UT
3
ER
T
WATPU
O UT
OR
AV T
FL PU
O UT
ER
T
WATPU
O UT
FIG. 6 MAIN ON/OFF SWITCH
P1516
Location:
The main ON/OFF switch is located on the lower
right front of the autofill box just above the power
cord.
BLK from Fuse Holder
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect the black wire from the fuse holder
and the white wire from pump #1.
3. Connect the dispenser to the power source. Check
for voltage across the black wire removed from the
switch and the white wire removed from pump #1.
The indication must be:
a) 120 volts ac for 2 wire 120 volt models.
b) 200 to 240 volts ac for two wire 200 or 240 volt
models.
c) 230 volts ac for two wire 230 volt models.
4. Disconnect the dispenser from the power source.
BLK to Solenoid (Water)
Switch #2 AFPO-2 & 3
FIG. 7 MAIN ON/OFF SWITCH TERMINALS
P1522
If voltage is present as described, proceed to #5.
If voltage is not present as described, refer to the
Wiring Diagrams and check the dispenser wiring harness.
12
28791 040501
SERVICE (cont.)
PROBE SYSTEM
Probe Boxes:
1. Disconnect the signal cable from the probe boxes.
2. Remove probe box from CDS hopper.
3. Place the switch on the probe box in the “HIGH”
position.
4. Check the resistance across the box cable pins,
see Fig. 10.
1
Mov
ing
Par
ts.
Do
not
unit ope
pan with thisrate
el rem
ove
d.
2
Ris
k Of
Ele
ctri
Disc
cal
Sho
servonnect
ck.
icin pow
g unit er
befo
.
re
De Typ
sig e
CH
n PreR2 AR
2,
ssu Am GE
Su
ctio res: oun
Hig t
n
Pre h _ _ OZ
ssu _ _
.
re
Low
__
_
300Ω
®
N
N
BU
_
AFPO-2 & 3
P1539
FIG. 10 PROBE BOX CABLE PINS
5. With the probe box switch in the “LOW” position
check the resistance across the pins, refer to Fig.
10.
6. If readings match the values indicated, the probe
system is functioning properly.
P1523
FIG. 8 PROBE SWITCHES
Location
The probe switches are located on the rear of each
hopper, mounted in the probe housing cover.
Switch Removal and Replacement:
1. Remove the #6 thread cutting screw securing the
probe housing cover to the probe housing.
2. Remove the wires from the switch terminals.
3. Compress the clips inside the probe box and
gently push the switch through the opening
4. Push the new switch into the opening and spread
the clips to hold switch in the autofill box.
5. Refer to Fig. 11 when reconnecting the wires.
Test Procedure
Probe System
1. Disconnect the dispenser from the power source.
2. With the probe boxes connected to the signal
cable, remove connector from refill box.
3. Place the switch on the probe box in the “HIGH”
position.
4. Check resistance across pins as shown in Fig. 9.
300Ω
300Ω
1
2
600Ω
4
3
300Ω
1
2
BLU (HIGH) from Probe
Box Cable
600Ω
300Ω
4
3
600Ω
AFPO-2
AFPO-3
FIG. 9 SIGNAL CABLE PINS
GRN (OFF) from Probe
Box Cable
P1538
TAN (LOW) from Probe Box
Cable
5. With the probe box switch in “LOW” position
check the resistance across the signal cable pins,
see Fig. 9.
6. If readings match values indicated, the probe
system is functioning properly. If not, check the
probe boxes.
FIG. 11 PROBE LOW/OFF/HIGH SWITCH
TERMINALS
13
P1540
28791 040501
SERVICE (cont.)
PUMP (FLAVOR) ON/OFF SWITCHES
3. Compress the clips inside the autofill box and
gently push the switch through the opening.
4. Push the new switch into the opening and spread
the clips to hold switch in the autofill box.
5. Refer to Fig. 13 when reconnecting the wires.
1
2
L
FIL
RE
3
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV UT
FL TP
OU
2
ER
UT
WAT
TP
OU
2
3
OR
AV UT
FL TP
OU
3
ER
UT
WAT
TP
OU
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
#1 AFPO-2 & 3
P1521
BLK from Solenoid Switch #1
BLK to Test/Operate Switch
RED/BLK to Test/Operate Switch
YEL to Pump #1
FIG.12 PUMP (FLAVOR) START SWITCHES
Location:
The pump (flavor) START switches are located on
the switch mounting plate inside the autofill box on the
right side.
#2 AFPO-2 & 3
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect the wires from the switch to be tested.
3. Check for continuity across the terminals on the
switch when it is held in the on position. Continuity
must not be present across these terminals in the
released position.
BLK from Solenoid Switch #1
BLK to Solenoid Switch #2
BRN/BLK to Test/Operate Switch
ORN to Pump #2
#3 AFPO-3 ONLY
If continuity is present as described, reconnect the
wires to the switch.
If continuity is not present as described, replace the
switch.
BLK from Solenoid Switch #2
BLK to Solenoid Switch #3
WHI/RED to Test/Operate Switch
RED to Pump #3
Removal and Replacement:
1. Disconnect the dispenser from the power source.
2. Remove the wires from the switch terminals.
P1524
FIG. 13 PUMP (FLAVOR) SWITCH TERMINALS
14
28791 040501
If continuity is present as described, reconnect the
wires to the switch terminals.
If continuity is not present as described, replace the
switch.
Removal and Replacement:
1. Remove the wires from the switch terminals.
2. Compress the clips inside the autofill box and
gently push the switch through the opening in the
front of autofill box.
3. Push the new switch into the opening and spread
the clips to hold switch in the autofill box.
4. Refer to Fig. 15 when reconnecting the wires.
SERVICE (cont.)
REFILL SWITCHES
1
RE
3
FIL
2
L
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV
FL TP UT
OU
2
ER
WAT UT
TP
OU
2
3
OR
AV
FL TP UT
OU
3
ER
WAT UT
TP
OU
OR
AV
FL TP UT
OU
ER
WAT UT
TP
OU
#1 AFPO-2 & 3
FIG. 14 REFILL SWITCHES
RED/BLK from Test/
Operate Switch
RED/BLK to Vacuum
Switch #1
P1516
Location:
The refill switches are located on the upper right
front of the autofill box.
#2 AFPO-2 & 3
Test Procedures:
1. Disconnect the dispenser from the power source.
2. Disconnect the two white/red, brown/black or red/
black wires from the switch terminals.
3. Connect the dispenser to the power source. Check
for voltage across the white/red, brown/black or
red black wires from the test/operate switch and
the white wire on solenoid #2 or #3. The indication
must be:
a) 120 volts ac for two wire 120 volt models
b) 200 to 240 volts ac for two wire 200 or 240 volt
models.
c) 230 volts ac for two wire 230 volt models.
4. Disconnect the dispenser from the power source.
BRN/BLK from Test/
Operate Switch
BRN/BLK to Vacuum
Switch #2
#3 AFPO-3 ONLY
WHI/RED from Test/
Operate Switch
WHI/RED to Vacuum
Switch #3
If voltage is present as described, proceed to #5.
If voltage is not present as described, refer to the
Wiring Diagrams and check dispenser wiring harness.
FIG. 15 REFILL SWITCH TERMINALS
P1522
5. Check for continuity across the switch terminals
with the switch in the “ON” position. Continuity
must not be present when the switch is in the
“OFF” position.
15
28791 040501
6. Check the solenoid valve for coil action. Connect
the dispenser to the power source. Press the
solenoid (water) switch and listen carefully in the
vicinity of the solenoid valve for a "clicking" sound
as the coil magnet attracts.
7. Disconnect the dispenser from the power source.
SERVICE (cont.)
SOLENOIDS
If the sound is heard as described and water will not
pass through the solenoid valve, there may be a
blockage in the water inlet before the solenoid valve or,
the solenoid valve may require inspection for wear,
and removal of waterborne particles.
If the sound is not heard as described, rebuild or
replace the solenoid valve.
1
2
L
FIL
RE
3
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV UT
FL TP
OU
2
ER
UT
WAT
TP
OU
2
3
OR
AV UT
FL TP
OU
3
ER
UT
WAT
TP
OU
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
FIG. 16 SOLENOIDS
Solenoid Valve Repair:
1. Remove the wires from the solenoid valve to be
repaired.
2. Turn off the water supply to the dispenser.
3. Remove the nut securing solenoid valve coil to the
base.
4. Remove guide, washer, o-ring, plunger/spring
and inspect for wear. Replace parts as required.
5. Install new o-ring in the solenoid valve base.
6. Install plunger/spring in the guide with spring to
the bottom of the guide.
7. Install guide with the plunger/spring into the solenoid valve base.
8. Install washer on guide.
9. Install coil with retainer on guide and secure with
nut.
10. Refer to Fig. 17 and reconnect the wires.
P1521
Location:
The solenoids are located inside the autofill box.
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect the white/violet, violet or blue wire and
the white wire from the solenoid.
3. Connect the dispenser to the power source. Press
the solenoid (water) switch, and with a voltmeter,
check the voltage between the white wire, and
either the blue, violet, or white/violet wire. The
indication must be:
a) 120 volts ac for two wire 120 volt models.
b) 200 to 240 volts ac for two wire 200 or 240 volt
models.
c) 230 volts ac for two wire 230 volt models.
4. Disconnect the dispenser from the power source.
WHI from Main Harness
WHI to Solenoid #2
BLU from Solenoid
Switch #1
If voltage is present as described, proceed to #5.
If voltage is not present as described, refer to the
Wiring Diagrams and check dispenser wiring harness.
WHI to Solenoid #1
WHI to Pump #1
VIO from Solenoid
Switch #2
WHI/VIO from
Solenoid Switch #3
WHI to Main
Harness
WHI to Pump #3
5. Check for continuity across the solenoid valve coil
terminals.
#1 AFPO-2
& 3 #2 AFPO-2
If continuity is present as described, refer to Fig. 17
and reconnect the wires to the solenoid.
If continuity is not present as described, replace the
solenoid valve.
& 3 #3 AFPO-3
ONLY
FIG. 17 SOLENOID TERMINALS
16
P1544
28791 040501
SERVICE (cont.)
SOLENOID (WATER) ON/OFF SWITCHES
Removal and Replacement:
1. Remove the wires from the switch terminals.
2. Compress the clips inside the autofill box and
gently push the switch through the opening.
3. Push the new switch into the opening and spread
the clips to hold switch in the autofill box.
4. Refer to Fig. 19 when reconnecting the wires.
1
2
L
FIL
RE
3
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV UT
FL TP
OU
2
ER
UT
WAT
TP
OU
2
3
OR
AV UT
FL TP
OU
3
#1 AFPO-2 & 3
ER
UT
WAT
TP
OU
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
BLK from Pump Switch #2
BLK to Pump Switch #1
BLU to Test/Operate Switch
BLU to to Solenoid #1
P1521
FIG.18 SOLENOID (WATER) ON/OFF SWITCHES
Location:
The solenoid (water) ON/OFF switches are located
on the switch mounting plate inside the autofill box on
the right side.
#2 AFPO-2 & 3
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect the wires from the switch to be tested.
3. Check for continuity across the terminals on the
switch when it is held in the on position. Continuity
must not be present across these terminals in the
released position.
BLK from Main ON/OFF Switch
BLK to Pump Switch #2
VIO to Test/Operate Switch
VIO to Solenoid #2
#3 AFPO-3 ONLY
If continuity is present as described, refer to Fig. 19
and reconnect the wires to the switch.
If continuity is not present as described, replace the
switch.
BLK from Pump Switch #3
WHI/VIO to Test/Operate Switch
WHI/VIO to Solenoid #3
P1524
FIG. 19 SOLENOID (WATER) SWITCH
TERMINALS
17
28791 040501
SERVICE (cont.)
a) 120 volts ac for two wire 120 volt models.
b) 200 to 240 volts ac for two wire 200 or 240 volt
models.
c) 230 volts ac for two wire 230 volt models.
5. Disconnect the dispenser from the power source.
TEST/OPERATE ON/OFF SWITCH
If voltage is present as described, proceed to #6.
If voltage is not present as described, refer to the
Wiring Diagrams and check the wiring harness.
1
2
L
FIL
RE
3
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV UT
FL TP
OU
2
ER
UT
WAT
TP
OU
2
3
OR
AV UT
FL TP
OU
3
6. Disconnect the wires from the left and center
terminals.
7. Check for continuity across the left and center
terminals in rows one, two, three and four with the
switch in the “operate” position. Continuity should
not be present across the left and center terminals
with the switch in the “center” or “Test” position.
ER
UT
WAT
TP
OU
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
FIG. 20 TEST/OPERATE SWITCH
P1521
Location
The test/operate switch is located inside the autofill
box on the right side.
If continuity is present as described, replace the wires.
The switch is operating properly.
If continuity is not present as described, replace the
switch.
Test Procedures:
1. Disconnect the dispenser from the power source.
2. Remove the black wire from the top center switch
terminal.
3. Connect the dispenser to the power source. With
a voltmeter, check the voltage across the black
wire removed from the test/operate switch and the
white wire from one of the solenoids. The indication must be:
Removal and Replacement:
1. Disconnect the dispenser from the power source.
2. Remove the wires from the switch terminals.
3. Compress the clips inside on the back of the switch
panel and gently push the switch through the
opening.
4. Push the new switch into the opening and spread
the clips to hold the switch in the switch panel.
5. Refer to FIG. 21 when reconnecting the wires.
BLK to Circuit
Board J4-3
BLK from Main
ON/OFF Switch
RED/BLK from Pump Switch #1
RED/BLK to Refill Switch #1
BRN/BLK from Pump Switch #2
BRN/BLK to Refill Switch #2
WHI/RED from Pump Switch #3
WHI/RED to Refill Switch #3
BLU from Solenoid Switch #1
VIO from Solenoid Switch #2
WHI/VIO from Solenoid Switch #3
FIG. 21 TEST/OPERATE SWITCH TERMINALS
18
P1525
28791 040501
SERVICE (cont.)
a) 120 volts ac for two wire 120 volt models.
b) 200 to 240 volts ac for two wire 200 or 240 volt
models.
c) 230 volts ac for two wire 230 volt models.
4. Disconnect dispenser from the power source.
VACUUM/SUPPLY PUMP ASSEMBLY
If voltage is present as described, replace the pump.
If voltage is not present as described, refer to the
Wiring Diagrams and check main wiring harness.
1
2
L
FIL
RE
3
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV UT
FL TP
OU
2
ER
UT
WAT
TP
OU
2
3
Pump Motor Removal and Replacement:
NOTE: Do not disconnect or damage tubing. Tubing is
not replaced when pump is replaced.
1. Disconnect the pump motor leads.
2. Remove the four thumb screws securing the pump
head to the pump body.
3. Remove the two #8-32 keps nuts securing the
pump motor/bracket assembly to the refill unit.
4. Remove the two #6-32 screws securing the pump
motor to its mounting bracket.
5. Reattach this bracket to the new motor, and reinstall the assembly in the refill unit.
6. Reattach the pump head to the rear of the pump
motor with the four thumbscrews.
7. Reconnect the pump motor leads. Refer to Fig. 23
to reconnect the wires.
OR
AV UT
FL TP
OU
3
ER
UT
WAT
TP
OU
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
P1521
FIG. 22 VACUUM/SUPPLY PUMP ASSEMBLY
Location:
The vacuum/supply pumps are located inside the
autofill box.
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect the white wire from the pump lead and
the yellow, orange or red wire from the other lead.
3. Connect the dispenser to the power source. With
a voltmeter, check the voltage across the white
wire and yellow, orange or red wire. Push the
pump (flavor) switch for the pump to be tested.
The indication must be :
Pump Removal and Replacement:
NOTE: Do not disconnect or damage tubing. Tubing is
not replaced when pump is replaced.
1. Remove the four thumb screws securing the pump
head to the pump body.
WHI from Solenoid #2
YEL from Pump Switch #1
WHI from Pump #1
ORN from Pump Switch #2
WHI from Solenoid #3
RED from Pump Switch #3
FIG. 23 VACUUM/SUPPLY PUMP TERMINALS
19
P1526
28791040501
SERVICE (cont.)
VACUUM/SUPPLY PUMP TUBE INSTALLATION
2. Separate the two halves of the pump and carefully
remove the tubing from within.
3. Hold the new pump head as shown in Fig. 24, with
rollers in the 2, 6, and 10 o’clock positions.
4. Reinstall tubing in the new pump head by wrapping the tubing around the rollers as shown in Fig.
24.
5. Insert the slot of the tubing key (supplied), as
shown in Fig. 24, on the rotor so the bottom edge
of the key is pressing the tubing into the pump
head cavity as shown in Fig. 25.
6. Hold the tubing and rotor in the pump head and
remove key carefully.
7. Squeeze the pump body and pump heads together
simultaneously until pump body and pump head
are touching on all sides. Be careful not to pinch
the tubing between the pump body and the pump
head.
8. Place the new pump body on the rotor shaft.
9. Replace the four thumbscrews to secure the pump
to the motor assembly. Finger-tighten only.
ROLLERS
PUMP
HEAD
TUBING
KEY
EXIT
PORT
ENTRY
PORT
FIG. 24 TUBING START POSITION
P1542
ROTOR
ROTOR
SHAFT
FIG. 25 TUBING LOADED
20
P1541
28791 040501
SERVICE (cont.)
VACUUM SWITCHES
If continuity is not present as described, the vacuum
switch is operating properly.
If continuity is present as described, without a vacuum
generated by an empty product concentrate, replace
the vacuum switch.
1
2
L
FIL
RE
3
Removal and Replacement:
1. Disconnect dispenser from the power source.
2. Removal of the two #8-32 keps nuts securing the
pump/motor assembly nearest the vacuum
switches is required to access the lower vacuum
switch.
3. Disconnect the input tubes from the vacuum switch
to be replaced.
4. Disconnect the wires on the vacuum switch to be
removed.
5. Remove the two #4-40 screws securing the vacuum
switch to the solenoid bracket. Remove and discard switch.
6. Using two #4-40 screws secure the new vacuum
switch to the solenoid bracket.
7. Refer to Fig. 27 and reconnect the wires.
8. Connect the input tubes to the new vacuum switch.
9. Restore the pump/motor assembly to its normal
mounting position using the two #8-32 keps nuts.
IN
MA ER
W
PO
ER
T
WAT
PU
IN
1
1
OR
AV
FL PUT
IN
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
2
2
3
OR
AV UT
FL TP
OU
3
ER
UT
WAT
TP
OU
OR
AV UT
FL TP
OU
ER
UT
WAT
TP
OU
FIG. 26 VACUUM SWITCHES
P1521
Location:
The vacuum switches are located inside autofill
box, mounted on the left side of the solenoid mounting bracket just below the needle valves.
Test Procedure:
1. Disconnect the dispenser from the power source.
2. Disconnect flavor input lines.
3. Disconnect white/red, red/black or the brown/
black wires on the vacuum switch.
4. Check for continuity across the switch leads.
WHI/RED from Refill ON/OFF Switch #3
WHI/RED to Circuit Board J4-2
RED/BLK from Refill ON/OFF Switch #1
RED/BLK to Circuit Board J4-5
#3
#1
BRN/BLK from Refill ON/OFF Switch #2
BRN/BLK to Circuit Board J4-4
#2
FIG. 27 VACUUM SWITCH TERMINALS
21
P1527
28791 040501
PIPING DIAGRAM (AFPO-2)
The only difference between the AFPO-2/3 and AFPO-2/3 SL is the SL models have a water output line and
a flavor output line for each product. This allows the water and flavor to be mixed in the CDS hoppers. Refer to
the illustration below when connecting the lines.
Mo
vin
g Pa
rts
Do
.
no
unit t op
era
panewith thi te
l rem s
oved
.
Risk
Of
Ele
ctr
Dis
ica
co
l Sh
servi nnec
oc
k.
cin t po
g un we
it. r befor
e
De Type
sig
n
CH
R2 AR
Pr
es 2, Am GE
Su sure
ou
cti
on s: Hi nt _
Pr gh
_ OZ
es _
su _
.
re
Lo
__ w
®
__
BU
N
N
1
2
L
FIL
RE
N
M AIER
W
PO
ER
WAT T
PU
IN
1
OR
AV
FL UT
TP
OU
2
1
OR
AV
FL T
INPU
2
ER
WAT UT
TP
OU
OR
AV
FL UT
TP
OU
ER
WAT UT
TP
OU
22
28791 011500
SCHEMATIC WIRING DIAGRAM
AFPO-2
L1
N
GRN
SW1D
BLK
MAIN ON/OFF
SWITCH
BLK
BLK
10
AMP
T
E
S
T
/
O
P
E
R
A
T
E
BRN/BLK
SW1C
SOLENOID
SWITCH #2
VIO
B
L
K
VIO
PUMP
SW ITCH #2
PUMP
#2
SOLENOID
SWITCH #2
W
H
I
RED/BLK
W
H
I
BRN/BLK
REFILL ON/OFF
SWITCH #1
RED/BLK
S
W
I
T
C
H
SW1B
SOL
#1
BLU
B
L
K
WHI
BRN/BLK
ORN
B
L
K
SOL
#2
REFILL ON/OFF
SWITCH #2
BLU
PUMP
SW ITCH #1
RED/BLK
YEL
PUMP
#1
WHI
SW1A
BLK
VACUUM
SWITCH #2
VACUUM
SWITCH #1
C
I
R
C
U
I
T
B
O
A
R
D
J4-1
2
3
4
5
6
7
8
9
J4-10
BLK
BRN/BLK
RED/BLK
WHI
ORN
WHI
VIO
YEL
1
2
4
3
GRN
RED
1
2
3
BLK
4
1
BLK
2
GRN 1
120 VOLT AC 2 WIRE
SINGLE PHASE
1
2
GRN 1
BLK
BLK
2
2
28814.0000D 6/98 ©1998 BUNN-O-MATIC CORPORATION
L
T
A
N
H
BLU
PROBE
CONTROL
1
23
L
T
A
N
H
BLU
PROBE
CONTROL
2
28791 051599
SCHEMATIC WIRING DIAGRAM
AFPO-3
N
GRN
L1
BLK
MAIN ON/OFF
SWITCH
BLK
BLK
REFILL ON/OFF
SWITCH #3
10
AMP
WHI/RED
SOLENOID
SWITCH #3
WHI/VIO
B
L
K
WHI/RED
SW1D
WHI/VIO
PUMP
SWITCH #3
SOL
#3
WHI
WHI/RED
RED
PUMP
#3
WHI
T
E
S
T
/
O
P
E
R
A
T
E
BRN/BLK
SW1C
B
L
K
SOLENOID
SWITCH #2
VIO
B
L
K
VIO
PUMP
SWITCH #2
PUMP
#2
W
H
I
SOLENOID
SWITCH #1
BRN/BLK
REFILL ON/OFF
SWITCH #1
RED/BLK
RED/BLK
W
H
I
S
W
I
T
C
H
SW1B
SOL
#1
BLU
B
L
K
WHI
BRN/BLK
ORN
B
L
K
SOL
#2
REFILL ON/OFF
SWITCH #2
VACUUM
SW ITCH #3
BLU
PUMP
SW ITCH #1
RED/BLK
YEL
PUMP
#1
WHI
SW1A
BLK
VACUUM
SWITCH #2
VACUUM
SWITCH #1
C
I
R
C
U
I
T
B
O
A
R
D
J4-1
2
3
4
5
6
7
8
9
J4-10
WHI/RED
BLK
BRN/BLK
RED/BLK
WHI
ORN
GRY
VIO
YEL
WHI
1
2
2
4
3
3
GRN
RED
1
BLK
4
1
1
1
BLK
2
BLK
2
2
WHI
1
GRN
GRN 1
120 VOLT AC 2 WIRE
SINGLE PHASE
BLK
2
GRN
1
BLK
BLK
2
2
28814.0001D 6/98 ©1998 BUNN-O-MATIC CORPORATION
L
H
T
A
N
BLU
PROBE
CONTROL
1
24
L
T
A
N
L
H
BLU
PROBE
CONTROL
2
T
A
N
H
BLU
PROBE
CONTROL
3
28791 051599
SCHEMATIC WIRING DIAGRAM
AFPOA-2
L1
L2
GRN
SW1D
BLK
MAIN ON/OFF
SWITCH
BLK
BLK
10
AMP
T
E
S
T
/
O
P
E
R
A
T
E
BRN/BLK
SW1C
SOLENOID
SWITCH #2
VIO
B
L
K
VIO
PUMP
SW ITCH #2
PUMP
#2
SOLENOID
SWITCH #2
W
H
I
RED/BLK
W
H
I
BRN/BLK
REFILL ON/OFF
SWITCH #1
RED/BLK
S
W
I
T
C
H
SW1B
SOL
#1
BLU
B
L
K
WHI
BRN/BLK
ORN
B
L
K
SOL
#2
REFILL ON/OFF
SWITCH #2
BLU
PUMP
SW ITCH #1
RED/BLK
YEL
PUMP
#1
WHI
SW1A
BLK
VACUUM
SWITCH #2
VACUUM
SWITCH #1
C
I
R
C
U
I
T
B
O
A
R
D
J4-1
2
3
4
5
6
7
8
9
J4-10
BLK
BRN/BLK
RED/BLK
WHI
ORN
WHI
VIO
YEL
1
2
4
3
GRN
RED
1
2
BLK
4
3
1
BLK
2
GRN 1
230 VOLT AC 2 WIRE
SINGLE PHASE
1
2
GRN 1
BLK
BLK
2
2
28814.0004A 11/99 ©1999 BUNN-O-MATIC CORPORATION
L
T
A
N
H
BLU
PROBE
CONTROL
1
25
L
T
A
N
H
BLU
PROBE
CONTROL
2
28791 011500
Download PDF

advertising