Invacare IRC9LXO2AWQ Oxygen Equipment Operator`s manual


Add to my manuals
36 Pages

advertisement

Invacare IRC9LXO2AWQ Oxygen Equipment Operator`s manual | Manualzz
Operator's Manual
Platinum™
Oxygen Concentrators Standard,
with
/SensO2® and HomeFill®
Dealer: This manual must be given to the
end user.
User: Before using this concentrator, read
this manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
0434
Model IRC5LXAW
Model IRC5LXO2AW
Model IRC5LXO2AWQ
Model IRC9LXO2AWQ
WARNING
WARNING
DO NOT USE THIS PRODUCT OR ANY AVAILABLE
OPTIONAL EQUIPMENT WITHOUT FIRST
COMPLETELY READING AND UNDERSTANDING
THESE INSTRUCTIONS AND ANY ADDITIONAL
INSTRUCTIONAL MATERIAL SUCH AS OWNER’S
MANUALS, SERVICE MANUALS OR INSTRUCTION
SHEETS SUPPLIED WITH THIS PRODUCT OR
OPTIONAL EQUIPMENT. IF YOU ARE UNABLE TO
UNDERSTAND THE WARNINGS, CAUTIONS OR
INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL
PERSONNELBEFORE ATTEMPTING TO USE THIS
EQUIPMENT - OTHERWISE, INJURY OR DAMAGE
MAY OCCUR.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
NOTE: Updated versions of this manual are available on www.invacare.com.
Platinum™
2
Part No 1134866
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES ............................................... 5
RECYCLING INFORMATION ............................. 6
SECTION 1—GENERAL GUIDELINES .....................7
Maintenance.......................................................................................... 9
Radio Frequency Interference.......................................................... 9
To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To
Persons .................................................................................................. 9
Oxygen Concentration .................................................................... 10
SECTION 2—FEATURES ......................................11
SECTION 3—SHIPPING AND HANDLING ............12
Unpacking............................................................................................ 12
Inspection ............................................................................................ 12
Storage................................................................................................. 12
SECTION 4—TYPICAL PRODUCT PARAMETERS .14
SECTION 5—OPERATING INSTRUCTIONS ..........19
Introduction........................................................................................ 19
Select a Location ............................................................................... 19
Recommended Guidelines for Optimum Performance............ 20
Plug in Power Cord .......................................................................... 20
Connect Humidifier (If So Prescribed)......................................... 21
Power Switch ..................................................................................... 23
Flowrate............................................................................................... 24
SensO2 Oxygen Purity Indicators (LX O2 Sensor Model) ....... 25
Indicators (LX Model) ...................................................................... 26
Main Power Loss Alarm System .................................................... 26
Elapsed Time Meter (Hour Meter) ............................................... 26
SECTION 6—MAINTENANCE .............................27
Part No 1134866
3
Platinum™
TABLE OF CONTENTS
Routine Maintenance........................................................................ 27
SECTION 7—TROUBLESHOOTING GUIDE ............31
LIMITED WARRANTY .................................... 33
Platinum™
4
Part No 1134866
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.
SIGNAL WORD
MEANING
DANGER
Danger indicates an imminently hazardous situation
which, if not avoided, will result in death or serious
injury.
WARNING
Warning indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
CAUTION
Caution indicates a potentially hazardous situation
which, if not avoided, may result in property
damage or minor injury or both.
NOTICE
The information contained in this document is subject to change without notice.
DANGER
DO NOT SMOKE while using this device. Keep all
matches, lit cigarettes or other sources of ignition out of
the room in which this product is located and away from
where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
Part No 1134866
5
Platinum™
RECYCLING INFORMATION
RECYCLING INFORMATION
This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/CE.
This product may contain substances that could be harmful to the environment if disposed of in places (landfills) that are not appropriate according to legislation.
The ʹcrossed out wheelie binʹ symbol is placed on this product to encourage you to recycle wherever possible.
Please be environmentally responsible and recycle this product through your recycling facility at its end of life.
Platinum™
6
Part No 1134866
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - GENERAL GUIDELINES contains
important information for the safe operation and use of
this product.
In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed.
CAUTION
“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the governmental agency in which he/she practices to use or order the use of this device.”
DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel.
Part No 1134866
7
Platinum™
SECTION 1—GENERAL GUIDELINES
WARNINGS
The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.
A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless expressly recommended by Invacare.
For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life. If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call Qualified Service Personnel for examination and repair. Keep the oxygen tubing, cord, and unit away from heated or hot surfaces, including space heaters, blankets, stoves and similar electrical appliances. DO NOT move or relocate concentrator by pulling on the cord.
NEVER drop or insert any object into any opening.
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.
Platinum™
8
Part No 1134866
SECTION 1—GENERAL GUIDELINES
Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill.
Keep unit at least 7.5 cm (3 in) away from walls, draperies, furniture, and the like. Invacare recommends that Crushproof oxygen tubing (supplied by Invacare) be used with this product and NOT exceed 15.2 m (50 ft) in length. Recommended use for Platinum 9 is up to 15.2 m (50 ft) in length of high flow tubing.
Maintenance
The Invacare Oxygen Concentrator was specifically designed to minimize routine preventive maintenance at intervals of once per year. Only professionals of the healthcare field or persons fully conversant with this process such as authorized or factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.
Radio Frequency Interference
Most electronic equipment is influenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment.
To Reduce The Risk Of Burns,
Electrocution, Fire Or Injury To Persons
Avoid using while bathing. If continuous usage is required by the physician’s prescription:
The concentrator must be located in another room at least 2.1 m (7 ft.) from the bath.
DO NOT come in contact with the concentrator while wet.
A product should NEVER be left unattended when plugged in.
DO NOT place or store product where it can drop into water or other liquid.
Part No 1134866
9
Platinum™
SECTION 1—GENERAL GUIDELINES
DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
This device is to be used only in accordance with the prescription of a physician and this Owners Manual. If at any time the patient or attendant conclude that the patient is receiving an insufficient amount of oxygen, the supplier and/or physician should be contacted IMMEDIATELY. No adjustments should be made to the flowrate unless prescribed by a physician.
Close supervision is necessary when this product is used near children or physically‐challenged individuals.
Use this product for only intended use as described in this manual. DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance.
In certain circumstances oxygen therapy can be hazardous. Invacare recommends that you seek medical advice before using this machine.
This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining. Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
Oxygen Concentration
Take care not to set the flow above RED ring. An oxygen flow greater than 5 L/min. (IRC5LXO2AW, IRC5LXAW, IRC5LXO2AWQ) and 9 L/min. (IRC9LXO2AWQ) will decrease the oxygen concentration.
Platinum™
10
Part No 1134866
SECTION 2—FEATURES
SECTION 2—FEATURES
FRONT VIEW
Oxygen Purity and Fault Indicator
Lights (SensO model)/Fault and
Power Indicator Lights (LX model)
Oxygen Outlet
2
Flowmeter
Power Switch
Circuit Breaker
Elapsed Time Meter
(Hour Meter)
Cabinet Filter
(one of two
shown)
*NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor. The outlet fitting does not affect concentrator performance. Refer to the HomeFill ownerʹs manual, part number 1145804, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill, contact Invacare.
Power Cord
Outlet Fitting*
ACCESSORIES
(NOT SHOWN):
HomeFill home oxygen
compressor
Contact your local Invacare
representative for available
accessory information.
TOP VIEW
Part No 1134866
11
Platinum™
SECTION 3—SHIPPING AND HANDLING
SECTION 3—SHIPPING AND
HANDLING
NOTE: For this procedure, refer to FIGURE 3.1 on page 13.
Before you install and operate the concentrator, please refer to Platinum Series Ownerʹs Manual.
The Platinum concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator.
If the concentrator is to be reshipped by common carrier, it should be packed in a new carton. Additional cartons are available from Invacare.
The air compressor suspension system has been engineered to withstand severe motion and orientation.
Unpacking
1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or Invacare.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
NOTE: Retain carton and packing materials for storage of the Platinum. Inspection
1. Examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place other objects on top of the repackaged concentrator.
Platinum™
12
Part No 1134866
SECTION 3—SHIPPING AND HANDLING
FIGURE 3.1 Unpacking
Part No 1134866
13
Platinum™
SECTION 4—TYPICAL PRODUCT PARAMETERS
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Regulatory
Listing:
EN/IEC 60601-1
Alternating Current
Type BF equipment
Unit running
Unit not running
Attention - Consider Accompanying Documents
DO NOT smoke
Class II, Double Insulation
Electrical Requirements:
230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
Rated Current Input:
1.4 A (IRC5LXO2AWQ)
2.0 A (IRC5LXAW, IRC5LXO2AW)
2.1 A (IRC9LXO2AWQ)
Sound Level:
39.5 dB Average (IRC5LXO2AWQ)
44 dB Average (IRC5LXAW, IRC5LXO2AW)
50 dB Average (IRC9LXO2AWQ)
Altitude:
Up to 1828 meters (6,000 ft.) above sea level
without degradation of concentration levels.
NOT RECOMMENDED FOR USE ABOVE
1828 meters (6,000 ft)
For IRC9LXO2AWQ - Up to 4,000 ft. (1230
meters) above sea level without degradation
of concentration levels. For 4,000 ft (1230
meters) to 13,129 ft (4000 meters) below
90% efficiency.
Platinum™
14
Part No 1134866
SECTION 4—TYPICAL PRODUCT PARAMETERS
*Oxygen Output
Concentration Levels:
All 5LXO2AWQ/5LXAW/
5LXO2AW/9LXO2AWQ
models
* (Stated concentration levels
achieved after initial warm-up
period (approximately 30
minutes).
IRC5LXAW, IRC5LXO2AW and
IRC5LXO2AWQ
93% minimum at 1 to 3 L/min.
91% minimum at 4 L/min.
87% minimum at 5 L/min. (maximum
recommended flow)
IRC9LXO2AWQ
92% minimum at 1 to 2 L/min.
91% minimum at 8 L/min.
87% minimum at 9 L/min.
93% minimum at 3 to 7 L/min.
Maximum Outlet Pressure:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
34.5 ± 3.45 kPa (5 ± 0.5 psi)
IRC9LXO2AWQ
62.0 kPa ± 3.45 kPa (9 ± 0.5 psi)
Flow Range:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 to 5 L/min. Flowrates less than 1 L/min.
are not recommend
IRC9LXO2AWQ
1 L/min. to 9 L/min. Flowrates less than 1
L/min. are not recommended.
Low Flow Alarm:
0 L/min. to 0.5 L/min. Rapid Audible Alarm
Beeping (No Accessories Connected).
Average Power Consumption: (IRC5LXO2AWQ)
300 W
280 W @ > 3 L/min.
(IRC5LXO2AW)
365 W
340 W @ > 3 L/min.
380 W (IRC5LXAW)
475 W (IRC9LXO2AWQ)
Pressure Relief Mechanism
Operational at:
Part No 1134866
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
15
Platinum™
SECTION 4—TYPICAL PRODUCT PARAMETERS
Change in maximum
recommended flow when
back pressure of 7k Pa is
applied:
0.7 L/min.
Filters:
Cabinet (2), Outlet HEPA and Compressor
Inlet.
Safety System:
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor
shutdown.
Low Pressure Alarm w/compressor
shutdown.
Battery Free Power Loss Alarm.
SensO2 Oxygen System (SensO2 Model) Low
Flow Alarm.
Width:
46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)
Height:
67.0 cm ± 1 cm (6 3/8 in. ± 3/8 in)
Depth:
36.5 cm ± 1 cm (14 3/8 in. ± 3/8 in)
Weight:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
23.6 kg ± 1 kg (52 lbs ± 2 lbs)
IRC9LXO2AWQ
24.0 kg ± 1 kg (53 lbs ± 2 lbs)
Shipping Weight:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
26.8 kg ± 1 kg (59 lbs ± 2 lbs)
IRC9LXO2AWQ
27.2 kg ± 1 kg (60 lbs ± 2 lbs)
Operating Ambient
Temperature:
10°C - 35°C (50°F - 95°F) at 20-60% relative
humidity.
Exhaust:
Less than Ambient +19°C (+ 35°F)
Oxygen Output:
Less than Ambient +2°C (+ 4°F).
Cabinet:
Impact Resistant flame-retardant plastic
cabinet that conforms to UL 94-V0.
Platinum™
16
Part No 1134866
SECTION 4—TYPICAL PRODUCT PARAMETERS
Regulatory Listing:
IRC5LXAW/IRC5LXO2AW/
IRC5LXO2AWQ/IRC9LXO2AWQ
ETL tested as complying with:
EN55011
EN61000-3-2
EN61000-3-3
IEC 60601-1, A1, A2
IEC 60601-1-2
IRC5LXO2AW, IRC5LXO2AWQ,
IRC9LXO2AWQ Models Only
ISO8359
MDD
CE marked models:
IRC5LXO2AW, IRC5LXO2AWQ,
IRC9LXO2AWQ
Electrical:
No extension cords.
Placement:
No closer than 7.5 cm (3 inches) from any
wall, furniture, draperies, or similar surfaces.
Tubing:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 M (50 ft) high
flow tubing with high flow cannula at all flow
rates.
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-29°C to 65°C (-20°F to 150°F) at 15-95%
relative humidity.
Part No 1134866
17
Platinum™
SECTION 4—TYPICAL PRODUCT PARAMETERS
SENSO2 INDICATORS
LABEL
SYMBOL
O2
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O2 over 85% (± 2%)
GREEN Indicator Light
O2 Between 73% (± 3%)
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O2 Below 73% (±3%)
RED Indicator Light
Refer to Troubleshooting Guide
on page 31.
LX INDICATORS
LABEL
SYMBOL
I/O
Platinum™
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
GREEN Indicator Light
SYSTEM FAILURE
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™ Compressor
Shutdown
Call a qualified technician
18
Part No 1134866
SECTION 5—OPERATING INSTRUCTIONS
SECTION 5—OPERATING
INSTRUCTIONS
Introduction
Your oxygen concentrator is intended for individual use in the home. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.
Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.
Select a Location
You may select a room in your house where using your oxygen concentrator would be most convenient. Your concentrator can be easily rolled from room to room on its casters.
Your oxygen concentrator will perform best when operated under the following conditions. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and/or fumes.
Part No 1134866
19
Platinum™
SECTION 5—OPERATING INSTRUCTIONS
Recommended Guidelines for Optimum
Performance
Temperature:
(10°C - 35°C) 50°F - 95°F
Electrical:
No extension cords.
Placement:
No closer than 7.5 cm (3 in) from the wall,
furniture, draperies, or similar surfaces.
Tubing and Cannula:
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 m (50 ft) high
flow tubing with high flow cannula at all flow
rates.
Environment:
Smoke and soot-free. No confined spaces
(Example: No closets).
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Flow Rate
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 L/min. to 5 L/min. Flow rates less than 1
L/min., are not recommended.
IRC9LXO2AWQ
1 L/min. to 9L/min. Flow rates less than 1
L/min. are not recommended.
Minimum Operating Time:
30 Minutes
NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away from walls, draperies or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Platinum™
20
Part No 1134866
SECTION 5—OPERATING INSTRUCTIONS
Connect Humidifier (If So Prescribed)
NOTE: For this procedure, refer to FIGURE 5.1,FIGURE 5.2, FIGURE 5.3 on page 22 and FIGURE 5.4 on page 23.
1. Fill humidifier with distilled water to the level indicated by the manufacturer. (Please contact Invacare for recommendations).
WARNING
DO NOT overfill humidifier.
Replace the humidifier cap and securely tighten.
DO NOT reverse the oxygen input and output
connections. Water from the humidifier bottle will travel
through the cannula back to the patient.
Humidifier Bottle
FIGURE 5.1 Filling the Humidifier
2. Place the humidifier bottle in the humidifier compartment.
Humidifier Bottle
FIGURE 5.2 Humidifier Compartment
Part No 1134866
21
Platinum™
SECTION 5—OPERATING INSTRUCTIONS
3. Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3.
4. The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3.
5. Attach it to the humidifier by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. Refer to Detail “C” in FIGURE 5.3.
DETAIL “A”
Plate Groove (Insert Flat
Head Screwdriver Here)
Filter Access Door
DETAIL “B”
Humidifier Bottle
Adapter
DETAIL “C”
Filter
Access
Door
Humidifier Bottle
Adapter
FIGURE 5.3 Attaching the Humidifier Bottle Adapter
Platinum™
22
Part No 1134866
SECTION 5—OPERATING INSTRUCTIONS
6. Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator.
7. Attach the cannula/patient supply tubing to the humidifier bottle outlet.
8. After assembly, ensure that oxygen is flowing through the cannula.
9. Replace the filter access door on the side of concentrator by snapping it back into the plate groove.
Oxygen Outlet
Connector
Oxygen
Tubing
Humidifier Outlet
FIGURE 5.4 Attaching the Oxygen Tubing
WARNING
For optimum performance, Invacare recommends that
each concentrator be on and running for a minimum of 30
minutes at a time. Shorter periods of operation may
reduce maximum product life.
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5 on page 24.
1. Press power switch to on position. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functioning properly.
Part No 1134866
23
Platinum™
SECTION 5—OPERATING INSTRUCTIONS
On/Off (I/O)
Power Switch
FIGURE 5.5 Power Switch
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6 on page 25.
1. Turn the flowrate knob to the setting prescribed by your physician or therapist.
WARNING
DO NOT change the L/min. setting on the flowmeter
unless a change has been prescribed by your physician or
therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed. 2. If the flowrate on the flowmeter ever falls below 0.5 L/min. for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min., the LOW FLOW audible alarm will go off.
Platinum™
24
Part No 1134866
SECTION 5—OPERATING INSTRUCTIONS
NOTE: The use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball
FIGURE 5.6 Flowrate
SensO2 Oxygen Purity Indicators (LX O2
Sensor Model)
This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.
When the unit is turned on, the GREEN light will come on (SYSTEM OK/O2 greater than 85%). After five minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values. Explanation of Oxygen Purity Indicator Lights
NOTE: For this procedure, refer to FIGURE 5.7.
Part No 1134866
25
Platinum™
SECTION 5—OPERATING INSTRUCTIONS
OXYGEN PURITY
Auto Shut Down
Use Backup
RED
Call Supplier
Below
Normal
Normal
YELLOW
O2
GREEN
FIGURE 5.7 SensO2 Oxygen Purity Indicators (LX O2 Sensor
Model)
GREEN light (O2) ‐ Normal Operation.
YELLOW light ( ) ‐ Call supplier IMMEDIATELY. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that backup oxygen is nearby.
RED light ( )‐ Total unit shutdown. Switch IMMEDIATELY to backup oxygen supply. Call supplier IMMEDIATELY (FIGURE 5.7).
GREEN light ‐ with YELLOW light flashing ‐ Call supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may continue to use the concentrator (FIGURE 5.7). Indicators (LX Model)
NOTE: For this procedure, refer to FIGURE 5.8.
Explanation of LX models Indicator Lights
RED light ( ) ‐ Total Unit Shut‐Down. Switch IMMEDIATELY to a back‐up oxygen supply. Call supplier IMMEDIATELY. GREEN light (I/O) ‐ On/Off. System okay.
Platinum™
26
Part No 1134866
SECTION 5—OPERATING INSTRUCTIONS
Auto Shut Down
Use Backup
RED
Call Supplier
Normal
I/O
GREEN
FIGURE 5.8 Indicators (LX Model)
Main Power Loss Alarm System
The alarm system contains no battery ‐ and is thereby maintenance free. The alarm system is powered by a capacitor that is continuously charged and ready to trigger the alarm in the event of main power loss. Elapsed Time Meter (Hour Meter)
The hour meter displays the cumulative number of hours the unit has operated.
Part No 1134866
27
Platinum™
SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to minimize routine preventive maintenance. In places with high dust or soot levels, maintenance may need to be performed more often. Only authorized or factory‐trained personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. Refer to Preventive Maintenance Record on page 30.
Routine Maintenance
NOTE: The following routine maintenance should be performed more frequently. See recommended intervals in each section and preventive maintenance record. Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1 on page 29. CAUTION
DO NOT operate the concentrator without the filters
installed.
NOTE: There are two cabinet filters. One filter located on each side of the cabinet.
1. Remove each filter and clean at least once a week depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
Platinum™
28
Part No 1134866
SECTION 6—MAINTENANCE
3. Dry the filters thoroughly before reinstallation.
Cabinet Filter
(located on both
sides of
concentrator)
FIGURE 6.1 Cleaning the Cabinet Filter
Cabinet
WARNING
Unplug the concentrator when cleaning. To avoid
electrical shock, DO NOT remove cabinet.
1. Clean the cabinet with a mild household cleaner and non‐abrasive cloth or sponge.
Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten parts water and one part vinegar.
3. Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.
WARNING
DO NOT overfill humidifier bottle.
Part No 1134866
29
Platinum™
Platinum™
30
*NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE:
2,160 hours are equivalent to usage 24 hours per day, for 90 days.
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Rebuild Top End of Compressor
Replace Exhaust Muffler
Clean Heat Exchanger
AS REQUIRED
Check Power Loss Alarm
Check Compressor Inlet Filter*
Check Outlet HEPA Filter*
Clean/Replace Cabinet Filter(s)
I RC
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
Check Oxygen Concentration
( UNITS WITH SensO 2 )
EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST
Check Oxygen Concentration
(UNITS WITHOUT SensO 2 )
MODEL NO.
EVERY 4,380 HOURS FOR IRC5LXA W
Check Prescribed L/min. Flow Rate
Clean Cabinet Filter(s)
Record Elapsed Hours On Hour Meter
Record Date of Service
ON EACH INSPECTION
PREVENTIVE MAINTENANCE RECORD
INVACARE PLATINUM OXYGEN CONCENTRATOR
SECTION 6—MAINTENANCE
SERIAL NO.
FIGURE 6.2 Preventive Maintenance Record
Part No 1134866
SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Main Power Loss:
Short beeps, long
pause
1. Power cord NOT
plugged in.
1. Insert plug into outlet.
Concentrator not
operating,
power switch On.
2. No power at outlet.
2. Inspect house circuit
breakers or fuses. If
problem recurs, use a
different outlet.
Beep......Beep......
3. Tripped circuit
breaker.
3. Call Service Provider.
Alarm:
Continuous
System Failure:
1. Unit overheating due
to blocked air intake.
1a. Remove and clean
cabinet filters.
1b. Move oxygen
concentrator at least 7.5 cm
(3 in) away from walls,
draperies or furniture.
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
Part No 1134866
2. Insufficient power at
outlet.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Internal repairs
required
3. Call Service Provider.
31
Platinum™
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM:
YELLOW or RED
light illuminated.
PROBABLE CAUSE:
1. Low oxygen purity.*
1. Clean or Replace filters.
2. Kinked or blocked
tubing, cannula or
humidifier.*
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power OFF
for 60 seconds and then
turn power back ON.
3. Flowmeter set at 0.5
L/min. or less.*
3. Check flowmeter is set to
1.0 L/min. or more. (Refer
to Recommended
Guidelines for Optimum
Performance on page 20.
4. Unit overheating due
to blocked air intake.
4.a. Remove and clean
cabinet filters.
Alarm:
Continuous
On red light only.
* Only applies to
SensO2 models.
SOLUTION:
4b. Move concentrator at
least 7.5 cm (3 in) from
walls, draperies, and
furniture.
5. Insufficient power at
outlet.
6. Internal repairs
required.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Call Service Provider.
GREEN light
YELLOW Light
Flashing.
1. Internal repairs
required.
1. Call Service Provider.
Alarm:
RAPID
Low Flow Alarm
1a. Kinked or blocked
tubing, cannula or
humidifier.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power OFF
for 60 seconds and then
turn power back ON.
1b. Reset flowmeter to
prescribed flowrate.
Flowrates less than1L/min.
are not recommended.
Beep....Beep...
Beep...Beep
1b. Flowmeter set at 0.5
L/min. or less.
Platinum™
32
Part No 1134866
LIMITED WARRANTY
LIMITED WARRANTY
Available At:
Belgium & Luxemburg:
Invacare nv, Autobaan 22, B‐8210 Loppem
Tel: (32) (0)50 83 10 10, Fax: (32) (0)50 83 10 11
[email protected]
Danmark: Invacare A/S, Sdr. Ringvej 37, DK‐2605 Brøndby
Tel: (45) (0)36 90 00 00, Fax: (45) (0)36 90 00 01
[email protected] Deutschland:
Invacare Aquatec GmbH, Alemannenstraße 10, D‐88316 Isny
Tel: (49) (0)75 62 7 00 0, Fax: (49) (0)75 62 7 00 66
info@invacare‐aquatec.com European Distributor Organisation:
Invacare, Kleiststraße 49, D‐32457 Porta Westfalica Tel: (49) (0)57 31 754 540, Fax: (49) (0)57 31 754 541
[email protected]
España:
Invacare SA, c/Areny s/n, Polígon Industrial de Celrà, E‐17460 Celrà (Girona)
Tel: (34) (0)972 49 32 00, Fax: (34) (0)972 49 32 20
[email protected] France:
Invacare Poirier SAS, Route de St Roch, F‐37230 Fondettes
Tel: (33) (0)2 47 62 64 66, Fax: (33) (0)2 47 42 12 24
[email protected] Ireland:
Invacare Ireland Ltd, Unit 5 Seatown Business Campus,
Seatown Road, Swords, County Dublin ‐ Ireland
Tel: (353) 1 810 7084, fax: (353) 1 810 7085
[email protected]
Part No 1134866
33
Platinum™
LIMITED WARRANTY
Italia:
Invacare Mecc San s.r.l., Via dei Pini 62, I‐36016 Thiene (VI)
Tel: (39) 0445 38 00 59, Fax: (39) 0445 38 00 34
[email protected] Nederland:
Invacare BV, Celsiusstraat 46, NL‐6716 BZ Ede
Tel: (31) (0)318 695 757, Fax: (31) (0)318 695 758
[email protected] [email protected] Norge:
Invacare AS, Grensesvingen 9, Postboks 6230, Etterstad, N‐0603 Oslo
Tel: (47) (0)22 57 95 00, Fax: (47) (0)22 57 95 01
[email protected] [email protected] Österreich:
Mobitec Mobilitätshilfen GmbH, Herzog Odilostrasse 101, A‐5310 Mondsee
Tel.: (43) 6232 5535 0, Fax.: (43) 6232 5535 4
office@mobitec‐austria.com, [email protected]
Portugal:
Invacare Lda, Rua Estrada Velha, 949, P‐4465‐784 Leca do Balio
Tel: (351) (0)225 1059 46/47, Fax: (351) (0)225 1057 39
[email protected] Sverige & Suomi:
Invacare AB, Fagerstagatan 9, S‐163 91 Spånga
Tel: (46) (0)8 761 70 90, Fax: (46) (0)8 761 81 08
[email protected]
[email protected] Platinum™
34
Part No 1134866
LIMITED WARRANTY
Switzerland:
Mobitec Rehab AG, Benkenstrasse 260, CH‐4108 Witterswil
Tel.: (41) (0)61 487 70 80, Fax.: (41) (0)61 487 70 81
office@mobitec‐rehab.ch
[email protected]
United Kingdom:
Invacare Ltd, South Road, Bridgend Industrial Estate, UK‐Bridgend CF31 3PY
Tel: (44) (0)1656 664 321, Fax: (44) (0)1656 667 532
[email protected] Part No 1134866
35
Platinum™
www.invacare.com
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
440-329-6000
800-333-6900
Technical Services
440-329-6593
800-832-4707
Manufacturer:
Invacare Medical
Equipment (Suzhou) Co.,
Ltd.
Unit 2B, Export
Processing Zone
No. 200 Suhong Road SIP
(Suzhou Industrial Park)
Suzhou, Jiangsu, PRC
215021
Tel: 86-512-62586192
Fax: 86-512-62586072
0434
Invacare Corporation
EC
REP
EU Representative
Invacare International Sarl
Route de Cité Ouest 2
1196 Gland
Switzerland
Tel: +41 22 354 60 10
Fax: +41 22 354 60 11
All rights reserved. Trademarks are
identified by ™ and ®.
All trademarks are owned by or
licensed to Invacare Corporation or
its subsidiaries unless otherwise
noted.
Part No 1134866
Rev E - 04/09
© 2009 Invacare Corporation