Masterscan 350 User`s Guide

Add to my manuals
113 Pages

advertisement

Masterscan 350 User`s Guide | Manualzz

Masterscan 350

User’s Guide

Masterscan 350 User’s Guide

Copyright © 2006 – 2009 Sonatest Limited

All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise, without the prior written permission of Sonatest Limited.

Sonatest Limited

Dickens Road

Milton Keynes

MK12 5QQ

England

Phone: +44 (0)1908 316345

Fax: e-mail:

+44 (0)1908 321323 [email protected]

Web: www.sonatest.com

Document Number: 147310

Sonatest Inc.

12775 Cogburn

San Antonio

Texas

78249

USA

Phone: +1 (210) 697 0335

Fax: +1 (210) 697 0767 e-mail: [email protected]

Issue 4 June 2009

Page i

Masterscan 350 User’s Guide ~ Contents

Contents

1  How to Use This User’s Guide .......................................................................................... 1 

1.1  Typographical Conventions .................................................................................. 1 

2  Disclaimers and Notices .................................................................................................... 2 

2.1 

2.2 

2.3 

2.4 

Specific Warnings .................................................................................................. 2 

General Warnings .................................................................................................. 2 

Operator Training .................................................................................................. 2 

Testing Limitations ................................................................................................ 3 

2.5  Critical Operating Factors ..................................................................................... 3 

2.5.1  Calibration of the Sound Velocity .................................................................... 3 

2.5.2  Transducer (Probe) Zero Procedure ................................................................ 3 

2.5.3  Flaw Detection Calibration ............................................................................... 3 

2.5.4  Effects of Temperature on Calibration ............................................................. 4 

2.5.5  Transducer Condition ...................................................................................... 4 

2.5.6  Use of Couplant ................................................................................................ 4 

2.6 

2.7 

Disclaimer of Liability ............................................................................................ 5 

Electromagnetic Compatibility .............................................................................. 5 

3  Quick Start ......................................................................................................................... 6 

3.1 

3.2 

3.3 

3.4 

Front Panel and Optional Keyboard Controls ...................................................... 6 

Flaw Detection ..................................................................................................... 12 

Thickness Gauging .............................................................................................. 13 

Masterscan 350 Memory ...................................................................................... 15 

3.4.1  Reset to factory defaults: ................................................................................ 15 

3.4.2  Clearing the Memory ...................................................................................... 15 

4  Detailed Menu Description ............................................................................................. 16 

4.1 

4.2 

Menu Tree ........................................................................................................... 16 

Main Menu ........................................................................................................... 18 

4.2.1  Main Menu Selection ...................................................................................... 18 

4.3  CAL Menu ............................................................................................................ 19 

4.3.1  Calibration Menu ............................................................................................ 19 

4.3.2  Amplifier Menu ............................................................................................... 19 

4.3.3  TX (Transmitter) Menu .................................................................................. 20 

4.3.4  GATE1 & GATE2 Menus ................................................................................. 21 

4.3.5  A-CAL Menu .................................................................................................... 22 

4.4  MEAS Menu .......................................................................................................... 23 

4.4.1  Measurement Menu ........................................................................................ 23 

4.4.2  CSC (Curved Surface Correction) .................................................................. 25 

4.4.3  PROBE Menu .................................................................................................. 25 

4.4.4  Distance Amplitude Correction (DAC) Menu ................................................ 26 

4.4.5  DGS/AVG ......................................................................................................... 27 

Page ii

Masterscan 350 User’s Guide ~ Contents

4.4.6  TCG Time Corrected Gain Menu ................................................................... 29 

4.4.7  AWS Menu ....................................................................................................... 30 

4.4.8  API Menu......................................................................................................... 31 

4.5  UTIL Menu ........................................................................................................... 32 

4.5.1  UTIL Menu ...................................................................................................... 32 

4.5.2  VIDEO Menu ................................................................................................... 32 

4.5.3  MISC Menu ..................................................................................................... 33 

4.5.4  BEA (Backwall Echo Attenuator) .................................................................... 33 

4.5.5  AGC (Automatic Gain Control) ...................................................................... 34 

4.5.6  PRINT Menu .................................................................................................... 35 

4.5.7  P_O/P Proportional Output Control Menu .................................................... 36 

4.5.8  CLOCK Menu .................................................................................................. 36 

4.6  MEMORY Menu ................................................................................................... 37 

4.6.1  PANEL Memory Menu ..................................................................................... 37 

4.6.2  A-LOG Memory Menu ..................................................................................... 38 

4.6.3  REF Menu ........................................................................................................ 39 

4.6.4  T-LOG Menu Numeric Mode .......................................................................... 39 

4.6.5  T-LOG Menu Sequential Mode ....................................................................... 40 

4.6.6  T-LOG Menu Delete Mode ............................................................................. 40 

4.7  Storage & Recall of Calibration Setups ................................................................ 41 

4.7.1  To store a panel set: ....................................................................................... 41 

4.7.2  To recall a panel set: ....................................................................................... 41 

4.7.3  To delete a panel set: ..................................................................................... 41 

4.7.4  Adding NOTES to PANEL and A-LOG sets ...................................................... 41 

4.7.5  Edit Notes Summary ....................................................................................... 43 

4.7.5.1 

4.7.5.2 

Using the Front Panel Keypad .............................................................. 43 

Using the Optional Keyboard ............................................................... 43 

5  Flaw Testing .................................................................................................................... 45 

5.1 

5.2 

5.3 

5.4 

5.5 

5.6 

5.7 

5.8 

Basic Flaw Testing ............................................................................................... 45 

Flaw Testing Calibration ...................................................................................... 46 

DAC Operation .................................................................................................... 50 

TCG Operation .................................................................................................... 53 

Weld Inspection Using Trigonometry Mode ....................................................... 55 

Weld Inspection Using the AWS Menu ................................................................ 57 

Evaluation of Pipe Imperfections using API 5UE ................................................ 59 

DGS/AVG Operation ............................................................................................ 62 

5.8.1  Theory ............................................................................................................. 62 

5.8.2  DGS/AVG ~ Frequency, TBVEL and ERS ....................................................... 64 

5.8.3  DGS/AVG ~ NFL, DELAY and PED ................................................................. 64 

5.8.4  DGS/AVG ~ T-Loss, REF and MAT ................................................................. 65 

5.8.5  DGS/AVG ~ dVK, REF TYPE and REFSIZE ..................................................... 66 

Page iii

Masterscan 350 User’s Guide ~ Contents

5.8.6  DGS/AVG Calibration ...................................................................................... 67 

5.8.6.1  Adjustment ............................................................................................ 69 

5.8.7  Measurement .................................................................................................. 70 

5.9  Backwall Echo Attenuator (BEA) ......................................................................... 71 

5.10  A-LOG, A-Scan Storage ......................................................................................... 73 

5.10.1  To Store An A-Scan: ........................................................................................ 73 

5.10.2  To Recall an A-Scan: ....................................................................................... 74 

5.10.3  To Delete a Stored A-Scan .............................................................................. 75 

5.10.4  REF, Reference Waveform Comparisons ........................................................ 75 

5.10.5  To Recall an A-Scan as a Reference Waveform: .............................................. 75 

5.10.6  Contour, Peak Echo & Hold Dynamics .......................................................... 76 

6  Thickness Gauging .......................................................................................................... 79 

6.1  Basic Thickness Gauging ..................................................................................... 79 

A-Cal ..................................................................................................................... 82  6.2 

6.3 

6.4 

6.5 

TCG for Reliable Gauging .................................................................................... 83 

T-LOG Thickness Storage (Numeric) .................................................................. 84 

T-LOG Thickness Storage (Sequential & Download) .......................................... 85 

T-Log Editing and Printing .................................................................................. 87  6.6 

6.6.1  To delete a single location: ............................................................................ 87 

6.6.2  To delete all Locations in a single Block: ....................................................... 87 

6.6.3  To delete all Blocks and all Locations: ........................................................... 87 

6.6.4  To print thickness logs ................................................................................... 88 

7  Power Supply .................................................................................................................. 89 

7.1  Lithium-Ion Battery Pack ..................................................................................... 89 

7.1.1  Battery Cautions ............................................................................................. 89 

7.2  Battery Charging .................................................................................................. 91 

7.2.1  Charging Caution ........................................................................................... 91 

8  Interface Connections ..................................................................................................... 93 

8.1 

8.2 

8.3 

8.4 

RS232 ................................................................................................................... 94 

Composite Video ................................................................................................. 95 

Proportional Outputs .......................................................................................... 95 

Synchronisation Outputs ..................................................................................... 95 

9  Features of the Masterscan 350 ....................................................................................... 96 

9.1  Functional Testing Methods ................................................................................ 96 

10  Specifications................................................................................................................... 97 

11  Warranty ........................................................................................................................ 101 

12  Index ............................................................................................................................. 102 

Page iv

Masterscan 350 User’s Guide ~ How to Use This User’s Guide

1 How to Use This User’s Guide

This user’s guide has been designed so that a person with a good knowledge of the basics of ultrasonic non-destructive testing may understand the operation and use of the features offered by the Masterscan 350. The user is advised, however, of the important nature of ultrasonic non-destructive testing and is referred to Section 2 for important information on the proper use of this technology.

Section 1: How to Use This User’s Guide is this section.

Section 2: Disclaimers and Notices contains important information that must be understood by users of the Masterscan 350

Section 3: Quick Start provides a user familiar with ultrasonic testing a means to operate the instrument’s basic functions and to quickly achieve familiarity without having to understand all of its features in detail.

Section: 4: Detailed Menu Description is an in-depth description of the menu items found in the Masterscan 350 with brief descriptions of each item.

Section 5: Flaw Testing gives guidelines on performing flaw testing with examples of how to calibrate the Masterscan 350 and how to set up the unit for DAC, TCG,

DGS/AVG and how to use the A-Scan storage facilities.

Section 6: Thickness Gauging gives guidelines on performing thickness measurements including the calibration of the Masterscan 350, using TCG and how to use the thickness measurement memory functions.

Section 7: Power Supply describes important aspects of using and caring for the battery power supply so as to get maximum battery duration time and life.

Section 8: Interface Connections is for users who desire to operate the Masterscan 350 with ancillary equipment.

Section 9: Features of the Masterscan 350 is a quick reference listing the pertinent characteristics of the instrument and the various functional testing methods that may be used with the instrument.

The function keys are shown throughout this manual as shown in Section 3.1, Front Panel and Optional Keyboard Controls.

1.1 Typographical Conventions

When describing front panel buttons the text is

éêáåíÉÇ=áå=íÜáë=Ñçåí

.

When describing settings in the Masterscan 350 the text is.

printed in this font

Page 1

Masterscan 350 User’s Guide ~ Disclaimers and Notices

2 Disclaimers and Notices

The following information must be read and understood by users of the Sonatest

Masterscan 350 ultrasonic flaw detector and thickness gauge. Failure to follow these instructions can lead to serious errors in test results or damage to the flaw detector.

Decisions based on erroneous results can lead to property damage, personal injury or death. Anyone using this instrument should be fully qualified by their organization in the theory and practice of ultrasonic testing, or under the direct supervision of such a person.

2.1 Specific Warnings

The Masterscan 350 contains a high-energy, precision pulser allowing optimum testing results to be obtained by matching the pulse width to the probe characteristics. This circuitry may be damaged by voltage spikes. It is recommended that the instrument be switched off, or the pulser stopped (by pressing the key) before changing transducers

(probes).

2.2 General Warnings

Proper use of the ultrasonic test equipment requires three essential elements:

Knowledge of the specific test or inspection and applicable test equipment.

Selection of the correct test equipment based on knowledge of the application.

Competent training of the instrument operator.

This user’s guide provides instruction in the basic operation of the Masterscan 350 flaw detector. In addition to the methods included, many other factors can affect the use of this flaw detector. Specific information regarding these factors is beyond the scope of this manual. The user should refer to appropriate textbooks on the subject of ultrasonic testing and thickness gauging for more detailed information.

2.3 Operator Training

Operators must receive adequate training before using this ultrasonic flaw detector.

Operators must be trained in general ultrasonic testing procedures and in the set-up and performance required by each specific test or inspection. Operators must understand:

Sound wave propagation theory.

Effects of the velocity of sound in the test material.

Behaviour of the sound wave at the interface of two different materials.

Sound wave spread and mode conversion.

Page 2

Masterscan 350 User’s Guide ~ Disclaimers and Notices

More specific information about operator training, qualification, certification and test specifications can be obtained from technical societies, industry groups and government agencies.

2.4 Testing Limitations

In ultrasonic testing, information is obtained only from within the confines of the sound beam as it propagates into the test material. Operators must exercise great caution when making inferences about the nature of the test material outside the limits of the sound beam. The condition of materials can vary significantly and the results can be erratic if operators don’t exercise good judgment.

2.5 Critical Operating Factors

The following procedures must be observed by all users of this ultrasonic flaw detector in order to obtain proper and accurate results.

2.5.1 Calibration of the Sound Velocity

An ultrasonic flaw detector operates on the principle of measuring the time of flight of a burst of high frequency sound through the test piece as well as evaluating the amplitude of reflected or transmitted echoes. The sound velocity of the test piece multiplies this time in order to obtain an accurate distance or thickness reading. Since the actual sound velocity in materials can vary from the published values, the best result is obtained when the instrument is calibrated on a reference block made from the same material as the test piece.

This block should be flat, smooth and as thick as the maximum thickness expected of the test piece.

Users should also be aware that the sound velocity might not be constant throughout the test piece due to effects such as heat-treating. This must be taken into consideration when evaluating the results of ultrasonic thickness testing. The calibration should always be checked after testing to minimize errors.

2.5.2 Transducer (Probe) Zero Procedure

The transducer calibration procedures must be performed as described in this manual. The calibration block must be clean, in good condition and free of noticeable wear. Failure to perform the transducer zero and calibration procedure will cause inaccurate thickness readings.

2.5.3 Flaw Detection Calibration

When performing flaw detection, it is important to note that the amplitude of indications is not only related to the size of the discontinuity; the depth of a discontinuity below the test piece surface will also have an effect on the amplitude due to characteristics of the sound beam spread and near field zone of the transducer. In addition, the characteristics of the discontinuity such as orientation and classification can alter the expected amplitude

Page 3

Masterscan 350 User’s Guide ~ Disclaimers and Notices response. For these reasons, calibration should be performed on test blocks made of the same material as the test piece with artificial discontinuities within the range of size and depth in the material to be detected. The user is again cautioned to refer to reference books which are beyond the scope of this manual.

2.5.4 Effects of Temperature on Calibration

The sound velocity in test pieces and the transducer wear face changes with temperature variations. All calibrations should be performed on site with test blocks at or near the same temperature as that expected on the test piece, to minimize errors.

2.5.5 Transducer Condition

The transducer used for testing must be in good condition, without noticeable wear of the front surface. The specified range of the transducer must encompass the complete range of the thickness to be tested and/or the types of discontinuities to be investigated. The temperature of the material to be tested must be within the transducer’s specified temperature range.

2.5.6 Use of Couplant

Operators must be familiar with the use of ultrasonic couplant. Testing skills must be developed so that couplant is used and applied in a consistent manner to eliminate variations in couplant thickness which can cause errors and inaccurate readings. Calibration and actual testing should be performed under similar coupling conditions, using a minimum amount of couplant and applying consistent pressure to the transducer.

Page 4

Masterscan 350 User’s Guide ~ Disclaimers and Notices

2.6 Disclaimer of Liability

All statements, technical information and recommendations contained in this manual or any other information supplied by Sonatest Limited in connection with the use, features and qualifications of the Masterscan 350 are based on tests believed to be reliable, but the accuracy or completeness thereof is not guaranteed. Before using the product you should determine its suitability for your intended use based on your knowledge of ultrasonic testing and the characteristics of materials. You bear all risk in connection with the use of the product.

You are reminded that all warranties as to merchantability and fitness for purpose are excluded from the contract under which the product and this manual have been supplied to you. The Seller’s only obligation in this respect is to replace such quantity of the product proved to be defective.

Neither the seller nor the manufacturer shall be liable either in contract or in tort for any direct or indirect loss or damage (whether for loss of profit or otherwise), costs, expenses or other claims for consequential or indirect compensation whatsoever (and whether caused by the negligence of the company, its employees or agents or otherwise).

2.7 Electromagnetic Compatibility

This product conforms to the following European Directives:

Directive 2002/95/EC on the Restriction of the Use of certain Hazardous Substances in

Electrical and Electronic Equipment (RoHS)

Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE)

Low Voltage Directive (LVD) 73/23/EEC

CE Marking Directive 93/68/EEC

EMC Directive 89/336/EEC. However, in order to ensure the equipment meets the requirements, the following should be read:

WARNING!

This is a “CLASS A” product. In a domestic environment, this product may cause radio interference. In which case the user may be required to take adequate measures.

Note: This product should not be connected to cables greater than three (3) meters in length. If this is necessary, the installation may require further

EMC testing to ensure conformity.

The Masterscan 350 also complies with EN 12668-1, Non-destructive testing and verification of ultrasonic examination equipment – Part 1: Instruments. For any questions relating to the proper use of this product, please contact the manufacturer at the number indicated on page i.

Page 5

Masterscan 350 User’s Guide ~ Quick Start

3 Quick Start

3.1 Front Panel and Optional Keyboard Controls

The front panel controls consist of a series sealed, pressure sensitive, tactile buttons that allow the instrument settings to be changed and the movement through the menus on the screen. An optional keyboard connected to the USB port can also be used to navigate through the menus and make changes to the parameter values. A more detailed description of each button is shown below followed by an overview of the front panel:

When the term “highlighted” is used, it refers to text with a bright background and dark letters which is the selected item.

Power on & off push button for switching the instrument on and off.

Operates as a toggle.

N.B Random lines or characters may be displayed for a second or two after switch on, before the memory is initialized, this is normal.

The Menu button is used to display the 4 main menu levels (

CAL

,

MEAS

,

UTIL

and

MEMORY

) on the right hand side of the screen. The jbkr button can be pressed at any time. To select the required sub-menu press the plain grey button associated with it. Corresponds to the “ESC” key when using a keyboard.

These buttons move the highlighted cursor along the top of the screen left and right to the sub menu to be selected. These are momentary buttons with no repeat. The parameter boxes along the right hand side of the screen change as the sub menu is selected. Corresponds to left and right arrows ( and ) when using a keyboard.

The four grey buttons on the right side of the display are used to select the menu boxes, which appear on the right side of the A-Trace. When a menu box is selected, it will be highlighted. Pressing the button a second time will, on certain parameters, display a drop-down list with a selection of values that can be chosen by using the and buttons. When using a keyboard F1 to F4 correspond to the 4 grey buttons.

This button operates in connection with the plain grey buttons next to the parameter boxes to increase the value or step the selection in the positive direction. In the case of transducer

Zero

or Delay, it moves the signals to the right. This is a repeat button with acceleration to facilitate quick scrolling of the value. Corresponds to up arrow () when using a keyboard.

Page 6

Masterscan 350 User’s Guide ~ Quick Start

This button operates in connection with the plain grey buttons next to the parameter boxes to decrease the value or step the selection in the reverse direction. In the case of transducer

ZERO

or

Delay

, it moves the signals to the left. This is a repeat button with acceleration to facilitate quick scrolling of the value. Corresponds to down arrow () when using a keyboard.

This button changes the size of the steps that are used to adjust the parameters; there are two sizes: small and large that are dependant on the parameter that is being changed. The button toggles between the two sizes and the current state is indicated by  for small and by  for the larger step in the parameter box that is being adjusted. Corresponds to the letter “I” when using the keyboard.

This button operates in conjunction with the memory menus to accept a store or to accept a recall of a memory already in storage. In the print mode, lh= acts as the print button. Corresponds to “Enter” or the return key.

Press this button once to select freeze mode for the A-Scan display. This is a useful feature for holding an echo for evaluation. When in this mode, a box is highlighted showing

FREEZE

below the graticule.

Pressing the button a second time selects peak mode, which holds and updates all echoes on the display during inspection. This feature allows an envelope or echo dynamic pattern to be drawn on the screen with the "live" signal shown within, which is useful for angle beam inspection to locate the peak signal. When in this mode, a box is highlighted showing

PEAK

below the graticule as sown below:

Pressing the key again invokes the Hold mode that captures the current A-

Scan in white allowing a comparison to be made with the live A-Scan that is shown in green. When this mode is selected a box is highlighted showing

Page 7

Masterscan 350 User’s Guide ~ Quick Start

HOLD

as shown below:

If key lock is enabled from within the

UTIL

menu, pressing the button a fourth time selects

KEYLOCK

mode, preventing accidental changing of parameters

Pressing the button a fifth time returns the A-Scan display to normal mode.

This is a momentary button with no repeat action. Corresponds to the letter

“F” when using the keyboard.

This button selects the help menu which overlays the display. The help menu explains how the Masterscan 350 operates with a choice of three options:

Using the Masterscan 350

Description of the active menu

Calibration

Pressing the ebim

button again at any point in the help screens returns the display to normal mode. The Help Screen also displays the Instrument serial number & software version. Corresponds to “F5” when using a keyboard.

Page 8

Masterscan 350 User’s Guide ~ Quick Start

The full screen button is used to toggle between showing the A-Scan on the full display and showing it with the menus and parameter boxes as shown below:

Full display is useful for maximising the size of the A-Scan while performing inspections. While in full screen mode the user can still change the gain in the normal way and use the cobbwbLmb^h

button; all other buttons are disabled. If split screen is enabled from the

UTIL

menu this button is used to display the A-Scan in the upper half of the display and parameters in the lower half as shown below:

In this mode the main parameters are shown in a grid that the user navigates by using the grey buttons next to the  &  arrows to scroll up or down and the and buttons to move form left to right. Once the required parameter has been selected it can be changed in the normal way by pressing the , and buttons.

Page 9

Masterscan 350 User’s Guide ~ Quick Start

Press to select the dB step value of amplifier gain as 0.1, 0.5, 1, 2, 6, 10, 14, or

20dB. The selected value is indicated at the top right-hand corner of the gain box, which is always located at the bottom right side of the screen. This is a momentary button with no repeat. Corresponds to the letter “X” when using a keyboard.

Press to increment or decrement the gain value indicated in the gain box.

This is always located at the bottom right side of the screen. This is a repeat button with acceleration to facilitate quick changing of the value.

Corresponds to “Page Up” and “Page Down” when using a keyboard.

Selects either reference or gain settings within the gain box. When reference is selected, the increment and decrement buttons will change both the reference and gain values together. This is a momentary button with no repeat. Corresponds to the letter “R” when using a keyboard.

This button toggles between single and double element transducers. The user should switch the unit off or stop the pulser by pressing the cobbwbLmb^h

key before changing transducers. Corresponds to the letter

“S” when using a keyboard.

BNC or LEMO 1 connector is the transmitter and receiver socket used for single transducers or as the transmitter only, for twin or dual transducer operation.

BNC or LEMO 1 connector is the receiver socket used for twin or dual transducer operation.

A socket to connect various USB devices such as a printer, keyboard or the user’s PC.

A socket used to connect the battery charger for recharging the battery pack.

A red dot is provided on both the socket and plug to facilitate alignment.

Refer to Power Supply for information on power supplies and charging of the battery pack.

Page 10

Changes between Normal

& Full screen views

Changes between

Normal, Freeze & Peak display of A-Scan

Displays context specific help

USB Connector

Charge Connector

On/Off Switch

Masterscan 350 User’s Guide ~ Quick Start

Used as acceptance of memory functions and when drawing various measurement curves

4 “Soft” keys

Move menu selection Left

Select main menu group

Move menu selection Right

Decrease parameter values

Increase parameter values

Change step size of parameter adjustments

Increase gain

Decrease gain

Switch between

Reference & Gain

Change gain adjustment step size

Switch between single and double element transducers

Page 11

Masterscan 350 User’s Guide ~ Quick Start

3.2 Flaw Detection

Perform the following steps to establish a basic flaw detection mode for the Masterscan 350.

Units shown are in inches. For metric units, select

METRIC UNITS

from the

UTIL

menu and use the corresponding values for the parameters.

1. Select a suitable transducer, preferably a 5MHz, half inch diameter narrow band.

2. When making a quick adjustment on a menu parameter, press the button until the double arrow appears next to the parameter name. This establishes the fast scroll using the and buttons.

3. In the

CAL

menu, set the following parameters:

ZERO

to

0.000

VEL

should be set the specimen material velocity.

RANGE

to

125

or other suitable value to cover the test range of interest.

DELAY

to

0.000

GAIN

to

50.0

4. In the

AMP

menu, set the following parameters:

DETECT

to

FULL

PRF

to

150Hz

5. In the

GATE1

menu, set the following parameters:

STATE

to

ON +VE

START

to

10.0

WIDTH

to

50

or other suitable value to cover the test range of interest.

LEVEL

to

50.0

6. In the

GATE2

menu, set the

STATE

to

OFF

7. In the

MEAS

menu, set the following parameters:

MODE

to

DEPTH

TRIGGER

to

FLANK

HUD

to

OFF

T-MIN

to

OFF

The Masterscan 350 is now configured for basic flaw detection. Using an appropriate calibration block, adjust the

GAIN

parameter to establish the correct sensitivity. Adjust other parameters as necessary to optimize the calibration. For more in-depth features of the

Masterscan 350, see Flaw Testing on page 45 .

Page 12

Masterscan 350 User’s Guide ~ Quick Start

3.3 Thickness Gauging

Perform the following steps to establish a basic thickness-gauging mode for the Masterscan

350. Units shown are in metric. For inch units, select

INCHES

from the

UTIL

menu and use the corresponding values for the parameters.

1. Select a suitable transducer, preferably a broadband, 5MHz, with a 10mm diameter.

2. Select an appropriate calibration block with at least three known thickness sections covering the range to be inspected and made from the same material as the test piece.

3. When making a quick adjustment on a menu parameter, press the button until the double arrow appears next to the parameter name. This establishes the fast scroll using the and buttons.

4. In the

CAL

menu, set the following parameters:

ZERO

can be left alone

VEL

should be set the specimen material velocity.

RANGE

to

125

or other suitable value to cover the test range of interest.

DELAY

to

0.000

GAIN

to

50.0

5. In the

AMP

menu, set the following parameters:

DETECT

to

+VE HW

PRF

to

150Hz

6. In the

GATE1

menu, set the following parameters:

STATE

to

+VE

START

to

10

WIDTH

to

50

or other suitable value to cover the test range of interest.

LEVEL

to

25.0

7. In the

MEAS

menu, set the following parameters:

MODE

to

DEPTH

TRIGGER

to

FLANK

HUD

to

ON

T-MIN

to

OFF

8. Calibrate the thickness readout on the selected calibration block using the procedure in

A-Cal on page 82 .

Page 13

Masterscan 350 User’s Guide ~ Quick Start

The Masterscan 350 is now configured for basic thickness gauging. Adjust parameters as necessary to optimize the calibration. For more in-depth features of the Masterscan 350, see

Thickness Gauging on page 79 .

Page 14

Masterscan 350 User’s Guide ~ Quick Start

3.4 Masterscan 350 Memory

The settings of the Masterscan 350 always remain in memory when the instrument is turned off, even if the battery pack is removed. That is, whatever the settings are just prior to turning the instrument off will be the settings in place the next time the instrument is turned on.

At times it may be desirable to start with default settings. This is especially true when beginning a new test procedure or going from flaw detection to a thickness gauging procedure. Otherwise, it may be necessary to go through all of the menus to reset various functions. A reset function is provided to facilitate the returning of all panel calibration settings to the factory defaults.

3.4.1 Reset to factory defaults:

1. Switch the instrument off.

2. Depress the crii=p`obbk

button and hold while switching the instrument on until the reset display is seen.

3. Press the button to reset the instrument to factory defaults

NOTE: Before performing this procedure, be sure to save any favourite settings to memory by using the procedure outlined in Storage & Recall of Calibration Setups on page 41 .

NOTE: When resetting to factory default settings the following dialogue is displayed allowing the user to choose imperial or metric defaults:

NOTE: The video always defaults to PAL.

3.4.2 Clearing the Memory

The Masterscan 350’s memory can store A-Scans, thicknesses and calibration set-ups.

Erasing these values accidentally could have serious consequences and should only be done after suitable consideration. To erase the memory follow this sequence:

1. Switch the instrument off.

2. Depress the jbkr

button and hold while switching the instrument on until the reset display is seen.

3. Press the button to clear the Masterscan 350’s memory

Page 15

Masterscan 350 User’s Guide ~ Detailed Menu Description

4 Detailed Menu Description

Before proceeding with this section, the user should be familiar with the front panel controls described in section 3.1

Front Panel and Optional Keyboard Controls on page 6 . It is also assumed that the user has a good understanding of the theory and practice of ultrasonic testing.

4.1 Menu Tree

This diagram represents the menu structure of the Masterscan 350; there are 4 main menu items with sub-menus below. The sub-menus are described in the following section.

CAL MEAS UTIL MEMORY

ZERO

VEL

RANGE

DELAY

FREQ

DETECT

REJECT

CONTOUR

WIDTH

EDGE

VOLTS

PRF MAX

STATE

START

WIDTH

LEVEL

STATE

START

WIDTH

LEVEL

DIST1

DIST2

ACCEPT

START

MONITOR

DEPTH

TRIGGER

HUD

T-MIN

E-E

TRIGGER

HUD

BLANK

TRIG

TRIGGER

HUD

THICK

G1-G2

TRIGGER

HUD

MODE

SURFACE

RAD

ANGLE

X_OFF

½ SKIP

UNITS

CLICK

ALARM

SMOOTH

COLOUR

BRIGHT

VIDEO

GRATICLE

LANG

SPLIT

BAUD

KEYLOCK

OFF

DRAW

CURSOR

ON

ATTEN

CURVE

OFF

% FSH

+-% TOL

ON

% FSH

+-% TOL

OFF

DISPLAY

CURR SET

A-LOG

T-LOG

BLOCK

LOC

SEQUENCE

STORE

STORE

STATE

EDIT

RECALL

STATE

EDIT

DELETE

STATE

EDIT

NOTES

STATE

STORE

STORE

STATE

EDIT

RECALL

STATE

EDIT

DELETE

STATE

EDIT

NOTES

STATE

STORE

OFF

STATE

RECALL

STATE

NOTES

STATE

NEXT

INSERT

DELETE

EXIT

NEXT

INSERT

DELETE

EXIT

OFF

SPLIT

DRAW

CURSOR

POINT

CURSOR

ON

CURVE

TRIGGER

MEAS

OFF

FREQ

TBVEL

ERS

SET PRB

NFL

DELAY

PED

SET ATT

T-LOSS

REF

MAT

SET REF dVK

REF TYPE

REFSIZ

SET SIG

CURSOR

CURSOR

SPLIT

ON

TRIGGER

ERS

T-LOSS

OFF

DRAW

CURSOR

POINT

CURSOR

ON

CURVE

MEAS

OFF

SET

REF %

CURSOR

CURSOR

MEAS

IL (A)

AF (C)

IR (D)

O/P 1

O/P 2

CAL

OFF

TIME

HOURS

MINS

DATE

DATE

MONTH

YEAR

OFF

NUMERIC

BLOCK

LOC

NO

DELETE

BLOCK

LOC

SEQUENCE

BLOCK

LOC

NOTES

OFF

CAL dr

START

MEAS

START k di

Page 16

Masterscan 350 User’s Guide ~Detailed Menu Description

When The Masterscan 350 is switched on, an information screen is displayed momentarily showing the instrument serial number and software version number while a self-test is performed as shown below:

Page 17

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.2 Main Menu

After the information screen disappears the main menu selections are shown on the right hand side of the screen. To select a particular menu press the plain grey button to the right of it. You can return to the main menu selection at any time by pressing the jbkr

button.

4.2.1 Main Menu Selection

CAL

MEAS

UTIL

Used to select the Calibration menu items:

CAL

,

AMP

,

TX

,

GATE1

,

GATE2

and

A_CAL

Used to select the Measurement menu items:

MEAS

,

CSC

,

PROBE

,

DAC

,

AVG

,

TCG

,

AWS

and

API

Used to select the Utilities menu items:

UTIL

,

VIDEO

,

MISC

,

BEA

,

AGC

,

PRINT

,

P_O/P

and

CLOCK

MEMORY

Used to select the Memory menu items:

PANEL

,

A-LOG

,

REF

and

T-LOG

.

Note: The

REF

/

GAIN

box always appears at the bottom right of the screen and can be changed using the pqbm

, obcLd^fk

and the and keys.

Throughout the menus there are parameter boxes that have a thick line underneath them as shown here:

Pressing the grey button next to one of these parameter boxes produces a dropdown list as shown here:

Press the grey button again to remove the drop-down list after the required value has been selected.

Page 18

Masterscan 350 User’s Guide ~Detailed Menu Description

4.3 CAL Menu

The CAL menu is the most used menu and contains those items that allow the Masterscan

350 to be easily calibrated. For full instruction on how to perform a calibration see Flaw

Testing Calibration on page 46

4.3.1 Calibration Menu

ZERO

VEL

RANGE

Used to calibrate the screen and thickness readout for zero offsets that are inherently different for each transducer. Units are microseconds in both mm and inch modes.

Used to calibrate the screen and thickness span readout based on the velocity of sound in the test material. Units are meters per second in mm mode and inches per microsecond in inch mode.

Used to set the full screen width of the horizontal A-Trace in mm, inches, or microseconds depending on the Units chosen in the

UTIL

menu. The range is 1mm to 20 meters (0.05 to

800 inches).

DELAY

Used to set the delay or offset of the left side of the A-Trace for viewing of a portion of a signal. The range is 0 to 20 meters (0 to 800 inches).

4.3.2 Amplifier Menu

FREQ

DETECT

REJECT

CONTOUR

Used to set the centre frequency band of the amplifier to match the transducer. Ranges are 500 kHz, 1, 2.5 5, 10 &

15MHz narrow bands and 1-35MHz WIDE band.

Used to set the display mode for the desired rectification of the signals from

FULL

(rectified full wave),

RF

( unrectified),

-VE HW

(negative half-wave) and

+VE HW

(positive half-wave)

Used to remove low level noise from the A-Trace. Reject is linear and is adjustable from 0 up to 80% of full screen height. The obgb`q

LED is lit when a reject is active.

Used to contour the shape of the echo signals, adjustable from 0 or no contouring to a level of 6 maximum contouring. See Contour , Peak Echo & Hold Dynamics on page 76 for a full description.

Page 19

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.3.3 TX (Transmitter) Menu

TX WIDTH

Used to adjust the width of the square wave pulser to match the transducer. Adjustable in 10 or 1ns steps from

20 to 2,000ns. Usually set to match the transducer frequency at ½ its wavelength. Thus, the nominal settings would be:

500kHz 1,000ns

1MHz 500ns

2MHz 250ns

5MHz 100ns

10MHz 50ns

15MHz 30ns

EDGE

Adjust to optimize the shape and amplitude of echoes.

Used to change the ActiveEdge™ parameter.

VOLTS

PRF MAX

Allows switching between 100 and 450 Volts in 50 Volt steps. Note that in 400V operation the pulse width is limited to a minimum of 100ns

Used to set the maximum pulse repetition frequency that may be reduced for large values of display range and delay.

Selectable values are from 15 to 100 in 5Hz steps, from 100 to 1,000 in 50Hz steps and 1,000 to 5,000 in 100Hz steps.

Lower values will reduce ghosting and noise echoes.

NOTE The TX has an overload protection circuit which monitors output power. If the output is overloaded a message is shown on the display:

If this occurs reduce either TX WIDTH, PRF MAX, or decrease VOLTS. No damage will occur in overload, but the output voltage may be reduced.

Page 20

Masterscan 350 User’s Guide ~Detailed Menu Description

4.3.4 GATE1 & GATE2 Menus

There are two gates available in the Masterscan 350 with slightly different options depending on the gate and the measurement mode chosen. The following describes the available settings for each gate with the differences between the gates stated. The gates can be identified by the fact that gate 1 has the number “1” above the line and gate 2 has the number “2” above it.

STATE

Set the state of the gate as follows:

OFF:

Switches the gate off.

+VE:

The alarm triggers when an echo in the gate exceeds the threshold level.

-VE:

The alarm triggers when an echo in the gate falls below the threshold level. Usually used to monitor for loss of back wall echo.

EXPAND:

Expands the gate width to fill the horizontal display width. Only applies to gate 1.

START

WIDTH

LEVEL

Used to set the start position of the gate relative to the initial pulse.

Units are mm or inches and range is from 0 to the full time base of the horizontal display.

Used to set the width of the gate. Units are mm or inches and range is from 0.15mm (0.001 inch), depending on the range selected, to the full time base of the horizontal display.

Used to adjust the alarm threshold level, which corresponds to the vertical height on the A-Trace. Adjustable in 0.5% or

2% steps from 0% to 100% full screen height.

Page 21

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.3.5 A-CAL Menu

This menu provides automatic calibration of sound velocity and transducer zero. Gate 1 is used to select the reference echoes. See A-Cal on page 82 for a full procedure.

DIST1

The actual distance to the first or thinnest reference echo in the calibration block.

DIST2

ACCEPT

The actual distance to the second or thickest reference echo in the calibration block

After pressing lh= to accept

DIST1

and

DIST2

. Press lh= when

ACCEPT CAL

is highlighted.

START

Used to adjust the start of the gate to assure that the first and second echoes are measured.

Page 22

Masterscan 350 User’s Guide ~Detailed Menu Description

4.4 MEAS Menu

The menus in this section allow the various measurement techniques to be configured; these include the general measurement set-up and specialist methods such as DAC, AVG,

TCG, AWS and API. Select the

MEAS

menu at any time by pressing the jbkr

button.

4.4.1 Measurement Menu

In the measurement menu, the top selection box shows the selected measurement mode and the remaining three selection boxes vary depending on the mode selected as follows:

MODE:

MONITOR

In this mode, gates 1 and 2 act as two independent monitor gates

MODE:

DEPTH

TRIGGER

In this mode, gate 1 functions as a depth or thickness monitor and displays the depth (D:) and height (H:) of the first signal after the start of the gate that reaches or exceeds the gate level threshold. Values are displayed in a highlighted box below the A-Trace.

Used to select the depth or thickness measurement to the

FLANK

(left edge) of the first echo after the start of the gate or the

PEAK

measurement within the gate i.e. the largest value in the gate.

HUD

When turned

ON

, provides a large, Head-Up Display of the depth or thickness reading at the top right of the A-Trace.

The button sequences the selections as follows:

G1-G2 E-E DEPTH TRIG

OFF OFF OFF OFF

DIST DIST DIST DIST

DIST + DIST + HEIGHT

DIST +

HEIGHT

DIST +

HEIGHT + HEIGHT +

+

+

T-MIN

When turned ON, the depth or thickness reading will freeze that last minimum or lowest value measured. To reset, toggle the function to OFF and then ON. Only available in DEPTH mode.

Page 23

Masterscan 350 User’s Guide ~ Detailed Menu Description

MODE:

E-E

BLANK

MODE:

TRIG

THICK

MODE:

G1-G2

In this mode, gate 1 functions as a thickness monitor and measures the thickness between the first signal in the gate and the second signal in the gate that reaches or exceeds the level threshold. A second bar is shown representing the blanking (see BLANK)

This function sets the blanking distance, as a percentage of the total gate width, which is a blind zone after the first echo, after which a second echo can be measured. This helps to eliminate undesired noise in the first echo from being measured, as thickness but will limit the minimum thickness capability if set too large

The Trigonometry mode is used with angle beam transducers for weld inspection to calculate the three important measurements based on the echo position: the Beam path distance (:), the Surface distance (:), and the Depth distance (:) from the index point of the transducer.

Set to the thickness of the material being tested to account for multiple skips of the angled sound beam in the test material.

This is similar in concept to E-E mode, but uses gates 1 & 2, allowing the thresholds for the two gate measurements to be completely independent.

Page 24

4.4.2 CSC (Curved Surface Correction)

Masterscan 350 User’s Guide ~Detailed Menu Description

This menu allows the user to configure the Curved Surface Correction (CSC) parameters.

This feature is used when testing pipes and other curved surfaces. In order for the correction to be made the correct wall thickness must be selected in the

THICK

option when

TRIG

is selected in the

MEAS

menu. CSC is only active when in TRIG measurement mode.

MODE

SURFACE

Turn CSC

ON

or

OFF

.

Choose CONCAVE for internal surfaces or CONVEX for external surfaces

RAD

Select the radius of the curvature.

4.4.3 PROBE Menu

ANGLE

X-OFFSET

Set to the nominal refracted angle of the transducer to calibrate the Surface and Depth measurements

Used to enter the distance from probe emission point to front of probe case. This is used by the TRIG function in

MEAS menu to give the surface distance.

1/2 SKIP

Used as a quick way to turn on the ½ skip graticule.

Page 25

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.4.4 Distance Amplitude Correction (DAC) Menu

This menu is used to create DAC curves using a series of reference echoes. Once drawn, the DAC curve acts as an alarm threshold level for the gate where the level varies to match the attenuation and field characteristics of the transducer and test material combination.

The DAC procedure is described fully in DAC Operation on page 50.

MODE:

OFF

Turns the DAC measurement mode off.

SPLIT

Activates or deactivates the split DAC function where an additional DAC curve is generated if the signal drops below

20%

MODE: ON

Displays the DAC curve on the screen

CURVE

TRIGGER

MEAS

MODE:

DRAW

Used to display the DAC curve alone, the -6/-12dB, -6/-14dB, -

2/-6/-10dB, -2dB, -6dB or -10dB reference curves.

Used to set the alarm threshold for the DAC curve, -2dB, -

6dB, -10dB, -12dB or -14dB curve or the gate.

Used to select the measurement value in dB, % Full Screen

Height (FSH) or %REF for any signal that is in the curve or gate.

Used to create the DAC curve.

CURSOR

POINT

CURSOR

Used to move the cursor over the reference echo for which a

DAC point is being set.

Display only. Shows the number of points created after pressing the lh= button.

Toggles the width of the selection cursor between 25 pixels

(default) and 5 pixels for more accurate selection of peaks.

Page 26

Masterscan 350 User’s Guide ~Detailed Menu Description

4.4.5 DGS/AVG

This menu is used to create a DGS/AVG curve allowing flaw sizing and distance compensation without requiring the reference standards for establishing a DAC curve or to set up a TCG correction. DGS/AVG is described fully in DGS /AVG Operation on page 62.

MODE:

OFF

Initial mode. Turns off the display of the DGS/AVG curve

FREQ

TBVEL

ERS

Used to set the transducer frequency taken from its datasheet.

Used to set the reference (test) block velocity in m/s

Used to set the Equivalent Reflector Size.

MODE:

SET PRB

NFL

DELAY

PED

Mode used to define curvature correction and probe delay

Used to set the Near Field Length taken from the transducer datasheet

Used to set the probe delay material velocity in m/s and is used in conjunction with the probe zero and specimen velocity to calculate the sound field equivalent length.

Calculated Probe Effective Diameter – This value is for information only and cannot be changed

MODE:

SET ATT

T-LOSS

REF

MAT

Used to set attenuation parameters

Used to set the transfer loss in dB due to surface condition, poor coupling etc…

Used to set the reference material attenuation in dB/m

Used to set the test material attenuation in dB/m

Page 27

Page 28

Masterscan 350 User’s Guide ~ Detailed Menu Description

MODE:

SET REF

Used to define the various reference values

dVK

Used to set the curvature correction factor in dB and is taken from the transducer datasheet.

REF TYPE

Used to select the reference reflector type: Flat Bottom Hole

(

FBH

), Side Drilled Hole (

SDH

) or Back Wall Echo (

BWE

)

REFSIZ

Used to define the size of the reference type. If

REF TYPE

is set to

BWE

,

REFSIZE

is set to

INFINITE

MODE:

SET SIG

CURSOR

CURSOR

SPLIT

Mode used to perform calibration

Select than use the and keys to move the selection over the echo and press

OK

. The Masterscan 380 will calculate the sensitivity and draw the curve on the screen

Used to change the width of the cursor between 25 pixels and 5 pixels

Select to activate or deactivate multiple DGS/AVG curves.

MODE: ON

Mode in which DGS/AVG is on.

TRIGGER

ERS

Used to select either the

GATE

or the

CURVE

as the measurement reference

Used to set the Equivalent Reflector Size.

T-LOSS

Used to set the transfer loss in dB due to surface condition, poor coupling etc…

Masterscan 350 User’s Guide ~Detailed Menu Description

4.4.6 TCG Time Corrected Gain Menu

This menu is used to create TCG curves using a series of reference echoes. Once drawn, the TCG curve acts as swept gain control on the amplifier to set different gain levels relative to distance. The TCG procedure is described fully in TCG Operation on page 53

MODE:

DRAW

CURSOR

Used to create the TCG curve. Cannot be drawn in RF display mode.

Used to move the cursor over the reference echo for which a

TCG point is being set

POINT

CURSOR

Display only. Shows the last point created after pressing the lh= button.

Used to change the width of the cursor between 25 pixels and 5 pixels

MODE: ON

CURVE

Activates the TCG curve to adjust the gain. Can be activated in RF or rectified display modes.

When switched ON, the TCG curve is displayed.

MEAS

Used to select between dB, %FSH and %REF for measurements

Page 29

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.4.7 AWS Menu

This menu is used when performing weld inspection in accordance with the American

Welding Society’s Structural Welding Code, ANSI/AWS D1.1-94. It provides a convenient method to automatically calculate the Indication Rating as defined in the code. The AWS procedure is described fully in Weld Inspection Using the AWS Menu on page 57

MODE:

SET

Used to set up the AWS measurement mode.

REF

CURSOR

MODE:

MEAS

Used to set the Indication Level.

Used to move the cursor over the reference echo to set the reference level.

Used to make measurements in accordance with the code.

IL (A)

AF (C)

IR (D)

The dB required setting the indication to the reference level.

The attenuation factor to correct for the depth of the indication.

The indication rating calculated in accordance with the code.

Also the difference the IL and the reference level with correction for attenuation.

Page 30

Masterscan 350 User’s Guide ~Detailed Menu Description

4.4.8 API Menu

This menu is used to assist users in performing tests according to the American Petroleum

Institute (API) Recommended Practice for Ultrasonic Evaluation of Pipe Imperfections using the Amplitude Distance Differential Method (ADDM). Users are referred to the publication: “API Recommended Practice 5UE, Second Edition, June 2005” The API procedure is described in Evaluation of Pipe Imperfections using API 5UE on page 59

MODE:

CAL

Used to set the initial calibration of the defect.

dr

START

LEVEL

MODE:

MEAS

START k di

Depth of the reference indicator; should be set by the user according to the size of the reference defect.

The position of the gate that is used to set the calibration.

The height of the gate that is used to set the calibration.

Used once the calibration has been done to make comparative measurements.

Used to position the start of the gate used for making measurements.

The k factor derived during the calibration phase and used to calculate the depth of the defect. It can’t be changed during the measurement phase and id displayed for reference.

Imperfection depth calculated by the Masterscan 350.

Page 31

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.5 UTIL Menu

Press the jbkr

button followed by the plain gray button next to

UTIL

to select these menu options. The items in this section allow the user to configure features such as units, display preferences, language, back wall echo attenuation (BEA), print options, proportional output configuration and date and time settings.

4.5.1 UTIL Menu

UNITS

CLICK

ALARM

SMOOTH

Selects

INCHES

,

µs

, or

METRIC

measurement units.

When in

µs

mode, the velocity is fixed at 2000 m/s

(5000in/μs) and is not adjustable.

When on, a beep will sound to confirm each button press.

When set to AUDIBLE, a buzzer sounds during any gate alarm.

When

SMOOTH

is selected the signal is displayed as an envelope. When

FILL

is selected the area under the trace is filled.

4.5.2 VIDEO Menu

COLOUR

BRIGHT

Selects one of eight colour schemes for the display

Controls the display brightness. The level selected affects the battery life. A value of 1 gives 16 hours, 10 gives 12½ hours and a value of 20 will give 9 hours of operation, assuming the battery is in good condition at 22°C.

VIDEO

Select NTSC (USA) or PAL (UK/Europe) composite video output modes. The screen update rate is 50Hz in PAL and

60Hz in NTSC.

GRATICULE

Selects graticule options from the following choices:

OFF

ON

: graticule.

A complete 100% graticule

SPARSE

: A row of dots at 10% intervals.

50%

: Divisions at 50% full screen height.

1/2 SKIP

: Sparse with vertical dotted lines representing the ½ skip, full skip and 1½ skip distances.

Press again to get the ½ skip lines with calculated distances shown

Page 32

Masterscan 350 User’s Guide ~Detailed Menu Description

4.5.3 MISC Menu

LANG

SPLIT

BAUD

Selects one of 6 built in languages or a user selectable language. The built in languages are selectable at time of ordering the Masterscan 350.

Allows the user to display a vertically compressed A-Scan in the upper half of the screen and a sub-set of the menus in the lower half. When set to

ACTIVE

the

=crii=p`obbk button switches between normal, full and split screen views.

Allows the user to select a Baud rate for the serial port to operate at. The choices are: 9,600 (

9K6

), 19,200 (

19K2

),

38,400 (

38K4

), 57,600 (

57K6

) or 115,200 (

115K2

).

KEYLOCK

Allows the user to “lock out” the keys on the unit to stop inadvertent changes. When set to

ACTIVE

the button toggles between freeze, peak, hold and keylock.

4.5.4 BEA (Backwall Echo Attenuator)

MODE

ATTEN

CURSOR

CURVE

Allows selection between

OFF

,

DRAW

and

ON

In

ON

mode this is used to select the level of attenuation to apply to the backwall. The value can be selected between

0dB and the REF gain value to a maximum of 40dB

In

DRAW

mode this is used to adjust the cursor so that it covers the area on the display that corresponds to the backwall. The cursor starts from the right hand side of the display and is made wider by using the button and made narrower by using the button. The attenuation is reduced by the selected

ATTEN

value for the width of the cursor.

Used to display a line representing the amount of attenuation applied

Page 33

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.5.5 AGC (Automatic Gain Control)

Automatic Gain Control feature is generally used in thickness logging.

By setting a percentage of full screen height (

% FSH

– default 50%) and ensuring that the signal is breaking gate 1, the gain will be adjusted automatically to bring the signal height to the user defined value subject to an adjustable tolerance (

+–% TOL

).

AGC MODE

Turn AGC

ON

of

OFF

% FSH

+-% TOL

Choose the screen height, between 10% and 90%, to which the user wants the gain to be adjusted.

Choose the tolerance on the % FSH to stop the AGC

“hunting” due to signal height jitter. The range can be set between 5% and 20% FSH

To Use:

 Set Gate 1

+ve

, to a height of, say 10% or 20% FSH and ensure that the signal will cross the gate.

 Ensure that

MEAS

mode is anything but

MONITOR

 Set the

%FSH

parameter to the signal height required.

 Enable AGC by changing AGC Mode to

ON

.

 For poor surface conditions, adjust the

+-% TOL

to increase the tolerance of variation in the signal before the gain is adjusted

 When a signal is present in Gate 1 and a suitable measurement mode is selected the gain will be automatically adjusted to bring the signal to the specified height on the screen.

NOTES:

 When the Masterscan 350 is switched on the AGC is turned OFF.

 The gain value that is last set by AGC remains when AGC is switched off.

 If the gate level is above the user defined %FSH, AGC is internally inhibited.

Page 34

Masterscan 350 User’s Guide ~Detailed Menu Description

4.5.6 PRINT Menu

This menu allows printing of information to a suitable USB printer connected to the USB port on the Masterscan 350. Please note that this requires a printer with a USB port from the Sonatest certified list of printers.

OFF

The print mode is switched OFF

DISPLAY

CURR SET

A-LOG

T-LOG

With the print mode in

DISPLAY

, the screen display will be sent to the printer when the lh= button is pressed.

With the print mode in

CURR SET

, the current calibration settings, the screen and all notes are sent to the printer when the lh= button is pressed.

In the

A-Log

print mode, all of the stored A-Scans that are valid will be printed when the lh button is pressed, along with the calibration settings and notes. This can take considerable time if all 800 A-Scans are stored.

In the

T-Log

print mode, all of the stored thickness measurements are printed when the lh

button is pressed.

This can take considerable time if all 8,000 thicknesses are stored

SEQUENCE

In the

SEQUENCE

print mode, all of the sequence thicknesses are printed along when the lh

button is pressed.

This can take considerable time if all 8,000 thicknesses are stored.

Page 35

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.5.7 P_O/P Proportional Output Control Menu

This menu controls the analogue and digital proportional outputs. Two outputs are provided. The outputs are active when the associated gate is triggered and are updated at the PRF.

O/P 1

DEPTH1

: (Gate1)

and

DEPTH2

: (Gate2)

O/P 2

E1-E2

: Distance between the two depth measurements

ALARM 1

: Voltage output when alarm 1 is active (O/P 1 only)

LEVEL1

: (Gate1)

LEVEL2

: (Gate2)

L1-L2

: The difference between the two amplitudes

ALARM 2

: Voltage output when alarm 2 is active (O/P 2 only)

CAL

Allows the user to change the voltage on the proportional outputs between 0V and 10.1V

4.5.8 CLOCK Menu

SET OFF

SET TIME

Do not display the time at the bottom of the display

Used to set the current time.

HOURS

MINS

Used to set the current hour in 24 hour format

Used to set the current minutes.

SET DATE

DATE

Used to set the current date.(Selected by pressing the plain grey button next to

SET

, then )

Used to set the current day.

MONTH

YEAR

Used to set the current month.

Used to set the current year.

Page 36

Masterscan 350 User’s Guide ~Detailed Menu Description

4.6 MEMORY Menu

Press the jbkr

button then the plain grey button next to the

MEMORY

option to access the Memory menu. These menus allow the user to save panel settings, A-Scans and thicknesses as well as recalling an A-Scan to be used as a reference.

4.6.1 PANEL Memory Menu

This menu provides storage and recall of up to 100 calibration settings. The use of this feature is described fully in Storage & Recall of Calibration Setups on page 41

STORE

Used to select a storage location (1-100). Pressing the grey button twice lists all of the locations along with the associated notes. Use & to scroll through the list; pressing the lh

key will recall the selected panel.

MODE:

STORE

MODE:

RECALL

MODE:

DELETE

MODE:

NOTES

STATE

EDIT

Stores the current calibration settings in memory to the selected Store set. Press lh= to store. Press lh= again after the confirmation prompt.

Used to recall a stored calibration set to the active memory.

Press lh= to recall

Used to delete a

VALID

(used) calibration set. Press lh= to delete. Press lh= again after the confirmation prompt

Shows the notes associated with the stored calibration set.

Pressing lh= will recall the calibration set at which time the

Notes can be edited.

Display only.

VALID

indicates that the store location is used.

EMPTY

indicates that it is not used. To change

VALID

to

EMPTY

, use the

DELETE

function.

(See Adding NOTES to PANEL and A-LOG sets on page 41 .)

Page 37

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.6.2 A-LOG Memory Menu

This menu provides storage and recall of up to 800 A-Scans with settings. The use of this feature is described fully in A-LOG , A-Scan Storage on page 73

STORE

Used to select a storage location (1-800). Pressing the grey button twice lists all of the locations along with the associated notes. Use & to scroll through the list; pressing the lh

key will recall the selected panel.

MODE:

STORE

MODE:

RECALL

Stores the current A-Scan and its settings in memory to the selected Store set. Press lh= to store. Press lh= again after the confirmation prompt.

Used to recall a stored A-Scan and settings to the active memory. Press lh= to recall. The A-Scan is displayed in

MODE:

DELETE

MODE:

NOTES

STATE

EDIT

FREEZE mode. Press to remove.

Used to delete a VALID (used) A-scan and its settings. Press lh= to delete. Press lh= again after the confirmation prompt.

Shows the notes associated with the stored A-Scan and its settings. Pressing lh= will recall the calibration set at which time the Notes can be edited.

Display only. VALID indicates that the store location is used.

EMPTY indicates that it is not used. To change VALID to

EMPTY, use the DELETE function.

(See Adding NOTES to PANEL and A-LOG sets on page 41 .)

Page 38

Masterscan 350 User’s Guide ~Detailed Menu Description

4.6.3 REF Menu

This menu allows a waveform stored in A-LOG memory to be displayed as a reference on the display. Before a waveform can be recalled, it must be stored in an A-LOG location as described above.

STORE

Used to select an A-LOG storage location (1-800). Pressing the grey button twice lists all of the locations along with the associated notes. Use & to scroll through the list; pressing the lh

key will recall the selected panel.

MODE:

OFF

MODE:

RECALL

Used to remove a Reference: waveform. Set

MODE

to

OFF

, and press lh

.

Used to recall the stored A-Scan and display it. Press lh= to recall.

MODE:

NOTES

Shows the notes associated with the: stored A-Scan. Pressing lh= will recall the waveform.

STATE

Display only.

VALID

indicates that the store location is used.

EMPTY

indicates that it is not used.

4.6.4 T-LOG Menu Numeric Mode

This menu provides storage and recall of up to 8,000 thickness readings. Numeric and

Sequential modes are available. This feature is described fully in T-LOG Thickness Storage on page 84

Readings are stored under a three level code, starting with BLOCK (1-14), followed by

Location (1-2000) and NO (1-2000).

MODE

Used to select the T-Logging mode, between OFF,

NUMERIC

, DELETE and SEQUENCE.

BLOCK

LOC

NO

Used to select the block number (1-14) for storage or viewing of a thickness reading.

Used to select the location number for storage or viewing of a thickness reading.

The individual thickness reading number into which a thickness reading is stored. Increments automatically when storing a reading by pressing the lh= button.

Page 39

Masterscan 350 User’s Guide ~ Detailed Menu Description

4.6.5 T-LOG Menu Sequential Mode

Readings are stored by block (1-14) and location (1-8000) number combination only. In this mode location and historical thickness data can be downloaded to the Masterscan 350 from a computer, a fixed series of four notes can be added to a location. This feature is described fully in T-LOG Thickness Storage (Sequential & Download) on page 85 .

MODE

Used to select the T-Logging mode, between OFF,

NUMERIC

, DELETE and SEQUENCE.

BLOCK

LOC

NOTES

Used to select the block number (1-14) for storage or viewing of a thickness reading.

Used to select the location number for storage or viewing of a thickness reading.

One of four notes can be selected or blank.

OBSTRC

for obstruction,

PITTING

,

POOR S/C

for poor surface contact, and

NO BWE

for no back wall echo.

4.6.6 T-LOG Menu Delete Mode

MODE

BLOCK

LOC

Used to select the T-Logging mode, between OFF,

NUMERIC

, DELETE and SEQUENCE.

Used to select the block number (1-14), this value can also be set to ALL – for deleting all readings. When block is set to

ALL

, location number is also set to ALL.

Used to select the location number (1-2000), this value can also be set to ALL – for deleting all readings under the specified BLOCK number.

Page 40

Masterscan 350 User’s Guide ~Detailed Menu Description

4.7 Storage & Recall of Calibration Setups

After the Masterscan 350 has been properly calibrated for a particular testing scheme, it is possible to store all of the panel settings for subsequent recall when performing the same test at a later time. Although the last panel settings are “remembered” by the Masterscan

350’s memory, this feature is useful when the instrument is being used for many different tests requiring substantially different panel calibration settings. In addition to the panel settings, a freeform note can be stored with the panel set that includes provision for a label, an operator name or number, a location identifier, and a transducer identifier. It is possible to store up to 100 sets of panel settings; each assigned a set number from 1 to 100. A set of panel settings can be printed after the set is recalled to active memory by using the LIST CPY function in the PRINT menu.

4.7.1 To store a panel set:

Press the jbkr

button followed by the plain grey button next to the

MEMORY

option to access the Memory menu.

Use the and buttons to highlight the PANEL menu.

If it is desired to store Notes with the panel settings, see the description for entering notes below.

Select the STORE box and, using the and buttons, select a desired STORE number from 1 to 100. Note that as the STORE number is changed, the STATE parameter indicates if the number is used (VALID) or EMPTY.

Select the MODE box and using the and buttons, select the STORE mode.

Press the OK button to store the panel settings.

If you forgot to enter Notes, create the Notes as described below and the re-save the panel settings starting at step 5 above.

4.7.2 To recall a panel set:

Follow the above procedure but set the MODE parameter to RECALL and then press the lh= button.

4.7.3 To delete a panel set:

Follow the above procedure but set the MODE parameter to DELETE and press the lh= button.

4.7.4 Adding NOTES to PANEL and A-LOG sets

When storing panel calibration settings or waveforms, it is often useful to add some notes to the set so it can later be identified, or to help the user recall the correct set. This is possible in the

PANEL

and

A-LOG

menus by using the

EDIT NOTES

feature.

Page 41

Masterscan 350 User’s Guide ~ Detailed Menu Description

When a panel calibration or waveform is to be stored, first calibrate the instrument as desired, or capture the waveform by pressing the button. Then select the

EDIT

NOTES

box in the menu and notice that the window shown below overlays the display area.

The

LABEL

line is 8 characters long, and is used by the SDMS PC program to allow quick identification. The

OPERATOR

,

LOCATION

and

PROBE

lines are 25 characters each, while the three

NOTES

lines allow up to 75 characters.

The and buttons are used to move the cursor from right to left along a line of text.

The plain grey button adjacent to the

NEXT LINE

box is used to advance the cursor down to the next line. When the cursor is at the bottom line, the

NEXT LINE

function moves it back to the top line.

To delete a character, position the cursor over it and press the plain grey button adjacent to the

DELETE CHAR

box.

To insert a character space, position the cursor to the right of the desired insert location and press the plain grey button adjacent to the

INSERT CHAR

box.

To select a character from the character map below the notes use the and buttons to move the cursor through the map, pressing the lh button to add the highlighted character to the currently selected position in the notes.

Once the desired notes have been added or edited, press the plain grey button adjacent to the

EXIT

box. The text is now entered but not stored. It is now necessary to store the notes using the methods outlined in the

PANEL

and

A-LOG

sections.

If a stored panel or A-LOG is recalled, its notes will be loaded into the window. The notes can then be viewed and edited using the same procedure detailed above.

Page 42

Masterscan 350 User’s Guide ~Detailed Menu Description

NOTE: When notes are displayed, the menu cursor buttons are disabled. Exit the Notes window to change to another menu.

The following shows a completed notes window:

4.7.5 Edit Notes Summary

The following sections provide a summary of the navigation techniques needed to edit notes using either the front panel keypad or the optional USB keyboard:

4.7.5.1 Using the Front Panel Keypad

EDIT NOTES:

Press to bring up the editing boxes as follows: Use the and buttons to move the cursor horizontally along a text line. Use the and buttons to highlight a character from the map.

Press lh= to write a character to a line.

NEXT LINE

Press to scroll through and select a line for editing.

INSERT CHAR

Inserts a blank character space before the one the cursor is currently on.

DELETE CHAR

Deletes the character under the cursor.

EXIT

Exits back to the PANEL or A-LOG menu. Store the calibration set to save the notes.

4.7.5.2 Using the Optional Keyboard

Keyboard Function Connect the optional keyboard to the USB socket on the front panel of the set. Press

EDIT NOTES

from the keypad to enter the keyboard function.

Small Character &

Numbers

Press any key on the keyboard to edit the characters, the same applies to numbers.

Page 43

Masterscan 350 User’s Guide ~ Detailed Menu Description

Capital Character

& Signs

Insert a Space

Press SHIFT + the wanted character or sign.

Press the space bar to insert a blank character at the position the cursor is currently on.

Delete Character Press the backspace to delete the previous character to the position the cursor is currently on.

Delete Character Press Delete to delete the character under the cursor

Up Arrow Press  to move the cursor up.

Down Arrow Press  to move the cursor down.

Left Arrow

Right Arrow

EXIT

Press  to move the cursor left

Press  to move the cursor right

Press the Esc to exit back to the PANEL or A-LOG menu. Store the calibration set to save the notes.

Page 44

Masterscan 350 User’s Guide ~Flaw Testing

5 Flaw Testing

5.1 Basic Flaw Testing

Reliable flaw detection requires three important considerations:

Proper transducer selection

An accurate reference block of the same material as that being tested with reference holes representing the orientation and sensitivity desired

Proper calibration of the instrument.

Transducers used for flaw detection are usually of the narrow band variety to provide the best possible sensitivity to the anticipated discontinuities. In some cases, broadband varieties are chosen to optimize near or far surface resolution – the ability to separate discontinuities from the front surface or back surface echo. In either case, the frequency is chosen so that the wavelength in the material is optimized for the orientation and size of expected discontinuities. The single element contact transducer is used for general-purpose flaw detection. The angle beam transducer is used for weld inspection, which is covered in

Weld Inspection Using Trigonometry Mode on page 55.

Other specialty transducers including dual and surface wave can be used but these topics are beyond the scope of the manual.

The next requirement for reliable flaw detection is a calibration or reference block. This block should be made of the same material as the parts to be inspected. In other words, it should have the same sound velocity and attenuation characteristics. The calibration block should have surfaces that mimic the parts to be inspected so that attenuation and sensitivity characteristic is similar. Most importantly, the calibration block should have a series of fabricated discontinuities representing those expected to found in the test piece. That is, the size and orientation of the fabricated discontinuities should match the expected natural discontinuities in the test piece. In some cases this may mean flat bottom holes perpendicular to the test surface, side drilled holes parallel to the testing surface, or in some cases, narrow notches representing planar surfaces parallel or perpendicular to the testing surface. Fabricated discontinuities must also be placed near the front and back surface of the calibration block to verify resolution of the transducer and instrument setting combination.

Finally, it is necessary to establish the proper calibration of the Masterscan 350 in order to assure reliable flaw detection.

Page 45

Masterscan 350 User’s Guide ~ Flaw Testing

5.2 Flaw Testing Calibration

The essence of this calibration is to set the pulser, amplifier and gate parameters to provide the necessary sensitivity and resolution. The Masterscan 350 has an automatic calibration technique that sets the necessary parameters by using a calibration block that has known thicknesses; see A-Cal on page 82 for details.

Alternatively a manual method can be used to perform the calibration and the following are the essential steps for basic flaw detection:

1. Select the appropriate single element contact transducer and calibration block that matches the material and expected discontinuities under test.

2. In the

CAL

menu, select the proper

RANGE

and

DELAY

so that several echoes from the expected material depth can be viewed.

3. From the

AMP

menu, set the

FREQ

to the nominal value of the transducer and set the detection mode to

RF

.

4. Use the button to set the transducer to

SINGLE

.

5. Couple the transducer to the calibration block and obtain an echo from a fabricated discontinuity or the back surface and adjust the

GAIN

to set the peak of the echo to about 80% screen height.

6. From the

TX

menu, adjust the

WIDTH

parameter so as to obtain the maximum amplitude from the echo while optimizing the symmetrical shape of the echo. The left hand image below shows the proper echo shape while the right hand image indicates an improper setting of the TX

WIDTH

parameter.

7. From the

TX

menu, adjust the ActiveEdge™ (

EDGE

) parameter to obtain desired echo shape and width for the near surface resolution desired. Check the resolution on a fabricated discontinuity in the calibration block that is near the front surface.

8. From the

AMP

menu, set the

DETECT

parameter to

FULL

and the

CONTOUR

parameter to a value to obtain the desired “filling” of the echo.

Page 46

Masterscan 350 User’s Guide ~Flaw Testing

9. Adjust the gain to set a reference echo from mid thickness of the calibration block to about 80% amplitude.

10. From the

MEAS

menu, set the

MODE

parameter to

G1-G2

and the

TRIGGER

parameter to

PEAK

.

11. From the

GATE1

menu, set the gate to

ON +VE

and adjust the

START

,

WIDTH

and

LEVEL

so that the gate covers the first echo. From the

GATE2

menu set the gate to

ON +VE

and adjust the

START

,

WIDTH

and

LEVEL

so that it covers an echo that is a known multiple of echoes from the first. For instance if a 25mm thick calibration block was being used the first gate would be set on the first backwall echo and the second gate on the 6 below: th

backwall giving a distance between them of 125mm as shown

12. The distance between the echoes is shown in the G-G measurement at the bottom of the display (in this case 125.54). It is unlikely that the value will be exactly correct as the velocity of sound in the material will not match one of the preset values. Select a velocity that is close to the material then adjust it so that the exact distance is obtained as shown below:

Page 47

Masterscan 350 User’s Guide ~ Flaw Testing

13. Next the calibration needs to take account of the distance between the transducer crystal and the front face of the transducer by setting the correct

ZERO

value: From the

MEAS

menu change the

MODE

to

DEPTH

.

14. From the

GATE2

menu change the

MODE

to

OFF

, from the

GATE1

menu change the

START

below:

,

WIDTH

and

LEVEL

values so that the gate covers the 6 th

peak as shown

15. The distance measurement is shown at the bottom of the display (in this case 151.04).

From the

CAL

menu adjust the

ZERO

value until the correct distance is shown as below:

Page 48

Masterscan 350 User’s Guide ~Flaw Testing

16. As a final check move the gate across the other echoes to ensure that the correct distances are shown; From the

GATE1

menu adjust the

START

and

LEVEL

parameters so that the gate covers each echo in turn. Check that the correct values are shown as below:

Now that the Masterscan 350 has been calibrated it can be used to perform inspections using any of the techniques described in the following sections.

For weld inspection using an angle beam transducer, follow the instructions in NOTE: A

TCG curve cannot be drawn while in RF display mode. It should be noted that the range of

TCG is limited only by the dynamic range of the set.

1. Use the MEAS parameter to display the echo height relative to Full Screen Height

(FSH), percentage of TCG Reference (%REF) or dB relative to TCG Reference.

Weld Inspection Using Trigonometry Mode on page 54 .

For storage and printing of A-Scans, refer to A-LOG , A-Scan Storage on page 73

Page 49

Masterscan 350 User’s Guide ~ Flaw Testing

5.3 DAC Operation

Distance Amplitude Correction curves act as monitoring gate triggering thresholds to compensate for attenuation and sound beam characteristics. When properly established, the DAC curve will provide consistent alarming from discontinuities of equal relative size at different depths in the test piece. Functionality is provided to automatically draw 6dB and

14dB or 6dB and 12dB reference curves below the calibrated curve.

It should be noted that since the DAC function only provides a dynamic alarm threshold capability, the dynamic range is limited to about 16dB. Therefore, for materials that attenuate more than 16dB over the depth range of interest, DAC should not be used.

Instead, the TCG function may be more applicable. See TCG Operation on page 53 for the use of the Time Corrected Gain feature.

Proper use of the DAC feature requires a reference block made from the same material as that being examined, with flat-bottom or side-drilled holes of the desired size placed at depths covering the range to be inspected.

To establish a DAC curve, follow the following instructions:

1. Establish the basic calibration of the Masterscan 350 using the desired transducer and the proper reference block. From the

MEAS

menu, select the

DAC

feature.

2. Choose SPLIT mode to be ACTIVE allowing up to 3 DAC curves, complying with

EN 583-2:2001.

3. Select the

MODE

parameter by pressing the plain grey button next to it and use the button to select

DRAW

. The other parameters will now be

CURSOR

and

POINT

.

Note: When in the

MODE

parameter, only press the button to turn the DAC off.

This will erase any logged echoes.

Page 50

Masterscan 350 User’s Guide ~Flaw Testing

4. The DAC system is now ready to accept reference points. Position the transducer to give a maximum echo from the first or topmost reference hole in the reference block.

Be careful to use a consistent amount of coupling and transducer pressure.

5. Use the and buttons to position the bright cursor bar over the echo from the reference hole.

6. Press the lh= button to accept the point. A small square will appear as a reference point on the display showing the peak of the signal just recorded.

7. Repeat steps 6 and 7 for each additional reference hole in the reference block, being careful to be consistent with the amount of coupling and transducer pressure. A maximum of 10 points can be recorded for a DAC curve.

8. Press the plain grey button adjacent to

MODE

and use the button to select

ON

mode. The DAC curve will now be displayed and the other parameters will change to

CURVE

and

TRIGGER

9. Use the

CURVE

parameter and the button to select just the DAC reference curve, the DAC curve along with the -2dB, -6dB, -10dB or -12dB curves, or the DAC curve along with the -6dB and -12dB curves, the -6dB and -14dB curves or -2dB, -6dB and -

10dB curves. These additional, automatically generated, reference curves provide a method for helping to evaluate discontinuities smaller than those in the reference block.

10. Use the

TRIGGER

parameter to select the alarm trigger from Gate 1, the DAC curve, or any of the additional curves listed in the curve parameter above that are currently selected.

If depth measurement is switched on with the DAC feature, the depth mode display below the A-Trace will show the distance to the echo peak and the relative amplitude to the DAC

Page 51

Masterscan 350 User’s Guide ~ Flaw Testing reference curve in dB. Note that the relative dB display is dependent on the REF gain setting, but not the

GAIN

setting.

With split DAC mode active, if the reference echoes drop below 20 %FSH another DAC curve will be generated at that point boosting the gain by 12dB each time, for up to 3 split

DAC curves.

To turn off the DAC feature, select the

MODE

parameter from the

DAC

menu and press the

button until

OFF

is indicated.

Page 52

Masterscan 350 User’s Guide ~Flaw Testing

5.4 TCG Operation

The Time Corrected Gain feature provides a means of creating an attenuation curve that sets the gain in a swept manner as a function of transit time. A properly constructed TCG curve will compensate for material attenuation and sound beam characteristics so that discontinuities of equal relative size at different depths in the test piece will give an equal amplitude response on the display.

Proper use of the TCG feature requires a reference block made from the same material as that being examined, with flat-bottom or side-drilled holes of the desired size placed at depths covering the range to be inspected.

To establish a TCG curve, follow the following instructions:

2. Establish the basic calibration of the Masterscan 350 using the desired transducer and the proper reference block.

3. From the

MEAS

menu, select the

TCG

feature.

4. Select the

MODE

parameter by pressing the plain grey button next to it and use the button to select

DRAW

. The other parameters will now be

CURSOR

and

POINT

.

5. The TCG system is now ready to accept reference points. Position the transducer to give a maximum echo from the first or topmost reference hole in the reference block.

Be careful to use a consistent amount of coupling and transducer pressure.

6. Use the and buttons to position the bright cursor bar over the echo from the reference hole.

7. Press the lh button to accept the point. The

POINT

parameter will then increment to indicate the next point. As each point is entered, the gain will change to place the echoes at about 80% amplitude as Shown below:

Page 53

Masterscan 350 User’s Guide ~ Flaw Testing

8. Repeat steps 5 and 6 for each additional reference hole in the reference block, being careful to be consistent with the amount of coupling and transducer pressure. A maximum of 10 points can be recorded for a TCG curve.

9. Press the plain grey button adjacent to

MODE

and use the button to select

ON

mode. The TCG curve will now be displayed, the amplifier gain will now swept, and the other parameters will change to

CURVE

and

MEAS

.

10. Use the

CURVE

parameter and the button to select ON or OFF for the display of the TCG curve.

NOTE: A TCG curve cannot be drawn while in RF display mode. It should be noted that the range of TCG is limited only by the dynamic range of the set.

11. Use the MEAS parameter to display the echo height relative to Full Screen Height

(FSH), percentage of TCG Reference (%REF) or dB relative to TCG Reference.

Page 54

Masterscan 350 User’s Guide ~Flaw Testing

5.5 Weld Inspection Using Trigonometry Mode

The trigonometry mode provides a convenient method for measuring the location of discontinuities when inspecting welds with angle beam transducers. Essentially, the trigonometry mode uses the thickness gauging features of the Masterscan 350 to calculate the surface distance and depth to a discontinuity from the actual beam path measured. This is accomplished by calibrating the trig mode with the actual refracted angle of the transducer being used. Additionally, the thickness of the test piece must be established in the menu in order to calculate depth and surface distance when using multiple skips in angle beam testing. The measurements being calculated are depicted in the following diagram:

These measurements are displayed below the A-Trace on the Masterscan 350 as follows:

Beam path distance B from the central exit point on the transducer to the discontinuity, the sum of B1 and B2 in double skip mode.

Surface distance S from the central exit point on the transducer to a point directly above the discontinuity on the test piece. By entering the distance from the Front of the Probe to the emission point in the PROBE menu, this distance may be displayed relative to the front of the probe, which is usually more convenient.

Depth distance D from the surface on which the transducer rests to the discontinuity.

Depending on the measurement technique being used there is an additional value displayed on the right of the measurement bar as shown below:

The example shows above the normal height value as a percentage of full screen; in this case 85%. If DAC is being used then the user has the choice of displaying either dB, %FHS or %DAC. If DGS/AVG is being used then ERS is displayed in the right hand side.

It is important to note that discontinuities in welds can be discontinuous and can extend over a distance. The measurements made will, therefore, be close approximations to the

Page 55

Masterscan 350 User’s Guide ~ Flaw Testing discontinuity if the user properly calibrates the Masterscan 350 and uses care to locate the desired peak amplitude signal to which a measurement is made.

To calibrate and use the trigonometry mode, follow the following instructions:

1. Using a suitable IIW or other calibration block, measure and establish the central sound beam exit point and actual refracted angle for the transducer being used.

2. Establish the zero and velocity (span) for the transducer and material under test using the A-CAL procedure described in A-Cal on page 82 . Use a suitable calibration block that has two drilled holes representing the thin to thick range of thickness expected in the test piece. In the

MEAS

menu set the

TRIGGER

parameter to

PEAK

. As in any ultrasonic thickness measuring method, it is important that the calibration be performed using a calibration block of the same material as that being inspected.

Alternately, if you know the velocity of sound in the test piece, you may calibrate using the A-CAL method on a different material and then set the velocity parameter.

3. Set the

START

and

WIDTH

of Gate1 to encompass the desired testing region.

4. From the Main menu, select

MEAS

and place the

MODE

to

TRIG

.

5. Set the

TRIGGER

parameter to

PEAK

.

6. Select the

ANGLE

parameter and using the and buttons, set the refracted angle of the transducer measured in step 1 above.

7. Select the

THICK

parameter and using the and buttons dial in the thickness of the test piece. You may have to readjust this parameter if the test piece had varying thickness and you are using a multiple skip mode of angle beam testing.

8. When measuring a discontinuity, position the transducer so as to obtain the maximum amplitude signal before recording the readings. A typical TRIG mode display is shown below.

Page 56

Masterscan 350 User’s Guide ~Flaw Testing

5.6 Weld Inspection Using the AWS Menu

The AWS menu provides a means for evaluating discontinuities when inspecting welds in accordance with the American Welding Society’s Structural Welding Code, ANSI/AWS D1.1-

94. The user is referred to the AWS standard for full details of the method.

The AWS menu provides a convenient method of automatically calculating the “Indication

Rating” (IR) as defined by the standard. The AWS menu can be used in conjunction with the trigonometry mode which will simultaneously indicate beam path, surface distance and depth distance at the bottom of the graticule. The AWS menu will not operate with either the DAC or TCG switched on.

To set up the AWS measurements, perform the following:

1. Calibrate the Masterscan 350 for weld testing and set up the trigonometry mode by following the steps in Weld Inspection Using Trigonometry Mode on page 55 .

2. From the

MEAS

menu, select

AWS

and place the

MODE

to

SET

.

3. Highlight REF on the menu and set to the desired reference level, usually 80% of full screen height.

4. Place the probe on the test block and obtain the maximum signal from the reference indication.

5. Place the cursor over the indication by selecting

CURSOR ADJUST

and using the and arrow keys.

6. With the signal maximized, press the lh

key.

The Masterscan 350 is now calibrated to make Indication Rating measurements in accordance with AWS D1.1-94. To use the AWS mode, do the following:

1. On the

AWS

menu, highlight

MODE

and select

MEAS

.

2. Ensure that the gate and measurement (

TRIG

) function are operative and properly adjusted.

3. When the AWS menu is selected and a signal is in the gate, the sub-menu boxes will show the resulting AWS measurements as follows:

IR

IL

RG

AF

(Indication Rating = Indication level – Reference Gain – Attenuation Factor)

Where:

IL: Indication Level The dB setting required to bring an indication to the reference level.

Page 57

Masterscan 350 User’s Guide ~ Flaw Testing

RG: Reference Gain The dB setting of the calibrated reference indication as a function of the reference standard and probe being employed.

AF: Attenuation Factor The attenuation factor required by the AWS standard and is:

Depth in inches minus 1, multiply by 2 and then rounded to the nearest ½ dB.

IR: Indication Rating The difference in dB between the indication and the reference gain with attenuation factor correction.

It is not necessary to bring the indication to the reference level to obtain the correct measurement information as the Masterscan 350 adjusts for gain offset. However, to ensure the best accuracy of calculation, it is advised to adjust the indication to be above 40% and below 100% of full screen height. In addition, in the AWS mode, gain offset is shown as +/- dB from reference as opposed to absolute gain.

Page 58

Masterscan 350 User’s Guide ~Flaw Testing

5.7 Evaluation of Pipe Imperfections using API 5UE

The Masterscan 350 is able to assist the user in making measurements in accordance with the American Petroleum Institute (API) Recommended Practice for Ultrasonic Evaluation of

Pipe Imperfections using the Amplitude Distance Differential Method (ADDM). Users are referred to the publication: “API Recommended Practice 5UE, Second Edition, June 2005”

The ADDT employs a combination of amplitude comparison and 6dB drop techniques to determine the radial depth of an imperfection. The technique is based on the premise that radial depth of an imperfection affects both the amplitude of the received echo signal and the differential time of flight of the transmitted ultrasonic wave as it passes over the imperfection.

The first step is to perform a standardisation using a reference indicator such as a notch of a known depth or a through drilled hole. It is assumed that the Masterscan had been setup with the correct values for the transducer and specimen being used for the inspection.

First turn on

GATE1

from the

CAL

menu and position it on the display so that it will cover the signal that will be produced by the reference indicator; the height of the gate is not important. Ensure that the signal will be between 5% and 100% of full screen height. Select

API

from the

MEAS

menu and select

PEAK

by pressing the button twice then capture the reference signal and the following display should be shown:

Once the desired signal is obtained press the lh

button and the k factor (a derived factor for calculating depth) is calculated and displayed as shown:

Page 59

Masterscan 350 User’s Guide ~ Flaw Testing

The Masterscan 350 automatically goes into

FREEZE

mode. Pressing the lh

button again puts the API facility into measurement mode as shown below. The k value is displayed but cannot be changed while in measurement mode.

To take measurements select

PEAK

mode by pressing the button twice and capture the echo dynamic signal; once the desired signal has been captured press the OK button.

The di value is calculated and displayed as shown below:

Page 60

Masterscan 350 User’s Guide ~Flaw Testing

Once the measurement has been made, pressing the lh

button clears the current reading.

To take the next measurement select peak mode by pressing the button twice.

Page 61

Masterscan 350 User’s Guide ~ Flaw Testing

5.8 DGS/AVG Operation

The DGS (Distance Gain Size) method, often known by its German acronym AVG

(Amplituden Vergleichs Größe), allows a theoretical model to be used for assessment of ultrasonic echoes.

The DGS/AVG method allows flaw sizing and distance compensation without requiring the large range of reference standards for each test which would be required in order to establish a measured DAC curve, or to set up a TCG correction.

The relative response from ‘Equivalent reflectors’ of different sizes at different distances can be predicted mathematically from the probe effective diameter, frequency and near field length (a function of the first two). Probe sensitivity can be calibrated from a single measurement (usually the back wall echo)

DGS/AVG has several limitations:

It is a predicative method, so only works in ‘standard’ configurations where the mathematical model is valid

It must be applied with care with non-circular probe elements as we have to approximate an

‘effective diameter’

It does not work well on broadband probes, where the frequency is not well defined.

It should only be applied to single element probes, as dual element probes have an inherent focus.

Corrections may be required for several factors:

Internal time delays within the probe

Attenuation with distance, both within the material to be tested and the reference block, if significant over the distances involved.

Transfer correction, e.g. if the tested material has a significantly rougher surface than the reference block

When using an angle probe and calibrating on a curved reflector (e.g. V1 or V2 Block) a correction must be applied.

NOTE: For DGS/AVG to work accurately the probe zero must be set correctly.

5.8.1 Theory

For a flat, disk shaped crystal of given frequency and size, the general beam profile can be defined in terms of a ‘normalised’ standard curve, scaled by the Probe ‘effective diameter’ and near field length.

The near field length is defined by the equation:

Page 62

Masterscan 350 User’s Guide ~Flaw Testing

N

D

2

EFF

4

 

2

Where

D

EFF

is the probe effective diameter and is the wavelength (Velocity/Frequency)

The echo height from a small (much less than sound field width) reflector will be proportional to its size.

The echo height from a given reflector in the far field (distance from the probe greater than

2-3 Near Field Lengths (NFL)) will be inversely proportional to the square of its distance from the probe.

It is therefore possible to define a standard relationship, knowing the basic probe parameters, which only need be corrected for the sensitivity of the probe by measuring the response from a known reflector.

The picture shows a ‘general’ DGS/AVG diagram plotted on logarithmic scales, Relative gain in dB, Distance as a multiple of Near Field Length, The various curves represent the signal from an Equivalent Reflector Size (ERS) expressed as a multiple of the effective probe diameter.

Note that the signal from the back wall echo (‘infinite’ reflector) falls off with distance, rather than with the square of distance.

Press jbkr

then the plain grey button next to

MEAS

and select the

AVG

menu,

MODE

should be

OFF

Page 63

Masterscan 350 User’s Guide ~ Flaw Testing

5.8.2 DGS/AVG ~ Frequency, TBVEL and ERS

Set

FREQ

from the probe data sheet

TBVEL

– Reference (Calibration) Block Velocity, in meters per second

This parameter is used internally and will take the same range of values as the Velocity parameter. This parameter MUST be entered. The Wavelength, which is calculated from this velocity and the Probe frequency, is critical to the DGS/AVG calculations.

Desired test sensitivity (

ERS

) may also be set now. (This can be adjusted later)

5.8.3 DGS/AVG ~ NFL, DELAY and PED

Set

MODE

to

SET PRB

NFL –

Near Field Length in mm.

This value should be copied from the probed data sheet (It is recommended that figures from specific probe tests are used rather than those from generic data sheets as typical manufacturing variations are sufficient to affect the accuracy of the results)

Page 64

Masterscan 350 User’s Guide ~Flaw Testing

DELAY

– Probe Delay Velocity, in meters per second

This is the velocity of compressive sound waves in the material of the probe wedge. It is used mostly in angle probes but also in zero degree compression probes which have a delay line. This wedge material is typically Perspex, which has a compressive sound wave velocity of 2700 m/s.

It is used, in conjunction with the probe zero and the specimen velocity, to calculate the

Sound field equivalent length, this being the effective distance of the crystal to the test piece/wedge interface.

PED

– Probe Effective Diameter, in mm

This is the calculated probe effective diameter and is for reference only; the value cannot be changed.

5.8.4 DGS/AVG ~ T-Loss, REF and MAT

Change MODE to ‘SET ATT’.

This menu allows the operator to enter attenuation parameters relating to the materials of the Calibration Block and the item under test, these parameters being:

T-LOSS

– Transfer Loss, in dB.

This is the loss due to surface roughness, poor coupling etc between the probe and the object under test. Generally it can be assumed that the Reference block will have a clean surface , but this may not apply to the object under test, methods exist for measuring this, experienced based ‘rules of thumb’ are also commonly used. This parameter can be changed once DGS/AVG is switched on.

REF

– Reference Material Attenuation, in dB per meter

Attenuation of ultrasound in the Reference (Calibration) block material in dB per Meter

MAT

– Test Material attenuation, in dB per meter

Page 65

Masterscan 350 User’s Guide ~ Flaw Testing

Attenuation of Ultrasound in the material of the object under test in dB per Meter

The

REF

and

MAT

are used where the ultrasound attenuation in these materials is such that it would cause a disparity in signal heights between similar reflectors in these materials.

At long path lengths these factors become very important.

5.8.5 DGS/AVG ~ dVK, REF TYPE and REFSIZE

Change

MODE

to

SET REF dVK

(V

K

) – Curvature Correction Factor in dB

This value allows the use of a curved surface to provide a back wall echo for use as a reference signal, thus the same reference block can be used irrespective of probe angle.

The amplitude correction factor shows by how many dB the echo from the cylindrical surface is larger or smaller than the echo from a flat backwall, normal to the probe field axis at the same distance.

This value is subtracted from the gain to compensate for the focusing effect of the curved reference reflector. It should be entered as a positive or negative value as expressed in the probe data sheets.

Probe datasheets typically express thus value as V

K1

, for use with the K1/V1 calibration block, or V

K2

for use with the K2/V2 calibration block. These calibration blocks are specified in standards EN12223 / DIN54122 / BS2704

REF TYPE

– Reference Reflector Type

The user can select one of a number of different reference reflectors.

In this implementation, these have been derived from EN 583-2:2001 Annex B.

These reflector types are:

BWE – Back Wall Echo

Page 66

Masterscan 350 User’s Guide ~Flaw Testing

FBH – Flat Bottom Hole

SDH – Side Drilled Hole

Reference type SDH is converted internally to the equivalent DSR diameter according to the formulae in EN 583-2:2001 Annex B.

When using reference SDH reflectors it is essential to select the correct calibration (test) block velocity (

TBVEL

). This is because the calculations for the equivalent DSR use the wavelength of sound that is calculated from the Calibration Block velocity and the probe frequency.

REFSIZ

– Reference Reflector Size, diameter in mm

Allows the user to enter a diameter, in mm, for a reflector. The reflector type is set using the

REF TYPE

parameter above.

The set will automatically compensate for the varying ‘reflectivity’ of different reflector types as specified by the

REF TYPE

parameter. Thus a 3 mm side drilled hole should be entered as

REF TYPE: SDH

and

REFSIZ: 3.0

In this implementation, a diameter of 0 (zero) is used to represent a reflector of infinite size and is expressed as

BACKWALL

or

INFINITE

.

5.8.6 DGS/AVG Calibration

Set

MODE

to

SET SIG

Place probe on reference block and maximise echo, select

CUSROR ADJUST

and use up/down keys to centre the selection bar over the echo.

If necessary change the CURSOR width to 5 pixels instead of 25 pixels to accurately select an echo that is close to others.

Page 67

Masterscan 350 User’s Guide ~ Flaw Testing

Selecting the SPLIT parameter to ACTIVE causes the AVG curve to recalculate and any point where the curve drops below 20 %FSH the gain is boosted by 12dB at that point. Up to 3 split DGS/AVG curves are possible.

The calibration of the DGS/AVG system sets the Probes ‘absolute sensitivity’ based on the response from the reference block.

Changing any parameters relating to the reference echo will invalidate the calibration, which must then be repeated. The current version of the software does not enforce this – it is the responsibility of the operator.

Check that the echo is still maximised and Press lh

; the instrument will adjust gain as necessary to get a good echo reference and then perform the necessary sensitivity calculation; the curve will then be drawn.

DGS/AVG is now on. In this screen it is possible to adjust the curve ERS level (the instrument will redraw the curve, but a new calibration is not required) and select either the

Curve or Gate as the measurement reference. It is also possible to change the T-LOSS value when DGS/AVG is on.

Page 68

Masterscan 350 User’s Guide ~Flaw Testing

5.8.6.1 Adjustment

When the DGS/AVG has been calibrated it is possible to leave the DGS/AVG menus and adjust other settings of the instrument, for example range and gate settings.

Gain may be adjusted with understanding – Changing the Gain setting will remove the correlation with the DGS/AVG curve, so should only be done in accordance with a specification , for example to compensate for varying material conditions (better done with the T-LOSS function). If it is desired to change the gain to give a larger signal (for example at long range) the REF gain should be adjusted. This will change the curve by an equivalent amount and preserve the measurement accuracy.

Increasing REF Gain by 6dB moves curve and preserves measurement

Page 69

Masterscan 350 User’s Guide ~ Flaw Testing

Decreasing Gain by 6dB returns to original height, but loses measurement

The Reference gain may be selected by pressing the obcLd^fk

Button. (DGS/AVG is only appropriate for Single probe mode)

When the Display Range or REF Gain is changed the DGS/AVG curve will recalculate at each step. If the settings must be changed by more than one or two steps it may be more convenient to Switch DGS/AVG off, Change REF gain and range etc as needed, and then switch DGS/AVG on again. This will not invalidate the curve provided the following are not changed:

FREQ

,

NFL

,

dVK

,

DELAY

,

REF dB/m

,

TBVEL

,

REF TYPE

,

REFSIZ

, (new

REFERENCE

Entered), Probe Zero, Filter, Contour, TX Parameters (Width, Damping,

Voltage,) or Probe Angle.

The following should not be used in DGS/AVG Mode:

Detect Modes other than Full Wave, non-zero REJECT, Dual probe mode, Smooth,

A_CAL (other than for initial setup before using DGS/AVG), DAC, TCG, AWS,

5.8.7 Measurement

Having set up the DGS/AVG system, it is possible to use the measurement system to calculate an Equivalent Reflector Size for the flaw.

This is achieved by using Gate 1 to identify the flaw indication and the DEPTH measurement mode to calculate its equivalent size. This Equivalent Size is calculated as the Test Sensitivity

(ERS value) multiplied by the Ratio of the Signal height to the DGS/AVG curve height at that distance. It is displayed in place of the signal amplitude (%FSH) as a diameter, in mm.

It is also possible to ‘size’ the flaw in terms of dB with respect to the DGS/AVG curve. This is achieved by increasing, or decreasing, the set gain until the peak of the flaw signal lies on the DGS/AVG curve. The difference in Gain required can then be read directly from the displayed Gain Value.

Page 70

Masterscan 350 User’s Guide ~Flaw Testing

5.9 Backwall Echo Attenuator (BEA)

The backwall echo attenuator allows the gain to be independently adjusted over a defined area of the A-Scan. To enable the backwall attenuator, select the

UTIL

main menu followed by the

BEA

feature. Changing the

MODE

to

DRAW

displays a solid block on the right hand side as shown below:

Select

CURSOR ADJUST

and use the and buttons to select the area over which the attenuator will be applied as shown below:

Once the required area has been selected, change the

MODE

to

ON

. The attenuator is switched on and a line is drawn on the screen showing where the attenuator becomes active as shown below:

Page 71

Masterscan 350 User’s Guide ~ Flaw Testing

Once the backwall attenuator is

ON

the size of the attenuator can be changed by adjusting the

ATTEN

parameter. The maximum area over which the backwall attenuator can be active is 50% and the maximum attenuation is the same as the available gain up to a maximum of 40dB.

Page 72

Masterscan 350 User’s Guide ~Flaw Testing

5.10 A-LOG, A-Scan Storage

The A-LOG feature provides for storage, recall and printing of A-Scans with all of the instrument settings. Up to 800 A-Scans and their settings can be stored. This feature is useful to record an indication for later review or for printing a hardcopy. In addition, recalling an A-Scan and its settings will allow for easy follow-up inspection of the same indication on the test piece. Simply unfreeze the display after recalling and the Masterscan

350 is ready to repeat the same inspection. Be sure to use the same transducer that was used for the original recording.

5.10.1 To Store An A-Scan:

1. Establish the basic calibration of the Masterscan 350 using the desired transducer and the proper reference block.

2. From the

MEMORY

menu, select the

A-LOG

feature.

3. Select the

STORE

parameter by pressing the plain grey button next to it and use the

and buttons to select a store location. The

STATE

parameter shows whether the store location is

VALID

(used) or

EMPTY

. Be careful not to store over a valid location unless you desire to erase what is there and replace it with the current A-Scan and settings.

4. Select the

MODE

parameter by pressing the plain grey button next to it and use the button to select

STORE

.

5. Obtain a desired echo from a discontinuity or reference block.

Optionally press the button to freeze the display before storing it.

Page 73

Masterscan 350 User’s Guide ~ Flaw Testing

6. Press the lh= button to store the waveform and instrument settings; the display will show a dialogue box:

7. At the

CONFIRM

prompt, press the lh= button again to store the A-Scan or press the

button to continue without storing the A-Scan.

8. After pressing the lh= button a second time, the

NOTES

window opens automatically to allow the creation of any desired notes for the stored A-Scan and its settings. See the section Adding NOTES to PANEL and A-LOG sets on page 41 . When finished adding notes, press the plain grey button next to the STORE parameter to return A-LOG menu.

5.10.2 To Recall an A-Scan:

1. From the

MEMORY

menu, select the

A-LOG

feature.

2. Select the

STORE

parameter by pressing the plain grey button next to it and use the

and buttons to select the desired store location to be recalled. The

STATE

parameter shows whether the store location is

VALID

(used) or

EMPTY

.

3. Select the

MODE

parameter by pressing the plain grey button next to it and use the button to select

RECALL

.

4. Press the lh= button to recall the waveform and instrument settings. This may take a couple of seconds at which time the display will be in

FREEZE

mode and the number of the A-LOG is displayed at the bottom of the display. To return to normal testing mode with the recalled settings, press the button.

5. Alternately, you may review the notes pages stored with each A-LOG to determine which set you want to recall. Set the

MODE

to

LIST

and using the

STORE

parameter, move through each one to view the

NOTES

page. When the desired A-LOG is found,

Page 74

Masterscan 350 User’s Guide ~Flaw Testing press the lh= button to recall it. To remove the notes review window, set the

MODE

back to

STORE

,

RECALL

or

DELETE

.

5.10.3 To Delete a Stored A-Scan

Follow the steps above but set the

MODE

parameter to

DELETE

and then press the lh= button. At the

CONFIRM

prompt, press the lh= button again to delete the A-Scan or press the button to continue without deleting the A-Scan.

To edit notes in an A-LOG, press the plain grey button next to the

EDIT NOTES

parameter and then follow the instructions in the section on Adding NOTES to PANEL and

A-LOG sets on page 41 .

NOTE: When notes are displayed, the menu cursor buttons are disabled. Switch off list mode to change to another menu.

5.10.4 REF, Reference Waveform Comparisons

This feature allows the user to select a waveform from one stored in the A-LOG menu, and to display this reference waveform on the graticule along with the real time waveform. This makes easy comparisons possible between an expected signal response (the reference waveform) and that obtained from the test piece.

5.10.5 To Recall an A-Scan as a Reference Waveform:

1. In order for a reference waveform to be recalled, the required trace must first be stored into an A-LOG store.

2. From the

MEMORY

menu, select the

REF

feature.

3. Select the

STORE

parameter by pressing the plain grey button next to it and use the

and buttons to select a store location. The

STATE

parameter shows whether the store location is

VALID

(used) or

EMPTY

.

4. Select the

MODE

parameter by pressing the plain grey button next to it and use the or button to select

RECALL

. If set to

LIST

, the note associated with the reference waveform will be displayed.

5. Press the lh= button to display the reference waveform on the graticule. The reference waveform will be a different colour as compared to the real time waveform, depending on which display colour is chosen.

6. To remove the reference waveform, set the

MODE

to

OFF

and press the lh= button.

NOTE: When the reference is recalled, no setting parameters are changed. Therefore, if the

RANGE

or

DELAY

are changed, the reference signal will not match the real time waveform.

This means that the

RANGE

and

DELAY

used for storing the reference waveform in A-LOG must match that used for the comparison of real time signals to the reference waveform.

Page 75

Masterscan 350 User’s Guide ~ Flaw Testing

The reference waveform cannot be displayed in video colours 0 and 1 (All white and all yellow).

5.10.6 Contour, Peak Echo & Hold Dynamics

Contouring is a technique that enables the significant echoes to be enhanced by extending the trailing edge of the echoes thus making them appear wider than they actually are making the Masterscan 350 operate more like an older analogue flaw detector. Only the trailing edge is modified so that depth measurements using FLANK and PEAK are not effected. From the AMP menu in CAL, select CONTOUR as shown below:

Change the value until the required shape is achieved, note how the measurement values do not change:

During flaw testing it is sometimes difficult to determine the exact peak of a discontinuity echo due to its geometry, coupling and surface anomalies and part geometry. The

Masterscan 350 incorporates a Peak feature that, while activated, stores all of the signal excursions on the display using a fill technique.

Page 76

Masterscan 350 User’s Guide ~Flaw Testing

In addition to providing an effective method of capturing peak echo information, the Peak mode is useful for crack tip diffraction methods in angle beam weld inspection. In the case of crack tip diffraction, the extent of a crack type discontinuity can be inferred by the positional excursions of the signal as it reflects from both tips of a crack. Thus, as the transducer is moved past the indication, the signals will be “filled-in” on the display and the resultant width of the envelope or “echo dynamic pattern” will infer the crack length.

To use the echo dynamic pattern feature, position the transducer just off one side of the indication. Press the button twice to activate the Peak mode which is indicated by a box-displaying PEAK below the graticule. Move the transducer past the indication to fill-in the envelope pattern. In the case of angle beam testing, it is advisable to rotate the transducer slightly and move it from side to side while moving forward past the indication.

This will assure the capturing of all amplitude and distance information. An example of peak echo mode capture is shown below:

To return the display to normal mode, press the button twice more. This will erase the envelope pattern and return the instrument to the instantaneous mode of display.

While the display is showing the dynamic echo pattern, it is possible to record the A-Scan as indicated in A-LOG , A-Scan Storage on page 73 .

Hold mode is similar to freeze mode except that it stores the currently displayed waveform as a reference trace and leaves the live waveform active. This is a quick and useful to way to make comparisons between echoes.

Page 77

Masterscan 350 User’s Guide ~ Thickness Gauging

Hold mode can be activated by pressing the button three times. Pressing the button again will return the Masterscan 350 to its normal mode of operation.

Page 78

Masterscan 350 User’s Guide ~Thickness Gauging

6 Thickness Gauging

6.1 Basic Thickness Gauging

Accurate and reliable thickness gauging requires three important considerations: a) Proper transducer selection b) An accurate reference block of the same material as that being tested, and c) Proper calibration of the instrument.

There are three types of transducers used in ultrasonic thickness gauging, all employing the straight or longitudinal beam method. The single element contact transducer is used for general purpose thickness gauging where material thickness is generally expected to be above 2.5mm. The delay line contact transducer is used for measuring thin material down to 0.5mm where the surfaces are clean and parallel. The dual element contact transducer is used for moderately thin materials down to about 1.0mm where the surfaces may be irregular and not necessarily parallel. The dual transducer is most commonly used for corrosion thickness inspection where its ability to obtain echoes from a pitted surface is superior. Nevertheless, dual transducer inspection on corroded materials is nowhere near as accurate as single element and delay line inspection, due to the nature of the material under test. In addition, the dual element transducer usually incorporates a slight included angle on both the transmit and receive side, resulting in a small non-linearity of the measurements. For this reason, dual transducer thickness testing is usually calibrated and performed over a limited thickness range.

Regardless of the type of transducer used, it is important that it be specifically designed for thickness gauging. This means that it must have broadband or highly damped characteristics that provide a sharp leading edge. Otherwise, amplitude variations of the signals can cause half-wavelength errors due to smaller “build-up”

The next requirement for thickness gauging is a calibration or reference block. This block should be made of the same material as the components to be inspected. In other words, it should have the same sound velocity and attenuation characteristics. The calibration block should have parallel machined surfaces representing the thickness range to be inspected.

Although two thicknesses representing the minimum and maximum values are sufficient for calibration, it is recommended that the calibration block have four sections covering the range anticipated. This will allow for verification of the calibration.

It is possible to calibrate the Masterscan 350 using a generic test block and only one known sample thickness of the material under inspection. This is a less desirable method because there is no way of verifying the calibration. With this method, the calibration would first be performed on the generic test block, which would have three to four sections covering the expected thickness range. Essentially, calibrating on the generic block would set the

Page 79

Masterscan 350 User’s Guide ~ Thickness Gauging transducer zero point. Then the Masterscan 350 Velocity would be adjusted on the known sample of the test material until the thickness readout matched the known thickness. This established the sound velocity for the test material.

Finally, it is necessary to establish the proper calibration of the Masterscan 350 in order to assure accurate and reliable thickness testing. The essence of this calibration is to set the pulser and amplifier characteristics to provide sharp leading edges on the echoes. The following are the essential steps for basic thickness gauging:

1. Select the appropriate broadband transducer and calibration block that matches the material under test.

2. In the

CAL

menu, select the proper

RANGE

and

DELAY

so those echoes from the range of thickness can be viewed.

3. From the

AMP

menu, set the

FREQ

to the nominal value of the transducer and set the detection mode to

RF

.

4. Use the button to select

SINGLE

for delay line and single element transducers and

DOUBLE

for dual element transducers.

5. Couple the transducer to a mid-range thickness section on the calibration block and obtain an echo.

6. From the

TX

menu, adjust the TX

WIDTH

parameter so as to obtain the maximum amplitude from the echo while maintaining a symmetrical shape to the echo.

7. From the

TX

menu, adjust the ActiveEdge™ (

EDGE

) if using a single element transducer. Adjust the value to reduce the width of the TX pulse. This is not necessary for dual or delay line transducers.

8. Repeatedly adjust both the TX

WIDTH

and the

EDGE

parameters until the echo shape is optimized.

9. From the

AMP

menu, set the

DETECT

parameter to either

–VE HW

or +VE HW, depending on which half of the RF provides the best half-cycle as determined in step 7 above.

10. Adjust the gain to set the echo at about 80 %FSH and increase the

REJECT

parameter to remove any undesirable noise from the baseline.

11. From the

GATE1

menu, set the gate to

ON +VE

and adjust the

START

and

WIDTH

so that the gate encompasses the thickness range of interest. Place the transducer between the thin and thick area on the calibration block to verify proper coverage.

Adjust the

LEVEL

parameter to about 30 %FSH.

Page 80

Masterscan 350 User’s Guide ~Thickness Gauging

12. From the

MEAS

menu, set the

MODE

parameter to

DEPTH

and the

TRIGGER

parameter to

FLANK

.

13. If desired, set the

T-MIN

parameter to

ON

so that the thinnest reading obtained will be held in the display. Reset by using the or buttons.

14. If desired, set the

HUD

parameter to

ON

to display the thickness in a large window at the top right of the graticule.

15. To calibrate the Masterscan 350 for thickness readings, follow the steps outlined in

A-Cal on page 82 . Alternately, you can use an iterative process from the

CAL

menu of setting the

ZERO

on the thinnest calibration block sample and the

VELOCITY

on the thickest sample. Using this technique, you must repeat

ZERO

to

VELOCITY

several times until the two values are correct.

When using a delay line transducer, there are two modes of measurement: a) Interface to first return echo, and b) multiple echo mode which usually measures from the first to the second return echoes after the interface echo.

1. Follow steps 1-10 above.

2. From the

MEAS

menu, set the

MODE

to

E-E

for echo to echo measurement. You will notice that a second gate bar appears below gate 1 and that its start point is slightly to the right of gate 1. This is the Blocking gate explained below.

3. For measuring from interface to first echo, position the start of gate 1 before the interface echo. Using the

BLANK

parameter in the

MEAS

menu, position the start of the second gate just after the interface echo but before the thinnest expected first echo.

4. For measuring multiple echoes after the interface, position the start of gate1 just after the interface echo. Using the

BLANK

parameter in the

MEAS

menu, position the start of the second gate just after the first echo but before the thinnest expected second echo.

5. The

BLANK

parameter determines the blanking or starting position of the second gate relative to the starting position of gate1. This prevents noise and spurious echoes associated with the interface or first echo from terminating the measurement and producing a false reading.

For storage and printing of thickness readings, refer to T-LOG Thickness Storage on page 84 and T-Log Editing and Printing on page 87 .

Page 81

Masterscan 350 User’s Guide ~ Thickness Gauging

6.2 A-Cal

A-CAL is an automated calibration feature for thickness gauging applications or when measuring depth to an indication in flaw testing.

In order to achieve proper calibration for measuring time-of-flight in ultrasonic testing, two factors must be known; the velocity of sound in the material under test, and the offset of the transducer caused by wear face and phase shifts.

Conventional thickness calibration technique usually requires setting the zero (offset) and span (velocity) using an iterative technique. This means alternately placing the transducer on a thin sample to set the zero and a thick sample to set the span. This process is repeated several times until both readings are correct. The reason for this is that raising the span will also raise the zero.

The A-CAL feature automates this process so that only two readings are required, one on the thin sample and one on the thick sample. The Masterscan 350 then calculates the correct offset and span factors and sets the velocity and zero. Any time the test material is changed

(velocity) or the transducer is changed (zero); this procedure must be repeated. Using this procedure will result in quicker calibration of the instrument and more accurate measurements.

The A-CAL calibration procedure is as follows:

1. Select a test block of the same material as that being inspected, with reference thickness covering the minimum to maximum thickness expected in the test parts. The difference between the thick sample and the thin sample should be no less than five to one.

2. From the main menu, select

CAL

and adjust the

RANGE

and

DELAY

so that the thin and thick reference echoes show on the display.

3. From the

CAL

menu, select

GATE1

and place the

STATE

in

ON +VE

. Adjust the

START

and

WIDTH

of the gate to encompass the echoes from the thin and thick samples.

4. From the main menu, select

MEAS

and place the

MODE

to

DEPTH

.

5. From the

CAL

menu, select the

A-CAL

menu.

6. Select

DIST1

using the adjacent plain grey button. Adjust the value using the and

buttons to the known thickness of the thin sample.

7. Place the transducer on the thin reference sample and obtain an echo. Adjust the gate start if necessary to obtain distance readout of the echo.

8. Press the lh= button to log the echo.

Page 82

Masterscan 350 User’s Guide ~Thickness Gauging

9. Select

DIST2

using the adjacent plain grey button. Adjust the value using the and

buttons to the known thickness of the thick sample.

10. Place the transducer on the thick reference sample and obtain an echo. Adjust the gate width if necessary to obtain distance readout of the echo.

11. Press the lh= button to log the echo.

12. Select

ACCEPT CAL

using the adjacent plain grey button.

13. Press the lh= button. The transducer zero and velocity will now be adjusted.

The

DIST1

and

DIST2

values can be stored when the calibration information in

PANEL

and

A-LOG

functions are stored. This overcomes the need to set these values and gate positions every time a calibration is desired. Once established, all that is required is to obtain each echo and press lh

, and then press lh on

ACCEPT CAL

and the instrument is calibrated.

If a delay transducer is used set

DIST1

to

0.0

and use the interface echo as the thin reference echo. The instrument will then set the transducer zero to compensate for the delay line.

Once calibrated, it might be desirable to record the velocity and zero values from the CAL menu for future reference. Note that this is a convenient method for determining the velocity of sound in an unknown material. Once the Masterscan 350 is calibrated on a known material, all that needs to be known is one fixed thickness point and the velocity can be determined.

6.3 TCG for Reliable Gauging

As previously mentioned, reliable thickness gauging requires that the shape of the echoes be consistent with sharp leading edges. When measuring the thickness of material with higher than normal attenuation to ultrasonic energy or a wide thickness range, it is difficult to maintain proper characteristics of the echoes. One way to overcome this is to use the Time

Corrected Gain feature of the Masterscan 350. TCG can improve the reliability of thickness gauging by increasing the amplitude of echoes from thicker sections of the test material so as to maintain the correct detection threshold on the leading edge. The objective is to equalize the amplitude of echoes over the entire thickness range to be expected.

To set up the TCG feature, refer to TCG Operation on page 53 .

Page 83

Masterscan 350 User’s Guide ~ Thickness Gauging

6.4 T-LOG Thickness Storage (Numeric)

The T-LOG menu provides a convenient method to store thickness readings for record keeping and analysis. The maximum number of readings that can be stored is 8,000. When using

NUMERIC

mode selected from the

T-LOG

submenu on the

MEMORY

menu, each thickness reading is stored under a three level code. The top level is the

BLOCK

number which can be 1 to 14. This is followed by the

LOC

(location) number which can be set between 1 and 2,000. Last is the reading number (

NO

) itself which can also be between 1 and 2,000.

This hierarchy of coding for a thickness reading allows readings to be associated with physical characteristics of the component under inspection. Thus, it is up to the user to devise a scheme of assigning blocks, locations and readings to the inspection, keeping in mind the maximum combination of 8,000.

To store readings, follow these simple steps:

1. Set up the instrument for the thickness measuring mode as described in Basic

Thickness Gauging on page 79 .

2. From the

MEMORY

menu, select the

T-LOG

menu.

3. Select the

MODE

to

NUMERIC

.

Note: The T-LOG table that appears at the bottom of the screen will display the stored values in the previous and currently selected location. The currently measured thickness value will be shown in the measurement line and in the HUD if switched on.

4. Use the corresponding plain grey buttons to select

BLOCK

,

LOC

, and

NO

in turn and

or the value to obtain the desire location number for storing the thickness.

5. Obtain the desired thickness reading and press the lh button to store the reading.

Notice that the

NO

value automatically increments to the next reading number.

6. To review a reading, select the desired

BLOCK

,

LOC

and

NO

to identify the location.

The value in the table at the bottom of the screen shows the reading stored for that location. Readings can be printed using the

PRINT

menu in utilities, see PRINT Menu on page 35.

Please note that numeric mode and sequence mode (see next page) occupy the same space in memory. This means that values stored in numeric mode may affect data stored in sequence mode and vice-a-versa.

Page 84

Masterscan 350 User’s Guide ~Thickness Gauging

6.5 T-LOG Thickness Storage (Sequential & Download)

The Sequential mode of thickness storage allows readings, locations and notes to be transferred between a computer and the Masterscan 350.

Typically, the user would create a thickness location sequence on the computer and download it to the Masterscan 350, which can then be used to guide the operator when collecting thickness readings. The collected readings could then be uploaded to the computer for further analysis and storage, releasing the Masterscan 350 to collect readings from another component. When a component is re-inspected, the historical readings can be downloaded to the Masterscan 350 along with the location information, providing the operator with an instant check of thickness variation while collecting new readings. The new readings can then be downloaded back to the computer for analysis, storage and printing.

To use the Sequence/Download mode, follow these steps:

1. Using suitable software or text editor, generate a location sequence on the host computer. Each location is comprised of a location number (1 to 8,000) and a text string of 32 characters. The text strings of 32 characters are subsequently displayed on the Masterscan 350 as two lines of 16 characters each in a table at the bottom of the screen, when the Masterscan 350 is in the Sequence mode. Generally, the first 16 characters are used to identify the location so as to direct the operator to the proper point for a thickness reading. The second 16 characters are for general use and can be used for the historical readings or baseline readings if this is the first inspection of the component.

2. Turn on the sequence mode by selecting the

T-LOG

function from within the

MEMORY

menu using the arrow buttons, pressing the plain grey select button next to the

MODE

option, and using the button to select

SEQUENCE

mode. Once selected, the graticule will reduce in size and a table is drawn along the bottom of the screen. The location and thickness data will be shown in the table if any data has been downloaded to the Masterscan 350. Alternately, the Sequence mode can be turned on by a command from the host computer.

3. Set up the instrument for the thickness measuring mode as described in Basic

Thickness Gauging on page 79 .

4. From the

MEMORY

menu, select the

T-LOG

menu.

5. Obtain the desired thickness reading and press the lh button to log the reading.

Notice that the

LOC

value automatically increments to the next reading number. Any historical reading downloaded from the computer will show in the window on the screen for comparison purposes.

Page 85

Masterscan 350 User’s Guide ~ Thickness Gauging

6. To attach one of the four available notes with the reading, press the plain grey select button next to the

NOTES

option, and using the or buttons, select

OBSTRC

(obstruction),

PITTING

,

POOR S/C

(poor surface contact), or

NO BWE

(no backwall echo).

7. To review a reading, select the desired

LOC

number to identify the location. The value in the bottom line of the table shows the reading stored for that location.

8. Once all of the readings for a component have been taken, they can be uploaded to the computer for further analysis and storage. The location data transferred is the 32 character string associated with the location, the thickness value, the units of measure, and the notes. Locations with a blank note will contain the characters “OK.”

Page 86

Masterscan 350 User’s Guide ~Thickness Gauging

6.6 T-Log Editing and Printing

The T-LOG menu provides capability for the deletion of thickness reading stored with the T-

LOG feature. With the T-LOG mode set to

DELETE

, it is possible delete a single location, all locations within a block, or all blocks (all readings).

6.6.1 To delete a single location:

1. Select the

T-LOG

menu from the

MEMORY

menu.

2. Select the

MODE

parameter by pressing the plain grey button next to it and pressing the

button to select

DELETE

mode.

3. Select the desired block number by pressing the plain grey button next to the

BLOCK

parameter using the or buttons.

4. Select the desired location number by pressing the plain grey button next to the

LOC

parameter using the or buttons.

5. Press the lh button to delete the thickness reading for the selected location.

6.6.2 To delete all Locations in a single Block:

1. Select the

T-LOG

menu from the

MEMORY

menu.

2. Select the

MODE

parameter by pressing the plain grey button next to it and pressing the

button to select

DELETE

mode.

3. Select the desired block number by pressing the plain grey button next to the

BLOCK

parameter using the or buttons.

4. Set the Location number to

ALL

by pressing the plain grey button next to the

LOC

parameter and using the button if not already selected.

5. Press the lh button to delete the thickness reading for the selected block.

6.6.3 To delete all Blocks and all Locations:

1. Select the

T-LOG

menu from the

MEMORY

menu.

2. Select the

MODE

parameter by pressing the plain grey button next to it and pressing the

button to select

DELETE

mode.

Page 87

Masterscan 350 User’s Guide ~ Thickness Gauging

3. Set the Block number to

ALL

by pressing the plain grey button next to the

BLOCK

parameter and using the button if not already selected. The location number will be automatically set to

ALL

.

4. Press the lh button to delete the thickness reading for all of the blocks and locations.

6.6.4 To print thickness logs

It is possible to print thickness logs by selecting the

PRINT

menu from the

UTIL

menu.

1. Select the

PRINT

menu from the

UTIL

menu.

2. Ensure printer is connected, switched on and contains paper.

3. Select the

MODE

parameter by pressing the plain grey button next to it and pressing the

button to select

T-LOG

mode to print normal thickness logs or

SEQUENCE

to print sequence mode data.

4. Set the block and location to the required values. Note it is possible to set the block and location numbers to

ALL

this will print all blocks and locations – all stored readings.

5. Press the lh button to print the thickness readings selected by the block and location numbers.

Page 88

Masterscan 350 User’s Guide ~ Power Supply

7 Power Supply

The Masterscan 350 is powered by a Lithium-Ion battery pack, which is shown below. The battery pack attaches to the rear of the Masterscan 350 with two thumbscrews and connects to the instrument via two terminal contact points.

7.1 Lithium-Ion Battery Pack

7.1.1 Battery Cautions

Misuse of Lithium-ion batteries may result in the following hazards: smoke, fire, or battery rupture. Misuse can also cause other battery damage or degradation of battery performance

DANGER! Use and charge only the specified Sonatest battery pack with the Masterscan

350. Only the Sonatest battery pack is tested and approved for use with the Masterscan 350.

Using third-party or counterfeit battery packs may cause smoke, fire, or cause the battery to burst.

DANGER! DO NOT hammer or otherwise impact the battery. Do not use the battery if it has been severely impacted or dropped, or if the battery has been subjected to heavy pressure. Battery damage may not be visible on the outside of the case. Even if the surface of the battery does not show cracks or any other damage, the cells inside the battery may rupture or catch fire.

DANGER! NEVER use or leave battery pack in areas with temperatures above +60°C.

High temperature build-up in the battery, such as could occur near fires or stoves, inside a sun-heated car, or by setting the battery in direct sunlight may cause the battery to rupture or catch fire. Excessive temperatures may also degrade battery performance or shorten battery life.

Page 89

Masterscan 350 User’s Guide ~ Power Supply

DANGER! DO NOT expose the battery to rain, snow, seawater, or any other liquids. Do not charge or use a wet battery. If the battery gets wet, be sure to wipe it dry before using.

The battery by itself is not waterproof.

DANGER! NEVER incinerate a used battery pack since internal battery gas may cause a rupture or explosion.

DANGER! NEVER solder the battery terminals, or modify the battery pack. This may cause heat generation, and the battery may rupture, emit smoke or catch fire.

DANGER! Use the battery only with the Masterscan 350 for which it is specified. Never use a battery with any other equipment, or for any purpose that is not specified in this instruction manual.

DANGER! If fluid from inside the battery gets in your eyes, blindness can result. Rinse your eyes with clean water, without rubbing them, and see a doctor immediately.

WARNING! Immediately stop using the battery if it emits an abnormal odour, heats up, or is discoloured or deformed. If any of these conditions occur, contact Sonatest.

WARNING! Immediately wash, using clean water, any part of the body that comes into contact with fluid from inside the battery.

WARNING! NEVER put the battery in a microwave oven, high-pressure container, or in an induction heating cooker. This could cause overheating, a fire, or cause the battery to rupture.

• CAUTION! Always use the battery within the specified temperature range for the

Masterscan 350 (-10°C to +55°C) and the battery itself (-20°C to +60°C). Using the battery out of its specified temperature range will reduce the battery's performance and battery life.

Please note that the specified temperature range of the battery may exceed that of the

Masterscan 350. In such cases, the Masterscan 350 may not work properly because it is out of its operating temperature range.

• CAUTION! Shorter battery life could occur if the battery is left fully charged, completely discharged, or in an excessive temperature environment (above +45°C) for an extended period of time. If the battery must be left unused for a long time, it must be detached from the Masterscan 350 after discharging. You may use the battery until the battery indicator shows half-capacity, and then keep it safely in a cool dry place with the temperature between -20°C to +25°C. If the battery pack is not going to be used for a long time it must be periodically charged every 4 months; failure to do so will render the battery inoperable and it will need to be returned to Sonatest for re-charging. If a number of spare battery packs are in use the easiest way to maintain optimum conditions is to rotate the battery packs so that when one is discharged it is swapped with a fully charged one.

Page 90

Masterscan 350 User’s Guide ~ Power Supply

7.2 Battery Charging

7.2.1 Charging Caution

DANGER! Only use the CH700-P charger to charge the battery. Using any other charger may cause smoke, fire, or cause the battery to burst.

DANGER! NEVER charge the battery pack in areas with extremely high temperatures, such as near fires or stoves, inside a sun-heated car, or in direct sunlight. In such environments, the safety/protection circuit in the battery will activate, causing the battery to stop charging.

WARNING! DO NOT charge or leave the battery in the battery charger beyond the specified time for charging. If the battery is not completely charged by the specified time, stop charging and remove the battery from the battery charger. Continuing to charge the battery beyond the specified time limit may cause a fire, overheating, or the battery may rupture.

WARNING! NEVER insert the battery charger into the Masterscan 350 or battery if it is wet or soiled. This could corrode the battery charger terminals or damage the charger. The charger is not waterproof.

• CAUTION! DO NOT charge the battery outside of the specified temperature range: O°C to

+40°C. Sonatest recommends charging the battery at +20°C. The battery may heat up or rupture if charged out of the specified temperature range. Additionally, battery performance or battery life may be reduced.

When fully charged, the battery pack should enable the unit to be operated for 16 hours when the brightness is at level 1 in the VIDEO menu under UTIL, for 12½ hours with a brightness of 10 (the default) or for 9 hours with a brightness of 20. The battery pack can be charged while mounted to the Masterscan 350 by use of the connector on the front panel. Alternately, the battery pack can be charged separate from the Masterscan 350 by using its own connector shown above allowing continued operation of the Masterscan 350 with the use of multiple battery packs.

When shipped the battery has only a nominal charge and must be fully charged before being used for the first time.

The Sonatest CH700-P battery charger shown below is suitable for use with main supply from 100 to 240 volts AC. The LED on the charger illuminates to indicate power on. When the battery level is low the LED will glow red. When the battery level is almost full the LED will glow yellow. When the battery level is full the LED glows green. The output voltage from the charger is 16.4 Volts DC. The output current when charging is a constant 2.0

Amps. The part number for a battery pack is Y-04.

Page 91

Masterscan 350 User’s Guide ~ Power Supply

WARNING – The CH700 is a battery charger only. Any attempt to use it to power a

Masterscan 350, without a battery fitted, WILL damage your CH700-P battery charger.

Page 92

Masterscan 350 User’s Guide ~ Interface Connections

8 Interface Connections

There is a 6-pin connector on the front of the Masterscan 350 as shown below:

This connector is used to connect the flaw detector to a PC using a modified USB cable; this allows information to be transferred to the computer using Sonatest’s program SDMS. For information on how to purchase SDMS please contact Sonatest using the details inside the front cover of this user’s guide. The USB connector also allows a USB keyboard to be connected and used for data entry in the notes fields. A printer can also be connected allowing information to be printed directly from the Masterscan 350. Please contact

Sonatest for a list for a list of supported printers.

At the rear of the Masterscan 350 there are various interface connections in a sealed compartment under a spring-loaded door. The connections are shown below and described in the following.

Page 93

Masterscan 350 User’s Guide ~ Interface Connections

8.1 RS232

The RS232 port on the Masterscan 350 allows for bi-directional communication between the instrument and a personal computer. Various control commands are available within three general categories; Write parameter values, Read parameter values, and Action commands.

The Write and Read parameters allow the computer to set a parameter and to read parameters current value. The Action commands provide special functions such as waveform and thickness logging transfer. The RS232 port is the D-shaped connector at the bottom.

The communication handshake is as follows:

Baud rate: 9,600, 19,200, 38,400, 57,600 or 115,200

Parity: None

Data Bits: 8

Stop Bits: 1

Handshake: Hardware

The cable connections for a 9-pin RS232 port are as follows:

The cable connections for a 25-pin RS232 port are as follows:

A 9-pin to 9-pin communications cable is available from Sonatest as part number 152120.

No cable is available for the 25-pin RS232 port.

Page 94

Masterscan 350 User’s Guide ~ Interface Connections

8.2 Composite Video

The composite video output is for connecting the Masterscan 350 to a compatible video monitor or projector. The output is either NTSC (USA/Japan) or PAL (UK/Europe), depending on the VIDEO selection in the UTIL menu. The screen update rate is 50Hz in

PAL and 60Hz in NTSC. The composite video connector is the second connector from the top with one centre pin.

8.3 Proportional Outputs

The Masterscan 350 has two proportional outputs which can be independently configured, both proportional outputs can be used to output the following readings: Depth 1 (gate 1),

Depth 2 (gate 2), E1-E2 (distance between gate 1 and gate 2 readings), Level 1 (gate 1),

Level 2 (gate 2), L1-L2 (difference between the amplitudes of signals through gate 1 and gate 2) and an Alarm signal (Alarm 1 (Gate 1 triggered) is available on proportional output 1 and Alarm 2 (Gate 2 triggered) is available on proportional output 2).

The choice of which reading to output is made using the

O/P 1

and

O/P 2

parameters in the

P_O/P

feature in the

UTIL

menu. The proportional outputs are updated at the currently selected PRF.

The voltage of the outputs is controlled by the

CAL

parameter in the

P_O/P

feature in the

UTIL

menu. This allows the voltage to be varied between 0V and 10.0V. The calibration voltage applies to all proportional outputs.

It is the small upper connector with 4 pins; the top 2 pins are used for the proportional outputs. The ground is achieved using the outer case of the connector.

8.4 Synchronisation Outputs

The Masterscan 350 provided two synchronisation outputs on the small 4 pin connector: S1 and S2.

S1 is synchronised with the TX pulse so that it goes high (3v3) for 50ns, at the start of the

TX signal to the analogue board.

S2 goes high (3.3V) for the duration of the acquisition period.

Page 95

Masterscan 350 User’s Guide ~ Features of the Masterscan 350

9 Features of the Masterscan 350

The Masterscan 350 is a user-friendly ultrasonic digital flaw detector and thickness gauge, which is simple to use and provides the experienced ultrasonic operator with a full-function device that incorporates many productivity enhancing features. All of the features of the

Masterscan 350 are accessed through a menu system using the front panel keypad.

9.1 Functional Testing Methods

The Masterscan 350 supports the following ultrasonic testing methods:

Pulse-Echo Flaw Detection

Transmit-Receive Flaw Detection

Time of flight measurement

Angle Beam (Shear Wave) Testing

Angle Beam (Surface Wave) Testing

Crack Diffraction Methods

Depth of Flaw Measurement

Single Transducer Thickness Measurement

Dual Transducer Thickness Measurement

Indirect Measurement of Sound Velocity in Materials

Time measurement in microseconds

Phase change display in un-rectified mode

Through transmission testing

Creep wave testing

Pitch and catch techniques

Page 96

Masterscan 350 User’s Guide ~ Specifications

10 Specifications

Feature Details

Test Range

Velocity

0 - 1mm (0.05in) up to 0 - 20000 mm (800 in.) at steel velocity.

Variable in 1,2,5 sequence or continuously in 1mm

256 to 16000m/s continuously variable.

Probe Zero

Delay

Gain

Test Modes

Pulser

ActiveEdge™

0 to 999.999 μs, continuously variable.

Calibrated delay from 0 -20000mm in 0.05 mm steps at steel velocity

(0-400in. in 0.002 in. steps).

0 to 110dB. Adjustable in 0.1, 0.5, 1, 2, 6, 10, 14 and 20dB steps.

Direct access to gain control at all times.

Pulse echo and transmit/receive.

100V - 350V square wave pulser. Pulse width from Spike to 2,000ns duration - rise/fall times <10ns into 50 ohms at 200V: Width adjustable in 2% of nominal width, minimum 1ns maximum 40ns.

Used to improve the pulse shape and improve near-surface resolution.

P.R.F Selectable 15 to 100 in 5Hz Steps; 100 to 1,000 in 50Hz steps and

1,000 to 5,000Hz in 100Hz steps.

Update Rate

Rectification

60Hz (NTSC Mode); 50Hz (PAL Mode).

Full wave, positive or negative halfwave and unrectified rf.

Frequency Range 6 narrow bands centred at 0.5 MHz, 1MHz, 2MHz, 2.25MHz, 5MHz,

10MHz and 15MHz. Broad band at 2 MHz to 22MHz (-6dB) and

1MHz to 35 MHz (-20dB).

System Linearity Vertical = 1% Full Screen Height (FSH). Amplifier Accuracy ±

0.1dB. Horizontal ±0.4% Full Screen Width (FSW).

Reject 80% linear reject. LED warning light when selected.

Units Metric (mm), inch (in) or time (μs).

Page 97

Masterscan 350 User’s Guide ~ Specifications

Feature Details

Display Colour

Transflective TFT

Display area 111.4 x 83.5 mm (4.39 x 3.29 in) 320 x 240 pixels. A-

Scan Area 255 x 200 pixels (315 x 200 expanded), 8 colour options and variable brightness.

Gate Monitor Two fully independent gates for echo monitoring and thickness measurement. Start and width adjustable over full range of unit, amplitude variable from 0 to 100% FSH. Bar presentation. Positive or negative triggering for each gate with audible and visual alarms.

Gate Expansion Expands range to width of Gate 1.

Selectable 0.6 seconds delay on gate 2 negative monitor tracking. Gate Monitor

Delay

Measurement

Modes

Mode 1

Mode 2

Mode 3

Mode 4

Mode 5

Signal Monitor

Depth and amplitude of first signal in gate.

Echo-to-Echo distance measurement. (single gate)

Trigonometric display of beam path, surface distance and depth of indication, curve surface correction and X-OFFSET for probe index.

Half skip indication on screen.

Gate to Gate distance measurement (independent gates).

Mode 6

Resolution

T-Min mode for holding minimum thickness reading. to 0.01mm (0.001in) for distance measurement or 1% FSH for amplitude measurement. Large display of measurement at top of A-

Scan display. Measurement mode selectable between peak and flank.

A-Scan Memory 800 waveforms can be printed or transferred to a PC using optional

SDMS software.

Panel Memory 100 stores for retaining calibrations.

Thickness

Logging

8,000 thickness readings configured either by Block / Location /

Number mode or pre-programmable work sheets in sequential

Page 98

Masterscan 350 User’s Guide ~ Specifications

Feature Details

mode. Readings can be exported to MS Excel using optional software.

AGC

DAC

Automatic Gain Control automatically sets the signal to a level between 10-90% FSH with tolerance between 5-20% accuracy.

DAC defined by up to 10 points and digitally drawn on screen. DAC curves meet requirements of EN 1714, JIS and ASME standards, selectable between -2, -6, -10, -12 and -14dB. Amplitude read out selectable between % DAC or relative dB.

TCG

Backwall Echo

Attenuation

Time Corrected Gain, also known as Swept Gain. 40dB dynamic range greater than 30dB per microsecond and up to 10 points may be used, setting all signals initially to 80% FSH.

0- 40dB attenuation.

AWS Built in software for evaluation of defect indications in accordance with AWS D1.1 structural weld code.

API

Auto-Cal

Flaw sizing complying with API 5UE

Provides automatic calibration from two echoes.

Reference

Waveform

This menu displays a waveform from one of the A-log stores as a reference or fingerprint display in a colour different from the active display highlighting differences from the reference.

Notes Alphanumeric labelling for panel and A-log allows the user to enter

Notes for storage with panel settings and A-scans.

Display Freeze For capturing the current A-scan image.

Peak Memory For echo-dynamic pattern determination.

Page 99

Masterscan 350 User’s Guide ~ Specifications

Feature Details

Help Key

Language

Support

Waveform

Smoothing

For instant operator guidance on using the Masterscan Series.

Supports 6 simultaneous languages, user selectable. English,

German, Spanish, French, Dutch. Others available on request.

Gives a smooth signal envelope, simulating analogue equipment. bi-directional readings and waveform memories. Composite video, PAL or NTSC compatibility. Analogue proportional outputs programmable to distance or amplitude of signal in the gate. Transmitter sync output

Front USB

Printers

For connection to printers, keyboards and PC.

Supports any printer with PCL support including HP Deskjet and

Epson.

Power Lithium Ion battery pack 14.4V, 5.0 ampere hours, gives up to 16 hours duration from a fully charged pack. Indication of low battery status. Recharge time 3-4 hrs.

Page 100

Masterscan 350 User’s Guide ~ Warranty

11 Warranty

Warranty/Defects after Delivery: Immediately upon receipt of the goods the buyer is required to check the goods carefully and thoroughly. In order to benefit from the guarantee, any defect in the product should be immediately reported in writing to Sonatest.

Sonatest will make good by repair or by the supply of a replacement or by equivalent adjustment of the price at our sole option, any defects which under proper use appear in the goods within the period of twenty-four (24) calendar months after the goods have been delivered and which arise solely from faulty design, material or workmanship, provided that the goods are carefully packed and promptly returned by you, free of charge, to the Sonatest works unless otherwise arranged. Said goods should be covered while in transit to us and must be accompanied by a written statement detailing the precise nature of the fault and the operating conditions under which the fault occurred. The repaired goods will be returned by Sonatest free of charge.

Save as in this Clause herein before expressed Sonatest shall not be under any liability in respect of defects in goods delivered or for any injury damage or loss resulting from such defects and our liability under this Clause shall be in lieu of any warranty or condition implied by law as to the quality, fitness or merchantability for any particular purpose of such goods.

Page 101

Masterscan 350 User’s Guide ~ Index

12 Index

A 

A-CAL Menu · 23

ActiveEdge · 21, 47, 81, 98

ActiveEdge™ · 21, 47, 81, 98

Adding NOTES to PANEL and A-LOG sets

· 38, 39, 42, 75, 76

Adjustment · 69

AGC · 19, 35, 100

AGC (Automatic Gain Control) · 35

A-Log · 19, 36, 39, 40, 42, 43, 44, 45, 50,

74, 75, 76, 78, 84

A-Scan Storage · 39, 50, 74, 78

Memory Menu · 39

American Welding Society · 31, 58

Amplifier · 10, 20, 30, 47, 55, 81, 98

Amplifier Menu · 20 angle · 7, 25, 26, 46, 50, 56, 57, 63, 66,

67, 78

Angle · 71, 97

API · 19, 24, 32, 60, 61, 100

Menu · 32

A-Scan · 1, 7, 9, 16, 34, 36, 38, 39, 40, 50,

72, 74, 75, 76, 78, 99

Attenuation · 27, 28, 31, 33, 34, 46, 51,

54, 58, 59, 63, 66, 67, 73, 80, 84, 100

AVG · 19, 24, 28, 63, 64, 68, 69, 71, 100

AWS · 19, 24, 31, 58, 59, 71, 100

AWS Menu · 31

B 

Backwall · 34, 48, 67, 72, 73, 87, 100

Backwall Echo Attenuation · 100

Basic Flaw Testing · 46

Basic Thickness Gauging · 80, 85, 86

Page 102

Battery · 1, 10, 16, 33, 90, 91, 92, 93, 101

Battery Charging · 92

Baud Rate · 34, 95

BEA(Backwall Echo Attenuator) · 19, 33,

34, 72

Beam · 3, 7, 25, 46, 50, 51, 54, 56, 57, 58,

63, 78, 80, 97, 99

Beam Path Distance · 25, 56

Blank · 41, 44, 45, 87

Block · 3, 12, 14, 15, 23, 28, 40, 41, 46,

47, 48, 57, 63, 65, 66, 67, 68, 72, 80,

81, 82, 88, 89, 99

BNC · 10

Broad Band · 98

C 

CAL · 6, 12, 14, 19, 20, 23, 32, 37, 47, 49,

57, 60, 71, 77, 81, 82, 83, 84, 96

Menu · 20

Calibration · 1, 3, 4, 8, 12, 14, 15, 16, 19,

20, 23, 29, 32, 36, 38, 39, 42, 43, 44,

45, 46, 47, 48, 49, 51, 54, 57, 65, 66,

67, 68, 69, 74, 80, 81, 82, 83, 84, 96,

100

Calibration Menu · 20

Calibration of the Sound Velocity · 3

Centre Frequency · 20

Certification · 3

Charge · 90, 91, 92, 102

Charger · 10, 92, 93

Clearing the Memory · 16

Click · 33

Clock · 19, 37, 100

CLOCK Menu · 37

Composite Video · 33, 96, 101

Concave · 26

Connector · 10, 92, 94, 95, 96

Contact · 5, 41, 46, 47, 80, 87, 90, 91, 94

Contents · ii

Contour · 20, 71, 77

Controls · 1, 6, 17, 33, 37

Convex · 26

Corrosion · 80

Couplant · 4

Coupling · 4, 28, 29, 52, 54, 55, 66, 77

Crack Diffraction · 97

Critical Operating Factors · 3

CSC (Curved Surface Correction) · 19, 26

Cursor · 6, 27, 29, 30, 31, 34, 43, 44, 45,

52, 54, 58, 76

Curve · 27, 28, 29, 30, 50, 51, 52, 53, 54,

55, 63, 68, 69, 70, 71, 99

D 

DAC · 1, 19, 24, 27, 28, 51, 52, 53, 56, 58,

63, 71, 100

Damping · 71 dB · 10, 27, 28, 29, 30, 31, 50, 53, 55, 56,

58, 59, 64, 66, 67, 71, 100

Default · 16, 27, 35, 92

Delay · 6, 7, 20, 21, 28, 65, 66, 80, 81, 82,

84, 98, 99

Delete a panel set: · 42

Delete a Stored A-Scan · 76

Depth Distance · 25, 56, 58

Depth of Flaw · 97

Detailed Menu Description · 1, 17

Detect · 71

DGS · 1, 28, 29, 56, 63, 64, 65, 66, 67, 68,

69, 70, 71, 100

DGS/AVG · 1, 28, 29, 56, 63, 64, 65, 66,

67, 68, 69, 70, 71

Diffraction · 78

Disclaimer of Liability · 5

Disclaimers and Notices · 1, 2

Discontinuity · 3, 47, 56, 57, 74, 77, 78

Masterscan 350 User’s Guide ~ Index

Dual Element · 63, 80, 81

Dual Probe · 71

Dynamic · 50, 51, 55, 78, 100

E 

Echo Dynamic · 7, 61, 78

Edit Notes · 44, 76

Summary · 44

Effects of Temperature on Calibration · 4

Electromagnetic Compatibility · 5

Envelope · 7, 33, 78, 101

Evaluation of Pipe Imperfections using

API 5UE · 32, 60

F 

Factory Default · 16

Far Field · 64

Far Surface · 46

Flank · 99

Flaw Detection · 3, 12, 97

Flaw Detection Calibration · 3

Flaw Detector · 2, 3, 77, 94, 97

Flaw Testing · 1, 13, 20, 46, 47

Flaw Testing Calibration · 20, 47

Focus · 63

Freeze · 7, 24, 34, 74, 78, 100

FREEZE/PEAK · 9, 10

Frequency · 64, 65, 98

Front Panel Controls · 6, 17

Front Surface · 4, 46, 47 full screen · 9, 20, 22, 33, 35, 56, 58, 59,

60

Full Screen · 27, 50, 55, 98

Function · 1, 16, 24, 25, 26, 27, 38, 39,

42, 43, 44, 51, 54, 58, 59, 63, 70, 86,

97

Functional Testing Methods · 97

Page 103

Masterscan 350 User’s Guide ~ Index

G 

Gain · 9, 10, 30, 34, 35, 48, 53, 54, 55,

58, 59, 63, 64, 67, 68, 69, 70, 71, 72,

73, 81, 98, 100

Gate · 22, 23, 24, 25, 27, 32, 33, 35, 37,

47, 48, 49, 50, 51, 52, 58, 60, 69, 71,

81, 82, 83, 84, 96, 99, 101

Gate 1 · 12, 14, 19, 22, 48, 49, 50, 60, 81,

83

Gate 2 · 12, 19, 22, 48, 49

Gates · 22, 24, 25, 99

General Warnings · 2

Ghosting · 21

H 

Handshake · 95

Hardcopy · 74

Help · 8, 42, 101

HUD · 12, 14, 24, 82, 85

I 

Included Angle · 80

Interface Connections · 1, 94

L 

LEMO · 10

Level · 20, 22, 24, 25, 27, 31, 33, 34, 35,

40, 58, 59, 69, 85, 92, 96, 100

Lithium-Ion Battery Pack · 90

Loc · 40, 41, 85, 86, 87, 88

Location · 38, 39, 40, 41, 42, 43, 56, 74,

75, 76, 85, 86, 87, 88, 89, 99

Longitudinal · 80

Page 104

M 

Main Menu · 19

Main Menu Selection · 19

MEAS · 6, 12, 14, 19, 24, 26, 27, 30, 31,

32, 35, 48, 49, 50, 51, 54, 55, 57, 58,

60, 64, 82, 83

MEAS Menu · 24

Measurement · 1, 19, 22, 24, 26, 27, 29,

31, 32, 33, 35, 48, 49, 52, 56, 57, 58,

59, 61, 62, 63, 69, 70, 71, 77, 82, 85,

97, 99

Measurement Menu · 19, 24

Memory · 1, 6, 7, 16, 19, 38, 39, 40, 42,

74, 75, 76, 85, 86, 88, 99, 100

Menu Tree · 17

Metric · 12, 14, 16, 98

Microseconds · 20, 97

MISC · 19, 34

MISC Menu · 34

Mode · 2, 7, 8, 9, 12, 14, 20, 22, 24, 25,

26, 27, 28, 29, 30, 31, 33, 34, 35, 36,

39, 40, 41, 42, 47, 50, 51, 52, 53, 55,

56, 57, 58, 59, 61, 62, 71, 75, 76, 78,

79, 81, 82, 85, 86, 88, 89, 97, 98, 99

Mode Conversion · 2

Multiple Echo · 82

N 

Near Field · 3, 28, 63, 64, 65

Notes · 36, 38, 39, 40, 41, 42, 43, 44, 45,

75, 76, 86, 87, 94, 100

Notes Window · 44

NTSC · 33, 96, 98, 101

O 

O/P · 19, 37, 96

Operator Training · 2, 3

Orientation · 3, 46

Outputs · 37, 96, 101

P 

P_O/P Proportional Output Control

Menu · 37

PAL · 16, 33, 96, 98, 101

Panel · 1, 6, 16, 17, 19, 38, 39, 40, 42, 43,

44, 45, 84, 92, 97, 99, 100

Parameter · 6, 7, 9, 12, 14, 19, 21, 35, 42,

47, 48, 50, 51, 52, 53, 54, 55, 57, 65,

66, 68, 73, 74, 75, 76, 81, 82, 88, 89,

95, 96

Parity · 95

Pipe · 32, 60

Pitting · 41, 87

Planar · 46

Point · 8, 25, 26, 27, 30, 52, 53, 54, 56,

57, 68, 81, 82, 84, 86

Power Supply · 1, 10, 90

PRF · 12, 14, 21, 37, 96

Print · 7, 19, 33, 36, 42, 85, 89

PRINT Menu · 36, 85

Printing · 36, 50, 74, 82, 86, 88

Probe · 2, 26, 28, 56, 58, 59, 63, 64, 65,

66, 67, 68, 71, 98, 99, 100

PROBE Menu · 26

Probe Zero · 28, 63, 66, 71, 98

Proportional Outputs · 96

Pulse · 2, 21, 22, 81, 96, 97, 98

Pulse Repetition Frequency · 21

Pulse-Echo · 97

Q 

Qualification · 3

Quick Start · 1, 6

Masterscan 350 User’s Guide ~ Index

R 

Range · 4, 12, 14, 20, 21, 22, 35, 47, 50,

51, 54, 55, 57, 63, 65, 69, 70, 71, 76,

80, 81, 83, 84, 91, 92, 98, 99, 100

Recall · 7, 38, 39, 40, 42, 74, 75, 76

Recall a panel set: · 42

Recall an A-Scan · 75

Recall an A-Scan as a Reference

Waveform: · 76

Receiver · 10

REF · 19, 27, 28, 29, 30, 31, 34, 40, 50,

53, 55, 58, 66, 67, 68, 70, 71, 76

REF, Reference Waveform Comparisons ·

76

REF/GAIN · 19, 71

Reference · 1, 3, 4, 10, 23, 27, 28, 29, 30,

31, 32, 38, 40, 46, 48, 50, 51, 52, 53,

54, 55, 58, 59, 60, 63, 65, 66, 67, 68,

69, 71, 74, 76, 78, 80, 83, 84, 100

Reference Block · 3, 46, 51, 52, 54, 55,

63, 66, 67, 68, 69, 74, 80

Reject · 20, 71, 81, 98

Reset · 16, 24, 82

Reset to factory defaults: · 16

Resolution · 46, 47, 98, 99

RS232 · 95

S 

Sensitivity · 12, 29, 46, 47, 63, 64, 65, 69,

71

Shear · 97

Shear Wave · 97

Single Probe · 71

Size · 3, 7, 9, 28, 29, 32, 46, 51, 54, 63,

64, 68, 71, 73, 86

Skips · 25, 56

Socket · 10, 44

Sound Field · 28, 64, 66

Page 105

Masterscan 350 User’s Guide ~ Index

Sound Velocity · 3, 4, 23, 46, 80, 81, 97

Specific Warnings · 2

Specifications · 3, 98 split screen · 9, 34

Start · 16, 22, 23, 24, 32, 82, 83, 96, 99

State · 7, 22

Stop Bits · 95

Storage · 16, 38, 40, 42, 82

Storage & Recall of Calibration Setups ·

16, 38, 42

Store · 7, 16, 38, 39, 40, 42, 43, 44, 45,

74, 75, 76, 85

Store a panel set · 42

Store An A-Scan · 74

Surface · 3, 25, 26, 28, 29, 35, 41, 46, 47,

56, 58, 63, 66, 67, 77, 80, 87, 90, 97,

98, 99

Surface Wave · 46, 97

Synchronisation Outputs · 96

T 

TCG · 1, 19, 24, 28, 30, 50, 51, 54, 55,

58, 63, 71, 84, 100

Temperature · 4, 90, 91, 92

Temperature Variations · 4

Test Blocks · 4, 58, 80, 83

Testing Limitations · 3

Theory · 63

Thickness Gauge · 2, 97

Thickness Gauging · 1, 14, 15, 80

Threshold · 22, 24, 25, 27, 51, 84

Time Corrected Gain · 30, 51, 54, 84, 100

Time of Flight · 3, 60, 97

Time-of-Flight · 83

T-Log · 36, 40, 41, 82, 88

Menu Numeric Mode · 40

Menu Sequential Mode · 41

Thickness Storage (Numeric) · 85

Page 106

Thickness Storage (Sequential &

Download) · 41, 86

T-Min · 99

Training · 2

Transducer (Probe) Zero Procedure · 3

Transducer Condition · 4

Transmit-Receive · 97

Transmitter · 10, 21, 101

Trigger · 52

Trigonometric · 99

Twin · 10

TX · 19, 21, 47, 71, 81, 96

Typographical Conventions · 1

U 

Ultrasonic · 1, 2, 3, 4, 5, 17, 32, 57, 60,

63, 80, 83, 84, 97

Ultrasonic Testing · 1, 2, 3, 5, 17, 83, 97

Unrectified · 98

USB · 6, 10, 36, 44, 94, 101

Use of Couplant · 4

Using the Front Panel Keypad · 44

Using the Optional Keyboard · 44

UTIL · 6, 7, 9, 12, 14, 19, 20, 33, 72, 89,

92, 96

Utilities · 19, 85

V 

Valid · 36, 63, 74

VEL · 12, 14, 20

Velocity · 2, 3, 12, 14, 20, 28, 33, 48, 57,

64, 65, 66, 68, 81, 83, 84, 98

Velocity of Sound · 2, 20, 48, 57, 83, 84

Video · 16, 19, 33, 77, 92, 96

VIDEO Menu · 33

W 

Warranty · 102

Wave · 2, 20, 21, 60, 66, 71, 97, 98

Wavelength · 21, 46, 64, 65, 68, 80

Wedge · 66

Weld · 25, 31, 46, 50, 56, 58, 78, 100

Weld Inspection · 25, 31, 46, 50, 56, 58,

78

Weld Inspection Using the AWS Menu ·

31, 58

Masterscan 350 User’s Guide ~ Index

Weld Inspection Using Trigonometry

Mode · 46, 50, 56

Width · 2, 20, 21, 22, 25, 27, 29, 30, 34,

47, 64, 68, 71, 78, 81, 84, 98, 99

Z 

Zero · 3, 6, 20, 23, 57, 66, 68, 71, 81, 83,

84

and Optional Keyboard Controls & Recall of Calibration Setups and Optional Keyboard

Controls Calibration , Peak Echo & Hold Dynamics Operation/AVG Operation Operation

Inspection Using the AWS Menu & Recall of Calibration Setups, A-Scan Storage Inspection

Using Trigonometry Mode, A-Scan Storage Operation Inspection Using Trigonometry Mode,

A-Scan Storage Editing and Printing Operation Menu

Page 107

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement